Abstract
This paper presents an anthropomorphic prosthetic hand using flexure hinges, which is controlled by the surface electromyography (sEMG) signals from 2 electrodes only. The prosthetic hand has compact structure with 5 fingers and 4 Degree of Freedoms (DoFs) driven by 4 independent actuators. Helical springs are used as elastic joints and the joints of each finger are coupled by tendons. The myoelectric control system which can classify 8 prehensile hand gestures is built. Pattern recognition is employed where Mean Absolute Value (MAV), Variance (VAR), the fourth-order Autoregressive (AR) coefficient and Sample Entropy (SE) are chosen as the optimal feature set and Linear Discriminant Analysis (LDA) is utilized to reduce the dimension. A decision of hand gestures is generated by LDA classifier after the current projected feature set and the previous one are “pre-smoothed”, and then the final decision is obtained when the current decision and previous decisions are “post-smoothed” from the decisions flow. The prosthetic hand can perform prehensile postures for activities of daily living and carry objects under the control of EMG signals.
Article PDF
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
References
Asghari O M, Hu H. Myoelectric control systems: A survey. Biomedical Signal Processing and Control, 2007, 2, 275–294.
Chan C C, Liao W H. Temporal gait parameters captured by surface electromyography measurement. Proceedings of 2012 IEEE International Conference on Robotics and Biomimetics, Guangzhou, China, 2012, 1056–1061.
Cannan J, Hu H S. Human-Machine Interaction (HMI): A Survey, Technical Report CES-508, University of Essex, Colchester, UK, 2011.
Chen W R, Xiong C H. On adaptive grasp with underactuated anthropomorphic hands. Journal of Bionic Engineering, 2016, 13, 59–72.
Connolly C. Prosthetic hands from touch bionics. Industrial Robot: An International Journal, 2008, 35, 290–293.
Dalley S, Wiste T, Withrow T, Goldfarb M. Design of a multifunctional anthropomorphic prosthetic hand with extrinsic actuation. IEEE/ASME Transactions on Mechatronics, 2009, 14, 699–706.
Cipriani C, Controzzi M, Carrozza M. The smarthand transradial prosthesis. Journal of Neuroengineering and Rehabilitation, 2011, 8, 29–42.
Zecca M, Micera S, Carrozza M C, Dario P. Control of multifunctional prosthetic hands by processing the electromyographic signal. Critical Reviews in Biomedical Engineering, 2002, 30, 459–485.
Lotti F, Tiezzi P, Vassura G, Biagiotti L, Palli G, Melchiorri C. Development of UB hand 3: Early results. Proceedings of IEEE International Conference on Robotics and Automation, Barcelona, Spain, 2005, 4488–4493.
Huang H, Jiang L, Liu Y, Hou L, Cai H, Liu H. The mechanical design and experiments of hit/dlr prosthetic hand. Proceedings of IEEE International Conference on Robotics and Biomimetics, Kunming, China, 2006, 896–901.
Zollo L, Roccella S, Guglielmelli E, Carrozza M, Dario P. Biomechatronic design and control of an anthropomorphic artificial hand for prosthetic and robotic applications. IEEE/ASME Transactions on Mechatronics, 2007, 12, 418–429.
Gaiser I, Pylatiuk C, Schulz S, Kargov A, Oberle R, Werner T. The fluidhand iii: A multifunctional prosthetic hand. Journal of Prosthetics and Orthotics, 2009, 21, 91–96.
Xu Z J, Tian Y T, Li Y. sEMG pattern recognition of muscle force of upper arm for intelligent bionic limb control. Journal of Bionic Engineering, 2015, 12, 316–323.
Scheme E, Englehart K. Electromyogram pattern recognition for control of powered upper-limb prostheses: State of the art and challenges for clinical use. Journal of Rehabilitation Research and Development, 2011, 48, 643–659.
Kent B A, Lavery J, Engeberg E D. Anthropomorphic control of a dexterous artificial hand via task dependent temporally synchronized synergies. Journal of Bionic Engineering, 2014, 11, 236–248.
Lorrain T, Jiang N, Farina D. Influence of the training set on the accuracy of surface emg classification in dynamic contractions for the control of multifunction prostheses. Journal of Rehabilitation Research and Development, 2011, 8, 25.
Hudgins B, Parker P, Scott R N. A new strategy for multifunction myoelectric control. IEEE Transactions on Biomedical Engineering, 1993, 40, 82–94.
