Introduction

The attractive features offered by magnetic microbeads (MBs) in the development of electrochemical immunosensors with improved performance [1,2,3] have made the number of applications grow considerably [4]. On the contrary, far fewer reports involving the use of magnetic nanoparticles (MNPs) have been found so far in the literature [5]. Immobilization of immunoreagents onto functionalized magnetic particles instead of the electrode surface minimizes some drawbacks related to practical application of immunosensors such as the long time required for immobilization, the steric hindrance of the electron transfer, or their reusability [4]. Moreover, the easy capturing of magneto-conjugates by simple application of a magnetic field diminishes nonspecific adsorptions and makes possible the analysis of complex samples by improving the selectivity and avoiding in a great extent matrix effects. Other interesting advantages of the use of MBs include the possibility to achieve high loadings of immobilized biomolecules due to the large surface area of these particles, which contributes to decrease the limits of detection, and faster assay kinetics because the beads are under continuous stirring in suspension thus reducing largely the assay time. The efficient magnetic capture of MBs on the electrode surface also implies that detectable products, usually generated through an enzyme reaction, are formed very close to the electrode surface, thus allowing the electrochemical response to be reached rapidly. Pividori’s group compared the application of magnetic particles with diameters of 1 μm and 300 nm for the construction of electrochemical immunosensors using magneto graphite-epoxy composite electrodes for the determination of Plasmodium falciparum [6]. An improved reactivity with enhanced efficiency in coupling the capture antibody and higher sensitivity were observed when the smaller diameter particles were employed. Same authors prepared immunomagnetic sensors for the determination of Salmonella Typhimurium in whole milk, where the smaller magnetic particles provided lower limits of detection but also suffered from higher matrix effects in the analysis of samples [7].

With the double objective to develop immunosensors using magnetic particles as solid supports where the immunoreactions are carried out and to give useful insights about the influence of the magnetic particle size on the immunosensor performance, in this work, we have compared two different electrochemical sandwich immunoassay strategies using magnetic particles similarly functionalized but with significant differences in size. Commercial carboxylated magnetic microbeads (cMBs) and magnetic nanoparticles (cMNPs), with diameters of 2.8 μm and 20 nm, respectively, were employed. It is worth to highlight at this point the differences between this study and those mentioned above [6, 7]. Apart from the use of disposable SPCEs instead of graphite-epoxy composite electrodes to perform the electrochemical transduction, the size of the used cMNPs is more than 10 times smaller than that tested before.

As the target protein for this study, we selected the tyrosine phosphorylated protein AXL [8]. The proteolytically processed extracellular fraction of this protein (sAXL) is considered a relevant biomarker in cancer [9], inflammatory processes [10] and pathophysiology of heart failure (HF) [11]. In fact, elevated serum levels of sAXL have been found in HF patients with an established cut-off value of 71 ng mL−1 with respect to healthy individuals [12], and they are also associated to inflammatory biomarkers in cardiovascular disease [12, 13].

Although various ELISA kits allow the detection of sAXL concentrations as low as 2 pg mL−1, the reproducibility of the measurements yields coefficients of variation around 10% and require long assay times (more than 4 h) and expensive and non-portable instrumentation. In this context, electrochemical immunosensors are considered as a very attractive alternative due to inherent high sensitivity and selectivity, precision and accuracy, relatively low cost, minimum sample requirement, simplicity of operation and possible integration in compact analytical devices [14, 15]. It is worth to mention at this point that the only one electrochemical immunosensor for sAXL reported so far was developed by our group [16] and involved a sandwich assay at screen-printed carbon electrodes modified with electropolymerized poly(pyrrole-propionic acid).

The comparison of the electrochemical sandwich immunoassay strategies involving the use of cMBs and cMNPs carried out in this work relied on the amperometric transduction, upon magnetic capture of the magnetoimmunocomplexes, on SPCEs at −0.20 V (vs Ag pseudoreference electrode) using the H2O2/HQ system. Once their analytical characteristics were critically compared, both immunosensors were applied to the analysis of human serum samples.

Experimental

Apparatus and electrodes

Amperometric measurements were performed with a CHI 812 B potentiostat (CH Instruments, www.chinstruments.com) controlled by CHI 812 B software. Cyclic voltammograms and electrochemical impedance spectroscopy (IES) measurements were made using a PGSTAT 101 potentiostat from Autolab controlled by Nova 1.6 electrochemical software, and an Antolab type III controlled by FRA2 software (EcoChemie, www.ecochemie.nl), respectively. The screen-printed carbon electrodes (SPCEs) (DRP-110, 4 mm ∅ DropSens, S.L., www.dropsens.com) included a carbon counter electrode and an Ag pseudo-reference electrode. A specific cable connector (DRP-CAC, DropSens, S.L.) acted as interface between the SPCEs and the potentiostat. All measurements were carried out at room temperature. A P-Selecta (Scharlab, http://scharlab.com) ultrasonic bath and a Thermomixer MT100 constant temperature incubator shaker (Universal Labortechmik, www.treffpunkt-labor.de/index.php?impressum) were also used. Magnetic separation during the incubation/washing steps was performed using a DynaMag™-2 (Dynal Biotech ASA, Thermo Fisher, www.thermofisher.com) magnetic particle concentrator. A neodymium magnet (AIMAN GZ, www.aimangz.es) was used to control the attraction of the modified cMBs and cMNPs to the SPCE surface. A Bunsen AGT-9 Vortex was also used for the homogenization of the solutions.

Reagents and solutions

All reagents used were of the highest available grade. Tween®20, hydroquinone (HQ), hydrogen peroxide (30%, w/v), N-(3-dimethylaminopropyl)-N′-ethylcarbodiimide (EDC), N-hydroxysulfosuccinimide (Sulfo-NHS) and ethanolamine were purchased from Sigma-Aldrich (www.sigmaaldrich.com). Sodium di-hydrogen phosphate, di-sodium hydrogen phosphate and sodium chloride were purchased from Scharlau. 2-(N-morpholino) ethane sulfonic acid (MES), from Gerbu Biotechnik, Gmbh (www.gerbu.de), was also used. Blocker casein solutions (blocking buffer) consisted of a ready-to-use 1% w/v purified casein prepared in phosphate-buffered saline (PBS 1×, pH 7.5) (Thermo Scientific). Deionized water was obtained from a Millipore Milli-Q purification system (18.2 MΩ cm).

Carboxyl-functionalized magnetic microbeads (cMBs, 2.8 μm, 10 mg mL−1, 2×109 particles mL−1, Dynabeads® M-270 Carboxylic Acid) were purchased from Dynal Biotech ASA (Thermo Scientific). Carboxyl-modified magnetic nanoparticles (cMNPs, 20 nm, 5 mg mL−1, 8.6×1014 particles mL−1, 79–02-201 nanomag®Dspio) were from Micromod Partikeltechnologie GmbH (www.micromod.de). Buffers used were 0.05 M sodium phosphate buffer of pH 7.5 supplemented with 0.1 M NaCl, PBS consisting of 0.01 M phosphate buffer containing 137 mM NaCl and 2.7 mM KCl (pH 7.5), 0.05 M phosphate buffer of pH 6.0, 0.1 M Tris-HCl (pH 7.2) and 0.025 M MES buffer of pH 5.0.

Mouse anti-human AXL antibody (anti-AXL), biotinylated goat anti-human AXL antibody (Biotin-anti-AXL), and recombinant human AXL were used. These bioreagents are components of the Human Total AXL DuoSet®IC ELISA kit from R&D Systems, Inc. (www.rndsystems.com) Catalog Number DYC 1643–2. This commercial ELISA kit was also used for comparison purposes. A high sensitivity streptavidin-horseradish peroxidase (HRP-Strept) conjugate (Ref 000000011089153001) from Sigma-Aldrich was also used. Solutions of anti-AXL antibodies were prepared in 25 mM MES buffer of pH 5.0, while blocking buffer was used to prepare Biotin-anti-AXL, AXL and HRP-Strept solutions.

Human cardiac troponins I and T (cTnI and cTnT), and N-terminal pro b-type natriuretic peptide (NT-proBNP), from HyTest Ltd. (www.hytest.fi); lipoprotein (a) (Lp(a)) and human C-reactive protein (CRP) from Audit Diagnostics; human tumor necrosis factor (TNF) (R&D Systems), and human interleukin 8 (IL-8) (Abcam) were tested as potential interferents.

Procedures

Preparation of HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs) immunoconjugates

All the incubation steps described below were performed at 25 °C and 950 rpm. After each incubation step, cMBs were isolated by placing the Eppendorf tube in the magnetic particle concentrator during 3 min while cMNPs were centrifuged at 14,000 rpm during 10 min before removing the corresponding liquid phase.

