Keywords

1 Introduction

In contrast to typical invasive laboratory testing of the blood sample, self-monitoring of glucose level provides a current scenario of diabetes with a continuous, dependable, and reliable approach for determining blood glucose concentration. Checking glucose levels frequently using invasive methods is a critical process for the treatment of diabetes. This invasive technique assists patients in preventing and detecting hypoglycaemic and hyperglycaemic conditions. Many commercial blood glucose monitors present in the market require a small drop of blood to be taken by pricking the skin with a lancet, which is generally a fingertip (often called a fingertip test). These are the intrusive blood glucose monitors that cause the patients a huge amount of discomfort because they are likely to be punctured several times a day for checking their blood glucose level.

Depending upon the amount and condition of the sample, human inaccuracy during sample collections, calibration errors, humidity, and poor cleanliness in the testing region, might increase the proportion of mistakes while using invasive monitoring techniques [1].

The non-invasive approach, a recent technology, requires glucose signals from the body. Non-invasive methods avoid “finger pricking” and help continuously monitor blood glucose levels. Techniques such as applying fluorescent light to the body in a specific location, in addition to the techniques involving the implantation of a sensor in the subcutaneous tissue, lead to a major disadvantage: the interference of nearby signals such as ultraviolet and visible light with the process.

Another approach that uses optical beams to detect glucose signals is acoustic spectroscopy. However, this process suffers from scattering effects, resulting in insensitivity. Near-Infrared (NIR) spectroscopy is also one technique that can also be used to measure the level of glucose in blood [2]. The level of glucose can also be evaluated with the help of multi-modal spectroscopy IC which combines impedance spectroscopy (IMPS) and multi-wavelength near-infrared spectroscopy (fNIRS) [1]. In Fig. 1 the flowchart of different blood glucose measurement techniques is shown.

Fig. 1
figure 1

Flowchart of blood glucose monitoring techniques

This technology employs indirect dielectric properties of the tissue surrounding the blood and the precision of the glucose levels obtained by the suggested microchip to eliminate various systemic sounds.

A recent method for identifying diabetes is to use the Gabor filter to analyze face contour features [3]. Human respiration can potentially be used to diagnose diabetes. This indicates a good link with blood sugar since human breath includes acetone, which can be analyzed simply by exhaling it directly into the monitoring device [4]. In Fig. 2, the non-invasive glucose monitoring techniques are shown.

Fig. 2
figure 2

Reproduced with permission from [5] CC By © 2019 by Gonzales et al., Licensee MDPI, Basel, Switzerland

Different non-invasive glucose monitoring.

2 NIR Spectroscopy

The spectroscopical region of 12,500–4000 cm−1 belongs to Near-infrared (NIR) region, and its wavelength is 800–2500 nm. Both wavenumber and wavelength are being considered in the present chapter. The region in-between IR and the visible region is NIR spectroscopical region. Generally, NIR spectroscopy deals with reflection, emission, diffusion, and light absorption. NIR spectroscopy has advanced characteristics that have played a significant role in basic and applied science applications over the last two decades [6]. In the spectral range of near-infrared, the ability of light to penetrate soft tissues and biofluids is relatively high (>0.5 mm) as, compared to visible light and ultraviolet light, it scatters less. Apart from this, this technique can achieve both reflection and transmission for the sensing measurement of light [7]. The mathematical formulation for calculating sample absorption is done by Beer-Lambert law (Eq. 1) from the already known thickness and concentration.

$$ I = I_o 10^{\left( { - l.\varepsilon .c} \right)} = I_o e^{ - l.\mu_a } $$
(1)

where I is the intensity of light at depth within the absorption medium W/cm2, initial light intensity is represented by I0 (W/cm2), absorption depth is represented by l, ε is the molar attenuation coefficient or molar extinction coefficient (L/mmol cm), it depends on the structure of absorbing molecules and the wavelength of incident light. Absorbing molecules concentration is represented by c (mmol/L). The absorption coefficient μa is proportional to the product of c and ε.

This model showcases the transmitted/reflected intensity of light as a function of concentration, the thickness of the sample, and absorption coefficient. In contrast, the scattered light effect is ignored in this phenomenon. Log(I0/I) is used to define absorbance [8]. NIR absorption spectroscopy can quantify the glucose absorbance and its dependence on wavelength in the aqueous medium. Incident light absorption by water must be considered as it is the most abundant species in biofluids. Two absorption peaks, one between 1350 and 1520 nm and the other in-between 1790 and 2000 nm, are revealed in the NIR spectrum range for water. To measure glucose in the NIR range, wavelength windows in the range of 700–1100 nm, 1500–1850 nm, and 2000–2400 nm can be used [9,10,11].

