1 Introduction

This paper focuses on the remediation phase of a CBRNe incident, which includes decontamination measures. This last phase of an incident is—in my view—a “stepchild” of the CBRNe industry, as it includes very long and technically challenging processes, such as the process of declaring an incident site safe. An incident site could be deemed “clean” when it “is no longer a hazard for the user”. However, this definition is not accurate, because the expression “hazard for the user” is on a sliding scale. This paper shall investigate what is considered as being safe.

The problems and the very time-consuming operations to render a site safe became very obvious—and also known to the general public—during the cleaning operations after the recent (2018) incidents in the UK, which involved a deadly substance—namely Novichok—a nerve agent belonging to the group of Chemical Warfare Agents (CWA).

The decontamination (or cleaning) of an affected site involves a process of testing, removal of items and materials that may have been contaminated, but also chemical cleaning and retesting. Testing and retesting are very time-consuming operations, as they involve the analysis of the samples taken, which is usually performed in (accredited) chemical laboratories. Can this process be shortened? This is another question that this paper tries to answer.

The conceptual approach of this paper is based upon a question that I have been faced with many times, in operational settings as well as in training environments: How can safe levels of surface contamination be established in a fast, real-time, and analytically recognized manner? In 2017, when I started working with the Proton Transfer Reaction Time of Flight Mass Spectrometry (PTR–ToF–MS) technology, I realized that instruments based upon this technology could finally allow the relevant actors to overcome the existing limitations and make them able to determine safe, residual contamination levels in a fast and reliable manner. I have engaged in a technical cooperation with an Austrian company, IONICON Analytik GmbH, that has developed that technology for industrial applications.

In August 2018, for the first time ever with that technology, we carried out field trials with Chemical Warfare Agents in a specialized laboratory in the Czech Republic. Such trials aimed at clarifying whether clearance decontamination could be supported by PTR–ToF–MS technology.

2 Chemical Warfare Agents

Chemical Warfare Agents—often referred to as Chemical Weapons—are chemicals used to cause intentional death or harm through their toxic properties; as such, they are defined in the Chemical Weapons Convention [1].

CW agents used in warfare are classified as follows [2]:

  • Nerve agents

  • Vesicants (blistering agents)

  • Bloods agents (cyanogenic agents)

  • Choking agents (pulmonary agents)

  • Riot-control agents (tear gases)

  • Psychomimetic agents

  • Toxins.

It is important to note that NATO AEP-58 [3] recommends decontamination testing to be performed with two main types of CWAs: nerve and blister agents.

Therefore, this paper (and its laboratory trials) focuses on those two categories of Chemical Weapons.

2.1 Residual contamination levels: field trials

The question to be answered is: Are there are any residual contamination levels after a completed cycle with current decontamination technologies and procedures? This level of thorough decontamination should partially or totally remove all levels of contamination. But what are the amounts that the “residual contamination levels” expression refers to? The results of some thorough decontamination field trials, carried out in 2009 and in 2014, will shed light on this point.

The trials were carried out at a specialized facility that allows the outdoor use of Chemical Warfare Agents, using decontamination equipment that had been tested against current NATO standards (AEP-58). The product names of various decontamination solutions are not mentioned, and the test protocols (set up of the tests) are not explained in detail in this paper.

Two field tests were carried out by the author of this paper with one military unit in 2014 and with an international arms control organization (Organisation for the Prohibition of Chemical Weapons) in 2009.

The samples were contaminated with Chemical Warfare Agents; more specifically, they were calculated by the equivalent of 10 g/m2. The surface areas of the contaminated painted metal plates were 5 cm by 5 cm, for a total of 25 cm2.

Totally, 8 out of 24 samples showed no residual contamination (below LOD, Limit of Detection), according to the analytical laboratory results of the samples taken (wipe samples).

It can be concluded that two-thirds of the decontaminated surfaces were still contaminated with small amounts of Chemical Warfare Agents. Some of those residual amounts can be considered as traces, as decontamination efficiency was in the range of 90–100%. Traces (or trace elements) of CWA are defined in a certain percentage of volume or weight, usually ranging from 0.1 to 0.01% [4], or as any element having an average concentration of less than about 100 parts per million atoms or less than 100 μg/g [5].

The wipe samples were analysed with GC–MS technology.

3 How clean is safe? NATO desirable detection levels for surface contamination

STANREC 4578 lists the essential and desirable surface contamination detection levels for some Chemical Warfare Agents. What are such desirable detection levels? They are defined as levels of detection—after completed decontamination—that will prevent exposure of individuals to have noticeable health effects; it is expected that detection equipment will detect contamination down to the levels shown in Table 1.

