Abstract
Mixing in enclosed culture micro-chambers is an important criterion on achieving the long-term cell growth with more consistent characteristics spatially distributed in the microfluidic environments. Here, we report a microfluidic mixer composed of multiple deformable membranes to drive circulation flows within the culture region in a peristaltic manner. This mechanically driven mixing scheme has the advantages over many others existing mixing schemes by inducing negligible shear stress over cells without side effects to the viability and growth characteristics. The membrane movements can induce moving boundary motions of the liquid volume contained in the culture chamber. The flow characteristics such as the velocity profile and shear stress are investigated by lumped-element modeling and computational simulations. Both experiments and simulations are performed to show the effectiveness of the mixing of both soluble substances and sub-microscale particles, including bacteria. The mixing performance under different operation parameters (e.g., membrane size and membrane switching time) is also investigated for the optimized operation. Further, the dental bacteria Streptococcus mutans are cultured for 2 days to demonstrate that the reported mixing scheme can generate a more even distribution of growth, which may be further applied for a uniform dental biofilm development in vitro for the related biofilm research. Additionally, this micro-mixer is highly compatible with the widely used soft lithography technique, and hence, it can be directly integrated with general microfluidic devices for extended biomedical diagnosis and bio-sample processing applications.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
In the past decades, microfluidics has been widely applied in chemical and biological applications in the microscales. For instance, microfluidics plays an important role in the long-term cell cultivation and analyses as micro-bioreactors, in which cells may grow and react with selected reagents for drug screening and discovery purposes (Materne et al. 2013; Wu et al. 2008). To maintain the long-term cell viability in the micro-bioreactors, fresh culture media should be regularly perfused into the incubation micro-chambers to refresh nutrients and growth factors. However, this essential process may generate gradients of soluble factors in the incubation region, i.e., with higher concentrations of nutrients and growth factors in the inlet, and thereby cells near the inlet may then grow faster than those in the rest of the chamber region. In the worst case, cell clusters near the inlet may eventually clog the chamber entrance. Therefore, it is important to develop a micro-mixer within an enclosed culture region in order to avoid the mentioned hurdles (e.g., inlet clogging and growth inconsistency) for the long-term cell incubation and analysis applications using microfluidics.
In recent years, many micro-mixers have been developed to achieve more rapid mixing (Capretto et al. 2011; Lee et al. 2011a; Mansur et al. 2008; Nguyen and Wu 2005). These mixers can be briefly categorized into two major groups: the ‘active’ mixers and ‘passive’ mixers for the designs with and without an additional actuator/energy source for the mixing enhancement, accordingly. For the passive mixers, Fu et al. (2014) combined a nozzle and a double-heart chamber with a Y-shaped injection channel. Due to the symmetrical rotating vortex structures, a significant perturbation of the species streams is induced to make more efficient mixing. Tofteberg et al. (2010) reported a lamination mixer with a planar lid on the top of a structured channel. With this structure, flows are made with a 90° rotation between each recombination and splitting. Abbas et al. (2013) developed a continuous-flow micro-mixer based on the channel deflection. Further by generating a chaotic advection using the channel deflection, the mixing efficiency was enhanced significantly. Ansari et al. (2010) designed a passive micro-mixer by splitting the main channel into two sub-channels with different widths. Due to the different flow rates in the sub-channels, unbalanced molecular collisions were created every time when two fluid streams combined. This kind of channel mixers (e.g., the T- or Y-shaped micro-mixers) could only operate with a continuous flow, where detached cells from clusters were removed with the flow. However, the continuous flow condition may not match the required microenvironments for some long-term cell analyses applications such as the biofilm development on tooth or bond surfaces, on which the liquids are static or nearly static.
