Abstract
Ion chromatography (IC) is widely used to quantify sulfate, nitrate, ammonium, sodium, chloride, and potassium from PM\(_{2.5}\) water extracts. IC hardware and software have progressed to allow a broader range of water-soluble compounds to be determined for the existing anion and cation programs and on the same solutions using analytical column, eluent, and detector modifications. Alkylamine, organic acid, and carbohydrate quantification by IC expands the number of source markers, especially for different types of biomass burning and secondary organic aerosols. Although modern systems are highly automated, internal quality control (QC) and external quality assurance (QA) programs are essential. QC includes detailed standard operating procedures, calibration over the range of expected concentrations, performance tests with independent standards, inspection of filters and chromatograms, and anion/cation balances. QA consists of independent system and performance audits, analysis of externally prepared performance samples, and interlaboratory comparisons. The additional water-soluble species provide compounds for speciated emission inventories, source markers to refine aerosol source apportionment, and increased understanding of global carbon, sulfur, and nitrogen cycles.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Measurements of PM\(_{2.5}\) (particulate matter with aerodynamic diameter less than 2.5 micrometers) elements, ions, and carbon fractions in chemical speciation networks are necessary, but not sufficient, for identifying and quantifying modern source contributions (Chow and Watson 2013; Watson and Chow 2013, 2015; Watson et al. 2016). Many elemental source markers are being eliminated from emission streams owing to their toxicity. As these primary PM emissions decrease, secondary organic aerosols (SOAs) are increasing in importance (Donahue 2013). Many of the SOA compounds absorb light at shorter (<600 nm) wavelengths, thereby affecting the Earth’s radiation balance as Brown Carbon (BrC) (Laskin et al. 2015). The broad diversity of biomass fuels and burning conditions (Chakrabarty 2016), especially in countries outside of North America and Europe, require more specific markers to better determine control strategies. Peat burning, a common Asian combustion source (Wang et al. 2004), is poorly studied, even though it may have an important effect on the global carbon (Turetsky et al. 2015) and nitrogen (Galloway et al. 2004) cycles.
Watson et al. (2016) identify several proven analytical methods that can be applied to samples already acquired in PM speciation networks (Dabek-Zlotorzynska 2011; IMPROVE 2016; U.S.EPA 20016a; Zhang et al. 2012) as illustrated in Fig. 1. Watson et al. (2016) also associate the measured species with potential sources. This article elaborates on further analysis of the water-extracted compounds using ion chromatography (IC). IC with conductivity detection was developed in the mid-1970s (Small et al. 1975) to obtain multipollutant measurements with a single analysis for water-soluble ions. Analytes include fluoride (F\(^{-})\), chloride (Cl\(^{-})\), nitrite (NO\(_{2}^{-})\), bromide (Br\(^{-})\), iodide (I\(^{-})\), nitrate (NO\(_{3}^{-})\), sulfate (SO\(_{4}^{2-})\), and phosphate (PO\(_{4}^{3-})\) for anions; and sodium (Na\(^{+})\), ammonium (NH\(_{4}^{+})\), magnesium (Mg\(^{2+})\), potassium (K\(^{+})\), and calcium (Ca\(^{2+})\) for cations. For PM\(_{2.5}\) deposits on filters, samples are first extracted in distilled-deionized water (DDW). IC separates the water-extracted solutes with an ion exchange resin that delays their passage owing to different ion sizes and charges relative to the ionic eluent. This results in different retention times for individual ion quantification with a detector, typically a conductivity detector that measures the electrical current traversing the eluted solution. Prior to detection, the ion exchange column effluent enters a suppressor (membrane or column) where the eluent is neutralized, reducing its conductivity while retaining that of the ions of interest. Ionic compounds are identified by their elution/retention times and quantified by the conductivity peak area. Peak areas are related to concentrations from standard solutions for quantification. Although IC instruments can be set up to calculate concentrations by either peak area or peak height, laboratory tests show that peak area provides better accuracy and precision than those for peak height. Peak area is not affected by baseline shifts which commonly occur in IC analysis.
While most applications involve sampling onto filters with extraction and analysis in the laboratory, a growing number of in situ IC-based instruments are being used to obtain real-time concentrations with high time resolution (ten Brink et al. 2007; URG 2016; Weber et al. 2001). Sub-diurnal time resolution can also be obtained using the recently developed Sequential Spot Sampler (Hecobian et al. 2016) that impacts PM\(_{2.5}\) into small wells on a carousel that can be subsequently extracted and analyzed with an automated laboratory IC.
Newer instrumentation and software have expanded the capability of IC to quantify selected organic compounds, several of which appear in the same spectra produced for inorganic ions and others that are detected using different columns, eluents, and detectors. This presentation updates a prior evaluation (Chow and Watson 1999) of IC methods applied to PM\(_{2.5}\) filter samples that are acquired in compliance and speciation networks. It provides a brief summary of recent reviews on liquid chromatography (LC) applications to environmental samples, then defines the key terms used in IC analyses. This work summarizes tests on water-soluble extraction efficiencies for particles collected on different filter media and describes procedures for calibration, standardization, and quality control/quality assurance (QC/QA). It compares chromatograms from different systems, demonstrating the progress of improved hardware and software to better resolve water-soluble components. It proposes instrumental configurations that can obtain more information from the sample extracts than is currently achieved by simple inorganic anion and cation analyses.
