Abstract
Purpose
The Self Powered Neutrons Detectors (SPND) have the advantage of not requiring a high voltage power supply for their operation and are small in size, enhancing the interest of these detectors in medicine.
Methods
In this context, we have developed a thermal neutron detection system based on SPND. This detector was placed in the thermal channel of our nuclear research reactor; where the values of the current for each detector have been recorded as a function of time, with a chain in a current mode where electrometers without HV were used.
Results
We performed the real-time measurement of neutron flux during boron neutron capture therapy or boron neutron therapy, the different materials constituting the SPND detectors have been carefully chosen for this application. These detectors were tested at a power of four MW corresponding to a neutron flux of 109 n cm−2 s−1.
Conclusions
The usefulness of 103Rh-SPND is for online measurement of thermal neutron flux on BNCT patients has been demonstrated based on an appropriate calibration of the thermal neutron spectrum.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Boron neutron capture therapy (BNCT) or boron neutron therapy is a technique that requires simultaneous presence of neutron fluxes with adequate energy and a neutron sensor, the boron-10 (10B) [1]. Boron and neutrons interact to attack tumor cells without causing significant damage to healthy cells because the destructive effect occurs mainly in tumor cells that have selectively accumulated boron, the capture reaction is as follows [2]
The path of alpha (4He) and lithium (7Li) particles in tumor tissue is around the diameter of the tumor cell (~ 10 µm) [3,4,5]. Therefore, the destructive action of the capture reaction occurs mainly in those cancer cells that have accumulated boron, while the normal cells with a low concentration of boron are not attacked. In order to optimize treatment by the BNCT technique, we are interested in determining the heat flux of neutrons in specific areas of patients. A few activation methods are currently used for this purpose, but they are not able to provide real-time information. Many authors have developed "in real-time" thermal neutron SPNDs using fast response time materials with neutron capture cross sections to supply current [6,7,8]; for example, the signal rhodium-based detector is the beta-current type of self-powered detector, which uses the activation reaction below to produce a current that can be measured according to Eq. (2).
When the neutron is captured by 103Rh, a 104Rh nucleus is formed. Then, it decays into 104Pd and \({\acute{\mathrm e} }\) by decay \({\beta }^{-}\). Thus, high-energy electrons are generated in the emitter. Figure 1 shows the 103Rh reaction chain.
A lot of self-powered neutron detectors were designed, developed and tested around the world [9]. Recently, many new self-powered neutron detectors types have been developed and tested for medical applications [10, 11]. Those detectors are now considered one of the most robust and reliable choices of in-core flux-detectors owing to their successful performance for decades.
In this work, we present the design and realization of new SPND detectors that we have developed in our measurement detectors laboratory in the nuclear research center of Birine, in order to use it in the measurement, online, of the thermal neutrons flux incident during therapy by neutron capture (BNCT). Compared to other in-core detectors, BNCTs feature many advantages, those types of detectors are small size, which can be placed directly under the skin or inside the patient's brain, for processing to determine exactly the actual flux used in the irradiated area; in addition, they are simple and robust structure. In addition, they show a good stability under temperature and pressure conditions. Another advantage of this type of detector is no high voltage is applied, which is very important for medical applications.
Conception, realization
Principle
The self-powered detector SPND is a coaxial cylindrical assembly of different materials, as shown in Fig. 2, which consists of an emitter, a center conductor with a material of relatively high cross section, an insulator and an external conductive sheath (collector). In this type of detector, the central emitter picks up the neutrons and emits the high-energy electrons, which reach the outer cladding after passing through the insulator. In the SPND detector, the current is proportional to the neutron flux that can be measured using an electrometer connected between the emitter and the sheath (Fig. 2). Table 1 shows the labels of our SPND prototype.
These kinds of detectors have a probability for a response to gamma radiation due to compton or photoelectric effects which can produce free electrons with enough energy to pass from the collector to the emitter or vice versa. This probability of occurrence of the two effects increases with Z and the mass of the electrode materials [2]. When designing the detector, the appropriate materials must be well chosen [13,14,15,16], so that the gamma response is as low as possible to eliminate the effect of gamma radiation [12, 17].
Material selection
In this work, several prototypes that aimed to optimize our detector design have been realized, then study and characterizing its response, and the materials of the components of the developed detector are chosen as follows:
Emitter: must have an effective neutron capture section, with a low mass, would provide a good level of signal at the output with a flux to be measured greater than 108 n cm−2 s−1. The current level (I) is being estimated according to the following formula [2]:
where C is constant relating to geometry and efficiency; q is charge released by absorbed neutron; σ is effective activation section of the emitter; N is number of atoms of the emitter; Φ is neutron flux.
Among all the materials analyzed, rhodium 103 (103 Rh) was the appropriate choice because this element has a fairly high capture cross section: 134 barns [18, 19], q = e (electron charge), C = 1, Φ = 108 n cm−2 s−1 in the worst case, and by determining N for a cylindrical emitter of dimensions (length = 1 cm and diameter = 1 mm), the current levels were obtained of the order picoammeter. In this picoammeter order, current levels can be easily measured with instrumentation very similar to low current measurement. In addition, although the rhodium signal is delayed due to the beta decay half-life of 42.8 s [18, 20], this time is negligible compared to the irradiation time of the treatment (more than 30 min); therefore, in these cases, in practice, the signal can be considered "online." For high-intensity beams or if several irradiation steps are used and the irradiation times may therefore be only a few minutes, it will be necessary to use dynamic compensation to correct for the delay.
