Introduction

During the last decades, the surgical management of kidney stones benefited of many technological advances and one of them is the development of flexible ureteroscopy (fURS).

This tool, ancillary equipment such as graspers and baskets, and lithotripsy technique with Holmium:YAG laser underwent many improvements leading to a widening of its indications with diagnostic and therapeutic management of upper urinary tract pathologies such as urolithiasis and urothelial tumors.

Firstly described in 1964 by Marshall, the uretroscope was only passively deflectable and did not include working channel. Although the first successful procedure in humans with a ureteroscope integrating active deflection has been reported by Takayasu, it was not until 1987 that Demetrius Bagley introduced flexible ureteroscopy as we know it today [1,2,3,4,5]. At this time, the fURS suffered from many weaknesses: low angle of deflection, poor resolution, difficult ureteral insertion and lack of irrigation [3]. But fURS benefited of many improvements between 1980 and 1990: fiberoptics for image relay and light transmission, miniaturization of the outer diameter, working channel, and passive and active deflections in both upward and downward directions. After 2000, another major development was introduced: the digital imaging system, improving the image quality substantially. Currently, we can distinguish two types of fURS: fiberoptic and digital fURS. The difference between them is the image relay and light transmission. In fiberoptic fURS, light and image are transmitted in analog format through fiberoptics bundles whereas illumination in digital scopes is made by fiberoptics or by a diode (DEL) and image capture charged by a digital sensor located at the tip of the endoscope: either charged coupled device (CCD) or complementary metal oxide semiconductor (CMOS). In both fiberoptic and digital fURS, most manufacturers have models with a 3.6 Fr working channel (for irrigation and use of accessory instruments) and at least one 270° active deflection of the tip. The most recent development in fURS was the introduction of single use digital fURS [6, 7].

Due to all these technological advances, fURS has gained in popularity and has become a treatment of choice for kidney stones.

The objective of this review is to describe the surgical technique for fURS as well as tips and tricks for the treatment of renal stones.

Indications

The different options of treatment offered to patients with kidney stones are:

  • Extracorporeal shockwave lithotripsy (ESWL),

  • Flexible ureteroscopy (fURS),

  • Percutaneous Nephrolithotomy (PCNL): standard and miniaturized PCNL (mini-PCNL, ultramini-PCNL, micro-PCNL),

  • Open surgical and laparoscopic removal.

In case of cystine or uric acid stones, ESWL is not an option because cystine stones are often poorly radio-opaque and resistant to ESWL fragmentation, and pure uric acid stones are radiolucent.

According to the last EAU guidelines, the indications for active stone removal of renal stones are: symptomatic stones (e.g., pain or hematuria), stones > 15 mm, stones < 15 mm if observation is not the option of choice, obstruction caused by stones, stone growth, stones in high-risk patients for stone formation, infection, patient preference, comorbidity, social situation of the patient (e.g., profession or travelling), choice of treatment [8]. AUA Guidelines support stone treatment for asymptomatic stones in cases of stone growth, associated infection, and specific situations such as vocational reasons. Otherwise, surgical treatment is indicated for patients with symptomatic stones and/or obstruction.

Once the indication for stone removal is taken, the choice of the treatment modality depends on several stone characteristics such as: size, location, composition (if available), collecting system anatomy; and other factors such as: complication rate, patient comorbidities, patient preference, available technical equipment, and economic aspects.

Both AUA and EAU provide recommendations for the selection of procedure for active removal of kidney stones [8,9,10].

AUA guidelines [9, 10]

The latest recommendations of the AUA on surgical treatment of kidney stones were published in 2016.

ESWL, fURS, and PCNL are the treatment options for symptomatic patients with < 20 mm non-lower pole renal stone. ESWL or fURS is recommended as first-line treatment for such stones because their stone-free rates are acceptable and have less morbidity compared to PCNL.

