1 Introduction

Carbonaceous-based materials are extensively used in many research fields, potential applications spanning from electric double-layer capacitors [1], energy production [2], field emission [3] or electrochemistry [4]. Carbon is a versatile element, and can be present as various allotropes including graphite, diamond and fullerene-like structures. Belonging to the carbon family, carbon nanotubes [5] (CNTs) consist of seamlessly rolled up graphene layer(s), meanwhile only graphene differentiates as a two-dimensional material, made of a single monolayer of carbon atoms [6]. Graphene is considered to be important in the electrochemistry field [7], since it has high electron conductivity [8], fast heterogeneous electron-transfer rate [9], high surface area [10] and also it can be obtained at affordable costs [11].

The relevance of β-nicotinamide adenine dinucleotide (NADH) as cofactor in naturally occurring enzymatic reactions, as well as the potential applications in developing NADH-based biosensors are still considered as key aspects in the bio-electro-analytical research field [1218]. The electrochemistry of the redox couple NAD+/NADH has been intensively studied by Gorton [19]; and the formal redox potential (E°′) of the NAD+/NADH is known to be −560 mV versus SCE [20]. Even if it is still controversial, however, a sequential electrochemical-chemical-electrochemical mechanism is widely accepted to be the path undertaken in NADH oxidation. Direct electrochemical oxidation of NADH to its corresponding oxidised form (NAD+) at the bare/unmodified electrodes is highly irreversible, requires large activation energy, and proceeds with coupled side reactions, poisoning the electrode [21, 22]. The decrease of the anodic oxidation overpotential may be achieved using redox mediators [2325], carbon in various hybridizations and forms [2628], boron-doped diamond [29], metallic sputtered CNTs [30], single crystal gold [31] or carbon paste electrode modified with nickel oxide nanoparticles [32]. The advancement of graphene-based biosensors has been reviewed [33, 34] and the NADH-based biosensors relaying on the use of graphene or graphene derived materials (such as GO or reduced GO) were reported by several worldwide research groups [3538].

In this paper, we report the preparation and analytical characterisation of multi-layered graphene and graphene oxide (GO)-modified glassy carbon (GC) electrode used for NADH. This electrochemical sensing probe was further utilised as transducer for the development of an l-lactate biosensor. This l-lactate biosensor, working at an applied potential of +250 mV versus Ag/AgCl reference electrode, was used to determine l-lactic acid in several types of yogurts, illustrating that the concentration of l-lactic acid was correlated with the percentage of fat in the analysed samples.

2 Experimental

2.1 Materials, reagents and instrumentation

Graphene was supplied by NanoIntegris, USA (PureSheets™ QUATTRO) as a water-based dispersion (0.05 mg/mL) containing 2 % (w/v) ionic surfactant. Graphite flakes (SP1) were from Bay Carbon, Inc., USA. Spectral graphite rods (d = 2 mm) were provided by Elektrokarbon a.s. (Slovak Republic).

NADH (purity ≥97 %) was purchased from Santa Cruz Biotechnology, Inc. (http://www.scbt.com/). l-lactate dehydrogenase (~140 IU/mg), NAD+ (>95 %, HPLC), l-lactic acid, Nafion® 5 % (v/v) solution, NaNO3, H2SO4, KMnO4, HCl, H2O2, hydrazine monohydrate, isopropanol, diethyl ether potassium phosphate dibasic, potassium dihydrogen phosphate, potassium chloride, potassium permanganate and all the other reagents used for buffers were from Sigma–Aldrich. PTFE-membrane filters (0.5 μm pore size) from Carl Roth GmbH (http://www.carlroth.com) were also used. All chemicals from commercial sources were of analytical grade. A colorimetric assay kit (R-Biopharm AG, Darmstadt, Germany) for l-lactic acid was also used.

Glassy carbon electrodes, d = 3.0 mm, were supplied by Bioanalytical Systems (BAS, West Lafayette, USA). PK-3 electrode polishing kit (0.05 μm alumina and 1 μm diamond) was purchased from Bio-Logic SAS (http://www.bio-logic.info). Electrochemical investigations were performed using the Autolab electrochemical system (Eco Chemie, The Netherlands). All electrochemical measurements were performed with a three electrode configuration. Potentials were measured versus the Ag/AgCl reference electrode obtained from BAS West Lafayette, USA. A platinum wire was used as the auxiliary electrode. Nova NanoSEM 630 (FEI Company, USA) was used for scanning electron microscopy (SEM) investigations and Allegra X-22 (Beckman Coulter) centrifuge was used for yogurt sample pre-treatment. Yogurt samples were from local suppliers.

