Abstract
For the sensitive and quantitative measurement of protein biomarkers, pathogens, and cells in clinical samples, magnetic nanoparticles (MNPs) offer unique advantages over traditional detection methods. Specifically, due to the inherently negligible magnetic background of biological material, MNPs can be used to obtain highly sensitive measurements in minimally processed samples. Our detection platform, termed diagnostic magnetic resonance (DMR), exploits MNPs to modulate the nuclear magnetic spin-spin relaxation time of water. Here, we review work done by our group to develop more effective MNP biosensors, advanced conjugational strategies to target the MNPs to molecular targets, and highly sensitive miniaturized NMR systems. We demonstrate this platform as a robust and easy-to-use system for the detection of a wide range of targets in clinical settings including whole cells, proteins, DNA/mRNA, metabolites, drugs, viruses, and bacteria.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
- Nuclear Magnetic Resonance
- Nuclear Magnetic Resonance Data
- Nuclear Magnetic Resonance Signal
- Solenoidal Coil
- Nuclear Magnetic Resonance Probe
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
1 Introduction
Robust, sensitive, and easy-to-use biosensors for the detection and quantification of rare biomarkers will have significant applications in both basic research and clinical practice. If made available, these platforms could aid in understanding of fundamental biology, in accurately detecting diseases at their early stage, and in evaluating and monitoring the efficacy of therapy [1–3]. To realize such sensors, the underlying detection technology should ideally (1) enable high sensitivity and accuracy, with minimal false positives and negatives; (2) support short assay time with minimal sample processing; and (3) allow for multiplexed detection in a single parent sample [4]. Different types of sensing platforms, fulfilling some of these requirements, have been developed based on optical [5, 6], electronic [7, 8], or mass-based [9] detection. These systems, however, often require lengthy sample purification, large sample volumes, or long assay time, which can potentially limit their clinical utility and adaption.
Biosensors based on magnetic detection have recently emerged as a promising diagnostic platform. Due to the intrinsically negligible magnetic susceptibilities of biological entities, magnetic detection experiences little interference from native biological samples; even optically turbid samples will often appear transparent to magnetic fields. Biomarkers of interests, when magnetically labeled, however, can attain a high contrast against the biological background. Recent progresses in the synthesis of magnetic nanoparticles (MNPs) have further advanced the magnetic detection technology. With their size scale similar to that of biological molecules, MNPs can efficiently and abundantly bind to biological targets, amplifying analytical signals [10–13]. Various detection technologies have been developed based on this magnetic-tagging concept. These include techniques that use magnetometers, such as superconducting quantum interference device (SQUID) [14–16], magnetoresistive sensors [17–20], and Hall sensors [21], all of which directly measure the magnetic fields arising from the magnetically labeled targets.
We have recently developed a new magnetic sensing platform, diagnostic magnetic resonance (DMR) [22]. Contrary to directly measuring the magnetic moments of the labeled targets, the DMR uses nuclear magnetic resonance (NMR) as the detection mechanism. When placed in NMR magnetic fields, MNPs create local magnetic fields and change the relaxation rate of surrounding water molecules [23]. The detection offers an intrinsic signal amplification mechanism, as more than millions of water molecules can be affected by a single MNP. Moreover, since the signal is generated from the entire sample volume, the assay procedure is significantly simpler than the direct magnetic detection in which MNP-labeled targets have to be closely positioned to the sensing elements.
By optimizing MNPs and miniaturizing NMR detectors, the DMR detection sensitivities for various target types have been considerably improved over the last few years. These developments enable rapid and multiplexed detection on a wide range of targets in microliter sample volumes, including nucleic acids [24], proteins [22], drugs, bacteria [25], and tumor cells [26–28]. With the recent integration of bioorthogonal targeting strategies [27, 29] as well as accurate and real-time control of device temperature [23], the DMR platform has become more robust and sensitive, allowing operation in clinical settings [26]. This chapter reviews the latest development of the DMR technology, focusing on its three major components: magnetic nanoagents, miniature NMR systems, and optimized assay protocols. Specific biomedical and clinical DMR applications will also be discussed.
2 Principle of DMR Detection
The DMR detection of MNP-labeled cells is realized by exploiting the “T 2-shortening” effect of MNPs in NMR measurements [30]. When placed in static, polarizing magnetic fields for NMR detection, MNPs produce local dipole fields with strong spatial dependence, which efficiently destroy the coherence in the spin-spin relaxation of water protons. MNP-labeled objects consequently cause faster decay of NMR signal, or shorter transverse relaxation time T 2, than nontargeted ones (Fig. 9.1).
The capability of MNPs to induce T 2 changes is defined as transverse relaxivity (r 2) [31]. With MNPs in solution, the relaxation rate (\({R}_{2} = 1/{T}_{2}\)) can be expressed as [28]
where R w is the relaxation rate of the background (usually water), V is the NMR detection volume, and N P is the total number of MNPs in V. If each biological cell has n MNPs and the total number of cells is \({N}_{C}(= {N}_{P}/n)\), the net change of \({R}_{2}(\Delta {R}_{2} = {R}_{2} - {R}_{w})\) is given as
where \({r}_{2}^{\mathit{cell}}(=\,n \cdot {r}_{2})\) is defined as the cellular relaxivity (transverse relaxivity per given cell concentration). Note that \({r}_{2}^{\mathit{cell}}\) is indicative of the abundance of relevant surface biomarkers. NMR thus can be used effectively for molecular profiling of target cells [27, 28]. Equation 9.2 provides valuable insights into how to increase the sensitivity and specificity of NMR-based sensors:
-
MNPs with high r 2 relaxivity. Pronounced R 2 changes will occur when cells are labeled with MNPs of high r 2 relaxivity [23]. Because r 2 is proportional to the magnetic moment (μ p ) of particles [32, 33], making magnetically stronger MNPs will benefit the measurements.
-
Maximal MNP labeling on cell. The R 2 changes are also directly proportional to the number of MNPs loaded onto cells. In addition to increasing the r 2 potency of individual MNPs, it is equivalently important to establish a labeling protocol to maximize and/or amplify MNP loading on cells.