For symptomatic ≤ 10 mm lower pole kidney stone, ESWL and fURS are the two options that urologists should offer. However, patients with a skin-to-stone distance greater than 9–10 cm or stone attenuation greater than 900–1000 Hounsfield units have less successful results with ESWL. An endoscopic approach (fURS or PCNL) is recommended for 10–20 mm lower pole renal stones because they appear to be less affected by stone burden than ESWL.

For stones > 20 mm, PCNL is recommended as first-line option because it offers a higher stone-free rate than ESWL or fURS and is less invasive than open surgery or laparoscopic/robotic-assisted procedures. PCNL should also be the first treatment utilized for most patients with staghorn calculi.

fURS should be offered as first-line therapy in most patients who require active stone removal with uncorrected bleeding diatheses or who require continuous anticoagulation/antiplatelet therapy.

The recommendations are summarized in Tables 1 and 2.

Table 1 AUA treatment algorithm for renal calculi (< 20 mm lower pole renal calculi excluded)
Table 2 AUA treatment algorithm for < 20 mm lower pole renal calculi

EAU guidelines [8]

The latest recommendations of the EAU on surgical treatment of kidney stones were published in 2017.

ESWL and fURS are the first-line treatment options for renal stone < 10 mm.

For 10–20 mm non-lower pole renal stone, ESWL, fURS, and PCNL are the treatment options that clinicians should offer. For 10–20 mm lower pole renal stone, PCNL or fURS is recommended in case of unfavorable factors for ESWL because its efficacy is limited. With favorable factors for ESWL, ESWL or an endoscopic approach (fURS or PCNL) should be offered. Unfavorable factors for ESWL are: shockwave-resistant stones (calcium oxalate monohydrate, brushite, or cystine), steep infundibular-pelvic angle, long lower pole calyx (> 10 mm), and narrow infundibulum (< 5 mm).

Kidney stone greater than 20 mm should be treated by PCNL first-line because it offers a higher stone-free rate than ESWL or fURS. In case PCNL is not an option or contraindicated, renal stone > 20 mm may be treated with fURS. However, in that case there is a higher risk of staged procedures and placement of a ureteral stent may be needed.

In case of an uncorrected bleeding disorder or continued antithrombotic therapy, fURS, in contrast to ESWL and PCNL, might offer an alternative approach since it is associated with less morbidity.

The recommendations are summarized in Tables 3 and 4.

Table 3 EAU treatment algorithm for renal calculi (10–20 mm lower pole renal calculi excluded)
Table 4 EAU treatment algorithm for 10–20 mm lower pole renal calculi

Contraindications

There are no specific contraindications to perform fURS apart from general problems such as general anesthesia or untreated preoperative positive urinalysis.

Technique, tips and tricks

Patient positioning, anesthesia

The patient is placed in lithotomy position. Although local, spinal anesthesia and intravenous sedation are feasible options; general anesthesia is suggested because it offers some advantages: renal movement caused by respiration may affect renal navigation and lithotripsy whereas with the mechanical ventilation the use of apnea stops temporarily these issues [11,12,13]. Furthermore, in case of large stone burden the time frame for spinal anesthesia can be exceeded, requiring a second general anesthesia to complete the procedure.

Access to the upper urinary tract and ureteral access sheath placement

The procedure starts with a rigid cystoscopy to look for location and configuration of the ureteral orifices. Then, a guidewire is placed into the renal cavities. Our preference is to use a 0.035″ stiff hydrophilic guidewire. They are other options than hydrophilic guidewire but a key point is that this latter has to be stiff in order to insert easily instruments and/or stents over it. Although the feasibility of performing fURS without a working or safety guidewire and without intraoperative complications has been reported, we suggest the use of a safety guidewire as it is still useful and does not add any morbidity. Moreover, the use of a safety guidewire ensures access to the collecting system and facilitates the insertion of a stent in case of ureteric or collecting system injury such as ureteral perforation or excessive bleeding [14,15,16,17,18]. Alternatively, the procedure may start with a first-look semirigid ureteroscopy, always using a safety guidewire, which has the advantage of passively dilating the whole ureter in order to choose the most appropriate ureteral access sheath (UAS) size [19]. Once the guidewire has been placed with or without performance of semirigid ureteroscopy, the fURS may be passed up to the kidney alongside this safety guidewire or over a second guidewire (inserted using a dual-lumen catheter) under fluoroscopic guidance. The other option is to place a UAS under fluoroscopic control and insert the fURS through the UAS. These different options for endoscope insertion depend on stone burden, upper urinary tract anatomy, and surgeon’s preference.

