Introduction

In the United States, 80% of adults experience lower back pain secondary to degenerative spinal diseases in the lumbar spine in their lifetime.31,42 Spinal fusion has been the gold standard surgical procedure for the treatment of severe lumbar degenerative diseases.18 However, radiographic follow-up studies revealed that 39–86% of fusion patients may develop adjacent segment degeneration (ASD),12,38 and 5–38% received revision surgery due to adjacent segment diseases within ten years after primary surgery.12,37 Although some believe that ASD is merely a natural progress of degenerative changes in the lumbar spine, altered segment biomechanics are widely assumed to be causative factors for ASD development.34,40

In vitro cadaveric tests and computational simulations have been widely used to investigate adjacent segment motion and mechanical behavior before and after various fusion surgical treatments using the load-control, displacement-control, and hybrid protocols.1,10,13,27 In general, these previous studies predicted that spinal fusion causes increasing ROMs at the adjacent segments using a displacement-control or hybrid protocol, but no or slight changes in ROMs using a load control protocol.48 However, an accurate representation of the in vivo physiological loading conditions in the lumbar spine is still a challenge in cadaveric tests and computational simulations.48 Previous in vivo studies investigated the changes of lumbar overall and treated/adjacent segment ROMs after fusion mainly in flexion-extension.5,6,9,26 Albeit a degree of variability in these reported in vivo data depending on different experimental setups, generally, no distinct kinematic changes occur at the segments adjacent to a fused segment, whereas the overall lumbar ROM after fusion substantially decreases after spinal fusion.33 That is, in vivo measurements reported in the literature did not strongly support the assumption that greater motion at adjacent segments occurring to compensate for the overall spinal ROMs in functional activity leads to the changes in the adjacent disc biomechanics. So far, the underlying mechanisms of how the biomechanics, including strains and stresses, in the adjacent discs are altered by spinal fusion have not been thoroughly understood.

In this study, we utilized a combined in vivo and in silico method (Fig. 1) to investigate the changes of biomechanical responses of adjacent segments after lumbar fusion in patients under physiological loading conditions. The in vivo adjacent disc heights under the weight-bearing upright standing condition were measured using a dual fluoroscopic imaging system (DFIS) with previously validated image registration techniques pre- and post-operatively. The measured in vivo intervertebral positions and orientations at adjacent segments were then used as displacement boundary conditions in three-dimensional (3D) patient-specific finite element (FE) disc models to simulate the changes of adjacent disc stress/strain distributions of patients caused by fusion. It has been suggested that excessive distraction at the fusion level could potentially increase pressurization of adjacent segments,24,25 as a consequence of the inter-segment interactions contributed by surrounding spinal structures which have adapted to the spinal degenerative status prior to fusion.56 Moreover, previous simulation of the lumbar spine during neutral standing using a musculoskeletal model showed that muscle damage due to the surgical procedure also plays a role in the increase of adjacent segment loading.32 Therefore, we hypothesized that fusion can elevate strains/stresses in the discs at the adjacent segments under the weight-bearing upright standing condition.

Figure 1
figure 1

A flowchart of the combined in vivo and in silico method for analysis of the biomechanics of the adjacent discs after lumbar fusion surgery. (BCs boundary conditions).

Methods

In vivo Image Collection and Processing

Five symptomatic patients (four males, one female; mean age: 57.0 ± 10.7 years; mean height: 169.7 cm; mean body weight: 65.7 Kg) were included after the approval by the internal review board (Protocol Number: 2011P000314) at Massachusetts General Hospital.41 All the patients suffered from low back pain due to degenerative spinal diseases at L4-5 and L5-S1 segments, and were scheduled for fusion of both the segments. Following the diagnosis, all patients underwent lumbar fusion treatments (2 with PLIF, 1 with ALIF + PLIF, and 2 with ALIF; where PLIF = posterior lumbar interbody fusion, and ALIF = anterior lumbar interbody fusion) at L4-5 and L5-S1 by two full-time fellowship-trained spine surgeons at Massachusetts General Hospital within 2 weeks.