Kim K S, Choi H H, Moon C S, Mun C W. Comparison of k-nearest neighbor, quadratic discriminant and linear discriminant analysis in classification of electromyogram signals based on the wrist-motion directions. Current Applied Physics, 2011, 11, 740–745.
Phinyomark A, Phukpattaranont P, Limsakul C. Feature reduction and selection for EMG signal classification. Expert Systems with Applications, 2012, 39, 7420–7431.
Phinyomark A, Hirunviriya S, Limsakul C, Phukpattaranont P. Evaluation of EMG feature extraction for hand movement recognition based on euclidean distance and standard deviation. Proceedings of the 2010 ECTI International Conference on Electrical Engineering, Electronics, Computer, Telecommunications and Information Technology, Chiang Mai, Thailand, 2010, 856–860.
Englehart K, Hudgins B, Parker P A. A wavelet-based continuous classification scheme for multifunction myoelectric control. IEEE Transactions on Biomedical Engineering, 2001, 48, 302–311.
Park S H, Lee S P. Emg pattern recognition based on artificial intelligence techniques. IEEE Transactions on Rehabilitation Engineering, 1998, 6, 400–405.
Khushaba R N, Kodagoda S, Takruri M, Dissanayake G. Toward improved control of prosthetic fingers using surface electromyogram (EMG) signals. Expert Systems with Applications, 2012, 39, 10731–10738.
Chen X, Wang Z J. Pattern recognition of number gestures based on a wireless surface EMG system. Biomedical Signal Processing and Control, 2012, 8, 184–192.
Liu Y H, Huang H P, Weng C H. Recognition of electromyographic signals using cascaded kernel learning machine. IEEE/ASME Transactions on Mechatronics, 2007, 12, 253–264.
Jolliffe I. Principal Component Analysis, Wiley Online Library, USA, 2005.
Artemiadis P K, Kyriakopoulos K J. EMG-based control of a robot arm using low-dimensional embeddings. IEEE Transactions on Robotics, 2010, 26, 393–398.
Chu J U, Moon I, Kim S K, Mun M S. Control of multifunction myoelectric hand using a real-time EMG pattern recognition. Proceedings of 2005 IEEE/RSJ International Conference on Intelligent Robots and Systems, Edmonton, Canada, 2005, 3511–3516.
Hargrove L J, Li G, Englehart K B, Hudgins B S. Principal components analysis preprocessing for improved classification accuracies in pattern-recognition-based myoelectric control. IEEE Transactions on Biomedical Engineering, 2009, 56, 1407–1414.
Chu J U, Moon I, Lee Y J, Kim S K, Mun M S. A supervised feature-projection-based real-time EMG pattern recognition for multifunction myoelectric hand control. IEEE/ASME Transactions on Mechatronics, 2007, 12, 282–290.
Khushaba R N, Kodagoda S, Takruri M, Dissanayake G. Toward improved control of prosthetic fingers using surface electromyogram (EMG) signals. Expert Systems with Applications, 2012, 39, 10731–10738.
Wang N F, Chen Y L, Zhang X M. The recognition of multi-finger prehensile postures using LDA. Biomedical Signal Processing and Control, 2013, 8, 706–712.
Wang N F, Chen Y L, Zhang X M. Realtime recognition of multi-finger prehensile gestures. Biomedical Signal Processing and Control, 2015, 13, 262–269.
Taylor C L, Schwarz R J. The anatomy and mechanics of the human hand. Artificial Limbs, 1955, 2, 22–35.
Lu Z. Practical Electromyography, People’s Medical Publishing House, Beijing, China, 2000. (in Chinese)
Konrad P. The ABC of EMG: A Practical Introduction to Kinesiological Electromyography, 1st ed, Noraxon U.S.A. Inc., Scottsdale, USA, 2005.
Englehart K, Hudgins B. A robust, real-time control scheme for multifunction myoelectric control. IEEE Transactions on Biomedical Engineering, 2003, 50, 848–854.
Smith L H, Hargrove L J, Lock B A, Kuiken T A. Determining the optimal window length for pattern recognition-based myoelectric control: Balancing the competing effects of classification error and controller delay. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2011, 19, 186–192.
Richman J S, Moorman J R. Physiological time-series analysis using approximate entropy and sample entropy. American Journal of Physiology-heart and Circulatory Physiology, 2000, 278, H2039–H2049.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Wang, N., Lao, K. & Zhang, X. Design and Myoelectric Control of an Anthropomorphic Prosthetic Hand. J Bionic Eng 14, 47–59 (2017). https://doi.org/10.1016/S1672-6529(16)60377-3
Published:
Issue Date: