Abstract
Measurement of arterial oxygen saturation by pulse oximetry was performed in two patients with acute and chronic anaemia (haemoglobin concentrations: 2.9 mmol/l (4.7 g/dl) and 1.9 mmol/l (3.0 g/dl), respectively) using a Radiometer OXI and a Nellcor N-200 pulse oximeter. The two oximeters read alternating different values in the two patients. In conclusion, pulse oximeters are able to give a value for oxygen saturation even at extreme anaemia, and when a high value is given, it possibly reflects arterial oxygen saturation. The value of pulse oximetry in severe anaemia is discussed.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Tremper KK, Barker SJ (1989) Pulse oximetry. Anesthesiology 70:98–108
Kelleher JF (1989) Pulse oximetry. J Clin Monit 5:37–62
Lee SE, Tremper KK, Barker SJ (1988) Effects of anemia on pulse oximetry and continuous mixed venous oxygen saturation monitoring in dogs. Anesth Analg 67:S130
Severinghaus JW, Koh SO (1990) Effect of anemia on pulse oximeter accuracy at low saturation. J Clin Monit 6:85–88
Kolesar R, Volgyesi G, Lerman J (1990) Effect of haemoglobin concentration on the accuracy of pulse oximetry. Can J Anaesth 37[Suppl]:S88
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Ramsing, T., Rosenberg, J. Pulse oximetry in severe anaemia. Intensive Care Med 18, 125–126 (1992). https://doi.org/10.1007/BF01705047