Introduction

Computational fluid dynamics (CFD) is a numerical method for modelling haemodynamics, which has been applied to the study of intracranial aneurysms. Image-based simulations of blood flow dynamics have been reported in different types of cerebral aneurysms [13] and a number of haemodynamic parameters studied to understand the influence of flow on their natural history [47]. More recently, researchers have attempted to simulate the haemodynamic consequences of endovascular interventional procedures using data from clinical cases [811].

In this study, we used a novel computational method, based on CFD, to simulate the endovascular coil embolisation of a recently ruptured cerebral aneurysm in an anatomically accurate 3-D computer model representation derived from a clinical case. The method used to model the coiling process of this aneurysm involved simulation of a porous medium within the aneurysm and decreasing its porosity as coils are inserted. The characteristics of the porous medium were modelled on the coil chart used in clinical practice. Subsequently, the haemodynamic changes induced during the different stages of the coiling process were analysed, as well as the local haemodynamic conditions which develop following aneurysm occlusion with a residual neck remnant. To our knowledge, this is the first report of such a study.

Material and methods

Aneurysm realistic computer model

The clinical data used for modelling purposes were derived from a ruptured large and wide-necked bifurcation aneurysm (11×9.4×7.8 mm) arising from the anterior communicating artery (Fig. 1). A series of 15 GDC-10 detachable platinum coils (Boston Scientific, Target, Fremont, Calif.) were deployed (Table 1). A packing density of 26.5% (i.e. percentage of sac lumen occupied by metal) was simulated but leaving a small aneurysm remnant at the neck area to ensure unobstructed flow through both distal anterior cerebral arteries (Fig. 2). 3-D imaging data were obtained using a Siemens Axiom-Artis angiography system (Siemens Medical Solutions, Erlangen, Germany). 3-D subtraction angiography images were reconstituted from a rotational acquisition. The protocol used included a 200° spin over 6 s at 40°/s and imaging at 1.5°f/s, followed by intraarterial injection of 15 ml of a nonionic iodinated contrast agent (Niopam 300 mg/ml, Nycomed, France) using an automatically triggered pump injection through a catheter positioned in the left internal carotid artery (ICA). The X-ray exposure parameters were: 70 kV, 155 mA, pulse width 10 ms and 1.20 μGy/pulse. A total of 150 mask and angiography images were acquired in 1,0242 matrices and reconstituted on a Leonardo workstation (Siemens) for 3-D display. The anonymized datasets were processed using Amira 3.1 (Mercury Computer Systems, Chelmsford, Mass.), a visualization system able to represent 3-D aneurysm imaging data as a triangular surface and create volumetric tetrahedral grids suitable for numerical simulations (Fig. 2). The data created were then imported into the Gambit 2.2.30 (Fluent, Lebanon, N.H.) software for surface and volume discretization.

Fig. 1
figure 1

Left ICA digital subtraction angiogram showing the ruptured anterior communicating artery aneurysm

Table 1 Coiling procedure for occlusion of our anterior communicating artery aneurysm
Fig. 2
figure 2

Aneurysm computer model with partition of its volume into a porous medium (blue) and a pure fluid part (red). Notice also the accommodation of the neck remnant (arrow)

Porous medium modelling and CFD

The main objective of this study was to model the endosaccular deployment of the platinum coils with the developing thrombus in the aneurysm volume as a porous medium of decreasing porosity. This idea was based on the uniformity and randomness of the coiling process and the relative scales involved. Moreover, it seems to be the only feasible option—within current and projected computational resources—since it can yield self-consistent numerical results (grid-independent).

Mathematical models describing transport phenomena through porous regions depend on a set of constituent characteristics of the porous medium. The methodology presented here requires the specification of the porosity, inverse permeability and the drag factor for flow through the porous medium. The porosity of the medium represents the volume occupied by the pores to the total volume of the system, while permeability is a measure of the fluid conductivity through the porous medium, representing the surface area to volume ratio of the porous matrix. Our porous media model followed the theory of Kozeny [12] and all changes in porosity, interstitial surface area and permeability during the coiling process (Table 2) were derived following the clinical coil chart (Table 1). The corresponding drag factor in the porous medium was estimated at CD≈2.2 [13]. The orthogonal dimensions of the aneurysm were axial 9.4 mm, sagittal 7.8 mm and coronal 11 mm, giving a total volume of Van=4.22293×10−7 m3. All the coils used were size 10, having a diameter, d, of 2.54×10−4 m. Assuming that blood flow has an average velocity of Vav=0.1 m/s, the Reynolds number (Re) at the inlet of the vasculature is 213 varying up to 363 close to the aneurysm (flow conditions appropriate to the anterior communicating artery [14]).

Table 2 Porosity characteristics for our anterior communicating artery aneurysm

A haemodynamic model was subsequently developed to describe the flow, pressure and force fields of the system. This was implemented on CFD-ACE software (ESI Group, Rungis, France) [15] assuming a Newtonian liquid of density ρ=1069 kg/m3 and constant dynamic viscosity μ=0.0035 kg/m-s, nonpulsatile flow, noncompliant walls and constant outlet pressure.

Results

We report the results of computational haemodynamic calculations for three different stages of the coiling procedure: following deployment of the 1st coil, the 7th coil (intermediate stage of the coiling simulation), and the 15th and final coil, leaving a small residual neck remnant. For every simulation, blood flow patterns and velocity as well as wall pressure distribution were computed and compared with the pretreatment simulation.

Introduction of the first coil had a dramatic effect on the blood flow pattern, although it occupied only 3.6% of the aneurysm volume (porosity 0.964, Tables 1 and 2). This first coil, as represented in the simulation by the porous medium, changed the endosaccular flow pattern from complex vortical to a concentrated inlet/outlet mode with more organized and defined streaming (Fig. 3). Similar changes in blood flow velocity were also detected at this stage (Figs. 4b and 5b).

