Abstract
Hip dislocation is a major and common complication of total hip arthroplasty (THA), which appears with an incidence between 0.3% and 10% in primary total hip arthroplasties and up to 28% in revision THA. The hip dislocations can be classified into three groups: early, intermediate and late. Approximately two-thirds of cases can be treated successfully with a non-operative approach. The rest require further surgical intervention. The prerequisite to developing an appropriate treatment strategy is a thorough evaluation to identify the causes of the dislocation. In addition, many factors that contribute to THA dislocation are related to the surgical technique, mainly including component orientation, femoral head diameter, restoration of femoral offset and leg length, cam impingement and condition of the soft tissues. The diagnosis of a dislocated hip is relatively easy because the clinical situation is very typical. Having identified a dislocated hip, the first step is to perform a closed reduction of the implant. After reduction you must perform a computed tomography scan to evaluate the surgical options for treatment of recurrent dislocation that include: revision arthroplasty, modular components exchange, dual-mobility cups, large femoral heads, constrained cups, elimination of impingement and soft tissue procedures. The objective is to avoid further dislocation, a devastating event which is increasing the number of operations on the hip. To obtain this goal is useful to follow an algorithm of treatment, but the best treatment remains prevention.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Total hip arthroplasty (THA) is one of the most successful orthopaedic procedures and is highly effective in relieving pain and improving function [1, 2]. Unfortunately, dislocation remains one of the common complications of THA. As published by Bozic et al. [2] and Sanchez-Sotelo and Berry [3], and according to the national registers of Sweden, Australia and United States [4–6], dislocation is the prime reason for early revision THA. In the national register of Italy [7], recurrent dislocation is the cause of revision THA in 9.3% overall, after aseptic loosening and infection. In the regional register of Emilia-Romagna [8], dislocation is the cause of revision in 9%, but in the first two years this percentage increased to 26%. Furthermore, dislocation is the first cause of multiple revision (22.5%).
The risk of dislocation is influenced by the surgical approach, the underlying diagnosis, the surgical technique, the lifetime of the prosthesis and the patient’s compliance with restrictions [9–12]. An improved understanding of the aetiology of dislocation and refinements in surgical techniques have led to a decrease in the rate of dislocation over time.
Although most dislocations after THA are single episodes that can be managed non-operatively [10, 13], some patients require surgical intervention to address recurrent dislocation. The choice of surgical technique to manage recurrent dislocation depends on the aetiology of the problem [14].
Revision arthroplasty for the treatment of recurrent dislocation is more likely to be successful when the cause for the dislocation has been identified [15, 16]. In addition, the timing of the onset of the dislocation influences the decision concerning treatment, especially with regard to operative intervention [17, 18].
Component malpositioning and abductor insufficiency are two of the most important recognised causes of recurrent dislocation [10, 19–21]. When malpositioning is the cause, revision of the component is the most effective type of surgical intervention [15, 21, 22]. However, when the aetiology of the dislocation is multifactorial or unknown, the best surgical technique with which to address it is often less obvious and dual mobility or constrained cup are the valid options.
Cause and classification
Prevalence of dislocation has been reported as being between 0.3 and 10% in primary THAs and up to 28% in revision THAs [16]. Approximately two-thirds of cases can be treated successfully with a non-operative approach. The remainder require further surgical intervention [15, 16].
Counting subluxation episodes in addition to dislocations would produce even higher rates. Subluxation is difficult to diagnose and therefore frequently overlooked, a fact that hinders an exact evaluation of its incidence rate, which has been estimated at 2–5.5% [20].
Dislocation rates vary with many factors, including the surgical approach and femoral head size, follow-up duration and many other factors such as female gender, advanced age, specific causes (avascular necrosis of the femoral head, proximal femoral fracture or non-union), obesity, comorbidities with an ASA score of 3 or more, neuromuscular impairments (Parkinson’s disease or stroke-related impairments), neurocognitive impairments (psychiatric disease or mental disability) and alcohol abuse. Finally, the dislocation rate is higher in patients with a history of surgery on the same hip, most notably previous THA procedures [23, 24].
Although most dislocations occur early after THA, the dislocation rate increases with follow-up duration with a cumulative risk of 1–1.39% each five years [25, 26].
Three categories can be defined based on time to occurrence [15, 16]:
-
Early dislocations within the first three (or 6) months after THA are the most common (50-70% ) and are accelerated by inadequate healing.
-
Intermediate dislocations occur after the resumption of previous activities, between three to six months and five years after THA, in relation to increased hip mobilisation; this category contributes 15-20% of all THA dislocations.
-
Late dislocations, occurring more than five years after THA, are often related to polyethylene wear; their rate may be underestimated and may reach 32%, with a mean time to occurrence of 11.3 years.
Evaluation of the unstable THA
The prerequisite to developing an appropriate treatment strategy is a thorough evaluation to identify the causes of the dislocation. The direction of the dislocation should be assessed based on the causative movement, femoral head position on the lateral view (which unfortunately is often not obtained) and, above all, direction of the instability as determined during reduction.
In addition to the above-listed risk factors, many factors that contribute to THA dislocation are related to the surgical technique, including the following: surgical approach, component orientation, femoral head diameter, restoration of femoral offset and leg length, cam impingements and condition of the soft tissues.
