Introduction

Cardiac magnetic resonance imaging (MRI) provides high temporal and reasonable spatial resolution, while images can be acquired in any anatomic orientation. Quantitative measurements of left-ventricular (LV) volumetric and functional parameters are accurate and reproducible, making cine MRI the standard of reference [13]. Thin-section multidetector-row computed tomography (MDCT) of the heart, providing an excellent longitudinal spatial resolution, is increasingly employed for the reliable diagnosis of obstructive coronary artery disease and coronary bypass grafts [47]. Image reformation can be performed in any desired plane. Retrospective electrocardiography (ECG) gating allows the image reconstruction in any phase of the cardiac cycle and subsequent LV volume measurements at end-diastolic and end-systolic phases [8].

Several methods have been employed for the determination of end-diastolic volume (EDV) and end-systolic volume (ESV) from MDCT short-axis reformations. The method of planimetry based on the manual tracing of endocardial contours on diastolic and systolic image series has been used [810]. Planimetric measurements are labor intensive and time consuming [9]. Most of the studies dealing with cardiac function assessment have used semi-automatic segmentation techniques that are available in commercial software [1118]. The obtained contours should be checked visually and manual adjustments to the true endocardial borders are often required [13].

To our knowledge, no reported experience exists about the stereological estimation of LV volumes using CT scans. Stereological volume measurements are based on the simple procedure of point counting and manual or semi-automatic segmentation of endocardial contours is not required. The stereological method has been successfully employed to provide efficient volume estimations of infarct and brain compartments [19], intracranial cavity [20, 21], intravertebral disc [22], liver [23], lung [24], and urinary bladder [25] from CT scans. Cardiac function has been evaluated by applying the stereological method on MR images [26, 27]. However, Roberts et al. [26] presented MR-based volume measurements of only one patient, whereas Graves and Dommett [27] made no attempts to optimize the volumetric method. The optimization procedure of the stereological method is a prerequisite whenever rapid and reliable volume assessments are needed [20, 2830].

The aims of the current study were (1) to optimize the stereological method for estimating EDV, ESV, stroke volume (SV), and ejection fraction (EF) from retrospectively ECG-gated MDCT data sets, and (2) to compare these stereological estimations with the reference values obtained by cine MRI.

Materials and methods

Patients

Seventeen consecutive patients with a mean age of 58.2 ± 8.8 years were prospectively enrolled in the current study. These patients with established or suspected coronary artery disease were referred for MDCT coronary angiography. An additional cine MRI study for the assessment of LV function was performed within 12  h following CT scanning. The institutional review board approved the study protocol. All patients gave written informed consent prior to imaging examinations.

MDCT examinations

All examinations were performed on a 16-detector-row MDCT scanner (Somatom Sensation 16, Siemens Medical Solutions, Forcheim, Germany). Scanning parameters were as follows: 420 ms gantry rotation time, 120 kV, 500 mAs, 0.75 mm beam collimation, and 3.5 mm table feed per rotation [31]. Images were acquired in the craniocaudal direction during inspiratory breath-hold while the patient’s ECG trace was recorded simultaneously. Iodinated contrast agent (120 ml; 300 mg I/ml; Xenetix, Guerbet, Aulnay, France) at a flow rate of 3.5 ml/s was administered via an 18-G access into the antecubital vein. Before MDCT examination, beta-blockers (Brevibloc, Baxter, Unterschleissheim, Germany) were injected intravenously in patients with heart rates exceeding 65 beats/min at a dosage of 0.5 mg/kg.

ECG-gated image reconstruction was performed in 5% steps through the entire RR interval. Twenty axial image series with a section thickness of 1.0 mm and a reconstruction increment of 0.5 mm were obtained. End-diastolic and end-systolic phases were determined as the axial images presenting the maximum and minimum cross-sectional LV areas, respectively. The axial diastolic and systolic image series were transferred to a workstation equipped with the Wizard software package (Siemens, Forcheim, Germany). Contiguous 8-mm multiplanar reformations in the short-axis orientation encompassing the entire LV from base to apex were calculated from the axial images. The pixel size in the short-axis images was 0.45 × 0.45 mm2. For the determination of LV volumetric and functional parameters from MDCT data, the most basal section to be included for analysis was defined as the image presenting LV myocardium in at least 50% of its perimeter. The most apical section was the image with a discernible LV lumen. Papillary muscles were included in the LV cavity.

