Abstract
Bearing surfaces for joint replacements have undergone many recent changes. The traditional metal-on-conventional polyethylene bearings, which were associated with a chronic inflammatory response to wear debris and osteolysis, have been replaced in many clinical scenarios, by highly cross-lined polyethylene, new ceramics, and metal-on-metal articulations. Each bearing couple has known advantages and disadvantages. The overall goal is to provide the patient with a long-lasting articulation that will allow full work and recreational activities, with minimal/no adverse events. Although we have not yet achieved this goal, highly cross-lined polyethylene and ceramic bearing couples in particular have facilitated new approaches to joint replacement that have demonstrated increased durability and improved patient outcomes.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
- Ultra High Molecular Weight Polyethylene
- Wear Particle
- Periprosthetic Osteolysis
- Large Femoral Head
- Bearing Couple
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
Total joint replacement (TJR) is one of the greatest technological advances in all of surgery. Hip , knee, and shoulder replacements, as well as reconstruction of smaller joints with artificial materials are currently performed worldwide. These procedures decrease pain and improve function in a cost-effective manner, and thereby improve the quality of life for millions of patients with end-stage arthritis.
Initially, most modern TJRs consisted of a bearing couple composed of a metallic alloy that articulated with conventional medical grade polyethylene [1]. This combination of materials functioned satisfactorily for many years in low demand, elderly patients for whom TJRs were originally designed. However, as joint replacement procedures were extended to younger more active higher-demand patients, wear of the polyethylene and the subsequent adverse biological reaction to wear byproducts became a serious concern [2].
Wear of the bearing materials of a TJR is a function of use, not time in vivo [3]. Higher-demand patients engage in greater numbers of gait cycles per day, and often participate in higher-impact sporting activities that increase wear [4]. Polyethylene wear particles generated at the articulation are pumped and distributed throughout the “effective joint space”, producing in some cases chronic synovitis, progressive bone loss (periprosthetic osteolysis), implant loosening and pathologic fracture [5]. Subsequent surgical reconstruction of loose TJRs with extensive periprosthetic bone loss is challenging; these surgical procedures are long and costly and have a higher complication rate and a poorer outcome compared to primary procedures [2]. These facts have stimulated intense research to improve the tribological characteristics of current materials, as well as develop newer more wear resistant bearing couples that potentially could last a lifetime [6]. Although this goal has not yet been realized, significant improvements in implant materials have been achieved in the last two decades. At the same time, unexpected obstacles have surfaced which have led, in some cases, to earlier revision surgery than with conventional materials.
The Inflammatory Reaction to Wear Debris
Wear particles are generated at all artificial joint articulations. These particles are largely in the micron and submicron range, with metallic particles being amongst the smallest [7–9]. Wear particles of polymethylmethacrylate (PMMA), polyethylene (PE) and ceramics evoke a nonspecific, non-antigenic chronic inflammatory and foreign body reaction [10]. The cellular components of this reaction commonly include the monocyte/macrophage cell lineage (macrophages, foreign body giant cells and osteoclasts), activated fibroblasts, with occasional polymorphonuclear leukocytes (PMNs) and lymphocytes [11–13]. Larger wear particles of metals such as stainless steel, cobalt chrome alloy and titanium alloy incite a similar chronic inflammatory reaction; however, recent evidence has demonstrated that metal byproducts may also produce a Type IV allergic reaction in some situations (see below) [14].
Macrophages and other cells phagocytize particles less than about 10 microns in diameter, as part of the innate immune response to foreign materials [2, 7, 8, 13, 15]. The wear debris is non-digestible and activates the cells to produce and release pro-inflammatory cytokines, chemokines, prostanoids, reactive oxygen species and other factors that, in the end, stimulate osteoclasts to degrade bone [15–17]. At the same time, homeostatic mechanisms are initiated that induce local bone formation [13, 18]. However, with ongoing production of wear debris, the balance between bone destruction and bone formation favours the former, leading to periprosthetic osteolysis, and potentially, implant loosening and fracture [18, 19]. Because of the cyclic nature of walking which induces high intra-articular pressures, the particles, cells and inflammatory factors are pumped and distributed around the prosthesis and insinuate into the adjacent cancellous bone along the bone–implant interface [20]. From this pumping and distribution, osteolysis can be seen adjacent to and remotely from the prosthesis bearing couple . Increased local fluid pressure also induces bone destruction [21]. The cells that phagocytize particles eventually die, liberating the particulate debris that continues to perpetuate the inflammatory cycle. Furthermore, recent in vivo studies have shown that wear particles induce a systemic biological response, rather than only a local response [22, 23]. Through the action of chemotactic cytokines or chemokines, inflammatory and reparative cells are mobilized to the site of particle generation to participate in the inflammatory cascade, attempt to contain this adverse reaction, and restore normal tissue architecture [22–27] .