Fig. 3
figure 3

Flow patterns before (a) and after (b) simulation of the deployment of the first coil. A significant change to a more constructed mode is identified

Fig. 4
figure 4

Velocity (m/s) contours at a sagittal slice before (a) and after simulation of the deployment of the 1st (b), 7th (c) and 15th (d) coils. A gradual reduction in blood flow velocity up to stagnation is obvious

Fig. 5
figure 5

Velocity (m/s) contours at a coronal slice before (a) and after simulation of the deployment of the 1st (b), 7th (c) and 15th (d) coils. There is still marked blood flow velocity at the coil mesh interface in the neck area (arrow), even after coiling is completed (d)

At an intermediate stage in the coiling simulation, i.e. during deployment of the 7th coil, the aneurysm occlusion was 19.2% and a significant decrease in blood flow velocity was noted with stagnation in some areas (Figs. 4c and 5c). At the end of the coiling simulation, a packing density of 26.5% (represented by a porosity of 0.735 in the porous medium, Tables 1 and 2) was achieved and the decrease in flow velocity was so profound that large areas, especially around the aneurysm dome, showed blood flow arrest and stagnation (Fig. 4d). At the same time, flow was still detected at the coil mesh/parent vessel interface, i.e. at the aneurysm neck area (Fig. 5d). This was despite a simulation of packing density above the average (a mean packing density of 24–25% is suggested as enough for a stable occlusion by most authors) [16, 17].

Although the gradual introduction of coils results in a stepwise relief of wall pressure at the aneurysm dome, a redistribution of wall pressure was identified at the aneurysm inflow zone during the intermediate stage of coil packing. This was also evident at the end-of-treatment simulation (Fig. 6) and may be a factor in determining if aneurysm recurrence and regrowth subsequently occurs.

Fig. 6
figure 6

Wall pressure (Pa) contours before (a) and after (b) coiling simulation. Although the high pressure at the aneurysm dome has been relieved (b straight arrow), an extension of the high pressure area to the neck remnant is also noted (b curved arrow)

Discussion

The main purpose of this study was to examine whether a porous medium approach, with porous characteristics derived to simulate the coils actually used during treatment, is capable of simulating the haemodynamic consequences of coiling. The porous medium representation of the coil mesh allows the simulation of blood flow circulating through the coil basket during all coiling stages and even after achieving a satisfactory packing density level. This seems to be close to a realistic representation of the in vivo situation, in which the coil mesh does not act as a solid barrier but rather obstructs and slows intraaneurismal blood flow circulation, to promote blood stagnation and clot formation.

Further validation and encouragement for our modelling approach is derived from observed consistency with clinical observations and practice. Follow-up clinical series [1822] have consistently shown significant recurrence and/or recanalization rates (17–20%), especially in large wide-necked aneurysms, but low rebleeding rates (0.5–0.8%). The favourable haemodynamic environment created by the coiling process facilitates intraaneurysm thrombosis, protecting the vulnerable aneurysm dome area and preventing rebleeding. However, the redistribution of wall pressure to the neck area and the persistent flow through the coil mesh, especially if a neck remnant coexists, may contribute to the relatively high rate of coil compaction and recanalization. It is obvious though that the complex phenomena of aneurysm thrombosis, recanalization and growth cannot be oversimplified and explained in haemodynamic terms only [23]. Furthermore, we know that the haemodynamic parameters can be seriously affected by many factors such as different aneurysm geometries and local haemodynamic conditions and thus general conclusions cannot be drawn from a single case only.

In the literature there have been attempts to simulate aneurysm coiling using different methods, but obtaining qualitatively similar results to ours. Byun and Rhee [24] reported CFD calculations for artificially constructed models focusing on different parent vessel geometries and coil placement assuming coils to be a sphere of blocked cells. Groden et al. [25] also employed an artificial CFD model using blocked cells to represent individual coils. Our methodology, based on porous media, follows a randomized process of coil simulation in the aneurysm lumen, eliminating the uncertainty introduced by the level of local grid refinement around individual coils and is therefore computationally efficient and self-consistent, since it leads to grid-independent solutions.

Most modelling attempts to simulate clinical situations—including the one described here—suffer from simplifying assumptions such as rigid walled vasculature, blood being considered as a Newtonian fluid, and use of steady-state blood flow. We adopted the last compromise in order to reduce the computational effort required for the simulations since the aim of the study was to test the utilization of the porous medium representation of the coil complex. However, our team has performed further additional simulations using the same method with different aneurysm geometries, porous media parameters and blood flow velocities, which resulted in similar haemodynamic alterations, enhancing further our confidence in the reliability of the proposed method.

By simulating a coiling scenario for a particular aneurysm, we obtained an overview of the haemodynamics in order to test the feasibility of porous media as a means for modelling coils in CFD. The use of porous media in this context is the novelty of this study and this method has not been used previously. The results of the simulations are encouraging and we plan to continue with further CFD studies using this technique. These will include induction of pulsatile flow, in addition to steady flow, as well as a series of aneurysm models based on different clinical cases. Furthermore, additional developments regarding the simulation of the coagulation process are also in progress by our team [26], which will further enhance the reliability of the modelling approach.

Conclusion

This study demonstrates that a computational simulation technique applied to patient-specific aneurysm geometry can be used to model the haemodynamic consequences of coiling. Although it seems that porous media may be a promising method for simulating coiling in such CFD models, it is obvious that there is a long way to go before simulations can be used as a basis for clinical decisions. However, it seems likely that in the future simulation studies will become a valuable tool in interventional treatment planning and decision-making.