Abstract
Magnetic actuation principles using superparamagnetic beads suspended in a fluid are studied in this paper. An experimental setup containing a sub-microliter fluid volume surrounded by four miniaturized electromagnets was designed and fabricated. On the basis of optical velocity measurements, the induced behavior of single beads and ordered chains was analyzed and compared to a theoretical model. This research can be used to develop new techniques for accelerated transportation in lab-on-a-chip bio-assays.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Miniaturization and integration are key aspects of future biomedical diagnostics technologies and are driven by the need for speed, reliability, sensitivity, reducing costs and to minimize human intervention (Tüdős et al. 2001). One of the challenges in miniaturized systems is to generate effective mixing and transportation, because molecular diffusion alone leads to unacceptably long reaction times. New miniaturized sensing technologies based on the usage of magnetic beads are currently under investigation in many research groups (Megens and Prins 2005; Neuberger et al. 2005). Because biological matter is barely magnetic, these techniques can be used in complex biological samples without interference. The combination of magnetic detection with magnetic bead transport seems very promising for sensitive, rapid and miniaturized bio-assays (Ottino 1990; Luxton et al. 2004; Wild 2001; Nguyen and Wu 2005; Gijs 2004).
Magnetic beads can range from tens of nanometers up to several micrometers in diameter (Fig. 1a). Generally, beads consist of many (∼105) separated small (8–15 nm) iron oxide grains that are embedded in a polymer matrix, for example polystyrene (Fig. 1b). This composition leads to a superparamagnetic behavior (Fig. 1c), giving a strong magnetizability in combination with a disappearance of magnetic moment after switching off the external magnetic fields (Gijs 2004; Dynal Biotech, http://www.dynalbiotech.com). Another important aspect is observed when the magnetic field is switched on: the beads may form chains aligned with the field lines.
In this paper we study the motion of individual beads and chains of beads in a sub-microliter magnetic bead manipulation device. We derive the equations of motion for single beads and chains, we describe the design and fabrication of the bead manipulator and we discuss the experimental results.
2 Equations of motion
The magnetic force \(\vec{F}_{\rm m} \) acting on a single bead, approximated by a point-like magnetic dipole with a magnetic moment \(\vec{m}_{\rm b}, \) is proportional to the gradient of the field \(\vec{B}\) and given by (Young and Freedman 1996; Pankhurst et al. 2003; Jackson 1998):
The magnetic moment of a superparamagnetic bead is induced by a field \(\vec{B} = \mu _{0} \cdot \vec{H}\) and can be expressed as \(\vec{m}_{\rm b} = V_{\rm b} \cdot \chi _{{\rm b,eff}} \cdot \vec{H}.\) Parameter V b is the volume of the bead and χb,eff the effective magnetic susceptibility of the bead, including demagnetization effects (the demagnetization factor for a sphere is 1/3). Because the magnetic susceptibility of the surrounding medium is assumed to be zero and there are no time varying fields or currents in the medium or bead, we can adapt Eq. 1 to the case of a superparamagnetic bead as (Pankhurst et al. 2003):
A vibrating sample magnetometer (VSM) measurement has been carried out for beads suspended in fluid (Fig. 1c) (Dynal Biotech). It is clear that at higher field intensities the magnetization does not increase linearly by slope χb,eff anymore. The magnetization reaches its saturation \(\vec{M}_{{\rm sat}}, \) where the magnetization is no longer dependent on the applied field. For this case, the magnetic force on a single bead is no longer given by Eq. 2 but can be expressed as:
The magnetic attraction forces given in Eqs. 2 and 3 generate movement of a bead in a fluid, which causes a hydrodynamic drag force. For a spherical single bead without surface roughness, the drag force can be calculated with Stokes law (Happel and Brenner 1973):
The dynamic viscosity of the surrounding fluid is given by ηf. The hydrodynamic radius of the bead is assumed to be equal to the material radius r b of the bead. The speed of the bead relative to the surrounding fluid (that is assumed to be static) is given by \(\vec{v}_{\rm b}.\) Equations 2 and 4 can now be combined using the equation of motion \(m \cdot \ddot{x} = F_{\rm m} + F_{\rm d} .\) The beads are accelerated in a small negligible time (∼10− 6 s) to their maximum velocity. Therefore \(\vec{v}_{\rm b} \) can be expressed as (Gijs 2004; Pankhurst et al. 2003):
In case of a bead in magnetic saturation, Eq. 4 has to be combined with Eq. 3, giving:
These equations are only valid if the induced motion is perpendicular to gravitational forces and if Brownian motion does not play a significant role. For beads with a diameter of approximately 1 μm, Brownian motion lead to superimposed velocities in random directions below 100 nm/s and can therefore be neglected.