-
Miniaturized NMR probes. Higher sensitivity can be achieved on the device level by decreasing the NMR detection volume (V ). This approach can effectively increase the analyte concentration (N C ∕ V ), leading to large \(\Delta {R}_{2}\). Furthermore, smaller NMR probes assume higher SNR (signal-to-noise) ratio due to the increased sample filling factor. It can be shown that the sensitivity of NMR coils (with a typical dimension of d) scales as \({d}^{-1/2}\) [34].
These recognitions motivated us to explore three major activities in DMR development: synthesis of new MNPs, optimization of such MNPs for cellular labeling, and miniaturization of NMR systems. The following sections will describe these accomplishments.
3 New Magnetic Nanoparticles
We have developed many different types of MNPs (Table 9.1) and engineered the particle size and composition to enhance the transverse relaxivity. According to the outer-sphere model of transverse relaxation, the r 2 value of an MNP is proportional to \({\tau }_{d} \cdot {M}^{2}\), where τ d is the residence time of water molecules around the particle and M is the particle magnetization [35]. The efforts to enhance r 2 were thus focused on synthesizing larger MNPs using magnetically stronger material. We herein introduce two types of such particles: metal-doped ferrite and iron (Fe)-based MNPs.
3.1 Ferrite-Based MNPs
With their excellent stability and biocompatibility, cross-linked iron oxide (CLIO) nanoparticles have been widely used for DMR applications [22, 36]. CLIO nanoparticles contain a superparamagnetic iron oxide core (3–5-nm monocrystalline iron oxide) composed of ferrimagnetic magnetite \(({\mathrm{Fe}}_{3}{\mathrm{O}}_{4})\) and/or maghemite (\(\gamma {\mathrm{Fe}}_{2}{\mathrm{O}}_{3}\)). The metallic core is encased with biocompatible dextran, which is cross-linked and functionalized with primary amine. Amine-terminated CLIO nanoparticles have an average hydrodynamic diameter of 38 nm, and about 60 amine groups are available for bioconjugation per nanoparticle. The r 2 of CLIOs is \(\sim 50\,{\mathrm{s}}^{-1}{\mathrm{mM}}^{-1}\) [Fe] [24, 37] as measured at \(4{0}^{\circ }\mathrm{C}\) and at the external field of B 0 = 0. 5 T.
Two main strategies have been employed to further improve the magnetization of ferrite nanoparticles and thereby the r 2 relaxivity: magnetic doping and nanoparticle sizing. Magnetic doping with ferromagnetic elements such as manganese (Mn), cobalt (Co), or nickel (Ni) has been known to modulate the overall magnetization of MNPs [38, 39]. Among these doped ferrite MNPs, \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) nanoparticles have the highest magnetization, as Mn2 + ions have the highest spin quantum number (5/2). Moreover, larger nanoparticles are also known to have increasing magnetization [40]. Spin canting, a feature that decreases the overall magnetic moment of small nanoparticle due to tilted surface spins, can be reduced in bigger nanoparticles to increase the overall magnetization. Concurrently, larger particle size further enhances the particle r 2 by increasing τ d .
We employed both magnetic doping and sizing strategies to produce \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) nanoparticles with superior r 2 relaxivity [28]. These particles were synthesized in the organic phase by reacting iron (III) acetylacetonate [Fe(acac)3], manganese (II) acetylacetonate [Mn(acac)2], and 1,2-hexadecanediol at elevated temperature (300 ∘ C). Through a seed-mediated growth approach, the particle size was stepwise increased from 10 nm to 12, 16, or 22 nm. \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) nanoparticles with diameter ≤ 16 nm were found to be highly monodisperse and superparamagnetic at 300 K (Fig. 9.2). Because of their larger magnetic core, these \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) nanoparticles assumed high relaxivities with r 2 values approaching \(420\,{\mathrm{s}}^{-1}.{\mathrm{mM}}^{-1}\)[metal] (equal to \(6 \times 1{0}^{-14}L \cdot \,{\mathrm{s}}^{-1}\) per particle), more than eight times greater than CLIO nanoparticles in metal basis (\(50\,{\mathrm{s}}^{-1}{\mathrm{mM}}^{-1}\)[metal] or \(7 \times 1{0}^{-16}L\,{\mathrm{s}}^{-1}\) per particle) [28].
3.2 Fe-Core MNPs
Ferromagnetic metals, instead of their oxides, have been suggested as an ideal constituent of MNPs for their superior magnetization. However, while Fe-core MNPs can achieve high r 2 relaxivities [41], these monometallic MNPs are extremely reactive and require protective layers to prevent rapid oxidation.
Recently, a unique 16-nm Fe-core/ferrite-shell MNP (Fe@ferrite) has been developed for DMR applications (Fig. 9.3) [42]. The particle consists of an elemental iron core (not iron oxide) and a protecting oxide shell. A novel synthetic route was established, as summarized below, that allows for the preparation of large yet monodisperse Fe MNPs and the growth of protective ferrite shells around existing Fe MNPs (Fig. 9.3a).
-
Larger Fe cores. Fe MNPs were formed by thermally decomposing metal complexes [Fe(CO)5] in the presence of surfactant (oleylamine) under air-free condition. By increasing the reaction temperature during Fe MNP synthesis, the particle size could be proportionally increased [42]. The phenomenon can be attributed to the higher reactivity of Fe ions at elevated temperatures during particle formation [43]. Applying this approach, Fe MNPs with diameters up to 18 nm could be prepared while maintaining the relative size variations < 5%.
-
Ferrite-shell overcoat. Native Fe MNPs undergo rapid oxidation, which necessitates the presence of protective shells. To prepare such shells, we carried out the reaction for ferrite synthesis in the presence of as-prepared Fe MNPs, based on the hypothesis that Fe MNPs could serve as nucleation sites for ferrite formation. Indeed, the resulting particles (Fe@ferrite, Fig. 9.3b) had Fe cores whose sizes were similar to that of the initial Fe MNPs (dashed circles in Fig. 9.3b). Importantly, Fe@ferrite MNPs maintained their shape and magnetic properties over time (Fig. 9.3c), verifying the sturdy protection against oxidation by the shell. Note that we further enhanced the overall magnetic moment and r 2 relaxivity by doping the ferrite with Mn2 + during the shell formation (Fe@MnFe\({}_{2}{\mathrm{O}}_{4}\) MNPs).