On the contrary, light absorption in a shorter wavelength range is lower for water. So to obtain selective NIR spectroscopic results with the minimization of interfering effects of water, one should essentially use shorter wavelengths [7, 12]. Figure 3 shows the spectra recording equipment. The equipment consists of a NIR spectrometer, a light source, and a fiber optical measuring head. In NIR spectrometer, a 128 pixels InGaAs photodiode array detector is attached to a glass block, and it uses a polychromator with a holographic imaging diffraction grating. According to the modified Beer's equation, near-infrared diffused reflection difference spectra are obtained at the skin tissue to constantly forecast the blood glucose content, which is proposed and investigated without multivariate analyses. The difference spectra are presumed to be generated from four primary elements in the human skin (glucose, protein, fat, and water) and a scattering equivalent component called the baseline. As a result, the morphological similarity of the absorption spectrum between glucose and baseline is one of the roots of inaccuracies in predicting blood glucose levels in the near-infrared region. An artificial component integrated with baseline and fat is revealed when extracting the glucose components from the distinction spectra at baseline using fat's specific wavelength. It is predicated on the notion that a change in skin scattering induces both the variation in fat contribution and leads to baseline development.

Fig. 3
figure 3

Reproduced with permission from [5] CC By © 2019 by Gonzales et al., Licensee MDPI, Basel, Switzerland

Schematic representation of the three modes of NIR spectroscopy. a Transmittance mode. b Reflectance mode. c Interactance mode.

We can reduce the blood glucose prediction mistakes by using the imaginary component. In contrast to multivariate analysis methods, the estimation procedure of blood glucose substances from observed reflection spectra is transparent, making it easier to evaluate the causes for fluctuations and contributions of the components in the observed reflection spectra. In Fig. 3, the typical blood glucose level profile is shown. Using radial-basis neutral networks (RBF) and partial least-squares regression (PLS), the calibration coefficient of matrices is calculated.

Glucose measurements are interfered with by chemical and physical parameters such as albumin, triglyceride, temperature, and pressure variation. Environmental changes such as variations in humidity, carbon dioxide, atmospheric pressure, skin hydration, and temperature also cause errors. The proposed technique may become a convenient tool for interpreting non-invasive blood glucose monitoring using near-infrared spectroscopy [13].

Rolamjaya Hotmartua et al. used NIR spectroscopy for Glucose detection from the earlobe. They first carried out the test by varying glucose concentration from 50 mg/dl to 2 g/dl, and 2nd test was carried out directly in the earlobe [14]. Samman et al. researched glucose monitoring for about 169 days and analyzed the accuracy as well as stability of the calibration method [15]. In brief, it can be said that the application of NIR spectroscopy in combination with regression analysis helps quickly to detect the glucose levels in a non-invasive manner, thereby providing an additional tool for the early detection of diabetic problems [9].

3 Raman Spectroscopy

The degree of scattering of monochromatic light is determined by Raman spectroscopy which is based on the Raman effect. Scattered light that travels in all directions is produced when a single wavelength light hits the target. Using the imaginary component can reduce the error in predicting the blood sugar level. Unlike multivariate analysis, this method makes it clearer to calculate blood sugar levels from the observed reflectance spectrum, making it easier to estimate the reason for the change and the component contribution in the reflectance spectrum. Rayleigh scattering occurs in visible radiation when the scattering is elastic. The phenomenon is known as Raman scattering if the scattering is not elastic [16]. The schematic representation of Raman scattering is shown in Fig. 4.

Fig. 4
figure 4

Reproduced with permission from [5] CC By © 2019 by Gonzales et al., Licensee MDPI, Basel, Switzerland

Schematic representation of Raman spectroscopy.

Raman shift is termed for such wavelength differences. The Raman shift represents the difference between the initial and final vibrational states of molecules under study [17]. The vibrational and rotational states of molecules are dependent on Raman spectroscopy. The functional group’s vibrational modes are shown in the peaks of Raman spectra. It consists of a lens that seizes the scattered radiation. It also filters the radiation and allows the Raman scattered radiation to the detector for getting sensed. The signals are processed by computer and provide Raman shift correspondingly.

Raman spectroscopy is a favorable non-invasive biomedical method and analyses the problems associated with metabolism. For example, clinical tests and self-monitoring of glucose levels based on a finger prick technique blood sample are no longer painful. Raman spectroscopy is based on the elastic scattering of photons by certain molecules in the sample. The energy shift of the scattered photons is determined by the bonds of the interacting molecule, thereby resulting in molecular fingerprints. The advantages and limitations of Raman scattering for glucose monitoring are shown in Table 1.

Table 1 The pros and cons of glucose detection using Raman spectroscopy

It also has important benefits in biomedical diagnostics, including non-invasiveness, a short procurement time, and the capability to offer quick results. It has been shown that Raman spectroscopy and principal component analysis (PCA) paired with support vector machine (SVM) has proven to be effective in classifying glycated hemoglobin levels using in vivo techniques [18].