Magnetic beads that aggregate in chains require a different analysis due to the large transformation in shape. In a first approximation, one could assume that the total magnetic moment of a chain is given by the summation of all magnetic moments of the beads in a chain. However, dipole interactions between neighboring beads induce a change in the demagnetization fields, represented by the demagnetization factor N d that is linked to the magnetic susceptibility by (Jackson 1998):
The intrinsic susceptibility χint is a material parameter and remains constant. The effective susceptibility χeff will change because of the change in demagnetization.
We performed numerical simulations in Comsol Multiphysics (http://www.comsol.com) to study the influence of the chain shape on the demagnetization factor and the resulting magnetization (Fig. 2a). The chain is modeled as a row of spheres each of 1 μm in diameter and with the same intrinsic susceptibility as the single beads. The spacing of 50 nm represents twice the non-magnetic shell of the beads (Fig. 1b). The simulations show that the demagnetization factor is reduced by increasing the number of beads in the chain. We defined a magnetization enhancement as αmag = χc,eff/χb,eff, where χc,eff is the effective susceptibility of a chain. This magnetization enhancement approaches a maximum already in the case of five beads in a chain (Fig. 2b). By defining the chain length as l c and using the already defined bead radius r b, the number of beads can be given by n = l c/(2 · r b). The magnetic force on a long chain can now be expressed as:
The hydrodynamic resistance on a chain also highly depends on its shape and orientation. The hydrodynamic resistance of elongated bodies, axially dragged in a viscous fluid can be approximated as follows (Happel and Brenner 1973):
In this approximation, C 1 and C 2 are dimensionless constants that describe the hydrodynamic shape of the body, which have to be determined experimentally. If we now combine Eqs. 8 and 9, the velocity of a chain relative to the medium can be derived:
The above equation shows that the chain velocity \(\vec{v}_{\rm c} \) is proportional to the velocity of a single bead \(\vec{v}_{\rm b}. \) The proportionality—or in this case enhancement—factor that is shown in brackets on the right in Eq. 10, depends on the aspect ratio l c/r b, the hydrodynamic constants C 1 and C 2 and the magnetic enhancement factor αmag, but not on the magnetic field.
3 The magnetic bead manipulator
To investigate the bead and chain behavior induced by external magnetic fields, an experimental setup was designed and fabricated (Fig. 3). This magnetic bead manipulator consists of a fluid container, surrounded by four miniaturized solenoids to generate a wide variety of magnetic field shapes.
Because future miniaturized biosensor applications will generally consume sub-microliter volumes, the magnetic bead manipulator has a cylindrical fluid volume with a diameter of 1 mm and a depth of 200 μm. The beads are observed with an optical microscope that is able to resolve beads with a diameter down to 300 nm. The working distance of the microscope limits the outer diameter of the solenoids. Soft iron cores are inserted in the solenoids and are coupled with a surrounding yoke to minimize air gaps and to increase the field intensity in the fluid volume. To suppress the influence of flaring field lines, the inner diameter of the solenoids is larger than the fluid volume diameter.
The four solenoids with an inner diameter of 1 mm, an outer diameter of 2.5 mm, a length of 5 mm and 294 windings were constructed using a Ø 100 μm bonding wire. The cores and surrounding yoke are machined out of soft-iron to minimize remanent magnetic fields. A casing to assemble all components in the exact position is machined out of aluminium with a very low magnetic permeability to minimize the influence on the magnetic fields.
