Abstract
A simple instrument for measuring adductor pollicis tension in children has been devised. The thumb sensor utilizes a thin conical rubber seal obtained from a 5, 10 or 20 ml Becton Dickinson syringe. A syringe barrel of similar size, fixed at right angles to the sensor, forms a handle for the fist. The sensor is linked to a pressure transducer with noncompliant tubing. Pressure output of the salinefilled system was found to be linearly related to the force exerted on each size of the sensor. In 20 children, aged two to eight years, the device, fitted in turn with the 5,10 and 20 ml syringe seal sensors, gave train-of-four (TOF) ratios which showed close correlation (coefficient r = 0.96, 0.93, 0.95) and agreement (regression slope b = 0.97, 0.95, 0.98) with those of a force transducer. Thus, inspection of the displayed TOF pressure pulses gives a rapid and reliable estimate of neuromuscular blockade.
Résumé
Void un dispositif simple pour mesurer chez les enfants la tension musculaire de l’adducteur du pouce. On détache l’obturateur conique du piston d’une seringue Becton Dickinson (5,10 ou 20 ml) et on le réinsère à l’envers dans le bardlet de la seringue. L’obturateur caoutchouté détecte les mouvements du pouce alors que le bardlet de la seringue sen de poignée. Avec une tubulure incompressible, on connecte le ditecteur à un transducteur de pression. Le système étant rempli de salin, la pression mesurée au transducteur est proportionnelle à la force exercée sur le détecteur. Chez 20 enfants de deux à huit ans, en utilisant des seringues de 5, 10 et 20 ml, nous avons observé pour les ratios du train-de-quatre ondes (TOF) une bonne corrélation entre notre dispositif et un transducteur de force (r = 0.96, 0.93 et 0.95 respectivement) avec une exactitude satisfaisante (pente b de la courbe de regression = 0.97, 0.95 et 0.98). Done, l’observation des courbes de pression produites par un TOF permet d’estimer adéquaiement le bloc neuromusculaire.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Donati F, Bevan JC, Bevan DR. Neuromuscular blocking drugs in anaesthesia. Can Anaesth Soc J 1984; 31:324–35.
Goudsouzian N. Muscle relaxants in children.In: Ryan J, Todres I, Cote C, Goudsouzian N (Eds.). A practice of anaesthesia for infants and children. Orlando: Grune and Stratton 1986: 105–14.
Viby Mogensen J, Jensen NH, Engbaek J, Ording H, Skovgaard LT, Chraemmar-Jorgensen B. Tactile and visual evaluation of the response to train of four nerve stimulation. Anesthesiology 1985; 63: 440–3.
Azar I, Wu WH, Turndorf H. “Twitch box”: a new device to monitor neuromuscular blockade. Med Instrum 1978; 12: 92–3.
Baraka A. Monitoring of neuromuscular transmission in anaesthetized man by a bulb transducer assembly. Anesth Analg 1973; 52: 36–8.
Armstrong JE, Goat VA, Loach AB. Measurement of neuromuscular blockade in man. A simple modification of a clinical pressure transducer. Anaesthesia 1977; 32: 480–2.
Gravenstein J, Paulus D. Arterial pressure.In: Gravenstein J, Paulus D (Eds.). Clinical monitoring practice. Philadelphia: Lippincott 1987: 43–104.
Viby Mogensen J. Clinical assessment of neuromuscular transmission. Br J Anaesth 1982; 54: 209–23.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Lyew, M.A., Pinto, S.R. & Bevan, J.C. A simple device for monitoring neuromuscular blockade in children. Can J Anaesth 36, 717–721 (1989). https://doi.org/10.1007/BF03005429
Issue Date:
DOI: https://doi.org/10.1007/BF03005429