4 Bio-impedance Spectroscopy

The bio-impedance analysis is a non-invasive, low-cost, and widely utilized method for determining body composition and assessing clinical status. There are numerous ways of interpreting the measured bio-impedance data. Apart from that, there are numerous applications of bio-impedance in the assessment of body composition and the evaluation of clinical status. In addition, bio-impedance spectroscopy is being used in various ways in healthcare institutions, including disease prognosis and monitoring of vital signs. Thus, we feel that this warrants a review of the most underlying facets and forecasts healthcare applications of bio-impedance spectroscopy [19].

The bio-impedance analysis is a low-cost, non-invasive method for evaluating body composition and helps in monitoring of clinical state. There are varieties of techniques to interpret measured bio-impedance data and a variety of bio-impedance applications in body composition assessment and clinical status assessment.

A wrist glucose monitor based on impedance spectroscopy has been invented. It is shown in Figs. 5 and 6. This monitor uses the skin as a dielectric to sample an LC resonance circuit data. There are numerous approaches for interpreting measured bio-impedance data and bio-impedance applications in body composition and clinical state assessment. The drawbacks and advantages are shown in Table 2. Bio-impedance is also used in healthcare facilities for various applications, such as disease prognosis and vital sign monitoring [20].

Fig. 5
figure 5

Reproduced with permission from [5] CC By © 2019 Gonzales et al., Licensee MDPI, Basel, Switzerland

Graphical abstract of glucose detection using bioimpedance spectroscopy.

Fig. 6
figure 6

Bioimpedance model measurement and architecture of gain-phase detector

Table 2 Advantages and disadvantages of glucose detection using bioimpedance spectroscopy

5 Thermal Emission Spectroscopy

Thermal Emission Spectroscopy (TES)—the based device is a novel, non-invasive hand-held BG monitor having the same dimensions and ease of use as an ear thermometer. Still, this technique is with more technological breakthroughs. The gadget is passive, and, in this technique, chronic external radiation does not injure the human tissues.

This gadget will help individuals with diabetes to live better lives since it will boost patient BG testing, which will lead to improved glycaemic management and lesser complications related to diabetes. Thermal emission spectroscopy (TES) detects infrared signals produced by changes in glucose content in the human body. According to this technique, the natural mid-infrared emission of the human body, particularly the eardrum, is regulated by the state of the emitting tissue. Its selectivity is based on the same premise as the absorption spectroscopy technique used for measurement analysis [21]. Figure 7 represents the pictorial representation of the principle of thermal emission spectroscopy for glucose detection.

Fig. 7
figure 7

Reproduced with permission from [5] CC By © 2019 by Gonzales et al., Licensee MDPI, Basel, Switzerland

Principle of thermal emission spectroscopy.

The detection of glucose level using a non-invasive prototype based on thermal emission in the mid-IR spectral area was measured satisfactorily. Individual daily calibrations are unnecessary with this technology, which is one of its benefits [21]. One disadvantage of this technique is that the strength of infrared radiation emitted by an eardrum is influenced by its thickness and temperature. The other drawbacks and advantages of this method for glucose detection are shown in Table 3. Although the clinical outcomes acquired with TES are encouraging, they do not yet meet clinical accuracy criteria [22].

Table 3 Pros and cons of glucose detection using thermal emission spectroscopy

6 Optical Polarimetry

Optical polarimetry works on the principle of chiral molecules, i.e., the plane of polarization is rotated by the molecules. The chiral molecule glucose can rotate the plane of polarisation of a beam of light by an angle α in the clockwise direction. This glucose detection method is one of the basic techniques of non-invasive technology. When the beam of polarized light is incident on the glucose solution, the plane of polarization of the incident light is rotated by the presence of glucose.

At present, the sum of glucose level is comparative to the directional angle concerning the original incident direction formed by the polarisation direction [23,24,25,26,27]. The rotation amount is proportional to the optical path length, the laser beam wavelength, the temperature, and the analyte concentration.

This wavelength will usually appear in the NIR upper region and the optical band is in the lower region (approximately 780–400 nm). The polarimeter measures the plane of polarised light when it passes through the sample. The pictorial representation of optical polarimetry is shown in Fig. 8. The maximum intensity of light is identified by a photodetector when the electric field rotation is comparable with the polarization axis of the analyzer. The photodetector cannot detect light when the polarization angle is perpendicular to the electric field rotation angle [28, 29].

Fig. 8
figure 8

Reproduced with permission from [5] CC By © 2019 by Gonzales et al., Licensee MDPI, Basel, Switzerland

Schematic representation of optical polarimetry in the eye for glucose detection.

