Introduction

Since the introduction of stent-assisted coiling, it was hypothesized that stents, besides preventing coil protrusion, could prevent perfusion of the aneurysm to some extent leading to the concept of endovascular flow diversion. Early studies could validate this concept in in vitro models [1,2,3]. Very recently this hypothesis was supported by several clinical observations when sole stenting was performed for treatment of intracranial aneurysms with braided stents [4, 5] and even large cell laser-cut stents [6]. Otherwise, because of high porosity of the first-generation laser-cut intracranial stents, in vivo evidence of aneurysm occlusion was not targeted and clinical application as a flow diverting device was limited [7].

Rapid technical progress led to the development of dedicated flow diverters (FDs) with a braided structure and considerably lower porosity than common stents, which have become available for endoluminal vessel reconstruction without the use of coils and were extensively evaluated in in vivo aneurysm models [8, 9]. However, besides the manufacturing differences in the devices, high material density is a potential determinant for thromboembolic complications and side branch occlusion [10,11,12,13].

Braiding technology at the same time allowed for manufacturing of stents for coil-assisting procedure with a lower porosity than traditional laser-cut stents, even if considerably higher than in FDs. The purpose of this in vivo study was to evaluate the stand-alone occlusive effect of a fine-meshed braided stent with porosity in the lower range of currently available (non-flow-diverter) microstents in an elastase aneurysm model. Besides the aneurysm occlusion rates, the degree and extent of neointima formation, the patency rate of parent vessel and covered branch arteries were investigated.

Materials and Methods

The investigated self-expanded stent (SX stent, Acandis, Pforzheim, Germany) is a prototype of the Accero® intracranial stent which obtained CE marking for the European Union in 2015 and aimed at coil-assisting technique. The prototype differs from the final Accero stent by different positions of the stent markers and slight different wire loop designs at both distal and proximal ends.

The braided stent consists of a braided structure with a single 50-µm-thick nitinol wire including a radiopaque platinum core and radiopaque platinum sleeves distributed in the middle and at both ends of the stent. With ten cells on the circumference, corresponding to 20 wires crossing the stent cross section, and an unconstrained diameter of 4.5 mm, the stent presents a variable porosity and pore density depending on the expansion diameter. If pore density is defined as the reciprocal of the ideal surface of the diamond limited by four wires and comprising the half of the wire width, the pore density amounts to 1.03/0.80/0.84 and 1.13 pores/mm2 at an unconstrained expanded configuration and in vessels with 4, 3 and 2 mm diameter, respectively. For the same configurations, porosity calculated as ratio between the free surface of the diamond and the surface comprising the half width of the wires corresponds to 84.3%/87.7%/87.1% and 82.7%. At both stent ends, close loops are flared with the aim of an improved vessel apposition. Delivery is performed within a microcatheter of 0.0165″ inner diameter. Foreshortening in a vessel with diameter of 2 mm, 3 mm and 4 mm is 7%, 18% and 36%, respectively.

Animal Experiment

All the experiments were approved by the local animal protection committee (no. 44/13) and conducted in accordance with the animal experiment guidelines. One wide-necked aneurysm with a mean dome-to neck-ratio of 0.92 ± 0.23 was created in each of ten New Zealand white rabbits according the method reported by Altes et al. [14]. Implantation of the SX stent was performed at least 21 days after aneurysm induction [15]. Three days prior to stent placement all animals were treated with aspirin (10 mg/kg orally) and clopidogrel (10 mg/kg orally) daily until the final experiment. The efficacy of dual platelet aggregation inhibition was not tested. Under general anaesthesia, the right femoral artery was surgically exposed and a 4-Fr sheath was inserted. The 2Fr microcatheter Neuroslider 21(Acandis, Pforzheim, Germany) with inner lumen 0.0165″ was placed into the aortic arch. Throughout the procedure, 300 U heparin was intravenously administered. For evaluation of the induced aneurysm, digital subtraction angiography (DSA) was performed. With a standard microguidewire, the Neuroslider 21 microcatheter was manoeuvered beyond the aneurysm neck into the right subclavian artery. Then, the SX stent was unsheathed under road-map guidance. This achieved through slow withdrawal of the delivery microcatheter while, at the same time, holding or gently pushing the delivery wire to achieve complete expansion and wall apposition of the SX stent for 2–3 min. The distal end of the stent was positioned distally to the right vertebral artery, and the proximal part of the stent was deployed within the brachiocephalic trunk. For adequate positioning, the middle radiopaque sleeve of the stent was held on the level of the aneurysm neck during the deployment. After removal of the transport wire, an angiogram of the brachiocephalic trunk was performed. The microcatheter was then retracted and placed in the abdominal aorta for DSA. An additional SX stent was placed in the aorta across the origin of lumbar arteries (Fig. 1C, D). Subsequent DSA was performed to ensure patency of the covered vessels. Endovascular materials were removed, and the femoral artery was ligated.