Table 1 NATO desirable detection levels

Below are the aforesaid levels—divided into essential and desirable ones [6]:

According to NATO, it is clear that some of said levels are particularly challenging and may be beyond current detection technologies [6]. This paper demonstrates that there is a technology (PTR–TOF–MS technology) that can detect surface contamination at those levels and therefore meet the desired NATO levels.

The detection and identification of CWA vapours, stemming from the surface contamination with low quantities, can also be considered as trace gas analysis.

4 PTR–ToF–MS technology

4.1 Description

Proton transfer reaction mass spectrometry (PTR–MS) is a technology that provides online monitoring of volatile organic compounds (VOCs). The main technical feature of this technology is that it offers a low detection threshold and a fast response time. PTR–MS commercial instruments are mainly based on the quadrupole analytical method. Their main limitation consists in the fact that they provide very little useful information, besides the nominal mass of ions detected. IONICON Analytik, instead, developed new instruments based on time-of-flight (PTR–TOF–MS) analysers. This technology gives the user a higher mass resolution, which subsequently leads to more analytical data contained in the spectra.

PTR–ToF–MS technology can fill current gaps already identified by NATO in 2015; the trials conducted with CWA, as described in this paper, clearly demonstrated the capabilities of this technology.

4.2 Hazard monitor

The hazard monitor is a purpose-built IONICON PTR–TOF–MS, with up to 2000 cps/ppbv sensitivity and a 1-s detection limit of around 10 pptv, which is housed in two interconnected 19″ cubic racks, with a total weight of about 90 kg, for maximum flexibility.

Furthermore, the instrument is equipped with a sophisticated automated data processing software, which switches the ion chemistry and performs pattern matching algorithms according to substances entries in a database. This means that selectivity is considerably improved, i.e. the risk of false positives and negatives is reduced, and the user gets concentrations readings of the respective CWAs with a time resolution of down to 1 s.

The trials with CWA allowed us to use—for the very first time—NH4+ reagent ions, which are produced without the need of any ammonia, in addition to H3O+, NO+ and O2+, respectively.

5 Trials description

5.1 Agents

The agents used for the trials belong to two of the groups mentioned earlier in this paper:

Nerve agents and blister agents.

  1. a.

    Nerve agents: tabun (GA); sarin (GB); soman (GD); cyclosarin (GF); VX

  2. b.

    Blister agents: mustard (HD); nitrogen mustard (HN1); nitrogen mustard (HN3); Lewisite (L1).

What is important to note at this point is that the proton affinity for CWA can only be assessed after evaluation of the trials. The proton affinity for CWA—to my knowledge—is not recorded in any scientific literature, as tests with real CWA have not been done to the extent it was carried out for this project.

5.1.1 Purity of agents

The equipment used for our trials (PTR–ToF–MS) would pick up impurities of the above-mentioned agents; it was therefore important to have the purity of agents assessed at the beginning.

The purities were as follows (Table 2):

Table 2 Agent purity

Analytical reports of all agents used during the trials are referenced. Detailed information can be found in the relevant analytical reports [7].

5.2 Primary ions

Primary ions (also called reagent ions) are used for chemical ionization of the analytes. When H3O+ is used as primary ions, the main ionization channel is proton transfer, i.e. the product ions are mainly protonated molecules (M.H+). However, it should be noted that, although the energy transfer upon proton transfer is relatively small, in some cases fragmentation of the analyte molecules can occur.

In the IONICON PTR–ToF–MS technology, the primary ions can also be switched to NH4+, O2+, NO+ and Kr+.

During our trials, we mainly used H3O+ and NH4+.

5.3 Solvents/dilution

Two solvents were used for the dilution of Chemical Warfare Agents—Hexane and Dichloromethane.

  • Hexane, a non-polar solvent, is commonly used in chromatography. Its chemical formula is C6H14, and it has a lower proton affinity than ammonia and water. Therefore, it does not get ionized via proton transfer from NH4+or H3O+ and can be used as a solvent for chemical analytical trials where the dilution of the main component (in our case CWA) is required.

  • The chemical formula of methylene dichloride (DCM, or methylene chloride, or dichloromethane) is CH2Cl2. It is widely used as a solvent and diluent. It also has a lower proton affinity than ammonia and water.

The decision on which solvent to use for our analytical trials was solely based on the initial results obtained. At times, we changed solvents to obtain better results.

5.4 Dilution series

The test samples were diluted, according to the requirements given to the laboratory staff.

The first sample always represented 1μl of the neat agent. The subsequent amounts were calculated based upon the surface area of the petri dish, namely 28 cm2.

The overview of the dilution steps is shown in Table 3. Series A was applied for 8 out of 9 agents. Series B had to be applied for the agent with the lowest vapour pressure—VX.

