Abstract
Modern computerized planning tools for periacetabular osteotomy (PAO) use either morphology-based or biomechanics-based methods. The latter rely on estimation of peak contact pressures and contact areas using either patient-specific or constant thickness cartilage models. We performed a finite element analysis investigating the optimal reorientation of the acetabulum in PAO surgery based on simulated joint contact pressures and contact areas using patient-specific cartilage model. Furthermore we investigated the influences of using patient-specific cartilage model or constant thickness cartilage model on the biomechanical simulation results. Ten specimens with hip dysplasia were used in this study. Image data were available from CT arthrography studies. Bone models were reconstructed. Mesh models for the patient-specific cartilage were defined and subsequently loaded under previously reported boundary and loading conditions. Peak contact pressures and contact areas were estimated in the original position. Afterward we used validated preoperative planning software to change the acetabular inclination by an increment of 5° and measured the lateral center-edge angle (LCE) at each reorientation position. The position with the largest contact area and the lowest peak contact pressure was defined as the optimal position. In order to investigate the influence of using patient-specific cartilage model or constant thickness cartilage model on the biomechanical simulation results, the same procedure was repeated with the same bone models but with a cartilage mesh of constant thickness. Comparison of the peak contact pressures and the contact areas between these two different cartilage models showed that good correlation between these two cartilage models for peak contact pressures (r = 0.634 ∈[0.6, 0.8], p < 0.001) and contact areas (r = 0.872 > 0.8, p < 0.001). For both cartilage models, the largest contact areas and the lowest peak pressures were found at the same position. Our study is the first study comparing peak contact pressures and contact areas between patient-specific and constant thickness cartilage models during PAO planning. Good correlation for these two models was detected. Computer-assisted planning with FE modeling using constant thickness cartilage models might be a promising PAO planning tool when a conventional CT is available.
Access provided by CONRICYT-eBooks. Download chapter PDF
Similar content being viewed by others
Keywords
- Hip dysplasia
- Periacetabular osteotomy (PAO)
- Planning
- Biomechanical simulation
- Finite element analysis (FEA)
- Image-guided surgery
- Joint preservation surgery
13.1 Introduction
Periacetabular osteotomy (PAO) is an established surgical intervention for treatment of hip dysplasia and acetabular retroversion [1, 2]. During the procedure, the acetabulum is reoriented in order to optimize the containment of the femoral head and the pressure distribution between acetabulum and femoral head for reduction of the peak contact pressures within the joint. The goal of acetabular reorientation is to restore or to approximate normal acetabular geometry. In order to achieve this, two types of planning strategies have been reported, which can be divided into morphology-based planning methods and biomechanics-based planning methods. Morphology-based planning uses standard geometric parameters, which have shown their importance for quantification of acetabular under- or overcoverage [3]. Several authors have described different morphology-based planning methods which range from simplified two-dimensional planning [4,5,6] to complex three-dimensional planning applications [7,8,9,10,11]. Other authors presented biomechanics-based planning methods. Different approaches have been presented using, for example, discrete element analysis (DEA) [12] or the more sophisticated finite element analysis (FEA) [13, 14]. In literature, both constant thickness cartilage models [14] and patient-specific cartilage models [15] have been suggested. In the clinical routine, knowledge of patient specific cartilage is rarely available, since special imaging protocol (e.g., CT arthrography or MRI with dGEMRIC, T1rho or T2 mapping) is necessary to retrieve this information. One alternative could be constant thickness cartilage model that is virtually generated from bony surface models derived from conventional CT scans. However differences between these two different cartilage models in planning of PAO using FE simulation have never been investigated. Previously, we have developed a morphology-based 3D planning system for PAO [16]. This system allows for quantification of the hip joint morphology in three dimensions, using geometric parameters such as inclination and anteversion angle, the lateral center-edge (LCE) angle, and femoral head coverage. It also allows for virtual reorientation of the acetabulum according to these parameters. In the current study, we enhanced this application with an additional biomechanics-based method for estimation of joint contact pressures employing FEA. In this study, we investigated the following research questions:
-
1.
What is the optimal position of the acetabulum based on simulated joint contact pressures using patient specific cartilage models in a FE analysis?
Are there significant differences in joint contact pressures between patient specific cartilage model and constant thickness cartilage model in the same hip model?
13.2 Materials and Methods
13.2.1 System Overview
The computer-assisted planning system for PAO uses 3D surface models of the pelvis and femur, generated out of DICOM (digital imaging and communication in medicine) data, using a commercially available segmentation program (AMIRA, Visualization Sciences Group, Burlington, MA). The system starts with a morphology-based method. Employing fully automated detection of the acetabular rim, parameters such as acetabular version, inclination, LCE angle, femoral head extrusion index (EI), and femoral head coverage can be calculated for a computer-assisted diagnosis [16]. Afterward, the system offers the possibility to perform a virtual osteotomy (Fig. 13.1a(1)) and reorientation of the acetabular fragment in a stepwise pattern. During the fragment reorientation, acetabular morphological parameters are recomputed in real time (Fig. 13.1a(2)) until the desired position is achieved. Our system is further equipped with a biomechanics-based FE prediction of changes of cartilage contact stresses, which occurs during acetabular reorientation. An optimal position of the acetabulum can be defined, once contact areas in the articulation are maximized, while at the same time peak contact pressures are minimized (Fig. 13.1b). The respective cartilage model for the biomechanics-based FE prediction is generated from either CT arthrography data (patient-specific) or using a virtually generated cartilage with predefined thickness (constant thickness).
13.2.2 Biomechanical Model of Hip Joint
13.2.2.1 Cartilage Models
In literature, both constant thickness cartilage models and patient-specific cartilage models have been employed. Zou et al. [14] used a constant thickness model and thus created a cartilage with a predefined thickness of 1.8 mm, a value derived from cartilage thickness data from the literature. In contrast Harris et al. [15] introduced a CT arthrography protocol allowing for excellent visualization of patient-specific cartilage. DICOM data of dysplastic hip joints, which have been CT scanned using this arthrography protocol, were provided by the open source dysplastic hips image data from the Musculoskeletal Research Laboratories, University of Utah [17]. The data provider has obtained IRB approval (University of Utah IRB #10983).We used our morphology-based planning system for calculation of the acetabular morphological parameters [18], verifying true dysplasia (Table 13.1). We used these datasets in order to retrieve the patient-specific cartilage models. The bony anatomy of the same ten specimens was then used to create the constant thickness cartilage models by expanding a constant 1.8 mm thickness using 3D dilation operation on the articular surface.
13.2.2.2 Mesh Generation
Bone and cartilage surface models of the reoriented hip joints were imported into ScanIP software (Simpleware Ltd., Exeter, UK) as shown in Fig. 13.2a, c. Surfaces were discretized using tetrahedral elements (Fig. 13.2b, d). Since the primary focus was the joint contact stresses, a finer mesh was employed for the cartilage than for the bone. Refined tetrahedral meshes were constructed for the cartilage models (∼135,369 elements for the femoral cartilage model and ∼92,791 elements for the acetabular cartilage model, using the ScanFE module (Simpleware Ltd., Exeter, UK). Cortical bone surfaces were discretized using coarse tetrahedral elements (∼149,120 elements for the femoral model and ∼188,526 elements for the pelvic model). The trabecular bone was not included in the models, as it only has a minor effect on the predictions of contact pressure as reported in another study [19].
13.2.2.3 Material Property
Acetabular and femoral cartilage were modeled as homogeneous, isotropic, and linearly elastic material with Young’s modulus E = 15 MPa and Poisson’s ratio ν = 0.45 [14]. The cortical bone of the pelvis and femur were modeled as homogeneous, isotropic material with elastic modulus E = 17 GPa and Poisson’s ratio ν = 0.3 [14].
13.2.2.4 Boundary Conditions and Loading
Tied and sliding contact constraints were used in Abaqus/CAE 6.10 (Dassault Systèmes Simulia Corp, Providence, RI, USA) to define the cartilage-to-bone and cartilage-to-cartilage interfaces, respectively. It has been reported that the friction coefficient between articular cartilage surfaces was very low (0.01–0.02) in the presence of synovial fluid, making it reasonable to neglect eventual frictional shear stresses [15, 20]. The top surface of pelvis and pubic areas were fixed, and the distal end of the femur was constrained to prevent displacement in the body x and y directions while being free in vertical z direction (Fig. 13.2e). The center of the femoral head was derived from a least-squares sphere fitting and was selected to be the reference node. The nodes of femoral head surface were constrained by the reference node via kinematic coupling. The fixed boundary condition model was then subjected to a loading condition as published before [21], representing a single leg stance situation with the resultant hip joint contact force acting at the reference node. Following the loading specifications suggested in another previous study [22](Fig. 13.2e), the components of joint contact force along three axes were given as 195 N, 92 N, and 1490 N, respectively. In order to remove any scaling effect of body weight on the absolute value of the contact pressure, we defined a constant body weight of 650 N for all subjects. The resultant force was applied, based on anatomical coordinate system described by Bergmann et al. [21], whose local coordinate system was defined with the x axis running between the centers of the femoral heads (positive running from the left femoral head to the right femoral head), the y axis pointing directly anteriorly, and the z axis pointing directly superiorly.
13.2.2.5 Study 1: FE Simulation for Biomechanics-Based Planning of PAO Using Patient-Specific Cartilage Model
In order to find the optimal acetabular position, the acetabular fragment was now virtually rotated around the y axis (Fig. 13.2e) in 5° increments in relation to the anterior pelvic plane (APP). This deemed to imitate a decrease in acetabular inclination, as performed during actual PAO surgery (Fig. 13.2c). For each increment, the predicted peak contact pressure and total contact area were directly extracted from the output of Abaqus/CAE 6.10. The resulting peak contact pressures and contact areas in the different acetabular positions were then compared and the corresponding LCE angle was measured. Optimal orientation was determined by the position yielding the maximum contact area and the minimum peak contact pressure.
13.2.2.6 Study 2: Evaluation the Influences of Using Different Cartilage Models on the Simulation Results
After the peak pressures and contact areas had been simulated using the patient-specific cartilage models, the same procedure was performed using the constant thickness cartilage models. Finally, comparison between peak pressures and contact areas between patient-specific and constant thickness cartilage models was performed. Linear regression analysis was used to determine associations between the results for peak pressures and contact areas for both cartilage types. Thus, the values for the constant thickness models were the independent variables, whereas the values obtained by the patient-specific models represented the dependent variables. Pearson’s correlation coefficient r was interpreted as “poor” below 0.3, “fair” from 0.3 to 0.5, “moderate” from 0.5 to 0.6, “moderately strong” from 0.6 to 0.8, and “very strong” from 0.8 to 1.0. Significance level was defined as p < 0.05.
13.3 Results
While the initial contact area in the dysplastic hip was primarily located in an eccentric superolateral region of the acetabulum, an increase in LCE angle led to an enlarged and more homogeneously distributed contact area (Fig. 13.3). At the same time, an increase in LCE angle resulted in decreased peak contact pressures. For each specimen, the optimal acetabular fragment reposition was defined as the position with minimum peak contact pressure and maximum contact area (Table 13.2).
Comparison of the peak contact pressures and the contact areas between the two different cartilage models showed similar results (Table 13.3). Regression analysis quantitatively showed that the results obtained by the constant thickness cartilage models have good correlation with those obtained by using the patient-specific cartilage models. Specifically, a moderately strong correlation was found between both cartilage models when analyzing peak contact pressures (r = 0.634 ∈ [0.6, 0.8], p < 0.001) (Fig. 13.4), while a very strong correlation was also found when analyzing the contact areas between the two different cartilage models (r = 0.872 > 0.8, p < 0.001) (Fig. 13.4b). For both cartilage models, the largest contact areas and the lowest peak pressures were found at the same position (Table 13.3).
13.4 Discussion
We used a previously validated morphology-based PAO planning system [16] to perform virtual acetabular reorientation. An additional biomechanics-based module then estimated contact areas and peak contact pressures within the joint. First we used hip joint models with patient-specific cartilage models and changed the LCE angle in order to increase femoral head containment and to find the optimal position with the largest contact area and lowest peak contact pressure. The same operation was then conducted with the bone models of the same hip joints by replacing the patient-specific cartilage models with virtually generated constant thickness cartilage models. In the patient-specific cartilage models, an increase in LCE angle led to enlarged and more homogeneously distributed contact areas and decreased peak contact pressures. Comparison of the peak contact pressures and the contact areas between the two different cartilage models showed similar results. Regression analysis quantitatively showed moderately strong correlation between both models for peak contact pressures while very strong correlation for contact areas.
In the light of our findings, several aspects need to be discussed. We did not include the acetabular labrum in our FE analysis; however the role of the labrum during load distribution is debatable in literature. While some authors promoted inclusion of the labrum [23], other authors denied the importance of its inclusion [24]. More interestingly, Henak et al. [17] showed that the labrum has a far more significant role in dysplastic hip joints biomechanics than it does in normal hips, since it supports a large percentage of the load transferred across the joint due to the eccentric loading in dysplastic hips. The same study group in a previous study [25], however, found that the labrum only supported less than 3% of the total load across the joint in normal hips. The final goal of our study was not to measure peak contact pressures and contact areas in the originally dysplastic state of our specimen but to find an optimal position resembling a “normal” hip joint during PAO. Hence, for this purpose disregarding the labrum was acceptable.