Fig. 1
figure 1

DSA immediately after stent implantation in the subclavian artery (A) and after 3 months follow-up (B) of the same animal shows inadequate occlusion of the aneurysm cavity resulting in grade I occlusion (arrow). DSA of the abdominal artery pre-stent implantation (C) and unsubtracted angiogram 6 months follow-up. Proximal and distal stent markers are highlighted with arrowheads (D). Redrawn margins of the contrast-filled aneurysm sac (E) on posterior–anterior projection on 2D angiogram for calculation of the aneurysm area (exemplarily case, animal no. 9)

Subjects were followed for 3 months (n = 5) and 6 months (n = 5). Under general anaesthesia, the left carotid artery was surgically exposed and a 4-Fr sheath was introduced in a retrograde manner with the purpose of better contrast (compared to intravenous injection), and since both femoral arteries were previously sacrificed following the procedure for aneurysm formation and stent implantation, respectively. DSA series were performed as follow-up angiogram, and the animals were sacrificed with a lethal injection of pentobarbital. After harvesting the subclavian artery with the aneurysm and parts of the abdominal aorta, the specimens were fixed in 10% solution of buffered formalin.

Data Analyses

Angiographic Evaluation

Three observers analysed angiograms using a five-point grading scheme to evaluate the occlusion of saccular aneurysms in consensus according to Kamran et al. [16] blinded to histology results. Grade 0: no reduction in aneurysmal flow; grade I: less than 50% and grade II: more than 50% flow reduction; grade III: almost complete occlusion; and grade IV: complete occlusion. In addition, the areas of the contrast-filled aneurysms were calculated as follows. The margins of the contrast-filled aneurysms were redrawn on posterior–anterior projection on 2D pre-interventional and follow-up angiograms (3 or 6 months) using the pencil tool of OsiriX 6.5.2 (Fig. 1E). Then, the areas of the aneurysms were converted from pixel2 to mm2 using a reference object (1 Cent coin), and the reduction in the areas is given as percentage.

Moreover, evaluation of parent vessel thrombosis, stenosis and side branch patency was performed.

Micro-CT

Micro-CT was performed for assessment of stent integrity and wire fractures. All devices were scanned by high-resolution CT (SkyScan 1172, Belgium). The X-ray tube operated at 75 kV, 10 W and 134 μA, with an aluminium filter and a focal spot size of 5 μm using a pixel size of 13 μm. Volume data were reconstructed with NRecon, volumetric software (1.6.2.0.) and postprocessed with OsiriX 6.5.2.

Histopathological Processing and Analyses

For each subclavian artery and abdominal aorta, transversal cuts were taken at three different regions (proximal, middle and central) for a total of 60 slices (n = 30 for each follow-up time, respectively). After microscopy (VHX-500F; Keyence; Neu-Isenburg, Germany) and digitization, an experienced pathologist (BD) performed histopathological evaluation, including stent apposition, neointima thickness, inflammation and potential foreign body reaction surrounding the stent struts and the presence of thrombotic material.

Morphometric measurements were performed as follows. Thickness of the neointima = distance between the outer surface of each strut and the luminal border. Neolumen = distance from luminal border to the opposite luminal border. Former vessel lumen = distance from the outside of a strut to the opposite outside of the strut across the vessel diameter. Diameter stenosis = (Neolumen ÷ Former lumen) × 100. For each parameter, maximum and minimum values were measured and average was calculated.

Statistical Analyses

Statistical analysis was performed using SPSS 20.0. Continuous variables are expressed as means and standard deviations (SD). Differences between the groups were compared with Student’s t-test (or Welch’s t-test when variance was inhomogeneous). The statistical significance of changes from baseline within each group was tested with paired t-test. Statistical significance was accepted at two-sided p ≤ .05.

Results

Angiography

Aneurysm characteristics and dimensions of harbouring arteries are given in Table 1.

Table 1 Aneurysm and harbouring artery characteristics

In all cases, the delivery and deployment of the devices were successful. No case of stent migration or inadequate wall apposition was observed.

The mean diameter of the stented aorta was 3.8 mm mm ± 0.49 mm in the 3 months’ group and 3.5 mm ± 0.33 mm in the 6 months group.

The mean area of the aneurysms in the 3 months group was 17.31 ± 7.48 mm2 before treatment and 13.45 ± 6.33 mm2 after treatment (reduction 22.5%, non-significant, p = .06). The mean area of the aneurysms in the 6 months group was 25.4 ± 23.87 mm2 before treatment and 15.37 ± 12.85 mm2 after treatment (reduction 42.1%, significant p = .02) (Fig. 2A, B). Overall occlusion rates of grades I and II were noted in 9 (90%) and 1 (10%) of ten aneurysms, respectively. In the 3 months group, grade I occlusion rate was noted in all five cases (Fig. 1A, B). In the 6 months group, grade I occlusion rate was noted in four cases (80%) and grade II occlusion in one case (20%).