Although biological approaches are currently being explored to improve the osseointegration of implants (to provide a more robust bone–implant interface) and to mitigate wear particle induced inflammation, perhaps a more direct approach is to develop more wear resistant materials. In essence this amounts providing bearing couples that generate fewer wear particles , with conceivably more benign biological physico-chemical properties, which will not perturb local tissue homeostasis. This goal would aim to provide a “permanent” joint replacement that would allow full activities (including impact loading) for the duration of the patient’s life .
New Polyethylenes
As stated above, metal-on-conventional ultra high molecular weight polyethylene has been the traditional bearing surface for many decades. This material has performed well in the very elderly, more sedentary population. However, in more active younger individuals with greater numbers of gait cycles per year, more wear particles are produced [3]. In general, polyethylene linear wear rates of less than 0.1 mm per year produce little osteolysis compared with higher wear rates [28]. Increased wear is produced by chain scission and oxidation of the linearly arranged polyethylene molecules. Recent attempts to improve the wear characteristics of polyethylene have included: altering the crystallinity of polyethylene, irradiating and packaging the product in an inert (non-oxygen containing) environment, irradiating and heating (above the melting point) and/or annealing the polyethylene to induce a more highly cross-linked end product that contains fewer free radicals, sequential irradiating and annealing protocols below the melting point of polyethylene , and adding surface coatings or free radical scavengers [29–31]. Although most of these new processes have shown highly encouraging early and intermediate clinical results after more than one decade of use, no long-term (20 + year) clinical outcomes have been reported [32]. Cross-linked polyethylene (XLPE) has less optimal mechanical properties (including toughness, ductility and resistance to fatigue) compared to conventional polyethylene [33–35]. Issues related to the use of larger femoral heads (to prevent dislocation) that articulate with thinner polyethylene acetabular liners have lead to reports of polyethylene rim fractures, necessitating revision surgery [36, 37]. This has been seen more commonly in implants with suboptimal positioning (for example, an excessively abducted or anteverted acetabular cup). Although in vitro studies have suggested potentially higher adverse biological reactions to wear particles from cross-linked polyethylene, compared to conventional polyethylene, the numbers of particles generated are decreased with the XLPE material as to almost negate this point [38–40]. However, not all XLPEs are exactly alike. The irradiation protocols, processing, packaging and other variables are different for each manufacturer [32]. Patients with XLPE components are still not encouraged to engage in impact loading activities that could damage the articular surface.
Ceramic Bearings
The use of ceramic-on-ceramic (CoC) bearings was popularized in France, Japan and Korea, but has been less popular in the United States. These bearings are biocompatible, display low friction, high-wear resistance and produce few wear particles with normal usage [41]. Intermediate term series have reported very encouraging results [42, 43]. The problem of catastrophic fracture of ceramic femoral heads in total hip replacement has largely been avoided with newer ceramics with smaller grain sizes. However, some new unanticipated problems have come to light with CoC bearings [44, 45]. Modular acetabular cups may be difficult to assemble, may seat incompletely, or dissociate from their metal backing. Third body interposition (with soft tissue, bone spicules, etc.) between modular components may be an issue in assembly. Chipping of the liner may also occur at surgery or with later impingement. Edge loading with striped wear may take place due to increased range of motion and cyclic micro-separation during gait, especially if the components are in suboptimal position [41, 44]. Troublesome and embarrassing audible squeaking has been noted with some implant designs. In addition, these implants are generally more expensive than metal-on-polyethylene (MoP) articulations. Nonetheless, CoC bearings facilitate the use of larger femoral heads and generally allow more normal activities, even high-impact sports according to surgeons who utilize them [43].
Metal-on-Metal (MoM) Bearings
MoM bearings were recently re-introduced for several reasons, including the high wear rates and high incidence of osteolysis with metal-on-conventional polyethylene bearings in younger patients, and for resurfacing arthroplasty [46]. MoM bearings depend on a high level of congruence of the articulating metallic surfaces to encourage fluid film lubrication [47]. This results in extremely low wear rates [41, 48]. The head sizes can be larger than with a MoP bearing, increasing the range of motion and overall stability of the joint. These points lead to a resurgence of MoM bearing surfaces, which at one point constituted about 25 % or more of the hip replacement market in the USA. The early and mid-term results for some MoM total hip and resurfacing implants were very encouraging [49]. However, the enthusiasm for this bearing couple has waned somewhat because of issues related to pain and adverse tissue reactions with some implants [48]. Indeed several suboptimal implant designs with unacceptably high failure rates have been withdrawn from the marketplace [50, 51].
