Introduction

Osteoporosis is the most common bone disease and represents a major public health problem because of its worldwide increasing incidence and heavy social economic burden [1].

Currently, it is estimated that over 200 million people worldwide suffer from this disease, and at least 40 % of women and about 20 % of men will sustain one or more fragility fractures in their lifetime [2].

Osteoporosis causes more than 8.9 million fractures annually worldwide, and over one-third of all osteoporotic fractures occur in Europe [3, 4]. By 2050, the worldwide number of hip fractures is estimated to rise from 1.66 to 6.26 million [5], and their incidence is projected to increase up to 240 % in women and 310 % in men [4].

Osteoporotic fractures cause increasing morbidity and mortality rates; in particular, hip fractures result in 10–20 % of mortality within 1 year [1, 6].

Osteoporosis has been defined on the basis of bone mineral density (BMD) assessed at the hip or lumbar spine level. BMD represents a cornerstone for the general management of osteoporosis, being used for diagnosis, fracture risk assessment, selection of patients for treatment and monitoring of treatment efficacy [7].

According to the World Health Organization (WHO) criteria, osteoporosis is defined as a BMD value less than or equal to 2.5 standard deviations below the young normal mean reference population.

Dual energy X-ray absorptiometry (DXA) is the current gold standard for BMD measurement.

BMD measured by DXA is highly correlated with biomechanical bone strength and, in longitudinal studies, with fracture risk [7].

High-quality BMD measurement and interpretation are important because physicians rely on these reports to make patient care decisions, including additional medical evaluations (laboratory or imaging), starting or changing drug therapy, or suggesting some dietetic and lifestyle habits [8].

Incorrect BMD acquisition or reporting may generate unnecessary medical expenses and lead to inappropriate clinical and therapeutic decisions that could have adverse effects on patients’ health [9]. Errors can result from improper equipment installation and maintenance, inadequate education and training in bone densitometry and insufficient knowledge of current guidelines.

BMD testing, interpretation and reporting are erroneously considered to be simple; instead, all these steps require training and experience for technologists and physicians.

For these reasons, mistakes in BMD assessment are commonly seen. In 2006, Lewiecki et al. [8] reported about 71 % of clinicians and 45 % of technologist see incorrect densitometric interpretation and acquisition at least once a month. These errors were due to a lack of adherence to manufacturers’ recommendations for device maintenance and quality control, incorrect acquisition and analysis of data (e.g., wrong demographic information, improper patient positioning, artifacts not removed from the scanned area, incorrect labeling of vertebral bodies), or misleading interpretation of results.

Hence, performing the periodic and daily machine calibration tests conforming to the manufacturers’ guidelines, an adequate acquisition and analysis of data, and correct interpretation and reporting of results are essential steps to help the management of patients [8, 10].

The aim of this review is to discuss the current practice of quality assurance (QA) procedures in bone densitometry.

Quality Assurance

QA is by definition “a programme for the systematic monitoring and assessment of the various aspects of a service or facility to ensure that standards of quality are being met.” QA in bone densitometry has to guarantee (1) the maintainance of adequate equipment performance with the lowest possible radiation dose to patients consistent with clinical imaging requirements and (2) the assurance of adequate diagnostic information provided at the lowest possible cost. A QA program should include (1) guidelines and requirements for equipment installation and its clinical use, (2) standardised periodic tests developed to maintain diagnostic quality (quality control, QC), (3) preventive maintenance procedures, (4) staff training and education, and (5) radiation safety policies and procedures.

Equipment and Installation

Manufacturers should provide all necessary and required equipment for DXA facilities (e.g., the positioning device for the anteroposterior lumbar spine scan acquisition; the hip positioning device for the femur scan acquisition).

The supplier of the equipment should describe all clinical applications of the system and specify the performance in terms of accuracy and precision [11∙∙]. Equipment packages should also include dedicated phantoms for calibration and QC procedures.

The installation of the equipment should be followed by procedures that guarantee correct functioning and satisfy all current radiation protection requirements and manufacturer specifications [12∙].

Diagnostic Quality and Quality Control: Phantom Scanning and Cross-Calibration

When performing DXA measurements, it is important to implement a QA program to guarantee that any observed density changes are real and not due to machine or to operator variability. The quality of the acquired data is defined by accuracy and precision [13].

Accuracy is the ability of a measurement to match the actual value of the quantity being measured and is expressed in terms of accuracy error. The accuracy error is the percentage difference between a measurement and a defined standard and is influenced by the machine. Generally, the accuracy error of the most widely used bone densitometers is about 5–10 % [14].

