Abstract
Introduction
Limb lengthening by external fixators is associated with many problems such as pain at the pin tracts, muscle transfixation, pin tract infections, reduced joint motion, and prolonged fixation time. The intramedullary skeletal kinetic distractor (ISKD) is a new internal, mechanically activated implant, which distracts by mild rotations of 3°.
Material and methods
In a prospective clinical study, four patients with an mean age of 29 years (18–36 years) underwent intramedullary lengthening via ISKD. The average lengthening of three femora and one tibia was 31 mm (26–40 mm).
Results
All patients performed the rotations for the distraction themselves without any significant problems. One patient took mild analgesics during the first days of distraction, whereas three patients did not require any analgesics. The average patient discharge occurred 10 days (8–11 days) postoperatively with no complications during the hospital stay. The planned length of distraction was achieved in all patients with normal alignment and normal joint orientation. Full weight bearing was performed on average after 10 weeks (7–14 weeks). Consolidation was noted 80 days (51–111 days) postoperatively with an average consolidation index of 2.9 days/mm. No complications were observed during the follow-up period of 14 months. The Enneking score was 26.8 points, and according to the classification of Paley all patients had an excellent result.
Conclusions
From these preliminary results we conclude that the comfort of limb lengthening with the ISKD is increased by the elimination of fixator-associated complications and by the simple distraction mechanism, which is well tolerated by the patients. Further advantages of the ISKD are early full weight bearing and excellent limb function.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Callus distraction by external fixation allows for regeneration of a high amount of new bone even in patients with severe bone loss [20]. However, callus distraction by external fixation is associated with many problems. Soft tissue transfixation by pins and wires can cause muscle contractures and joint stiffness [15, 20], pain [11, 29], and infections [6, 20]. Furthermore, lengthening by external fixators can lead to secondary axial deformity [20, 28] and refractures of the regenerated bone [7, 20, 27]. The long period of time for external fixation delays rehabilitation and return to normal daily activities. The total rate of problems and complications during limb lengthening by external fixation ranges from 1.0 to 2.8 per patient [6, 9, 18, 20, 21, 28].
A combination of an intramedullary nail and a temporary external fixator, which is only applied during distraction, can reduce the time of external fixation and the risk of infection [21, 24]. Due to early removal of the external fixator, return to normal range of motion is accelerated and the risk of refractures is reduced [21]. However, the risk of pin tract infections continues to represent a significant problem. Severe osteitis may occur if treated improperly [19, 26]. This concern is also known from intramedullary nailing after previous external fixation of femoral fractures [25].
Fully implantable devices have been developed within the last few decades to solve these problems. Betz and Baumgart developed the first motorized lengthening device with a subcutaneous receiver [4], which was successfully implanted to many patients for limb lengthening and bone transport [3]. This motor-driven device is not authorized for general use to our knowledge.
Guichet et al. designed fhe first mechanically activated lengthening device (Albizzia nail). In this device, lengthening is achieved by rotations of 20° around the longitudinal axis of the bone. Mechanical testings of the nail revealed comparable characteristics to conventional intramedullary nails [12]. One major drawback of this device is the induction of severe pain in several individuals, induced by the large degree of rotation necessary for distraction [10, 13].
The intramedullary skeletal kinetic distractor (ISKD) is a mechanically distracting intramedullary nail designed for the femur and tibia [5]. Its proximal and distal part is internally connected with a threaded rod by two one-way clutches. These clutches are activated by rotations of 3–9°. These oscillations are part of the physiological gait process. A distraction of 1 mm is achieved by 160 rotations of 3°. The actual amount of distraction is controlled by an external handheld monitor, which measures the orientation of a magnet on the distal part of the internal threaded rod (Fig. 1). The patients measure the daily distraction at a minimum of five times per day. If the distraction length is insufficient, the leg is rotated under the control of the monitor until the desired distraction length is reached. The maximal distraction length by the ISKD is 80 mm. The ISKD is available for both the femur and the tibia [5, 10, 12, 13].
Patients and methods
After approval of the local ethics committee (approval no. 2948), a prospective study was performed on four patients with a leg discrepancy who were operated between 16 July and 30 July 2002. Inclusion criteria were patients with the need of 20–80 mm lengthening for limb length discrepancy due to short femur or tibia, a minimal patient age of 18 years, compliant patients who understand the nature of the device, and no history of osteitis for at least 1 year. Exclusion criteria were patients who cannot bear weight on the contralateral limb, patients in whom an osteotomy cannot be made in the proximal or mid-third of the shaft of the bone, deformities that require multilevel osteotomies, and patients with systemic bone diseases.