In general, patients with MoM total hip replacements have a higher incidence of adverse tissue reactions compared with those with MoP or CoC bearings. Some MoM failures are the result of a type IV hypersensitivity reaction to metal particles and their byproducts [41, 47]. The clinical presentation may vary from a diffusely painful joint with chronic synovitis and no other abnormal radiographic features to loosening, osteolysis or pseudotumor formation . Registry data from several countries have shown a higher revision rate for MoM bearing THRs [48, 52, 53]. Larger head sizes (> 28 mm) appear to increase these adverse events compared to smaller head sizes.
Willert and colleagues published a seminal study on adverse tissue reactions to MoM bearings and implicated a hypersensitivity reaction to metallic byproducts [14]. They noted prominent perivascular lymphocytic cuffing in the periprosthetic tissues and implicated immune processes for the adverse clinical outcomes in some patients. Patients with high wear rates of MoM hip implants, especially those with suboptimal alignment leading to edge loading , may have increased metal ion levels of cobalt and chromium in the blood . In vitro and in vivo studies have demonstrated that metal particulates and their byproducts may be associated with cytotoxicity, DNA damage (DNA-strand breaks, inhibition of DNA repair, chromosomal aberrations, etc.), metal hypersensitivity reactions and pseudotumors [47, 54]. Metal particles are about 30–200 nm in size; ionic complexes may form due to corrosion and other processes that degrade the alloys. The numbers of these smaller particles are often 2–3 orders of magnitude greater than with MoP articulations. These small metallic particles are small enough to cross the placenta. Although some hematopoietic abnormalities have been noted with MoM bearings, the incidence of different cancers in patients with MoM bearing surfaces does not appear to be higher compared to conventional MoP bearing surfaces [55].
In the last several years, the number of new MoM resurfacing arthroplasties has decreased dramatically, especially in younger women with smaller implant sizes [56]. These higher-risk patients are particularly susceptible to adverse immunological events due to wear byproducts from MoM implants [47]. Resurfacing arthroplasty is reported to have a much higher success rate in younger males with good bone stock and little deformity.
Other Bearing Couples
Other novel, so-called “hard-on-hard” bearing couples (such as ceramic-on-metal etc.) have recently been introduced to avoid the metallic byproduct issue altogether [46]. Longer-term studies are needed to determine their importance as a practical articulation for hip replacement.
Summary
As the general population continues to age, and high demands are placed on joint replacements to function for prolonged periods of time, issues related to implant materials become more prominent. Thorough preclinical assessment of newly introduced materials must be rigorous to avoid some of the pitfalls noted during the last one to two decades. Although advances have been made, the long-lasting, high-performance joint replacement that will function normally in vivo is still elusive.
This work was supported in part by National Institute of Health (NIH) grants 2R01 AR055650-05 and 1R01 AR063717-01.
References
Charnley J (1979) Low friction arthroplasty of the hip. Springer, New York
Goodman SB, Gomez Barrena E, Takagi M, Konttinen YT (2009) Biocompatibility of total joint replacements: a review. J Biomed Mater Res A 90(2):603–618
Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Fa’vae F et al (2000) The John Charnley award. Wear is a function of use, not time. Clin Orthop Relat Res 2000(381):36–46
Marshall A, Ries MD, Paprosky W (2008) How prevalent are implant wear and osteolysis, and how has the scope of osteolysis changed since 2000? J Am Acad Orthop Surg 16(Suppl 1):S1–S6
Schmalzried TP, Kwong LM, Jasty M, Sedlacek RC, Haire TC, O’Connor DO et al (1992) The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin Orthop Relat Res 1992(274):60–78
Campbell P, Shen FW, McKellop H (2004) Biologic and tribologic considerations of alternative bearing surfaces. Clin Orthop Relat Res 2004(418):98–111