2.2 Graphene oxide synthesis and graphene preparatory

Graphene oxide was prepared accordingly to Hummers method [39], slightly modified [40]. Basically 2 g of graphite flakes and 1 g of NaNO3 were mixed with 46 mL of concentrated H2SO4, under continuous stirring and controlled temperature (5 °C). KMnO4 (6 g) was added and the temperature was maintained below 5 °C for at least 60 min; the reaction was allowed to proceed by carefully monitoring the temperature in order not to exceed 35 °C. The reaction slurry was gradually diluted using 300 mL of deionized water and the unreacted KMnO4 was neutralized by adding 5 % v/v solution of H2O2 (50 mL). The reaction mixture was filtered on a sintered glass filter (medium pore size); sulphate salts were removed by using HCl solution (5 % v/v) and plenty of deionized water. The obtained cake was dried under vacuum, at 80 °C, during 48 h. A stable dispersion of GO (1 mg/mL) was achieved after ultrasonication, for about 30 min in water, and further used to modify the electrodes as described in Sect. 2.3.

Surfactants are necessary to avoid agglomeration and precipitation of graphene platelets [41], but removal of surfactant is a key aspect for graphene to preserve its intrinsic properties. The commercial available graphene consisted of: 27 % single layer, 48 % double layer, 20 % triple layer and 5 % four or more layers and does not have any graphite impurities. A detailed technical specification sheet, including also atomic force microscopy characteristics is available at: www.nanointegris.com. Since graphene was supplied as water-based dispersion containing 2 % (w/v) ionic surfactant, we proceed with the removal of the surfactant. For such a reason a certain volume of the received graphene dispersion was mixed with isopropanol and diethyl ether (1:1:2 v/v/v), by using a glass rod. Allowing few minutes to settle down, graphene platelets gathered in a fine layer, at the interface between the organic (i.e. diethyl ether) and the aqueous phase (i.e. mixture of isopropanol and water), by selective partition and floated at the interface due to the large surface available (2,630 m2/g) [10]. The graphene suspension was removed and dispersed again by ultrasounds in 50 mL of distilled water; again the above described process was repeated five times, and in the end, the purified graphene platelets were dried under vacuum, at 80 °C, during 48 h. A stable dispersion of graphene (1 mg/mL) was achieved in water:ethanol (1:1 v/v), with ultrasonication for 30 min, and further used to modify the electrodes as described in Sect. 2.3.

2.3 Preparation of modified electrodes

Several types of electrodes were prepared for our electrochemical studies. Prior modification, GC electrodes (d = 3.0 mm) were mechanically polished (PK-3 Electrode Polishing kit), immersed in distilled water in an ultrasonic bath (20 s.) and electrochemically cleaned, in 0.1 M H2SO4 solution using cyclic voltammetry, by performing 35 scans, at 50 mV/s, sweeping between −300 mV and +1,200 mV. Glassy carbon electrodes were modified by using 4 μL from the respective graphene or GO dispersions and allowed to dry under a moderate stream of nitrogen. Lactate biosensors were prepared by adsorbing l-lactate dehydrogenase on graphene-modified electrode and thereafter 5 μL of 1 % (v/v) Nafion® neutralized solution were casted and allowed to dry. The l-lactate biosensor, when not used, was stored in 10 mM phosphate buffer saline (PBS), at 4 °C. After investigating the effect of enzyme loading, 20 IU of l-lactic dehydrogenase (l-LDH)/electrode were selected to be the best enzyme loading among 5, 10, 20 and 40 IU of L-LDH. The concentration of the cofactor (NAD+) was also optimised. It was found that 3 mM NAD+ was the best concentration, (from 0.5, 3.0 and 6.0 mM investigated concentrations of cofactor), to be used for the analytical determination of l-lactic acid, with the proposed biosensor. The optimised pH value was found to be 7.5. The biosensor working applied potential was +250 mV versus Ag/AgCl reference electrode.

2.4 Sample preparation

Commercially available yogurt samples were bought from local markets. Samples were centrifuged during 15 min at 6,000 rpm, and the supernatant was carefully collected, filtered through 0.5 μm PTFE-membrane filters and diluted, prior analysis, in 10 mM PBS.

