Abstract.
The measurement of the spectral width of ultrashort light pulses using a Fabry–Pérot interferometer (FPI) is investigated. It is shown, numerically and experimentally, that the measured width critically depends on the pulse properties (such as pulse shape, pulse duration, frequency chirp and wavelength) and on the properties of the FPI (such as the mirror spacing and the mirror reflectivities). The obtained results are of particular importance if the spatial length of the short light pulses is comparable or even shorter than the distance between the FPI mirrors. The derived guideline indicate that the actual spectral width of the ultrashort light pulses is measured with good accuracy only if the finesse F≥40 and the round trip time of the light pulses inside the Fabry–Pérot interferometer is approximately one to three times the pulse duration.