Stereological estimations from MDCT data sets

In accordance with the Cavalieri principle, the volume of an object can be estimated by cutting it into equally spaced sections from end to end and measuring the area of the object on each section [32]. The estimate of the volume is given by the formula:

$$V = T\sum\limits_{i = 1}^m {A_i } $$

where Tis the section thickness, m is the number of sections and A i is the object area of a section i. The area is usually measured by means of the planimetric technique where the user manually delineates the borders of the object of interest on a section-by-section basis and the software program counts the voxels encompassed by the generated contours. A more efficient method for such area measurements is the stereological method of point counting [32].

Stereological measurements were carried out using the Analyze software (Mayo Foundation, Rochester, MN, USA). According to stereological method, a computer-generated square grid containing an array of test points was superimposed on the first end-diastolic image depicting the LV (Fig. 1). The orientation of the grid was randomly selected in the first short-axis image and it remained unchanged in all subsequent diastolic images. The user had to select all points falling inside the LV. To perform this procedure, the “passive pick” option was enabled. In the passive collection mode, the left mouse button was clicked once and then movement of the cursor over a grid point automatically caused it to be collected. The button did not have to be held down during the entire volume measurement. The procedure would stop any time after a second button click. The total number of selected points was automatically counted by the Analyze software. The above procedure was repeated for estimating the ESV.

Fig. 1
figure 1

A square grid of test points is placed on a short-axis MDCT image for the stereological estimation of end-diastolic volume (EDV). The point spacing is equal to 25 pixels

The software provided LV volume calculations using the following formula:

$$V = TA_p \sum\limits_{i = 1}^m {P_i } $$

where A p is the test point area and P i is the number of points counted on a section i. The test point area in the square grid is given by the formula:

$$A_p = \left( {u \times u} \right)S$$

where u is the distance between the test points in pixels and S is the pixel area. The SV and EF were calculated by the following equations:

$$\begin{array}{*{20}c}{SV = EDV - ESV} \hfill \\{EF = \frac{{SV}}{{EDV}} \times 100\% } \hfill \\\end{array} $$

The precision of the estimated EDV and ESV was expressed as coefficient of error (CE), which contains contributions due to both sectioning and point counting. The CE was calculated using the formula suggested by Cruz-Orive [33]:

$$CE = \left( {\sum\limits_{i = 1}^m {P_i } } \right)^{ - 1} \left[ {\frac{1}{{12}}\left( {3\sum\limits_{i = 1}^m {P_i^2 } + \sum\limits_{i = 1}^{m - 2} {P_i P_{i + 2} - 4\sum\limits_{i = 1}^{m - 1} {P_i P_{i + 1} } } } \right) + 0.0543\frac{B}{{\sqrt A }}\left( {m\sum\limits_{i = 1}^m {P_i } } \right)^{0.5} } \right]^{0.5} $$

where B is the mean boundary length and A is the mean LV area. For all study participants, the quantity A was determined using the point counting process, whereas B was found by the number of intersections between the LV and a square grid of test lines [24]. The mean value of \(\frac{B}{{\sqrt A }}\), known as the shape coefficient, was equal to 4.7. The above mean value was used for all CE calculations. Reported experience has suggested that a precision of 5% or less can be considered as sufficient in stereological applications [32]. The CE of SV and EF estimations was calculated using error propagation analysis.

Optimization of the stereological method

The appropriate point spacing of the test grid and the optimum sampling intensity of short-axis images for the stereological estimation of EDV and ESV were defined. The LV volumes of all patients were measured for six different distances between the test points of the grid. The following point-separation distances were used: 10, 15, 20, 25, 30, and 35 pixels. Moreover, the EDV and ESV were estimated for the sample types of 1/2 and 1/3. A sample type of 1/2 means that two samples can be systematically drawn from the total number of short-axis images depicting the LV during the end-diastolic or end-systolic phases. For example, assume an LV cavity of a patient depicted in ten end-diastolic images numbered 1,2,…,9,10. A sample type of 1/2 will yield two samples with size five. These samples will contain the images numbered {1, 3, 5, 7, 9} and {2, 4, 6, 8, 10}. Systematic sampling of MDCT image reformations was performed for the first seven patients participating in this study.

The reproducibility of the optimized stereological method was evaluated. The EDV, ESV, SV, and EF of all patients were estimated one more time by the same observer. The interval between the first and the second set of measurements was at least 25 days. To find the interobserver variability, a second independent observer experienced in both MDCT imaging and stereological applications determined the LV volumes and EF. The second observer was blinded to the results of all prior experiments. Reproducibility measurements of EDV and ESV were performed using the optimum point-spacing settings.

