Abstract
The feasibility of stenting the ductus arteriosus with a balloon-expandable vascular endoprosthesis was tested in 8 newborn lambs. Tantalum wire and stainless steel mesh coronary stents were implanted antegrade or retrograde by percutaneous transfemoral catheterization. One lamb died during the procedure from perforation of the aorta. In 7 lambs, the ductus arteriosus was crossed using endhole catheters and wires, and stents mounted on angioplasty catheters were expanded in the ductus arterisus. Six lambs had successful implantation and had maintained a sizeable patent ductus arteriosus at 2 h. We conclude that the feasibility of percutaneous stenting of the newborn ductus was demonstrated. By providing patency of the ductus arteriosus, stents may offer nonsurgical alternatives for palliation of cyanotic congenital heart disease and hypoplastic left heart syndrome
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Schatz RA (1989) A view of vascular stents (editorial). Circulation 79:445–457
Schatz RA, Palmaz JC, Tio FO, Garcia F, Garcia O, Reuter SR (1987) Balloon-expandable intracoronary stents in the adult dog. Circulation 76:450–457
Sigwart U, Puel J, Mirkovitch V, Joffre F, Kappenberger L (1987) Intravascular stents to prevent occlusion and restenosis after transluminal angioplasty. N Engl J Med 316:701–706
Levine MJ, Leonard BM, Burke JA, Nash ID, Safian RD, Diver DJ, Baim DS (1990) Clinical and angiographic results of balloon-expandable intracoronary stents in right coronary artery stenoses. J Am Coll Cardiol 16:332–339
Palmaz JC, Sibbitt RR, Reuter SR, Tio FO, Rice WJ (1985) Expandable intraluminal graft: A preliminary study. Radiology 156:73–77
Palmaz JC, Garcia OJ, Schatz RA, Rees CR, Roerren T, Richter GM, Noeldge G, Gardiner GA, Becker GJ, Walker C, Stagg J., Katzen BT, Dake MD, Paolini, RM, McLean GK, Lammer J, Schwarten DE, Tio FO, Root HD, Rogers W (1990) Placement of balloon-expandable intraluminal stents in iliac arteries: First 171 procedures. Radiology 174:969–975
Mullins CE, O'Laughlin MP, Vick GW, Mayer DC, Myers TJ, Kearney DL, Schatz RA, Palmaz JC (1988) Implantation of balloon-expandable intravascular grafts by catheterization in pulmonary arteries and systemic veins. Circulation 77:188–199
Ellis SG, Savage M, Baim D, Hirshfeld J, Cleman M, Teirstein P, Toppol EJ (1990) Intracoronary stenting to prevent restenosis: Preliminary results of a multicenter study using the Palmaz-Schatz stent suggest benefit in selected high risk patients. JACC 15:118A
Author information
Authors and Affiliations
Additional information
The opinions expressed are those of the authors and are not to be construed as official or as reflecting the views of the Army or the Department of Defense.
Rights and permissions
About this article
Cite this article
Moore, J.W., Kirby, W.C., Lovett, E.J. et al. Use of an intravascular endoprosthesis (stent) to establish and maintain short-term patency of the ductus arteriosus in newborn lambs. Cardiovasc Intervent Radiol 14, 299–301 (1991). https://doi.org/10.1007/BF02578454