Three μL of cMBs (or 1 μL of cMNPs) commercial suspension were washed twice with 50 μL of MES buffer for 10 min and isolated. Thereafter, cMBs (or cMNPs) were incubated in 25 μL of a 100 mM EDC and 100 mM Sulfo-NHS solution (prepared in MES buffer) for 35 min. Next, activated particles were washed twice with 50 μL MES buffer and after isolation they were re-suspended in 25 μL of a solution of 10.0 μg mL−1 (cMBs) or 5.0 μg mL −1 (cMNPs) anti-AXL (prepared also in MES buffer) and incubated for 15 min (cMBs) or 30 min (cMNPs). Thereafter, two washing steps with 50 μL MES buffer were performed and the modified magnetic particles were incubated with 25 μL of a 1 M ethanolamine solution prepared in 0.1 M phosphate buffer of pH 8.0, used as blocking solution, for 1 h. The resulting anti-AXL-cMBs or anti-AXL-cMNPs were washed once with 50 μL of Tris-HCl of pH 7.2, and twice with blocking buffer. The so prepared anti-AXL-cMBs or anti-AXL-cMNPs were stored (in filtered PBS pH 7.5 at 4 °C) until they were used for performing the immunoassay. The magnetic conjugates were re-suspended in 25 μL of standard AXL or sample solutions (prepared in blocking buffer) and incubated for 30 min. The resulting AXL-anti-AXL-magnetoconjugates were washed twice with 50 μL of blocking buffer and sandwiched by incubation in 25 μL of 0.5 μg mL−1 (cMBs) or 3.0 μg mL−1 (cMNPs) Biotin-anti-AXL solution (prepared in blocking buffer) for 30 or 15 min, respectively. After washing twice with 50 μL of blocking buffer, incubation was performed in 25 μL of a 1/1000 (cMBs) or 1/5000 (cMNPs) diluted HRP-Strept prepared in the same buffer for 15 or 30 min, respectively. Finally, HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs or HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMNPs were washed twice with 50 μL of phosphate buffer of pH 7.5.

Amperometric measurements

Magnetic immunoconjugates were re-suspended in 50 μL of phosphate buffer (pH 6.0) and captured on the working electrode surface of the SPCE by keeping the SPCE in a horizontal position after placing it in the homemade magnet holding block with an encapsulated neodymium magnet. The ensemble magnet holding block/SPCE with the captured cMBs or cMNPs immunoconjugates was immersed into an electrochemical cell containing 10 mL of 0.05 M phosphate buffer of pH 6.0 and 1 mM HQ (freshly prepared). The amperometric measurements in stirred solutions were performed at a potential value of −0.20 V vs. Ag pseudo-reference electrode upon addition of 50 μL of 0.1 M H2O2 solution until the steady-state current was reached (approx. 1 min).

Analysis of serum samples

Human serum samples were provided from Hospital Clinic of Barcelona. These samples were analyzed with the HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs or HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMNPs immunosensors after just a ten-fold dilution with blocking buffer. Validation of the method was performed by comparison of the results with those provided by the commercial ELISA kit involving the same immunoreagents.

Results and discussion

The immunoassays involved the covalent immobilization of capture antibodies onto cMBs or cMNPs and implementation of sandwich-type configurations using Biotin-anti-AXL and HRP-Strept as the detection enzyme conjugate (Fig. 1). The protocols used for the preparation of the immunosensors are detailed in the Experimental section. In brief, once cMBs or cMNPs were activated by carbodiimide chemistry, anti-AXL antibodies were covalently immobilized and a blocking step of the remaining free active sites on the particles surface was carried out with ethanolamine. Sandwich-type assays involved the target protein and biotinylated detection antibodies. A further conjugation with HRP-Strept and the subsequent capture of magnetoimmunoconjugates onto SPCEs allowed the amperometric determination of sAXL to be performed using H2O2 as HRP substrate and hydroquinone (HQ) as the redox mediator.