Prior to fusion treatments, each subject underwent computed tomography (CT) scanning (SOMATOM Definition AS+, Siemens, Forchheim, Amsterdam, Germany) of the lumbar spine (from L1 to S1) in a supine, relaxed position. Parallel sagittal CT images with a space of 0.6 mm between two slides and a resolution of 512 × 512 pixels each having a pixel size of 0.519 × 0.519 mm were obtained. CT segmentation of L3, L4, L5, and S1 (Fig. 2a) were performed in the Rhinoceros 5.0 3D modelling software (Robert McNeel & Associates, Seattle, WA) using a previously reported segmentation technique,49 to reconstruct 3D anatomical models of these vertebrae (Fig. 2b).

Figure 2
figure 2

Image data collection and processing. (a) Segmentation of L3, L4, L5, and S1 from CT images. (b) The segmented 3D L1-S1 lumbar spine model, in which the L3-4 intervertebral space was zoomed to present the shortest distances at selected locations, i.e., anterior (A), center (C), posterior (P), left (L), and right (R) sites. (c) The upright, weight-bearing standing position adopted in biplane fluoroscopic imaging of patients’ lumbar spines. (d) Registration of 3D vertebral models to the pre-operative biplane fluoroscopic images in upright standing. (e) Registration of 3D vertebral models to the post-operative biplane fluoroscopic images in upright standing following fusions at L4-5 and L5-S1.

Following CT scanning, the lumbar spine of each subject in the upright, weight-bearing standing position (Fig. 2c) was imaged using a dual fluoroscopic imaging system (DFIS), in which two C-arm fluoroscopes (BV Pulsera, Philips, Netherlands) were orthogonally aligned. The biplanar fluoroscopic images captured in the standing position were imported into the Rhinoceros modelling software, in which a virtual DFIS (Fig. 2d) was created using a well-established 3D-to-2D registration technique.30,50 The vertebral positions and orientations of the lumbar segments (L3 ~ S1) in the standing position were reproduced by matching the projections of the 3D vertebral models to their radiographic vertebral outlines/features on the pair of fluoroscopic images (Fig. 2d). Three years after the fusion treatments, the alignments of these patients’ lumbar segments (L3 ~ S1) in the weight-bearing standing position were evaluated again using the same dual fluoroscopic imaging protocol with the 3D-to-2D registration procedure (Fig. 2e).

To minimize the non-surgical effects on the pre- and post-operative standing positions, a standard standing posture (Fig. 2c) was defined with detailed requirements: (1) keep the head level and do not push the head forward, backward, or to either side; (2) stand straight with the shoulders pulled backward; (3) keep both the hands on the thighs and lie down straight; (4) keep both feet together. After the patients sat down for a rest of at least 30 min, they were asked to achieve the standard standing posture as possible as they could during fluoroscopic imaging; they could hold a handle settled on the C-arm of a fluoroscope if discomfort or pain occurred. The lumbar spine of each subject was imaged by the DFIS at 30 Hz. In data processing, 3D vertebral models were registered to multiple pairs of pre-/post-operative fluoroscopic images, when the captured frame series became steady. We eventually selected the pre-operative and post-operative registered lumbar models with minimal differences in their alignments between each other.

A previous validation of the registration technique using the same fluoroscope modality showed that the overall accuracy in determining intervertebral translations was 0.4 mm by comparing the in vitro measurements of a testing machine for ovine cadaveric spine vertebral models, and the average repeatability of the DFIS technique was ±0.3 mm and ±0.7° in determining in vivo intervertebral kinematics.50 Thus, it was expected that the registration technique could be feasible and repeatable in determining in vivo intervertebral alignment of the human lumbar spine.

In vivo Measurement of Intervertebral Heights at the Adjacent Segment

In the CT supine lumbar spine model, two right-hand Cartesian local coordinate systems were defined at the disc proximal and distal endplates of each segment, respectively. The geometric center of the endplate was chosen as the origin of the local coordinate system. The relative positions/orientations of the proximal disc endplate in six degrees of freedom (DOFs) with respect to the distal disc endplate at the adjacent L3-4 level were determined using these local coordinate systems. Subsequently, the local coordinate systems were mapped to the dual fluoroscopic image-registered models to measure L3-4 intervertebral positions/orientations in the pre-operative and post-operative standing positions.