Table 1 lists the analytes that can be obtained from the IC methods described here and associates them with potential pollution sources. The major ionic components of Cl\(^{-}\), NO\(_{3}^{-}\), SO\(_{4}^{-2}\), Na\(^{+}\), and NH\(_{4}^{+}\) are essential for reconstructing PM mass (Chow et al. 2015) and estimating chemical light extinction to improve visibility (Cheng 2015; Pitchford et al. 2007; Watson 2002). Other ions, such as F\(^{-}\), Br\(^{-}\), I\(^{-}\), PO\(_{4}^{3-}\), Mg\(^{2+}\), K\(^{+}\), and Ca\(^{2+}\) which often go unreported, also have value as source markers. Water-soluble K\(^{+}\) is a common biomass burning marker (Calloway et al. 1989), while water-soluble Ca\(^{2+}\) is found in transported Asian dust (Duvall et al. 2006). The addition of alkylamines, organic acids, and carbohydrates quantification by IC expands the number of source markers for biomass burning; fuel combustion; biodiesel, diesel, and gasoline engine exhaust; meat cooking; marine aerosol; and bioaerosol. Although several sources are potential contributors, these can be narrowed down by the study context and network design (Chow et al. 2002). Agricultural sources such as fertilizer and animal husbandry contributions would not be expected at urban sites, and biomass burning would favor SOA formation during cold winter periods.
Recent Reviews of Liquid Chromatography (LC) Methods
Liquid chromatography (LC) includes both IC and high performance liquid chromatography (HPLC) that have wide application to analyses of compounds dissolved in water. IC is used for most inorganic constituents, whereas HPLC speciates organic molecular constituents (Michalski 2016). HP uses a high-pressure pump [\(\sim \)3000 to \(\sim \)6000 psi (pounds per square inch) or \(\sim \)21,000 to \(\sim \)41,000 kilopascals (kPa)] (Hartwell et al. 2013) to push the sample and eluent through the separating columns while lower pressures (<2,000 psi or \(\sim \)14,000 kPa) are applied in IC. The types of stationary phase matrices, column packing methods, eluent compositions, and detection modes vary between IC and HPLC. IC variants, such as ion-exclusion chromatography (IEC), ion-pair chromatography (IPC), and reversed-phase liquid chromatography (RPLC), have also been used to separate hydrophilic molecules, biomolecules, and metal complexes (Fritz 1991; Gama et al. 2012; Gennaro and Angelino 1997; Michalski 2014; Nakatani et al. 2012). In tandem with mass spectrometric (MS) detection, as implemented in HPLC–MS, LC–MS, and LC–MS/MS, more complex water-soluble organic compositions, molecular structures, molar weights, and functional groups (Fenn 2002, 2003; Michalski 2014) can be quantified at ultra-low detection limits.
Recent reviews (Butt and Riaz 2009; Hartwell et al. 2013; Karu et al. 2012; Lucy and Wahab 2013; Michalski 2013, 2014; Nakatani et al. 2012; Olariu et al. 2015; Pohl 2013; Zhang et al. 2015) highlight the technological advances in IC systems and applications for environmental, pharmaceutical, clinical, and food chemistry samples. The sensitivity, selectivity, and linearity of IC systems can be modulated based on the selection of stationary phase properties and eluent compositions. Although IC methods began with conductivity detection (Small et al. 1975), other advanced detection techniques such as UV–VIS or diode array absorbance, amperometry, and potentiometry have been used to quantify additional water-soluble compounds (Buchberger 2001).
There are several suppliers of IC hardware (Lachat 2016; Metrohm 2016; Thermo-Dionex 2016; Waters 2016), with Thermo/Dionex being the major one. Figure 2 illustrates a generic IC set-up, but there are countless combinations of columns, eluents, and detectors that can be selected from the vendors’ websites. The configurations described here are specified in Table 2. Basic components and terminology used in IC systems are documented in Table 3.
Filter Preparation, Receipt, Acceptance Testing, and Storage
Each batch of unexposed filters should be acceptance-tested with a visual inspection of every filter over a light table, with rejection of blank filters showing visual flaws, pinholes, or discoloration. One filter from each batch of 100 is extracted and analyzed for the intended species and the batch is rejected if excessive blank levels exceeding 0.5 \(\upmu \)g per filter are detected. Nylon-membrane filter (Pall 2016) acceptance testing is especially important as it passively absorbs nitric acid vapors when exposed to the atmosphere (Bytnerowicz et al. 2005). Acceptance-tested filters are then placed in clean Petri slides or Petri dishes with bar-coded IDs, then sorted and sealed into ziplock bags prior to being loaded into sealed filter holders for shipment to and from field sites. This sealing intends to minimize passive absorption or deposition onto the filters. It is good practice to store filters and extracts at \({<}4\,{^{\circ }}\)C in laboratory refrigerators with controlled access before ion analyses to minimize evaporation and contamination.
Filter Extraction
If available, the entire filter is immersed in 15 mL of DDW after placement in a polystyrene extraction tube that is labeled with the barcoded filter ID. If portions of the filter are to be used for other analyses, it can be sectioned into halves or quarters with one section used for extraction. Less than 15 ml of DDW can be used as the solvent, but it is important that the filter is completely immersed within the vial.
Extraction tubes (FisherScientific 2016) are capped, placed in extraction racks, and sonicated for 60 min in a temperature-controlled (\({<}25\,{^{\circ }}\)C) sonication bath. Sonicator bath water is recirculated to minimize temperature increases from the dissipation of ultrasonic energy. After sonication, the extraction tubes are mechanically shaken [at a speed setting on 20 revolutions per minute (rpm)] for 60 min. Agitation assists the extraction of water-soluble particles embedded within the filter. The extracted filters are then aged for >12 h in a refrigerator at \({<}4\,{^{\circ }}\)C. This process allows for complete dissolution of water-soluble components and the settling of any solid materials to the bottom of the vial (Stevens et al. 1978).