Hallab NJ, Jacobs JJ (2009) Biologic effects of implant debris. Bull NYU Hosp Jt Dis 67(2):182–188
Ingham E, Fisher J (2002) Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H 214(1):21–37
Shanbhag AS, Jacobs JJ, Glant TT, Gilbert JL, Black J, Galante JO (1994) Composition and morphology of wear debris in failed uncemented total hip replacement. J Bone Joint Surg Br 76(1):60–67
Meneghini RM, Hallab NJ, Jacobs JJ (2005) The biology of alternative bearing surfaces in total joint arthroplasty. Instr Course Lect 54:481–493
Holt G, Murnaghan C, Reilly J, Meek RM (2007) The biology of aseptic osteolysis. Clin Orthop Relat Res 460:240–252
Jacobs JJ, Roebuck KA, Archibeck M, Hallab NJ, Glant TT (2001) Osteolysis: basic science. Clin Orthop Relat Res 2001(393):71–77
Revell PA (2008) The combined role of wear particles, macrophages and lymphocytes in the loosening of total joint prostheses. J R Soc Interface 5(28):1263–1278
Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G et al (2005) Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am 87(1):28–36
Purdue PE, Koulouvaris P, Potter HG, Nestor BJ, Sculco TP (2007) The cellular and molecular biology of periprosthetic osteolysis. Clin Orthop Relat Res 454:251–261
Tuan RS, Lee FY, Konttinen Y, Wilkinson JM, Smith RL (2008) What are the local and systemic biologic reactions and mediators to wear debris, and what host factors determine or modulate the biologic response to wear particles? J Am Acad Orthop Surg 16(Suppl 1):S33–38
Goodman SB, Huie P, Song Y, Schurman D, Maloney W, Woolson S et al (1998) Cellular profile and cytokine production at prosthetic interfaces. Study of tissues retrieved from revised hip and knee replacements. J Bone Joint Surg Br 80(3):531–539
Kadoya Y, Revell PA, al-Saffar N, Kobayashi A, Scott G, Freeman MA (1996) Bone formation and bone resorption in failed total joint arthroplasties: histomorphometric analysis with histochemical and immunohistochemical technique. J Orthop Res 14(3):473–482
Willert HG (1977) Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res 11(2):157–164
Aspenberg P, Van der Vis H (1998) Migration, particles, and fluid pressure. A discussion of causes of prosthetic loosening. Clin Orthop Relat Res 1998(352):75–80
Aspenberg P, Vis H van der (1998) Fluid pressure may cause periprosthetic osteolysis. Particles are not the only thing. Acta Orthop Scand 69(1):1–4
Ren PG, Irani A, Huang Z, Ma T, Biswal S, Goodman SB (2011) Continuous infusion of UHMWPE particles induces increased bone macrophages and osteolysis. Clin Orthop Relat Res 469(1):113–122
Fritton K, Ren PG, Gibon E, Rao AJ, Ma T, Biswal S et al (2012) Exogenous MC3T3 preosteoblasts migrate systemically and mitigate the adverse effects of wear particles. Tissue Eng Part A 18(23–24):2559–2567
Ren W, Markel DC, Schwendener R, Ding Y, Wu B, Wooley PH (2008) Macrophage depletion diminishes implant-wear-induced inflammatory osteolysis in a mouse model. J Biomed Mater Res A 85(4):1043–1051
Jiranek WA, Machado M, Jasty M, Jevsevar D, Wolfe HJ, Goldring SR et al (1993) Production of cytokines around loosened cemented acetabular components. Analysis with immunohistochemical techniques and in situ hybridization. J Bone Joint Surg Am 75(6):863–879
Goodman SB (2007) Wear particles, periprosthetic osteolysis and the immune system. BioMaterials 28(34):5044–5048
Gibon E, Batke B, Jawad MU, Fritton K, Rao A, Yao Z et al (2012) MC3T3-E1 osteoprogenitor cells systemically migrate to a bone defect and enhance bone healing. Tissue Eng Part A 18(9–10):968–973
Dumbleton JH, Manley MT, Edidin AA (2002) A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 17(5):649–661
Kurtz SM, Gawel HA, Patel JD (2011) History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res 469(8):2262–2277
Oral E, Muratoglu OK (2011) Vitamin E diffused, highly crosslinked UHMWPE: a review. Int Orthop 35(2):215–223
Gordon AC, D’Lima DD, Colwell CW Jr (2006) Highly cross-linked polyethylene in total hip arthroplasty. J Am Acad Orthop Surg 14(9):511–523
Jacobs CA, Christensen CP, Greenwald AS, McKellop H (2007) Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. J Bone Joint Surg Am 89(12):2779–2286
Atwood SA, Van Citters DW, Patten EW, Furmanski J, Ries MD, Pruitt LA (2011) Tradeoffs amongst fatigue, wear, and oxidation resistance of cross-linked ultra-high molecular weight polyethylene. J Mech Behav Biomed Mater 4(7):1033–1045