Regional disc heights of the adjacent disc at the L3-4 level were evaluated based on the relative alignment of the disc endplates in the pre- and post-operative weight-bearing standing positions. The definition of local disc heights conformed to the previous description.51 As shown in Fig. 2b, we first positioned the anterior, center, posterior, left, and right points of the proximal disc footprint in the CT configuration; these points were mapped to registered models in the pre-operative and post-operative standing positions to measure weight bearing disc heights. The corresponding points on the distal disc endplate surface, which can be located at the vertices, edges, and facets of the distal disc endplate surface mesh, were determined by searching the shortest distances using an online available algorithm19 in MATLAB R2017b (MathWorks, Natick, MA). At last, local disc heights were represented by the projections of these shortest distances along the normal direction of a plane fitted to the distal endplate surface.

In Silico Evaluation of Pre- and Post-Operative Adjacent Disc Biomechanics

The adjacent disc geometry at the L3-4 segment in the CT supine position was reconstructed in the Rhinoceros software. The disc shape was simply modeled by lofting the marked profiles of adjacent vertebral bony endplates,4 so disc bulging at the annular internal and external walls36 was not considered. The nucleus pulposus (NP) and annulus fibrosis (AF) were scaled, such that the ratio of the NP volume to the total disc volume was 44%, with the NP center approximately 2 mm posterior to the disc center.45 The disc geometry was meshed to 8-node hexagonal solid finite elements (which could be reduced to 6-node wedge elements in the NP) using the Altair HyperWorks V13.0 software (Altair Engineering Inc., Troy, MI), as presented in Fig. 3a. To ensure appropriate element aspect ratios, the solid element size of the AF was gradually increased from the posterior to the anterior. The solid element counts and sizes of both the AF and NP in the disc FE models were summarized in Table 1. The resulting disc FE model was imported into the ANSYS 15.0 MAPDL software (ANSYS Inc., Canonsburg, PA) to perform pre-processing, solving, and post-processing, as described below.

Figure 3
figure 3

Finite element modeling of the adjacent disc at the L3-4 segment. (a) The transverse and sagittal cross-section of the disc model. (b) Six layers (Lry 1–Lry 6) of fiber lamellae which were embedded within the AF. The alternating fiber angles (°) of the innermost and outermost lamellae measured with respect to the disc transverse plane in different polar sectors were marked; fiber angles at other sites were linearly interpolated. (c) Radially and circumferentially varying fiber stress-stretch relations from outer (Lyr 1) to inner (Lyr 6) layers in different polar sectors.

Table 1 The solid element counts and sizes of both the AF and the NP in the five adjacent (L3-4) disc FE models

Well-documented normal disc material properties were assigned to the FE models of adjacent discs, assuming that no tissue deterioration/disruption at the adjacent segments occurred after fusion. Linear elasticity (Young’s modulus, \(E\) = 24 MPa and Poisson’s ratio, \(\upsilon\) = 0.4)15 was assumed for two cartilaginous endplates, each with a thickness of 0.5 mm.39 The Mooney-Rivlin hyperelastic material model was used to simulate the NP material (\(c_{1}\) = 0.05 MPa, \(c_{2}\) = 0.01 MPa, and the bulk modulus, \(\kappa\) = 50 MPa) and the AF ground substance material (\(c_{1}\) = 0.20 MPa, \(c_{2}\) = 0.01 MPa, and the bulk modulus, \(\kappa\) = 6 MPa).55 Six layers of fiber lamellae (modeled using ANSYS smear reinforcing elements, Fig. 3b) in a crossing fiber pattern were embedded within the AF ground substance.58 Radially and tangentially varying fiber angles (marked in Fig. 3b)57 and tangentially varying lamellar thicknesses (anterior: 0.73 mm, lateral: 0.56 mm, posterior: 0.39 mm)57 were assigned to reinforcing elements. To simulate the gradual transition of Type I and II collagen content within the annulus,8 regional fiber stress-stretch relations (implemented by an ANSYS USERMAT subroutine, Fig. 3c)58 were fitted and interpolated from previously reported single lamellar tensile test data.21