Micro-mixers functioning inside enclosed microfluidic regions often require complex fabrication, integration and operation (Lin et al. 2007; Zhu and Kim 1998), limiting their practical applications. Typically, these mixers are ‘active’ mixers requiring additional flow induction. The commonly used actuating sources are the electric field (Park and Wereley 2013; Yan et al. 2009) and the magnetic field (Peng et al. 2011; West et al. 2002). Yet, the extra electromagnetic fields may cause direct effects to the cell behaviors and viability, due to factors such as the abundant ions in the growth media and the ion channels over cell membranes. For example, the electric field would enhance the growth of human keratinocyte cells (Vianale et al. 2008), and an electromagnetic field (frequency: 60 Hz) could alter the activity of ornithine decarboxylase in human lymphoma cells and mouse myeloma cells (Byus et al. 1987). Kim et al. (2009) demonstrated a thermal technique by alternating the actuations of two heaters to induce natural convection in the micro-chamber to enhance the mixing rate. This technique should not be applicable in long-term cell applications as the cell growth is very sensitive to the microenvironmental temperature fluctuations. Alternatively, mechanical flow actuation seems to be a more feasible approach. For example, Lee et al. (2011b) worked on a micro-mixer composed of three interconnected deformable growth chambers. The corresponding mixing was achieved through deflections of the PDMS membranes in a peristaltic manner. While highly functional, it was very challenging to integrate the components into general micro-bioreactors since each culture region was split into three deformable chambers. Hsiung et al. (2007) and Koh and Marcos (2015) demonstrated mixers utilizing moving membrane structures, activated pneumatically by the external air pressure. The mixing efficiency was reasonably good, but cells could be flushed away by the continuous, and hence, the long-term cell culture application was challenging. Xia et al. (2014) utilized a micro-oscillator to generate oscillatory flows and further the induced flow segmentation in order to enhance mixing efficiency. Fu et al. (2013) proposed to apply the pressure disturbances, functioning as the chaotic vortex micro-mixer. Lin et al. (2014) proposed a bubble-driven mixer by driving an air flow through a wooden diffuser to generate bubbles to mix the fluids. However, the gaseous contents in the bubbles could change the culture environment by altering the concentration of dissolved oxygen and carbon dioxide. The piezoelectric micro-mixers were developed to exert a directed physical oscillation at the ultrasound frequencies on the fluid body (Nguyen and Wu 2005). Liu et al. (2002) and Marcos et al. (2014) developed a mixing technique based on the principle of cavitation micro-streaming and bubble-induced acoustic micro-streaming. Nevertheless, an excessive power level of the generated ultrasonic wave may lead to physical damages on the cells. Jang et al. (2007) employed a piezoceramic/silicon diaphragm to generate acoustic streaming to homogenize mixing solution by rapidly switching the resonant frequencies, but the ultrasound caused significant effects on the cell growth (Pitt and Ross 2003). Altogether, there is still a demand for the design of a microfluidic mixer, which can be integrated into high-throughput micro-bioreactors and induces only negligible effects on the growth of individual cells, biofilms or tissues.
Here, we introduce a microfluidic mixer operating in a closed liquid volume, consisting of six deformable membranes, located around the outer area of the ceiling of a cylindrical micro-chamber. The membranes have been grouped as two three-membrane peristaltic flow drivers, in which each membrane was driven by an external computer-controlled air pressure. We may consider the membrane movements as moving boundary motions of the culture chamber, in order to generate circulation flows within the chamber to enhance mixing. We have run simulations to investigate the corresponding flow characteristics. We have also conducted experiments for further verification and quantification of the mixing performance. Dental bacteria named Streptococcus mutans have been cultured in a micro-chamber, in order to demonstrate improvements of cell growth with the assist of the reported mixing scheme.
2 Methods
2.1 Device fabrication
The microfluidic mixer included mainly four layers of microstructures: the control channels, the flow channels, the chambers and a glass substrate (located from top to bottom). The design layout is shown in Fig. 1a. The microfabrication was based on multilayer soft lithography (Unger et al. 2000) (the more detailed procedures are described in Supplemental Information and Fig. S1). A fabricated device is shown in Fig. 1b.
2.2 Simulation for flow profiles
We performed computational studies to validate and analyze liquid velocity and pressure profiles generated by the mixer operation. We first applied a lump-element model to compute the membrane deflections during the mixing operations. These estimated deflections and velocities of the membranes were then imposed as moving boundaries and boundary fluxes of the micro-chamber, respectively. Then, we used commercial simulation software (COMSOL 4.3b, COMSOL Inc., Burlington, MA, USA) to compute the transient flow profiles in the three-dimensional finite-element chamber model. We adopted the laminar flow model instead of solving the full form of Navier–Stokes equations to reduce the computational cost. In fact, the flows can be considered as laminar due to the low Reynolds number (Re ≪ 1), calculated as described later in flow characteristics. Water (density: 103 kg/m3; viscosity: 10−5 cm2/s) was set as the flowing liquid in the chamber. The purpose of these simulations was to obtain the repeating flow profiles during mixing. In practice, with the zero-velocity initial conditions, simulations were implemented for multiple mixing cycles (typically seven cycles should be sufficient) until the stabilized repeating profiles were obtained. The resultant flow profiles were then used for the following simulations to obtain the transient concentration profiles of both soluble molecules and particles.
2.3 Experiments for mixing characterization
We applied the microfluidic device to mix color dye (McCormick) and fluorescent beads (Cat# FP-0545-2, carboxyl fluorescent light yellow particles, Spherotech. Inc., Lake Forest, IL) to characterize the mixing for soluble molecules and particles (with diameters bigger than 100 nm including bacteria). The bead samples were diluted in 5 % pluronics (Sigma) at a concentration of 0.01 % w/v in order to prevent the adhesion of beads on the chamber walls. In the experiments, we first filled the chamber and channels with water. The selected samples with ~20 % of the chamber volume were injected to the chamber through one of the solution inlets. We then captured videos under a fluorescence microscope during the mixing operations of different mixing parameters for further analyses.
