Abstract
A valve-switching ion chromatography method coupled with conductivity and pulsed amperometric detection has been set up for the simultaneous determination of organic acids, inorganic anions and alditols in wine. The system, based on valve-switching technique, consisted of two pumps, two analytical columns and a trap column connected in series via one six-port valve. In addition, the optimal valve-switching time and chromatographic conditions were explored to obtain a good linearity, precision, accuracy, and sensitivity with a mean correlation coefficient of > 0.99, repeatability of 0.62–6.18% for eight replicates, and the average spiked recovery of 96.66%. The new method was successfully applied in the analysis of six kinds of wine from different regions.
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Organic acids, inorganic anions, and alditols play a significant role in wine for its properties (e.g., taste, aroma, and colour), nutrition, and biological stability. Therefore, they are regarded as chemical markers of ripeness, bacterial activity, or storage conditions. The content and kinds of organic acids are the decisive factors on acidity which are related to the region and climate during the growth and ripening of grapes [1, 2]. Similarly, alditols are seen as an emerging functional sweetener with less thermal, indigestibility, as well as low sweetness, which can make the wine more mellow and aromatic [3]. In addition, the main anions such as chloride, sulfate, and phosphate must be controlled within a certain extent, because they can influence the flavor of wine. As a result, the accurate determination of organic acids, inorganic anions, and alditols is highly valuable to optimize the production process and control product quality.
Nowadays, chromatographic analysis [4,5,6,7,8,9,10,11,12,13,14] or tandem mass spectrometry [15,16,17,18,19] is the most used to identify and quantify ions, organic acids, sugars, and alditols. Nevertheless, the analysis of some organic acids and alditols needs to employ complicated derivation by high-performance liquid chromatography (HPLC) [7,8,9,10], or gas chromatography–mass spectrometry [5]. Ion chromatography as a sensitive and selective method without derivatisation was widely employed to determine these polar compounds [11,12,13,14]. However, simultaneous determination of organic acids, inorganic anions, and alditols has not been reported. One same sample needs to be tested many times, respectively, so that it will cost more time to obtain all information of target substances. In the present study, valve-switching technique has offered a good pathway for the elimination of matrices interference and enrichment of the target substance in the chromatographic system. It not only overcomes the limitation in resolution of one-dimensional ion chromatography, but also simplifies tedious operation and decreases a great deal of testing time [20,21,22].
Therefore, we developed and validated an efficient ion chromatography method by valve-switching technique for the simultaneous determination of organic acids, inorganic anions, and alditols in this paper. The system consisted of two pumps, a six-port valve, two ion-exchange columns, conductivity detector, and pulsed amperometric detector and automatically switched by only one six-port valve. The new method was successfully applied in the analysis of six kinds of wine from different regions. Furthermore, it saved much working time, so that greatly improved the efficiency during wine-making process monitor and quality control.
Experimental
Apparatus and Chromatography Conditions
Analyses were conducted using an ICS-5000 system (Thermo Fisher Scientific, Sunnyvale, CA, USA) equipped with two gradient pump, an eluent generator (EG40) with potassium hydroxide (KOH) cartridge and CR-TC, a pulsed amperometric detector (PAD), a conductivity detector (CD), a self-regenerating suppressor (AERS 500 4 mm), an AS40 auto-sampler with a 25 μL sample loop, and a six-port valve. System control and data acquisition were performed by Chromeleon 6.8 software.
CarboPac MA1 Guard column (4 × 50 mm, Thermo Fisher Scientific) was used as trap column. An IonPac AS11-HC Analytical column (4 × 250 mm, Thermo Fisher Scientific) with an IonPac AG11-HC Guard column (4 × 50 mm, Thermo Fisher Scientific) was employed for the separation of organic acids and inorganic anions. A CarboPac MA1 (4 × 250 mm, Thermo Fisher Scientific) analytical column was used to separate alditols.