The Fe@ferrite MNPs assumed higher saturation magnetization (796 kA m − 1) and r 2 (\(7 \times 1{0}^{-14}\,L\,{\mathrm{s}}^{-1}\) per particle, \(430\,{\mathrm{s}}^{-1}{\mathrm{mM}}^{-1}\)[metal]) than similarly sized ferrite MNPs, primarily due to the large Fe cores. It is noteworthy that the Fe core is in a thermally stable ferromagnetic state with nonzero coercivity. The ferrite shell, which is superparamagnetic, however, effectively reduces the overall coercivity of particles by leading the magnetization processes at small external magnetic fields [42]. The resultant Fe@ferrite MNPs thus displayed a unique magnetic feature, namely, the presence of hysteresis with negligible coercivity (Fig. 9.3d). This property is crucial in preventing interparticle aggregations from magnetic interactions. When applied for DMR assays, these Fe@ferrite MNPs achieved superior performance, capable of detecting picomolar avidin and single cancer cells in whole blood samples.
4 Optimizing MNPs for DMR Applications
In addition to the above-mentioned strategies to improve nanoparticle relaxivities through inorganic chemistry, postsynthesis modifications such as better particle surface chemistry and new labeling approaches have also been developed for DMR applications. These novel postsynthesis modifications not only improve the detection sensitivities but also simplify the targeting assays, making the DMR platform easily applicable to detect a wide range of biological entities and translatable for effective clinical utility.
4.1 Biocompatible Coating on Hydrophobic MNPs
Most MNPs, synthesized via the thermal decomposition method, are suspended in nonpolar solvents and coated with hydrophobic surfactant. For biological applications, these particles should be transferred into aqueous phase. We have traditionally used a small bifunctional molecule DMSA (meso-2,3-dimercaptosuccinic acid) to replace hydrophobic capping layers (e.g., oleic acid or oleylamine) on MNP surfaces [44, 45]. The resulting particles, however, displayed short-term stability ( < 3 month), gradually precipitating out in physiological buffers [44].
Overcoming the issue, we have established a new, polymer-based surface coating that can render MNPs hydrophilic with superb stability under varying pH and ionic strength (Fig. 9.4a) [46]. As a coating substrate, we selected polyvinyl alcohol (PVA) since the material is synthetic, inexpensive, hydrophilic, and biodegradable [47–49]. The polymer was further modified into carboxymethyl polyvinyl alcohol (CMPVA); we hypothesized that multidentate carboxylic (-COOH) groups would allow CMPVA to strongly bind to the metal oxide surface. Indeed, when hydrophobic \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) MNPs were coated with CMPVA, the particles showed excellent, long-term solubility ( > 12 months) in aqueous buffers, whereas PVA-coated particles precipitated spontaneously (Fig. 9.4b). CMPVA coating also provided free amine ( − NH2) groups for bioconjugation.
When tested for cellular labeling, the CMPVA coating exhibited extremely low nonspecific binding. Figure 9.4c shows an example of cellular labeling with CMPVA-MNPs. Cancer cells (SkBr3) were first labeled with biotinylated anti-HER2/neu monoclonal antibodies, followed by an incubation with neutravidin-modified, fluorescent CMPVA-MNPs. Control samples were prepared in the same way but without antibody injection. Strong fluorescence signal could be visualized on the surface of targeted cells, whereas the signal from control samples was negligible. This obvious difference between the control and positively labeled samples is highly important for DMR assay in reducing false positives in biosensing.
4.2 New Labeling Method for High MNP Loading
Besides improving the physical properties of MNPs, equally important for the DMR assay is to develop an efficient method for MNP labeling on target cells. A conventional way is to use MNPs preconjugated with target-specific affinity ligands [22], which often requires extensive optimization of the affinity ligands and the conjugation method for each new target.
We have developed a new targeting strategy, BOND (bioorthogonal nanoparticle detection), that is modular and broadly applicable and can amplify MNP binding to biological objects [27]. The BOND is based on [4 + 2] Diels-Alder cycloaddition, especially between tetrazine (Tz) and trans-cyclooctene (TCO; Fig. 9.5a) [50]. The reaction is fast and irreversible (covalent) and can be performed at room temperature without using any catalyst (copper). We have adapted the chemistry for MNP labeling of cells (Fig. 9.5b), wherein cells are pretargeted with TCO-modified antibodies and subsequently incubated with Tz-loaded MNPs (Tz-MNPs). Multiple TCO tags (usually ∼ 20) could be incorporated onto an antibody, and such modified antibodies were found to maintain their affinity. Consequently, the antibodies functioned as a scaffold to promote multiple attachments of Tz-MNPs. Indeed, flow cytometry measurements showed that the BOND yielded ∼ 15-fold improvement in MNP loading on cells, compared to labeling with antibody-MNP direct conjugates (Fig. 9.5c). The trend was further confirmed in NMR-based cell detection (Fig. 9.5d); the BOND method yielded more pronounced T 2 changes and improved the cellular detection limit.
BOND has been successfully adapted for DMR molecular profiling of experimental cellular samples [27] and clinical fine-needle aspirate samples [26]. Recently, the BOND technology was further generalized by developing newer two-step detection schemes based on complementary oligonucleotide approaches [29], alternative cycloaddition chemistries [51], and cyclodextrin/adamantine supramolecular interactions [52].
5 Miniaturized NMR System
The development of miniaturized nuclear magnetic resonance (μNMR) systems [22] represents a key milestone in sensitive detection in DMR. Device miniaturization brings several distinctive advantages for sensitive detection and clinical translation. First, it provides a promising way to improve the detection sensitivity. Smaller NMR systems reduce the detection volume, which in turn effectively increases the concentration of MNP-targeted cells for large R 2 changes (Eq. 9.3) Second, miniaturized NMR probes (coils) produce much stronger radio-frequency (RF) magnetic fields per unit current, leading to higher signal-to-noise per unit sample volume [53]. Third, with smaller RF coils, the requirement for spatial homogeneity of static magnetic fields becomes less stringent, making it possible to use small, portable magnets [22]. The entire measurement system can be realized as a portable device for point-of-care operations.