2.4 Image analysis
We customized MATLAB scripts (version 8.2, MathWorks, Novi, MI, USA) to quantitatively analyze the mixing performance during mixing based on intensity changes in a micro-chamber. Briefly, we applied image processing on the snapshots at different time points during the mixing operations. We calculated the normalized standard deviations (SD) (Voth et al. 2003) of the snapshots in the chamber region, with the value at the beginning of mixing to be ‘1’ and at the stabilized status to be ‘0.’
2.5 Device preparation for cell culture
To prepare for cell culture experiments, each microfluidic mixer was first sterilized by baking at 100 °C for >10 h and UV exposure for >2 h. All control channels and flow regions were initially primed with pressurized DI water with an upstream pressure of ~20.7 kPa. Culture media, distilled water and phosphate-buffered saline (PBS) were then inserted into the medium inlet, and solution inlets one and two, respectively. We placed the mixer on a microscope fitted with an enclosed temperature-controlled shield to provide a stable operating temperature of 37 °C. We then inserted the S. mutans bacteria sample (ATCC 25175, ~104 cells/ml) to fill up the micro-chamber using the peristaltic micro-pump lying along the outlet channel of the device. Afterward, we isolated the bacteria by closing the valves at the chamber entrance and exit and incubate for 30 min, to achieve the initial cell attachment.
3 Results and discussion
3.1 Mixer design and basic operation
We have developed a microfluidic peristaltic mixer composed of six micro-membranes fabricated around the ceiling of a micro-chamber. The layout of a proof-of-concept device, which contained a layer of control channels overhead the flow-channel layer, is shown in Fig. 1a. As described in Fig. 1c, the six micro-membranes were grouped as a pair of three-membrane flow drivers working simultaneously with a repeating peristaltic sequence of pressurization of the three membranes: Membrane 1, Membrane 2 and Membrane 3 in the left or right side of the chamber. The pressurization was achieved by connecting the inlet control channel for each membrane to a compressed-air source to drive the membrane deflection. While the micro-chamber (diameter: 1 mm; height: 80 µm) was closed by the entrance and exit valves, peristaltic actuation cycles were continuously applied to each three-membrane flow driver. Each mixing cycle contained three pressurization steps: step 1 with Membrane 1 on and Membrane 2 and Membrane 3 off; step 2 with Membrane 2 on, and Membrane 1 and Membrane 3 off; and step 3 with Membrane 3 on, and Membrane 1 and Membrane 2 off. Accordingly, the mixing operation generated two symmetrical flow circulations inside the chamber (indicated with arrows).
As mentioned, micro-valves were fabricated at the entrance and exit of the micro-chamber to control for the liquid accessibility. A side channel was located beside the culture chamber such that liquids could flow bypass the micro-chamber. Additionally, there existed a medium channel for injections of fresh culture media on one side of the chamber (right in Fig. 1a). A microfluidic peristaltic pump which consisted of three valves was fabricated downstream of the micro-chamber to drive liquid flows from the medium channel, the buffer solution channel, or one of the eight inlet channels controlled by a microfluidic multiplexer. The eight solution inlets were designed as reserves for potential cell research applications, such as buffer solutions (e.g., PBS), growth factors, supplementary biochemical stimuli for cell responses, and reagents for analyses of cell characteristics.
3.2 Lumped-element model of membrane deflection
To reduce the computational cost and complexity, we decomposed the simulation into three major sub-tasks: (1) determination of moving boundaries of the micro-chamber using a lumped-element model for the membrane dynamics, (2) simulation of the flow velocities in the chamber using the computed moving boundaries and (3) calculation of transient distributions of solutions and particles based on the simulated velocity profiles.
To implement the first sub-task, we simplified the model by considering the physical symmetry of the chamber geometry and operation and taking only a half chamber with the three moving membranes shown in Fig. 2a. Major effects of the membrane dynamics should be the membrane stiffness and the fluidic resistances along the connecting control channels prefilled with water. Here we represented the integrative effects by the equivalent electrical circuit composed of lumped elements. Considering the analogy of constitutive relations in mechanical and electrical domains, we expressed the flow rates as electrical currents and the hydraulic pressures as voltages. Hence, the fluidic resistances along three control channels were represented by electrical resistors (R 1, R 2 and R 3), and the mechanical stiffnesses of the membranes were equivalent to capacitors (C 1, C 2 and C 3). R 1, R 2 and R 3 had identical flowing liquid (water) filled with the same length and channel dimensions, and hence, they were equal to the value R (Senturia 2001),
where μ is the viscosity of fluid along channel and L C , W C and H C are the length, width and height of the water sections filled in the channels, respectively. Further, we estimated the membrane stiffness (Timoshenko et al. 1959) as the same value C, due to the same membrane material (PDMS) and dimensions (the calculation is described in Supplemental Information):
where L M , W M and H M are the membrane length, width and thickness, E PDMS [=750 kPa (Armani et al. 1999)] is the Young’s modulus of PDMS and ν PDMS (≈0.5) is the Poisson’s ratio of PDMS, respectively.