The disposable fluid containers are fabricated out of polydimethylsiloxane (PDMS) (Fig. 3a). By photolithography and subsequent wet etching steps, an inversely shaped mold has been created on a silicon wafer that consists of an array of circular posts. A pre-mixture of a silicone elastomer (Sylgard 184 base) and a curing agent elastomer (Dow Corning) is poured over the whole wafer and placed in a sandwich curing plate equipped with three adjustable micro-meter screws to set the final height of the containers.
After curing, the PDMS slice is torn off the wafer and laser-cut into individual containers with outer dimensions of 2 mm × 2 mm. Bovine serum albumin (BSA) is applied on all walls of the fluid container to increase the hydrophilicity of the PDMS surface (originally highly hydrophobic). To close the containers, glass cover sheets are used that were sawed into the same lateral dimensions.
The complete magnetic bead manipulator is mounted on a PCB board to facilitate the handling and to create solid electrical connection points (Fig. 3b). The magnetic bead manipulator is placed under a microscope (objective M = 65, NA = 0.70, WD = 1.30 mm) equipped with a digital camera (resolution 720 × 576 pixels, frame rate 25 fps), connected to a DVD recorder. A four-channel home-built current source (I max = 250 mA over R = 3 Ω) is used to power the solenoids separately. With use of a D/A converter, the current source is controlled with scripts that are written in Matlab. A height gauge is mounted just below the microscope base-plate to control the position of the view plane in the fluid volume.
Numerical simulations were carried out in Comsol Multiphysics (http://www.comsol.com) in order to investigate a variety of magnetic fields induced by the electromagnets. In the first simulation, one solenoid is powered with a low current of 75 mA (Fig. 4a). The maximum magnetic field intensity is 15 mT at the border of the fluid volume, as illustrated by the color map (gray scale). The bead trajectories are calculated with use of Eq. 5 and 6 and illustrated by the arrows. In this model, a maximum velocity of 5 μm/s is calculated for beads with a diameter of 1 μm. In a second simulation, the left and top solenoids are both powered by a high current of 130 mA (Fig. 4b), generating 39.7 mT at maximum. Two neighboring solenoids can be used this way to interpolate the field characteristics between the directions of the solenoids, preserving a bead velocity of 20 μm/s. In the third simulation, every solenoid is powered by a high current of 130 mA (Fig. 4c), generating maximum magnetic field intensities of 48.5 mT. The bead trajectories show a large variation, with maximum velocities of 40 μm/s. Many more (complex) field variations have been found by similar simulations, which demonstrate the high performance and flexibility of the magnetic bead manipulator.
4 Experiments
Experiments were carried out to verify the travel velocities of single beads and chains as expected by Eqs. 5 and 10, using the magnetic field prediction from the simulations (Fig. 4a). By applying the current of 75 mA through the left solenoid, single beads and chains are attracted towards the solenoid. Beads with a diameter of 1 μm were used, commercially available from Dynal (Dynal Biotech). Their magnetic saturation occurs at much higher field intensities than reached in these experiments and therefore the magnetic susceptibility (χb,eff = 1.52) can be assumed to be constant (Fig. 1c).
The stock solution of beads was diluted with water into two different concentrations. A low concentration of 200 μg/ml (≈2 · 108 beads/ml) minimizes chain formation, suitable to study the travel of single beads. A higher concentration of 500 μg/ml (≈5 · 108 beads/ml) forces the beads to aggregate in chains under the application of a magnetic field, which is used in the study on chains. By adding the inorganic salt sodium polytungstate (SPT), the bead solution was tuned to have a matching density as the beads itself (ρ = 1.8 g/cm3). This avoids out-of-plane traveling of the beads, to enable a complete bead tracking in the whole imaging plane of the microscope. SPT does not affect the neutral properties of the fluid, but increases the viscosity η to a value of 2.1 · 10−3 kg/m s, lowering the expected experimental velocities (already taken into account in the simulations).
The path of every single bead or chain was tracked from a recorded movie, using home developed processing software that is written in Matlab (Fig. 5). Every movie frame is averaged with the 25 surrounding frames to suppress background noise and subsequently all intensity peaks in the frame are identified. Neighboring intensity peaks are joined to one area and divided by the actual area of a single bead, giving the number of beads in the chain or identifying a single bead. The bead or chain position is determined with sub-pixel resolution by an estimation of the center of mass.
