Stereotactic radiosurgery using an adapted linear accelerator (linac-SRS) is an established treatment option for brain metastases, a variety of benign brain tumors and arteriovenous malformations (AVM) [57, 14, 16, 17, 25, 26]. Focused isocentric irradiation techniques and a set of 24 1.5-mm graded circular collimators as well as the use of a micromultileaf collimator (microMLC) with 1.5 mm optimally wide blades allow the generation of dose distributions with high conformity [4, 22, 28]. Here we investigated whether at least an equivalent treatment plan can be achieved with a CyberKnife (Accuray) with 12 circular collimators (graduated: 2.5–5 mm) and a non-isocentric irradiation technique [1, 2, 13, 18, 24]. Our aim was to compare the established clinical standard with clinically realizable CyberKnife plans. Therefore randomly chosen clinically realized linac-SRS plans were used in our comparison since they constitute an unbiased representation of current treatment plan quality and CyberKnife planning was performed with a uniform planning procedure and with subject to clinical constraints.

For this investigation treatment plans were compared based on the multicriteria decision theory as outlined in a metastudy on plan comparison by Phillips and Holdsworth [23]. Uniform dosimetric decision objectives for plan comparison were defined. Comparison was performed separately for each decision objective and on a multiobjective basis by scoring decision outcomes of the decision objectives for each patient and each target volume.

Material and methods

Patients treated in our clinic with intracranial linac-SRS in August 2012 (n = 16) were retrospectively selected as reference candidates. We purposely chose a period prior to the establishment of the CK technique in our hospital to avoid any bias with regard to planning strategies.

The treatment of the patients was performed as single dose irradiation after invasive head fixation in a stereotactic frame in all cases. The basis for the treatment planning was an intraoperative stereotactic CT and preoperative (frameless) magnetic resonance imaging. The irradiation planning was for circular collimators with STP3.5 (Leibinger) and for microMLC plans with Virtuoso 3.0.3 (Leibinger). Circular collimators were mainly used for small targets because of the superior accuracy in field definition and microMLC for large and complex shaped target volumes. An adapted linear accelerator SL25 (Elekta) with 6 MV photon radiation was used. Treatment time was about 45 min for each target volume. The patient’s head stayed fixed in the stereotactic frame during the whole procedure to ensure high geometrical accuracy of the treatment. This allowed to use a CTV–PTV (clinical target volume–planning target volume) margin of 0 mm. The treatment method has been described in detail previously [11, 25, 26].

All cases were re-planned with the CyberKnife planning program Multiplan 4.5.0 (Accuray) for a CyberKnife with 6 MV and without a flattening filter. Identical PTV as in linac-SRS were used with zero margins. Only the 12 fixed circular collimators were used; the Iris Collimator [10] was not used due to the broader penumbra area. Like linac-SRS, the CyberKnife was commissioned for patient use and all dose measurements were performed with comparable methods and similar equipment [2, 8, 31]. Constraints in planning were boundary and maximal dosage taken from the linac-SRS plans, as well as an irradiation time of less than 60 min per target volume. Parameters for estimating the irradiation time were 6 min for the setup time per collimator and intervals of 60 s for imaging. For the CyberKnife planning sequential optimization was selected [27] with the inverse planning objectives: (1) target coverage, (2) conformity, and (3) minimal monitor units (MU). Since collimator size was not included in sequential optimization, for each target volume several collimator settings were specified manually and then the best solution in the context of the decision objectives (see below) was selected.

The irradiation plans were compared pair-wise in terms of the decision objectives (1) dosage conformity (CI), (2) minimal dose in target volume (D min), (3) coverage (Cov), and (4) volume of brain tissue irradiated by more than 10 Gy (V 10). Coverage was defined by \( Cov = {{{V_{{\textrm{T,}}TD}}}\mathord{\left/ {\vphantom {{{V_{{\textrm{T,}}TD}}}{{V_{\textrm{T}}}}}}\right. \kern-\nulldelimiterspace} {{V_{\textrm{T}}}}} \), where V T represents the target volume, and V T, TD the proportion of the target volume irradiated with a dose of TD or more. The conformity index is defined by \( CI = {{V_{{\textrm{T,}}TD}^2} \mathord{\left/ {\vphantom {{V_{{\textrm{T,}}TD}^2} {({V_{TD}}}}}\right. \kern-\nulldelimiterspace} {({V_{TD}}}}\cdot {V_{\textrm{T}}}) \), where V TD represents the size of the volume irradiated with a dose of TD or more [21, 34]. In a generalization of these definitions, for patients with more than one target volume the total coverage Cov total of all n T target volumes is calculated by \( Co{v_{{\textrm{total}}}}= {{\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}TD{\textrm{,}}i}}}}\mathord{\left/ {\vphantom {{\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}TD{\textrm{,}}i}}}}{\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}i}}}}}}\right. \kern-\nulldelimiterspace} {\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}i}}}}} \), where V T,i stands for the volume of the i-th target and V T,TD,i for the proportion of the i-th target volume irradiated with a dose of TD or more. Correspondingly the total conformity index CI total is calculated according to \( C{I_{{\textrm{total}}}}= {{{{\left({\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}TD,i}}}}\right)}^2}}\mathord{\left/ {\vphantom {{{{\left({\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{{\textrm{T,}}TD,i}}}}\right)}^2}}{\left({\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{TD,i}}}\cdot \sum\nolimits_{i = 1}^{{n_T}}{{V_{{\textrm{T}},i}}}}\right)}}}\right. \kern-\nulldelimiterspace} {\left({\sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{TD,i}}}\cdot \sum\nolimits_{i = 1}^{{n_T}}{{V_{{\textrm{T}},i}}}}\right)}} \). Furthermore, the minimal dose D min, total in the case of several target volumes is calculated by \( {D_{{\textrm{min,total}}}}= \mathop {\min }_{i = 1,{n_{\textrm{T}}}}\left\{{{D_{\min,i}}}\right\} \) and the total volume V 10,total of brain tissue irradiated by more than 10 Gy by \( {V_{10,{\textrm{total}}}}= \sum\nolimits_{i = 1}^{{n_{\textrm{T}}}}{{V_{10,i}}} \).