We defined the liquid flow rates along control channels as the electrical currents (i 1, i 2 and i 3), the driving pressure at the control channel inlets as the input voltages (S 1, S 2 and S 3) and the pressure differences across the membranes as voltages (V 1, V 2 and V 3). During the mixing operations, the chamber was isolated (without net inflow and outflow) and the fluidic resistance of the chamber should be negligible, compared to the resistances of the control channels. We expressed this effect by linking the membrane capacitors at a common node following the Kirchhoff’s current law. We then obtained the following relations:
Alternatively, we could write all these relations as a linear system in the state-space form:
We defined S 1, S 2 and S 3 as functions of time, during one mixing cycle (300 ms) were in the form as plotted in Fig. 2b, including the switching of membrane pressurization every 100 ms. The driving pressures at the actuation states were set at a consistent level (~137.9 kPa), matching with the experiment configurations. We then computed for the corresponding stabilized one-cycle transient responses of the half-chamber mixer model to obtain the transient V 1, V 2 and V 3, as functions of time. In practice, we obtained the stabilized responses (Fig. 2c) by simulating 20 mixing cycles with zero initial conditions and extracted only the last-cycle response.
V 1, V 2 and V 3 were then converted to spatial deflections of the membranes (Θ1, Θ2 and Θ3) in the local membrane coordinate frames with local origins at the membrane centers as described in Fig. 2a:
Θ1, Θ2 and Θ3 should be reasonable approximations as long as the membranes do not contact with the chamber base at any time during the mixing cycle. For this reason, we chose S 1, S 2 and S 3 at a sufficiently low level, i.e., 137.9 kPa.
3.3 Flow characteristics
We used commercial simulation software (COMSOL 4.3b, COMSOL Inc., Burlington, MA, USA) to perform simulations of flows in a chamber under moving boundaries calculated by the lump model. We first considered a template mixing design with the membrane dimensions: r M = 400 µm, L M = 200 µm, W M = 150 µm and H M = 20 µm. However, both the simulation and experiment studies have shown that this design could not induce significant mixing performance (data not shown). Based on this template design, we considered two modifications of the membrane dimensions by either reducing the membrane thickness (model I: r M = 400 µm, L M = 200 µm, W M = 150 µm and H M = 10 µm) or enlarging the membrane areas (model II: r M = 360 µm, L M = 230 µm and W M = 220 µm and H M = 20 µm), in order to reduce the membrane stiffness. The simulated profiles of membrane displacements, flow velocities and shear stress over the chamber base of model I are shown in Fig. 2d–f, whereas the profiles for model II are available in Fig. S2. Because the generated circulating liquid flows (average velocity for model I: 0.363 × 103 µm/s and average velocity for model II: 0.178 × 103 µm/s) were sufficiently slow, the Reynolds numbers were ~10−20 and the corresponding acoustic Reynolds numbers (Friend and Yeo 2011; Sayar and Farouk 2011, 2012, 2015; Yeo and Friend 2009) were also very small (≪1). Notably, fluctuations in hydraulic pressure in the chamber (<10 Pa for both models) were much smaller than the changes of driving pressures (137.9 kPa) because the pressure drops happened mainly in the control channels (R 1, R 2 and R 3). These pressure fluctuations induced only negligible further deflections of C 1, C 2 and C 3. Additionally, the maximum membrane downward deflections were 29.6 and 38.6 µm for model I and model II, respectively. This double confirmed that there was no mechanical contact between the membranes and the chamber base. Further, we have calculated also the shear stress over the bottom surface of the micro-chamber τ(x, y, t)| z=0:
where u and v are the local velocity components along x- and y-directions, respectively. We chose Δz = 1 µm such that Δz is sufficiently small for a reasonable approximation. Results (Fig. 2f) show that the maximum shear stresses (model I: ~0.173 kPa and model II: ~0.102 kPa) should not cause any harmful effects on the cell viability (Lange et al. 2001; Legazpi et al. 2009; Scragg et al. 1988; Shiragami and Unno 1994) and shear-induced movements (Hyon et al. 2012; Koh and Marcos 2015; Marcos et al. 2009, 2011, 2012, 2014; Marcos and Stocker 2006).
