Introduction

Electrophysiological studies, radiofrequency ablation, and implantable devices have thoroughly revolutionized the treatment of patients with symptomatic tachyarrhythmia and bradyrhythmia. Benefits range from quality-of-life improvement to preventing sudden death [110]. Just three decades ago, patients with symptomatic supraventricular tachycardia that failed drug therapy would have attempted open-heart surgical treatment with high morbidity and with unacceptable morbidity and mortality. Complete endovascular approaches to map and ablate accessory pathways and atrial input to the atrioventricular (AV) node have all but nearly obviated the need for open surgical procedures for these disorders.

In the last 10 years, major strides in the invasive treatment of atrial fibrillation have occurred [1]. The last 10 years have seen major strides in the management of atrial fibrillation including the appreciation of the role of the great thoracic veins including the pulmonary veins and the genesis of atrial fibrillation and thereby presenting a target for ablation [11]. Further and even more recent approaches in managing atrial fibrillation have included targeting the retroatrial ganglionated plexii, the vein of Marshall, and fragmented signals in the atria to improve success rate with invasive procedures.

With each new advance, new opportunities to innovate and find solutions that enhance success rates in each of these subtechniques abound.

The newer methods of treatment have also been associated with unrecognized complications including catastrophic thromboembolism, atrial esophageal fistula [12] formation, and pacemaker-related myocardial dysfunction. Since we need to maintain the success associated with these new vistas of treatment, these complications demand yet further innovation.

Another important requirement for the type of innovation necessitated by the advances, further needs, and complications associated with present solutions for arrhythmia patients is the need for platform technology. Such technology involves not only an enabling invention that allows the use of further multiple creative solutions for the same disorder but also for applying to a completely different organ system.

In this paper, we present the following technologies developed at Mayo Clinic Rochester, MN, USA.

  1. 1.

    A novel circuit and technique to reduce coagulum formation on indwelling catheters and on electrodes including during radiofrequency ablation energy delivery

  2. 2.

    A percutaneous subxiphoid intrapericardial technique to ligate the left atrial appendage to prevent thromboembolism in patients with atrial fibrillation

  3. 3.

    A newly developed intramyocardial bipolar pacing and sensing lead and a novel method to synchronously pace the ventricle without crossing the tricuspid valve or entering the coronary sinus

  4. 4.

    Technique, method, and devices to map and ablate in the central nervous system via the venous drainage of the brain

For each of these inventions, we describe the clinical problem attempting to be solved, results of our experience with the invention, and the scope of the problem solved with the innovation.

Reducing Invasive Procedural Risk by Preventing Coagulum Formation

Background

Thromboembolic stroke resulting from coagulum formation on catheters and ablation electrode surfaces is a serious complication seen with radiofrequency ablation procedures [13]. The primary method used by electrophysiologists to minimize this risk is intravenous heparinization prior to ablation in the left atrium and ventricle. Despite documented therapeutic activation clotting times from heparinization, about 2% of complex ablations result in thromboembolism [14, 15]. The reason why coagulum formed despite complete heparinization is likely due to the thrombin-independent nature of this process. Coagulum primarily results from direct heat-related denaturation of fibrinogen to form the fibrin clot, i.e., coagulum. Prior studies have shown that fibrinogen binds to the surface of catheters and sheaths within seconds of exposure [16]. Thus, a critical step in preventing coagulum formation on surfaces is preventing the adherence of fibrinogen itself. At neutral pH, fibrinogen is negatively charged, and this binding is possibly through electrostatic mechanisms [17]. Based on these facts, we hypothesized that maintenance of negative charge on catheters and ablation electrode surfaces during ablation or exposure of blood will reduce the risk of coagulum formation and thus also decrease subsequent thromboembolism.

The Invention

WeFootnote 1 developed a novel circuit (Fig. 1) that allows concurrent application of negative charge onto the ablation and mapping electrodes during radiofrequency ablation [18]. Important variations of the novel circuit and technique include the ability to simultaneously deliver and maintain negative charge on all electrode surfaces as well as the catheter body.