Surgical approach
The postero-lateral approach is associated with a higher dislocation rate (mean, 6.9%) compared to the antero-lateral approach (3.1%) and anterior approach (0.6-1.3%) [27].
Component orientation
Suboptimal component position often results in early or secondary dislocation but may be difficult to demonstrate.
Acetabular component
Lewinnek et al. [19] defined a safe zone of 40 ± 10° of inclination and 15 ± 10° of anteversion. However, many patients have cups outside this safe zone yet do not experience dislocation [28, 29]. A cup that is too vertical and/or anteverted increases the risk of anterior dislocation, whereas a cup that is too horizontal and/or insufficiently anteverted increases the risk of posterior dislocation. Inclination can be reliably measured on a standard radiograph, whereas anteversion cannot. Although numerous radiograph and computed tomography (CT) protocols have been developed, the values often differ between two measurements. Several groups [30, 31] have therefore developed CT measurement protocols that take into account the degree of pelvic tilt measured on a standing lateral radiograph and involve measurements of anteversion in the lying, sitting and standing positions. This point is important, as the orientation of the pelvis and, therefore, of the cup, changes with body position.
Navigation may optimise component positioning, producing substantially lower proportions of cups outside the safe zone compared to conventional surgery, as demonstrated in a prospective study reported by Parratte and Argenson [32].
Femoral component
Anteversion of the femoral component is easier to determine, by measuring the angle between the prosthetic neck axis and the line tangent to the posterior femoral condyles on a CT scan. This value often differs from the surgeon’s estimate, by a mean of 16.8° according to a study by Dorr et al. [33].
Also, for the femoral component, navigation may optimise positioning, improving management of limb-length discrepancy and off-set restoring versus freehand techniques [34].
Femoral head diameter
Femoral head diameter influences the stability of the prosthetic joint. In a study by Berry et al. [27], femoral heads measuring 22 mm in diameter were associated with higher dislocation rates; however, the posterior approach was used in most patients. In a study of 2,020 THAs performed via the antero-lateral approach with heads measuring 36 mm or more in diameter, the authors [35] identified a single case of dislocation (0.05%) after a mean follow-up of 31 months. In a prospective randomised controlled trial to compare 28-mm to 36-mm heads implanted via the postero-lateral approach, the dislocation rate was five times higher with the smaller heads [36]. A larger head diameter (36 mm or more) increases the head/neck ratio and delays contact between the neck and the cup; in addition, the ‘jumping’ distance is increased, allowing for a greater range of ‘subluxation’ before complete dislocation occurs [37–39]. It has also been demonstrated that a larger metallic head may produce greater fretting damage owing to an increased head-neck moment arm with a risk of trunnionosis [40]. Moreover, for a larger ceramic head it is necessary to use HXPLE to reduce wear and liner breakage.
Restoration of femoral offset
This point is crucial to hip stability, mobility, and optimal abductor muscle efficiency. Femoral offset measured on radiographs varies with femoral anteversion. Merle et al. [41] reported underestimation of femoral offset on antero-posterior pelvic radiographs, with improved accuracy on antero-posterior hip radiographs. Computer-assisted techniques may improve accuracy in restoration of the native offset [34].
Restoration of leg length
Leg-length restoration also improves stability, by maintaining muscle tension. A clinical and radiological evaluation of leg length, although mandatory, is associated with numerous pitfalls and errors, most notably in patients with a controlateral prosthetic hip [42].
Cam impingements
The principles regarding impingement in the natural hip put forth by Ganz et al. [43] are similar in concept to what can occur in the prosthetic hip. To understand impingement, it is helpful to recognise the common mechanisms that cause mechanical abutment in both anatomic and prosthetic hips. In the anatomic hip joint, impingement is a mechanical abutment conflict between the bone of the femur and the pelvis; in a total hip replacement, it is contact between the metal femoral neck and the cup liner or bone-to-bone contact such as between the greater trochanter and the pelvis [39].
The femoral head-neck ratio, which is the relationship between the diameter of the femoral head and the diameter of the femoral neck, influences impingement. Cam impingement is caused by a reduced femoral head-neck ratio. An example is the pistol-grip deformity that is created by a decreased offset of the femoral head-neck junction [43]. Cam impingement in a prosthetic hip is caused by any implant feature that reduces the head-neck ratio. A skirt on the metal femoral head or a large circular femoral neck can cause mechanical abutment in a prosthetic hip through this mechanism [39, 44].
Pincer impingement in the anatomic hip is a mechanical abutment caused by acetabular retroversion, protrusio, or coxa profunda. Pincer impingement in the prosthetic hip is caused by hooded and constrained liners or by placement of a small femoral head in a big acetabular cup [45, 46]. Failure to remove acetabular osteophytes so that the metal neck or the femoral bone abuts on the osteophytes is another cause of pincer impingement.
Because impingement is a dynamic process, it has been difficult to identify it and to define its prevalence on the basis of clinical evaluations or plain radiographs. In the clinical setting, some causes of failure such as wear or dislocation can be related to impingement [10, 38], but a direct relationship with impingement has been difficult to document.
