Abstract
The accuracy of LDA measurements depends on the optical alignment of the laser beams. Improperly designed optical systems lead to fringe distortion in the measurement volume and in earlier investigations this effect has always been taken as the main cause of optical inaccuracy in LDA measurements. In the present work a different cause of fringe distortion is considered: astigmatism due to beam refractions. A quantitative theory for the astigmatism of laser beams is derived for both single and multiple refractions. Parameter calculations with regard to the size of the astigmatism effect have been carried out. It is shown that astigmatism is a relevant parameter which influences the fringe uniformity and fringe distortion in an LDA measurement volume and affects the measurement accuracy of measurements in internal flow. The equations derived enable the change in cross sections of the refracted laser beams to be determined. The spatial deviations of the diverse focusing points of refracted laser beams relative to the position of the LDA measurement volume are found to depend strongly on the incident angle of the beams and therefore on the off-axis alignment angle of the LDA probe (off-axis from the normal to the flow-wall-interface).
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Durst F; Stevenson WH (1975) Moiré patterns to visually model laser-Doppler signals. The Accuracy of Flow Measurements by Laser Doppler Methods. Proceeding of the LDA-Symposium Copenhagen, 183–205
Ghatak AK; Thyagarajan K (1978) Contemporary Optics. Plenum Press. New York and London
Hanson S (1975) Visualization of alignment errors and heterodyning constraints in laser Doppler velocimeters. The Accuracy of Flow Measurements by Laser Doppler Methods. Proceeding of the LDA-Symposium Copenhagen, 176–182
Miles PC; Witze PO (1994) Fringe field quantification in an LDA probe volume by use of a magnified image. Exp Fluids 16: 330–335
Ruck B (1991) Distortion of LDA fringe pattern by tracer particles. Exp Fluids 10: 349–354
Zhang Z (1993) Theoretische Untersuchungen der Einsatzmöglichkeit für 3-dimensionale Laser-Doppler-Messungen in Kanalströmungen. Sulzer Innotec Bericht. Nr.: STT. TB93.022, Winterthur, Schweiz
Zhang Z (1995) Einfluß des Astigmatismus auf Laser Doppler Messungen. Sulzer Innotec Bericht. Nr.: STT. TB95.022, Winterthur, Schweiz
Zhang Z; Eisele K (1995a) Off-axis alignment of an LDA-probe and the effect of astigmatism on the measurements. Exp Fluids 19: 89–94
Zhang Z; Eisele K (1995b) Einfluß des Astigmatismus auf LDA- Messungen. Gala-Tagung Rostock, Deutschland
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Zhang, Z., Eisele, K. The effect of astigmatism due to beam refractions on the formation of the measurement volume in LDA measurements. Experiments in Fluids 20, 466–471 (1996). https://doi.org/10.1007/BF00189387
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00189387