Abstract
Hickman catheters were the major venous access devices utilized at the University of Maryland Cancer Center from November 1978 to 1987. This study provided an opportunity to standardize insertion technique, to manage catheter-related activities and daily maintenance procedures in order to examine the progression of Hickman-catheter-related problems, to identify those factors that may minimize them, and to develop guidelines for the management and prevention of complications and malfunctions. In all, 690 Hickman catheters (368 double lumens) were placed in patients with acute leukemia and other cancers: 401 catheters were placed in patients with leukemia; 269 were placed during neutropenia; and 230 at platelet counts of <50 000/μl. Two surgeons inserted 490 catheters, and the remaining 200 were placed by a group of rotating surgeons. All catheters were placed with the intention that they would remain in place as long as clinically necessary. Total Hickman catheter days were 134273. Infectious complications included exit site infections (160), tunnel infections (46) and bacteremias (397). There were 438 instances of noninfectious complications including thrombosis, lack of function, catheter migration, fracture and hemorrhage. Recommendations for prevention and treatment of Hickman-catheter-related complications include the development of a select group committed to placement, daily maintenance and management of problems; prompt removal of catheters with Candida sp. fungemia and bacteremia due to Bacillus sp. or a bacteremia that persists for >48 h after initiation of appropriate antibiotics, tunnel infections or Hickman-catheter-associated thrombosis. The majority of bacteremias and exit site infections can be effectively treated with antibiotics and local care.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Dugdale DC, Ramsey PC (1990) Staphylococcus aureus bacteremia in patients with Hickman catheters. Am J Med 89:137–141
Hickman RO, Buckner CD, Clift RA, Sanders JE, Stewart P, Thomas ED (1979) A modified right atrial catheter for access to the venous system in marrow transplant recipients. Surg Gynecol Obstet 148:871–875
Khabbaz RF, Wade JC (1992) The alimentary tract as a source of S. epidermidis bacteremia in patients with cancer: clues from molecular epidemiology. J Infect Dis (in press)
Lowder JN, Lazarus HM, Herzig RH (1982) Bacteremias and fungemias in oncologic patients with central venous catheters: changing spectrum of infection. Arch Intern Med 142:1456–1459
Maki DG (1991) Infection caused by intravascular devices: pathogenesis, strategies for prevention. In: Improving catheter site care. (International congress and symposium series 179) Royal Society of Medicine Services, London, pp 3–28
Marrie TJ, Costerton JW (1984) Scanning and transmission electron microscopy of in situ bacterial colonization or intravenous and intraarterial catheters. J Clin Microbiol 19:687–693
Newman KA, Reed WP, Bustamante CI, Schimpff SC, Wade JC (1989) Venous access devices utilized in association with intensive cancer chemotherapy. Eur J Cancer Clin Oncol 24:1–4
Press OW, Ramsey PG, Larson EB, Fefer A, Hickman RO (1984) Hickman catheter infections in patients with malignancies. Medicine 63:189–200
Reed WP, Newman KA (1983) An improved technique for the insertion of Hickman catheters in patients with thrombocytopenia and granulocytopenia. Surg Gynecol Obstet 156:355–358
Tenney JH, Moody MR, Newman KA, Schimpff SC, Wade JC, Costerton JW, Reed WP (1986) Clinical significance of microorganism adherent to the surface of the lumen of silicone intravenous catheters removed from cancer patients. Arch Intern Med 146:1949–1954
Schimpff SC (1990) Infection in the leukemia patient: diagnosis, therapy and prevention. In: Henderson ES, Lister TA (eds) Leukemia, 5th edn. Saunders, Philadelphia, p 702
Wade JC, Newman KA, Schimpff SC, VanEcho DA, Gelber RA, Reed WP, Wiernik PH (1981) Two methods for improved venous access in acute leukemia patients. JAMA 140–144
Wade JC, Schimpff SC, Newman KA, Wiernik PH (1982) Staphylococcus epidermidis: an increasing cause of infection in granulocytopenic patients. Ann Intern Med 97:503–508