The goals of the UAS are: to facilitate multiple passages for stone fragments removal and most importantly providing irrigation with better fluid outflow, thereby decreasing the intrarenal pressure [2, 20,21,22,23]. The choice of the UAS, its size and diameter depends on the patient’s anatomy, the endoscope used, and surgeon’s preference [24, 25]. Instead of using two guidewires, some new designed UAS use only one guidewire, meaning the working guidewire turns into safety guidewire [26, 27] (Fig. 1). Attention has to be paid to the force applied during UAS placement because ureteral lesions may be the consequence of forced manoeuvers. Furthermore, in case of second session procedure, the presence of stone fragments, which may have migrated into the ureter, has to be removed before any UAS insertion. If any resistance is encountered during UAS placement, this latter should be stopped and the endoscope introduced to explore the ureter. If not possible, it is safer to place a ureteral stent for passive ureteral dilation, making a secondary fURS procedure much easier to perform at a later date. The best position of the UAS is with its tip in the proximal ureter or just below the ureteropelvic junction but not through it since this is the portion of the ureter at greatest risk for avulsion because of the least muscular tissue support. Taking these data into consideration, the use of the smaller UAS as possible in accordance with the fURS shaft size fitting is suggested. Depending on the fURS used, we suggest the 10/12-Fr UAS as first-line choice as it is the best compromise in terms of intrarenal pressure and irrigation [25].

Fig. 1
figure 1

Ureteral access sheath. The removal of the inner tip automatically converts the working guidewire as a safety guidewire laterally to the sheath

Once the fURS placed into the kidney, the collecting system is explored starting with the upper calices followed by the middle and lower calices. In the right kidney, the calyces are seen on the left of the endoscope screen. Thus, supination is the essential movement to explore the right renal cavities for urologist with right dominant hand. Conversely, the left renal calyces are seen on the right of the endoscope screen, meaning that pronation is essential when exploring left kidney (Fig. 2).

Fig. 2
figure 2

Dominant-hand pronation and supination to change the direction of the distal tip of the scope. Supination is essential when the scope is in the right renal cavities and pronation is essential when the fURS is in the left renal cavities

Lithotripsy and fragment extraction

Laser lithotripsy, using holmium:yttrium-aluminum-garnet (Ho:YAG) laser and laser fiber, has become the standard for stone lithotripsy during fURS procedure as it proved its efficacy and safety [28]. It delivers a pulsatile energy and operates at a wavelength of 2100 nm which is the infrared spectrum and is invisible to the human eye. The holmium laser is effective for any composition of stones. Stones are ablated via a photothermal effect whereby longer laser pulses (100–1000 μs) induce a temperature rise inside the stone. Laser radiation is absorbed by residual water in the stone, creating a vapour pressure that breaks the stone [29,30,31].

Although most of the manufacturers recommend to wear proper eye protection to avoid any corneal or retinal damage in case of unintentional laser exposure, unintentional Ho:YAG laser emission in air over distances of more than 5 cm may not be harmful, even if high pulse energies such as 2 J are used and for accidental emission periods of up to 3 s. Only corneal lesions may occur in the absence of eye protection and only when laser is set to high energy (2 J) and activated from a short distance (0–5 cm). Thus, it will be more beneficial to wear lead-lined glasses as eye lens is highly radiosensitive and prolonged radiation exposure increases the risk of developing cataracts. Furthermore, these lead-lined glasses are as protective as the laser safety glasses to prevent any eye damage [32].

