Abstract
Purpose
Few studies investigated whether trochlear and patellar design enhancements improve long-term outcomes of total knee arthroplasty (TKA). This study aimed to compare the long-term survival and complication rates of two consecutive generations of the same TKA system with identical tibiofemoral geometry, but different patellofemoral designs.
Methods
The authors retrieved the records of 93 patients (104 knees) operated with the HLS II system and 116 patients (122 knees) operated with HLS Evolution system. Patients were evaluated preoperatively and at a minimum of 10 years noting all complications. Kaplan–Meier (KM) survival was compared for two endpoints: (1) revision of all components and (2) revision of any component.
Results
From the HLS II series, the incidence of revision of all components was 6.4%, and of any component was 9.8%. From the HLS Evolution series, the incidence of revision of all components was 4.1%, and of any component was 5.1%. Comparing the survival at equivalent follow-up of 14 years, considering revision of all components, the HLS II had higher survival than the HLS Evolution (98.9% vs 95.9%), while considering revision of any component, the HLS II had lower survival than the HLS Evolution (93.0% vs 94.9%). The differences in survival of the two implants were not significant, neither at equivalent follow-up of 14 years (n.s.), nor at maximum follow-up of each cohort (n.s.). The complication rate was higher for the HLS II series compared to the HLS Evolution (28% vs 12%, p = 0.009), but patellofemoral complications were not more frequent (8% vs 6%, n.s.).
Conclusions
Though the differences in survival of the two implants were not significant, conflicting findings are observed due to partial revisions for patellar fractures (5 in the HLS II series and 1 in the HLS Evolution series) which could be related to patellofemoral design enhancements. This study highlights the importance of patello-femoral geometry, which is often overlooked in TKA.
Level of evidence
Retrospective comparative study, Level III.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Long-term survival of total knee arthroplasty (TKA) was demonstrated by landmark studies on historic implants [12, 35]. Despite its success, many patients remain dissatisfied with TKA, often due to patellofemoral complications, which remain reasons for revision [9, 22, 32, 39]. Patellofemoral pain and instability are often caused by extensor mechanism malalignment [7, 31, 39] or inappropriate implant design [10, 17]. Numerous biomechanical studies highlighted the importance of trochlear and patellar geometries [1, 29, 40, 41], which led several manufacturers to develop more patella-friendly implants [10, 36], though few studies investigated the clinical benefits of such design enhancements [43].
The HLS (Hôpital Lyon Sud) system is a unique TKA design, postero-stabilized (PS) by a central ‘third condyle’, that successfully reproduces native knee kinematics [45]. It has provided excellent clinical and radiological short-term outcomes [28, 38], with satisfactory survival up to 10 years [13, 24, 38]. The HLS system underwent several incremental improvements since its introduction in 1987, which appeared to reduce the incidence of short-term patellofemoral complications [13]. Notably, the HLS Evolution™ (third generation) had identical tibiofemoral geometry, but enhanced trochlear design in the inter-condylar region compared to its predecessor, the HLS II™ (second generation), to improve patellar tracking. Furthermore, the HLS Evolution also featured three small fixation pegs for its patellar button, to replace the large central peg of the HLS II that rendered the patella susceptible to fracture [13]. However, the long-term effects of these design changes on implant survival and patellofemoral complications remain unknown.
Few studies reported the survival of the HLS II and HLS Evolution TKAs beyond 10 years, and none investigated whether the design enhancements were effective at reducing patellofemoral complications. The purpose of this study was therefore to report the long-term survival of the HLS II and HLS Evolution TKA systems, and to compare their complication rates, with particular focus on patellofemoral complications.
Materials and methods
The authors analysed the survival and compared the complication rates of two TKA series of the 2nd and 3rd generations of the HLS system (Tornier, St-Ismier, France), HLS II and HLS Evolution.