Standard CT analysis

Left ventricular volumetric and functional parameters were obtained by standard CT analysis employed in everyday clinical practice. The analysis was performed using a commercially available software (Argus, Siemens, Forcheim, Germany). The software supports the semi-automatic detection of endocardial contours in diastolic and systolic MDCT images with a discernible LV cavity. All the segmented contours were studied visually and the LV borders were manually corrected using software tools if necessary. The software provided calculations of all parameters.

MRI study

MRI studies were generated on a 1.5-Tesla whole-body scanner (Magnetom Sonata, Siemens, Erlangen, Germany). The phased-array torso coil and the table-integrated spine coil were employed for signal reception. All patients were examined in supine position during inspiratory breath-hold. A prospectively ECG-gated steady-state free precession cine sequence (TR/TE: 3.0/1.5, flip angle: 60°) was acquired in the short-axis orientation. The entire heart was encompassed with 8-mm sections with no interslice gap. An independent observer with experience in cardiac MRI analyzed short-axis images using the Argus software (Siemens, Forcheim, Germany). The papillary muscles were regarded as being part of the LV cavity. The most basal section to be included for analysis had to cover more than 50% of the LV circumference.

Statistical analysis

Stereological estimations were compared with those by standard CT analysis. Moreover, LV volumetric and functional parameters estimated by cine MRI were compared with the respective values determined by MDCT using the stereological method or standard analysis. Linear regression analysis was performed to examine the relation between different modalities or different volumetric CT methods. A Wilcoxon matched-pairs test was employed to detect possible statistical differences. The Bland-Altman statistical method was used to illustrate the limits of agreement [34]. For each LV parameter, the difference between the values obtained by MDCT and MRI or by the two methods applied on MDCT short-axis reformations was plotted against their mean. Based on the Bland-Altman analysis, the 95% limits of agreement were defined as the mean difference ±1.96SD, where SD is the standard deviation of the differences.

For the optimized stereological method, the SD of measurements obtained by stereological and MRI estimations was plotted against the mean of measurements in order to examine whether the agreement depends upon the size of the volumetric or functional parameter. Intra- and interobserver variability in the stereological estimation of EDV, ESV, SV, and EF was expressed as coefficient of variation (CV) values. A P-value of less than 0.05 was considered to indicate statistical significance. All statistical analyses were performed using the software GraphPad Prism version 4.0 (GarphPad Software, CA, USA).

Results

Optimization of stereological method

The mean CE and the mean number of counted points for the six different distances between the test points of the grid that were employed during stereological EDV and ESV estimations are given in Table 1. The increase in point spacing resulted in a considerable reduction in the number of counted points and, therefore, in a reduction of the time required. The mean time needed to estimate EDV with a point spacing of 10, 15, 20, 25, 30, and 35 pixels was equal to 3.7, 2.2, 1.6, 1.2, 1.1, and 0.9 min, respectively. The corresponding times for ESV estimation were 2.0, 1.1, 1.0, 0.9, 0.9, and 0.8 min, respectively. Based on the results of Table 1, all point distances of 25 pixels or smaller provided acceptable EDV estimations with a CE of less than 5%. The point spacing of 25 pixels was considered as the optimum one because it could provide volume estimates with acceptable levels of precision, while simultaneously reducing the time measurement in comparison with that required for point distances of 10, 15, and 20 pixels. For the same reasons, we decided that the optimum point spacing for ESV estimation is equal to 15 pixels. Stereological EDV and ESV measurements with the aforementioned optimum point-spacing settings were performed in a total mean time of 2.3 ± 0.5 min with a range of 1.8–2.9 min. All above volume estimations were obtained by applying the stereological method on the total number of end-diastolic or end-systolic images depicting the LV cavity. For all point separation distances, the CEs of EDV estimations arising from a sample type of 1/2 and 1/3 were more than 5.6 and 8.9%, respectively. The corresponding CEs for ESV estimation exceeded 7.6 and 10.8%, respectively.

Table 1 Mean values (±SD) of the coefficient of error (CE) and the number of counted points associated with the stereological estimation of end-diastolic volume (EDV) and end-systolic volume (ESV) from MDCT data sets using six different distances between test points of the grid

The mean CEs of the obtained SV and EF estimations were 6.6 ± 1.0 and 5.6 ± 1.5%, respectively. The intra- and interobserver CV values associated with the stereological measurement of all LV volumetric and functional parameters from MDCT data sets are presented in Table 2.