To exclude chains that do not move in an axial direction, the relative alignment with the direction of movement is determined, which was not allowed to exceed a difference of 5 degrees. The obtained tracks are visualized in the movie as trajectory lines.
In a first experiment, the travel of single beads has been studied. For each single bead position located with the movie processing software, the actual velocity is calculated. We have plotted the experimental velocities as function of the calculated values of \(\vec{\nabla }\vec{B}^{2} \) (Fig. 6). The measured velocities are on average noticeable lower than predicted by Eq. 5, quantified by an average relative error (RE) of 0.33. Furthermore, the measurements show a large spread that is expressed as a coefficient of variation (CV) of 0.22. This RE and CV can be caused by uncertainties in the following parameters (in brackets the form in Eq. 5 is given):
-
The radius and hydrodynamics of the bead including shape and surface roughness (r 2 b);
-
The viscosity of the surrounding fluid (1/ηf);
-
The magnetic susceptibility of the beads (χb,eff);
-
The magnetic field characteristics \((\vec{\nabla }\vec{B}^{2} );\)
-
(computational) errors in the velocity measurement itself \((\vec{v}_{\rm b} ).\)
Light scattering experiments show a small distribution in bead diameter (Dynal Biotech), affecting the experimental velocities with 20% at maximum that could contribute to the CV. The viscosity of the surrounding fluid has been increased by the density matching process. Measurements determined the fluid viscosity with an uncertainty of 5%. However, a small increase of the fluid temperature by heat dissipation of the solenoids and the illumination could lower the viscosity existential. The viscosity is therefore estimated to affect the resulting velocities at most with 20%, which could contribute to the RE. The average magnetic susceptibility per bead has been estimated by a VSM measurement in a concentrated volume of magnetic beads (Dynal Biotech). Magnetic interactions between beads in the suspension may occur during this measurement. The obtained value may therefore not reflect the single bead magnetic susceptibility (Häfeli et al. 2005; van Ommering et al. 2006), which could contribute to the RE. The measurement also conceals a contribution to the CV that could exist because of a possible but unknown variation in the iron oxide content (Häfeli et al. 2005).
The estimation of the magnetic field characteristics is based on simulations. Deviation could occur by magnetic saturation of the cores and surrounding yoke. Simulations predicted maximum field intensities of approximately 100 mT in the flux guides (applying the same parameter values as in the experiment), which is far below the saturation level of the core and yoke material. In this point in time, we approximate the deviation in the experimental velocities because of the field characteristics to be 15% (contributing to the RE and CV), but this assumption should be validated with a magnetic field calibration. The velocity measurement and numerical calculations could also be subject of experimental error. A deviation of 10% on the experimental velocities is estimated, contributing to both the RE and CV. Considering the many potential sources of deviation, the single bead experiments were in good agreement with the established equations of motion.
The following experiment has investigated the utility of chain formation for the acceleration of transport. Because chains will align with the field lines, an axial movement of the chains can be expected, according to the simulations (Fig. 4a). For all obtained trajectory lines, the resulting velocities as a function of \(\vec{\nabla }\vec{B}^{2} \) are calculated (Fig. 7a). The shape of these curves is comparable with the curve of the single bead experiments (roughly linear). Quantitatively, we observe for the velocity of each chain length an enhancement factor normalized to the velocity of a single bead: \({\rm VEF} = {\vec{v}_{\rm c} }/{\vec{v}_{\rm b} }.\) This velocity enhancement factor depends in a logarithmic way only on the number of beads in the chain (n), approximated by (Fig. 7b):
The obtained function is in agreement with the velocity enhancement factor we found in Eq. 10. Using αmag = 1.2, we find a corresponding fit with C 1 ≈ 9 and C 2 ≈ 0.56. A comparison with values for an ordinary cylinder dragged in an axial direction through the fluid (C 1 = 4 and C 2 = −0.72) (Happel and Brenner 1973) indicates that the hydrodynamic drag force of a chain of beads is evidently higher. The undefined surface roughness and the undulating shape of the chain of beads is most likely the main reason for the higher hydrodynamic drag force.