The gradient elution program for organic acids and inorganic anions was: 1 mmol L−1 (1–15 min); 5 mmol L−1 (15–20 min); 20 mmol L−1 (20–31 min); 22.4 mmol L−1 (31–45 min); 40 mmol L−1 (45–55 min); 45 mmol L−1 (55–60 min); and 1 mmol L−1 (60–65 min). The isocratic 500 mmol L−1 NaOH was utilized for alditols at a flow rate of 0.4 mL min−1. The pulsed amperometric detector was equipped with a triple-electrode system that included a gold working electrode, an Ag/AgCl reference electrode, and a titanium cell body as the counter electrode. The waveforms are listed in Sect. 2.1 of Ref. [20].
Reagents and Samples
Organic acid standards were obtained from Sigma-Aldrich (St. Louis, MO, USA). The anion standards (1000 mg L−1) were purchased from the National Institute of Metrology. The alditol standards were supplied by Aladdin (Shanghai, China). Sodium hydroxide (50%, w/w, certified grade) was offered by Acros Organics (New Jersey, USA). All the solutions were prepared in 18.2 MΩ cm water (Milli-Q system, Millipore, Mosheim, France). The working standard solutions were prepared by diluting Stock standard solutions (1000 mg L−1).
The wine samples of six brands were collected from different countries and regions including red wine from Italy Primitivo (R1), red wine from South Africa Remhoogte (R2), red wine from France La Tour Argent (R3), white wine from France Chateau Rievssec (W1), white wine from France Latour Laguens (W2), and white wine from Portugal AiQuan (W3). All samples were diluted 100 times and then filtered through 0.22 μm nylon membrane and were handled with Cleanert IC-RP column (Agela Technologies, Tianjin, China) for removing the macromolecular organic matters before analysis.
Valve-Switching Program
Four steps were operated to accomplish the valve-switching procedure (Fig. 1). First, the sample was loaded into the loop. Second, it was the extracting phase, in which alditols was eluted from the AS11-HC column and extracted on the trap column. Third, it was the analysis phase, where organic acids and inorganic anions were continually separated on the AS11-HC column and detected on conductivity detector, while alditols were eluted from the trap column, separated on MA-1 column, and analyzed by pulsed amperometric detector, respectively. Finally, it involves re-equilibration. Once the system reached re-equilibration, the cycle restarted from the beginning. The time and the status of valves are listed in Table 1.
Method Validation
The method was validated for linearity, repeatability, reproducibility, limits of detection (LODs), and quantification (LOQs). The linearity was established by the injection of the standard mixtures at concentrations varying from 0.02 to 50 mg L−1. The calibration curves were constructed by plotting peak areas versus different concentrations of standard mixtures. Besides, the repeatability was estimated using eight replicates of 2 mg L−1 standard mixtures within 1 day, whereas reproducibility was calculated by analyzing three replicates at three different concentrations levels (2, 5, and 10 mg L−1) on three consecutive days. Furthermore, the LODs and LOQs were calculated as the amounts for which the signal-to-noise ratio (S/N) was 3 and 10, respectively. Eventually, the recoveries represented the mean of three levels (1, 5, and 10 mg L−1) of spiked recoveries.
Results and Discussion
The Valve-Switching Time
At appropriate valve-switching time, alditols should be completely transferred to the trap column and the weak retain anion must not be eluted from AS11-HC column. Therefore, in this section, glycerol of 10 mg L−1 was selected as the standard to confirm the optimal valve-switching time as it was first eluted from the MA-1 column. Since glycerol did not appear before 3.2 min., the optimization of switching time started from 3.2 min and ended to 4.0 min (3.2, 3.4, 3.5, 3.6, 3.7, 3.8, and 4.0 min). The peak area of glycerol reached a maximum value at 3.5 min suggested that the optimal valve-switching time was 3.5 min.
Analytical Properties
Table 2 lists the linear equation, range of linearity, and correlation coefficient (r2) of calibration curves. Satisfyingly, the method showed good linearity, which the mean correlation coefficient was 0.996. As displayed in Fig. 2a, b, all components were successfully separated without obvious interference under the optimal conditions. The relative standard deviation (RSD) of repeatability and reproducibility ranged from 0.62 to 6.18 and from 0.34 to 3.48%, respectively. The LODs were 0.01–0.03 mg L−1 while LOQs 0.01–0.10 mg L−1. Table 3 summarizes the average recoveries of three levels with 83–113%.