Patient characteristics and surgical data
The records of all TKAs performed using the HLS II system at the authors’ centre between January 1992 and December 1993 were reviewed. From a total of 93 patients (104 knees), 1 patient (1 knee) was excluded due to prior surgery on the patella, leaving 92 patients (103 knees). (Table 1). The indications for surgery were primary osteoarthritis in 68 knees (66%) and rheumatoid arthritis in 20 knees (19%). The surgical approach was medial parapatellar in 93 knees (90%), lateral parapatellar in 4 knees (4%), midvastus in 1 knee (1%), and unspecified in 5 knees (5%).
The records of all TKAs performed using the HLS Evolution system at the authors’ centre between January 1998 and December 2000 were reviewed. From a total of 116 patients (122 knees), 1 patient (1 knee) was excluded because the patella was not resurfaced, leaving 115 patients (121 knees). (Table 1). The indications for surgery were significantly different, with primary osteoarthritis in 99 knees (82%).The surgical approach was medial parapatellar in 79 knees (65%), lateral parapatellar in 8 knees (7%), midvastus in 34 knees (28%).
Clinical assessment
All patients were evaluated preoperatively, noting their knee alignment, Charnley score [15, 34] to assess the patients’ walking ability, and the preoperative Knee Society Score (KSS). Patients were contacted by telephone and/or mail to update their records. If patients were deceased, their general practitioner or next of kin was contacted to confirm the date and cause of death, and whether any of their TKA components had been revised. All patients living with their original TKA components were invited for a clinical evaluation, and those unable to travel were evaluated by telephone. The clinical questionnaire noted the KSS and Oxford Knee Score (OKS), as well as any complications [16], with particular focus those related to the patella or extensor mechanism.
From the HLS II cohort of 92 patients (103 knees), 3 patients (3 knees) had two or more TKA components revised, and 4 patients (4 knees) had only one TKA component revised (Fig. 1). A total of 51 patients (58 knees) had died with their original components in place, and 7 patients (8 knees) could not be reached, but their most recent follow-up records (at 8.6 ± 1.4 years) indicated that none had revision surgery. This left a cohort of 29 patients (30 knees) living with their original TKA components. Twenty-seven patients (28 knees) were evaluated at a minimum follow-up of 15 years: 15 patients (15 knees) were examined at the clinic, whereas 12 patients (13 knees) were surveyed by telephone as they were unable to travel because of limited mobility or poor health, and 6 were not surveyed as they were bedridden or confined to nursing homes.
From the HLS Evolution cohort of 115 patients (121 knees), 4 patients (4 knees) had two or more TKA components revised, and 1 patients (1 knee) had had only one TKA component revised (Fig. 1). A total of 36 patients (39 knees) had died with their original components in place, and 8 patients (8 knees) could not be reached, but their most recent follow-up records (at 2.5 ± 3.0 years) indicated that none had revision surgery. This left a cohort of 67 patients (69 knees) living with their original TKA components. Sixty-five patients (67 knees) were evaluated at a minimum follow-up of 10 years: 33 patients (35 knees) were examined at the clinic, whereas 32 patients (32 knees) were surveyed by telephone as they were unable to travel because of limited mobility or poor health, and 13 were not surveyed as they were bedridden or confined to nursing homes.
All patients provided informed consent to use their data for research and publishing purposes. As the study used retrospective data, collected for routine clinical assessment, the Institutional Review Board (Lille University Hospital, France) waived the standard ethical approval process.
Statistical analysis
Descriptive statistics were used to summarize the data. For non-Gaussian quantitative data, between group differences were evaluated using Wilcoxon rank sum tests (Mann–Whitney U test). Categorical data were analysed using Chi-square tests. Survival analysis was performed for 2 different endpoints: (1) revision of all components, and (2) revision of any component. The authors calculated implant survival using the Kaplan–Meier (KM) method and using the cumulative incidence function (CIF), both of which were deemed important for this study. The KM survival is most frequently reported for joint arthroplasty and enables direct comparison with published survival data, whereas the CIF is recommended for reporting survival at 10 or more years, where risks of revision are overestimated by KM because of greater proportions of patients deceased and/or lost to follow-up [4, 11, 20, 21, 33, 44]. The differences between the KM curves of both implants and their endpoints were evaluated using the log-rank tests. The cumulative incidence of competing risks of the implants and their endpoints was evaluated using K-sample tests [14]. Statistical analyses were performed using R, version 3.3.2 (R Foundation for Statistical Computing, Vienna, Austria). p values < 0.05 were considered to be statistically significant.