Plan ranking was achieved by multiobjective comparison of the decision objectives [23]. The values of the decision objectives for each target volume were compared pair-wise. The ranges within the decision objectives seen as equivalent were as follows: ± 0.5 % for Cov, ± 0.5 Gy for D min, ± 0.03 for CI, and ± 0.5 ml for V 10. Larger deviations, each according to their signs, were classified as better or worse and the outcomes were scored for a target volume or patient for all decision objectives.

Differences in the decision objectives of CyberKnife and linac-SRS plans were tested for significance with the Wilcoxon signed rank test on paired samples. The software IBM SPSS Statistics V20 was used.

Results

A total of 16 patients and 23 target volumes were treated with linac-SRS. Of these patients, 8 had one or more brain metastases, 3 an acoustic neuroma, 3 a pituitary adenoma, 1 a meningioma, and 1 AVM. Circular collimators were used in 9 patients and 15 target volumes and the microMLC in 7 patients and 8 target volumes. The median target volume size (V T) was 1.99 ml (mean 2.44 ml, range 0.01–7.9 ml). The treatment dose (TD) had a median of 18 Gy (mean 17.5 Gy, range 12–25 Gy) and the median treatment isodose was 64.9 % (mean 66.1 %, range 58.7–80.0 %).

In all cases generating focal and conformal dose distributions with CyberKnife was no problem. For all 16 patients and for 20 out of 23 target volumes the CyberKnife plans were superior to the linac-SRS plans. In 4 out of 23 target volumes, the CyberKnife plan was better for all four decision objectives, in 8 cases for three decision objectives, in 6 cases for two decision objectives, and in 2 cases for one decision objective. For three target volumes, the plans were equivalent. In 3 out of 16 patients the CyberKnife plan was better for all four decision objectives, in 8 patients for three decision objectives, in 3 patients for two decision objectives, and in 2 patients for one decision objective. These results were robust against changes in the range limits, within which two decision objectives were classified as equivalent.

In linac-SRS, the mean coverage of the target volumes was 98.9 % (range 93.3–100 %), the mean minimal dose was 15.3 Gy (range 8.3–27.1 Gy), the mean conformity index was 0.57 (range 0.05–0.81), and the mean V 10 was 4.7 ml (range 0.3–10.3 ml). In the CyberKnife plans the mean coverage of the target volumes was 99.6 % (range 98.8–100 %), the mean minimal dose was 16.6 Gy (range 10.8–25.1 Gy), the mean conformity index was 0.72 (range 0.32–0.91), and the mean V10 was 4.5 ml (range 0.3–10.0 ml). With the CyberKnife plans the coverage was significantly higher (p = 0.006) by 0.7 % (mean, range − 0.5  - +6.5 %), the minimal dose was significantly higher (p = 0.001) by 1.3 Gy (mean, range − 6.7  - +4.4 Gy), the conformity index was significantly higher (p < 0.001) by 0.15 (mean, range − 0.03  - +2.7), and V 10 was less (p = 0.157) by 0.2 ml (mean, range − 2.2  - +2.0 ml). The expected treatment duration with CyberKnife was on average 40 min/target volume (range 20–59 min). Similar values were obtained for patient-related decision objectives. On average Cov total was significantly higher (p = 0.003) by 1.0 % (mean, range − 0.5  - +6.5 %), D min,total was significantly higher (p = 0.001) by 1.9 Gy (mean, range − 0.0  - +4.4 Gy), CI total was significantly higher (p < 0.001) by 0.14 (mean, range + 0.06  - +0.28), and V 10,total was less (p = 0.155) by 0.3 ml (mean, range − 2.2  - +1.4 ml). The expected treatment duration with CyberKnife was on average 58 min/patient (range 20–115 min/patient). In comparison, the treatment duration with linac-SRS was on average 83 min/patient (range 30–205 min/patient).

Distributions of the decision objectives of individual target volumes (Fig. 1) as well as individual patients (Fig. 2) were directly compared between the SRS plans and CyberKnife plans.