The application of optical polarimetry in the skin is not feasible as the scattering of light in tissue and skin is high, decreasing the glucose's optical rotation [30]. The optical polarimetry can be used in the eyes anterior chamber as it has a good optical property [30, 31]. Eye glucose monitoring is shown in Fig. 8. For this method, a satisfied error accuracy is not obtained even though this method can be detected by visible light, easy operation, and obtaining results [32,33,34]. Currently, there are two new mechanisms for resolving the interference issues of the birefringence of the polarization of glucose overlooked in blood. Dual-wavelength polarizer and birefringence compensator are the two mechanisms used to resolve this issue. Dual-wavelength polarimetric glucose detection is shown in Fig. 9.

Fig. 9
figure 9

Dual-wavelength polarimetric glucose detection

7 Fluorescence

Fluorescence technology is associated with the principle of fluorescent light emission. This fluorescent light emission results in stock shift, wavelength difference due to the effect of the light emission.

Unlike optical approaches, this approach of utilizing fluorescence requires contact between sample and sensor. Recently, new techniques have been developed for analyzing glucose and vitamin A using field instruments. Pickup et al. [47] studied glucose monitoring using the intrinsic tissue fluorescence. This technique utilizes the product of glucose metabolism, nicotinamide adenine dinucleotide phosphate (reduced) (NAD(P)H), a fluorescent cofactor. The test was carried out in in-vivo analysis, and later they were studied in vitro. The fluorescence sensitivity is highly required as the glucose concentration in the fluid is in the micromolar range.

The fluorescence techniques can be categorized into affinity-binding and glucose oxidase (GOx) based sensors. GOx based sensors have a certain limitation as the results depend not only on the concentration of glucose but also on the oxygen tension under in-vivo conditions.

Fluorophores, a specialized molecule that release fluorescent light with certain characteristics, are proportional to the analyte concentration under study. Some fluorophore molecules can bind with glucose molecules directly, but they are associated with interference, irreversibility, analyte depletion, and low selectivity. Therefore, the use of intermediatory molecules like receptors binds with glucose molecules more effectively, leading to changes in their local properties reversibly. This results in fluorescence alteration [67]. Different types of receptors are used for this such as boronic acid derivatives, enzymes, and glucose binding proteins, which are nature-derived and fabricated synthetic materials such as quantum dots and carbon nanotubes.

Fluorescence resonant energy transfer (FRET) has gained much attention based on binding assays. The energy transfer in this technique occurs between the donor and acceptor molecules, the light-sensitive molecules. When the acceptor molecules bind with glucose, the bond between acceptor and donor gets disrupted, increasing fluorescence due to less electron sharing. This is shown in Fig. 10.

Fig. 10
figure 10

Fluorescence glucose monitoring

The FRET and fluorescence lifetime techniques were used to measure glucose concentration by Lakowicz and co-workers [50]. They used reagents such as maltose-insulin-malachite green and ruthenium-Con A to fabricate a similar sensor. The fluorescence lifetime and fluorescence intensity of ruthenium dye is increased with an increase in glucose concentration. The advantage of this type of fluorescence sensor is that they are highly specific and highly sensitive to the analyte under study. Apart from that, it also eliminates the potential interferences. The drawback of this technique is the short lifespan of the fluorophore, foreign materials in biological media which can cause potential toxicity, etc.

There are so many other techniques that can non-invasively detect glucose concentration, such as Metabolic Heat Conformation (MHC), Photoacoustic Spectroscopy (PAS), Optical Coherence Tomography (OCT), Millimeter and Microwave Sensing, etc. Table 4 shows the list of recent research developments on MI and NI techniques in Glucose detection. Table 5 compares non-invasive and minimally invasive glucose monitoring devices that are currently accessible or about to be released in the market soon. Figure 11 depicts the position of non-invasive and minimally invasive methods and devices for monitoring glucose in the spectrum.

Table 4 List of recent research on MI and NI techniques for monitoring glucose levels
Table 5 Comparison table of non-invasive and minimally invasive glucose monitoring devices
Fig. 11
figure 11

Reproduced with permission from [5] CC By © 2019 Gonzales et al., Licensee MDPI, Basel, Switzerland

The chart with location of non-invasive and minimally invasive methods and devices for glucose in the frequency spectrum.

8 Conclusion

The scientific advancement of non-invasive glucose monitoring technologies in recent years is discussed in this chapter. Optical, microwave, and electrochemical methods are the three types of non-invasive blood glucose monitoring technologies available. The advantages of optical and microwave technologies, in general, are their non-invasive nature and ability to monitor continuously without causing discomfort to the human body. In terms of detection, there are still certain issues, such as sophisticated detection means, harsh detecting components, a time-consuming detection procedure, high detection equipment needs, and significant background signal interference. Future research will see if other physical characteristics and other indicators linked to blood glucose may be integrated to increase the accuracy of non-invasive skin glucose test results.