Results
From the original HLS II cohort of 92 patients (103 knees), 29 patients (30 knees) were still living with their original implants at a mean follow-up of 20.2 ± 1.1 years (range, 15.3–21.4) (Fig. 1; Table 2). Considering revision of all components as endpoint, the revision incidence calculated using the KM method (1-survival) was 6.4% (CI, 2.0–19.7%), whereas using the CIF it was 3.2% (CI, 0–6.8%) (Fig. 2). Considering revision of any component as endpoint, the revision incidence calculated using the KM method was 9.8% (CI, 4.4–20.7%), whereas using the CIF it was 7.1% (CI 1.9–12.0%) (Fig. 3). For the 7 revised TKAs, the causes of revision were:
-
Aseptic patellar fracture in 3 knees, revised at 8 months, 15 months and 11.2 years;
-
Septic patellar fracture in 2 knees, revised at 2.9 years and 18.8 years;
-
Severe polyethylene wear in 1 knee, revised at 17 years;
-
Deep infection in 1 knee, revised at 2 years.
From the original HLS Evolution cohort of 115 patients (121 knees), 67 patients (69 knees) were still living with their original implants at a mean follow-up of 13.3 ± 0.9 years (range 10.0–14.4) (Fig. 1; Table 2). Considering revision of all components as endpoint, the revision incidence calculated using the KM method (1-survival) was 4.1% (CI 1.5–10.5%), whereas using the CIF it was 3.5% (CI 0.1–6.8%) (Fig. 2). Considering revision of any component as endpoint, the revision incidence calculated using the KM method was 5.1% (CI 2.2–11.9%), whereas using the CIF it was 4.4% (CI 0.6–8.0%) (Fig. 3). For the 5 revised TKAs, the causes of revision were:
-
Deep infection in 2 knees, revised at 8 months and 2 years;
-
Aseptic tibiofemoral loosening in 2 knees, revised at 5.5 years and 8.9 years;
-
Aseptic patellar fracture in 1 knee, revised at 5.5 years.
Comparing the survival at equivalent follow-up of 14 years, considering revision of all components, the HLS II had higher survival than the HLS Evolution (98.9% vs 95.9%), while considering revision of any component, the HLS II had lower survival than the HLS Evolution (93.0% vs 94.9%). The differences in survival of the two implants were not significant, neither at equivalent follow-up of 14 years (n.s.), nor at maximum follow-up of each cohort (n.s.).
Complications
Comparing complications at equivalent follow-up of 14 years, the complication rate was significantly higher for the HLS II series compared to the HLS Evolution series (26% vs 12%, p = 0.024) (Table 3), notably stiffness (11% vs 2%, p = 0.007), patellar fracture (5% vs 2%, n.s.) and tibial or femoral fracture (4% vs 1%, n.s.). The overall incidence complications related to the patellofemoral joint or extensor mechanism was only slightly greater for the HLS II than the HLS Evolution (7% vs 6%, n.s.). Post hoc analysis (using α = 0.05) revealed that the study was slightly underpowered (2-tailed, β = 0.76) to detect significant differences in complication rates.
Discussion
The main finding of the present study was that, although the difference in overall survival for either endpoint was not statistically significant, comparing KM survivals of the HLS II and HLS Evolution at equivalent follow-up of 14 years, the perceived superiority of one model over the other depends on the endpoint of interest. Considering revision of all components, the HLS II had higher survival than the HLS Evolution (98.9% vs 95.9%, both at 14 years), while considering revision of any component, the HLS II had lower survival than the HLS Evolution (93.0% vs 94.9%, both at 14 years). The differences in KM survivals of the two models was not statistically significant, possibly due to insufficient cohorts or events, though it remains interesting to understand the reasons behind our seemingly paradoxical findings. While the rates of patellofemoral complications did not differ significantly among the two series, the HLS II had five partial revisions due to patellar fracture, compared to the HLS Evolution which had only one partial revision due to patellar fracture.
