Introduction

A major concern using cone-beam computed tomography (CBCT) in dentistry is the potential increase of cancer risk due to ionizing radiation stochastic effects. As the risk of cancer induction has no threshold, the radiation dose that patients are exposed to is crucial, as any dose can increase the probability of cancer development [1,2,3,4,5]. Patient radiation dose reduction by optimization is one of the tenets of the ALADAIP (As Low As Diagnostically Acceptable being Indication-oriented and Patient-specific) principle [6]. In CBCT, the appropriate selection of field of view (FOV) to the region of interest and diagnostic task can substantially reduce patient radiation dose.

A wide variety of CBCT devices are currently available in Dentistry [7]. Technological improvements in hardware and software (flat panel detectors, iterative reconstruction, X-ray beam collimation, optimized filtering, tube current modulation, and low dose protocols) allow for maintaining image quality while decreasing patient dose [8, 9]. The Veraviewepocs 3D R100 (R100) and Veraview X800 (X800) CBCT devices (J. Morita MFG. Corp. Kyoto, Japan) feature a 180° rotation convex triangular Reuleaux triangle) field of view (FOV) in addition to the typical cylindrical FOV. This Reuleaux triangle shape is observed in the axial reconstruction of the FOV. The Reuleaux triangle is a non-circular, equilateral shape with curves that maintain a constant width developed by Franz Reuleaux in the 19th century [10]. According to the manufacturer, the convex triangular FOV, which covers more volume than the 80 mm cylindrical FOV, may decrease patient dose by up to 15% while adapting to the parabolic shape of the dental arches.

Pauwels et al. (2012) estimated the effective doses numerous CBCT devices using different exposure protocols and projection geometries and reported a substantial range in doses between different CBCT devices [11]. Rottke et al. (2018) suggested that the effective dose of the convex triangular FOV of the R100 device was lower than the cylindrical FOV, but only assessed the 80 mm height convex triangular and cylindrical FOVs [9]. Furthermore, previous in vitro studies showed an influence of the image geometry formation of the convex triangular FOVs in the image quality [12,13,14]. A newer model, the X800, also provides a convex triangular FOV and, as of yet, no comparable dosimetry for either device or for cylindrical FOVs of similar height have been reported.

Understanding the dosimetry of the different CBCT devices allows consideration of different protocols to reduce patient dose via optimization based on scientific evidence. The present study is aimed at assessing the dosimetry of the convex triangular FOVs and comparing them with the cylindrical FOVs of similar dimensions in the R100 and X800 (J. Morita, Kyoto, Japan) CBCT devices.

Materials and methods

The head and neck portion of an anthropomorphic Rando phantom (Alderson Research Laboratories, Stanford, CT) representing an average adult male, embedded in synthetic rubber for tissue simulation and horizontally sliced, was used for the dosimetry scans (Fig. 1a). Optically stimulated luminescence dosimeters (OLSDs) (NanoDots, Landauer, Glenwood, IL) were calibrated using a calibration set provided by the manufacturer for the acquisition parameters used in the study. For each CBCT protocol, 24 OSLDs (10 × 10 × 2 mm) (Fig. 2) were placed in different anatomical locations according to previous studies (Table 1), and 3 OSLDs were placed outside the room to measure background radiation [15,16,17,18]. The phantom was previously scanned to determine the location and slice level of the anatomical structure that each OSLD represented. Vertical slots were created with a high-speed handpiece to hold the OSLDs in their determined positions (Fig. 1b).

Fig. 1
figure 1

Anthropomorphic phantom used in the different protocol scans. a Diagonal view of the phantom in the R100 device. b Numbered slots (arrows) at level 2 for holding the OSLDs within the phantom: 1. calvarium anterior; 2. calvarium left; 3. calvarium posterior; 4. midbrain)

Fig. 2
figure 2

Close-up view and relative size of the Optically Stimulated Luminescent Dosimeters (OSLDs)

Table 1 Anatomical locations and slice level of the OSLDs

Dosimetry scans

CBCT scans were performed using the R100 and X800 CBCT devices, both using continuous exposition radiation. Acquisition parameters are shown in Table 2. The FOV position for each scan protocol was determined by two-direction scout to simulate clinical conditions. Three exposures were performed for each scan protocol to ensure enough radiation reached the OSLDs. After each protocol, the OSLDs were read in a microSTARii reader (Landauer, Glenwood, IL), which had been previously calibrated using a set of 80-kVp reference dosimeters provided by the manufacturer (a beam energy calibration factor was used to correct variations in dosimeter sensitivity across a range of beam energies). The measured background radiation was subtracted from the measured radiation of the experimental dosimeters. Each reading was divided by three to determine the absorbed dose of each OSLD per exposure. Before re-irradiation of the OSLDs, they were placed on a 150W 15,000 Lumen LED light for 30 minutes. This process of optical bleaching clears the stored signal in the device allowing for new measurements to be made. The OSLDS were then inserted into the same corresponding slots in the phantom slices for all the subsequent scans to reduce angular dependence [19]. Twelve protocols among two CBCT devices were scanned for dose calculations (Table 3).