In case of use of reusable laser fiber, most of the manufacturers recommend to prepare the laser fiber tip before initiating laser lithotripsy by stripping several millimeters from the terminal portion of the polymer coating, preventing any possible performance degradation from previous laser emission. However it has been shown that stripping the laser fiber tip reduces lithotripsy performance compared to coated laser fibers, resulting in a lower ablation volume and higher fiber tip degradation. Moreover, using metal scissors to cleave coated laser fibers does not significantly influence fiber performance compared to ceramic scissors [33, 34].

Once the laser fiber is inserted into the working channel, it is recommended to place the laser fiber tip until reaching one-quarter of the endoscope screen to avoid any fURS tip damage caused by the cavitation bubble created by laser activation (Fig. 3) [35].

Fig. 3
figure 3

Laser fiber at one-quarter of the screen in a digital fURS

The majority of laser lithotripters allow the urologist to control two parameters: the pulse energy and pulse frequency (Fig. 4). The total power output results from the product of these two parameters: total power (W) = pulse energy (J) × pulse frequency (Hz). By adjusting these parameters, the urologist decides the energy intensity that is delivered at the tip of the laser fiber to ablate the stone. With the release of a new generation of laser lithotripters, a third parameter can be set: the pulse duration. The laser machine allows choosing between short-pulse and long-pulse modes (Fig. 3). The difference between these two modes is that for a same amount of energy, its distribution is over a short (180–330 μs) or a longer period of time (650–1215 μs) [36]. At the same power levels, high frequency–low pulse energy is less efficient than low frequency–high pulse energy lithotripsy. Stone ablation volume directly correlates with pulse energy, whereas frequency, especially high frequency, and total power setting play more minor roles [37]. An in vitro study showed that the short-pulse mode is 25.0 and 9.9% more ablative than the long-pulse mode when associated with low frequency–high pulse energy and high frequency–low pulse energy, respectively [36,37,38]. Low pulse energy, like high frequency–low pulse energy settings, produces smaller stone fragments (also called “dusting effect”) than high-pulse energy, such as low frequency–high pulse energy settings [39,40,41]. Thus, two options for stone laser lithotripsy are described: “dusting” where the laser lithotripsy reduces the stone to fine dust and to small fragments 2 mm or less in diameter which are supposed to pass spontaneously avoiding the use of basket, or stone fragmentation producing stone pieces which have subsequently to be removed individually with a basket [39,40,41,42] (Fig. 5). Dusting lithotripsy requires to work over the stone’s surface and avoid being always in contact at the same place by “panting” with the laser fiber [42]. If fragmentation is performed and a 10/12-Fr UAS used for example, the stones have to be reduced to 3 mm fragments, which is the maximum stone size that can be removed through this sheath. The removal of such fragments is time consuming and may require a specific laser lithotripsy technique called “popcorn effect” to reduce the stone burden, thereby avoiding numerous entries/reentries with basket [37, 43, 44]. Each method has advantages and disadvantages, but surgeon’s preference, stone size, composition, location and intrarenal anatomy determine which ablative technique is the most appropriate. Furthermore, the two techniques can be combined starting with dusting and finishing with fragmentation to get stone pieces for stone analysis. For large stones, the fragmentation alone will produce a multitude of small fragments whose removal may be time consuming. Conversely, complete dusting of such stones may also increase operative time. Thus, both techniques may be used for large stones as well as the “popcorn effect” when significant stone pieces are remaining at the end of the procedure to avoid basketing of all the fragments.

Fig. 4
figure 4

Laser parameters

Fig. 5
figure 5

Laser lithotripsy. a Dusting effect, b fragmentation

The different laser settings for stone lithotripsy are:

  • Fragmentation: high energy (1–2 J), low frequency (3–5 Hz) and if needed short pulse duration.

  • Dusting: low energy (0.2–0.5 J), high frequency (10–20 Hz) and if needed long pulse duration.

  • “Popcorn effect”: high energy (1 J), high frequency (10–20 Hz) and if needed long pulse duration.