5.1 System Concept
NMR-based cell detection has been previously performed on benchtop relaxometers (e.g., minispec, Bruker) [36, 54–56]. The systems are equipped with permanent, low-field ( < 1 T) magnets for field generation, which simplifies operation and housing of the equipments; however, the main drawbacks of the benchtop system include the use of relatively large sample volumes (\(\sim 100\,\mu \mathrm{L}\)) and the lack of capability for parallel measurements.
Overcoming these limits, the first μNMR prototype was designed and tested for the feasibility of miniaturization. Figure 9.6 shows the main features of the μNMR system: planar microcoils, microfluidic networks, onboard NMR spectrometer, and a portable magnet. The microcoils are used for NMR detection and are arranged in an array format for parallel measurements. The microfluidic networks facilitate the handling and distribution of small volumes of samples. A small, portable magnet (NdFeB, B 0 = 0. 5 T) was employed to generate NMR field. The system measured the T 1 relaxation time using inversion recovery pulse sequences; for T 2 measurements, Carr-Purcell-Meiboom-Gill (CPMG) spin-echo pulse sequences were used to compensate for the inhomogeneity of the polarizing magnetic field. To generate versatile pulse sequences while using minimal electronic parts, we devised a new circuit schematic for NMR electronics that has served as a blueprint for subsequent NMR systems.
5.2 Optimal NMR Probe Design
Reducing sample volume requirements can lead to the effective increase of cell concentrations. However, it can also lead to degradation of the signal-to-noise ratio (SNR), as the absolute level of the NMR signal is proportional to the sample volume. System miniaturization thus should be accompanied by measures to maintain or enhance SNR to truly improve the detection sensitivity. In the second generation of μNMR system, we focused on improving SNR by engineering the NMR probes.
The SNR of a NMR probe can be expressed as
where κ is the fraction of the coil volume (V c ) occupied by the samples (filling factor), M 0 is the nuclear magnetization of the sample, μ0 is the vacuum permeability, ω0 is the Larmor frequency, Q is the quality factor of an NMR coil, and \(\Delta f\) is the bandwidth of a receiver electronics. For a given NMR setup (i.e., the same magnets and electronics), SNR could be improved by increasing κ and Q, which are properties of the NMR probes. Indeed, we have demonstrated a new probe design that achieves both maximal \(\kappa (\approx 1)\) and high Q ( ∼ 25) [28]. In this design, the probe consisted of a solenoidal microcoil embedded in a microfluidic structure (Fig. 9.7a). Solenoidal coils were chosen for their higher SNR than planar or birdcage coils [53]. To increase κ, we adopted the cast-molding technique in device fabrication. First, the coils were wound around polyethylene tubes and subsequently immersed into a polymer (PDMS). After PDMS cure, the tubes were withdrawn to open up the fluidic channels. With this design, the entire bore of the coil was available for samples; in one example, the new probe displayed > 350 % larger SNR than a similar coil wrapped around a tube (Fig. 9.7b). Compared to the lithographically patterned planar coils in our previous systems [22], the improvement in SNR was much more significant ( > 20-fold enhancements); the solenoidal coil excited larger volumes of samples by producing more homogeneous radio-frequency magnetic fields and had smaller electrical resistance (e.g., 0.3 vs. \(3.2\Omega \) at 20 MHz).
We have further modified the embedded coil to improve the overall SNR and to streamline the assay procedure. In the new probe design, a cell-capturing membrane filter was incorporated at the outlet side of the coil (Fig. 9.7c). The filter serves two essential functions. First, it size selectively captures cells and concentrates them inside the NMR detection coil, which leads to more pronounced R 2 changes (Eq. 9.2). The filter also provides a way to detect a small number of targets from large sample volumes. Second, the filter enables on-chip separation of cells from unbound MNPs, therefore obviating the need for separate off-chip purification steps (e.g., centrifugation). Together with the highly magnetic Fe@ferrite MNPs, this new probe has been applied to diagnose tuberculosis [25]. Using the attenuated bacillus Calmette-Guérin (BCG) as a surrogate for Mycobacterium tuberculosis, we detected as few as 20 colony-forming units (CFUs) in sputum (1 mL). Importantly, the entire detection procedure was performed in a single-chip format, minimizing sample loss and making the assay simple and fast (in less than 30 min).
5.3 Clinical System
Our third-generation μNMR system is designed for practical applications in a clinical environment (Fig. 9.8) [57]. Its small size, easy accessibility, and high robustness provide end users with fast and stable measurements of biological samples. The new μNMR system is composed of three core parts: a newly designed probe for clinical samples, NMR electronics, and a user-friendly software.
The system design is based on that of previous generations of μNMR and consists of a small portable magnet (B 0 = 0. 5 T) and a solenoidal coil for higher SNR (Fig. 9.8 left). A custom-made PMMA (polymethyl methacrylate) housing cages the magnet, the microcoil, and RF matching circuit. With its major improvement focused on clinical translatability, this new system uses disposable thin-walled polyimide tubes to load biological samples for measurement, thereby eliminating potential contamination of the NMR probe. The disposable tubes are filled with samples ( ∼ 5 μL) and are inserted into the coil bore for NMR detection. Modular coils made in a variety of sizes can be plugged into the system to optimally accommodate available sample volumes (\(1\mbox{ \textendash }100\,\mu \mathrm{L}\)).