Table 3 Steps of dilution

The levels of contamination are always (NATO) given in amounts per cm2. It was therefore extremely important for us to be able to calculate the results on that surface area.

The dilution was carried out by laboratory staff. They prepared the petri dishes and their content based on the values given to them.

5.5 Trials nerve agents

Step 0 was performed with pure (undiluted) CWA. All the figures below, titled “Agent type net”, are referring to this step (Table 4).

Table 4 Dilution nerve agents

Diluted CWA was used for the subsequent trials, for agents GA, GB, GD, and GF from step 1 to step 2, with a dilution factor of 103. From step 2 to step 3 the factor was 102, from 3 to 4 101, from 4 to 5 again 101, and the same applied to steps 5–6.

The last dilution step always represents the LOD, Limit of Detection. This is further explained in chapter 6.

VX had to be dealt with differently, due to its relatively low vapour pressure compared to all other agents. It was always diluted by a factor of 101.

5.6 Trials blister agents

Step 0 was performed with pure (undiluted) CWA. All the figures below, titled “Agent type net”, refer to this step (Table 5).

Table 5 Dilution blister agents

Diluted CWA was used for the subsequent trials, for the agents HD, HN1 and HN3 from step 1 to step 2 with a dilution factor of 103, from step 2 to step 3 the factor was 102 and from 3 to 7 101, respectively.

This series was somewhat different for Lewisite, as the dilution factor was 103 from Step 1 to 2, and only 101 from 2 to 3.

The last dilution step always represents the LOD, Limit of Detection. This is further explained in chapter 6.

6 Evaluation of field trials

This chapter describes and evaluates the test results, based on:

  • detectable concentration in the air and

  • detectable concentration on a surface.

It also compares the test results to existing standards.

6.1 Detectable concentration in air

The concentration values were calculated based on the results of the laboratory trials. The instrument gives concentration values, based on calculations expressed in parts per trillion (volume). In order to compare said values to existing CWA health and safety standards, I transformed the pptv values into ng/m3, by using the following simplified formula:

$$ {\text{mg}}/{\text{mg}}^{3} = {\text{ppm}} \times \frac{\text{MW}}{22.41} $$

whereas MW stands for molecular weight; 22,41 is a conversion constant.

Detailed results are shown in Tables 6, 7, 8, 9, 10 and Figs. 1 and 2.

Table 6 LOD air nerve agents
Table 7 LOD air blister agents
Table 8 NATO safe residual contamination levels (NATO Standard NATO STANREC 5784 Edition November 2015)
Table 9 LOD surface nerve agents
Table 10 LOD surface blister agents
Fig. 1
figure 1

HD LOD air

Fig. 2
figure 2

LOD surface HD

6.1.1 Nerve agents

6.1.1.1 Limit of detection (LOD): concentration in air

The results are summarized in the table below.

6.1.2 Blister agents

6.1.2.1 Limit of detection (LOD): concentration in air

It has to be noted that the concentration values below refer to the volume of the funnel (30 cm3), measured at a distance of 10 cm from the surface of the petri dish.

The results are summarized in the table below.

6.2 Detectable concentration on surface

The LoD on surface areas has been identified through our field trials. The results are described hereunder.

NATO defines its essential and desirable levels of residual and safe surface contamination as follows:

The results of the field trials have shown that the detection and identification levels of the PTR–ToF–MS technology lie well within these desirable NATO limits, apart from VX and L, for which the instrument meets the essential limit.

6.2.1 Nerve agents

6.2.1.1 Limit of detection: concentration on surface

The values in the table below represent the area where the noise (background) signal could still be clearly distinguished from the background. At the subsequent step of dilution, this was no longer possible.

6.2.2 Blister agents

6.2.2.1 Limit of detection (LOD): concentration on surface

The values in the table below represent the area where the obtained signal could still be distinguished from the background (noise). At the subsequent step of dilution, this was no longer possible.

7 Conclusions: clearance decontamination supported by PTR–ToF–MS—a way forward?

Multiple decontamination cycles might be required to render contaminated surfaces safe to a level that they do not pose any threat to humans anymore, which means that such levels do not cause any noticeable health effect. How do the relevant authorities perform checks on potentially high residual contamination levels? For the moment being, this process is time-consuming and involves elaborate analytical procedures and laboratories. The IONICON PTR–ToF–MS technology can be used on site and can produce real-time—immediate—results. It does not require taking any samples, is easy to use, and analytically reliable. Field trials have clearly shown that NATO essential and desirable detection levels can be reached. In my view, the Proton Transfer Reaction–Time of Flight–Mass Spectrometry trace detection technology is an invaluable asset supporting civilian authorities in determining safe levels of surface contamination in real time, during decontamination cycles, and after the completion of decontamination operations.