Fig. 2
figure 2

A and B Area of the aneurysms measured on posterior–anterior projection of DSA, baseline (initial aneurysm area before stent implantation) and after 3 months (A) and 6 months (B) follow-up. Results of neointima thickness of the stented aorta and subclavian artery (defined as aneurysm) after 3 months (C) and 6 months (E). D and F show their corresponding stenosis after each time point

All branch arteries (e.g. vertebral, renal or lumbar arteries) covered by the SX stent remained patent at follow-up. Mean diameter of vertebral, renal and lumbar artery origin was 1.3 ± 0.1, 2.3 ± 0.3 mm and 1.2 ± 0.2 mm, respectively. Angiographically, no stenosis in parent vessels or stent thrombosis was assessed.

Micro-CT

All stents were intact in both subclavian and aorta. No wire fractures or distortion was assessed. A light bulging of the stents in the central part was observed in all aorta specimens, while in the subclavian artery, a tapering with continuous diameter reduction extended from the proximal to the distal part, according to the vessel geometry. A representative picture of both specimens in the same animal is shown in Fig. 3A, B.

Fig. 3
figure 3

CT scans of the Accero stent prototype explanted from the abdominal aorta (A) and subclavian artery (B) from the same animal after 3 months. Photomicrograph (H&E) of neointima (arrows) being thinner in the abdominal aorta (C) than in the subclavian artery (D)

Histology

Histologically, in all analysed tissue sections stent material was observed at the interface between the media and the neointima leading to complete wall apposition in all cases (Fig. 3C, D).

The neointima thickness of the stent harbouring vessels was significantly higher in the subclavian artery than in the abdominal aorta at follow-up 3 months (119.3 ± 59.0 µm vs 53.0 ± 7.0 µm, Fig. 2C). This difference was not significant after 6 months (92.0 ± 40.2 µm vs 49.2 ± 10.3, Fig. 2E). Stenosis grade was significantly higher in the subclavian artery compared with abdominal aorta at follow-up 3 months (p = .03); after 6 months there was no significant difference between aorta and subclavian artery (p = .6) (Fig. 2D, F).

In the regions of the aneurysm neck, incomplete neointima formation with scattered neointima islands was observed over the stent wires. Inflammation and foreign body reaction were seen in 1 of 30 slices, both at 6 months follow-up (3.3%). Calcification islands were observed in 4 and 11 slices at 3 and 6 months (13.3% and 36.7%), (Fig. 4).

Fig. 4
figure 4

Photomicrograph (H&E) of subclavian artery. Neointima islands (hash) at the aneurysm neck at 3 months follow-up (A). Beginning calcification (starlet) (B) and foreign body giant cell (arrow) (C) after 6 months follow-up

Discussion

In this study, we report the in vivo evaluation of a fine-meshed braided stent as a sole device in the treatment of wide-neck aneurysms without coiling. In comparison with common FDs like Pipeline embolization device (PED, Medtronic, Irvine, California, USA), Silk flow diverter (SFD, Balt Extrusion, Montmorency, France) or Derivo (Acandis, Pforzheim, Germany), the investigated SX stent is characterized by a higher porosity.

Since use of FDs can be associated with thromboembolic events and might cause delayed brain parenchymal lesions [17, 18], recent clinical studies underline the potential role of high-porosity implant, i.e. for small diameter vessels and vessels with high perforator density. In the study of Matsuda et al. [19], low-neck metal coverage correlated to less perforator coverage in a comparison between a stent and a flow diverter in a cadaver. Möhlenbruch et al. [20] presented the results of the implantation of a 0.021″ compatible FD with porosity of 70% in a multicentre clinical study, showing an aneurysm occlusion rate amounting to 100% at 12 months despite a 12% rate of thrombus related events.

Besides FD treatment, the use of multiple stents in telescoping technique is another option for aneurysm treatment. Xu et al. [21] showed the feasibility of this procedure in a retrospective clinical study. Aneurysm occlusion rate was achieved in 80% in the case series. But this technique holds the risk of thromboembolic events and in-stent stenosis due to the increased material and overlaying struts.

The main idea of our investigation was to assess the flow remodelling performance and vessel reaction of the Accero stent prototype with less material coverage and deliverability within a 0.0165″ microcatheter in an experimental in vivo model. Due to the decrease in the metal-to-artery ratio compared to FDs, the SX stent could potentially reduce thromboembolic complications.

