Abstract
End-tidal carbon dioxide (ETCO2) values obtained from awake nonintubated patients may prove to be useful in estimating a patient’s ventilatory status. This study examined the relationship between arterial carbon dioxide tension (PaCO2) and ETCO2 during the preoperative period in 20 premedicated patients undergoing various surgical procedures. ETCO2 was sampled from a 16-gauge intravenous catheter pierced through one of the two nasal oxygen prongs and measured at various oxygen flow rates (2, 4, and 6 L/min) by an on-line ETCO2 monitor with analog display. Both peak and time-averaged values for ETCO2 were recorded. The results showed that the peak ETCO2 values (mean = 38.8 mm Hg) correlated more closely with the PaCO2 values (mean = 38.8 mm Hg; correlation coefficient r = 0.76) than did the average ETCO2 values irrespective of the oxygen flow rates. The time-averaged PaCO2-ETCO2 difference was significantly greater than the PaCO2-peak ETCO2 difference (P < 0.001). Values for subgroups within the patient population were also analyzed, and it was shown that patients with minute respiratory rates greater than 20 but less than 30 and patients age 65 years or older did not differ from the overall studied patient population with regard to PaCO2-ETCO2 difference. A small subset of patients with respiratory rates of 30/ min or greater (n = 30) did show a significant increase in the PaCO2-ETCO2 difference (P < 0.001). It was concluded that under the conditions of this study, peak ETCO2 values did correlate with PaCO2 values and were not significantly affected by oxygen flow rate. However, obtaining peak ETCO2 values is clinically more difficult, especially when partial air-way obstruction is present.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Shankar KB, Moscley H, Kumar Y, Vemula V. Arterial to end tidal carbon dioxide tension difference during caesarean section anesthesia. Anaesthesia 1986;41:698–702
Badgwell JM, McLeod ME, Lerman J, Creighton RE. End-tidal PCO2 measurements sampled at the distal and proximal ends of the endotracheal tube in infants and children. Anesth Analg 1987;66:959–964
McEvedy BAB, McLeod ME, Mulera M, et al. End-tidal, transcutaneous, and arterial PCO2 measurements in critically ill neonates: a comparative study. Anesthesiology 1988;69:112–116
Ibarra E, Lees DE. Mass spectrometer monitoring of patients with regional anesthesia. Anesthesiology 1985;63:572–573
Norman EA, Zeig NJ, Ahmad I. Better designs for mass spectrometer monitoring of the awake patient. Anesthesiology 1986;64:664
Huntington CT, King HK. A simpler design for mass spectrometer monitoring of the awake patient. Anesthesiology 1986;65:565–566
Pressman MA. A simple method of measuring ETCO2 during MAC and major regional anesthesia. Anesth Analg 1988;67:905–906
Bowe EA, Boysen PG, Broome JA, Klein EF. Accurate determination of end-tidal carbon dioxide during administration of oxygen by nasal cannula. J Clin Monit 1989;5:105–110
Goldman JM. A simple, easy, and inexpensive method for monitoring ETCO2 through the nasal cannula. Anesthesiology 1987;67:606
Louwsma DL, Silverman DG. Reproducibility of end tidal CO2 measurements in sedated patients receiving supplemental O2 by nasal cannula (abstract). Anesthesiology 1988;69:A268
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307–310
Fordyce WE, Kanter RK. Arterial-end tidal PCO2 equilibration in the cat during acute hypercapnia. Respir Physiol 1988;73:257–272
Piiper J. Blood-gas equilibrium of carbon dioxide in lungs; a continuing controversy. J Appl Physiol 1986;60:l-8
Severinghaus JW. Water vapor calibration errors in some capnometers: respiratory conventions misunderstood by manufacturers? Anesthesiology 1989;70:996–998
Algora-Weber A, Rubio JJ, Dominguez de Villota E, et al. Simple and accurate monitoring of end-tidal carbon dioxide tensions during high-frequency jet ventilation. Crit Care Med 1986; 14:895–897
Fletcher R. Smoking, age and the artificial-end-tidal PCO2 difference during anaesthesia and controlled ventilation. Acta Anaesthesiol Scand 1987;31:355–356
Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the differences. Anesth Analg 1981;60:508–512
Dunphey JA. Accuracy of expired carbon dioxide partial pressure sampled from nasal cannula. II. Anesthesiology 1988;68:960–961
From RP, Scamman FL. Ventilatory frequency influences accuracy of end-tidal CO2 measurements: analysis of seven capnometers. Anesth Analg 1988;67:884–886
Zupan J, Martin M, Bcnumof JL. End-tidal CO2 excretion waveform and error with gas sampling line leak. Anesth Analg 1988;67:579–581
Koebert RF, Munster R. One-way leak in mass spectrometer sampling system. Anesthesiology 1987;67:606–607
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
McNulty, S.E., Roy, J., Torjman, M. et al. Relationship between arterial carbon dioxide and end-tidal carbon dioxide when a nasal sampling port is used. J Clin Monitor Comput 6, 93–98 (1990). https://doi.org/10.1007/BF02828284
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02828284