Fig. 1
figure 1

Illustration of the novel concurrent charge application circuit. The system can be used with any ablation or mapping catheter and can simultaneously place negative charge on the distal as well as the proximal electrode (see text for details)

Experimental Results

We first demonstrated, using atomic force microscopy, that fibrinogen molecules were weakly bound to a negatively charged surface. Conversely, with the application of a layer of positively charged nickel ions, fibrinogen adherence to the surface with high affinity resulted [19, 20] (Fig. 2).

Fig. 2
figure 2

Atomic force microscopy images acquired with a Veeco MultiMode AFM (Plainview, NY, USA). Top panel shows an image acquired in liquid with fibrinogen molecules deposited on a surface that attracts fibrinogen with high affinity. The clusters of white are fibrinogen molecules undergoing spontaneous thrombosis. Bottom panel: atomic force microscopy image showing fibrinogen on a negatively charged surface with streaks indicating movement of the fibrinogen molecule being displaced by the atomic force microscope probe

We then assessed [20] the effects of the application of charge during catheter placement without ablation using scanning electron microscopy (Fig. 3). Without negative charge, moderate amounts of coagulum were seen. When positive charge was applied, thick confluent coagulum coating the entire catheter tip was observed. When negative charge was applied, however, markedly decreased coagulum was observed on the catheter tip with large portions of the catheter tip being entirely free of coagulum [20]. Using high-magnification scanning electron microscopy, the classic lace-like pattern of fibrin clot was confirmed on the neutral or positively charged catheter tips. Following this, we assessed efficacy of charge application during ablation. In ex vivo experiments with ablation with negatively charged catheter tips, scanned surface clot coverage was noted similar to the results obtained without energy delivery (Figs. 4 and 5). These experiments were done by deliberately producing variable amounts of charge injection using variable resistors to define a therapeutic range in this model for the beneficial effects of charge delivery.

Fig. 3
figure 3

Close-up photographs of the electrode surface without concurrent charge application (a, d), with positive charge application (b), and negative charge application (c). An ex vivo experiment application of negative charge significantly decreased the surface electrode coverage with coagulum

Fig. 4
figure 4

Close-up photographs of the electrode surface with positive charge (left panel) and negative charge being applied during contact with blood. The bottom panels show the corresponding scanning electron micrographs. Confluent and multilayer coagulum is seen with positive charge being applied, whereas minimal coagulum is seen with negative charge with areas on high-power magnification showing no clot coverage

Fig. 5
figure 5

ad Similar data as shown in Fig. 4 with radiofrequency energy being delivered through the distal electrode. Marked decreased in surface electrode coverage with coagulum is noted when negative charge was applied (c) when compared to either neutral or positive charge

We then performed a series of in vivo canine experiments to test our technology in the beating heart (Lim et al., manuscript in progress). In these experiments, we have further confirmed the decreased adherence of fibrinogen on ablation electrodes and thus decreased char formation during electrode heating with radiofrequency ablation when negative charge was concurrently applied to the electrode surface (Fig. 6). These experiments have also verified that ablative lesions and electrogram integrity were similar with and without charge delivery. Importantly, the in vivo experiments have allowed us to demonstrate a therapeutic range for concurrent charge application with beneficial effects seen at significantly lower charge delivery than levels at which any arrhythmia, including minimal, ectopy is seen. An additional benefit in preventing coagulum on surfaces during direct current charge application documented in our study is the presence of surface electrolysis possibly acting as a physical “cleansing” agent supplementing the demonstrated fibrinogen-stabilizing effects of negative charge application.

Fig. 6
figure 6

Close-up photograph of the distal and proximal electrodes of an ablation catheter. In the left panel, negative charge was applied through our novel system during radiofrequency energy delivery with the catheter placed in the left atrium. In the right panel, negative charge was not delivered with the catheter placed at a similar location with identical power, temperature, and duration settings during radiofrequency energy delivery. Coagulum formed in this instance when negative charge was not being concurrently applied

Potential Application and Impact of Innovation

Thromboembolism remains a catastrophic complication despite present best efforts to prevent it during invasive left heart procedures. Since coagulum formation is not directly prevented with heparinization, our novel system of placing negative charge on the catheter and electrode holds promise for mitigating this risk during clinical ablation.

