Abstract
We report on the determination of choline and acetylcholine via biocentrivoltammetry. This method combines centrifugation and voltammetry and is based on a carbon paste electrode modified with acetylcholinesterase and choline oxidase. The electrode was placed at the bottom of a biocentrivoltammetric cell. Acetylcholine and choline are accumulated on the enzyme electrode via centrifugative forces, upon which a direct voltammetric scan is applied. Reaction time, pH values, quantities of enzyme and centrifugation parameters were optimized. A linear response is obtained in the 0.07 to 10 μM concentration range of acetylcholine, and a limit of detection as low as 0.5 μM. The linear range is between 0.1 and 500 μM for choline. The method was applied to the determination of acetylcholine and choline in spiked serum samples.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Centri-voltammetry enables to preconcentrate analyte on the electrode surface with proper carrier via centrifugation. Then it provides direct analyte detection without needing any dissolution or filtration processes. It allows a practical way for voltammetric analysis in application of coprecipitation. As a result, preventing of analyte loss provides the sensitive analysis compared with other preconcentration techniques. Centri-voltammetry is firstly used by us to detect trace amounts of Pb (II) ions. Pb (II) ions were accumulated by means of Al(OH)4 and XAD-7 resin carrier materials on the Pt planar electrode. The Pt planar working electrode was placed at the bottom of the cell. LOD value of lead (II) ions was calculated as 5.2 × 10−9 M with XAD-7 resin while with Al(OH)4 this value was calculated as 2.2 × 10−9 M [1, 2]. In another application, mercury (II) ions were preconcentrated on gold film electrode by applying centri-voltammetry [3].
Centri-voltammetry and biosensing system were first used together by our group. We recommended the name “biocentri-voltammetry”. The first biocentri-voltammetric application includes the xanthine detection where xanthine oxidase was immobilized onto planar Pt working electrode. Xanthine accumulation onto this electrode was achieved via centrifugation force without using any carrier precipitate [4]. In another application, acetlycholinesterase (AChE) activity was determined by using biocentri-voltammetry [5]. For this purpose, again carbon paste electrode (CPE) was used as the working electrode. After the optimization of experimental parameters, the developed method was applied for Donepezil-based drug analysis by following inhibition signal of the developed system [5].
Carbon based electrodes are widely utilized in electrochemical analysis because of wide anodic potential window, short response time, low background current and ease in the modification. They were utilized as transducers in many sensor/biosensor systems like glucose, xanthine, ethanol, aspartame and catechol [6–18]. Especially the composite carbon based electrodes were preferred since they can be easily modified and renew [9–18].
Acetylcholine (ACh) is the one of the most important neurotransmitters which was firstly discovered in central nervous system. ACh and its metabolite choline (Ch) play vital roles in brain chemistry. They affect learning, memory and attention. An abnormal amount of ACh causes some of neuropsychiatric disorder such as Alzheimer and Parkinson diseases, progressive dementia and schizophrenia. Therefore sensitive and accurate detection of ACh is very important in clinical applications [19–22]. Determination of ACh and Ch is based on the following enzymatic reactions as shown below:
We describe here the first usage of biocentri-voltammetry for ACh and Ch determination. ACh and Ch molecules were accumulated on the composite electrode via centrifugation force. Accumulated analytes were then detected by applying direct voltammetric scan. For this purpose, a specially designed cell compatible to both voltammetry and centrifuge was constructed. This cell contains AChE/ChO modified CPE as a working electrode at the bottom (Fig. 1b, c and d). After optimization of workings conditions, developed method was tested for ACh detection in artifical serum sample.
Experimental
Apparatus
Voltammetric studies were carried out with the AUTOLAB PGSTAT 12 electrochemical measurement system from ECO CHEMIE Instruments B.V. (The Netherlands) (www.ecochemie.nl) driven by GPES software. Sigma3–16PK centrifuge was used for centrifugation (www.sigma-zentrifugen.de). The cell, made from a delrin tube, was constructed to be compatible to the centrifugation system (Fig. 1A and B). Studies were made with CPE. CPE was placed at the bottom of the cell where reference (Ag/AgCl) and counter (Pt rod) electrodes were immersed in the same cell.
Reagents
AChE from Electrophorus electricus (518 Units (U)/mg solid), ChO from Alcaligenes (14 Units (U)mg−1 solid), acetylcholine chloride, choline chloride and CaCl2 were obtained from Sigma (www.sigmaaldrich.com). Graphite powder, glucose, MgCl2 and Tris–HCl were purchased from Merck and mineral oil was obtained from Aldrich. NaCl was obtained from Pancreac, while KCl was purchased from Riedel-de Haen. Urea was obtained from Horasan Chemistry. Phosphate buffer system was (50 mM pH 7.0) served as supporting electrolyte. All solutions were prepared by dilution of proper amount of the reagents with phosphate buffer.