Although many tests of water extraction efficiency from filters have been conducted over the decades, few results from these tests have been published (Fosco and Schmeling 2007; Hoffer et al. 1979; Talebi and Abedi 2005). These tests involve either spiking the filter with known amounts of the ions using a microsyringe, comparing a less aggressive extraction with a more proven aggressive method, or by submitting an extracted filter to a second extraction (Mouli et al. 2003). Derrick and Moyers 1981 examined extraction methods for Teflon-membrane and quartz-fiber filters. Teflon is hydrophobic, and DDW does not wet the surface or penetrate within the membrane (Wolfson 1980). Pure DDW, sodium carbonate/bicarbonate eluent, and perchloric acid by themselves yielded <80% efficiencies after 15 min of sonication and <2 h of aging for nitrate, sulfate, and ammonium. However, the DDW extraction yielded >95% efficiencies for these ions when analyzed after 24 h of aging following sonication. Chow and Watson (1999) noted the hydrophobic nature of Teflon-membrane filters that require applications of a wetting solution (e.g., 200 \(\upmu \)L ethanol or methanol) onto the filter deposit surface to ensure complete dissolution of the analyte ions.
Nylon-membrane filters are hydrophilic, as evidenced by their soaking up of a water droplet, in contrast to the Teflon-membrane on which the droplet remains on the surface. Yu et al. (2005) found equivalent nitrate and sulfate concentrations on Nylon-membrane filters extracted in DDW and IC eluent, both with 30–45 min of sonication. They did not report the aging time prior to analysis, but it is evident that the eluent extraction is unnecessary. This is a useful result, as the sodium carbonate/bicarbonate IC eluent for anions would interfere with possible cation analysis on the extracts. A pure DDW extraction is preferable
Extraction efficiencies determined from two sets of 20 Nylon-membrane filters spiked with known concentrations of mixed anion standards are compared in Fig. 3. Recovery rates range from 100–103.9% and 92–96% for Laboratories A and B, with averages and standard deviations of \(100.2\pm 0.8\%\) and \(94\pm 2.1\%\), respectively.
IC for Anions, Cations, Organic Acids, and Carbohydrates
Thermo/Dionex ICs are most commonly used, although alternatives are also available. The older models (e.g., DX-500, DX-600, and ICS-2000) have been used in the past (Chow et al. 1990; Watson 1991) and are adequate for inorganic anion and cation analyses. These instruments are still available on the used-instrument market, along with replacement parts and supplies. They have been largely superseded by later model hardware and software.
Anions
Anion columns use carbonate or hydroxide-selective anion exchange resins that can accommodate a reasonable (\(\sim \)2000) sample throughput. An anion electrolytically regenerated suppresser (AERS) is used to reduce background conductivity by exchanging the K\(^{+}\) with hydrogen ion (H\(^{+})\) to produce water in the case of hydroxide (or similarly convert carbonate to bicarbonate and/or carbonic acid). An eluent generator is used to produce a 15 millimole (mM) potassium hydroxide solution for isocratic elution followed by conductivity detection.
Anion chromatograms from two systems are compared in Fig. 4. The ICS-5000\(^{+}\) configuration (Fig. 4b) has higher sensitivity with improved chromatographic resolution. This is partially due to the carbonate/bicarbonate buffer system used by the ICS-3000, which does not completely eliminate background conductivity. This is most evident for the fluoride or chloride peak, which often interferes with the DDW dip. As the conductance of DDW is often less than that of the suppressed eluent, fluoride or chloride eluting near the DDW dip is co-eluted or suppressed. Background subtraction and manual adjustments are required for the older IC systems to quantify potential co-eluted species and peak overlaps. This is labor-intensive and results in large uncertainties in ion concentrations (Chow and Watson 1999). Nitrite was also found in this sample, though it was below ICS-3000 minimum detection limits (MDLs) (Fig. 4a).
The carbonate peaks in Fig. 4b are dominated by artifacts rather than components of the aerosol deposit. Despite de-gassing of the water by helium sparging, this interference results from carbon dioxide (CO\(_{2})\) dissolved in the DDW or eluent. The carbonate peak is separable from the other anions and causes no bias to their quantification. It may be reduced by purging the sample chamber with an inert gas and minimizing the time between extraction and analysis.
Figure 5 compares nitrate and sulfate concentrations for the same sample extracts analyzed by the ICS-5000\(^{+}\) and the ICS-3000 systems. There is good agreement (\(0.98 \le R^{2} \le 1\)) for the replicate analyses, indicating that the eluent difference does not affect the results for these important components.
Cations
Cation columns (i.e., either the guard column or analytical column) use a hydrophilic and carboxylate functionalized cation exchange resin that can accommodate a reasonable (\(\sim \)2500) sample throughput. A cation electrolytically regenerated suppressor (CERS) is used to reduce background conductivity by exchanging the buffer anion (e.g., methane sulfonate or sulfate) with hydroxide ion (OH\(^{-})\), and an eluent generator is used to produce 40 mM methanesulfonic acid (MSA) solution for isocratic elution followed by conductivity detection. Figure 6 shows an example of the cation chromatogram with \(\sim \)20-min analysis time per sample.
Both gaseous ammonia (NH\(_{3})\) and particle ammonium are important atmospheric constituents that contribute to the global nitrogen cycle (Fowler 2015), elevated PM\(_{2.5}\) (Wang et al. 2013), and ecosystem deposition (Granath et al. 2014). If ammonium is the major cation of interest, a fast run cation column (e.g., 4\(\,\times \,\)250 mm CS12A) can be used with a 33 mM MSA eluent composition for a 3-min analysis time. Although there are potential savings on analysis time (i.e., from 20 to 3 min per run), these are not necessarily achieved when other cations are present, as additional rinsing time is needed to elute the remaining cations (e.g., water-soluble magnesium, potassium, and calcium); sample carry-over may cause cross-contamination. The \(\sim \)20-min analysis time ensures clean separation. The only cost savings would be the time reduction in system calibration and chromatogram validation, as fewer ion species are needed to prepare calibration standards and only a single peak is subject to verification.