Heat-denatured fibrinogen (coagulum) once formed can be a nidus for clot propagation and larger soft thrombus formation. Further, even without coagulum statis-enlarged sheaths or catheter in the circulation may form thrombus. Since soft thrombus formation is proven to be inhibited with heparinization, initial use of our system will continue to require heparinization during atrial fibrillation ablation.

We are also evaluating our system [18] by placing charge on the catheter body, various sheaths, and on other nonablation implanted electrode surfaces attempting to prevent the deleterious effects of coagulum and subsequent thrombus on such surfaces.

Reducing Stroke Incidence by Ligating the Left Atrial Appendage

Background

The left atrial appendage is a common site for left atrial thrombus formation and subsequent stroke in patients with atrial fibrillation. Some studies have suggested that greater than 95% of thromboembolism based on TEE studies [21, 22] in patients with atrial fibrillation arise from this structure. Given the widespread prevalence of atrial fibrillation (expected to afflict 16 million by 2050) [23], this becomes significant. At present, the best established therapy for preventing thromboembolism has been warfarin which has been proven to reduce stroke risk in patients eligible for anticoagulation. Multiple problems with warfarin therapy including frequent blood draws, bleeding risk, and drug–dietary interactions result in significant underutilization of this therapy [24].

Because of these difficulties, new strategies have been directed at preventing left atrial appendage thrombus by excluding this structure from the circulation. These include surgical exclusion [25]; however, because of the morbidity associated with invasive surgical procedures, percutaneous approaches have been the nidus around which recent innovation has been based. A self-expanding nitinol cage with permeable fabric covering placed close to the ostium of the left atrial appendage as well as epicardial approaches without thoracotomy are being evaluated [2628]. An ideal approach, perhaps, is to have closure of the atrial appendage at the level of the ostium but without any portion of the device being exposed to the left-sided circulation.

The Invention

WeFootnote 2 have developed a system that involves ligation of the left atrial appendage via a percutaneous subxiphoid approach using a novel hollow-suture concept [28]. With the aid of local electrograms and fluoroscopy, the appendage is grabbed with an electrode-containing device. The ligator is then placed over the appendage and the hollow suture tightened (Fig. 7) with transesophageal echocardiographic documentation of adequate appendage closure ([29]; Friedman et al., submitted manuscript).

Fig. 7
figure 7

Illustration showing pericardial access with sheath placement. Through the sheath, the novel electrogram-directed grabber is advanced to stabilize the left atrial appendage. Over this, a loop which converts to a ligature at the base of the appendage is advanced

Experimental Results

In four live acute closed-chest canine experiments, we successfully deployed the device ligating the left atrial appendage. In this initial proof of concept study, there was no evidence of trauma to the left atrial appendage or pericardium[30] (Fig. 8). In our experiments, we used transesophageal echocardiography and fluoroscopic visualization to ensure closure of the appendage ostium when the ligature was applied. Additional parameters included the loss of atrial electrograms measured by the electrical mapping plant when closure was achieved. We are presently evaluating the system in closed-chest chronic survival experiments.

Fig. 8
figure 8

Following closed-chest pericardial placement of the ligature, the chest was opened to reveal successful placement of the ligature at the appendage base

Potential Application and Impact of Innovation

If further validated with chronic animal and human data, this minimally invasive percutaneous epicardial approach to ligate the left atrial appendage may eliminate or decrease the risk of thromboembolism in patients with atrial fibrillation. Further, this may be done without having an endovascular structure in the left-sided circulation.

Synchronous Ventricular Pacing Without Crossing the Tricuspid Valve or Entering the Coronary Sinus

Background

Most pacemakers and defibrillator leads are placed near the right ventricular apex. An additional atrial lead may be placed in the right atrial appendage with dual-chamber systems [31]. Despite the clear lifesaving advances that have resulted from endocardial pacing and defibrillation, some problems remain.

Present pacing and defibrillator leads sometimes record electrical signals from neighboring structures (far-field signals). With defibrillators, such abnormal sensing can give rise to inappropriate device therapy.