In FE analysis, a pilot node was created at the origin of the proximal disc endplate local coordinate system in the CT-scanned disc configuration, which represented the reference (unstrained/unstressed) geometry (Fig. 4a). All nodes of the proximal endplate surface were rigidly coupled to the pilot node using ANSYS multipoint constraint techniques. To simulate disc deformation in the pre-operative/post-operative standing positions, the changes in the six DOFs on the proximal endplate surface, including three translations (\(T_{x}\), \(T_{y}\), \(T_{z}\)) and three rotations (\(\theta_{x}\), \(\theta_{y}\), \(\theta_{z}\)) were applied to the pilot node, while all nodal DOFs of the distal disc endplate surface were fully constrained. A detailed description regarding the calculation of six DOFs can be found in Appendix A. We have validated that the resulting nodal positions of the proximal disc endplate surface in FE analysis could exactly match the proximal disc endplate footprint of the deformed disc geometries in the pre-operative and post-operative standing positions, as shown in Figs. 4b and 4c, respectively.

Figure 4
figure 4

Validation of the application of six DOFs in FE analysis by examining if the nodes of the proximal disc endplate surface exactly align on the proximal disc footprint in the reference (CT supine) and deformed (pre-/post-operative standing) L3-4 segment models. (a) The reference configuration segmented from CT images; (b) The pre-operative standing configuration after the registration to dual fluoroscopic images; (c) The post-operative standing configuration after the registration to dual fluoroscopic images.

Data Analysis

The regional in vivo disc heights of the adjacent discs in the pre- and post-operative standing positions were measured and compared. Using FE analysis, von Misses strains, annular fiber stresses, and intradiscal pressures (i.e., the average hydrostatic pressure in the NP) of the adjacent disc in the weight-bearing pre- and post-operative standing positions were simulated. The differences in disc strains and annulus stresses before and after lumbar fusion were also calculated using the corresponding nodal solutions of the pre- and post-operative disc FE models, and the distributions of both strain and stress differences were mapped to the post-operative deformed disc FE models to visualize. All results were presented using “median (CI)”, where CI is the 95% confidence interval. A paired, two-sided Wilcoxon signed-rank test (IBM SPSS Statistics Version 24, Chicago, IL) was used to compare the disc heights, disc strains, annulus stresses, and intradiscal pressures (IDPs) at the adjacent level before and after lumbar fusion. Statistical significance was set as p-value less than 0.05, and the powers of significant differences were calculated using post hoc power analysis (G*Power Version 3.1.9.6, Universität Kiel, Germany).

Results

Measurements of Adjacent Disc Heights

In the weight-bearing standing position, the adjacent disc heights were 12.8 (11.2, 13.7) mm, 9.7 (7.9, 10.2) mm, 4.6 (3.9, 6.3) mm, 7.1 (6.1, 9.2) mm, and 8.3 (6.4, 9.7) mm at the anterior, center, posterior, left, and right locations, respectively, before fusion, and were 11.3 (9.5, 13.1) mm, 9.4 (7.8, 9.7) mm, 3.6 (3.1, 5.5) mm, 6.5 (5.6, 8.5) mm, and 8.3 (6.7, 10.1) mm, correspondingly at the five locations after fusion (Fig. 5). Only the disc height at the posterior location was significantly decreased by − 0.8 (− 1.2, − 0.4) mm (p < 0.05, power = 73%) after the fusion treatment (Table 2 and Fig. 5).

Figure 5
figure 5

Regional in vivo disc heights (mm) at the adjacent L3-4 segment measured at anterior, center, posterior, left, and right locations (Fig. 2b) in the pre- and post-operative weight-bearing standing postures. Significant differences (p < 0.05) before and after fusion were marked by asterisk signs, and their statistical powers (%) were presented in parentheses.

Table 2 A summary of the changes in the adjacent disc heights and simulated responses (represented by medians and 95% CIs) at different disc regions from the pre-operative to the post-operative standing positions

Simulation Results of Adjacent Disc Responses

Strains of the Adjacent Disc

The maximal strains of the whole disc in the standing positions before and after fusion were 0.41 (0.32, 0.56) mm/mm and 0.70 (0.57, 0.75) mm/mm, respectively, and both were located near the outermost and posterior annulus portion (Figs. 6b and 7). The maximal increase of the strain after fusion was 0.32 (0.21, 0.43) mm/mm (p < 0.05, power = 89%), which occurred at the posterior annulus portion, followed by the peak changes of 0.24 (0.19, 0.43) mm/mm (p < 0.05, power = 67%) at the left annulus portion and 0.29 (0.18, 0.35) mm/mm (p < 0.05, power = 96%) at the nucleus/center portion (Fig. 6c and Table 2).