5 Conclusions and outlook
We performed studies on the dynamics of single magnetic beads and chains of beads inside a sub-microliter device with four surrounding miniaturized electromagnets. Experiments on single beads showed velocities that can differ with a coefficient of variation (CV) of 0.22, most probably by bead-to-bead variations. The average measured velocity deviates from theoretical predictions with an average relative error (RE) of 0.33. One hypothesis is that this deviation is partially caused by the use of a susceptibility value from bulk VSM measurements, which may not reflect the actual average susceptibility of individual beads. Further experiments should be carried out to get a good overall understanding of single bead properties. Furthermore, the magnetic field in the fluid volume generated by the bead manipulator should be measured as function of its position and compared to the simulations to confirm the accuracy of this parameter.
Equations for the influence of chain formation on the magnetization and the hydrodynamic drag force have been established and compared to experimental data. We observe a logarithmic dependence of velocity on the chain length. For a given magnetic field, the velocity of chains is proportional to the velocity of single beads with an enhancement factor that depends on the number of beads in the chain and the hydrodynamic properties of the chain.
Our experiments show that chain formation can in principle be used for accelerated transport in fluids. We envisage that the controlled rapid manipulation of chains can be beneficial for bio-assays, for example when target molecules are to be bound to magnetic particles or when enhanced mixing or transport is required. Next steps will involve the study of chain manipulation in dynamic fields, for example to generate a rotational motion of the chains. The application of bead manipulation should in further research be confirmed in biological samples.
References
Comsol Multiphysics, Comsol 3.2.0.222, Comsol AB. http://www.comsol.com
Dynal Biotech ASA, Norway. http://www.dynalbiotech.com
Gijs M (2004) Magnetic bead handling on-chip: new opportunities for analytical applications. Microfluid Nanofluid J 1:22–40
Häfeli U, Lobedann M, Steingroewer J, Moore L, Riffle J (2005) Optical method for measurement of magnetophoretic mobility of individual magnetic microspheres in defined magnetic field. J Magn Magn Mater 293:224–239
Happel J, Brenner H (1973) Low Reynolds number hydrodynamics, with special applications to particulate media. Noordhoff International Publishing, Gronigen. ISBN 9-001-37115-9
Jackson J (1998) Classical electrodynamics, 3rd edn. Wiley, New York. ISBN 0-471-30932
Luxton R, Badesha J, Kiely J, Hawkins P (2004) Use of external magnetic fields to reduce reaction times in an immunoassay using micrometer-sized paramagnetic particles as labels (magnetoimmunoassay). Anal Chem 76:1715–1719
Megens M, Prins M (2005) Magnetic biochips: a new option for sensitive diagnostics. J Magn Magn Mater 293:702–708
Neuberger T, Schöpf B, Hofmann H, Hofmann M, von Rechenberg B (2005) Superparamagnetic nanoparticles for biomedical applications: possibilities and limitations of a new drug delivery system. J Magn Magn Mater 293:483–496
Nguyen N, Wu Z (2005) Wu micromixers—a review. J Micromech Microeng 15:1–16
van Ommering K, Nieuwenhuis J, van IJzendoorn L, Koopmans B, Prins M (2006) Confined Brownian motion of individual magnetic nanoparticles on a chip—characterization of magnetic susceptibility. Appl Phys Lett (in press)
Ottino J (1990) The kinematics of mixing: stretching, chaos and transport. Cambridge University Press, Cambridge. ISBN 0-521-36335-7
Pankhurst Q, Connolly J, Jones S, Dobson J (2003) Applications of magnetic nanoparticles in biomedicine. J Phys D Appl Phys 36:167–181
Tüdős A, Besselink G, Schasfoort R (2001) Trends in miniaturized total analysis systems for point-of-care testing in clinical chemistry. Lab Chip 1:83–95
Wild D (2001) The immunoassay handbook, 2nd edn. Nature Publishing Group, London. ISBN 0-333-72306-6
Young H, Freedman R (1996) University physics, 9th edn. Addison-Wesley Publishing Company, Reading. ISBN 0-201-31132-1
Acknowledgments
The authors would like to thank Erik Homburg for his assistance on the simulations and electronics, Kim van Ommering for her great input in developing the movie processing software, and Frans van Gaal for the realization of the experimental setup.