The conductivity/concentration relationship is non-linear for ammonium at high concentrations and a quadratic curve is needed to fit the calibration standards. Figure 7a shows an upper limit of 10 \(\upmu \)g/mL standard may underestimate the ammonium concentration. The linear curve is improved with a calibration curve from 0.005 to 2 \(\upmu \)g/mL (Fig. 7b), with best linearity found with an upper limit of 1 \(\upmu \)g/mL (Fig. 7c). For cation analyses, the upper limit can be set at 10 \(\upmu \)g/mL for water-soluble sodium, magnesium, potassium, and calcium ions and at 1 \(\upmu \)g/mL for ammonium. Values above these require dilution and re-analysis.
Alkylamines
Alkylamines, derivatives of ammonia, can serve as markers for agricultural sources (Ge et al. 2011) and are determined following the cation procedure. Dimethylamine may co-elute with the magnesium ion and trimethylamine may co-elute with the calcium ion; however, their concentrations are usually small compared to the inorganic ions. Using gradient eluent mode, Fig. 8 shows that the three alkylamines can be quantified along with ammonium and four other inorganic cations within a single \(\sim \)30-min run.
Organic Acids
Organic acids derive from a mixture of anthropogenic sources, including biomass burning; biodiesel, diesel, and gasoline engine exhaust; marine aerosols; meat cooking; biogenic; as well as primary organic aerosols (POAs) and SOAs (Hawkins et al. 2010; Kundu et al. 2010; Millet 2015). Figure 9 shows the extension of anion analysis with nine organic acids in a single run. The organic acids most commonly found in ambient samples includes four mono-carboxylic acids (i.e., lactate, acetate, formate, and MSA) and five di-carboxylic acids (i.e., glutarate, succinate, malonate, maleate, and oxalate) (Brent et al. 2014; Karthikeyan et al. 2007) that can be speciated along with commonly measured anions as specified in Table 2.
Some of these organic acids could possibly interfere with inorganic anion peaks if present in sufficient quantities. With proper gradient elution, Fig. 9 shows that early eluting mono-carboxylic acids (e.g., acetate, formate, or MSA) do not interfere with the chloride peak and dicarboxylic acids (e.g., maleate or oxalate) do not interfere with the sulfate peak.
Carbohydrates
Carbohydrates are markers for biomass burning, bioaerosols, biogenic POAs, and SOAs. The commonly used biomass burning marker, levoglucosan, as well as its isomers mannosan and galactosan, can be quantified by IC using a pulsed amperometric detection (PAD) (Engling 2006; Garcia et al. 2005; Iinuma et al. 2009; Zhang 2013), with applications in several field studies (Caseiro et al. 2007; Ho 2015; Iinuma et al. 2009; Sullivan et al. 2011; VandenBoer et al. 2012; Yttri 2015; Zhang 2013). IC-PAD integrates HPLC and IC techniques by combining anion exchange with electrochemical detection. It is based on the principle that multiple hydroxyl (OH) groups are detected by amperometry, using specific voltages to induce oxidation of individual species and to elute from the anion exchange column. Carbohydrate species are separated upon ionization with a strong alkaline eluent (see Table 2).
IC-PAD is cost-effective as it allows for detection of polar organic compounds using portions of the same DDW sample extract available for anion or cation analyses without the need for chemical derivatization, solvent extraction, and/or extract volume reduction (as is required for gas chromatography–mass spectrometry (GC–MS) analysis of polar organic compounds). Other analytical methods used for polar organic compounds require a separate sample for solvent extraction followed by GC–MS (Mazurek et al. 1989). An additional advantage of the IC-PAD method is its sensitivity and selectivity for multiple poly-hydroxy compounds (i.e., anhydrosugars and sugar alcohols). The electrochemical detector is optimized in pulsed amperometric mode, allowing selective measurement of the anhydrosugars, along with other carbohydrates and sugar alcohols. Figure 10 shows the chromatogram of markers for biomass burning (e.g., levoglucosan, mannosan, and galactosan); fungi (e.g., arabitol, mannitol, and trehalose); bacteria (e.g., adonitol); and biogenic POAs (e.g., erythritol, inositol, glucose, galactose, arabinose, fructose, and sucrose). Biogenic SOA markers (e.g., 2-methyl-erythritol) derived from oxidation of isoprene can also be identified and quantified with a single \(\sim \)60-min run.
IC Analysis Procedures, Quality Control (QC), and Quality Assurance (QA)
Several standard operating procedures (SOPs) are available that can serve as starting points for establishing an IC laboratory (Pfaff 1997; U.S.EPA 2000, 2007). SOPs codify the actions that are taken to implement a measurement process over a specified time period (Watson et al. 2001). The latest scientific information should be incorporated into SOPs with each revision. QC and QA activities are important parts of each SOP, as specified in Tables 4 and 5, respectively.
System calibration relates the sensor output to known concentrations as determined by standard solutions that span the range of expected concentrations. Although these standards can be prepared by carefully weighing and dissolving dehydrated salts in known amounts of DDW, it is now possible to purchase certified solutions (AlfaAesar 2016; ERA 2016) that can be diluted to the appropriate concentrations. The stock solution of 100 \(\upmu \)g per milliliter (mL) is diluted to 10 levels of working standards ranging from 0.005 to 10 \(\upmu \)g per mL. It is good practice to use solutions from one supplier for calibration and from another supplier for QC performance tests to assure that the concentrations are accurate. Table 1 includes MDLs and lower quantifiable limits (LQLs) achievable with the Table 2 configurations. MDLs are typically much lower than the LQLs derived from field blanks, which include passive deposition and adsorption of gases.