A small form-factor (20 ×12 ×5 cm) NMR electronics is implemented using off-the-shelf integrated circuit (IC) chips to achieve cost-effective (\(<\$200\)) and highly programmable NMR platform (Fig. 9.8 middle). The NMR electronics generates the NMR pulse sequences, acquires the NMR signal, and communicates with external terminals (computer, mobile devices). It has three main parts: a microcontroller unit (MCU), an RF transmitter, and a signal receiver. The MCU (TMS320F28235, Texas Instruments) controls overall RF transceiver operations as well as data communication with external terminals. As an RF transmitter, a direct digital synthesis chip (AD9954, Analog Devices) is employed to generate two RF signals with 90 ∘ phase difference that are modulated by voltage-controlled switches (ADG1419BRMZ, Analog Devices). The RF heterodyne system is implemented to process the NMR signal. First, the signal is amplified by low-noise amplifier (AD604, Analog Devices) and down-converted to baseband (1–10 kHz) by a mixer (ADE-6, Mini-Circuits). Baseband signals subsequently pass a low-pass filter and are digitized by an analog-to-digital converter (AD7625, Analog Devices). Mobile devices (e.g., iPhoneTM, iPadTM), which are connected to the MCU via a standard communication channel (e.g., USB, Bluetooth), receive digital data and show graphical outputs.
The NMR software incorporates graphical user interface (GUI), data acquisition/process, data logging/sharing, and a temperature compensation engine, which deliver a user-friendly interface and contribute to the robustness of the new μNMR system (Fig. 9.8 right). The software is programmed with Objective-C using Cocoa and Cocoa Touch frameworks and operates on iOSTM and OS XTM. In order to implement a graphical representation of R 2 relaxation curve, open-source plotting framework (Core Plot) is cross-linked with Cocoa and Cocoa Touch frameworks. Time domain NMR data are acquired at the negative edge of spin echoes and processed real time. Negative edge-triggered data acquisition reduces the size of the data to be transferred and consequently decreases the data acquisition time via USB/Bluetooth connections by more than 10-fold. Transferred NMR data are then processed to obtain T 2 relaxation time. The overall signal pattern is displayed, and the data is stored in the terminal device. In addition, wireless communication module is embedded in the software for data logging/sharing over an encrypted wireless network, promoting usability in a remote clinical site.
To ensure stable and reliable T 2 relaxation time measurements, a temperature compensation engine has been implemented to run independently from the main thread. This temperature compensation engine transforms NMR data from time domain to frequency domain using fast Fourier transformation (FFT) and detects any changes in the NMR frequency due to environmental temperature fluctuation (described below).
5.3.1 Temperature Compensation
The NMR frequency f 0 is the parameter that requires most frequent adjustments and affects the measured NMR signals most significantly. The frequency f 0 changes as the magnetic field (B 0) from the permanent magnet drifts with temperature [58]. For example, with a 1 ∘ C increase in temperature, B 0 field from a NdFeB magnet will drop ∼ 0. 1 % from its initial value, and f 0 will proportionally decrease by 0.1 %; when the initial f 0 is 20 MHz (B = 0. 47 T), the frequency change is then ∼ 20 kHz. Such changes can place down-converted NMR signals near or beyond the low-frequency cutoff in the amplification stage, distorting the measured signal. Commercial benchtop NMR systems address the problem by housing the entire magnet block inside a heated container. This solution, however, significantly undermines the portability of the system due to the use of bulky and power-consuming parts. In the new μNMR system, we employed a dynamic control approach. Namely, programmable hardware in the NMR electronics and temperature compensation engine in the NMR software are designed to track and compensate for temperature dependency of the system. These implementations ensure optimal measurement settings for reliable and robust performance.
Figure 9.9a and b show the algorithm for temperature compensation. The feedback loop tracks the Larmor frequency f 0 and reconfigures the frequency f of NMR excitation. Coarse-tuning mode starts with initial NMR excitation frequency f i and increases \(f\) by \(\Delta f\). When the spectral power (P) of NMR spin echo reaches a predefined threshold P th , a fine-tuning mode takes over to measure the frequency offset \({f}_{d}(=\vert f - {f}_{0}\vert )\). The fine tuning iterates until f d reaches a target value. The target f d value is carefully selected to keep down-converted NMR signal within the passband of the low-pass filter. Once the new NMR excitation frequency f has been established, CPMG pulse sequence is used to measure the T 2 relaxation time of the sample.
Figure 9.9c demonstrates the effectiveness of the developed temperature compensation method. When f 0 was allowed to drift but the RF frequency (f) for sample excitation was fixed, T 2 values varied up to 200 % relative to its starting value with typical fluctuation of room temperature \((\Delta T \sim {2}^{\circ }\mathrm{C})\). When the temperature compensation engine was activated, however, T 2 variations were significantly reduced to < 1 %. We further tested the system in environmental settings with a wide range of temperature differences (\(4\mbox{ \textendash }5{0}^{\circ }\mathrm{C}\)). To determine the measurement accuracy, the linear dependence of T 2 on temperature was utilized. Figure 9.9d shows the T 2 values of an MNP solution monitored at different temperature. For a given environment setting, the μNMR system was operated with the temperature tracking activated to compensate for minute temperature variations (\(\sim {1}^{\circ }\mathrm{C}\)). The results show a linear relationship (R 2 > 98 %) as theoretically predicted, demonstrating the capacity for reliable T 2 measurements in various settings.
6 Biological Applications
6.1 Cancer Detection and Profiling
Sensitive detection and rapid characterization of tumor cells in minimally processed biological samples will have significant impact on both biomedical research and clinical practice. Using the first-generation DMR device (μNMR-1) [22], DMR molecular profiling of cancer markers (Her2/neu, EGFR, and EpCAM) on human cells was demonstrated with CLIO nanoparticles directly conjugated to monoclonal antibodies. With the development of the second-generation μNMR-2 and the highly magnetic \({\mathrm{MnFe}}_{2}{\mathrm{O}}_{4}\) nanoparticles, subsequent cellular detection sensitivity was remarkably improved to approximately single-cell level, far surpassing the sensitivity of other conventional clinical methods (Fig. 9.10a). Furthermore, a new assay protocol was established that reports the expression level of a specific biomarker and the target cell density.