Fig. 1
figure 1

Schematic display of the preparation and transduction of amperometric immunosensors constructed for AXL involving cMBs and cMNPs. Relative sizes of the components are not drawn on real scale in order to visualize all of them

Optimization of the experimental variables

The different variables affecting the preparation of HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs) conjugates were optimized taking as a criterion the largest ratio between the currents measured with the immunosensors in the presence (S) or in the absence (N) of the target protein. These optimization studies involved evaluation of: a) the anti-AXL loading and the corresponding incubation time on cMBs or cMNPs; b) the incubation time of AXL onto anti-AXL-cMBs (or -cMNPs); c) the Biotin-anti-AXL loading and incubation time on AXL-anti-AXL-cMBs (or -cMNPs); d) the HRP-Strept loading and incubation time on Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs). Details about these optimization studies are provided in the Supporting Information and in Figs. S1S4. Table 1 summarizes the tested ranges and the selected values. Other variables such as the composition of the H2O2/HQ system or the detection potential used were optimized in previous works [17]. It is remarkable that, under the optimal conditions, the immunosensors prepared with cMBs and cMNPs using 5.0 or 25.0 ng mL−1 AXL standard solutions, respectively, yielded specific-to-unspecific ratios (S/N) slightly higher than 9 and 6, respectively, with unspecific cathodic currents (obtained in the absence of AXL) lower than 200 nA in both cases.

Table 1 Optimization of the experimental variables involved in the preparation of HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs)/SPCE immunosensors

Analytical performance of the immunosensors

Calibration plots for AXL standard solutions were constructed with the HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs)/SPCE immunosensors under the optimized working conditions (Fig. 2). The corresponding analytical characteristics are summarized in Table 2.

Fig. 2
figure 2

Calibration plots for AXL constructed with electrochemical HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (a), or -cMNPs (b) /SPCE immunosensors. See the text for the other conditions. Error bars were estimated as triple of the standard deviation

Table 2 Analytical characteristics for AXL obtained from calibration plots constructed with HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs)/SPCE immunosensors

Interestingly, both linear ranges are suitable for the determination of sAXL in human serum after an adequate dilution, where the established cut-off value is 71 ng mL−1 [12]. As it is observed, the immunosensor prepared with cMBs exhibited a shorter range of linearity but provided a 3.7-times higher sensitivity. This observation contradicts the few reported works where the performance of electrochemical immunoassays using different sized magnetic particles is compared [6, 7]. This discrepancy may be attributed to the use of much smaller nanomagnetic particles (20 vs 300 nm) than those employed in the previous works which may cause the cMNPs agglomerate more and be captured less efficiently on the surface of the SPCE. This poorer capture efficiency would justify the slightly worse behavior of these immunosensors with respect to those prepared with the cMBs. These results, taken together with those reported previously by other authors, show that while magnetic particles of hundred nanometers seem to allow a better sensitivity with the immunosensors than that obtained using conventional microbeads, sensitivity is worsened when particles of tens of nanometers are involved.

The immunosensors constructed with cMBs or cMNPs allow the determination of 4.5 times lower AXL concentrations than that achieved with the only immunosensor reported so far for AXL [16]. Moreover, the assay time is significantly reduced (2 h 15 min) as well as with respect to commercial ELISA kits (even more than 4 h counting from coating of the of capture antibody).

The reproducibility of the amperometric measurements for 1.0 ng mL−1 AXL was tested with ten different immunosensors fabricated using cMBs or cMNPs on the same day and different days. Relative standard deviation (RSD) values of 3.8 (cMBs) and 3.6% (cMNPs) were calculated for measurements made in the same day, whereas RSD values of 3.8 (cMBs) and 6.5% (cMNPs) were obtained when measuring in different days.

The storage stability of the anti-AXL-cMBs and anti-AXL-cMNPs conjugates was checked by keeping them at 4 °C in eppendorf tubes containing 50 μL of filtered PBS (pH 7.5). Control charts constructed by taking the mean value of ten measurements made in the presence of 1.0 ng mL−1 AXL on the first day of the sets of experiments as the central value, and ±3 × s of this value as the upper and lower control limits, respectively, showed as that the amperometric responses obtained with the immunosensors prepared using the stored conjugates remained within the control limits for at least 25 days (no longer times were assayed) using both types of magnetic particles.

