Abstract
Modern semiconductor sensors are based on an MOS (metal–oxide–semiconductor) structure and their further development is tending to combine them together with UV irradiation. In this paper, a system is presented that can pave the way towards gas analyzers that rely on visible spectrum light. The article presents a semiconductor gas sensor induced with safe optical irradiation delivered from the LED matrix to the semiconductor surface. Different irradiation wavelengths (440, 530, 600, 710 nm) under constant flux were used separately to find the condition for the best sensor response on gas sorption. An output signal was recorded in zero-level emission and in the presence of ethanol. Changes in sensor response and signal rise/relaxation time constant in the presence of saturated ethanol vapor were observed. Sensor response to the ethanol vapor was detected for each used irradiation wavelength and it is approximated by the non-linear falling regression curve with the highest sensor response at 440 nm (R = 104%). In addition, the different rise and relaxation time constant of the signal trace was detected depending on the irradiation wavelength. The time constant ratio is approximated by the non-linear rising regression curve with a maximum value at 710 nm for both signal rise (τ = 0.61) and relaxation (τ = 2.08) parts.
Access provided by Autonomous University of Puebla. Download conference paper PDF
Similar content being viewed by others
Keywords
1 Introduction
Nowadays, there is increasing concern about the environmental pollution which increases each year, thereby causing irreparable damage and influencing every inhabitant on Earth. Air pollution from industrial gases, from fuel combustion products and other sources, inevitably increases the incidence of diseases, including: lung cancer, asthma, heart attacks and allergies [1, 2].
Modern methods are directed to prevent and even foresee the problem. Studies show that the determination of biomarkers in exhaled air can become a powerful tool for medical diagnostics. Currently, there are certain biomarkers in exhaled air that are accompanied by diseases such as obesity, lung cancer, kidney failure, heart attacks and diabetes mellitus [3, 4].
Along with this, the ethanol concentration in exhaled air may correspond as an indicator of the glucose level in blood [5]. Moreover, the potential endogenous source of ethanol is intestinal bacterial flora [6]. This is a challenging area, as the ethanol concentration in exhaled air is normally much lower than the concentration level in human breath after alcohol ingestion [7].
In the past decade, semiconductor gas sensors have been rapidly developed. The sensors are cheap, easy to use, able to detect a variety of different gases and are stable. Because of this, the sensor gained widespread use. Gas sensor technologies are providing significant progress in the detection of hazardous substances and exhaled biomarkers [8].
Among the actual technologies widely distributed, MOS sensors have several significant drawbacks. They are not appropriate for the detection of inflammable gases with a low autoignition temperature as the sensing area must be heated up to 500 °C. Hence, heating to a higher temperature demands higher power consumption [9, 10]. More often the sensors are combined with UV irradiation. As a result, gases adhere to the MOS surface and the current via the MOS transistor is influenced [11,12,13]. However, UV light is potentially harmful for humans and the environment.
As an alternative, optically induced semiconductor gas sensors were designed to provide both high sensitivity without heating the sensing area and a possibility to differentiate the presence of different gases by measuring the output signal and processing time constant at different supply voltages [14]. Furthermore, the electrophysical properties of an optically induced semiconductor gas sensor depending on the irradiation type and measurement of low acetone concentration under UV stimulation at room temperature were shown previously [15]. Despite the progress already made, several knowledge gaps in the working principle still exist.
2 Materials and Methods
The aim for the current research was to conduct sensor response measurements and time constant ratio calculations using optical stimulation under different irradiation wavelengths of visible spectrum light.
For those purposes, a test setup (Fig. 1) was built. The LED matrixes (wavelengths equal to 440, 530, 600, 710 nm) were used to provide homogeneous optical irradiation on the semiconductor sensor surface. By using an LED power supply, the equal photon flux was adjusted for each irradiation wavelength. Arduino controller was assembled to carry out measurements, process signal trace and export data for further processing to a PC.
2.1 Sensor Response Measurement
During each measurement in zero-level emission and in the presence of saturated ethanol vapor, an output signal was recorded. Finally, the sensor response on ethanol was calculated:
where S(ethanol)—is the signal peak value in an ethanol environment; S(ambient)—is the zero-level emission signal peak value.
2.2 Signal Rise and Relaxation Time
Trace of the sensor response was recorded and then processed to the time constant through the Arduino microcontroller (Fig. 2).
Signal rise and relaxation time can be characterized using the following definition:
-
Rise time (tr)—the interval of time it takes an output signal to increase from 10 to 90% of its peak value;
-
Relaxation time (tx)—the time interval required for an output signal to decrease from 90 to 10% of its peak value.
Time constant ratio defined as the ratio between the time constant of the signal in the presence of ethanol and the time constant of the signal in zero-level emission was calculated:
where τ(ethanol)—is the time consonant measured in an ethanol environment; τ(ambient)—is the time constant measured in a zero-level emission.
3 Results and Discussion
3.1 Sensor Response Measurement
The obtained result indicates the non-linear dependence of the semiconductor gas sensor response to the optical irradiation wavelength and can be approximated by a falling regression curve. The maximal signal response was detected at 440 nm and equals R = 104% and the minimum sensor response at 710 nm, R = 39% (Fig. 3).