Condition of the soft tissues
Soft-tissue damage (failed healing of a posterior or antero-lateral approach) explains many dislocations but is difficult to demonstrate before revision surgery. Healing failure rates increase with the number of surgical procedures. Non-union of the greater trochanter results in slackness of the abductor muscles, thereby increasing the risk of dislocation, most notably when the greater trochanter is displaced upwards. Thus, THA instability can be related to a single cause, and the treatment is then relatively easy. In most cases, however, multiple factors are involved, which complicates the therapeutic management.
Surgical techniques
Surgical options for treatment of recurrent dislocation include:
-
Revision arthroplasty
-
Modular components exchange
-
Dual mobility cup
-
Large femoral head
-
Constrained cup
-
Elimination of impingement
-
Soft-tissue procedures
Revision arthroplasty
As mentioned previously, component malpositioning is one of the primary causes of recurrent dislocation, and component revision is mandatory to successfully treat this type of dislocation (Fig. 1) [10, 15, 16, 19, 21].
However, identifying the malpositioned component is not always straightforward. Plain radiographs provide limited information regarding the orientation of the acetabular and femoral components.
Computed tomography (CT) may be needed to more accurately assess component positioning (Fig. 2), especially with regard to version of the acetabulum [30, 31]. Although little component malpositioning is difficult to detect on plain radiographs, the radiographs of a dislocated hip should be scrutinised carefully. In addition to showing the direction of the dislocation, these radiographs can convey other important information. Other critical parameters, such as abductor strength and the overall neurological status of the patient, can be gleaned from clinical examination. Limb-length discrepancy detected on clinical or radiographic examination can be an important finding as it can be associated with component malpositioning [42], because intra-operative instability, which may have been a result of a suboptimally positioned component such as a retroverted socket, could have been addressed by lengthening of the femoral neck to increase the soft-tissue tension.
Correction of the malpositioned component can simultaneously correct the limb-length inequality (Fig. 1c). The ultimate and most accurate information regarding component positioning is obtained from intra-operative inspection of the components during revision surgery. When poor position is demonstrated, changing the component is a logical procedure that has produced highly variable success rates.
Carter et al. [47] obtained a higher success rate of 86% but often increased the head diameter (36 mm), in addition to changing the cup. Changing a poorly positioned femoral component is desirable but considerably more invasive than cup revision; in the study by Carter et al., this procedure was performed in only 11.5% of 156 hips. Always in this study, changing only the liner may carry an even higher failure rate, of 34-55%, as a result of unrecognised suboptimal cup position.
If the component malposition is confirmed, it should be revised to address the recurrent dislocation. An exception may be made for any frail, elderly or infirm patient, in whom a slightly malpositioned component may be accepted to prevent prolonged surgery due to revision of a well-fixed acetabular or femoral component. In these patients, a constrained liner may be utilised in an effort to prevent recurrent dislocation.
Modular components exchange
Several studies have demonstrated success with the use of modular component exchange for correction of recurrent dislocation after THA [48, 49]. This surgical treatment involves exchanging the acetabular liner and the femoral head with the main intention being to “upsize” the femoral head and/or use an elevated liner. This treatment can be successful only if the patient has well-positioned and well-fixed acetabular and femoral components. In addition, the acetabular component in place must be sufficiently large to allow an adequate thickness of polyethylene (a minimum of 4 mm) to be used with the larger femoral head.
Toomey et al. [48] described a series of 13 patients treated with exchange of the femoral head and/or acetabular liner. One patient was lost to follow-up, and only one of the remaining 12 patients had recurrent dislocation at a mean of 5.8 years; thus, this surgical treatment had a success rate of 92%, with less extensive morbidity. However, the authors recommended that modular component exchange may be used in only selected cases and that each patient be evaluated thoroughly to identify all factors contributing to the dislocation (Figs. 3 and 4). Additionally, adequate intra-operative stability must be achieved. Despite the success reported by Toomey, modular components can be problematic [50]. In a report on complications related to 20 hip replacements, Barrack et al. [51] described 15 complications that were related to failure of the modular interface. Complications were attributed to detachment of the femoral head from the trunnion, dislodgment of the polyethylene liner from the shell and asymmetrical rotation of the polyethylene liner. Thus, modular component exchange requires meticulous surgical technique and should be reserved for specific cases.
Cups with elevated rims decrease the dislocation rate [52, 53], they also place considerable mechanical stresses on the cup and can result in cam impingement. McConway et al. [54] reported a 1.6% recurrence rate in 302 cases after a mean follow-up of nearly seven years and with less than 2% of radiolucent lines accentuation.
Dual-mobility cup
The principle of the dual-mobility cup was developed by Gilles Bousquet [55], then used by many different manufacturers. A 22- or 28-mm head is held captive in a polyethylene cup, which is mobile within a cemented or press-fit metal cup made of steel or cobalt-chromium. Either standard or highly reticulated polyethylene is used, and the head can be made of ceramic. Dual-mobility cups provide an increased range of movement and may reduce the risk of dislocation [55–60]. Outcomes after the implantation of dual-mobility cups for THA instability have been evaluated in numerous studies, of which the earliest was conducted by Leclercq et al. [58]. After 2.5 years of follow-up, 13 hips with recurrent dislocation were stabilised without correction of the causes of the instability. Other authors [59] studied 59 cases and found a 1.7% recurrence rate and a 98% eight year survival rate. In 54 cases, Guyen et al. [60] reported a 5.5% recurrence rate after a mean follow-up of four years and in 47 cases. Prudhon et al. [56] studied 79 cases of revision with two different types of cementless dual-mobility cup. In this series, 21 cases were treated for recurrent dislocation with no re-dislocation at three years of follow-up.