Application to Wine Samples
The new method was applied to determine organic acids, inorganic anions, and alditols in wine samples of three kinds of red wines and three white wines from six regions. The chromatograms of six samples demonstrated a complete separation of organic acids, inorganic anions, and alditols (Fig. 2a, R1, R2, R3, W1, W2, and W3 and Fig. 2b, R1, R2, R3, W1, W2, and W3). Their contents are listed in Table 3.
First, 11 organic acids were found in these samples. The major acids were lactic acid, tartaric acid, succinic acid, citric acid, maleic acid, fumaric acid, and isocitric acid. Among them, lactic acid originates from the alcoholic fermentation and malolactic fermentation, in which the procedure gets the original wine more smooth. As the “wine diamond”, tartaric acid has considerably higher concentration than other acids. In addition, the red wine had a larger amount of lactic acid than the white, because the white wine is lack of malolactic fermentation process [8]. Second, the data revealed all samples presented low concentration of chloride and phosphate. However, the content of sulfate was obviously higher than the other anions due to the addition of sulfur dioxide or sulfite for killing the bacteria and ensure the integrity of wine flavor. Finally, the amount of glycerol of sample R1 was remarkably higher than others. In addition, three isomerides (sorbitol, mannitol, and dulcitol) are well separated and qualified. In brief, the content changes in organic acids, inorganic anions, and alditols of six samples demonstrated the diversities in process, quality, or geographical origin.
Conclusions
The work provided a reliable, sensitive, and effective method coupling with conductivity and pulsed amperometric detection for the simultaneous determination of organic acids, inorganic anions, and alditols in wine. It demonstrated that the method had a good linearity, precision, accuracy, and sensitivity with a mean correlation coefficient of > 0.99, repeatability of 0.62–6.18% for eight replicates, the average spiked recovery of 96.66%, and LODs of 0.01–0.03 mg L−1. Results of real samples are satisfactory and highly valuable for wine-making process monitor and quality control.
References
Lasik M (2013) The application of malolactic fermentation process to create good-quality grape wine produced in cool-climate countries: a review. Eur Food Res Technol 237:843–850
Zeravik J, Fohlerova Z, Milovanovic M, Kubesa O, Zeisbergerova M, Lacina K, Petrovic A, Glatz Z, Skladal P (2016) Various instrumental approaches for determination of organic acids in wines. Food Chem 194:432–440
Tang KT, Lin L, Liang LN, Cai YQ, Mou SF (2008) Simultaneous determination of sugars and sugar alcohols in sugar-free and low-sugar food with high performance anion-exchange using pulsed amperometric detection. Food Sci 29:327–331
Mato I, Suarez-Luque S, Huidobro JF (2005) A review of the analytical methods to determine organic acids in grape juices and wines. Food Res Int 38:1175–1188
Zhou HB, Xiong ZY, Yu Y, Wan R, Li P, Shen B (2013) Determination of alditols in wine by gas chromatography–mass spectrometry after acetate derivatization. Chin J Chromatogr 31:786–790
Jurado-Sanchez B, Ballesteros E, Gallego M (2011) Gas chromatographic determination of 29 organic acids in foodstuffs after continuous solid-phase extraction. Talanta 84:924–930
Kemmei T, Kodama S, Yamamoto A, Inoue Y, Hayakawa K (2015) Reversed phase liquid chromatographic determination of organic acids using on-line complexation with copper(II) ion. Anal Chim Acta 886:194–199
Zong YY, Lin J, Xu H, Jia ZH, Yang XP, Choi MF (2015) Optimization and validation of an HPLC-photodiode array detector method for determination of organic acids in vinegar. J AOAC Int 98:422–430
Gao NF, Zhang J, Han YS (2004) High performance liquid chromatography (HPLC) method for determination of organic acids in wine. Liquor Mak 30:67–69