The decrease in the incidence of patellar fracture observed is likely due to the design improvements from one implant generation to the other. The HLS Evolution was designed to prevent the patellofemoral complications observed with its predecessor, the HLS II, notably patellar loosening and fracture. The design improvements included a deeper trochlear groove in the inter-condylar region, and a patellar button with three small fixation pegs instead of one large central peg. Our findings, therefore, suggest that the design enhancements were effective at preventing patellar fractures that require revision, though they did not eliminate less severe patellofemoral complications, such as patellar instability and failure of the extensor mechanism. Our data does not enable us to determine which of the design enhancements contributed to preventing patellar fractures. The replacement of the large patellar fixation peg seems likely to prevent weakening the patellar bone after resurfacing [27], though it is possible that the enhanced trochlear groove prevented patello-femoral overload and fracture.
The HLS II series had significantly more overall complications than the HLS Evolution series. While it is possible that the design improvements are partially responsible for this reduction, we found significant differences in the cohorts which could explain this difference. First, the patients in the HLS Evolution cohort had better preoperative KSS knee and function scores than those in the HLS II cohort. Second, the surgical technique differed between the series, as HLS Evolution patients had fewer MCL releases. Finally, the postoperative rehabilitation for HLS Evolution cohort could have been improved from that of the HLS II cohort, with shorter immobilization and more effective exercises, which may have reduced the incidence of stiffness.
The survival rates beyond 15 years of the HLS implants were satisfactory: considering revision of all components as endpoint, the KM survival was 93.6% for the HLS II, compared to 95.9% for the HLS Evolution. Likewise, considering revision of any component as endpoint, the KM survival was 90.2% for the HLS II, compared to 94.9% for the HLS Evolution. Thus, the 15–20 years survival of the HLS implants is well within the range of 90–98.1% reported for postero-stabilized TKA [2, 7, 13, 21, 26, 28, 37, 42] and of 81.7–91% for cruciate-retaining TKA [3, 5, 8, 18], as well as of 85.7–98.9% for other TKA systems [2, 6, 19, 23, 30] (Table 4).
Both the KM survival of the implants and the cumulative incidence of revision using the CIF are reported in this study. Since our cumulative risk of mortality was greater than the cumulative risk of revision, the traditional KM estimates could be somewhat misleading because the competing risks exaggerate the perceived revision rates [4, 44]. The CIF is, therefore, recommended as an alternative or complement, although this method was recently introduced in arthroplasty studies [21, 25, 44]. Comparing revision rates for the present series using the KM and CIF estimates reveals that the former exaggerates the revision rates by an average of 17.2–18.2% for the HLS Evolution series at 14 years and by an average of 38.4–97.5% for the HLS II series at 20 years. Despite its limitations, especially on longer follow-up periods, KM remains the most commonly reported measure of survival in arthroplasty studies and the only option for comparison with other TKA studies on the literature.
The present study has several limitations inherent to long survival studies involving patients of advanced age at index surgery: (1) despite efforts to reach these patients, 9% of HLS II patients and 7% of HLS Evolution patients were lost to follow-up, which remains acceptable for long-term survival studies, (2) the small number of patients evaluated postoperatively, (3) the patients who were unable to travel to the clinic because of advanced age or limited mobility and had to be assessed by means of telephone consultations, which does not permit full appreciation of the range of motion and function, and (4) the small cohort sizes that rendered insufficient observations to determine statistical significance. It is also worth noting that the comparisons made between the two HLS series may not be valid, considering the differences in preoperative demographics, surgical techniques and time lapse between their inclusion periods. Nevertheless, the study has a number of strengths: (1) its long follow-up, being the first study to report results of the HLS TKA system beyond 15 years, and (2) that all surgeries were performed at one center by the same surgical team using the same techniques and instruments, which allows a better consistency between the two series, (3) the use of validated outcome assessments including the Oxford knee score that is rarely reported in long-term studies. It is also one of few studies [13, 42] to compare survival and complication rates between several generation of a TKA system. Finally, this study highlights the importance of patello-femoral geometry, which is often overlooked in TKA.