Dual-mobility cups are now very widely used to prevent instability after revision THA, also in Italy. Civinini et al. [61] prospectively followed 33 patients who underwent isolated acetabular revisions with a minimum of two years’ follow-up. There were no dislocations. Survivorship rates of acetabular components were 97% at five years; the re-revision rate for any reason was 3%. No patients had progressive osteolysis, component migration or loosening on radiographs. Their data suggest that the use of a dual-mobility cup reduced the risk of dislocation without increasing loosening from two to five years. In a prospective study of 2107 revision THAs including 62% with dual-mobility cups, the authors [62] found a three month dislocation rate of only 4%. Enthusiasm about these very good results must be tempered by knowledge of the fixation problems seen with press-fit dual-mobility cups. Massin et al. [63] found a lower eight-year survival rate of press-fit grit-blasted cups made of hydroxyapatite-coated steel, compared to the original tripod cup and to cobalt-chromium cups (91% versus 98% and 100%, respectively). However, the fixation problems reported in this series are related to use of the first generation of dual-mobility cup.
Finally, the deeper concavity of dual-mobility cups is associated with greater wear [35], which carries a risk of intra-prosthetic dislocation (IPD) due to locking-system failure. Mean time to IPD is 10 years. Besides, the risk of this complication can be minimised by avoiding neck-cup impingement and is extremely low with the newest-generation implants [37, 63, 64]. Nevertheless, even with the original tripod cup, 15-year survival rates were greater than 85% in the long-term studies [65, 66], each of which included more than 400 cases.
Large femoral head
Large-diameter heads are associated with decreased dislocation rates [27, 35–37] and are therefore useful for the treatment of THA instability [67–69]. Stability increases with the diameter of the femoral head up to 38 mm; with larger diameters, variable results have been reported [67–69]. In a study by Amstuz et al. [68] of 29 THA revisions for dislocation managed with head diameters of 36 mm or more, the recurrence rate was 13.7%. Suboptimal cup orientation was noted in the six hips with persistent instability; after correction of cup position, none of the hips was unstable after a mean follow-up of 5.5 years. Dislocation rates after revision THA for reasons other than instability with 36- or 40-mm heads was lower than 32-mm heads after a mean follow-up of five years [69]. Most of the available studies focused on the impact of large head diameter on primary THA stability and wear debris volume. However, there are various other friction-causing combinations for which the medium- and long-term outcomes are unclear. Now metal-on-metal implants are not used due to the high risk of pseudo-tumour formation [70].
Constrained cup
The constrained acetabular liner is an invaluable tool in the armamentarium for surgical treatment of recurrent dislocations. Its success has been widely demonstrated [71–74]. This device is especially suited for the treatment of recurrent dislocation secondary to soft tissue (abductor) deficiency. It is also an excellent option for patients with recurrent dislocation of unknown aetiology, elderly patients in whom the components are well fixed and patients with neurological impairment. In other words, constrained liners are used as a salvage treatment option in the most difficult subset of cases.
Anderson et al. [71] was the first author to describe the use of constrained liners in patients with recurrent dislocation.
He reported a success rate of 72% in a study of 18 patients followed for a mean of 31 months (range, 24-64 months) after the use of this device.
The only factor predictive of failure was an increased acetabular abduction angle of the metallic acetabular cup. At the time of follow-up, there was no radiographic or clinical evidence of loosening of the acetabular component. One of the main advantages of a constrained liner is its ability to provide stability without the need to revise a well-fixed and well-positioned acetabular component. Callaghan et al. [75] reported the clinical and radiographic outcomes of 31 revision THAs in which a constrained liner had been cemented into a well-fixed cementless acetabular shell. At an average of 3.9 years posto-peratively, 29 constrained liners (94%) remained securely fixed in the cementless shell and only two liners had failed. One of the failed liners had separated from the cement, and the other had failed as a result of fracture of the capturing mechanism. Each hip was successfully revised with another cemented constrained liner. No acetabular component showed radiographic evidence of progressive loosening or associated osteolysis. The authors drew attention to the importance of proper preparation of the liner, correct sizing of the component, and the use of optimal cementing technique. Their meticulous surgical technique may explain the good results despite the suboptimal outcomes of this technique in other studies [74, 76].
Noble et al. [76] examined 57 constrained cups of four different designs recovered during THA revision surgery performed 36 months on average after the primary procedure. Locking ring failure explained 51% of revisions and cup loosening a further 28%.
Cooke et al. [74] identified three types of early failure of a constrained acetabular implant. Eight patients (14%) required a re-operation because of failure of the constrained liner, with seven of them having had recurrent dislocations. Failures were described on the basis of the mechanism of implant failure. There were three Type-I failures (of the bone-prosthesis interface), two Type-II failures (of the liner locking mechanism) and one Type-III failure (of the femoral head locking mechanism).
In conclusion, hip arthroplasty with use of a constrained acetabular cup should be considered to be an option for the surgical treatment of patients with extensive soft-tissue deficiency, deficiency of the abductor mechanism, dislocation with no discrete or identifiable cause, recurrent dislocation despite prior attempted surgical correction, also in case of failure of a dual-mobility cup, and finally in old and frail patients with well-fixed metallic shell.