Schroder DT, Kelly NH, Wright TM, Parks ML (2011) Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE. Clin Orthop Relat Res 469(2):387–394
Rimnac C, Pruitt L (2008) How do material properties influence wear and fracture mechanisms? J Am Acad Orthop Surg 16(Suppl 1):S94–S100
Furmanski J, Anderson M, Bal S, Greenwald AS, Halley D, Penenberg B et al (2009) Clinical fracture of cross-linked UHMWPE acetabular liners. BioMaterials 30(29):5572–5582
Waewsawangwong W, Goodman SB (2012) Unexpected failure of highly cross-linked polyethylene acetabular liner. J Arthroplasty 27(2):323 e1–e4
Fisher J, McEwen HM, Tipper JL, Galvin AL, Ingram J, Kamali A et al (2004) Wear, debris, and biologic activity of cross-linked polyethylene in the knee: benefits and potential concerns. Clin Orthop Relat Res 2004(428):114–119
Illgen RL 2nd, Bauer LM, Hotujec BT, Kolpin SE, Bakhtiar A, Forsythe TM (2009) Highly crosslinked vs conventional polyethylene particles: relative in vivo inflammatory response. J Arthroplasty 24(1):117–124
Illgen RL 2nd, Forsythe TM, Pike JW, Laurent MP, Blanchard CR (2008) Highly crosslinked vs conventional polyethylene particles—an in vitro comparison of biologic activities. J Arthroplasty 23(5):721–731
Zywiel MG, Sayeed SA, Johnson AJ, Schmalzried TP, Mont MA (2011) State of the art in hard-on-hard bearings: how did we get here and what have we achieved? Expert Rev Med Devices 8(2):187–207
D’Antonio JA, Capello WN, Naughton M (2012) Ceramic bearings for total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res 470(2):373–381
Hannouche D, Zaoui A, Zadegan F, Sedel L, Nizard R (2011) Thirty years of experience with alumina-on-alumina bearings in total hip arthroplasty. Int Orthop 35(2):207–213
Jeffers JR, Walter WL (2012) Ceramic-on-ceramic bearings in hip arthroplasty: state of the art and the future. J Bone Joint Surg Br 94(6):735–745
Barrack RL, Burak C, Skinner HB (2004) Concerns about ceramics in THA. Clin Orthop Relat Res 2004(429):73–79
Cigada A, Cotogno G, Chiesa R (2011) The ceramic-on-metal coupling in total hip replacements for young patients: a review study. J Appl Biomater Biomech 9(1):2–10
Gill HS, Grammatopoulos G, Adshead S, Tsialogiannis E, Tsiridis E (2012) Molecular and immune toxicity of CoCr nanoparticles in MoM hip arthroplasty. Trends Mol Med 18(3):145–155
Bozic KJ, Browne J, Dangles CJ, Manner PA, Yates AJ Jr, Weber KL et al (2012) Modern metal-on-metal hip implants. J Am Acad Orthop Surg 20(6):402–406
Shetty V, Shitole B, Shetty G, Thakur H, Bhandari M (2011) Optimal bearing surfaces for total hip replacement in the young patient: a meta-analysis. Int Orthop 35(9):1281–1287
Steiger RN de, Hang JR, Miller LN, Graves SE, Davidson DC (2011) Five-year results of the ASR XL acetabular system and the ASR hip resurfacing system: an analysis from the Australian orthopaedic association national joint replacement registry. J Bone Joint Surg Am 93(24):2287–2293
Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P et al (2011) Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br 93(8):1011–1106
Smith AJ, Dieppe P, Vernon K, Porter M, Blom AW (2012) Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet 379(9822):1199–1204
Graves SE, Rothwell A, Tucker K, Jacobs JJ, Sedrakyan A (2011) A multinational assessment of metal-on-metal bearings in hip replacement. J Bone Joint Surg Am 93(Suppl 3):43–47
Mabilleau G, Kwon YM, Pandit H, Murray DW, Sabokbar A (2008) Metal-on-metal hip resurfacing arthroplasty: a review of periprosthetic biological reactions. Acta Orthop 79(6):734–747
Makela KT, Visuri T, Pulkkinen P, Eskelinen A, Remes V, Virolainen P et al (2012) Risk of cancer with metal-on-metal hip replacements: population based study. BMJ 345:e4646
Macpherson GJ, Breusch SJ (2011) Metal-on-metal hip resurfacing: a critical review. Arch Orthop Trauma Surg 131(1):101–110
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2014 Springer Science+Business Media New York
About this chapter
Cite this chapter
Goodman, S. (2014). Bearing Surfaces for Joint Replacement: New Materials or New Problems. In: Jones, L., Haggard, W., Greenwald, A. (eds) Metal-on-Metal Bearings. Springer, New York, NY. https://doi.org/10.1007/978-1-4614-8999-3_2