Fig. 1
figure 1

Distribution of the decision objectives of the linac-SRS plans (SRS) and CyberKnife plans (CK) for individual target volumes (n = 23) represented as box plots

Fig. 2
figure 2

Distribution of the decision objectives for the linac-SRS plans (SRS, horizontal axis) and CyberKnife plans (CK, vertical axis) for individual patients (n = 16) represented as scatter plots. The areas within which two values were classified as equivalent are indicated by dashed lines

Discussion

Application of high single doses of typically 12–20 Gy or more in intracranial stereotactic radiosurgery poses the highest demands on both the quality of the irradiation plan and the target accuracy of the dose application [30, 32]. The limit for the maximal target point deviation lies at 1 mm [33], a challenge that can be met with a CyberKnife [2]. The aim of this study was to investigate whether the plan quality with a CyberKnife can also be at least equivalent to that achieved with established linac-SRS with circular collimators and with microMLC.

Several studies are already published on dosimetric comparison of linac-SRS and CyberKnife [3, 9, 15, 20, 29, 36]. However, none of these studies were suitable to answer the questions we posed here. Either the studies concentrated on purely geometric investigations with an elliptic target volume [36] or focused on special indications such as acoustic neuroma [9, 15] and AVMs [3, 15]. Alternatively they investigated very specific questions such as volume load with irradiation of multiple metastases [20]. Also the aim of one study was not “to do a strict dosimetric study” but “to compare actual treatment plans from institutes” in the Netherlands [29].

Some of the studies based dosimetric comparison on one individual case [20, 36]. The study by Dutta et al. [9] only investigated CyberKnife plans with multiple isocenters and the study by Gevaert et al. [15] with CyberKnife exclusively used the Iris-collimator instead of circular collimators. Thus, in both cases the dosimetric possibilities of CyberKnife were not adequately exploited. Finally, several of the studies did not meet the quality standards for dosimetric comparison studies defined by Phillips and Holdsworth in a meta-analysis [23]. For example, the comparison of decision objectives sometimes used the t-test [3, 9]. However the t-test assumes normal distribution and the Wilcoxon signed rank test would probably have been more appropriate for the data [23].

As a basis for the comparative study presented here, we selected a random sample from 16 patients and 23 target volumes, which had been treated in our clinic with state-of-the-art linac-SRS procedures. This sample was thought to represent typical clinical cases encountered in intracranial radiosurgery. The methodical procedure in our comparative study was carried out according to the multi-criteria decision theory [23].

The definition of the decision objectives was as follows. According to a guideline of the Radiation Therapy Oncology Group (RTOG) for radiosurgery the parameters coverage, homogeneity index, and conformity index should be used to evaluate the irradiation plan [32]. In radiosurgery today, depending on kind and location of the target volume, it is common practice to use the homogeneity of the dose distribution as part of the dose prescription in treatment planning. We therefore used the homogeneity index of the linac-SRS plans as a constraint for the CyberKnife plans.

The term coverage is not uniformly used in the literature and is sometimes dosimetrically defined as the ratio of the minimal dose in the target volume to the prescribed dose, but often also volumetrically defined as the ratio of the proportion of the target volume covered by the prescribed isodose to the total volume of the target [12, 19]. Due to the steep dose gradient in radiosurgery, both aspects of coverage have independent meanings [19]. To take this into account, we have not only chosen the volumetrically defined coverage Cov but also the minimal dose D min in the target volume as decision objectives. We counteract these two parameters that characterize coverage with two similarly complementary parameters that characterize conformity, i.e., the conformity index CI and the volume V 10. Risk analyses of stereotactic radiosurgery have shown that the volume of healthy brain tissue that is irradiated with a single dose of more than 10 Gy must be regarded as an important parameter [35].

Our results show that for the multiobjective comparison of the plans, the CyberKnife is significantly superior to the classic linac-SRS with circular collimators or microMLC. The larger graduation of the CyberKnife collimators of ≥ 2.5 mm, compared to those according to Bortfeld et al. [4] with “optimal” graduation of 1.5 mm for 6 MV photon irradiation with the linac-SRS, is apparently more than compensated by the possibility of nonisocentric beam steering with the CyberKnife and perhaps by the slightly improved lateral dose distributions due to the flattening filter free design of the CyberKnife. Cov, D min, and CI for all patients and target volumes were in general clearly significantly better with the CyberKnife plans, with at the same time moderate improvement in V 10. In particular, the mean higher minimal dose of 1.9 Gy per patient in the target volume, with concurrently a better conformity index of 14 % in the target volume, appears to be an improvement with clear clinical relevance.

In all cases generating the CyberKnife plans was unproblematic using simple and uniform optimization scripts. The question remains as to what extent the generated CyberKnife plans are already optimal or whether there is even more room for improvement. Compared to linac-SRS there is an increased time window for treatment planning with the CyberKnife as frameless and image-guided treatments can even be performed on an outpatient basis. With frame-based linac-SRS, the possibility to generate optimal plans is definitely limited simply due to the time pressure in treatment planning during the treatment of patients fixed in a stereotactic frame. Taken together, irradiation planning for radiosurgery with CyberKnife seems to be clearly superior to the classic linac-SRS.