3 Results and discussion

3.1 Characterisation of graphene oxide and graphene

Scanning electron microscopy (SEM) investigations revealed that there was a significant difference in the morphology of graphene and GO. SEM micrographs illustrated in Fig. 1a, b depicted wrinkled graphene sheets with uneven and swollen margins, while home-made GO was flat with even edges. Energy dispersive X-ray spectra indicated that the ratio of C to O was ~6:1 in graphene, meanwhile for GO this ratio was roughly 1:1, due to the presence of oxygen functionalities (i.e. hydroxyl, epoxy, carboxyl groups, etc.) on the surface of graphene sheets, and also indicated that the oxidation process has been achieved. Graphene supplied by NanoIntegris is supposed to be entirely pristine graphene platelets that have not been oxidised, reduced or chemically modified in anyway, and we assumed that the presence of oxygen in EDX spectrum is due to adsorbed oxygen. Moreover, the presence of sodium indicated that traces of surfactant were still present, even after five washing steps.

Fig. 1
figure 1

Microscopic images illustrating: a wrinkled graphene with uneven edges after surfactant removal (inset the energy dispersive X-ray spectrum is depicted), b flat graphene oxide with sharp edges (inset the energy dispersive X-ray spectrum is also depicted; Si and Au peaks are ascribed to the support substrate), c graphene oxide atomic force microscopy with height profiles

Atomic force microscopy (Fig. 1c) performed on GO illustrated that the flake thickness (~0.8 − 0.9 nm) was typical for single-layered GO [42].

3.2 Electrochemical characterisation of the modified electrodes

In order to rationalise on the electrocatalytic behaviour of the modified GC electrodes, as a preliminary assay, we used potassium ferricyanide (1 mM) in supporting electrolyte to investigate peak separation. As it can be seen from Fig. 2a, in the case of the unmodified electrode the peak potential separation (ΔE = EpaEpc) was 110 mV (Epa = +283 mV, Epc = +173 mV), while for the GO-modified GC electrode ΔE = 100 mV (Epa = +273 mV, Epc = +173 mV). Instead graphene-modified GC electrode (ΔE = 70 mV, Epa = +263 mV, Epc = +193 mV) illustrated to be able to drive the electron transfer reaction faster than both bare or GO-modified electrode. The obtained results are in good agreement with the results reported in other studies, illustrating the fast electron transfer behaviour on graphene, for single-electron-based electrochemical reaction [33]. Brownson et al. [43, 44] observed that heterogeneous electron kinetics is influenced by the global coverage of graphene on the electrode surface. Two distinct cases were described: “Zone I”—corresponding to an incomplete coverage of the underlying electrode surface, predominance of graphene basal planes and reduced electrochemical reactivity, and “Zone II”—made of layered graphene, governance of edge planes and thus increase in heterogeneous electron transfer and improved electrochemical response. By doing this coverage study, we were able to illustrate the case of “Zone II” (Fig. 2b), where increased amounts of graphene casted on the GC electrodes produced a decrease in peak potential separation, approaching the graphite case (Fig. 2b—inset). Peak potential separation for bare graphite rod was 60 mV and for 2, 3 and 4 μL graphene (1 mg/mL graphene stock solution) casted GC electrodes ΔE varied from 120 to 90 mV and finally to 70 mV, respectively. As reported in [43] the ΔE changed with the amount of graphene immobilised on the surface of the underlying electrode, reduced peak-to-peak separation illustrating improved electron kinetic ascribed to electrochemically active edge plane sites in “Zone II.” In order to discriminate between thin-layer effects and diffusional response at the multi-layered graphene-modified electrode, scan rate was varied in the presence of 1 mM potassium ferricyanide as redox probe. Plots illustrating the logarithm of the anodic peak current versus the logarithm of the scan rate (Fig. 2c) inset c1) gave a slope coefficient of 0.445 close to 0.5. Corroborating this data with the linear correlation obtained from plotting peak current versus the square root of the applied scan rate (Fig. 2c) inset c2), we can assume that the response of the electrode is governed by diffusion.