Patients should be selected carefully for treatment with this technique, only after detailed examination for the cause of the dislocation and after it has been determined that other interventions are unlikely to be successful. Prior to insertion of a constrained liner, it is crucial that the positions of the femoral and acetabular component be scrutinised.
Elimination of impingement
Because impingement is a dynamic process, it has been difficult to identify it and to define its prevalence on the basis of clinical evaluations or plain radiographs. In the clinical setting, some causes of failure, such as wear or dislocation, can be related to impingement [10, 38], but a direct relationship with impingement has been difficult to document. To avoid the impingement is the goal during the primary implant: elimination of skirts on femoral heads, chamfering of the polyethylene rim, narrow femoral necks, large femoral heads, and perhaps modular femoral necks, are the tricks.
However, when the cause of impingement is the presence of an osteophyte detected on the CT scan is possible to remove it. Furthermore, liner exchange and larger femoral head is an option to eliminate impingement when the components are correctly positioned.
Soft tissue procedures
Other, currently less-used surgical procedures for addressing recurrent dislocation include reinforcement of the soft tissues, the abductor mechanism, around the hip [15, 16], trochanteric advancement and trochanteric lowering [77, 78]. The main problem associated with these procedures is the variability in outcome. These procedures can also be technically demanding and are likely to fail if used in patients with component malpositioning. Therefore, soft-tissue reinforcement and trochanteric advancement are being used with less and less frequency, and only in cases in which component position has absolutely been determined to be acceptable.
Often these procedures are consistently used in combination with the above-described techniques.
Discussion
Dislocation after THA is a common complication. As published by Bozic et al. [2] and Sanchez-Sotelo and Berry [3], and according to national register of Sweden, Australia and United States [4–6], dislocation is the prime reason for early revision THA. Also, in the regional register of Emilia-Romagna [8], a region of Italy, dislocation is the cause of revision in 26% in the first two years. Furthermore dislocation is the first cause of multiple revision (22.5%).
Surgical treatment for THA instability is usually considered after the second or even the third dislocation episode in patients without obvious component malposition, failed closed reduction, or significant greater trochanter displacement.
In a patient who is younger than 70 years of age and has no risk factors, the cup should be changed and properly positioned. If allowed by the implants, the head diameter should be increased. In cases of isolated wear, a change of liner, with a preference for an elevated lip design but with the risk of controlateral dislocation by cam effect deserves consideration. In patients who are older than 70 years and/or have risk factors for instability (multiple surgical procedures or neurological impairments for instance), a dual-mobility cup seems the best solution.
A constrained cup may be considered, particularly in the oldest patients. Furthermore, in these old patients, when they have a well-fixed cementless cup, a polyethylene cup can be cemented in the appropriate position into the shell. The five year survival rate with this method was 78% [79].
In older frail patients with acetabular loosening and stable Charnley femoral stem, Civinini et al. [61 purpose isolated acetabular revision with assembly of the dual-mobility cup in situ on the non-modular head of the stem].
With a cementless cup, the liner can be exchanged for an elevated lip liner and the head diameter can be increased if allowed by the implants. With a cemented cup, a lip augmentation device can be considered. In young patients, however, a change of cup with an increase in head diameter seems preferable, also with the use of a dual-mobility cup, helpful in multifactorial instability, in compromisation of soft tissue and in multiple revisions.
In the event of inadequate femoral offset or excessive anteversion or retroversion, replacement of the femoral component should be considered. A lateralised component or a component with a greater degree of varus is usually required to restore offset. A less invasive means of restoring offset consists of changing the length of the neck and/or the diameter of the head (if the cup liner can be changed); care should be taken, however, to avoid excessive lengthening of the lower limb. When the primary THA includes a removable neck, this component alone can be changed, and its length and orientation can be modified. When changing a femoral component, the use of a removable neck design can be considered to optimise anteversion, offset and neck length. The stem should be changed if it is loose or unstable.
The soft tissues must always be repaired to the extent possible. In addition to the above-described procedures, suturing the fascia lata to the greater trochanter and vastus lateralis muscle with the hip in abduction is useful. A fracture or non-union of the greater trochanter should be treated. When no cause is identified, lowering of the greater trochanter can be considered but carries a risk of non-union.
Immobilisation can be helpful in patients with severe soft tissue alterations or with factors such as agitation or poor motor co-ordination [13].
Conclusions
Surgical management of recurrent dislocation following THA is a challenging problem.
Identification of the aetiology is critical for successful treatment. Close scrutiny of component position is a crucial step in the management of these patients.
Unrecognised component malposition should always be suspected. Revision of the malpositioned component is perhaps the most effective type of surgical intervention in the treatment of recurrent dislocation. During revision surgery, it is imperative that the components be carefully inspected to determine if they are optimally positioned. When revision surgery is planned, the necessary equipment should always be available for revision of the malpositioned component.
When dislocation is multifactorial or idiopathic, the potential surgical options include modular component exchange, dual mobility cup, use of a larger femoral head and lipped liner, and insertion of a constrained acetabular component.
Soft-tissue reinforcement and trochanteric advancement have variable and less successful results and should be used in only carefully selected cases.