To independently measure cell numbers by μNMR, we exploited a phenomenon of low-grade phagocytosis of nontargeted MNPs by tumor cells [59]. When mammalian cells were incubated (15 min at \(3{7}^{\circ }\mathrm{C}\)) with unmodified MNPs (MNP-\(\Phi \)), linear and cell-number-dependent R 2 changes (\(\Delta {R}_{2}^{\Phi }\)) were observed (Fig. 9.10b). Interestingly, these changes were similar across a wide variety of cell types. The results were fitted to Eq. 9.2; \(\Delta {R}_{2}^{\Phi } = {r}_{2}^{\Phi } \cdot {n}_{C}\), where \({r}_{2}^{\Phi }\) is the cellular relaxivity for MNP-\(\Phi \) and n C is the cell concentration (N C ∕ V ). The cellular relaxivities (\({r}_{2}^{\Phi }\)) were statistically identical (p > 0. 99) among different tumor cell lines, suggesting that the method may provide a universal measure for estimating n C .
Using Eq. 9.2 and the cell density information (above), the expression level (ξ) of a select marker was defined as
where \(\Delta {R}_{2}^{\mathit{Ab}}\) and \({r}_{2}^{\mathit{Ab}}\) are R 2 changes and the cellular relaxivity, respectively, with a marker-specific MNP. In this normalized form, ξ now reports the cellular expression level of a targeted marker, providing a way to molecularly profile target cells regardless of cell numbers in a sample. The method was extensively verified by comparing μNMR measurements to other standard methods (flow cytometry, Western blotting) [22, 27, 28]. In one set of experiments, we measured the expression level of HER2 in breast cancer cell lines (Fig. 9.10c). The measured \({\xi }^{\mathit{HER}2}(= \Delta {R}_{2}^{\mathit{HER}2}/\Delta {R}_{2}^{\Phi })\) from μNMR (requiring \(\sim 1{0}^{3}\) cells) showed good agreement (R 2 > 98 %) with both flow cytometry (requiring ∼ 105 cells) and Western blotting (requiring ∼ 107 cells), validating the analytical capability of DMR. Note that DMR detection was much faster ( ∼ 15 min) and performed using > 102 times fewer cells.
6.2 Clinical Trial
Through the integration of the complementary DMR and BOND technologies, this chip-based NMR detection platform has been applied in clinical trials of cancer cell profiling [26]. A total of 50 patients with suspected abdominal malignancies were enrolled. Each patient underwent fine-needle aspiration (FNA) using a 22-G needle, followed by routine core biopsies (17-G needle) for conventional analysis. The FNA samples were aliquoted and profiled for 11 predefined cellular markers: 9 cancer-related markers (EpCAM, MUC-1, HER2, EGFR, B7H3, CK18, Ki-67, p53, Vimentin), CD45 for leukocyte counts, and control (MNP-\(\Phi \)) to measure the total cell density. A priori selection of cancer markers was based on current practice (e.g., EpCAM, CK18) [60, 61] or on reports of clinically relevant overexpression [62–64]. On average, 3,850 cells were obtained per patient via FNA. Of these, approximately one-third were CD45-positive leukocytes (mean: 1,273 cells), and the remaining were nonleukocytic, primarily tumor cells as determined by flow cytometry. The aliquots (containing ∼ 350 cells) were labeled using the BOND method and measured by the clinical μNMR system. The R 2 data with CD45 was used to account for the contribution by leukocytes; for each cancer marker, its cellular expression level (ξ) was obtained based on Eq. 9.3.
Figure 9.11 shows the profiling results of 9 cancer markers for 50 patients. When plotted for each patient, the expression level of the markers showed considerable heterogeneity across samples, which reconfirmed the importance of multiple-marker screening in cancer detection. Indeed, when analyzed for the diagnostic sensitivity, specificity, and accuracy (Table 9.2), a single marker showed accuracy of < 70 %. The highest accuracy for cancer diagnosis with our cohort was obtained with a quadruple marker combination (\(\mathrm{MUC}1 + \mathrm{EGFR} + \mathrm{HER}2 + \mathrm{EpCAM}\), 96 % accuracy), followed closely by weighted triple markers (\(\mathrm{MUC}1 + \mathrm{EGFR} + \mathrm{HER}2\), 92 % accuracy). Interestingly, the EpCAM and CK18 combination, which is routinely used in detecting circulating tumor cells in peripheral blood, achieved an overall diagnostic accuracy of 80 %, wherein high sensitivity (84 %) was offset by low specificity (50 %).
The DMR results were also compared to those by standard-of-care methods. Conventional cytology on FNA specimen was performed in 49 of 50 cases and was diagnostic in 36 cases with 11 misdiagnoses (accuracy 74 %). Conventional histology on all 50 biopsy samples correctly diagnosed 37 cases out of 45 diagnostic samples (accuracy 84 %), with the remaining results nondiagnostic (5 cases). DMR consistently outperformed the other methods, with 2 misdiagnoses in all 50 samples (accuracy up to 96 %). Also note that DMR permitted fast detection (in less than 60 min for each patient), whereas the mean clinical turnaround time (from sample submission to final report) was 3 days for cytology (1–8 days) and 4 days for surgical pathology (1–11 days).
7 Conclusions and Outlook
As a novel biosensing platform, DMR offers many synergistic advantages over traditional approaches, such as high detection sensitivity, multiplexed capability, rapid measurement, and small sample volume requirement with minimal sample processing. Indeed, DMR thrives through the complement of several cutting-edge technologies, namely, magnetic nanomaterials, bioconjugation chemistry, and microfabrication. With new developments such as chip-based μNMR devices, optimized magnetic nanoparticles, and advanced labeling techniques, the DMR technology has proven itself as a robust and sensitive approach for quantitative and molecular analyses for biomedical research. Moreover, our recent clinical trial has confirmed that the DMR is capable of detecting and molecularly profiling cells with minimal false negatives. Its remarkable performance and potential impact on clinical disease management would no doubt accelerate the advance of personalized treatment by providing valuable information on molecular signature of individual patients.
We further envision broader application of the DMR in global healthcare. The DMR technology does not require extensive sample purification and can be packaged as a portable device. The system thus is well suited for rapid and point-of-care (POC) testing, especially in resource-limited primary clinics where majority of diagnoses are made based on physical symptoms only. Indeed, we plan to evaluate the developed system for TB (tuberculosis) detection in fields. Successful completion of this research will be a cornerstone for realizing POC technology for TB detection, which will bring significant societal benefits worldwide.