Since most ventricular leads are now placed in the right ventricle, they necessarily cross the tricuspid valve giving rise in some patients with symptomatic tricuspid regurgitation that may subsequently necessitate lead removal, repositioning, or valve surgery[32].

The most significant problem with right ventricular pacing, however, that has recently been described is the detrimental effects on myocardial function as a result of chronic right ventricular apical pacing ([33]; Henz et al., manuscript submitted to Circulation). Cardiac resynchronization therapy (CRT) using a left ventricular pacing lead in addition to the right ventricular lead attempts to overcome the problem of ventricular dyssynchrony and eventual deterioration in cardiac function. While CRT has shown significant benefits in improving quality of life, decreasing hospitalization, and improving symptomatic congestive heart failure[9, 34, 35], despite this major advance (CRT), problems still remain since cannulation of the coronary sinus in placing a left ventricular pacing lead is not straightforward, and a right ventricular lead is used in addition to the left ventricular lead, thus still necessitating crossing and possibly affecting the tricuspid valve [36].

To overcome these present limitations of endocardial pacing (far-field sensing, tricuspid regurgitation, and ventricular dyssynchrony promotion), we hypothesized that a totally intramyocardial bipolar pacing and sensing lead will limit far-field sensing, and, by placing this lead at a novel location (the atrioventricular septum), we would decrease ventricular dyssynchrony without the need to cross the tricuspid valve or enter the coronary sinus.

In the normal heart, the septal leaflet of the tricuspid valve is not at the same level as the mitral valve. This discrepancy results in a portion of the intraventricular septum being proximal to the tricuspid valve (Fig. 9). In this location, the right atrium and left ventricle myocardium are opposed. This unique anatomic location has left ventricular fibers oriented in a manner that would facilitate conduction to the free wall of the left ventricle [37, 38].

Fig. 9
figure 9

Left panel: photograph of the human heart atrioventricular septum. The septal leaflet of the tricuspid valve (arrow) was displaced more apically than the mitral valve (hatched arrow). Thus, a portion of the intraventricular septum is adjacent to and accessed from the right atrium. Right panel: histological section showing the layers of ventricular myocardium clearly present proximal to the insertion of the tricuspid valve leaflet

The Invention

WeFootnote 3 developed an intramyocardial pace-sense lead where both the anode and cathode are located within the myocardium (Henz et al., manuscript submitted to Circulation; [39, 40]; Fig. 10).

Fig. 10
figure 10

Left panel shows illustrations of placement of one iteration of the intramyocardial bipolar pacing lead onto the atrioventricular septum. The inset shows another iteration with a central pin serving as one electrode and an ultra helix serving as the other with both portions placed into the myocardium

The atrioventricular septum is a unique anatomical location (Fig. 9) which can be accessed without crossing the tricuspid valve and essentially has left ventricular myocardium exposed to the atrial endocardium. Since far-field sensing can be a major potential detriment of pacing this site, the intramyocardial bipole is necessary and enables pacing this novel anatomical site. We have designed and tested various bipolar pacing screw designs including housing the anode on a helix and cathode on a central pin, vice versa, two separate helices, and two deployed electrodes on a single screw. Several dimensions of the screw were tested given the variability in atrioventricular septal thickness in humans. If a screw larger than the septal dimension is deployed, a portion of the electrode would be exposed to the left-sided circulation. To avoid this complication and ensure stability of the lead being accurately deployed on the atrioventricular septum, we used a combination of intracardiac ultrasound visualization and fluoroscopy.

Experimental Results

  1. 1.

    We initially performed an open-chest experiment in the canine model verifying the improved sensing capability of intramyocardial bipolar pacing [39].

  2. 2.

    Following this verification, we performed in vivo endocardial deployment of the novel lead on the atrioventricular septum (Fig. 11) using a combination of high-density electroanatomic mapping from the endocardium, electroanatomic mapping from the epicardial surface of both right and left ventricles, electrocardiography, and tissue Doppler imaging via intracardiac ultrasound [41] (Fig. 12). We demonstrated relatively synchronous pacing (shorter QRS duration and relative preexcitation of the left ventricular free wall) when compared to right ventricular apical pacing. These findings were confirmed in three in vivo closed-chest canine experiments (Henz et al., manuscript submitted to Circulation; Fig. 13). This relative synchronous pacing occurred without tricuspid regurgitation, atrioventricular conduction abnormality, or entering the coronary sinus (Figs. 14 and 15).