The objective is to avoid further dislocation, a devastating event which is increasing with the number of surgeries on the hip. To obtain this goal is useful to follow an algorithm of treatment (Fig. 5), but the best treatment remains prevention.
References
Chang RW, Pellisier JM, Hazen GB (1996) A cost effectiveness analysis of total hip arthroplasty for osteoarthritis of the hip. JAMA 275:858–865
Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ (2009) The epidemilogy of revision total hip arthroplasty in the United States. J Bone Joint Surg Am 91:128–133
Sanchez-Sotelo J, Berry DJ (2001) Epidemiology of instability after total hip replacement. Orthop Clin N Am 32(4):543–552
American Joint Replacement Registry (2014) Annual Report. Avaible via http://www.ajrr.net/images/annual_reports/AJRR_2014_Annual_Report_final_11-11-15.pdf
Swedish Hip Arthroplasty Register. Annual Report (2011). Avaible via http://www.shpr.se/en/default.aspx
Australian Orthopaedic Association National Joint Replacement Registry. Annual Report (2012). Avaible via http://aoanjrr.dmac.adelaide.edu.au/fr
Torre M, Bellino S, Luzi I, Ceccarelli S, Salvatori G, Balducci MT, Piffer S, Zanoli G, Romanini E, Boniforti F, Carrani F (2016) Italian Register Arthroplasty Project. Third Report. Il Pensiero Scientifico Editore. Avaible via http://www.iss.it/binary/riap2/cont/3_Report_RIAPcompleto_2016.pdf
Report of R.I.P.O, Regional Register of Orthopaedic Prosthetic Implantology. Annual report (2014). Available via https://ripo.cineca.it/pdf/RIPO_REPORT_2015_english_rev1.pdf
Morrey BF (1992) Instability after total hip arthroplasty. Orthop Clin N Am 23:237–248
Dorr LD, Wan Z (1998) Causes of and treatment protocol for instability of total hip replacement. Clin Orthop Relat Res 355:144–151
Hedlundh U, Karlsson M, Ringsberg K, Besjakov J, Fredin H (1999) Muscular and neurologic function in patients with recurrent dislocation after total hip arthroplasty: a matched controlled study of 65 patients using dual-energy X-ray absorptiometry and postural stability tests. J Arthroplasty 14:319–325
Soong M, Rubash HE, Macaulay W (2004) Dislocation after total hip arthroplasty. J Am Acad Orthop Surg 12:314–321
Dewal H, Maurer SL, Tsai P, Su E, Hiebert R, Di Cesare PE (2004) Efficacy of abduction bracing in the management of total hip arthroplasty dislocation. J Arthroplasty 19:733–738
Bourne RB, Mehin R (2004) The dislocating hip: what to do, what to do. J Arthroplasty 19(Suppl 1):111–114
Charissoux JL, Asloum Y, Marcheix PS (2014) Surgical management of recurrent dislocation after total hip arthroplasty. Orthop Traumatol Surg Res 100(1 Suppl):25–34
Parvizi J, Picinic E, Sharkey PF (2008) Revision total hip arthroplasty for instability: surgical techniques and principles. J Bone Joint Surg Am 90(5):1134–1142
Conroy JL, Whitehouse SL, Graves SE, Pratt NL, Ryan P, Crawford RW (2008) Risk factor for revision for early dislocation in total hip arthoplasty. J Arthroplasty 23(6):867–872
Von Knoch M, Berry DJ, Harmsen WS, Morrey BF (2002) Late dislocation after total hip arthroplasty. J Bone Joint Surg Am 84:1949–1953
Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am 60:217–220
Ritter MA (1976) Dislocation and subluxation of the total hip replacement. Clin Orthop Relat Res 21:92–94
Parvizi J, Kim KI, Goldberg G, Mallo G, Hozack WJ (2006) Recurrent instability after total hip arthroplasty: beware of subtle component malpositioning. Clin Orthop Relat Res 447:60–65
Daly PJ, Morrey BF (1992) Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am 74:1334–1343
Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty. A multivariate analysis. J Arthroplasty 17:282–288
Kim YH, Choi Y, Kim JS (2009) Influence of patient-, design-, and surgery-related factors on rate of dislocation after primary cementless total hip arthroplasty. J Arthroplasty 24:1258–1263
Berry DJ, Von Knoch M, Schleck CD, Harmsen WS (2004) The cumulative long-term risk of dislocation after primary Charnley total hip arthroplasty. J Bone Joint Surg Am 86:9–14
Caton J, Aslanian T, Prudhon JL, Ferreira A, Descamps L, Dehri G, Puch JM (2016) Dual Mobility Cup: a new THA revolution. E-Mémories de l’Académie Nationale de Chirurgie 15(1):004–010
Berry DJ, Von Knoch M, Schleck CD, Harmsen WS (2005) Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am 87:2456–2463
Berend KR, Sporer SM, Sierra RJ, Glassman AH, Morris MJ (2010) Achieving stability and lower-limb length in total hip arthroplasty. J Bone Joint Surg Am 92:2737–2752
Rittmeister M, Callitsis C (2006) Factors influencing cup orientation in 500 consecutive total hip replacements. Clin Orthop Relat Res 445:192–196