3.4 Mixing of soluble substances
We performed simulations by COMSOL 4.3b, with the convection–diffusion equations as the governing equations in the chamber model. We assumed the liquid diffusivity was close to pure water (~10−9 m2/s). To reduce the computation complexity, we decoupled the simulation into two stages: (1) liquid flows induced by the membrane movements and (2) transient concentration profiles under the flow. We adopted only the stabilized velocity profile because the flow stabilization took <2 s (a few mixing cycles), which was negligible comparing to the entire mixing duration (≥30 s). The velocity profile at time t should be equal to the extracted stabilized velocity profile at t s with the relation t = 0.3 N + t s , where N is the number of previous mixing cycles.
We further compared the results between the simulations and the experiments. To facilitate the comparison, we converted the 3D simulation results into spatial 2D profiles, by averaging concentrations along the vertical (z) direction at different ‘top-view’ positions (i.e., x and y). We quantified the simplified concentration profiles using standard deviations over the chamber top views, SD(t), at different mixing time t. We normalized SD(t) as SD′(t) = [SD(t) − SD(t → ∞)]/[SD(0) − SD(t → ∞)], such that SD′ equals ‘1’ at the beginning and drops toward ‘0’ throughout mixing (Voth et al. 2003). We defined the ‘mixing time’ as the moment where SD′ drops to ‘0.1’ to characterize the mixing performance.
Additionally, we conducted experiments to characterize SD′ and the mixing time under different values of the membrane switching time. Results show reasonable agreements between simulations and experiments, for mixing time in the cases without mixing (simulation: 157 s; experiment: 179 s. Fig. 3a), model I (simulation: 87 s; experiment: 67 s. Fig. S3) and model II (simulation: 30 s; experiment: 43 s. Fig. 3b). Due to the better mixing performance, we adopted the mixer model II as our model device for the following analyses. As shown in Fig. 3c, the optimal switching time for shortest mixing time of soluble substances was ~100 ms for model II.
3.5 Mixing of particles
Experiments were also conducted for the mixing of microparticles driven by the microfluidic mixer for different pumping switching times. The selected fluorescent beads (Cat# FP-0545-2, carboxyl fluorescent light yellow particles, Spherotech. Inc., Lake Forest, IL) had a comparable size (diameter: ~500 nm) with some bacteria (e.g, Escherichia coli and S. mutans); therefore, these results of the fluorescent beads should be, to certain extent, representative to mixing of the bacteria.
On the other hand, computation of the particle tracing was implemented to investigate the mixing for sub-microscale particles (e.g., bacteria and fluorescent micro-beads). We applied the previously computed transient velocity profiles and performed the particle simulation for one mixing cycle (duration: 300 ms) in COMSOL. In this simulation, a particle was released from each node of the mesh model of the chamber. The final positions (time: 300 ms) of all particles were then exported and stored. Owing to the cyclic membrane movements during the mixing operation, the chamber geometry should restore at the end of every mixing cycle. Therefore, we may neglect the membrane dynamics within the cycle by considering only the particle dislocations for every time step of the mixing duration (300 ms). For the particle tracking over a longer period, we have developed a customized MATLAB script to compute the particle trajectories with a time step of 300 ms. The particle dislocations inside each grid volume were approximated by the trilinear interpolation scheme using the nodal values of the enclosing grid. Both the simulation and experimental results (Fig. S4) revealed that two chamber regions were ineffective for the particle mixing.
To solve the mentioned limited mixing problem, we extended the mixing operation by implementing a mixing mode with repeatedly changing grouping of the six membranes as shown in Fig. 4b. Three stages of the membrane grouping were switched every 12 s, in order to generate circulation flows to enhance mixing in different directions. The simulation and experimental results (Fig. 4c) have demonstrated that the extended operations can induce an even distribution of the mixed particle profile.
Analogous to the analyses for the soluble substances, we performed repeated experiments (at least 20 trials for each switching time interval) to further investigate the transient SD′ for different switching time (Fig. 4d) and the mixing time as a function of the switching time (Fig. 4e). Results show that the particle mixing should operate with a switching time interval ~50 ms for the more efficient mixing. Indeed, the extended operation can be used also for the soluble substances to ensure a thorough mixing.