Fig. 2
figure 2figure 2

Cyclic voltammetry recordings illustrating: a the behaviour of bare glassy carbon electrode (GC_bare), graphene (GC_G) and graphene oxide (GC_GO)-modified electrodes, b the electrode coverage study and relationship between the mass of graphene deposited upon the electrode surface and the resultant peak separation (inset) and c the behaviour of GC_G electrode at different scan rates illustrating that the response of the electrode is purely diffusional (inset C1 and C2)

3.3 Electrochemical behaviour of NADH at the modified electrodes

Considering the performance of an electrochemical sensor, the properties of the electrode material are the most significant. The electrochemical response of graphene sheets (single-, few- and multi-layer graphene) has shown to be independent of the number of layers [7] and does not have tremendous advantages over graphite micro-particles [45]. Nonetheless the combination of its interesting electrochemical properties [43, 44, 4648] and its peculiar physiochemical properties made graphene beneficial to be utilised in a range of sensing and energy related electrochemical applications. Keeley et al. [48] used exfoliated graphene, obtained in dimethylformamide after prolonged ultrasonication, to investigate the detection of NADH. Under Ar atmosphere, electrochemistry experiments were achieved using a GC electrode modified with graphene nano-sheets. Chronoamperometry, conducted at +0.4 V versus Ag/AgCl, illustrated a LOD in the low micromolar range (1.9 μM) for NADH and a stability improvements over bare GC electrode.

In our experiments peak oxidation (Ep) of NADHat bare GC electrode occurred at +625 mV versus Ag/AgCl. Graphene oxide (GC_GO)-modified electrodes were able to detect NADH at Ep = +525 mV versus Ag/AgCl, lowering by 100 mV the potential but with a concomitant decrease (roughly 50 %) in current intensity. Graphene-modified GC electrode was able to drop further the overpotential at Ep = +280 mV versus Ag/AgCl, indicating a decrease with ~350 mV when compared with the bare GC (Fig. 3a). Also in this case, as reported by Keeley et al. [48] and Goh and Pumera [7], the electrocatalytic effect was ascribed to the presence of large electrochemically active area of graphene [10] and edge-plane sites on multi-layered graphene, since it is no particular advantage in using monolayer graphene instead of few- or multi-layer graphene as electrode material. Ascorbic acid (AA) is one of the most common interfering specie in the electrochemical determinations of NADH. Nonetheless, ascorbic acid capitalized also on the large electrochemically active area of the graphene and its edge-plane sites, the anodic oxidation peak being situated at ~+400 mV versus Ag/AgCl (Fig. 3b). In amperometry ascorbic acid produced a significant positive interference (+125 %, for 10−4 mol L−1 ascorbic acid and NADH). This relevant issue is acute for the majority of the carbon-based modified electrodes and could be minimized either by electrochemical pre-treatments [49] or by using ascorbate oxidase [50].

Fig. 3
figure 3

Cyclic voltammetry recordings (50 mV/s) illustrating: a the behaviour of bare glassy carbon electrode (GC_bare), graphene (GC_G) and graphene oxide (GC_GO)-modified electrodes in the presence of 2 mM NADH, b the behaviour of graphene-modified electrode in the presence of NADH and ascorbic acid

The reluctant behaviour of GO to further promote the electron transfer reaction or to improve the electrochemical properties of the GC electrode can be ascribed to its conductivity, smaller than that of graphene; moreover GO has a resistivity of about 188–413 kΩ/m [51] and, by nature, it is considered as an insulating nanomaterial [52]. Amperometric measurements carried out at constant applied potential (+250 mV vs Ag/AgCl reference electrode), performed in a stirred solution at pH 7.4, clearly indicated that the modified electrode is suited as NADH amperometric sensor (Fig. 4). It can be observed that a nonlinear dependence of the anodic current on NADH cofactor concentration in the whole calibration range was obtained, as would be anticipated for a Michaelis-Menten type process. When considering GO-modified GC electrode, the linear working range (LWR) was very narrow, comprised between 1 × 10−4 ÷ 2.5 × 10−3 M (RSD = 6.5 %, n = 3, y = 5.8 × 10−8 + 0.1 × 10−3 X, r = 0.9969), while the multi-layered graphene-modified GC electrode displayed improved analytical features (LWR = 5 × 10−6 ÷ 1 × 10−4 M, RSD = 4.5 %, n = 3, y = 4 × 10−9 + 11.1 × 10−3 X, r = 0.99458). The limit of detection (LOD) was calculated as 3s b/S, where s b is the standard deviation of the blank signal and S is the sensitivity. For the GO-modified GC electrode, the calculated LOD was 1 × 10−4 M, meanwhile for the graphene-modified GC electrode, the LOD was 1.0 × 10−6 M.