Conclusion
At equivalent follow-up of 14 years, considering revision of all components, the HLS II had better survival than the HLS Evolution (98.9% vs 95.9%), while considering revision of any component, the HLS II had lower survival than the HLS Evolution (93.0% vs 94.9%). Though the differences in survival of the two implants were not significant, conflicting findings are observed due to partial revisions for patellar fractures (5 in the HLS II series and 1 in the HLS Evolution series), which could be related to patellofemoral design enhancements.
References
Anouchi YS, Whiteside LA, Kaiser AD, Milliano MT (1993) The effects of axial rotational alignment of the femoral component on knee stability and patellar tracking in total knee arthroplasty demonstrated on autopsy specimens. Clin Orthop Relat Res (287):170–177
Argenson JN, Boisgard S, Parratte S, Descamps S, Bercovy M, Bonnevialle P, Briard JL, Brilhault J, Chouteau J, Nizard R, Saragaglia D, Servien E (2013) Survival analysis of total knee arthroplasty at a minimum 10 years’ follow-up: a multicenter French nationwide study including 846 cases. Orthop Traumatol Surg Res 99(4):385–390
Attar FG, Khaw FM, Kirk LM, Gregg PJ (2008) Survivorship analysis at 15 years of cemented press-fit condylar total knee arthroplasty. J Arthroplasty 23(3):344–349
Biau DJ, Latouche A, Porcher R (2007) Competing events influence estimated survival probability: when is Kaplan-Meier analysis appropriate? Clin Orthop Relat Res 462:229–233
Bistolfi A, Massazza G, Rosso F, Deledda D, Gaito V, Lagalla F, Olivero C, Crova M (2011) Cemented fixed-bearing PFC total knee arthroplasty: survival and failure analysis at 12–17 years. J Orthop Traumatol 12(3):131–136
Bouras T, Bitas V, Fennema P, Korovessis P (2017) Good long-term results following cementless TKA with a titanium plasma coating. Knee Surg Sports Traumatol Arthrosc 25(9):2801–2808
Brown EC 3rd, Clarke HD, Scuderi GR (2006) The painful total knee arthroplasty: diagnosis and management. Orthopedics 29(2):129–136 (quiz 137–128)
Callaghan JJ, Beckert MW, Hennessy DW, Goetz DD, Kelley SS (2013) Durability of a cruciate-retaining TKA with modular tibial trays at 20 years. Clin Orthop Relat Res 471(1):109–117
Cheng T, Zhu C, Guo Y, Shi S, Chen D, Zhang X (2014) Patellar denervation with electrocautery in total knee arthroplasty without patellar resurfacing: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 22(11):2648–2654
Dejour D, Ntagiopoulos PG, Saffarini M (2014) Evidence of trochlear dysplasia in femoral component designs. Knee Surg Sports Traumatol Arthrosc 22(11):2599–2607
Fennema P, Lubsen J (2010) Survival analysis in total joint replacement: an alternative method of accounting for the presence of competing risk. J Bone Joint Surg Br 92(5):701–706
Font-Rodriguez DE, Scuderi GR, Insall JN (1997) Survivorship of cemented total knee arthroplasty. Clin Orthop Relat Res (345):79–86
Gaillard R, Lustig S, Peltier A, Villa V, Servien E, Neyret P (2016) Total knee implant posterior stabilised by a third condyle: design evolution and post-operative complications. Orthop Traumatol Surg Res 102(8):1061–1068
Gray RJ (1988) A class of K-sample tests for comparing the cumulative incidence of a competing risk. Ann Stat 16(3):1141–1154
Halley DK, Charnley J (1975) Results of low friction arthroplasty in patients thirty years of age or younger. Clin Orthop Relat Res (112):180–191
Healy WL, Wasilewski SA, Takei R, Oberlander M (1995) Patellofemoral complications following total knee arthroplasty. Correlation with implant design and patient risk factors. J Arthroplasty 10(2):197–201