Another parameter which may facilitate laser lithotripsy is the influence of the working channel position. Most of the fURS have a 3 or 9 o’clock working channel position. When performing laser lithotripsy in a right kidney, the calyces are seen on the left of the endoscope screen and the stones are located on the right of the screen because of the gravity. Thus, a 3 o’clock working channel position is better to ablate stones. However, to better fragment stones located in right anterior calyces, a 9 o’clock working channel position is advised. Conversely, a 9 o’clock working channel is advised for left renal stones except for posterior calyces where a 3 o’clock working channel is preferable (Fig. 6).

Fig. 6
figure 6

Working channel position. Left: 9 o’clock working channel, right: 3 o’clock working channel

The last parameter to take into account is the laser fiber diameter. Large fibers (550 µm) produce wider fissures, while small fibers (200 µm) create deep fissures, but neither surpassed the other in terms of ablation volume [37, 38]. Although large diameter laser fibers are more resistant to fiber tip degradation or the ‘burn-back’ effect, their stiffness reduces fURS maximal deflection and their diameter decreases irrigation flow. Conversely, small diameter laser fibers are more fragile and prone to fiber tip degradation, but this is compensated for by better irrigation, better endoscope deflection, and less retropulsion [37, 45, 46].

Finally, when stone removal has to be performed, tipless nitinol baskets have become the standard devices and many designed are available on the market (Fig. 7). The advantages of nitinol baskets are their flexibility and ability to not damage the scope and intrarenal cavities during fragments extraction. The diameter smaller than 2.2-Fr allows maximum irrigation flow and deflection of the endoscope, whereby all parts of the collecting system can be reached [46]. Blind basketing has to be avoided as it may increase the risk for ureteral avulsion. Thus, stone retrieval with baskets, even if a UAS is used, has to be endoscopically guided. The stone and basket have to be kept under direct vision as they are removed; the size of the stone can be continually compared with the size of the ureteral lumen until the entry of the sheath if used. If the stone appears to be too large to be removed intact, it should be fragmented into smaller pieces that may be subsequently removed [47].

Fig. 7
figure 7

Examples of nitinol baskets

Irrigation

Although fURS is often performed with continuous saline irrigation through the working channel, an additional device such as syringe or manual pressure pump may be needed to improve the visibility when an instrument is inserted into the working channel, while dusting a stone or small venous bleeding happens. Furthermore, these devices may be helpful to flush stone fragments to facilitate basketing in difficult cases. However, although these irrigation pressure devices are helpful and often used, they may generate excessive intrarenal pressure, especially with an empty working channel. Depending on the irrigation system used, the level of intrarenal pressure may vary. Another factor affecting the intrarenal pressure is the physical strength of the assistant using these irrigations devices [48]. While working in the renal cavities for a long time, placing a UAS may prevent the elevation of pressure but this has to be proven (Fig. 8).

Fig. 8
figure 8

Hand-assisted irrigation device

Lower pole stones

When considering lower pole stones, we have to emphasize that differences in accessing the lower pole exist between the fURSs. A recent study found that digital fURSs were less effective in accessing the sharp angled inferior calix and had lesser end-tip deflection compared with the fiberoptic fURSs. Thus, it might be better to use a fiberoptic fURS when approaching a difficult inferior calix [49].