Figure 6
figure 6

Simulated strains (mm/mm) of the adjacent discs of the five subjects (viewed from the top). The five patients were denoted by #1 ~ #5. (a) The disc reference geometries in the CT supine position; (b) The disc strain distributions in the pre- and post-operative standing positions; (c) The strain difference distributions between both standing positions.

Figure 7
figure 7

Predicted regional peak strains (mm/mm) measured in anterior annulus, nucleus (center), posterior annulus, left annulus, and right annulus (Fig. 3b) in the weight-bearing standing postures before and after lumbar fusion. Significant differences (p < 0.05) before and after fusion were marked by asterisk signs, and their statistical powers (%) were presented in parentheses.

Annular Stresses of the Adjacent Disc

The maximal stress across the whole annulus before fusion was 2.10 (0.86, 4.26) MPa, occurring in the outermost lamellae of the posterior annulus portion in the standing position (Figs. 8b and 9). After fusion, the peak stress at the posterior annulus portion was 2.99 (2.14, 4.12) MPa, followed by the peak annulus stress of 2.87 (2.38, 4.61) MPa occurring at the left annulus portion. It was observed that the annular stresses in the outermost lamella were significantly increased by 1.70 (1.07, 3.60) MPa at the left annulus portion (p < 0.05, power = 69%), and by 1.20 (0.57, 2.25) MPa at the anterior annulus portion (p < 0.05, power = 61%) (Fig. 8c and Table 2).

Figure 8
figure 8

Simulated annulus stresses (MPa) of the adjacent discs of the five subjects (viewed from the posterior). For better visualization, the fiber stresses in the fiber lamellar elements (Fig. 3b) have been mapped to the annular ground substance elements. The five patients were denoted by #1 ~ #5. (a) The annulus reference geometries in the CT supine position; (b) The annular stress distributions in the pre- and post-operative standing positions; (c) The stress difference distributions between both standing positions.

Figure 9
figure 9

Predicted regional peak stresses (MPa) measured in anterior, posterior, left, and right annulus portions (Fig. 3b) in the weight-bearing standing postures before and after lumbar fusion. Significant differences (p < 0.05) before and after fusion were marked by asterisk signs, and their statistical powers (%) were presented in parentheses.

Intradiscal Pressure of the Adjacent Disc

As shown in Fig. 10, the intradiscal pressure (IDP) of the adjacent disc in the standing position was 0.37 (0.20, 0.42) MPa before fusion and 0.66 (0.43, 0.80) MPa after fusion. The intradiscal pressure was significantly increased by 0.29 (0.13, 0.47) MPa (p < 0.05, power = 78%) after the fusion treatment (Table 2).

Figure 10
figure 10

The predicted intradiscal pressures (MPa) in the weight-bearing standing postures before and after lumbar fusion. The significant difference (p < 0.05) before and after fusion was marked by an asterisk sign, and its statistical power (%) was presented in parentheses.

Effects of Different Fusion Treatments

The five fusion patients investigated in this study underwent different fusion approaches at L4-5 and L5-S1. The effects of these fusion approaches on the anatomic and biomechanical changes of the adjacent (L3-4) discs were presented in Table 3. It was observed that especially at the L4-5 segments, ALIF caused greater distractions than PLIF. Correspondingly, greater changes in the disc heights, strains, and IDPs were measured at the immediately adjacent segments for those patients who underwent ALIF than PLIF.

Table 3 The effects of different fusion approaches performed at L4-5 and L5-S1 of the five patients on the changes of disc heights, strains, and IDPs of the adjacent (L34) disc in the upright standing position