Fig. 4
figure 4

Amperometric recordings of the graphene (GC_G) and graphene oxide (GC_GO)-modified electrodes in the presence of successive additions of NADH in supporting electrolyte, inset a illustrates the response of GC_G electrode in the presence of 0.001 and 0.005 mM NADH and inset b depicts the linear working ranges for GC_G and GC_GO electrodes

3.4 l-lactic acid determination in yogurts

l-lactate biosensor was developed relaying on the GC electrode modified with multi-layered graphene. l-lactic acid in the presence of the oxidised form of β-nicotinamide adenine dinucleotide (NAD+) is enzymatically converted to pyruvic acid, in the presence of (l-LDH) (Equation 1).

$${{\sc{l}}}{\text{-lactic acid}} + {\text{ NAD}}^{+} \mathop {\Leftrightarrow} \limits^{{{\sc{l}}{\text{-LDH}}}} {\text{Pyruvic acid}} + {\text{NADH}} + {\text{H}}^{+}$$
(1)

Amperometric sensing of l-lactic acid in samples was achieved by coupling the l-lactate dehydrogenase with the sensor for NADH. Based on equation 2, NADH was oxidised at the electrode, thus producing an oxidation current directly proportional with the l-lactic acid concentration.

$${\text{NADH }}\Rightarrow {\text{NAD}}^{ + } + {\text{H}}^{ + } + 2{\text{e}}^{ - }$$
(2)

Finally, a calibration plot for l-lactic acid, in buffer, was generated (Fig. 5). The linear working range was comprised between 3 × 10−5 and 6 × 10−4 M, and the calculated LOD was 1.9 × 10−5 M (LOD was calculated as 3sb/S, where sb is the standard deviation of the blank signal and S is the sensitivity). Table 1 summarise the analytical characteristics of the developed biosensor compared to previously l-lactic biosensors reported in literature. The results obtained, using the l-lactate biosensor, were compared with the data from the colorimetric kit for l-lactic acid determination acid in foodstuffs, obtaining a good degree of correlation. The content of l-lactic acid in yogurt samples is illustrated in Table 2.

Fig. 5
figure 5

Amperometric recordings of the L-Lactate biosensor in the presence of successive additions (0.5, 1.0, 2.0, 4.0, 8.0, 16, 32, 64, 128, 256 μL of 0.1 M l-lactic acid in 5 mL constant cell volume) in supporting electrolyte (10 mM phosphate buffer saline (PBS) + 0.1 M KCl, pH 7.4); inset the linear working range is depicted

Table 1 Analytical features of several l-lactic acid biosensors reported in literature
Table 2 l-lactic acid content in several yogurt types from local supermarkets

Lactose, milk fat, proteins, citrates and their enzymatic degradation products are considered as the source of flavour compounds in yogurt. During milk incubation with yogurt cultures, minimum 15 % of lactose is fermented to lactic acid, the degree of lactose utilisation and the rate of lactic acid production depending on bacteria and their symbiotic effect [5355]. It has been reported that lactic acid sources in yogurts can be considered both proteins and fat, beside lactose [5658]. Taking this into consideration, we consider the estimated l-lactic acid content in the analysed yogurts correlates well both with the increasing fat content (from 0 % up to 5 %) and with the data reported in literature [5658].

4 Conclusions

Pristine graphene platelets and GO were used as electrode modifiers, aiming the investigation of their electrochemical efficacy towards NADH. The peak potential separation (ΔE) of the unmodified GC electrode in the presence of 1 mM ferricyanide was 110 mV, while after the modification of GC with graphene and GO, the electrode peak potential separation was 70 mV and 100 mV, respectively. When NADH, our analyte of interest was used, cyclic voltammetry experiments revealed that the graphene-modified GC was able to decrease the anodic peak potential with ~350 mV, when compared with the bare GC, meanwhile in the case of GO-modified electrode the decrease was relatively moderate, i.e.~100 mV. Relying on the graphene-modified transducer, (l-LDH) was successfully immobilised in a 1 % Nafion® membrane. The developed biosensor, working at +250 mV versus Ag/AgCl reference electrode, was used to assess l-lactic acid in four different types of yogurts, revealing an l-lactic acid concentration ranging between 0.3 and 0.6 %. The analytical performances of the biosensor reported were more or less similar in terms of linear range and detection limits (in the micromolar range) and were comparable with those of the l-lactic acid biosensor reported in literature [5963].