To evaluate the accuracy of a DXA, a daily calibration measurement, according to the manufacturer’s instructions, is performed using a standard phantom supplied with the system. In addition to the daily calibration, the manufacturers recommand a periodic QC (generally every 3–6 months) using a different phantom to independently monitor the scanner stability [13]. The various manufacturers have developed different types of QC phantoms. Several of these consist of vertebrae made of simulated bone material encased in plastic; others are made of an aluminum bar to be submersed in water. During the calibration process, the phantom has to be located in the middle of the scanning table, at the top side of the table. At the end of the calibration, results have to be analyzed and recorded in the QC database [13].

Precision is defined as the reproducibility and is expressed in terms of precision error. Precision error provides a quantitative measure of the ability of densitometric systems to detect changes in BMD over time and thus is an important component of clinical decision making. This is influenced by several parameters, such as the skill and training of the technician performing the scan, the skeletal site of measurement and the clinical features of patients [12∙].

Knowledge of the precision error is essential in clinical practice to guarantee a correct diagnosis and management of patients, because it allows distinguishing between true change in BMD and random variability in the measurement procedure. The least significant change (LSC) enables the identification of statistically significant changes in BMD by considering the precision error [coefficient of variation (CV)] of the measurement [15∙]. LSC should to be calculated using the International Society for Clinical Densitometry (ISCD) precision calculating tool at www.iscd.org.

The various densitometer manufacturers often establish precision errors for their equipment on the order of 1 % [16]. Since the precision errors estimated by the manufacturer do not usually represent the error in the clinical setting, the ISCD recommends that in vivo precision errors and LSC for each DXA center should be determined using repeated measurements in 30 subjects (representative of the clinic’s patient population) with repositioning or in 15 subjects scanned three times [17, 18∙∙].

Moreover, if a DXA facility has more than one technologist, every technologist should perform an in vivo precision assessment using patients. Then an average precision error combining data from all technologists should be used to establish precision error and consequently LSC for the facility.

The ISCD also recommends that each technologist should complete precision assessment after having performed approximately 100 patient scans [18∙∙].

The LSC of the most commonly used DXA machines are reported in Table 1 [15∙].

Table 1 Precision of bone measurement expressed in terms of least significant change (LSC) using the most common DXA technologies [15∙]

Differences in calibration exist among the densitometry systems from different manufacturers. To overcome this relevant limit, cross-calibration techniques have been developed. In the last 20 years cross-calibration was attempted using a European Spine Phantom containing three simulated vertebrae of different densities (low, normal and high density) composed of calcium hydroxyapatite. Results were compared to the spine BMD of a group of females and then were used to define and introduce the standardized BMD (sBMD). sBMDs were used to correct the calibration differences among the three major distributors of DXA (Hologic, Lunar and Norland) [19, 20]. The ISCD guidelines actually affirm that is not possible to quantitatively compare BMD or to calculate a LSC between different facilities without cross-calibration.

The ISCD also recommends performing a cross-calibration when the DXA hardware, but not the entire system, have been changed, or when a system with the same technology (manufacturer and model) has been replaced, in order to allow a quantitative comparison to the prior machine. The cross-calibration procedure consists of ten phantom scans, with repositioning, before and after hardware change [18∙∙]. If a greater than 1 % difference in mean BMD is observed, the manufacturer should be contacted for service or correction. When changing an entire system to one made by the same manufacturer using a different technology, or when changing to a system made by a different manufacturer, cross-calibration should be performed in 30 patients representative of the facility’s patient population once on the initial system and then twice on the new system within 60 days. All cross-calibration measures have to be performed at the hip and lumbar site. The average BMD and LSC between the initial and new machine should be calculated using the ISCD DXA Machine Cross-Calibration Tool [18∙∙]. Then LSC for comparison between the previous and new system should be used. Inter-system quantitative comparisons can only be made if cross-calibration is performed on each commonly measured skeletal site. Once a new precision assessment has been performed on the new system, all future scans should be compared to scans performed on the new system using the newly established intra-system LSC. If a cross-calibration assessment is not performed, no quantitative comparison to the prior machine can be made. Consequently, a new baseline BMD and intra-system LSC need to be established [18∙∙].

Preventive Maintenance Procedures

In addition to QC, a periodic assessment of the condition of the DXA devices should be executed (i.e., checking the electrical safety, equipment emergency stop, radiation warning lights, and the availability and adequacy of patient support tools). Moreover, all test results should be recorded to detect needed changes [12∙].