References
R. Etzioni et al., The case for early detection. Nat. Rev. Cancer 3, 243–252 (2003)
R. Fan et al., Integrated barcode chips for rapid, multiplexed analysis of proteins in microliter quantities of blood. Nat. Biotechnol. 26, 1373–1378 (2008)
D.A. Giljohann, and C.A. Mirkin, Drivers of biodiagnostic development. Nature 462, 461–464 (2009)
M.M. Cheng et al., Nanotechnologies for biomolecular detection and medical diagnostics. Curr. Opin. Chem. Biol. 10, 11–19 (2006)
A.G. Tibbe et al., Optical tracking and detection of immunomagnetically selected and aligned cells. Nat. Biotechnol. 17, 1210–1213 (1999) 1999.
M.M. Wang et al., Microfluidic sorting of mammalian cells by optical force switching. Nat. Biotechnol. 23, 83–87 (2005)
E. Stern et al., Label-free immunodetection with CMOS-compatible semiconducting nanowires. Nature 445, 519–522 (2007)
G. Zheng et al., Multiplexed electrical detection of cancer markers with nanowire sensor arrays. Nat. Biotechnol. 23, 1294–1301 (2005)
S.C. Bendall et al., Single-cell mass cytometry of differential immune and drug responses across a human hematopoietic continuum. Science 332, 687–696 (2011)
J. Cheon, and J.H. Lee, Synergistically integrated nanoparticles as multimodal probes for nanobiotechnology. Acc. Chem. Res. 41, 1630–1640 (2008)
N.A. Frey et al., Magnetic nanoparticles: synthesis, functionalization, and applications in bioimaging and magnetic energy storage. Chem. Soc. Rev. 38, 2532–2542 (2009)
Y.W. Jun, J.W. Seo, and J. Cheon, Nanoscaling laws of magnetic nanoparticles and their applicabilities in biomedical sciences. Acc. Chem. Res. 41, 179–189 (2008)
Q.A. Pankhurst et al., Applications of magnetic nanoparticles in biomedicine. J. Phys. D: Appl. Phys. 36, 167–181 (2003)
Y.R. Chemla et al., Ultrasensitive magnetic biosensor for homogeneous immunoassay. Proc. Natl. Acad. Sci. U.S.A. 97, 14268–14272 (2000)
R.L. Millen et al., Giant magenetoresistive sensors. 2. Detection of biorecognition events at self-referencing and magnetically tagged arrays. Anal. Chem. 80, 7940–7946 (2008)
V. Schaller et al., Towards an electrowetting-based digital microfluidic platform for magnetic immunoassays. Lab Chip 9, 3433–3436 (2009)
D.R. Baselt et al., A biosensor based on magnetoresistance technology. Biosens. Bioelectron. 13, 731–739 (1998)
R.S. Gaster et al., Matrix-insensitive protein assays push the limits of biosensors in medicine. Nat. Med. 15, 1327–1332 (2009)
R.L. Millen et al., Giant magnetoresistive sensors and superparamagnetic nanoparticles: a chip-scale detection strategy for immunosorbent assays. Anal. Chem. 77, 6581–6587 (2005)
S.J. Osterfeld et al., Multiplex protein assays based on real-time magnetic nanotag sensing. Proc. Natl. Acad. Sci. U.S.A. 105, 20637–20640 (2008)
T. Aytur et al., A novel magnetic bead bioassay platform using a microchip-based sensor for infectious disease diagnosis. J. Immunol. Methods 314, 21–29 (2006)
H. Lee et al., Chip-NMR biosensor for detection and molecular analysis of cells. Nat. Med. 14, 869–874 (2008)
M. Gueron, Nuclear-relaxation in macromolecules by paramagnetic-ions – novel mechanism. J. Magn. Reson. 19, 58–66 (1975)
J.M. Perez et al., DNA-based magnetic nanoparticle assembly acts as a magnetic relaxation nanoswitch allowing screening of DNA-cleaving agents. J. Am. Chem. Soc. 124, 2856–2857 (2002)
H. Lee, T.J. Yoon, and R. Weissleder, Ultrasensitive detection of bacteria using core-shell nanoparticles and an NMR-filter system. Angew. Chem. Int. Ed Engl. 48, 5657–5660 (2009)
J.B. Haun et al., Micro-NMR for rapid molecular analysis of human tumor samples. Sci. Transl. Med. 3, 71ra16 (2011)
J.B. Haun et al., Bioorthogonal chemistry amplifies nanoparticle binding and enhances the sensitivity of cell detection. Nat. Nanotechnol. 5, 660–665 (2010)
H. Lee et al., Rapid detection and profiling of cancer cells in fine-needle aspirates. Proc. Natl. Acad. Sci. U.S.A. 106, 12459–12464 (2009)
M. Liong et al., Multiplexed magnetic labeling amplification using oligonucleotide hybridization. Adv. Mater. 23, H254–H257 (2011)
P. Gillis, and S.H. Koenig, Transverse relaxation of solvent protons induced by magnetized spheres: application to ferritin, erythrocytes, and magnetite. Magn. Reson. Med. 5, 323–345 (1987)
Y. Gossuin et al., Magnetic resonance relaxation properties of superparamagnetic particles. Wiley Interdiscip. Rev. Nanomed. Nanobiotechnol. 1, 299–310 (2009)
R.A. Brooks, F. Moiny, and P. Gillis, On T2-shortening by weakly magnetized particles: the chemical exchange model. Magn. Reson. Med. 45, 1014–1020 (2001)
P. Gillis, F. Moiny, and R.A. Brooks, On T(2)-shortening by strongly magnetized spheres: a partial refocusing model. Magn. Reson. Med. 47, 257–263 (2002)
C. Massin et al., High-Q factor RF planar microcoils for micro-scale NMR spectroscopy. Sens. Actuat. A 97–98, 280–288 (2002)
R.A. Brooks, T(2)-shortening by strongly magnetized spheres: a chemical exchange model. Magn. Reson. Med. 47, 388–391 (2002)
J.M. Perez et al., Magnetic relaxation switches capable of sensing molecular interactions. Nat. Biotechnol. 20, 816–820 (2002)
L. Josephson et al., High-efficiency intracellular magnetic labeling with novel superparamagnetic-Tat peptide conjugates. Bioconjug. Chem. 10, 186–191 (1999)