Regarding loading conditions, a fixed body weight of 650 N [21] was used, which is not patient-specific. However, Zou et al. [14] justified the use of constant loading, since the relative change of contact pressure before and after PAO reorientation planning is assessed, regardless of the true patient weight. Also, the applied loading conditions were derived from in vivo data from patients who underwent total hip arthroplasty (THA) [21] and thus might be just an approximation to the true loading conditions in the native joint. For simplification reasons we also did not simulate typical motion patterns such as sitting-to-standing or gait cycle. Since we only performed static loading, the conchoid shape of the hip joint, which is important when performing dynamic loading, was also disregarded. This might be a limitation, when interpreting our results. Finally, although the CT scans were performed in the supine position and the loading condition is based on one-leg stance situation, this is not an infrequent practice, [26] and previous work [27] has shown that there was no significant difference between the contact pressure in the one-leg stance reference frame and those in the supine reference frame.
Our results are reflected conclusively in the current literature. Zhao et al. [13] conducted a 3D FE analysis investigating the changes of von Mises stress distribution in the cortical bone before and after PAO surgery. They showed the favorable stress distribution in the normal hips compared to dysplastic hips. One limitation of this study might be that the specimens were not truly dysplastic hips. The authors created dysplasia by deforming the acetabular rim of normal hip joints. Hence, their depiction of the stress distribution in the dysplastic joint is rather an approximation. Furthermore, they used a constant thickness cartilage model. They did not estimate pressure distribution in the cartilage model but in the underlying subchondral cortical bone. Another group developed a biomechanical guiding system (BGS) [12, 26, 28]. In 2009 they presented a manuscript reporting on three-dimensional mechanical evaluation of joint contact pressure in 12 PAO patients with a 10-year follow-up. They measured radiologic angles and joint contact pressures in these patients pre- and postoperatively. The authors were able to show that after a 10-year follow-up, peak contact pressures were reduced 1.7-fold and that lateral coverage increased in all patients. One limitation of their study is the use of discrete element analysis (DEA). Since the system was not only used for preoperative planning but also as an intraoperative guidance system, the DEA represents a computationally efficient method for modeling of cartilage stress by neglecting underlying bone stress. The cartilage models however remain largely approximated, since neither patient-specific nor constant cartilage models are used, but a simplified distribution of spring elements is employed for cartilage simulation. Recently, Zou et al. [14] also developed a 3D FE simulation of the effects of PAO on contact stresses. They validated their method on five models generated from CT scans of dysplastic hips and used constant thickness cartilage models. The acetabulum of each model was rotated in 5° increments in the coronal plane from the original position, and the relationship between contact area and pressure, as well as von Mises stress in the cartilage, was investigated, looking for the optimal position for the acetabulum. One limitation of this study is that acetabular reorientation was roughly performed with commercial FE analysis software (Abaqus1, Dassault Systèmes Simulia Corp, USA). Unlike our morphological-based planning application, their method is thus unvalidated and does not have a precise planning tool for an accurate quantification of patient specific 3D hip joint morphology.
In conclusion, our investigation contributes well to the current state of the art. First, to the best knowledge of the authors, this is the first study to use a patient-specific cartilage model for biomechanics-based planning of PAO allowing for estimation of changes of contact areas and peak pressures in truly dysplastic hips. Previous studies had investigated either normal or dysplastic hips, but never the true change during virtual reorientation of the latter. Furthermore, our results seem conclusive, since the optimal position with the largest contact areas and lowest peak pressures were found within the predefined normal values [3, 29] for the investigated LCE angle. This range for safe positioning is especially important, since in real-time surgery, reorientation toward the one “perfect” position might not be feasible. Finally, the comparison to constant thickness cartilage models is another novelty. Strong correlation was found for biomechanical optimization results between these two cartilage models. This is encouraging, since acquisition of patient-specific cartilage requires special multiplanar arthrography imaging (e.g., CT arthrography or MRI with dGEMRIC, T1rho or T2 mapping), while constant thickness cartilage is basically always available. Although our study has its limitations and further investigation is needed, computer-assisted planning with FE modeling using constant thickness cartilage might be a promising PAO planning tool providing conclusive and plausible results.