Every 2 weeks during distraction, and every 4 weeks during consolidation, radiographs were taken in anteroposterior (AP) and lateral directions. Additionally, at these time points the use of analgesics was recorded, range of motion of the hip and knee joint, and any problems or complications. The duration of the hospital stay and time to full weight bearing was analyzed.
Preoperatively and at follow-up examination, the functional status was assessed by the Enneking score (function, pain, emotional acceptance, supports, walking, and gait 5 points each, maximum 30 points) [8] and by Paley’s classification (range of motion, leg discrepancy, gait, joint orientation, pain, and ability to perform activities; maximum 100 points; excellent=95–100 points, good=75–94 points, fair=40–74 points, and poor=less than 40 points) [21].
Consolidation was defined when the distraction gap was corticalized on three of four sides as seen on AP and lateral radiographs. The distraction index was calculated from the length of the radiographic distraction gap divided by the time between beginning to the end of the distraction. The consolidation index was calculated from the interval of the operation date to the date of radiographic consolidation divided by the length of the distraction gap. Pre- and postoperative alignment, joint orientation, and leg length were measured on digital bilateral leg standing radiographs with special computer software (MediCAD version 2.0, Hectec, Altfrauenhofen, Germany) and compared to the normal values described by Paley et al. [22].
Three femoral and one tibial ISKD (Orthofix, Valley, Germany) were implanted into one female and three male patients with an average age of 29 years (18–36 years). Two femora had a complex deformity with a combined 7° varus and 28° rotational deformity and a 12° varus and 41° rotational deformity. A third femur had a mild valgus deformity of 5°. Apart from these three post-traumatic deformities, one congenital tibial shortening was addressed with an ISKD. Preoperatively, the adequate length and diameter of the ISKD was determined on plain radiographs. Different from conventional femoral nails, the femoral ISKD is straight, which has to be considered in preoperative planning. The final follow-up investigation was made 14.2 months postoperatively (14.0–14.5 months).
Intraoperatively, the rotation was controlled by two 3.0-mm Kirschner wires, which were placed parallel in the proximal and distal segments. For femoral osteotomy a multiple drill hole technique was used: with control by the image intensifier, transverse drillings were made through a lateral stab incision, and the osteotomy was completed with a small osteotome. For tibial osteotomy, a Gigli saw was used [23]. Special attention was paid to create transverse, straight osteotomies, which allowed rotational movements for the distraction of the ISKD. If necessary, rotation was corrected under the control of the angle between the Kirschner wires in the proximal and distal fragments. Corrections in the frontal plane were performed with the “cable technique” [16]. Due to the straight design of the femoral ISKD, the medullary canal of the femur has to be overreamed 2.0 mm, whereas the tibia has to be overreamed 1.5 mm. Postoperatively, the patients controlled distraction by the external monitor at a minimum of five times per day.
Results
The mean operation time was 108 min (range: 90–145 min) and total intraoperative blood loss 230 ml (100–320 ml). No intraoperative problems or complications occurred. In three patients additional operations were performed (removal of osteophytes n=1, implant removal n=2, tenotomy n=1).
Mobilization was started on the 1st postoperative day. From the 5th day, a daily distraction of 1 mm was intended. One patient required analgesics during the first 3 days of the distraction. The other patients had only mild pain during distraction and did not need any analgesics. The patients were discharged from hospital after an average of 10 days (8–11 days) postoperatively.
All patients completed the planned lengthening with an average distraction index of 1.2 mm/day (0.9–1.8 mm/day) (Fig. 2, Fig. 3, Fig. 4). The average lengthening was 31 mm (26–40 mm). At an average of 10 weeks (7–14 weeks) full weight bearing was allowed. Average hip flexion measured 123° (105–135°) preoperatively, 116° (95–135°) after distraction, and 122° (105–140°) at follow-up examination. Mean knee flexion decreased from 126° (105–140°) preoperatively to 107° (70–125°) after distraction and normalized to 124° (105–135°) at follow-up.
Complete radiographic consolidation was observed 80 days (51–111 days) postoperatively. The mean consolidation index was 2.9 days/mm (1.8–4.1 days/mm). Leg standing radiographs on follow-up examination showed normal alignment and joint orientation [22]. The mechanical lateral distal femur angle (mLDFA) averaged 92.2° (88.5–96.5°) prior to surgery and 88.3° (85.5–90°) at follow-up. Mechanical axial deviation (MAD) was corrected from 21.7 mm (12–31 mm) preoperatively to 8.2 mm (3–13 mm). No hardware failures, infections, non-unions, or malunions were observed.