Electrode preparation
AChE/ChO-modified CPE was prepared by hand-mixing of proper amount of graphite powder with mineral oil, AChE and ChO. A portion of the resulting paste was then packed firmly into the electrode cavity. Electrical contact was established via a copper wire. The paste surface was smoothed on a weighing paper and rinsed carefully with double distilled water.
Procedure
Desired amount of ACh or Ch was put into biocentri-voltammetric cell for 5 mins. After the completion of enzymatic reaction, the cell was centrifuged for 4 min at 4000 rpm. Then the cell was carefully placed in the voltammetric stand and reference and counter electrodes were immersed into the solution. D.P voltammograms were recorded in the range 0 to −0.9 V. The decrease of the reduction current of molecular oxygen through the subsequent enzymatic reaction was monitored for detection of ACh and Ch.
Sample application
Artificial serum solution was prepared by using 140 mM NaCl, 4.5 mM KCl, 2.5 mM CaCl2, 0.8 mM MgCl2, 2.5 mM urea and 4.7 mM glucose in 10 mM Tris–HCl. The final pH of the solution was adjusted to 7.3 with the addition of concentrated HCl [23]. Differential pulse voltammetric measurements were performed after transferring the corresponding solution to the biocentri-voltammetric cell. The potential range was between 0 and −0.9 V versus Ag/AgCl.
Result and discussions
Effect of reaction time
In order to increase substrate and ezyme interaction, effect of reaction time was investigated. 10 mL of 50 mM phosphate buffer solution including 0.25 mM ACh was put into biocentri-voltammetric cell. Then this solution was allowed to be reacted at room temperature for 0.5, 1, 2, 5, 8 and 10 min. After that, the cell was centrifuged for 5 min. at 3000 rpm. Results were expressed in terms of % biosensor response. The current that was obtained at the optimal working conditions was assumed as 100 % and other measured values were calculated relative to this value [24]. As shown in Fig. 2a, response current of enzyme electrode increases from 0.1 to 5 min. and then a decrease is obtained. Therefore 5 min. was selected as optimum reaction time and used for further studies. From the results it is obvious that the reaction completes at 5 min. and after this period, there is a slight activity loss. At this point, it could be stated that the enzyme could loss the catalytic activity for longer reaction times. The formation of un-reactive species could be the reasons for that. Thiocholine might form a dimer or enzyme can react with cysteine residues and as a result these species can be formed [5]. On the other hand, 5 min. is a very short measuring period demonstrating the practicality of the developed method.
pH influence
The pH of the buffer solution effects the biosensor response because of altering protonation state of enzyme. Therefore, influence of pH on the biosensor response was investigated using phosphate buffer solution in a range from pH 5.5 to 8.0. From Fig. 2b, it is clearly be seen that peak values increase from pH 5.5 to 7.0 and then a sharp decrease is obtained which might be due to the enzyme denaturization. As a result pH 7.0 was chosen as optimum pH for further studies. This value is in accordance with previous study, concerning the detection of ACh [19].
Effect of enzyme amount
Amount of enzymes in the bienzymatic sensor were investigated for examining the effect towards ACh and Ch response. ChO amount was firstly tested (1.4 U to 7.0 U) for 0.25 mM Ch (Fig. 3). As clearly can be seen from Fig. 3F, response current of the enzyme elecrode to Ch increases from 1.4 U to 5.6 U, and then a sharp decrease is obtained at 7.0 U. As a result 5.6 U was used as the optimum enzyme amount for this bienzymatic sensor.
For ACh biosensor, AChE amount effect was searched between 51.8 and 259 U and ChO concentration was keep constant at 5.6 U (Fig. 4). As you can see from Fig. 4F, the highest current value was obtained at 207 U. So this AChE amount was used for further experiments.
The decreases after certain enzyme amount for both enzymes might be attributed to the electrode passivation due to the presence of higher protein amount that might also cause diffusion problems [5]. As mentioned above, further experiments were conducted by using these optimum amounts.
Effect of centfifugation parameters
Centrifugation force is a critical key at centri-voltammetric/biocentri-voltammetric systems [1, 2, 4, 5]. Because in these techniques, the analyte was accumulated on the working electrode with centrifugation force. Hence centrifugation parameters such as centrifugation time and speed are important and needed to be optimized.
Effect of centrifugation time was examined between 0 and 6 min. with an increment of 1 min. for 0.25 mM ACh. Results were presented at Fig. 5. Response current of electrode to Ach, increases from 0 min. to 4 min (Fig. 5). Then a decrease is obtained. The decrease at higher centrifugation time longer than 4 min. is attributed to dispersion of accumulated analyte on the electrode surface [4, 5]. So further experiments were conducted by utilizing 4 min. as the optimum centrifugation time.