Fig. 11
figure 11

Fluoroscopic image of intramyocardial pace lead placement during in vivo canine experiment. Arrows point to the lead placed near the right ventricular apex and on the atrioventricular septum. The intracardiac ultrasound probe is also seen

Fig. 12
figure 12

Intracardiac ultrasound image showing the atrioventricular septum (AVS). The linear-phased array probe is placed in the right atrium looking down on the atrioventricular septum (AVS). RA = right atrium; RV = right ventricle; TV = tricuspid valve; AVS = atrioventricular septum; LV = left ventricle

Fig. 13
figure 13

Photograph at necropsy showing placement of the pacing lead on the atrioventricular septum. Note that the results of synchronous ventricular activation were obtained without the lead crossing the tricuspid valve or entering the coronary sinus (CS). TV annulus = tricuspid valve annulus

Fig. 14
figure 14

Isochronal electroanatomic maps obtained during right ventricular apical pacing. The left panel shows endocardial activation viewed in an anterior projection. Note that activation begins (red area) and proceeds first to the septum of the left ventricle with the lateral wall of the left ventricle being activated last of all. The right panel shows similar activation in a posterior view with both endocardial and epicardial activation maps superimposed. Once again, the sequenced and dyssynchronous electrical activation (septum much before lateral wall) is seen

Fig. 15
figure 15

In contrast with Fig. 14, this electroanatomic map was obtained when pacing at the atrioventricular septum. Note the near-simultaneous activation of points on the lateral walls (blue) with a large portion of the anterior wall and septal apex, suggesting more synchronous activation

Potential Application and Impact of Innovation

  • Intramyocardial pacing may decrease far-field sensing that has plagued implanted defibrillators and removed some of the obvious advantages in having an atrial lead to help distinguish between supraventricular and malignant ventricular arrhythmia.

  • Since the atrioventricular septum is proximal to the septal leaflet of the tricuspid valve, tricuspid regurgitation would not occur with ventricular pacing.

  • If further chronic animal and human data confirm the relative ventricular synchrony that we showed in acute experiments by pacing the atrioventricular septum, the deleterious effects of right ventricle apical pacing may be prevented.

Noncardiac Arrhythmias—Different Locations, Similar Solutions: Mapping and Ablation in the Central Nervous System for Seizures via the Venous Drainage of the Brain

Background

Epilepsy is a prevalent disorder with over 50/100,000 persons affected per year [42]. Presently, when drug therapy for this disorder fails and patients have intractable and life-altering symptoms, surgical removal of the epileptogenic cortex is attempted. Although highly successful in eliminating the seizure focus, the procedure is highly invasive associated with significant morbidity. As a result of this, surgical management of seizures is presently uncommon. Adequate localization of the epileptogenic cortex and distinguishing this abnormal tissue from the neighboring eloquent cortex involves intracortical craniotomy-based mapping [12].

The evolution of cardiac arrhythmia management may provide insights into how epilepsy innovations might evolve. Three decades ago, treatment of what is now considered straightforward cardiac rhythm disorders involved open-heart surgery [43, 44]. The revolution in management of cardiac arrhythmia occurred when percutaneous endovascular approaches (mapping and radiofrequency ablation) evolved, initially as an adjunct to surgical therapy and now as a stand-alone curative approach. At the present time, surgical treatment for cardiac rhythm is rarely, if ever, required for disorders such as Wolf–Parkinson–White syndrome and AV node reentry (Fig. 16).