After analysis, each chromatogram is examined for: (1) proper operational settings and peak identification; (2) correct peak shapes and integration windows; (3) peak overlaps; (4) correct background subtraction (if needed); and (5) QC standard comparisons. Level I data validation (Watson et al. 2001) involves chromatogram inspection for each analysis, applying range checks to batches of data and investigating outliers, ensuring differences in replicate analysis comparable to those of differences in long-term averages, and examining consistency between calibration standards and performance tests with independent QC standards. Analysts need to re-calculate ion concentrations from existing chromatograms when there is evidence that peaks were not adequately defined by the peak-processing software, and re-analyze batches of samples that do not pass validation tests. Level I data validation flags are assigned that indicate filter appearance, filter damage, unusual deposits, excessive filter loading, and other anomalies observed during the analysis. Cation/anion balances (Chow et al. 1994) are also a useful validation tool when both are measured, although these may not balance when unquantified hydrogen ions are present, as for sulfuric acid.
Concentrations for each species are reported for each filter after normalizing to the extraction volume. Analysis precisions are calculated for different concentration intervals as:
where CV is the coefficient of variation for each ion; N is the number of samples, \(c_{i}\) is the concentration of initial analysis of sample ifor each ion, \(c_{i,r}\) is the concentration of sample “i” replicate analysis for each ion; MDL is defined as three standard deviations of laboratory blanks for each ion; \(\sigma _{\mathrm{cion}}\) is the precision of \(c_{\mathrm{ion}}\).
Data are transmitted to a master data base in any specified format, which usually includes a separate record for each sample with the sample ID, filter lot number, analysis date, validation flags, filter loading values, and precision for each element. These are associated with field data, such as air sample volume, through the sample ID for a final output in \(\upmu \)g/m\(^{3}\).
QA consists of external examinations of the measurement process, as delineated in Table 5. These include laboratory certifications (TNI 2016), inspections and systems audits (Taylor 2005, 2007), and performance audits/interlaboratory comparisons (Smiley 2005, 2007, 2009, 2010; Taylor 2008). These are often required by an external sponsor, but they are good practice even if not required.
Summary and Conclusions
IC has progressed from simple analyses of inorganic anions and cations to a versatile laboratory tool applicable to a large range of water-soluble inorganic and organic compounds. Modern hardware and software allows several organic compounds to be discerned along with the normal anion and cation analyses. Modifications of columns, eluent compositions, concentrations, and detectors allow further characterization of alkylamines, organic acids, and carbohydrates that are markers for biomass burning, fuel combustion, engine exhaust, meat cooking, marine aerosol, and bioaerosol (e.g., fungi and bacteria) as well as primary and secondary organic aerosols. Labor is reduced by autosampling and sample routing, such that a single sample tray could be used for multiple analyses described here. These additional water-soluble species are useful to better understand their abundances and variations in source profiling and to refine aerosol source apportionment.
References
AlfaAesar (2016) Ion chromatography standards. AlfaAesar, Tewksbury
Baker SC, Kelly DP, Murrell JC (1991) Microbial degradation of methanesulfonic acid: a missing link in the biogeochemical sulfur cycle. Nature 350:627–628. doi:10.1038/350627a0
Brent LC, Reiner JL, Dickerson RR, Sander LC (2014) Method for characterization of low molecular weight organic acids in atmospheric aerosols using ion chromatography mass spectrometry. Anal Chem 86:7328–7336
Buchberger WW (2001) Detection techniques in ion chromatography of inorganic ions Trac-Trends. Anal Chem 20:296–303. doi:10.1016/s0165-9936(01)00068-1
Butt SB, Riaz M (2009) Determination of cations and anions in environmental samples by HPLC: review. J Liq Chromatogr Relat Technol 32:1045–1064. doi:10.1080/10826070902841299
Bytnerowicz A, Sanz MJ, Arbaugh MJ, Padgett PE, Jones DP, Davila A (2005) Passive sampler for monitoring ambient nitric acid (HNO3) and nitrous acid (HNO2) concentrations. Atmos Environ 39:2655–2660
Calloway CP, Li SM, Buchanan JW, Stevens RK (1989) A refinement of the potassium tracer method for residential wood smoke. Atmos Environ 23:67–69
Caseiro A, Marr IL, Claeys M, Kasper-Giebl A, Puxbaum H, Pio CA (2007) Determination of saccharides in atmospheric aerosol using anion-exchange high-performance liquid chromatography and pulsed-amperometric detection. J Chromatogr A 1171:37–45
Chakrabarty RK et al (2016) Brown carbon aerosols from burning of boreal peatlands: microphysical properties, emission factors, and implications for direct radiative forcing Atmos. Chem Phys 16:3033–3040. doi:10.5194/acp-16-3033-2016
Cheng Z et al (2015) Estimation of aerosol mass scattering efficiencies under high mass loading: case study for the megacity of Shanghai. China Environ Sci Technol 49:831–838