Once fURS is in the renal cavities, attempt to relocate the stone with a nitinol basket to a more favorable position (upper pole or renal pelvis) has to be performed [50] (Fig. 9). This practice reduces the risk of fURS damage and makes the stone targeting during laser lithotripsy easier and more effective. Furthermore, according to the stone lithotripsy technique used, the stone fragments produced will be easier to grasp with the basket or able to pass spontaneously through the ureter instead of accumulating in the lower pole. However, stones larger than 10 mm may not be possible to grasp and relocate, necessitating laser treatment in situ, with the fURS maximally deflected, potentially increasing the risk of laser fiber failure and ureteroscope damage [51]. In that case, it is recommended to insert the laser fiber with the minimal deflection of the endoscope to avoid any working channel damage and subsequently access the lower pole to treat the stone in situ. However, once the laser fiber is in place, this results in a loss of the fURS maximal deflection and access to the lower pole may be difficult [46, 52]. For this reason has been developed a new design of laser fibers with a ball-tip shape, which reduces friction within the working channel in a completely deflected fURS, thereby reducing the likelihood of endoscope damage [53]. However, after the first minute of laser emission, the ball-shaped tip fibers lose their special feature due to fiber tip degradation, especially at high pulse energy settings [54]. Thus, this allows the insertion of a ball-tip fiber only once in a fully deflected fURS. A recent study reported that cleaving the tip of standard laser fibers with metal scissors results in equivalent passage capabilities through the deflected endoscopes than the ball-tip shaped fibers [52]. A recent work showed that laser fiber diameter and curve as well as laser settings are parameters to take into account when treating a lower pole renal stone in situ [55]. The 272 µm fibers broke at the curve diameter of 9 mm and below for laser dusting parameters (high frequency and low energy) and 12 mm for the laser fragmentation parameters (low frequency and high energy). The 365 µm fibers broke at 9 mm for the dusting parameters and required a smaller curve (15 mm) to break with the fragmentation parameters. Thus, to prevent laser fiber breakage, it is advised to use small 272 µm fibers, at a dusting parameter, to reduce the stone diameter until it can be relocated as soon as possible.

Fig. 9
figure 9

Relocation of a lower pole renal stone

Caliceal diverticulum

The procedure begins as standard fURS. Once the endoscope placed into the renal cavities; the principle consists of injecting contrast through the working channel to identify the diverticular neck with the assistance of fluoroscopic guidance (Fig. 10a, b). In difficult cases, it has been reported the blue test technique (Fig. 10c). Instead of injecting contrast alone, a mix of contrast and indigo carmin is injected. If opacification of the diverticulum is obtained, this means that its neck is patent. Then, the collecting system has to be washout using the saline irrigation through a manual irrigation device. Once the cavities are clear, a leakage of blue by the diverticular neck is identifiable. The next step is the laser incision of the diverticular neck and subsequently treatment of the stone burden [56].

Fig. 10
figure 10

a Diverticulum at the middle portion (arrow). b Flexible ureteroscope at the diverticulum. c Blue test technique

The use of apnea during laser lithotripsy

During laser lithotripsy of renal stones or in case of precise laser ablation (upper tract urothelial carcinomas, endopyelotomy, incision of diverticular neck, stone in narrow calyx), the renal movement caused by respiration and diaphragmatic excursion makes it challenging as the surgeon has to treat a moving target without injuring urothelium or papillae. For these reasons, apnea may be needed during fURS procedure to facilitate the treatment. A recent report described a protocol to perform apnea during fURS allowing its maintenance for approximately 5 min. This illustrates the role of the collaboration between anesthesiologists and surgeons in order to achieve the treatment [13].

Ureteral stenting

Careful ureteral exploration should be performed at the end of the procedure to retrieve the scope and the UAS while keeping the scope’s tip a few centimeters out of the sheath to detect eventual ureteral lesions due to UAS placement and stone fragments [57].

To prevent the incidence of postoperative ureteric obstruction and renal colic secondary to ureteral edema, a ureteral stent may be placed. This stent placement has also been advocated to facilitate the passage of stone fragments secondary to the passive ureteric dilation [58,59,60,61,62]. Although this common practice is still debated, it has been reported in large series that approximately 80% of urologists place a ureteric stent after uncomplicated ureteroscopy for stone disease [63,64,65,66,67,68,69,70,71]. The last AUA guidelines recommend ureteral stenting among patients with ureteric injury during URS, those with evidence of ureteral stricture or other anatomical impediments to stone fragment clearance, such as ureteral wall edema, a large stone burden (> 15 mm), those who have an anatomically or functionally solitary kidney or renal functional impairment, and in those in whom another ipsilateral URS is planned [9, 10].

Conclusion

Flexible ureteroscopy has become widely used due to the technical advancements and refinements in the endourologic armamentarium made these last two decades. Its high efficacy and low morbidity in the treatment of renal stones up to 20 mm have made it a treatment of choice for the management of renal calculi.