Some in vitro and numerical investigations were previously performed to assess intra-aneurysmal fluid dynamical effect of implants with different porosities, including intracranial stents and flow diverters [22, 23]. Although a certain flow modification was proven also for high-porosity stents, biological factors triggering intra-aneurysmal clot formation and vessel reconstruction are obviously neglected in these models.

In both follow-up 3 and 6 months, we observed an incomplete aneurysm occlusion with a reduction in aneurysm filling in both groups, being significant after 6 months follow-up compared to baseline. The progressive occlusion can be explained by the increase in induced scattered neointima islands covering the device struts across the aneurysm neck (Fig. 4). These results support the findings of several clinical studies using low-profile braided stents as a stent monotherapy in aneurysm treatment. In Cagnazzo et al. [5], a 70% complete aneurysm treated with a braided stent was reported. In Kim et al. [24], implantation of laser-cut stents led to a complete occlusion of 37.5% of aneurysms treated.

Previous investigations with state-of-the-art FDs in similar rabbit models with comparable aneurysm dimensions achieved high aneurysm occlusion rates. Kallmes et al. [25] observed complete occlusion in 40% and 80% of aneurysms at 3 and 6 months follow-up using implants with 70% porosity. Sadasivan et al. [26] compared three devices with porosity between 65 and 70%, achieving higher occlusion rate for implants with higher pore density. In Ley et al [27], two devices with porosity 65% and 80% showed a significantly higher rate of complete occlusion at a lower porosity.

Also Wang et al. [28] showed a positive correlation between aneurysm occlusion rate and metal coverage measured by micro-CT.

The investigated device presents porosity in the lower range of currently available intracranial stents. According to the calculation presented before, a mean subclavian artery diameter of 2 mm to 3 mm leads to porosity between around 82.8% and 87.1% in our study. For comparison, Cho et al. [29] investigated the porosity of different intracranial stents at 0.5 mm oversizing, amounting to 86–88.5% for braided, 90% for an open-cell laser cut and 95% for a close-cell laser-cut stent.

The higher porosity compared to FDs is presumably responsible for a weaker or rather slower intra-aneurysmal flow reduction and thus occlusion rate. A previous in vitro study showed a relationship between flow diverter porosity and intra-aneurysmal thrombus formation, hypothesizing (1) a material-triggered thrombus initiation and (2) a thrombus growing dependent on intra-aneurysmal stagnation points [30]. However, in our study the area of residual aneurysm filling was progressively decreasing and significantly lower at 6 months follow-up compared to baseline, indicating the potential for a slow aneurysm occlusion. Longer follow-up results would be needed to evaluate the further progression of the occlusion.

The histological evaluation revealed a slight foreign body reaction without significant inflammation. The focal calcification observed in some slides was also reported in previous studies [31, 32].

The higher neointima thickness and diameter stenosis in the subclavian artery compared to the stented aorta is in agreement with previous studies [8, 33] and is presumably caused by the increased metal coverage and different mechanical behaviour due to the device oversizing in small vessel diameter. Furthermore, the devices in the subclavian arteries are exposed to a higher shear stress due to the movement of the legs of the animals.

In Kallmes et al., implantation of flow diverter in the subclavian artery resulted in a neointima and stenosis grade ranging between 100 and 200 ± 100 µm and between 17 and 23% at 6 months follow-up. In Sadavisan et al., neointima thickness was between 129 and 175 µm depending on the implanted flow diverter. In Ley et al., neointima thickness was between 125 and 138 µm in different subclavian regions, with the stenosis grade amounting to 7.3% to 8.3% at 6 months follow-up. Considering the similarity of the animal models, the lower neointima thickness and stenosis rate in the investigated stent can be presumably ascribed to the lower metal density and thus lower vessel stress.

Our preclinical study had some limitations. We did not include a control group of untreated aneurysms or a group of aneurysms treated with commercial available FDs, since experience with this animal model at our department has proven patency of untreated aneurysms and adequate occlusion rates after 6 months with flow diverting implants [27, 32]. The study population was small. The induced aneurysms in a rabbit model differ from spontaneously grown intracranial aneurysms in humans regarding morphological properties such as size and neck diameter as pointed out by Kallmes [25]. Further, the induced aneurysms are more similar to a side-wall aneurysm. Flow diversion properties of the tested stent may differ in the treatment of bifurcation aneurysms due to hemodynamic variations.

Conclusion

The Accero stent is biocompatible. The implantation of the Accero stent alone, without coil occlusion, does not allow a reliable occlusion of the target aneurysms. Using the same model (elastase aneurysms in rabbits), a comparison of coiling alone versus Accero-assisted coiling would allow a better understanding of the potential benefit of said stent, beyond the mechanical coil retention.