3.6 Experimental verification with cell culture
Cell culture experiments were further conducted to demonstrate the improvements of cell growth consistency and viability in the long term. The dental bacterium S. mutans was selected here. In particular, the S. mutans species is well known of its key role in the early development of human dental biofilms (Loesche 1986). The attachment of S. mutans on the tooth/pellicle substrates in human oral cavity forms a firm foundation for the following cell–cell adhesion between S. mutans and other bacteria (e.g., Fusobacterium nucleatum) involved in the later development of mature dental biofilm. Therefore, the culture of an evenly distributed layer of S. mutans can be further applied to the quantitative study of early dental biofilm formation and the corresponding mechanisms.
In our experiments, S. mutans cells were injected into the micro-mixer chamber following the procedures described in “Methods” and subsequently maintained under the microenvironment under the ambient gas condition, ~37 °C temperature and a regular injection of fresh culture media. Considering that the overall salivary flow rate in human oral cavity (volume: ~100 ml) is ~0.22 ml/min (Watanabe and Dawes 1990), we have set the average medium flow rate in the micro-chamber (capacity: ~62.8 pl) at ~0.363 pl/min for a reasonable liquid velocity. Furthermore, the flow rate of the medium injection operation was set as ~30 pl/min for 30 s once every 30 min, corresponding to average flow rate 0.5 pL/min and a maximum shear over the chamber base of ~0.02 Pa. On the other hand, we applied the device configuration ‘model II’ for the more effective mixing (switching time: 100 ms; driving pressure: 137.9 kPa) and the mixing operation was applied for 3 min after every medium injection.
An experiment was performed to show that S. mutans would grow faster near the chamber inlet for the medium injection without the peristaltic mixing operation (Fig. 5a). We analyzed the phase-contrast microscopic images of the cell distribution in the micro-chamber using an image processing scheme as reported in our previous works (Cui et al. 2014; Lam et al. 2009). Briefly, the background pattern of chamber region was estimated by the convolution of the image with a Gaussian mask (size: 20 pixel × 20 pixel, radius: 10 pixel) as a low-pass image filter. After subtracting the image intensities by the filtered background, the bacteria region would then be shown as darker patterns. Thresholding was then performed identify bacterial cells (black) as shown in Fig. 5a (right).
Furthermore, culture experiments of S. mutans were also applied with the medium injection and mixing operations. As an example, time-lapsed images of the S. mutans population (at 24, 32, 40 and 48 h) show the cells grew with an even distribution, induced by the mixing operation of freshly injected media (Fig. 5b). This result indicates that the peristaltic mixing scheme is effective to generate a more consistent microenvironment for even cell growth and to avoid cell clogging at chamber inlet.
4 Conclusion
In this study, a microfluidic mixer operating inside an enclosed micro-chamber was reported on the device design, function optimization and cell culture application. This mixer consisted of six microscale PDMS membranes, which were paired up as two peristaltic flow drivers. Circulation flows were induced based on the membrane movements (driven by external compressed-air pressures applied in a peristaltic manner) over the chamber boundaries. We have characterized multiple device configurations and demonstrated that the membrane parameters r M = 360 µm, L M = 230 µm and W M = 220 µm and H M = 20 µm can induce an effective mixing. Further, we introduced an extended mixing operation by changing grouping of the membranes for driving the circulation flows in different directions repeatedly. Quantitative analyses on the transient concentration profiles of color dyes and fluorescent beads have shown that the mixing can be optimized with the switching time of 100 and 50 ms for soluble substances and microparticles, respectively. As the comparable diameters between the fluorescent beads and the bacteria, the optimized operations for beads were further applied for the culture of bacteria with the equivalent diameters. We applied the mixing scheme to the culture of S. mutans to achieve a more evenly distributed bacteria clusters during their growth up to 2 days. The improvement on the growth consistency of S. mutans in the micro-chamber is obvious, comparing to the control case (no mixing) that the cell growth accumulated near the chamber inlet. In essence, this work has demonstrated the effectiveness and applicability of the microfluidic peristaltic mixing scheme based on moving boundaries of the chambers, driven by six deformable micro-membranes. As the soft lithography-based fabrication enables the direct integration of the multiple micro-mixers in microfluidics, this work can induce improvements and extended applications of microfluidics such as biomedical diagnosis and high-throughput bio-sample processing.