Huang CH, Hsu L-I, Chang T-K, Chuang T-Y, Shih S-L, Lu Y-C, Chen C-S, Huang C-H (2014) Stress distribution of the patellofemoral joint in the anatomic V–shape and curved dome–shape femoral component: a comparison of resurfaced and unresurfaced patellae. Knee Surg Sports Traumatol Arthrosc 25(1):263–271
Huizinga MR, Brouwer RW, Bisschop R, van der Veen HC, van den Akker-Scheek I, van Raay JJ (2012) Long-term follow-up of anatomic graduated component total knee arthroplasty: a 15- to 20-year survival analysis. J Arthroplasty 27(6):1190–1195
Karachalios T, Varitimidis S, Bargiotas K, Hantes M, Roidis N, Malizos KN (2016) An 11- to 15-year clinical outcome study of the advance medial pivot total knee arthroplasty: pivot knee arthroplasty. Bone Joint J 98-b (8):1050–1055
Keurentjes JC, Fiocco M, Schreurs BW, Pijls BG, Nouta KA, Nelissen RG (2012) Revision surgery is overestimated in hip replacement. Bone Joint Res 1(10):258–262
Lacny S, Wilson T, Clement F, Roberts DJ, Faris PD, Ghali WA, Marshall DA (2015) Kaplan–Meier survival analysis overestimates the risk of revision arthroplasty: a meta-analysis. Clin Orthop Relat Res 473(11):3431–3442
Lygre HLS, Espehaug B, Furnes O (2011) Failure of total knee arthroplasty with or without patella resurfacing A study from the Norwegian Arthroplasty Register with 0–15 years of follow-up. Acta Orthop 82:282–292
Macheras GA, Galanakos SP, Lepetsos P, Anastasopoulos PP, Papadakis SA (2017) A long term clinical outcome of the medial pivot knee arthroplasty system. Knee 24(2):447–453
Magnussen RA, Weppe F, Demey G, Servien E, Lustig S (2011) Residual varus alignment does not compromise results of TKAs in patients with preoperative varus. Clin Orthop Relat Res 469(12):3443–3450
Martin CT, Callaghan JJ, Gao Y, Pugely AJ, Liu SS, Warth LC, Goetz DD (2016) What can we learn from 20-year followup studies of hip replacement? Clin Orthop Relat Res 474(2):402–407
McCalden RW, Hart GP, MacDonald SJ, Naudie DD, Howard JH, Bourne RB (2017) Clinical results and survivorship of the GENESIS II total knee arthroplasty at a minimum of 15 years. J Arthroplasty 32(7):2161–2166
Meding JB, Fish MD, Berend ME, Ritter MA, Keating EM (2008) Predicting patellar failure after total knee arthroplasty. Clin Orthop Relat Res 466(11):2769–2774
Nakamura S, Sharma A, Nakamura K, Ikeda N, Zingde SM, Komistek RD (2014) Can post-cam function be replaced by addition of a third condyle in PS TKA? J Arthroplasty 29(9):1871–1876
Ostermeier S, Buhrmester O, Hurschler C, Stukenborg-Colsman C (2005) Dynamic in vitro measurement of patellar movement after total knee arthroplasty: an in vitro study. BMC Musculoskelet Disord 6:30
Ouanezar H, Franck F, Jacquel A, Pibarot V, Wegrzyn J (2016) Does computer-assisted surgery influence survivorship of cementless total knee arthroplasty in patients with primary osteoarthritis? A 10-year follow-up study. Knee Surg Sports Traumatol Arthrosc 24(11):3448–3456
Petersen W, Rembitzki IV, Brüggemann G-P, Ellermann A, Best R, Gösele- Koppenburg A, Liebau C (2014) Anterior knee pain after total knee arthroplasty: a narrative review. International Orthopaedics 38
Pilling RW, Moulder E, Allgar V, Messner J, Sun Z, Mohsen A (2012) Patellar resurfacing in primary total knee replacement: a meta-analysis. J Bone Joint Surg Am 94(24):2270–2278
Porcher R (2015) CORR insights (R): Kaplan–meier survival analysis overestimates the risk of revision arthroplasty: a meta-analysis. Clin Orthop Relat Res 473(11):3443–3445
Roder C, Staub LP, Eichler P, Widmer M, Dietrich D, Eggli S, Muller U (2006) Avoiding misclassification bias with the traditional Charnley classification: rationale for a fourth Charnley class BB. J Orthop Res 24(9):1803–1808