Ala Eddine T, Migaud H, Chantelot C, Cotton A, Fontaine C, Duquennoy A (2001) Variations of pelvic anteversion in the lying and standing positions: analysis of 24 control subjects and implications for CT measurement of position of a prosthetic cup. Surg Radiol Anat 23:105–112
Puri L, Lapinski B, Wixson RL, Lynch J, Hendrix R, Stulberg SD (2006) Computed tomographic follow-up evaluation of operative intervention for periacetabular lysis. J Arthroplasty 21(6 Suppl 2):78–82
Parratte S, Argenson JN (2007) Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am 89:494–498
Dorr LD, Malik A, Dastane M, Wan Z (2009) Combined anteversion technique for total hip arthroplasty. Clin Orthop Relat Res 467:119–127
Confalonieri N, Manzotti A, Montironi F, Pullen C (2008) Leg length discrepancy, dislocation rate and offset restoration in total hip replacement using a short modular stem: navigation vs conventional freehand. Orthopedic 31(10 Suppl 1). pii: orthosupersite.com/view.asp?rID=35541
Lombardi AV, Skeels MD, Berend KR, Adams JB, Franchi OJ (2011) Do large heads enhance stability and restore native anatomy in primary total hip arthroplasty? Clin Orthop Relat Res 469:1547–1553
Howie DW, Holubowycz OT, Midelton R (2012) Large femoral heads decrease the incidence of dislocation after total hip arthroplasty. A randomised controlled trial. J Bone Joint Surg Am 94:1095–1102
Triclot P, Gouin F, Richter D, Musset T, Bonnan J, Ollivier H et al (2011) Update ‘big-head’: the solution to the problem of hip implant dislocation? Orthop Traumatol Surg Res 97S:120–127
Malik A, Maheshawari A, Dorr LD (2007) Impingement with total hip replacement. J Bone Joint Surg Am 89:1832–1842
Cooper HJ, Della Valle CJ (2014) Large diameter femoral heads: is bigger always better? Bone Joint J 96B (11 Suppl. A):23–26
Del Balso C, Teeter MG, Tan SC, Howard JL, Lanting BA (2015) Trunnionosis: does head size affect fretting and corrosion in total hip arthroplasty? J Arthroplasty 31(10):2332–2336
Merle C, Waldstein W, Pegg E, Streit MR, Gotterbarm T, Aldinger PR et al (2012) Femoral offset is underestimated on anteroposterior radiographs of the pelvis but accurately assessed on anteroposterior radiographs of the hip. J Bone Joint Surg (Br) 94:477–482
Parvizi J, Sharkey PF, Bissett GA, Rothman RH, Hozack WJ (2003) Surgical treatment of limb-length discrepancy following total hip arthroplasty. J Bone Joint Surg Am 85:2310–2317
Ganz R, Parvizi J, Beck M, Leunig M, Notzli H, Siebenrock KA (2003) Femoroacetabular impingement: a cause for osteoarthritis of the hip. Clin Orthop Relat Res 417:112–120
Barrack RL, Butler RA, Laster DR, Andrews P (2001) Stem design and dislocation after revision total hip arthroplasty: clinical results and computer modeling. J Arthroplasty 16(8 Suppl 1):8–12
Earll M, Fehring T, Griffin WL, Mason JB, McCoy TH (2004) Early osteolysis associated with trunion-liner impingement. Clin Orthop Relat Res 418:153–156
Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM (1998) Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res 355:163–170
Carter AH, Sheehan EC, Mortazavi J, Purtill JJ, Sharkey PF, Parvizi J (2011) Revision for recurrent instability. What are the predictors of failure? J Arthroplasty 26:46–52
Toomey SD, Hopper RH Jr, McAuley JP, Engh CA (2001) Modular component exchange for treatment of recurrent dislocation of a total hip replacement in selected patients. J Bone Joint Surg Am 83:1529–1533
Earll MD, Fehring TK, Griffin WL, Mason JB, McCoy T, Odum S (2002) Success rate of modular component exchange for the treatment of an unstable total hip arthroplasty. J Arthroplasty 17:864–869
Blom AW, Astle L, Loveridge J, Learmonth ID (2005) Revision of an acetabular liner has a high risk of dislocation. J Bone Joint Surg (Br) 87(12):1636–1638
Barrack RL, Burke DW, Cook SD, Skinner HB, Harris WH (1993) Complications related to modularity of total hip components. J Bone Joint Surg (Br) 75:688–692
Morrey BF (2004) Results of reoperation for hip dislocation: the big picture. Clin Orthop Relat Res 429:94–101
Insull PJ, Cobbett H, Frampton CM, Munro JT (2014) The use of a lipped acetabular liner decreases the rate of revision for instability after total hip replacement: a study using data from the New Zealand Joint Registry. Bone Joint J 96-B(7):884–888
McConway J, O’Brien S, Doran E, Archbold P, Beverland D (2007) The use of a posterior lip augmentation device for a revision of recurrent dislocation after primary cemented Charnley/Charnley Elite total hip replacement. Results at a mean follow-up of six years and nine months. J Bone Joint Surg (Br) 89:1581–1585
Bousquet G, Argenson C, Godeneche JL, Cisterne JP, Gazielly DF, Girardin P, Debiesse JL (1986) Recovery after aseptic loosening of cemented total hip arthroplasties with Bousquest’s cementless prosthesis. Apropos of 136 cases. Rev Chir Orthop Reparatrice Appar Mot 72(Suppl 2):70–74