References
Abbas Y, Miwa J, Zengerle R, von Stetten F (2013) Active continuous-flow micromixer using an external braille pin actuator array. Micromachines 4:80–89
Materne E-M et al (2013) Dynamic culture of human liver equivalents inside a micro-bioreactor for long-term substance testing. In: BMC Proceedings of BioMed Central Ltd. Suppl 6:P72
Ansari MA, Kim K-Y, Anwar K, Kim SM (2010) A novel passive micromixer based on unbalanced splits and collisions of fluid streams. J Micromech Microeng 20:055007
Armani D, Liu C, Aluru N (1999) Re-configurable fluid circuits by PDMS elastomer micromachining. In: Twelfth IEEE International Conference on Micro Electro Mechanical Systems, 1999. MEMS’99, Ieee, pp 222–227
Byus CV, Pieper SE, Adey WR (1987) The effects of low-energy 60-Hz environmental electromagnetic fields upon the growth-related enzyme ornithine decarboxylase. Carcinogenesis 8:1385–1389
Capretto L, Cheng W, Hill M, Zhang X (2011) Micromixing within microfluidic devices. Microfluidics. Springer, New York, pp 27–68
Cui X, Yip HM, Zhu Q, Yang C, Lam RHW (2014) Microfluidic long-term differential oxygenation for bacterial growth characteristics analyses. RSC Adv 4:16662–16673. doi:10.1039/C4RA01577K
Friend J, Yeo LY (2011) Microscale acoustofluidics: microfluidics driven via acoustics and ultrasonics. Rev Mod Phys 83:647
Fu L-M, Ju W-J, Tsai C-H, Hou H-H, Yang R-J, Wang Y-N (2013) Chaotic vortex micromixer utilizing gas pressure driving force. Chem Eng J 214:1–7
Fu L-M, Fang W-C, Hou H-H, Wang Y-N, Hong T-F (2014) Rapid vortex microfluidic mixer utilizing double-heart chamber. Chem Eng J 249:246–251
Hsiung S-K, Lee C-H, Lin J-L, Lee G-B (2007) Active micro-mixers utilizing moving wall structures activated pneumatically by buried side chambers. J Micromech Microeng 17:129
Hyon Y, Marcos, Powers TR, Stocker R, Fu HC (2012) The wiggling trajectories of bacteria. J Fluid Mech 705:58–76
Jang L-S, Chao S-H, Holl MR, Meldrum DR (2007) Resonant mode-hopping micromixing. Sens Actuators A Phys 138:179–186
Kim S-J, Wang F, Burns MA, Kurabayashi K (2009) Temperature-programmed natural convection for micromixing and biochemical reaction in a single microfluidic chamber. Anal Chem 81:4510–4516
Koh JBY, Marcos (2015) The study of spermatozoa and sorting in relation to human reproduction. Microfluid Nanofluid pp. 1–20
Lam RHW, Kim M-C, Thorsen T (2009) Culturing aerobic and anaerobic bacteria and mammalian cells with a microfluidic differential oxygenator. Anal Chem 81:5918–5924. doi:10.1021/ac9006864
Lange H, Taillandier P, Riba JP (2001) Effect of high shear stress on microbial viability. J Chem Technol Biotechnol 76:501–505
Lee C-Y, Chang C-L, Wang Y-N, Fu L-M (2011a) Microfluidic mixing: a review. Int J Mol Sci 12:3263–3287
Lee KS, Boccazzi P, Sinskey AJ, Ram RJ (2011b) Microfluidic chemostat and turbidostat with flow rate, oxygen, and temperature control for dynamic continuous culture. Lab Chip 11:1730–1739
Legazpi L, Laca A, Díaz M (2009) Kinetic analysis of hybridoma cells viability under mechanical shear stress with and without serum protection. Bioprocess Biosyst Eng 32:717–722
Lin C-H, Tsai C-H, Pan C-W, Fu L-M (2007) Rapid circular microfluidic mixer utilizing unbalanced driving force. Biomed Microdevices 9:43–50
Lin Y-H, Wang C-C, Lei KF (2014) Bubble-driven mixer integrated with a microfluidic bead-based ELISA for rapid bladder cancer biomarker detection. Biomed Microdevices 16:199–207
Liu RH, Yang J, Pindera MZ, Athavale M, Grodzinski P (2002) Bubble-induced acoustic micromixing. Lab Chip 2:151–157
Loesche WJ (1986) Role of Streptococcus mutans in human dental decay. Microbiol Rev 50:353–380
Mansur EA, Ye M, Wang Y, Dai Y (2008) A state-of-the-art review of mixing in microfluidic mixers. Chin J Chem Eng 16:503–516
Marcos, Stocker R (2006) Microorganisms in vortices: a microfluidic setup. Limnol Oceanogr Methods 4:392–398
Marcos, Fu HC, Powers TR, Stocker R (2009) Separation of microscale chiral objects by shear flow. Phys Rev Lett 102:158103
Marcos, Seymour JR, Luhar M, Durham WM, Mitchell JG, Macke A, Stocker R (2011) Microbial alignment in flow changes ocean light climate. Proc Natl Acad Sci 108:3860–3864
Marcos, Fu HC, Powers TR, Stocker R (2012) Bacterial rheotaxis. Proc Natl Acad Sci 109:4780–4785