Discussion

This study utilized a combined in vivo (DFIS) and in silico (3D FE modelling) technique (Fig. 1) to investigate the in vivo biomechanical changes of the adjacent disc at the L3-4 segment, after L4-S1 underwent spinal fusion in five patients. Our measurements and simulation results revealed decreases in adjacent disc heights, as well as elevations in disc strains, annular stresses, and intradiscal pressure after spinal fusion. It was observed that the maximal increases in strains and stresses were located within the outermost annular lamellae at the posterior and lateral portions of the adjacent disc, respectively (Figs. 6 and 8). The mismatch between the locations of maximal strain and stress changes could result from the non-homogeneous annular material properties (Fig. 3c); fibers in the lateral annulus are tauter (narrower toe regions) and stiffer (steeper linear regions) than the posterior annulus. Overall, we demonstrated the feasibility of the combined in vivo and in silico technique, and the preliminary data supported our hypothesis that spinal fusion reduced adjacent disc heights but elevated strains/stresses in the adjacent discs. In addition, it appeared that ALIF was related to greater changes in the disc heights and biomechanical responses at the adjacent segments than PLIF. However, care should be taken to interpret the effects of surgical approaches due to the small sample size in this study, as it has been reported that ALIF resulted in a lower incidence of ASD than PLIF in previous clinical studies.6

In this study, the anatomical dimensions of the adjacent segments were determined in the weight-bearing standing position that was consistently performed by each patient before and after fusion surgery. The disc heights were represented by the projections of minimal distances measured between the proximal and distal endplates of the disc at selected locations along the normal direction of the distal endplate surface.49,51 Therefore, this enabled evaluating the anatomical changes of the discs in the weight-bearing positions of patients before and after the fusion treatment in a consistent manner. Furthermore, the measured in vivo intervertebral positions and orientations were introduced to patient-specific FE models as the displacement boundary conditions to simulate the stress/strain responses of the adjacent discs. It avoided assumptions regarding the preloading in the neutral position (e.g., introducing the follower preload43 or standing-specific compressive loads)54 in computational FE simulations. Therefore, this combined in vivo and in silico method could quantitatively evaluate anatomic and biomechanical changes in adjacent discs after fusion treatments under physiological loading conditions.

According to clinical results ten years after 360° fusions, Schulte et al. showed that the average reduction in the disc height at the proximally adjacent segment amounts approximately 20%.46 In contrast, in the group of the patients three years after PLIF and ALIF in our study, the adjacent disc heights only decreased by ~ 3% at the disc center but by ~ 17% at the posterior disc location. A recent study by Lay et al. evaluated adjacent disc height changes following extreme lateral interbody fusion using 3D CT scanning of patients’ lumbar spines pre-operatively and post-operatively within seven days.29 They reported that the proximally adjacent segment was significantly decreased by an average of 0.50 mm, with significant decreases in local disc heights at the nucleus portion, the posterior annulus portion, and the right annulus portion. In general, our data are in agreement with their findings that fusion could result in decreases of the adjacent disc heights, depending on different measurement locations. However, differences need to notice; we evaluated the adjacent disc height changes during standing upright, and the follow-up was three years after fusion.

In terms of disc biomechanical responses, our simulation predicted that the pre-operative intradiscal pressure was 0.37 (0.20, 0.42) MPa at the L3-4 segments in the standing posture; by comparison, it was reported that those at the L4-5 segments of the normal spines were 0.40 MPa (range: 0.23–0.50 MPa) predicted using a similar combined in vivo and in silico technique in a recent study4 and 0.5 MPa measured in vivo using a needle pressure transducer.52,53 However, a consistent comparison of the responses (strains/stresses) of the adjacent discs before and after the fusion treatment lacks in the literature. It is worth noting that three aspects could result in the lower pre-operative IDP at L3-4 in our simulation compared to published data. First, the IDPs of discs in the lumbar spine are level-dependent, and the L3-4 disc could be less loaded and pressurized than the L4-5 disc.7 Second, the previous simulation modeled the nucleus as fluid,4 but we considered the compressibility of the nucleus by assigning a bulk modulus of 50 MP,55 since the nucleus of the adjacent discs in fusion patients might have undergone a loss of water content. At last, healthy discs in vivo are swelled by high water content in the nucleus,35,47,58 thus higher IDPs were measured in vivo using needle transducers.52,53