According to Paley’s classification all patients had excellent results. The functional outcome measured by the Enneking score increased from 13.3 points preoperatively to 26.8 points postoperatively.
Discussion
Callus distraction by external fixation has become a widely accepted treatment method for limb lengthening [1, 2, 11, 17, 18, 21]. However, it was discussed by experts that “complications are the rule rather than the exception” [14]. Removal of the fixator is a critical step, and malalignment and refractures can occur [20]. Many patients dislike the external fixator because of the painful soft tissue transfixation, decreased joint mobility, and long disability caused by external fixation accompanied by delayed return to normal daily activities [15, 29]. Studies of external lengthening devices have demonstrated a rate of severe complications between 24% and 117% [1, 6, 9, 18, 20, 21].
Potential advantages of intramedullary lengthening devices include the reduced risk of contractures and infections, prophylaxis of axis deviation and refractures, reduction of pain due to the elimination of soft tissue transfixation, and earlier return to daily activities. To the best of the authors’ knowledge, the Albizzia nail and the ISKD are the only intramedullary lengthening devices, which are now in general use for limb lengthening [10]. The complication rate of external limb lengthening ranges from 24% to 117% depending in particular on the length of distraction, experience of the surgeon, and age of the patients [1, 6, 9, 18, 20, 21]. In contrast, the complication rate after treatment with the Albizzia nail ranges only from 22% to 29%, if the need for general anesthesia for distraction is not considered [10, 13]. The designer of the ISKD reported even fewer complications with a rate of 11% after 18 lengthenings [5].
In a study of Guichet et al., a considerable number of patients with an Albizzia nail were readmitted to the hospital and rotations of the nail were performed under general anesthesia at some stage of the lengthening because the extensive rotations at the osteotomy site caused severe pain and discomfort [13]. In another the study, 12% of the femoral lengthenings with the Albizzia nail were incomplete because these patients refused epidural anesthesia for the distraction [10].
A limitation of this study is the small number of patients. However, only one publication exists about the ISKD so far, which was published by the designer of the implant [5]. Furthermore, this study analyses on the distraction phase, which can be a very painful period of time for the patients [10, 11, 13, 29]. The analysis of the first four European ISKD patients showed that the ISKD is a relatively simple device, which was well understood by the patients. The mild rotations of 3° were well tolerated enabling the patients to perform the distraction themselves. Furthermore, the ISKD offered the advantages of early rehabilitation, early full weight bearing, and excellent functional results. It provided a high quality of regenerated bone and stable intramedullary fixation.
Nevertheless, several aspects have to be considered in the treatment with an ISKD. Since the ISKD cannot be shortened or corrected in the postoperative course, exact preoperative planning and implantation is essential. Furthermore, good compliance of the patients and understanding are mandatory for successful treatment. The costs of this new implant are higher than those of external fixators. In consideration of the low rate of complications and the reduction in hospital stay due to ambulatory distraction, the ISKD appears to be a very attractive and cost-effective implant. However, prospective, randomized studies have to prove this hypothesis.