Chow JC, Engelbrecht JP, Watson JG, Wilson WE, Frank NH, Zhu T (2002) Designing monitoring networks to represent outdoor human exposure. Chemosphere 49:961–978
Chow JC, Fujita EM, Watson JG, Lu Z, Lawson DR, Ashbaugh LL (1994) Evaluation of filter-based aerosol measurements during the. South Calif Air Qual Study Environ Mon Assess 30:49–80
Chow JC, Lowenthal DH, Chen L-WA, Wang XL, Watson JG (2015) Mass reconstruction methods for PM\(_{2.5}\): a review. Air Qual Atmos Health 8:243–263
Chow JC, Watson JG (1999) Ion chromatography in elemental analysis of airborne particles. In: Landsberger S, Creatchman M (eds) Elemental analysis of airborne particles, vol 1., Advances in environmental industrial and process control technologies. Gordon and Breach Science, Amsterdam, pp 97–137
Chow JC, Watson JG (2013) Chemical analyses of particle filter deposits. In: Ruzer L, Harley NH (eds) Aerosols handbook : measurement, dosimetry, and health effects, 2nd edn. CRC Press/Taylor & Francis, New York, pp 179–204
Chow JC et al (1993) A sampling system for reactive species in the western United States. In: Winegar ED, Keith LH (eds) Sampling and analysis of airborne pollutants. Lewis Publishers, Ann Arbor, MI, pp 209–228
Chow JC, Watson JG, Bowen JL, Frazier CA, Gertler AW, Hinsvark BA, Fung KK (1990) Laboratory operations manual for the California Acid Deposition Monitoring Program: Sample pretreatment, sample preparation, and chemical analysis, Final report. Desert Research Institute, Reno, NV
Dabek-Zlotorzynska E et al (2011) Canadian National Air Pollution Surveillance (NAPS) PM\(_{2.5}\) speciation program: methodology and PM\(_{2.5}\) chemical composition for the years 2003–2008. Atmos Environ 45:673–686
Davison B et al (1996) Dimethyl sulfide, methane sulfonic acid and physicochemical aerosol properties in Atlantic air from the United Kingdom to Halley Bay. J Geophys Res 101:22855–22868
Derrick M, Moyers J (1981) Precise and sensitive water-soluble ion extraction method for aerosol samples collected on polytetrafluoroethylene filters. Anal Lett Part A Chem Anal 14:1637–1652
Donahue NM et al (2013) Why do organic aerosols exist? Understanding aerosol lifetimes using the two-dimensional volatility basis set. Environ Chem 10:151–157
Duvall RM, Majestic BJ, Shafer MM, Chuang PY, Simoneit BRT, Schauer JJ (2008) The water-soluble fraction of carbon, sulfur, and crustal elements in Asian aerosols and Asian soils. Atmos Environ 42:5872–5884
Engling G et al (2006) Determination of levoglucosan in biomass combustion aerosol by high-performance anion-exchange chromatography with pulsed amperometric detection. Atmos Environ 40:S299–S311
ERA (2016) Custom standards simplified. ERA: A Waters Company, Golden, CO
Fenn JB (2002) Electrospray ionization mass spectrometry: how it all began. J Biomol Tech 13:101–118
Fenn JB (2003) Electrospray wings for molecular elephants (Nobel lecture). Angewandte Chem Int Edition 42:3871-3894. doi:10.1002/anie.200300605
FisherScientific (2016) Falcon\(^{\rm TM}\) 15mL Conical Centrifuge Tubes. Thermo Fisher Inc, Waltham, MA
Fosco T, Schmeling M (2007) Determination of water-soluble atmospheric aerosols using ion chromatography. Environ Mon Assess 130:187–199
Fowler D et al (2015) Effects of global change during the 21st century on the nitrogen cycle Atmos. Chem Phys 15:13849–13893. doi:10.5194/acp-15-13849-2015
Fritz JS (1991) Principles and applications of ion-exclusion. chromatogr J Chromatogr A 546:111–118
Galloway JN, Dentener FJ, Capone DG, Boyer EW, Howarth RW, Seitzinger SP, Asner GP, Cleveland CC, Green PA, Holland EA, Karl DM, Michaels AF, Porter JH, Townsend AR, Vorosmarty CJ (2004) Nitrogen cycles: past, present, and future. Biogeochemistry 70:153–226
Gama MR, Silva RGD, Collins CH, Bottoli CBG (2012) Trac Trends Anal Chem. Hydrophilic interaction chromatography 37:48–60. doi:10.1016/j.trac.2012.03.009
Garcia CD, Engling G, Herckes P, Collett JL Jr, Henry CS (2005) Determination of levoglucosan from smoke samples using microchip capillary electrophoresis with pulsed amperometric detection. Environ Sci Technol 39:618–623
Ge XL, Wexler AS, Clegg SL (2011) Atmospheric amines - Part I. A review. Atmos Environ 45:524–546
Gennaro MC, Angelino S (1997) Separation and determination of inorganic anions by reversed-phase high-performance liquid chromatography. J Chromatogr A 789:181–194. doi:10.1016/s0021-9673(97)00971-0
Granath G, Limpens J, Posch M, Mucher S, De Vries W (2014) Spatio-temporal trends of nitrogen deposition and climate effects on Sphagnum productivity in European peatlands. Environ Pollut 187:73–80
Hartwell SK, Kehling A, Lapanantnoppakhun S, Grudpan K (2013) Flow injection/sequential injection chromatography: a review of recent developments in low pressure with high performance chemical separation. Anal Lett 46:1640–1671
Hawkins LN, Russell LM, Covert DS, Quinn PK, Bates TS (2010) Carboxylic acids, sulfates, and organosulfates in processed continental organic aerosol over the southeast Pacific Ocean during VOCALS-REx 2008. J Geophys Res-Atmos 115
Hecobian A, Evanoski-Cole A, Eiguren-Fernandez A, Sullivan AP, Lewis GS, Hering SV, Collett JL (2016) Evaluation of the Sequential Spot Sampler (S3) for time-resolved measurement of PM\(_{2.5}\) sulfate and nitrate through lab and field measurements Atmospheric. Meas Tech 9:525–533. doi:10.5194/amt-9-525-2016
Ho KF et al (2015) Characteristics of water-soluble organic nitrogen in fine particulate matter in the continental area of China. Atmos Environ 106:252–261
Hoffer EM, Kothny EL, Appel BR (1979) Simple method for microgram amounts of sulfate in atmospheric particles. Atmos Environ 13:303–306. doi:10.1016/0004-6981(79)90173-2
Iinuma Y, Engling G, Puxbaum H, Herrmann H (2009) A highly resolved anion-exchange chromatographic method for determination of saccharidic tracers for biomass combustion and primary bio-particles in atmospheric aerosol. Atmos Environ 43:1367–1371
IMPROVE (2016) Interagency Monitoring of Protected Visual Environments. National Park Service, Ft., Collins, CO
Karthikeyan S, See SW, Balasubramanian R (2007) Simultaneous determination of inorganic anions and selected organic acids in airborne particulate matter by ion chromatography. Anal Lett 40:793–804
Karu N, Dicinoski GW, Haddad PR (2012) Use of suppressors for signal enhancement of weakly-acidic analytes in ion chromatography with universal detection methods. Trac Trends Anal Chem 40:119–132
Kundu S, Kawamura K, Andreae TW, Hoffer A, Andreae MO (2010) Molecular distributions of dicarboxylic acids, ketocarboxylic acids and alpha-dicarbonyls in biomass burning aerosols: implications for photochemical production and degradation in smoke layers Atmos. Chem Phys 10:2209–2225
Lachat (2016) QuikChem with ion chromatography option. Lachat Instruments, Loveland, CO
Laskin A, Laskin J, Nizkorodov SA (2015) Chemistry of atmospheric brown carbon. Chem Rev 115:4335–4382
Lucy CA, Wahab MF (2013) Advances in high-speed and high-resolution ion chromatography. LC GC Eur 31:38-42
Mazurek MA, Cass GR, Simoneit BRT (1989) Interpretation of high-resolution gas chromatography and high-resolution gas chromatography/mass spectrometry data acquired from atmospheric organic aerosol samples. Aerosol Sci Technol 10:408–420
Metrohm (2016) Ion chromatography. Metrohm AG, Herisau, Switzerland
Michalski R (2014) Application of ion chromatography in clinical studies and pharmaceutical industry. Mini Rev Med Chem 14:862–872
Michalski R (2016) Application of IC-MS and IC-ICP-MS in environmental research. Wiley, New York, NY
Millet DB et al (2015) A large and ubiquitous source of atmospheric formic acid. Atmos Chem Phys 15:6283–6304
Miyazaki Y, Fu PQ, Ono K, Tachibana E, Kawamura K (2014) Seasonal cycles of water-soluble organic nitrogen aerosols in a deciduous broadleaf forest in northern Japan. J Geophys Res Atmos 119:1440–1454
Mouli PC, Mohan SV, Reddy SJ (2003) A study on major inorganic ion composition of atmospheric aerosols at Tirupati. J Hazard Mater 96:217–228
Nakatani N, Kozaki D, Mori M, Tanaka K (2012) Recent progress and applications of ion-exclusion/ion-exchange chromatography for simultaneous determination of inorganic anions and cations. Anal Sci 28:845–852
Olariu RI, Vione D, Grinberg N, Arsene C (2015) Applications of liquid chromatographic techniques in the chemical characterization of atmospheric aerosols. J Liq Chromatogr Relat Technol 38:322–348
Pall (2016) Nylasorb\(^{\rm TM}\) nylon membrane disc filters. Pall Corporation, East Hills, NY
Pfaff JD, Hautman DP, Munch DJ (1997) Methd 300.1: Determination of inorganic anions by ion chromatography. US Environmental Protection Agency, Research Triangle Park, NC
Pitchford ML, Malm WC, Schichtel BA, Kumar NK, Lowenthal DH, Hand JL (2007) Revised algorithm for estimating light extinction from IMPROVE particle speciation data. JAWMA 57:1326–1336
Pohl C (2013) Recent developments in ion exchange columns for ion chromatography. LC GC Eur 31:16–22
Small H, Stevens TS, Bauman WC (1975) Novel ion exchange chromatographic method using conductimetric detection. Anal Chem 47:1801–1809
Smiley J (2005) Technical memorandum: experimental intercomparison of speciation laboratories. US Environmental Protection Agency, Montgomery, AL
Smiley J (2007) Technical memorandum: experimental intercomparison of speciation laboratories. US Environmental Protection Agency, Montgomery, AL
Smiley J (2009) Technical memorandum: experimental intercomparison of speciation laboratories. US Environmental Protection Agency, Montgomery, AL
Smiley J (2010) Technical memorandum: experimental intercomparison of speciation laboratories. US Environmental Protection Agency, Montgomery, AL
Stevens RK, Dzubay TG, Russwurm GA, Rickel DE (1978) Sampling and analysis of atmospheric sulfates and related species. Atmos Environ 12:55–68
Sullivan AP, Frank N, Onstad G, Simpson CD, Collett JL Jr (2011) Application of high-performance anion-exchange chromatography-pulsed amperometric detection for measuring carbohydrates in routine daily filter samples collected by a national network: 1. Determination of the impact of biomass burning in the upper Midwest. J Geophys Res-Atmos 116:D8. doi:10.1029/2010JD014166
Talebi SM, Abedi M (2005) Determination of atmospheric concentrations of inorganic anions by ion chromatography following ultrasonic extraction. J Chromatogr A 1094:118–121. doi:10.1016/j.chroma.2005.07.118
Taylor S (2005) Technical memorandum: DRI laboratory audit. US Environmental Protection Agency, Montgomery, AL