Table 2 Technical parameters according to CBCT device and FOV shape and size
Table 3 Scan protocols according to CBCT device and FOV shape, size, and region

Statistical analysis

Mean equivalent and effective doses and standard deviation values were calculated based on the OLSD absorbed dose readings and according to previous studies [16,17,18]. The equivalent doses for each of the following organs or tissues were calculated and multiplied by their estimated fraction irradiated in the exam: bone marrow (16.5%), esophagus (10%), thyroid (100%), bone surface (16.5%), brain (100%), salivary glands (100%), skin (5%), and remainder (oral mucosa, and extrathoracic tissue—100%, lymphatic nodes, and muscle—5%). Other tissues that comprise the remainder (i.e., adernals, gall bladder, heart, kidneys, pancreas, prostate, small intestine, spleen, and thymus) were considered to yield no radiation for these scans. The tissue weighting factors recommended by the International Commission on Radiological Protection Publication 103 were applied to calculate the effective doses for each protocol [1]. The mean effective doses were compared by the two-way analysis of variance (ANOVA) with the Tukey post hoc test to evaluate the effect of the FOV and device. All analyses were performed using IBM SPSS Statistics software v. 24.0 (IBM Corp., Armonk, NY), adopting a significance level of 5%; a p value lower than 0.05 was considered significant. The null hypothesis was that the FOV, and device would not influence the effective dose.

Results

Table 4 shows the equivalent doses to each organ or tissue and mean effective dose and standard deviation for the different FOVs in the CBCT devices assessed. The effective doses ranged between 69 and 324 μSv for the convex triangular FOVs and between 76 and 332 μSv for the cylindrical FOVs. When comparing the effective dose between the convex triangular FOVs and the cylindrical FOVs of similar height (intra-device), the R100 device showed a reduction in the effective dose ranging from 2.3 to 15.3% (p < 0.05) and from 8.8 to 11.8% for the X800 device (p < 0.05). The lowest effective doses were found in the convex triangular FOVs restricted to the maxillary regions (p < 0.05). In contrast, the highest doses were observed in the cylindrical FOVs involving both maxillary and mandibular regions (p < 0.05). Furthermore, when comparing the R100 and X800 convex triangular FOVs with the same dimensions, doses were 48 to 57% higher in the R100 device (p < 0.05). The highest dose for the convex triangular FOVs corresponded to the R100 device (p < 0.05).

Table 4 Equivalent doses to organs/tissues (μSv) and effective doses (μSv) according to FOV and CBCT device

The average contribution of each organ to the effective dose is shown in Table 5. The highest equivalent doses to organ and tissues were measured in the salivary glands among all FOVs and devices, ranging from 1525 (X800) to 7760 μSv (R100 device). Furthermore, an average increase of 60% in the absorbed dose was observed in the thyroid among all FOVs of the R100 when compared with the X800. The highest contributions come from the remainder organs, followed by the salivary glands, thyroid gland, and bone marrow. The brain, bone surface, esophagus, and skin have minimal contributions.

Table 5 Average and relative contribution of each organ to the effective doses according to the CBCT device and FOV

Discussion

The present study evaluated organ/tissue equivalent doses and effective doses of convex triangular FOVs of the R100 and X800 CBCT devices and compared them with cylindrical FOVs of similar height and diameter of the same devices under standardized energetic parameters. It is essential to highlight that comparing devices and FOVs helps to understand the effective dose to which patients are exposed. This can assist oral and maxillofacial radiologists and dental practitioners in selecting the most appropriate FOV for a specific diagnostic need.

The convex triangular FOVs showed discreetly lower effective doses than the cylindrical FOVs of the same devices. A previous study suggested that the convex triangular FOV decreases the effective dose by modifying the characteristic isocentric trajectory observed in cylindrical FOVs and its consequent momentaneously augmentation of the radiation source to object distance, which diminishes the intensity of the X-ray beam [9]. Our study corroborates that hypothesis by assessing convex triangular FOVs of different heights on two CBCT devices and contrasting them with corresponding cylindrical FOVs of similar size. In this sense, the discreet dose reduction could be important when assessing larger extensions of the maxillomandibular complex since the convex triangular FOVs (100 × 50 mm and 100 × 80 mm) of the R100 and X800 CBCT devices cover a larger region than the cylindrical FOVs (80 × 50 mm and 80 × 80 mm) of the same devices.

A tendency in the effective dose results was observed among all FOVs and devices: the bimaxillary FOVs showed a greater effective dose, followed by the mandibular FOVs, while the maxillary FOVs showed the lowest effective dose. This tendency reflects the importance of the absorbed dose of the salivary glands into calculation of the effective dose, even when the remainder tissues are weighted twelve times higher [11]. Furthermore, this reinforces the impact of the location of the FOV in the calculation of effective dose [16]. The effective dose increases when the FOV height involves a considerable portion of the salivary glands.