The selectivity of the immunosensors toward AXL was tested using various non-target proteins (cTnI, cTnT, NT-proBNP, Lp(a), CRP, TNF, IL-8) as negative controls. The amperometric responses obtained with the immunosensors for 0.0 and 5.0 ng mL−1 AXL were compared with those measured both in the absence and in the presence of the potential interfering compound at the concentrations expected in serum from healthy people [18, 19]. As an example, the results obtained with the immunosensor prepared using cMBs are displayed in Fig. 3. As it is shown, similar S/N ratios were measured for AXL and in the presence of cTnI, cTnT or CRP (at 500 ng mL−1), NT-proBNP (7.5 ng mL−1), Lp(a) (2 ng mL−1), TNF (20 pg mL−1), or IL-8 (30 pg mL−1), indicating the absence of significant interference in the presence of all these proteins at the concentrations assayed. Interestingly, this ratio is also similar in the presence of 30 pg mL−1 IL-8, although the amperometric currents for AXL were slightly higher in the presence of IL-8, probably because of a slight cross-reactivity of the used antibodies used toward this protein. As expected, and due to the same immunoreagents were employed when the immunosensors were prepared with cMNPs, similar results were obtained with these immunosensors since the selectivity depends mainly on the antibodies used.

Fig. 3
figure 3

Selectivity tests for HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs immunosensors. Currents measured in the absence (white bars) and in the presence (grey bars) of 5 ng mL−1 AXL in blocking buffer, pH 7.5, and in the presence of cTnI, cTnT and CRP (500 ng mL−1), NT-proBNP (7.5 ng mL−1), Lp(a) (2 ng mL−1), TNF (20 pg mL−1) and IL-8 (30 pg mL−1)

Analysis of human serum

The practical usefulness of the methods implemented with immunosensors prepared with both types of magnetic particles was evaluated by determining the endogenous content of sAXL in serum collected from 5 different HF patients. The possible existence of a matrix effect was evaluated by constructing a calibration plot from the samples, ten-times diluted with blocking buffer, and spiked with growing concentrations of a sAXL standard solution up to 7.5 ng mL−1 (cMBs) or 10 ng mL−1 (cMNPs). The slope values of the corresponding linear calibration plots obtained were 127 ± 9 and 43 ± 1 nA mL ng−1 with the immunosensors prepared with cMBs and cMNPs, respectively, which indeed are not statistically different than the slope values of the calibration plots constructed with sAXL standard solutions: 158 ± 6 nA mL ng−1, and 51.0 ± 0.2 nA mL ng −1 (texp = 1.299 and 2.870, respectively, ttab = 3.355). Accordingly, no significant matrix effect was apparent upon the above- mentioned samples dilution, and the sAXL endogenous concentration was determined in a straightforward mode by interpolating the amperometric responses measured in the ten-times diluted samples into the calibration plot prepared with sAXL standards.

The results obtained by triplicate with the immunosensors for the serum samples collected from the five patients are summarized in Table 3. In addition, and for validation purposes, these samples were also analyzed with a commercial ELISA kit that used the same immunoreagents. A paired samples t-test demonstrated that no significant differences (α = 0.05) existed for the results provided by the two methods (p-value = 0.35). These results demonstrated that, on the contrary to that reported previously [6, 7] describing higher matrix effect using nanomagnetic particles, there is not a significant difference in the potential matrix effect when the determination of sAXL is carried out in ten-times diluted human serum samples with immunosensors prepared with both types of particles used in this work.

Table 3 Determination of sAXL (in ng mL−1) in human serum samples with HRP-Strept-Biotin-anti-AXL-AXL-anti-AXL-cMBs (or -cMNPs)/SPCE immunosensors

Conclusions

Amperometric sandwich immunoassay strategies using SPCEs as electrodes and micro- and nano- magnetic beads as solid supports to perform the immunological reactions were implemented and compared for the determination of sAXL. The analytical characteristics obtained from the calibration plots for sAXL showed a sensitivity 3 times higher and a shorter linear range for the immunosensors prepared using cMBs when compared with those prepared with cMNPs. However, both strategies provided very similar LOD values which were three orders of magnitude lower than the clinical cut-off value established in serum for patients at risk of heart failure. In addition, both strategies demonstrated to be useful for the accurate determination of the target protein in serum samples from HF patients providing results in good agreement with those obtained by applying a conventional ELISA.

Taking together the results presented in this paper and those reported previously, it can be concluded that the sensitivity, linear range and agglomeration and matrix effects observed with immunosensors prepared with magnetic beads are dependent on the size of the particles.

Therefore, since the tendency to suffer particle agglomeration and the potential matrix effects are strongly dependent both on the size of the magnetic beads and the sample complexity and dilution applied, it is difficult to provide in advance general rules about the performance of the resulting immunosensors. Accordingly, comparative experiments should be carried out for particular applications, especially those involving analysis in scarcely pretreated complex samples.