Centrifugation speed is another centrifugation parameter that is needed to be investigated in biocentri-voltammetry. For this purpose, 0, 1000, 2000, 3000, 4000, 5000 and 6000 rpm were applied to system. As you can see at Fig. 6, best current value was obtained at 4000 rpm. This increment can be attributed to the increase in convection process that causes more analyte accumulation on the electrode surface. On the other hand, higher centrifugation speeds more than 4000 rpm might cause removal of the substrate from the accumulation layer [5]. As a result, 4000 rpm was used for further experiments as centrifugation speed parameter.
Analytical characteristics and sample application
Under the above established experimental results, linear range for ACh (Fig. 7a) was obtained between 0.5 and 10 μM (R 2 = 0.9943) with RSD value of 4.56 (n = 5). LOD, based on s/n = 3, was calculated as 5.1 × 10−7 M. For comparison, a linear graph was formed for ACh without applying any centrifugation (Fig. 7b). As a result, a linear range between 2 and 25 μM with R 2 = 0.7035 were obtained. As it is clearly be seen biocentri-voltammetry possess wider linear range with better R 2 value. The comparison of biocentri-voltammetric methods with other electrochemical biosensor was also done in terms of linear range and LOD (Table 1). As can be seen from the Table, biocentri-voltammetry offers comparable linear range with better sensitivity compared to even nanoparticle based systems. Though some other systems like reference [19] in Table 1 offers wider linear range for ACh detection, when enzyme amounts were considered, biocentri-voltammetry could be accepted as more advantageous technique.
The bienzymatic biosensor was also used for Ch detection. Figure 8 shows the calibration curves of the biocentri-voltammetric biosensor for Ch. A linear range was obtained between 1 × 10−5 and 5.0 × 10−4 M. Wider linear range with better LOD value were obtained for ACh detection. This difference is attributed AChE and ChO enzyme amount in the biosensor. Since 207 U AChE activity is much higher than that of 5.6 U ChO activity, it is expected to get better results for ACh [19].
On the other hand, developed biocentri-voltammetric system was applied for ACh and Ch detection in artifical serum sample. Known amounts of ACh and Ch were added to the biocentri-voltammetric cell from stock solution that was prepared by artifical serum solution as described in experimental part. Recovery values were calculated as 104.7 ± 0.15 (n = 3) for ACh and 100.9 ± 0.14 (n = 3) for Ch.
Conclusion
In the present study, biocentri-voltammetry was used for the first time to detect ACh and Ch. ACh and Ch molecules were deposited via centrifugation force onto the AChE/ChO modified CPE. Concerning analytical characteristic values, it can be concluded that biocentri-voltammetry can be utilized for detection of these two analytes. Moreover when compared with even nanomaterial included systems, this technique provides accesible linear range and higher sensitivity (Table 1). Developed system was also applied for detection of ACh and Ch in artificial serum sample. From the promising recovery values given above section, it is clear that developed system can be applied to complex natured samples and not showed any matrix effect. In our previous stıudies, biocentri-voltammetry was applied for xanthine detection [4] and also acetylcholineesterase activity detection. [5] These studies demonstrate the practicability of developed method. On going works concerning application of the technique for different analytes continue in our lab.