Rodriguez JA, Bhende H, Ranawat CS (2001) Total condylar knee replacement: a 20-year followup study. Clin Orthop Relat Res (388):10–17
Saffarini M, Demey G, Nover L, Dejour D (2016) Evolution of trochlear compartment geometry in total knee arthroplasty. Ann Transl Med 4(1):7
Schiavone Panni A, Falez F, D’Apolito R, Corona K, Perisano C, Vasso M (2017) Long-term follow-up of a non-randomised prospective cohort of one hundred and ninety two total knee arthroplasties using the NexGen implant. Int Orthop 41(6):1155–1162
Tayot O, Ait Si Selmi T, Neyret P (2001) Results at 11.5 years of a series of 376 posterior stabilized HLS1 total knee replacements. Survivorship analysis, and risk factors for failure. Knee 8(3):195–205
van Jonbergen H-PW, Reuver JM, Mutsaerts EL, Poolman RW (2014) Determinants of anterior knee pain following total knee replacement: a systematic review. Knee Surg Sports Traumatol Arthrosc:478–499
Varadarajan KM, Freiberg AA, Gill TJ, Rubash HE, Li G (2010) Relationship between three-dimensional geometry of the trochlear groove and in vivo patellar tracking during weight-bearing knee flexion. J Biomech Eng 132(6):061008
Varadarajan KM, Rubash HE, Li G (2011) Are current total knee arthroplasty implants designed to restore normal trochlear groove anatomy? J Arthroplasty 26(2):274–281
Victor J, Ghijselings S, Tajdar F, Van Damme G, Deprez P, Arnout N, Van Der Straeten C (2014) Total knee arthroplasty at 15–17 years: does implant design affect outcome? Int Orthop 38(2):235–241
Werth L, Saffarini M, Amsler F, Abdelkafy A, Hirschmann MT (2017) The need for secondary resurfacing is affected by trochlear height in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25(12):3818–3823
Wongworawat MD, Dobbs MB, Gebhardt MC, Gioe TJ, Leopold SS, Manner PA, Rimnac CM, Porcher R (2015) Editorial: Estimating survivorship in the face of competing risks. Clin Orthop Relat Res 473(4):1173–1176
Zaffagnini S, Bignozzi S, Saffarini M, Colle F, Sharma B, Kinov PS, Marcacci M, Dejour D (2014) Comparison of stability and kinematics of the natural knee versus a PS TKA with a ‘third condyle’. Knee Surg Sports Traumatol Arthrosc 22(8):1778–1785
Acknowledgements
The authors are grateful to Dr. Gerjon Hannink and Mr. Luca Nover for their assistance with statistical analyses.
Funding
No funding was received for this study.
Author information
Authors and Affiliations
Contributions
NJ participated in study design, data collection and manuscript preparation. CF participated in study design and data collection. JV participated in literature review, statistical analysis and manuscript writing. MS participated in statistical analysis and manuscript writing. HM participated in study design, data collection and manuscript editing. GP Participated in study design and data collection. SP participated in study design, data collection and manuscript editing. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
JV and MS received fees for support in study design and manuscript preparation from ReSurg SA (Switzerland).
Ethical approval
All patients provided informed consent to use their data for research and publishing purposes. As the study used retrospective data, collected for routine clinical assessment, the Institutional Review Board (Lille University Hospital, France) waived the standard ethical approval process.
Rights and permissions
About this article
Cite this article
Jan, N., Fontaine, C., Migaud, H. et al. Patellofemoral design enhancements reduce long-term complications of postero-stabilized total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 27, 1241–1250 (2019). https://doi.org/10.1007/s00167-018-5137-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00167-018-5137-6