J.H. Lee et al., Artificially engineered magnetic nanoparticles for ultra-sensitive molecular imaging. Nat. Med. 13, 95–99 (2007)
S. Sun et al., Monodisperse MFe2O4 (M = Fe, Co, Mn) nanoparticles. J. Am. Chem. Soc. 126, 273–279 (2004)
Y.W. Jun et al., Nanoscale size effect of magnetic nanocrystals and their utilization for cancer diagnosis via magnetic resonance imaging. J. Am. Chem. Soc. 127, 5732–5733 (2005)
S. Peng et al., Synthesis and stabilization of monodisperse Fe nanoparticles. J. Am. Chem. Soc. 128, 10676–10677 (2006)
T.J. Yoon et al., Highly magnetic core-shell nanoparticles with a unique magnetization mechanism. Angew. Chem. Int. Ed Engl. 50, 4663–4666 (2011)
J. Park et al., Ultra-large-scale syntheses of monodisperse nanocrystals. Nat. Mater. 3, 891–895 (2004)
Z. Chen et al., Preparation and characterization of water-soluble monodisperse magnetic iron oxide nanoparticles via surface double-exchange with DMSA. Colloid. Surf. A 316, 210–216 (2008)
N. Fauconnier et al., Thiolation of maghemite nanoparticles by dimercaptosuccinic acid. J. Colloid Interface Sci. 194, 427–433 (1997)
M. Liong et al., Carboxymethylated polyvinyl alcohol stabilizes doped ferrofluids for biological applications. Adv. Mater. 22, 5168–5172 (2010)
J. Panyam, and V. Labhasetwar, Biodegradable nanoparticles for drug and gene delivery to cells and tissue. Adv. Drug Deliv. Rev. 55, 329–347 (2003)
N.A. Peppas, and R.E.J. Benner, Proposed method of intracordal injection and gelation of poly (vinyl alcohol) solution in vocal cords: polymer considerations. Biomaterials 1, 158–162 (1980)
K.M. Rosenblatt, and H. Bunjes, Poly(vinyl alcohol) as emulsifier stabilizes solid triglyceride drug carrier nanoparticles in the alpha-modification. Mol. Pharm. 6, 105–120 (2009)
N.K. Devaraj et al., Fast and sensitive pretargeted labeling of cancer cells through a tetrazine/trans-cyclooctene cycloaddition. Angew. Chem. Int. Ed Engl. 48, 7013–7016 (2009)
M.R. Karver, R. Weissleder, and S.A. Hilderbrand, Synthesis and evaluation of a series of 1,2,4,5-tetrazines for bioorthogonal conjugation. Bioconjug. Chem. 16, 2263–2270 (2011)
S.S. Agasti et al., Supramolecular host-guest interaction for labeling and detection of cellular biomarkers. Angew Chem Int Ed Engl 51, 450–454 (2012)
A.G. Webb, Radiofrequency microcoils in magnetic resonance. Prog. Nucl. Magn. Reson. Spectrosc. 31, 1–42 (1997)
J. Grimm et al., Novel nanosensors for rapid analysis of telomerase activity. Cancer Res. 64, 639–643 (2004)
C. Kaittanis, S.A. Naser, and J.M. Perez, One-step, nanoparticle-mediated bacterial detection with magnetic relaxation. Nano Lett. 7, 380–383 (2007)
J.M. Perez, L. Josephson, and R. Weissleder, Use of magnetic nanoparticles as nanosensors to probe for molecular interactions. Chembiochem 5, 261–264 (2004)
D. Issadore et al., Miniature magnetic resonance system for point-of-care diagnostics. Lab Chip 11, 2282–2287 (2011)
E. Danieli et al., Small magnets for portable NMR spectrometers. Angew. Chem. Int. Ed Engl. 49, 4133–4135 (2010)
A. Moore, R. Weissleder, and A.J. Bogdanov, Uptake of dextran-coated monocrystalline iron oxides in tumor cells and macrophages. J. Magn. Reson. Imaging 7, 1140–1145 (1997)
S. Nagrath et al., Isolation of rare circulating tumour cells in cancer patients by microchip technology. Nature 450, 1235–1239 (2007)
L.V. Sequist et al., The CTC-chip: an exciting new tool to detect circulating tumor cells in lung cancer patients. J. Thorac. Oncol. 4, 281–283 (2009)
S.B. Ho et al., Heterogeneity of mucin gene expression in normal and neoplastic tissues. Cancer Res. 53, 641–651 (1993)
T.J. Roth et al., B7-H3 ligand expression by prostate cancer: a novel marker of prognosis and potential target for therapy. Cancer Res. 67, 7893–7900 (2007)
C.L. Vogel et al., Efficacy and safety of trastuzumab as a single agent in first-line treatment of HER2-overexpressing metastatic breast cancer. J. Clin. Oncol. 20, 719–726 (2002)
Acknowledgements
We thank T. J. Yoon, J. H. Chung, J. B. Haun, and N. Sergeyev for assistance with experiments and R. M. Westervelt (Harvard) and C. Castro (MGH) for many helpful discussions. This work was supported in part by the following NIH grants: U54CA151884, R01EB010011, R01EB004626, P01CA069246, P50CA86355, U01CA141556, U24CA092782, and R21CA14122. H. Shao acknowledges financial support from the B.S.-Ph.D. National Science Scholarship awarded by the Agency for Science, Technology and Research, Singapore.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2013 Springer-Verlag Berlin Heidelberg
About this chapter
Cite this chapter
Min, C., Shao, H., Issadore, D., Liong, M., Weissleder, R., Lee, H. (2013). Diagnostic Magnetic Resonance Technology. In: Issadore, D., Westervelt, R. (eds) Point-of-Care Diagnostics on a Chip. Biological and Medical Physics, Biomedical Engineering. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-29268-2_9
Download citation
DOI: https://doi.org/10.1007/978-3-642-29268-2_9
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-29267-5
Online ISBN: 978-3-642-29268-2
eBook Packages: Physics and AstronomyPhysics and Astronomy (R0)