Kelebek H, Selli S, Canbas A, Cabaroglu T (2009) HPLC determination of organic acids, sugars, phenolic compositions and antioxidant capacity of orange juice and orange wine made from a Turkish cv. Kozan. Microchem J 91:187–192
Pan BZ, Liu Q, Pang SQ, Zhang S, Li ZY, Li X, Chen WR (2013) Determination of organic acids and mineral anions in liquor by ion chromatography. Mod Food Sci Technol 29:876–881
Silveira ELC, de Caland LB, Tubino M (2014) Simultaneous quantitative analysis of the acetate, formate, chloride, phosphate and sulfate anions in biodiesel by ion chromatography. Fuel 124:97–101
Hanko VP, Rohrer JS (2000) Determination of carbohydrates, sugar alcohols, and glycols in cell cultures and fermentation broths using high-performance anion-exchange chromatography with pulsed amperometric detection. Anal Biochem 283:192–199
Wu FY, Jia ZS, Zhu Y (2006) Determination of organic acid and inorganic anion by suppressed electric conductance ion chromatography. J Zhejiang Univ 33:312–315
Regueiro J, Vallverdu-Queralt A, Simal-Gandara J, Estruch R, Lamuela-Raventos R (2013) Development of a LC–ESI–MS/MS approach for the rapid quantification of main wine organic acids in human urine. J Agric Food Chem 61:6763–6768
Cui Y, Li Q, Liu ZZ, Geng LL, Zhao X, Chen XH, Bi KS (2012) Simultaneous determination of 20 components in red wine by LC–MS: application to variations of red wine components in decanting. J Sep Sci 35:2884–2891
Ehling S, Cole S (2011) Analysis of organic acids in fruit juices by liquid chromatography–mass spectrometry: an enhanced tool for authenticity testing. J Agric Food Chem 59:2229–2234
Sandín-España P, Mateo-Miranda M, López-Goti C, De Cal A, Alonso-Prados JL (2016) Development of a rapid and direct method for the determination of organic acids in peach fruit using LC–ESI–MS. Food Chem 192:268–273
Xiong ZY, Dong Y, Zhou HB, Wang H, Zhao YX (2014) Simultaneous determination of 16 organic acids in food by online enrichment ion chromatography–mass spectrometry. Food Anal Methods 7:1908–1916
Vermeiren K (1085) Trace anion determination in concentrated hydrofluoric acid solutions by two-dimensional ion chromatography I. Matrix elimination by ion-exclusion chromatography. J Chromatogr A 2005:60–65
Fa Y, Yang HY, Ji CS, Cui H, Zhu XS, Du J, Gao J (2013) Simultaneous determination of amino acids and carbohydrates in culture media of Clostridium thermocellum by valve-switching ion chromatography. Anal Chim Acta 798:97–102
Johns C, Shellie RA, Pohl CA, Haddad PR (2009) Two-dimensional ion chromatography using tandem ion-exchange columns with gradient-pulse column switching. J Chromatogr A 1216:6931–6937
Acknowledgements
This research was supported by the National Natural Science Foundation of China (Nos. 21405166, U1507203) and the National Public Welfare Special Fund for “Food Chemical Risk Substances Screening and Warning Demonstration Research” (No. 2012424012).
Funding
This study was funded by the National Natural Science Foundation of China (Nos. 21405166, U1507203) and the National Public Welfare Special Fund for “Food Chemical Risk Substances Screening and Warning Demonstration Research” (No. 2012424012).
Author information
Authors and Affiliations
Corresponding authors
Ethics declarations
Conflict of interest
All authors declare that they have no conflict of interest.
Ethical approval
This article does not contain any studies with human participants performed by any of the authors.
Rights and permissions
About this article
Cite this article
Li, F., Bao, Y., Yu, Y. et al. Simultaneous Determination of Organic Acids, Inorganic Anions, and Alditols in Wine with Valve-Switching Ion Chromatography. Chromatographia 81, 1103–1108 (2018). https://doi.org/10.1007/s10337-018-3563-8
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10337-018-3563-8