Taylor S (2007) Technical memorandum: DRI laboratory audit. US Environmental Protection Agency, Montgomery, AL
Taylor S (2008) Technical memorandum: experimental intercomparison of speciation laboratories. US Environmental Protection Agency, Montgomery, AL
ten Brink HM, Otjes R, Jongejan P, Slanina S (2007) An instrument for semi-continuous monitoring of the size-distribution of nitrate, ammonium, sulphate and chloride in aerosol. Atmos Environ 41:2768–2779
Thermo-Dionex (2016) Ion chromatography. Thermo Scientific Inc, Waltham, MA
TNI (2016) National Environmental Laboratory Accreditation Program (NELAP). The NELAC Institute (TNI), Weatherford, TX
Turetsky MR, Benscoter B, Page S, Rein G, van der Werf GR, Watts A (2015) Global vulnerability of peatlands to fire and carbon loss. Nature Geosci 8:11–14
U.S.EPA (2000) NAREL standard operating procedure—Cation analysis for the PM\(_{2.5}\) chemical speciation QA program. U.S. Environmental Protection Agency, Las Vegas, NV
U.S.EPA (2007) Method 9056A: Determination of inorganic anions by ion chromatography. Environmental Protection Agency, USA
U.S.EPA (2016a) Chemical speciation. US Environmental Protection Agency, Research Triangle Park, NC
U.S.EPA (2016b) PM\(_{2.5}\) speciation lab audit reports and assessments. US Environmental Protection Agency, Research Triangle Park, NC
URG (2016) Ambient nitrate and sulfate monitor. URG Corporation, Chapel Hill, NC
VandenBoer TC, Markovic MZ, Petroff A, Czar MF, Borduas N, Murphy JG (2012) Ion chromatographic separation and quantitation of alkyl methylamines and ethylamines in atmospheric gas and particulate matter using preconcentration and suppressed conductivity detection. J Chromatogr A 1252:74–83
Wang Y, Zhang QQ, He K, Zhang Q, Chai L (2013) Sulfate-nitrate-ammonium aerosols over China: response to 2000–2015 emission changes of sulfur dioxide, nitrogen oxides, and ammonia Atmos. Chem Phys 13:2635–2652
Wang YH, Field RD, Roswintiarti O (2004) Trends in atmospheric haze induced by peat fires in Sumatra Island, Indonesia and El Nino phenomenon from 1973 to 2003. Geophys Res Lett 31
Waters (2016) Acquity UPC2 system. Waters Corporation, Milford, MA
Watson JG (2002) Visibility: science and regulation—critical review. JAWMA 52:628-713
Watson JG, Chow JC (2013) Source apportionment. In: El-Shaarwi AH, Piegorsch WW (eds) Encyclopedia of environmetrics. Wiley, Chichester, pp 1–8
Watson JG, Chow JC (2015) Receptor models and measurements for identifying and quantifying air pollution sources. In: Murphy BL, Morrison RD (eds) Introduction to Environmental Forensics, 3rd edn. Elsevier, Amsterdam, pp 677–706
Watson JG et al (1991) Measurements of dry deposition parameters for the California Acid Deposition Monitoring Program. Desert Research Institute, Reno, NV
Watson JG, Chow JC, Engling G, Chen L-WA, Wang XL (2016) Source apportionment: principles and methods. In: Harrison RM (ed) Airborne particulate matter: sources, atmospheric processes and health. Royal Society of Chemistry, London, pp 72–125
Watson JG, Turpin BJ, Chow JC (2001) The measurement process: precision, accuracy, and validity. In: Cohen BS, McCammon CS Jr (eds) Air sampling Instruments for Evaluation of Atmospheric Contaminants, 9th edn. American Conference of Governmental Industrial Hygienists, Cincinnati, OH, pp 201-216
Weber RJ, Orsini D, Daun Y, Lee YN, Klotz PJ, Brechtel F (2001) A particle-into-liquid collector for rapid measurement of aerosol bulk chemical composition. Aerosol Sci Technol 35:718–727
Wolfson JM (1980) Determination of microgram quantities of inorganic sulfate in atmospheric particulates. JAPCA 30:688–690
Yttri KE et al (2015) An intercomparison study of analytical methods used for quantification of levoglucosan in ambient aerosol filter samples. Atmos Measurement Tech 8:125–147
Yu XY, Lee T, Ayres B, Kreidenweis SM, Collett JL Jr, Malm WC (2005) Particulate nitrate measurement using nylon filters. JAWMA 55:1100–1110
Zhang FF, Shen GB, Ji SL, Yang BC (2015) Recent advances of stationary phases for hydrophilic interaction liquid chromatography and ion chromatography. J Liq Chromatogr Relat Tech 38:349–352. doi:10.1080/10826076.2014.941258
Zhang XY, Wang YQ, Niu T, Zhang XC, Gong SL, Zhang YM, Sun JY (2012) Atmospheric aerosol compositions in China: spatial/temporal variability, chemical signature, regional haze distribution and comparisons with global aerosols Atmos. Chem Phys 12:779–799
Zhang ZS et al (2013) Determination of isoprene-derived secondary organic aerosol tracers (2-methyltetrols) by HPAEC-PAD: Results from size-resolved aerosols in a tropical rainforest. Atmos Environ 70:468–476
Acknowledgements
This research is funded in part by the US National Science Foundation (CHE-1214163 and CHE-1464501) and the US EPA PM\(_{2.5}\) Chemical Speciation Network (CSN) Laboratory Analysis Program (EP-D-15-0250). The authors would like to thank Mr. Patrick Hurbain, Dr. Guenter Engling (now with California Air Resources Board), Dr. Paul Cropper, and Mr. Steve Kohl for technical input and laboratory analyses and Ms. Katherine Greenblatt of the Desert Research Institute (DRI) for assembling and editing the manuscript.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflicts of interest
On behalf of all authors, the corresponding author states that there is no conflict of interest.