3.2 Signal Rise and Relaxation Time
Figure 4 shows the processed signal rise and relaxation time from the signal shapes recorded at a different irradiation wavelength.
Obtained results indicate non-linear dependence of processed time constant ratio to optical irradiation wavelength for both rise and relaxation time. Both results can be approximated by a rising regression curve. For both time constant ratios, the minimal value was detected at 440 nm and maximal value at 710 nm.
4 Conclusions
Visible spectrum LED light may be used for an optically induced semiconductor gas sensor to carry out the sensor response measurement of ethanol vapor. Highest sensor response was detected at the lowest tested irradiation wavelength of 440 nm.
Processed values of signal rise and relaxation time ratio may be used to sense the presence of ethanol vapor in air.
A correlation between the sensor response and time constant ratio was not observed and, therefore, both methods can be used jointly to increase the detection quality.
Although results demonstrate potential application of visible light usage to detect presence of ethanol, current results are achieved for saturated ethanol vapor and, therefore, concentration comparison to modern progress cannot be done.
References
Kampa,M., Castanas,E.: Human health effects of air pollution. Environmental Pollution 151(2), 362–367 (2008).
Katsouyanni,K.: Ambient air pollution and health. British Medical Bulletin 68(1), 143–156 (2003).
Risby,T.H., Sehnert,S.S.: Clinical application of breath biomarkers of oxidative stress status. Free Radical Biology and Medicine 27(11–12), 1182–1192 (1999).
de Zwart,L.L., Meerman,J.H., et al.: Biomarkers of free radical damage: Applications in experimental animals and in humans. Free Radical Biology and Medicine 26(1–2), 202–226 (1999).
Galassetti,P.R., Novak,B., et al.: Breath ethanol and acetone as indicators of serum glucose levels: an initial report. Diabetes Technology & Therapeutics 7(1), 115–123 (2005).
Cope,K., Risby,T., Diehl,A.M.: Increased gastrointestinal ethanol production in obese mice: Implications for fatty liver disease pathogenesis. Gastroenterology 119(5), 1340–1347 (2000).
Miekisch,W., Schubert,J.K., Noeldge-Schomburg,G.F.: Diagnostic potential of breath analysis—focus on volatile organic compounds. Clinica Chimica Acta 347(1–2), 25–39 (2004).
Arshak,K., Moore,E., et al.: A review of gas sensors employed in electronic nose applications. Sensor Review 24(2), 181–198 (2004).
Kim,H.J., Lee,J.H.: Highly sensitive and selective gas sensors using p-type oxide semiconductors: Overview. Sensors and Actuators, B: Chemical 192, 607–627 (2014).
Karmakar,M., Mondal,B., et al.: Acetone and ethanol sensing of barium hexaferrite particles: A case study considering the possibilities of non-conventional hexaferrite sensor. Sensors and Actuators B: Chemical 190, 627–633 (2014).
Herrán,J., Fernández-González,O., et at.: Photoactivated solid-state gas sensor for carbon dioxide detection at room temperature. Sensors and Actuators B: Chemical 149(2), 368–372 (2010).
Comini.E., Faglia,G., Sberveglieri,G.: UV light activation of tin oxide thin films for NO2 sensing at low temperatures. Sensors and Actuators B: Chemical 78(1–3), 73–77 (2001).
Chen,H., Liu,Y., et al.: A comparative study on UV light activated porous TiO2 and ZnO film sensors for gas sensing at room temperature. Ceramics International 38(1), 503–509 (2012).
Dekhtyar,Y., Sneiders,M., et al.: Towards Optically Induced Semiconductor Human Exhalation Gas Sensor. In: XIV Mediterranean Conference on Medical and Biological Engineering and Computing 2016, IFMBE Proceedings, vol. 57, pp. 482–485. Springer, Cham (2016).
Dekhtyar,Y., Sneiders,M., et al.: Optically Induced Semiconductor Gas Sensor: Acetone Detection Range using Continuous and Cyclic Optical Irradiation Types. In: EMBEC & NBC 2017. EMBEC 2017, NBC 2017. IFMBE Proceedings, vol. 65, pp. 330–333. Springer, Singapore (2018).
Conflict of Interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2019 Springer Nature Singapore Pte Ltd.
About this paper
Cite this paper
Dekhtyar, Y., Komars, M., Sneiders, M. (2019). Semiconductor Ethanol Sensor Inducted with Visible Light. In: Lhotska, L., Sukupova, L., Lacković, I., Ibbott, G. (eds) World Congress on Medical Physics and Biomedical Engineering 2018. IFMBE Proceedings, vol 68/3. Springer, Singapore. https://doi.org/10.1007/978-981-10-9023-3_1
Download citation
DOI: https://doi.org/10.1007/978-981-10-9023-3_1
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-10-9022-6
Online ISBN: 978-981-10-9023-3
eBook Packages: EngineeringEngineering (R0)