Abstract
A better repeatability and accuracy in the quantitative determination of trace elements in mussel shells or carbonate-based materials by LA-ICP-MS was achieved by using a series of multielement calibration standards prepared by co-precipitation of twelve elements into a CaCO3 matrix in order to improve the homogeneity of the resulting powder samples. Pressed powder discs of good mechanical stability could be obtained at a pressure of 50 MPa, without the addition of a binder. An UV laser (modified Nd:YAG, 266 nm) was used in the Q-switched mode at a repetition rate of 10 Hz and an energy level of 3.5 mJ. Correlation coefficients (R) for the linear calibration graphs (concentration range: 1.5–400 μg/g) for Cr, Mn, Co, Cu, Zn, As, Cd, Sn, Ba, and Pb are generally better than 0.997. The detection limits for all elements investigated are in the sub-μg/g range. Incorporation of elements into the matrix by co-precipitation has shown as a superior method for producing calibration standards than the simple mixture of the analytes (in carbonate or oxide form) with the matrix (CaCO3) or addition of standard solutions to a carbonate powder base. Two examples of the quantitative determination of toxic elements in mussel shells will be presented.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Author information
Authors and Affiliations
Additional information
Received: 10 January 2000 / Revised: 16 March 2000 / Accepted: 26 March 2000
Rights and permissions
About this article
Cite this article
Bellotto, V., Miekeley, N. Improvements in calibration procedures for the quantitative determination of trace elements in carbonate material (mussel shells) by laser ablation ICP-MS. Fresenius J Anal Chem 367, 635–640 (2000). https://doi.org/10.1007/s002160000421
Issue Date:
DOI: https://doi.org/10.1007/s002160000421