Prudhon JL, Steffan F, Ferreira A, Verdier R, Aslanian T, Caton J (2014) Cementless dual mobility cup in total hip arthroplasty revision. Int Orthop 38:2463–2468
Caton JH, Prudhon JL, Ferreira A, Aslanian T, Verdier R (2014) A comparative and retrospective study of three hundred and twenty primary Charnley type hip replacements with a minimum follow up of ten years to assess wether a dual mobility cup has a decresead dislocation risk. Int Orthop 38:1125–1129
Leclercq S, El Bidi S, Aubriot JH (1995) Bousquet’s prosthesis for recurrent total hip prosthesis dislocation. Rev Chir Orthop 81:389–394
Leiber-Wackenheim F, Brunschweiler B, Ehlinger M, Gabrion A, Mertl P (2011) Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 9 years’ follow-up. Orthop Traumatol Surg Res 97:8–14
Guyen O, Pibarot V, Vaz G, Chevillotte C, Béjui-Hugues J (2009) Use of dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res 467:465–472
Civinini R, Carulli C, Matassi F, Nistri L, Innocenti M (2012) A dual-mobility cup reduces risk of dislocation in isolated acetabular revisions. Clin Orthop Relat Res 470(12):3542–3548
Delaunay C, Hamadouche M, Girard J, Duhamel A (2013) What are the causes for failures of primary hip arthroplasties in France? Clin Orthop Relat Res 471(12):3863–3869
Massin PH, Orain V, Philippot R, Farizon F, Fessy MH (2012) Fixation failures of dual mobility cups. A mid-term study of 2601 hip replacements. Clin Orthop Relat Res 470:1932–1940
Fabry C, Langlois J, Hamadouche M, Bader R (2016) Intra-prostethic dislocation of dual-mobility cups after total hip arthroplasty: potentialcauses from a clinical and biomechanical perspective. Int Orthop 40(5):901–990
Philippot R, Farizon F, Camilleri JP, Boyer B, Derhi G, Bonnan J et al (2008) Survival of dual mobility socket with a mean 17 years follow-up. Rev Chir Orthop 94:43–48
Lautridou C, Lebel B, Burdin G, Vielpeau C (2008) Survival of the cementless Bousquet dual mobility cup: minimum 15-year follow-up of 437 total hip arthroplasties. Rev Chir Orthop Reparatrice Appar Mot 94:731–739
Rodriguez JA, Rathod PA (2012) Management factorials in THA. Large diameter heads. Is bigger always better? J Bone Joint Surg Br 94 Suppl A:52–54
Amstutz HC, Le Duff MJ, Beaulé PE (2004) Prevention and treatment of dislocation after total hip replacement using large diameter balls. Clin Orthop Relat Res 429:108–116
Garbuz DS, Masri BA, Duncan CP, Greidanus NV, Bohm ER, Petrak MJ et al (2012) Dislocation in revision THA. Do large heads (36 and 40 mm) result in reduced dislocation rates in a randomised clinical trial? Clin Orthop Relat Res 470:351–356
Bosker BH, Ettema HB, Boomsma MF, Kollen BJ, Maas M, Verheyen CC (2012) High incidence of pseudotumour formation after large-diameter metal-metal total hip replacement. A prospective cohort study. J Bone Joint Surg (Br) 94:755–761
Anderson MJ, Murray WR, Skinner HB (1994) Constrained acetabular components. J Arthroplasty 9:17–23
Goetz DD, Bremner BRB, Callaghan JJ, Capello WN, Johnston RC (2004) Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A concise follow-up of a previous report. J Bone Joint Surg Am 86:2419–2423
Shapiro GS, Weiland DE, Markel DC, Padgett DE, Sculco TP, Pellicci PM (2003) The use of a constrained acetabular component for recurrent dislocation. J Arthroplasty 18:250–258
Cooke CC, Hozack W, Lavernia C, Sharkey P, Shastri S, Rothman RH (2003) Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty. J Arthroplasty 18:827–833
Callaghan JJ, Parvizi J, Novak CC, Bremner B, Shrader W, Lewallen DG, Johnston RC, Goetz DD (2004) A constrained liner cemented into a secure cementless acetabular shell. J Bone Joint Surg Am 86:2206–2211
Noble PC, Durrani SK, Usrey MM, Mathis KB, Bardakos NV (2012) Constrained cups appear incapable of meeting the demands of revision THA. Clin Orthop Relat Res 470:1907–1916
Ekelund A (1993) Trochanteric osteotomy for recurrent dislocation of total hip arthroplasty. J Arthroplasty 8:629–632
Kaplan SJ, Thomas WH, Poss R (1987) Trochanteric advancement for recurrent dislocation after total hip arthroplasty. J Arthroplasty 2:119–124
Beaulé PE, Ebramzadeh E, Le Duff M, Prasad R, Amstutz C (2004) Cementing a liner into a stable cementless acetabular shell: the double-socket technique. J Bone Joint Surg Am 86:929–934
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they did not receive any financial payments or other benefits from any commercial entity related to the subject of this article.
Ethical standards
All procedures were in accordance with the ethical standards of the institutional and/or national research committee.