Table 2 Intraobserver and interobserver coefficient of variation (CV) values associated with the stereological estimation of left-ventricular volumetric and functional parameters from MDCT data sets

Comparison of stereological method with standard CT analysis

An excellent correlation was found between the stereological method and standard CT analysis based on the semi-automatic tracing of endocardial borders (Fig. 2). The spread of the differences between the two methods is presented in Fig. 3. The mean difference between EDV, ESV, SV, and EF values derived by standard CT analysis and the stereological method was 4.8 ± 7.7 ml, 2.9 ± 6.4 ml, 0.8 ± 6.9 ml, and −1.7 ± 3.4%, respectively. There was no statistical difference between the two methods (EDV: P = 0.057; ESV: P = 0.071; EF: P = 0.061; SV: P = 0.878). The mean duration for EDV and ESV measurements as performed by an experienced radiologist in standard CT analysis was 4.2 ± 1.2 min.

Fig. 2
figure 2

Linear regression analysis of a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by the stereological method and standard CT analysis

Fig. 3
figure 3

Bland-Altman scatter plots showing the difference in a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by standard CT analysis and the stereological method against the mean value of each parameter. The solid and dotted lines denote the mean difference and 95% limits of agreement, respectively

Comparison of standard CT analysis with MRI

Left-ventricular volumetric and functional parameters obtained by standard CT analysis correlated well with the respective MRI estimations (Fig. 4). Bland-Altman statistical method for EDV, ESV, SV, and EF displayed a mean difference of −17.7 ± 13.9 ml, −11.7 ± 8.1 ml, −7.0 ± 14.0 ml, and 3.6 ± 6.2%, respectively (Fig. 5). Statistically significant differences were found for EDV and ESV (EDV: P = 0.0008; ESV: P = 0.0004). EF and SV estimations obtained by standard analysis of MDCT short-axis reformations and MRI did not differ significantly (EF: P = 0.051; SV: P = 0.064).

Fig. 4
figure 4

Linear regression analysis of a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by standard CT analysis and MRI

Fig. 5
figure 5

Bland-Altman scatter plots showing the difference in a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by MRI and standard CT analysis against the mean value of each parameter. The solid and dotted lines denote the mean difference and 95% limits of agreement, respectively

Comparison of stereological and MRI estimations

Stereological estimations of all volumetric and functional parameters from MDCT data sets and MRI-based measurements are summarized in Table 3. An excellent correlation was found between stereological and MRI estimations (Fig. 6). The Bland-Altman scatter plots are shown in Fig. 7. For each plot, no more than one data point was outside the 95% limits of agreement. Based on Bland-Altman analysis, the mean difference between MRI and the stereological method for estimating EDV, ESV, SV, and EF was −15.0 ± 13.2 ml, −8.8 ± 7.7 ml, −6.2 ± 12.0 ml, and 2.5 ± 5.6%, respectively. For each parameter, the SD of stereological and MRI estimations versus the mean of these estimations is presented in Fig. 8. No significant correlation was found except for SV (EDV: R = −0.09, P = 0.73; ESV: R = −0.25, P = 0.33; EF: R = −0.36, P = 0.15; SV: R = 0.56, P = 0.02), where R is the correlation coefficient. The Wilcoxon test showed that stereological estimations on short-axis image reformations significantly overestimated the LV volumes (EDV: P = 0.0011; ESV: P = 0.0013). For EF and SV, there was no statistically significant difference between the stereological method and MRI (EF: P = 0.067; SV: P = 0.064).

Fig. 6
figure 6

Linear regression analysis of a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by the stereological method and MRI

Fig. 7
figure 7

Bland-Altman scatter plots showing the difference in a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by MRI and the stereological method against the mean value of each parameter. The solid and dotted lines denote the mean difference and 95% limits of agreement, respectively

Fig. 8
figure 8

Plots showing the standard deviation (SD) of a end-diastolic volume (EDV), b end-systolic volume (ESV), c stroke volume (SV), and d ejection fraction (EF) estimations obtained by the stereological method and MRI against the mean value of each parameter

Table 3 Mean values (±SD) and range of the left-ventricular volumetric and functional parameters estimated by the stereological method and cine MRI