Marcos, Tran NP, Saini AR, Ong KCH, Chia WJ (2014) Analysis of a swimming sperm in a shear flow. Microfluid Nanofluid 17:809–819
Nguyen N-T, Wu Z (2005) Micromixers—a review. J Micromech Microeng 15:R1
Park C, Wereley ST (2013) Rapid generation and manipulation of microfluidic vortex flows induced by AC electrokinetics with optical illumination. Lab Chip 13:1289–1294
Peng Z-C, Hesketh P, Mao W, Alexeev A, Lam W (2011) A microfluidic mixer based on parallel, high-speed circular motion of individual microbeads in a rotating magnetic field. In: Solid-State Sensors, Actuators and Microsystems Conference (TRANSDUCERS), 16th International, 2011. IEEE, pp 1292–1295
Pitt WG, Ross SA (2003) Ultrasound increases the rate of bacterial cell growth. Biotechnol Prog 19:1038–1044
Sayar E, Farouk B (2011) Acoustically generated flows in microchannel flexural plate wave sensors: effects of compressibility. Sens Actuators A Phys 171:317–323
Sayar E, Farouk B (2012) Multifield analysis of a piezoelectric valveless micropump: effects of actuation frequency and electric potential. Smart Mater Struct 21:075002
Sayar E, Farouk B (2015) Bulk acoustic wave piezoelectric micropumps with stationary flow rectifiers: a three-dimensional structural/fluid dynamic investigation. Microfluid Nanofluid pp. 1–13
Scragg A, Allan E, Leckie F (1988) Effect of shear on the viability of plant cell suspensions. Enzyme Microb Technol 10:361–367
Senturia SD (2001) Microsystem design, vol 3. Kluwer Academic Publishers, Boston
Shiragami N, Unno H (1994) Effect of shear stress on activity of cellular enzyme in animal cell. Bioprocess Eng 10:43–45
Timoshenko S, Woinowsky-Krieger S, Woinowsky-Krieger S (1959) Theory of plates and shells, vol 2. McGraw-hill, New York
Tofteberg T, Skolimowski M, Andreassen E, Geschke O (2010) A novel passive micromixer: lamination in a planar channel system. Microfluid Nanofluid 8:209–215
Unger MA, Chou H-P, Thorsen T, Scherer A, Quake SR (2000) Monolithic microfabricated valves and pumps by multilayer soft lithography. Science 288:113–116
Vianale G, Reale M, Amerio P, Stefanachi M, Di Luzio S, Muraro R (2008) Extremely low frequency electromagnetic field enhances human keratinocyte cell growth and decreases proinflammatory chemokine production. Br J Dermatol 158:1189–1196
Voth GA, Saint T, Dobler G, Gollub JP (2003) Mixing rates and symmetry breaking in two-dimensional chaotic flow. Phys Fluids (1994-present) 15:2560–2566
Watanabe S, Dawes C (1990) Salivary flow rates and salivary film thickness in five-year-old children. J Dent Res 69:1150–1153
West J et al (2002) Application of magnetohydrodynamic actuation to continuous flow chemistry. Lab Chip 2:224–230
Wu M-H, Huang S-B, Cui Z, Cui Z, Lee G-B (2008) Development of perfusion-based micro 3-D cell culture platform and its application for high throughput drug testing. Sens Actuators B Chem 129:231–240
Xia H, Seah Y, Liu Y, Wang W, Toh AG, Wang Z (2014) Anti-solvent precipitation of solid lipid nanoparticles using a microfluidic oscillator mixer. Microfluid Nanofluid pp. 1–8
Yan D, Yang C, Miao J, Lam Y, Huang X (2009) Enhancement of electrokinetically driven microfluidic T-mixer using frequency modulated electric field and channel geometry effects. Electrophoresis 30:3144–3152
Yeo LY, Friend JR (2009) Ultrafast microfluidics using surface acoustic waves. Biomicrofluidics 3:012002
Zhu X, Kim ES (1998) Microfluidic motion generation with acoustic waves. Sens Actuators A Phys 66:355–360
Acknowledgments
We sincerely thank for the supports during the early development of this research from Department of Mechanical Engineering in Massachusetts Institute of Technology. We appreciate the invaluable advices from Dr. T. Thorsen. We thank for the financial supports from City University of Hong Kong (Seed Grant; project #7003019), Croucher Foundation Scholarship and General Research Grant of Hong Kong Research Grant Council (project# RGC115813).
Author information
Authors and Affiliations
Corresponding author
Additional information
Chengpeng Yang and Dinglong Hu have contributed equally.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Yang, C., Hu, D., Sun, B. et al. Mixing in an enclosed microfluidic chamber through moving boundary motions. Microfluid Nanofluid 19, 711–720 (2015). https://doi.org/10.1007/s10404-015-1596-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10404-015-1596-y