References
Ganz R, Klaue K, Vinh TS, Mast JW (1988) A new Periacetabular osteotomy for the treatment of hip dysplasias technique and preliminary results. Clin Orthop Relat Res 232:26–36. PMID: 3383491
Steppacher SD, Tannast M, Ganz R, Siebenrock KA (2008) Mean 20-year followup of Bernese periacetabular osteotomy. Clin Orthop Relat Res 466(7):1633–1644. https://doi.org/10.1007/s11999-008-0242-3. PMID: 18449617
Tannast M, Hanke MS, Zheng G, Steppacher SD, Siebenrock KA (2015) What are the radiographic reference values for acetabular under-and overcoverage? Clin Orthop Relat Res 473(4):1234–1246. https://doi.org/10.1007/s11999-014-4038-3
Tannast M, Siebenrock KA, Anderson SE (2007) Femoroacetabular impingement: radiographic diagnosis—what the radiologist should know. Am J Roentgenol 188(6):1540–1552. https://doi.org/10.2214/AJR.06.0921. PMID: 17515374
Tannast M, Mistry S, Steppacher SD, Reichenbach S, Langlotz F, Siebenrock KA et al (2008) Radiographic analysis of femoroacetabular impingement with Hip2Norm-reliable and validated. J Orthop Res 26(9):1199–1205. https://doi.org/10.1002/jor.20653. PMID: 18404737
Clohisy JC, Carlisle JC, Trousdale R, Kim YJ, Beaule PE, Morgan P et al (2009) Radiographic evaluation of the hip has limited reliability. Clin Orthop Relat Res 467(3):666–675. https://doi.org/10.1007/s11999-008-0626-4. PMID: 19048356
Klaue K, Wallin A, Ganz R (1988) CT evaluation of coverage and congruency of the hip prior to osteotomy. Clin Orthop Relat Res 232:15–25. PMID: 3383480
Millis MB, Murphy SB (1992) Use of computed tomographic reconstruction in planning osteotomies of the hip. Clin Orthop Relat Res 274:154–159. PMID: 1729000
Dutoit M, Zambelli P (1999) Simplified 3D-evaluation of periacetabular osteotomy. Acta Orthop Belg 65(3):288–294. PMID: 10546351
Janzen D, Aippersbach S, Munk P, Sallomi D, Garbuz D, Werier J et al (1998) Three-dimensional CT measurement of adult acetabular dysplasia: technique, preliminary results in normal subjects, and potential applications. Skelet Radiol 27(7):352–358. https://doi.org/10.1007/s002560050397. PMID: 9730324
Dandachli W, Kannan V, Richards R, Shah Z, Hall-Craggs M, Witt J (2008) Analysis of cover of the femoral head in normal and dysplastic hips NEW CT-BASED TECHNIQUE. J Bone Joint Surg Br Vol 90(11):1428–1434. https://doi.org/10.1302/0301-620X.90B11.20073
Armand M, Lepistö J, Tallroth K, Elias J, Chao E (2005) Outcome of periacetabular osteotomy: joint contact pressure calculation using standing AP radiographs, 12 patients followed for average 2 years. Acta Orthop 76(3):303–313. PMID: 16156455
Zhao X, Chosa E, Totoribe K, Deng G (2010) Effect of periacetabular osteotomy for acetabular dysplasia clarified by three-dimensional finite element analysis. J Orthop Sci 15(5):632–640. https://doi.org/10.1007/s00776-010-1511-z. PMID: 20953924
Zou Z, Chávez-Arreola A, Mandal P, Board TN, Alonso-Rasgado T (2013) Optimization of the position of the acetabulum in a ganz periacetabular osteotomy by finite element analysis. J Orthop Res 31(3):472–479. https://doi.org/10.1002/jor.22245. PMID: 23097237
Harris MD, Anderson AE, Henak CR, Ellis BJ, Peters CL, Weiss JA (2012) Finite element prediction of cartilage contact stresses in normal human hips. J Orthop Res 30(7):1133–1139. https://doi.org/10.1002/jor.22040. PMID: 22213112
Liu L, Ecker T, Schumann S, Siebenrock K, Nolte L, Zheng G (2014) Computer assisted planning and navigation of periacetabular osteotomy with range of motion optimization. Springer, Heidelberg, pp 643–650
Henak CR, Abraham CL, Anderson AE, Maas SA, Ellis BJ, Peters CL et al (2014) Patient-specific analysis of cartilage and labrum mechanics in human hips with acetabular dysplasia. Osteoarthr Cartil 22(2):210–217. https://doi.org/10.1016/j.joca.2013.11.003. PMID: 24269633