References
Aldegheri R (1999) Distraction osteogenesis for lengthening of the tibia in patients who have limb-length discrepancy or short stature. J Bone Joint Surg Am 81:624–634
Aronson J (1997) Limb-lengthening, skeletal reconstruction, and bone transport with the Ilizarov method. J Bone Joint Surg Am 79:1243–1258
Baumgart R, Betz A, Schweiberer L (1997) A fully implantable motorized intramedullary nail for limb lengthening and bone transport. Clin Orthop 343:135–143
Betz A, Baumgart R, Schweiberer L (1990) First fully implantable intramedullary system for callus distraction—intramedullary nail with programmable drive for leg lengthening and segment displacement. Principles and initial clinical results. Chirurg 61:605–609
Cole JD, Justin D, Kasparis T, DeVlught D, Knobloch C (2001) The intramedullary skeletal kinetic distractor (ISKD): first clinical results of a new intramedullary nail for lengthening of the femur and tibia. Injury 32 [Suppl 4]:129–139
Dahl MT, Gulli B, Berg T (1994) Complications of limb lengthening. A learning curve. Clin Orthop 301:10–18
Danziger MB, Kumar A, DeWeese J (1995) Fractures after femoral lengthening using the Ilizarov method. J Pediatr Orthop 15:220–223
Enneking WF, Dunham W, Gebhardt MC, Malawar M, Pritchard DJ (1993) A system for the functional evaluation of reconstructive procedures after surgical treatment of tumors of the musculoskeletal system. Clin Orthop 286:241–246
Faber FW, Keessen W, van Roermund PM (1991) Complications of leg lengthening. 46 procedures in 28 patients. Acta Orthop Scand 62:327–332
Garcia-Cimbrelo E, Curto A, Garcia-Rey E, Cordero J, Marti-Ciruelos R (2002) The intramedullary elongation nail for femoral lengthening. J Bone Joint Surg Br 84:971–977
Garcia-Cimbrelo E, Olsen B, Ruiz-Yague M, Fernandez-Baillo N, Munuera-Martinez L (1992) Ilizarov technique. Results and difficulties. Clin Orthop 283:116–123
Guichet JM, Casar RS (1997) Mechanical characterization of a totally intramedullary gradual elongation nail. Clin Orthop 337:281–290
Guichet JM, Deromedis B, Donnan LT, Peretti G, Lascombes P, Bado F (2003) Gradual femoral lengthening with the Albizzia intramedullary nail. J Bone Joint Surg Am 85:838–848
Herzenberg JE, Paley D (1997) Tibial lengthening over nails (LON). Tech Orthop 12:250–259
Herzenberg JE, Scheufele LL, Paley D, Bechtel R, Tepper S (1994) Knee range of motion in isolated femoral lengthening. Clin Orthop 301:49–54
Krettek C, Miclau T, Grun O, Schandelmaier P, Tscherne H (1998) Intraoperative control of axes, rotation and length in femoral and tibial fractures. Technical note. Injury 29 [Suppl 3]:29–39
Maffulli N, Lombari C, Matarazzo L, Nele U, Pagnotta G, Fixsen JA (1996) A review of 240 patients undergoing distraction osteogenesis for congenital post-traumatic or postinfective lower limb length discrepancy. J Am Coll Surg 182:394–402
Noonan KJ, Leyes M, Forriol F, Canadell J (1998) Distraction osteogenesis of the lower extremity with use of monolateral external fixation. A study of two hundred and sixty-one femora and tibiae. J Bone Joint Surg Am 80:793–806
Oedekoven G, Jansen D, Raschke M, Claudi BF (1996) The monorail system-bone segment transport over unreamed interlocking nails. Chirurg 67:1069–1079
Paley D (1990) Problems, obstacles, and complications of limb lengthening by the Ilizarov technique. Clin Orthop 250:81–104
Paley D, Herzenberg JE, Paremain G, Bhave A (1997) Femoral lengthening over an intramedullary nail. A matched-case comparison with Ilizarov femoral lengthening. J Bone Joint Surg Am 79:1464–1480
Paley D, Herzenberg JE, Tetsworth K, McKie J, Bhave A (1994) Deformity planning for frontal and sagittal plane corrective osteotomies. Orthop Clin North Am 25:425–465
Paley D, Tetsworth K (1991) Percutaneous osteotomies. Osteotome and Gigli saw techniques. Orthop Clin North Am 22:613–624
Raschke MJ, Mann JW, Oedekoven G, Claudi BF (1992) Segmental transport after unreamed intramedullary nailing. Preliminary report of a “Monorail” system. Clin Orthop 282:233–240
Siebenrock KA, Gerich T, Jakob RP (1997) Sequential intramedullary nailing of open tibial shaft fractures after external fixation. Arch Orthop Trauma Surg 116:32–36
Simpson AH, Cole AS, Kenwright J (1999) Leg lengthening over an intramedullary nail. J Bone Joint Surg Br 81:1041–1045
Simpson AH, Kenwright J (2000) Fracture after distraction osteogenesis. J Bone Joint Surg Br 82:659–665
Tjernstrom B, Olerud S, Rehnberg L (1994) Limb lengthening by callus distraction. Complications in 53 cases operated 1980–1991. Acta Orthop Scand 65:447–455
Young N, Bell DF, Anthony A (1994) Pediatric pain patterns during Ilizarov treatment of limb length discrepancy and angular deformity. J Pediatr Orthop14:352–357
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Hankemeier, S., Pape, HC., Gosling, T. et al. Improved comfort in lower limb lengthening with the intramedullary skeletal kinetic distractor. Arch Orthop Trauma Surg 124, 129–133 (2004). https://doi.org/10.1007/s00402-003-0625-6
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-003-0625-6