Some device differences must be pointed out to understand the differences in the effective doses between the similar dimension FOVs of the devices assessed. First, the R100 and X800 devices are hybrid, i.e., they can be used to acquire panoramic and CBCT images. In both R100 and X800 panoramic image acquisition, the x-ray source is negatively angulated so that the projection beam can pass through the object in an inferior-superior direction. However, for the X800, the radiation source adjusts to become perpendicular to the object and detector, whereas in the R100, it remains negatively angled. Thus, the X-ray beam in the R100 device passes through more radiosensitive organs (thyroid, salivary glands, esophagus’ soft tissue, and skin) than in the X800, which increases the final effective dose to the patient. In addition, the source-to-detector distance and consequently the source-to-patient distance are different between devices. The X800 device presents longer distances (source-to-detector 60 cm and source-to-patient 40 cm) than the R100 device (source-to-detector 51.85 cm and source-to-patient 34.06 cm). In this sense, according to the inverse square law, the shorter the source-to-patient, the higher the intensity of the X-ray beam, and, consequently, higher the effective dose. Furthermore, when the source-to-detector distance is smaller, the beam needs to be more divergent to acquire the same field of view size as when the distance is larger. This increased divergence results in more exposure of the irradiated area, which leads to a higher effective dose due to direct and scattered radiation of radiosensitive organs. These differences explain the higher dose (up to 57%) in the convex triangular FOVs for the R100 device than in the X800 device. However, an earlier study revealed that when subjected to standardized energetic parameters (90 kV and 5 mA), the R100 device demonstrated a greater tendency towards shape distortion in high-density materials than the X800 device [13]. This observation suggests that reducing the tube current in the R100 device to compensate for the shorter source-to-detector distance may decrease image quality.

During the imaging process, both devices emitted continuous radiation, which could significantly impact the final effective dose. However, to avoid bias and prevent doubling the dose in the cylindrical FOVs, all FOVs were set to a 180° rotation angle. It is worth noting that the 100 × 80 mm convex triangular FOV of device R100 had fewer basis images than the FOV of similar dimensions for device X800, which also resulted in lower spatial resolution (according to raw data). The authors hypothesize that the acquisition velocity in the 100 × 80 mm convex triangular FOV of device R100 was faster than that of device X800. However, the effective dose in the 100 × 80 mm convex triangular FOV of device R100 was higher than that of device X800. This finding highlights the impact of the angulation of the X-ray beam on the final effective dose.

Concerning the contribution of organs to the effective dose, we found that in addition to the remainder tissues, the salivary glands, and thyroid glands, the bone marrow also significantly contributes to the effective dose, which agrees with a previous study assessing different CBCT devices [ 11]. However, we found a greater contribution of the esophagus to the effective dose. This could be explained because the most recent ICRP weighted tissue factors ponder these tissues with a higher impact in the final effective dose [1].

According to the manufacturer, the actual value of the convex triangular field of view is to cover more volume than the 80 mm cylindrical FOV while reducing the radiation dose. This is particularly useful for analyzing third molar regions that cannot be covered by the 80 mm cylindrical FOV. As the larger FOV in the R100 device is the convex triangular FOV, this is a reasonable option for covering both third molar regions simultaneously, above all, when it is necessary to compare both sides. On the other hand, the X800 device presents a 150 mm diameter FOV, but its resolution is lower than the convex triangular FOV. Even when our objective was not assessing image quality, it is crucial to consider this topic when discussing radiation dose. Thus, previous in vitro studies showed that the image quality of the convex triangular FOVs is lower than the cylindrical FOVs [12,13,14]. In this sense, the magnitude of artifacts in the convex triangular of the X800 device was greater than in the cylindrical FOVs [12]. Furthermore, the volumetric alteration artifact was influenced by the convex triangular FOV in both CBCT devices R100 and X800, with different manifestations among the materials assessed [14]. Finally, the shape distortion was even greater in the convex triangular FOVs than in the cylindrical FOVs of both R100 and X800 CBCT devices [13]. For this reason, it is essential to emphasize that even when the convex triangular FOVs assessed in the present study showed lower effective doses than the cylindrical FOVs, this reduction may be irrelevant. Therefore, there is still the need to evaluate in clinical conditions the relationship radiation dose-image quality for an appropriate dose optimization and the ALADAIP principle [6]. Thus, assessing diagnostic tasks related to the third molar regions is crucial to determine the real benefit of the convex triangular FOVs.

The present study used OSLDs for the dosimetry measurements, which present advantages over thermoluminescent dosimeters (TLDs). OSLDs can be re-used after each cycle of scans and dose readings, which could be beneficial for data recovery [20, 21]. One potential disadvantage of OSLDs is their angular dependence due to their geometric shape. However, a previous study has shown that the angular dependence of the OSLDs for radiation dose measurements is subtle [19]. In the current study, the OSLDs were placed in the same position for all scan protocols to ensure consistency.

Conclusion

The convex triangular fields of view delivered slightly lower effective doses than the cylindrical fields of view of similar dimensions in the R100 and X800 CBCT devices.