Staff Training and Education

Incorrect acquisition and analysis of data or incorrect interpretations of results occur frequently in daily clinical practice [8] and are responsible for misleading diagnoses and inadequate patient management.

Therefore, training of technicians performing the examination as well as education of radiologists writing the reports is needed to prevent and minimize these kinds of errors.

Errors in performing DXA acquisitions and data analysis are usually due to (1) improper patient positioning (e.g., the spine is not parallel to the edges of the DXA table or the hip is not sufficiently internally rotated); (2) using the wrong scan mode (e.g., the scan mode may alter the BMD and is manually or automatically selected); (3) artifacts not being removed from the scanned area (e.g., the spine is scanned when the patient is wearing an underwire bra or has a bellybutton ring); (4) obtaining incorrect demographic information (e.g., a man is entered as a woman or the incorrect age is used); (5) failure to correct improper default identification of bone edges and regions of interest (e.g., the computer includes large osteophyte or vascular calcifications in the area of the measured spine); (6) incorrect labeling of vertebral bodies [9, 21].

Errors in the interpretations of BMD values also frequently occur and are due to (1) an incorrect application of WHO diagnostic T-score criteria and ISCD official positions (e.g., reporting T-scores in a healthy premenopausal woman and applying the WHO diagnostic criteria); (2) an invalid BMD comparison (LSC not known, different instruments used, left hip compared with right hip, comparing T-scores instead of BMD, different scan modes); (3) fracture risk incorrectly represented [9].

Moreover, body composition may influence the precision and accuracy of DXA and therefore affect interpretations of DXA examinations. In particular, caution must be used when interpreting BMD results in subjects who have undergone large weight changes [22].

However, training courses for both physicians and technologists are provided by several national and international societies.

Dose Radiation Exposure

One of the main strengths of DXA is the low radiation exposure for patients [10, 23]. Nevertheless, recent advances in DXA technology (e.g., spatial resolution improvement, small time consumption) have involved an increase in radiation dose [24]. However, radiation exposure is much lower than other diagnostic techniques and lower than the average daily dose from natural background radiation (6.6 μSv/day; 2,400 μSv/year) [12∙, 24].

Many variables affect the DXA patient dose, such as the site of measurement (e.g., lumbar spine, forearm, hip), patient body size, acquisition technique, X-ray tube filtration, efficiency of detection system, number of scans, exposure parameters, scan speed and scan size [11∙∙].

Table 2 shows the effective dose (calculated using the organ- or tissue-absorbed doses and the relative radiosensitivity assigned to each of these organs or tissues [25]) of several models of DXA.

Table 2 Effective dose of different DXA devices

Pregnancy is a contraindication to DXA. Damilakis et al. [26] estimated conceptus doses and risks related to DXA performed during pregnancy. A careful screening for pregnancy is essential to minimize the number of accidental exposures of pregnant patients by DXA. When a pregnant woman has a DXA examination, communication with her about the dose to the fetus is very important. Because the risks from DXA are negligible, usually abortion due to radiation exposure is not recommended [25, 26].

The radiation burden is increased in follow-up studies. A clinical justification of these examinations should be provided to ensure that the expected benefits exceed the potential risks [11∙∙].

The occupational radiation dose should also be considered. The radiation exposure for DXA technologists is generally very low if all procedures are performed according to the manufacturer’s guidelines. The reported scatter dose rates at 1 m from the central axis of the patient table range from a few tenths of 1–5 μSv/h depending on the scanner model. From these values, recent calculations and measurements indicate that the annual dose from fan-beam systems for an average workload (20 patients/day) at 1 m from the scanner will be between 0.1 and 1.5 mSv, depending on the scanner model [10, 27].

If a female technician becomes pregnant, she should inform her employer. A pregnant staff member can continue working in a DXA scanning room as long as the fetal dose is kept below 1 mGy for the term of the pregnancy [28].

An increase in the operator’s distance from the patient and the use of mobile or immobile shielding can reduce the operator’s radiation exposure [10]. The need for shielding mainly depends on the patient workload, size of the examination room and type of DXA scanner (especially for fan-beam technology) [12∙].

Fan-beam and cone-beam systems may require shielding if the operator has to be seated close to the table or if the walls are close to the table [27].

Moreover, if the room configuration does not allow the necessary distance between the DXA and operator, a transparent lead-acrylic shielding may be needed to protect the technicians during the examination [10].

Conclusions

QA procedures in bone densitometry should be performed carefully and strictly conform to current guidelines to guarantee adequate diagnostic performance and a consequent correct overview and management of osteoporotic patients.