Discussion

In the current study, the stereological method was used to determine the LV volumetric and functional parameters from MDCT data sets. The stereological method can be optimized either by altering the point spacing of the square test grid or by systematically sampling short-axis images. The CE of the LV volumes derived by the systematical sampling procedure exceeded the target precision of 5%. To obtain an acceptable level of precision, stereological volume estimations should be carried out on all images depicting the LV cavity during the end-diastolic and end-systolic phases. Analytical measurements revealed that the optimum point spacing for EDV and ESV estimation is 25 and 15 pixels, respectively. These separation distances between test points of the grid can be employed in all stereological measurements irrespective of the LV size. This may be attributed to the population of this study including consecutive patients with a wide variation in cardiac volumes. The use of optimum point spacing results in quick and acceptable LV volume estimations with a CE below 5%.

The optimized stereological method was found to be reproducible, as reflected by the relatively small intra- and interobserver variability values. The interobserver variability for the stereological estimation of all volumetric and functional parameters was similar to the values of previous studies that performed manual or semi-automatic segmentation of endocardial contours on MDCT image reformations. They reported that the interobserver variability for the measurement of EDV, ESV, SV, and EF was 2.0–7.2, 4.0–9.5, 3.0–9.0, and 2.0–8.7%, respectively [10, 14, 3538]. To our knowledge, only one study of Sugeng et al. [37] evaluated the intraobserver variability in EDV, ESV, and EF measurements from MDCT data. Their CV values of 2.0–2.2% are comparable with those presented here.

Statistical analysis revealed that the stereological method and standard CT approach based on the semi-automatic segmentation of endocardial contours in diastolic and systolic images are in good agreement. Both volumetric methods applied on MDCT short-axis reformations overestimated EDV and ESV in comparison with the reference standard cine MRI. Several previous studies have concluded that LV volume measurements from MDCT agree well with MRI [13, 17, 35, 38]. Van der Vleuten et al. [39] reported that MDCT volume measurements are interchangeable with those obtained by MRI. However, other studies have reported similar findings to ours [10, 31, 37, 40]. The overestimation of the LV volume during the end-systolic phase should be attributed to the limited temporal resolution of MDCT. The 16-MDCT system used in the current study provides a temporal resolution of approximately 210 ms whereas that of MRI is reduced to 40 ms. A temporal resolution of 30–50 ms is necessary to define the peak systolic constriction and depict the minimum systolic LV volume [3]. The limited temporal resolution of MDCT can not lead to increased EDV values compared with MRI. The significant EDV overestimation might be due to intermodality differences related to the depiction of LV in a series of short-axis images. Regarding the SV and EF, the differences between stereological and MRI estimations were found to be not statistically significant. Therefore, the minor underestimation of EF and overestimation of SV by the stereological method should not be considered as clinically important.

The proposed method for LV volumetric and functional analysis from MDCT data sets has several limitations. One limitation is patient exposure to ionizing radiation during CT. Sixteen-slice MDCT coronary angiography on the same scanner with that employed in our study results in a patient effective dose range from 9 to 13 mSv [15, 17, 18]. The use of ECG controlled tube current modulation can considerably reduce the above radiation doses [41]. Moreover, a relatively small number of consecutive patients were examined, and the majority of them presented a normal LV function. The optimized stereological measurements, as described here, can not provide information about the ventricular mass. Further research is required to adapt the stereological method and MDCT data for LV mass determination.

The efficiency of stereological volumetric estimations on short-axis reformations is strongly affected by the temporal resolution of the 16-detector-row MDCT scanner employed. Recent technological developments in CT systems resulting in reduced gantry rotation times down to 330 ms may considerably improve the temporal resolution. Flohr et al. [42] reported that dual-source CT provides a temporal resolution as low as 83 ms, independent of the patient’s heart rate. The new generation of MDCT systems permitting the simultaneous acquisition of 64 submillimeter slices can yield a temporal resolution of 42–165 ms [43]. However, the better temporal resolution of 64-MDCT scanners is associated with a higher patient dose than that observed during 16-slice cardiac angiography [41, 44, 45].

In conclusion, the optimized stereological method allows the quick determination of EDV and ESV from MDCT short-axis reformations. The good agreement of stereological SV and EF estimations with the respective cine MRI calculations gives strong evidence about the accuracy of the above parameters. The analysis of LV function based on stereological volume estimations obtained from MDCT data sets is reproducible.