Liu L, Ecker T, Xie L, Schumann S, Siebenrock K, Zheng G (2015) Biomechanical validation of computer assisted planning of periacetabular osteotomy: a preliminary study based on finite element analysis. Med Eng Phys. https://doi.org/10.1016/j.medengphy.2015.09.002
Anderson AE, Ellis BJ, Maas SA, Peters CL, Weiss JA (2008) Validation of finite element predictions of cartilage contact pressure in the human hip joint. J Biomech Eng 130(5):051008. https://doi.org/10.1115/1.2953472. PMID: 19045515
Caligaris M, Ateshian GA (2008) Effects of sustained interstitial fluid pressurization under migrating contact area, and boundary lubrication by synovial fluid, on cartilage friction. Osteoarthr Cartil 16(10):1220–1227. https://doi.org/10.1016/j.joca.2008.02.020. PMID: 18395475
Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J et al (2001) Hip contact forces and gait patterns from routine activities. J Biomech 34(7):859–871. https://doi.org/10.1016/S0021-9290(01)00040-9. PMID: 11410170
Phillips A, Pankaj P, Howie C, Usmani A, Simpson A (2007) Finite element modelling of the pelvis: inclusion of muscular and ligamentous boundary conditions. Med Eng Phys 29(7):739–748. https://doi.org/10.1016/j.medengphy.2006.08.010
Ferguson S, Bryant J, Ganz R, Ito K (2003) An in vitro investigation of the acetabular labral seal in hip joint mechanics. J Biomech 36(2):171–178. https://doi.org/10.1016/S0021-9290(02)00365-2. PMID: 12547354
Konrath GA, Hamel AJ, Olson SA, Bay B, Sharkey NA (1998) The role of the acetabular labrum and the transverse acetabular ligament in load transmission in the hip. J Bone Joint Surg 80(12):1781–1788
Henak CR, Ellis BJ, Harris MD, Anderson AE, Peters CL, Weiss JA (2011) Role of the acetabular labrum in load support across the hip joint. J Biomech 44(12):2201–2206. https://doi.org/10.1016/j.jbiomech.2011.06.011. PMID: 21757198
Armiger RS, Armand M, Tallroth K, Lepistö J, Mears SC (2009) Three-dimensional mechanical evaluation of joint contact pressure in 12 periacetabular osteotomy patients with 10-year follow-up. Acta Orthop 80(2):155–161. https://doi.org/10.3109/17453670902947390. PMID: 19404795
Niknafs N, Murphy RJ, Armiger RS, Lepistö J, Armand M (2013) Biomechanical factors in planning of periacetabular osteotomy. Frontiers in bioengineering and biotechnology; 1. https://doi.org/10.3389/fbioe.2013.00020. PMID: 25152876
Lepistö J, Armand M, Armiger RS (2008) Periacetabular osteotomy in adult hip dysplasia–developing a computer aided real-time biomechanical guiding system (BGS). Suomen ortopedia ja traumatologia = Ortopedioch traumatologi i Finland = Finn J Orthop Traumatol 31(2):186. PMID: 20490364
Haefeli P, Steppacher S, Babst D, Siebenrock K, Tannast M (2015) An Increased Iliocapsularis-to-rectusfemoris Ratio Is Suggestive for Instability in Borderline Hips. Clin Orthop Relat Res 473(12):3725–3734. https://doi.org/10.1007/s11999-015-4382-y
Acknowledgments
This work was supported by the open source dysplastic hips image data from the University of Utah [17] and partially supported by Natural Science Foundation of SZU (grant no. 2017089) and Japanese-Swiss Science and Technology Cooperation Program.
This chapter was modified from the paper published by our group in PLoS One (Li et al., PLoS One 2016; 11(1):e0146452). The related contents were reused with permission.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer Nature Singapore Pte Ltd.
About this chapter
Cite this chapter
Liu, L., Siebenrock, K., Nolte, LP., Zheng, G. (2018). Biomechanical Optimization-Based Planning of Periacetabular Osteotomy. In: Zheng, G., Tian, W., Zhuang, X. (eds) Intelligent Orthopaedics. Advances in Experimental Medicine and Biology, vol 1093. Springer, Singapore. https://doi.org/10.1007/978-981-13-1396-7_13