This study also reinforces the notion that there is a link between distractions at the treated segments and anatomical/biomechanical alterations at the immediately adjacent segments after spinal fusion.24,25 Contemporary fusion surgery primarily aims to restore the degenerative intervertebral space to “normal” disc heights and lordosis,22 whereas the “normal” height/lordosis for a degenerative disc is unknown, since degenerative changes in patients have been developed at the diseased segment for decades prior to treatments. As a result, it is difficult to avoid an excessive distraction at the instrumented level in spinal fusion, and such operation may increase compression of the adjacent discs, which could be associated to progressive disc tissue disruption and degenerative changes.2,3,23 It should be noted that even a slight decrease in the disc height can result in a marked increase in disc compression. A previous in vitro experiment showed that the disc load soared to approximately 1000 N, when only a compressive displacement of 1 mm was applied to a disc specimen.36 However, it is also acknowledged that not all fusion patients would experience the development of post-operative ASD.28 A clinical study reported that the incidence of post-operative radiographic changes at adjacent segments reached 42.6% in a minimal 5-year follow-up of a total of 188 fusion patients.11 Hence, it is important to quantitatively determine the threshold values of the altered loading that could accelerate the ASD development. The combined in vivo and in silico evaluation provides a useful approach in clinical prognosis of ASD following spinal fusion, by establishing thresholds in various biomechanical measures as used in this study.

Several limitations in this study should be acknowledged when interpreting our data. Only five patients were included for the evaluation of adjacent segment biomechanics after fusion in this study, due to revision surgery or losing follow-up, albeit our initial recruitment of 14 fusion patients. Therefore, the power of statistical analysis and analysis of the effects of different fusion treatments on adjacent disc heights and biomechanical responses were limited by the small sample size. Since it is challenging to collect the data of patients with the treatments at the same levels, future studies should include a large group of fusion patients through a collaboration of multiple medical centers, to further examine how fusion surgery affects adjacent disc responses and establish their thresholds causing ASD. Additionally, we only investigated the patients in the weight-bearing upright standing position in this study. It is also necessary to more comprehensively evaluate pre- and post-operative adjacent disc responses in functional spinal activity, weight lifting, and level walking.

Previous validation studies of the 3D-to-2D registration technique for the same fluoroscope modality showed that the accuracy in determining translations was 0.4 mm;50 in this study, the change of the adjacent disc height especially at the center site (Table 3), to which the simulated IDP changes could be sensitive, was near or within the accuracy limit. As generally required in computational simulation of spinal responses using in vivo boundary conditions,4,17 higher accuracy (less than 0.1 mm) in determining in vivo DOFs17 is desirable to detect radiographic and biomechanical changes following fusion. Moreover, changes in disc responses do not only depend on changes in axial translations, but also on changes in disc inclinations in both the sagittal and coronal planes. Hence, it also requires high accuracy in determining rotational DOFs. Future work should adopt newer dual fluoroscopes with clearer radiographic imaging to improve the reliability of image registration, and relevant validation of the accuracy and repeatability using upgraded modalities is also warranted.

In this study, well-documented material properties of normal discs were adopted for all the FE models of adjacent discs. Since the MRI images of the patients before and after fusion were not available, it limited our capability of creating high-fidelity disc FE models in terms of the disc geometry and material properties. Correspondingly, we did not consider tissue deterioration/disruption potentially developed in adjacent discs. It should be interpreted that the biomechanical responses simulated in this study are the upper bounds of the actual responses (e.g., creep in the adjacent disc may reduce disc stresses). In the future, it is anticipated to establish the correlation of disc material properties with MRI signals (e.g., T2 mapping values),16 such that patient-specific tissue properties can be incorporated to enhance the biofidelity of the disc FE models. In terms of upright standing, disc behavior was almost not influenced by the anterior and posterior longitudinal ligaments when displacement boundary conditions were applied (Appendix B), although both ligaments closely attach to the anterior and posterior portions of the disc annulus. However, these surrounding spinal structures should be considered in the simulation of disc responses in functional spinal motions, which would cause large disc deformation and bulging.

Conclusion

In conclusion, a combined in vivo and in silico method was proposed to investigate the changes in adjacent disc anatomy and biomechanical responses before and after lumbar fusion. This study demonstrated the feasibility of this proposed method, and showed that fusion could lower the adjacent disc height, and elevate intradiscal pressures and disc stress/strain primarily in the outermost lamellae of the posterolateral annulus portion of the disc. The proposed method can help future studies more systematically and comprehensively investigate biomechanical factors associated with the accelerated development of ASD in fusion patients.