Fig. 16
figure 16

Top panel: endovascular mapping of cardiac arrhythmia to find the earliest site of activation or location of abnormal substrate and then targeted for ablation as revolutionized cardiac rhythm care. Bottom panel: similarly, with seizures, there is often an area of abnormal electrical activity that proceeds (arrow) global abnormalities and can potentially be targeted for focal energy delivery as envisioned in our invention

Since seizure mapping prior to surgical removal of abnormal cortex involves the recording and analysis of electrical signals (electrocorticograms), there may be avenues to use intravenous cardiac electrophysiological approaches and modify and apply them to the central nervous system[45, 46] (Fig. 17).

Fig. 17
figure 17

Illustration of the principle of accessing the jugular vein and cerebral venous system in this instance with an over-the-wire mapping and ablation catheter moved towards the focus of seizure activity

Further, in the heart, the primary diagnosis of myocardial infarction (a vascular phenomenon) is from the electrocardiogram (electrical parameters). Similarly, with vascular occlusion (stroke), there is a change in the nature of the electrical signals arising from this structure that is ischemic or infarcting in the brain. Whether or not similar approaches (electrical diagnosis for vascular events in the brain) could facilitate early diagnosis of patients with cerebrovascular disease and decrease the present devastating consequences of stroke is not known.

The Invention

WeFootnote 4 have developed a system that includes sheaths and specifically designed mapping and ablation catheters to access proximate cortical brain tissue via the central nervous system venous network [47]. This novel technique involves endovascular mapping using the superficial and deep-penetrating cortical veins along with the petrosal sinus, other sinuses, and the vein of Galen. These new techniques needed to cater to the unique structure of the brain, tortuosity of the vessels, and high impedance from brain tissue with delivering radiofrequency energy.

We further have developed a technique to detect via venous-based mapping early electrogram changes occurring as a result of cortical ischemia (Fig. 18).

Fig. 18
figure 18

Illustration showing various potential applications of the invention. Indwelling catheters, stimulators, and suppression devices for seizures and other disorders may all be accessed through the venous system (see text for details)

Experimental Results

  • Ex vivo cadaveric experiments were performed to successfully assess the feasibility of placing catheters via the jugular vein to reach the petrosal sinus and various branches up to and including the cortical veins.

  • Using fluoroscopic and angiographic guidance (Fig. 19), we performed an in vivo experiments with an entirely venous-based approach in the swine model and successfully mapped and placed histologically confirmed ablative lesions in the cortex of the pig [48] (Fig. 20).

  • We performed an open craniotomy experiment in the canine model with simultaneous venous, intracortical, and surface cortical electrogram recordings in the baseline state and with pacing-induced seizure activity and confirmed with a mathematical transform model the essential identical results obtained with venous-based mapping and plunge electrode cortical mapping.

  • We are presently assessing specifically designed catheters in other animal models to map and ablate cortical tissue and assess electrogram characteristics associated with early cerebral ischemia.

Fig. 19
figure 19

Fluoroscopic images showing venography performed to help access the venous system in the swing model. Catheters were placed through the vein to successfully map and create ablation lesions [48]

Fig. 20
figure 20

At necropsy, ablative lesions were found at the targeted sites over the cortex of the pig (left panel). Ablative lesions were further confirmed on histological examination (right panel)

Potential Application and Impact of Innovation

If further validated in other animal and seizure models, this technology may change the way in which patients with drug refractory seizures are treated. Further validation alongside present surgical mapping is required prior to fulfilling this promise.

The venous-based approach may provide a platform for placement of deep-brain-stimulating electrodes, other indwelling electrodes, and electrodes used to detect early ischemia and allow prompt treatment for impending stroke.

Summary

Despite the major strides made in the management of patients with rhythm disorders over the last few decades, innovative solutions continue to be required to improve the quality of these patients’ lives. We have briefly reviewed our group’s work attempting to reduce thromboembolism during invasive procedures [20] and decreasing stroke from atrial fibrillation with a percutaneous epicardial appendage ligation system. We have also described our work with a novel intramyocardial pacing lead placed in a unique anatomical position (atrioventricular septum) to synchronously pace the ventricles without crossing the tricuspid valve or needing to cannulate the coronary sinus. Finally, we discuss a novel technique to modify and use our knowledge obtained from endovascular approaches to cardiac arrhythmia management to the central nervous system to treat and detect common disorders such as seizures and stroke.