References
Kirgoz AU, Tural H, Ertas FN (2004) A new procedure for voltammetric lead determination based on coprecipation and centrifugation preconcentration. Electroanalysis 16:765–768
Kirgoz AU, Tural H, Ertas FN (2005) Centri-voltammetric study with amberlite XAD-7 resin as a carrier system. Talanta 65:48–53
Ürkmez İ, Gökçel HI, Ertas FN, Tural H (2009) Centrifugation: an efficient technique for preconcentration in anodic stripping voltammetric analysis of mercury using a gold film electrode. Microchim Acta 167:225–230
Anik U, Çevik S (2011) Centri-voltammetry for biosensing systems: biocentri-voltammetric xanthine detection. Microchim Acta 174:207–212
Çevik S, Timur S, Anik Ü (2012) Biocentri-voltammetry for the enzyme assay: a model study. RSC Advances 2:4299–4303
Anik Ü, Çubukçu M, Çevik S, Timur S (2010) Usage of bismuth film electrode as biosensor transducer for alkaline phosphatase assay. Electroanalysis 22:1519–1523
Merkoçi A, Anik U, Çevik S, Çubukçu M, Guix M (2010) Bismuth film combined with screen-printed electrode as biosensing platform for phenol detection. Electroanalysis 22:1429–1436
Anik Ü, Cevik S, Timur S (2011) Bismuth film electrode as sensing platform for IgE-anti-IgE interactions. Electroanalysis 23:2379–2385
Wang J, Kırgöz ÜA, Mo JW, Lu J, Kawde AN, Muck A (2001) Glassy carbon paste electrodes. Electrochem Commun 3:203–208
Kırgöz ÜA, Timur S, Wang J, Telefoncu A (2004) Xanthine oxidase modified glassy carbon paste electrode. Electrochem Commun 6:913–916
Kırgöz ÜA, Odacı D, Timur S, Merkoçi A, Pazarlıoğlu N, Telefoncu A, Alegret S (2006) Graphite epoxy composite electrodes modified with bacterial cells. Bioelectrochemistry 69:128–131
Kırgöz ÜA, Odacı D, Timur S, Merkoçi A, Alegret S, Beşün N, Telefoncu A (2006) A biosensor based on graphite epoxy composite electrode for aspartame and ethanol detection. Anal Chim Acta 570:165–169
Çubukcu M, Timur S, Anık Ü (2007) Examination of performance of glassy carbon paste electrode modified with gold nanoparticle and xanthine oxidase for xanthine and hypoxanthine detection. Talanta 74:434–439
Anik U, Cubukcu M (2008) Examination of the electroanalytic performance of carbon nanotube (CNT) modified carbon paste electrodes as xanthine biosensor transducers. Turk J Chem 32:711–719
Anik U, Cevik S (2009) Double-walled carbon nanotube based carbon paste electrode as xanthine biosensor. Microchim Acta 166:209–213
Anık Ü, Çevik S, Pumera M (2010) Effect of nitric acid “washing” procedure on electrochemical behavior of carbon nanotubes and glassy Carbon μ-particles. Nanoscale Res Lett 5:846–852
Çevik S, Anık Ü (2010) Banana tissue-nanoparticle/nanotube based glassy carbon paste electrode biosensors for catechol detection. Sens Lett 8:667–671
Çubukçu M, Ertaş FN, Anık Ü (2012) Metal/metal oxide micro/nanostructured modified GCPE for GSH detection. Curr Anal Chem 8:351–357
Yang M, Yang Y, Yang Y, Shen G, Yu R (2005) Microbiosensor for acetylcholine and choline based on electropolymerization/sol–gel derived composite membrane. Anal Chim Acta 530:205–211
Xue W, Cui T (2008) A thin-film transistor based acetylcholine sensor using self-assembled carbon nanotubes and SiO2 nanoparticles. Sensor Actuator B Chem 134:981–987
Shimomura T, Itoh T, Sumiya T, Mizukami F, Ono M (2009) Amperometric biosensor based on enzymes immobilized in hybrid mesoporous membranes for the determination of acetylcholine. Enzyme Micro Tech 45:443–448
Lopez MS, Perez JPH, Lopez-Cabarcos E, Lopez-Ruiz B (2007) Amperometric biosensors based on choline oxidase entrapped in polyacrylamide microgels. Electroanalysis 19:370–378
Coldur F, Andac M, Isildak I (2010) Flow-injection potentiometric applications of solid state Li+ selective electrode in biological and pharmaceutical samples. J Solid State Electrochem 14:2241–2249
Kırgöz ÜA, Timur S, Odacı D, Perez B, Alegret S, Merkoçi A (2007) Carbon nanotube composite as novel platform for microbial biosensor. Electoanalysis 19:893–898
Doretti L, Ferrara D, Lora S, Schiavon F, Veronese MF (2000) Acetylcholine biosensor involving entrapment of acetylcholinesterase and poly(ethylene glycol)-modified choline oxidase in a poly(vinyl alcohol) cryogel membrane. Enzym Microb Technol 27:279–285
Shimomura T, Itoh T, Sumiya T, Mizukami F, Ono M (2009) Amperometric biosensor based on enzymes immobilized in hybrid mesoporous membranes for the determination of acetylcholine. Enzym Microb Technol 45:443–448
Houa S, Oub Z, Chenc Q, Wub B (2012) Amperometric acetylcholine biosensor based on self-assembly of gold nanoparticles and acetylcholinesterase on the sol–gel/multi-walled carbon nanotubes/choline oxidase composite-modified platinum electrode. Biosens Bioelectron 33:44–49
Acknowledgments
The grant from The Technical and Scientific Council of Turkey (TUBİTAK) Project no: 109T885 and Mugla Sıtkı Kocman University BAP project no 12/08 are gratefully acknowledged.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Çevik, S., Timur, S. & Anik, Ü. Biocentri-voltammetric biosensor for acetylcholine and choline. Microchim Acta 179, 299–305 (2012). https://doi.org/10.1007/s00604-012-0895-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00604-012-0895-1