Insulation: we have chosen acrylic as the insulating material in all designs; it is generally used in all detectors for applications in the medical field. It is a tissue-equivalent material for neutrons, which does not increase the radiological risk during or after irradiation, because it does not alter the treatment conditions and does not activate after the irradiation is terminated. In addition, the properties of acrylic, for this type of application, do not present any risk of appreciable modification under gamma or neutron irradiation up to 50 kGy [21, 22] and 1017 n cm−2 [23,24,25], respectively. Therefore, with the typical fluxes used for BNCT treatment, the acrylic insulation shows no significant degradation even in the long term.
The collector (outer sheath): the sheath material must not be toxic to avoid contamination of the patients during medical application, must be sterilized before each use, must have a very low effective capture section, and have the lowest possible residual radioactivity. For that, two materials were evaluated: graphite and Zircaloy-4 [25,26,27], and the main advantage of using graphite is the possibility of obtaining very thin conductive layers, but it is a water-soluble material, which complicates its implantation in tissues. Zircaloy-4 is a zirconium-based alloy that must be subjected to a purification process to reduce impurities present in the base metal and which can have a high effective cross section. In addition, zirconium is a biocompatible metal [28, 29], used in biological prosthesis. With this material (zircaloy), we cannot get thin conductive layers like with graphite, and this will result in a larger outside diameter of the detector [30, 31]. Table 2 shows the characteristics of detectors SPND1, SPND2 and SPND3.
Characterization
For the neutron characterization on the nuclear research reactor, the equipment used is as follows: three detectors: SPND1, SPND2 and SPND3, acrylic support, thin activation sheets for measuring neutron flux, electrometer model Keithley 602, electrometer model Keithley 617 and the last electrometer model Keithley 602 and PC for data acquisition. The detectors are placed on acrylic support as described in the diagram of the irradiation device and the device for irradiating SPNDs (Figs. 3, 4 and 5). In order to measure the heat flux in the areas considered, very thin sheets (activation detectors) are placed in the irradiation area. One of the detectors (SPND2) was placed in a cadmium capsule to estimate the contribution of the entire signal of thermal neutrons for comparison with the naked detector (SPND1) without a capsule. The device was placed in the thermal column of the reactor. We recorded the value of the leakage current, Io, produced by each detector in the absence of radiation. We performed the irradiation on our SPND detector and by recording the results of the current produced by these detectors as a function of time (Figs. 6, 7). The first level of the curve corresponds to the intensity of the heat flow for a power of 2 MW and the second level to a power of 4 MW.
Discussion
Estimation of the gamma rate
Our prototype SPND detectors have been calibrated under gamma radiation from a 60Co source, we deduced the sensitivity of each prototype, and the results obtained are:
We notice that the results of the sensitivity of detector SPND1 and SPND2 are five times higher than SPND3, this is mainly due to the composition of the material of the transmitter, and as the main material of the SPND3 transmitter being Zircaloy (Zy4). This material has a low effective capture section for thermal neutrons, so its low contribution is expected. However, if we assume that the signal from this detector comes from gamma radiation, the value obtained is:
However, the gamma flux cannot be as high in this irradiation position, which is why we consider that the signal produced by the SPND3 can come mainly from the contribution of the coaxial cable.
Evaluation of the neutron sensitivity of SPNDs with Rhodium emitter
Simple model
Suppose that the signal of each SPND is of the form:
where \(S_{{{\text{nt}}}}\) is the sensitivity to thermal neutrons, \(\emptyset_{t}\) is the heat flux, \(I_{0}\) is the leakage current. The net current of the detectors is equal to the total current measured minus the leakage current, namely:
The values obtained during the experiment are \(\emptyset_{t}^{1} = (1.39 \pm 0.02) \times 10^{12} {\text{ n}}\;{\text{cm}}^{ - 2} \;{\text{s}}^{ - 1}\), \(I^{{{\text{net}}}} = \left( {3.24 \pm 0.01} \right) \times 10^{ - 9} {\text{A }}\), and the sensitivity is therefore equal to \(S_{{{\text{nt}}}} = \left( {2.33 \pm 0.03} \right) \times 10^{ - 21} {\text{A/ n}}\;{\text{cm}}^{ - 2} \;{\text{s}}^{ - 1}\).
General model
In this model, we assume that the signal of each SPND is of the form:
\({S}_{\mathrm{nt}}\) is the sensitivity to thermal neutrons, \({\Phi }_{t}\) is the heat flux, \({S}_{n>t}\) is the sensitivity to neutrons of energy greater than the thermal, \({\Phi }_{>t}\) is the neutron flux of energy greater than thermal, \({S}_{ \gamma }\) is the gamma sensitivity, \({\Phi }_{\gamma }\) is the gamma flux, \({I}_{\mathrm{cable}}\) is the response of the cable, which behaves like an SPND with a different emitter material than the active area and \({I}_{0}\) is the leakage current. The net current is therefore the difference between the total current and the current of escape that is to say:
After calculations, we get the \({S}_{\mathrm{nt}}\) sensitivity:
Conclusion
In this work, an implantable SPND-based system was developed in compliance with all design requirements (materials, dimensions and sensitivity). It can be used to obtain online the thermal neutron doses delivered to patients, and to recalculate treatment parameters for their optimization and correction during irradiation. From the results, it can be seen that SPND1 exhibited a strong neutron response, comparing to the response of SPND2 and SPND3, and this leads to the conclusion that most signal comes from the interaction between neutrons and rhodium, compared to Zy-4 that has a low interaction contribution with neutron flux. In addition, it is shown that the level of leakage current of the two detectors SPND1 and SPND2 was less than 1% of that of the neutrons, which is low enough to be considered negligible. This leakage is mainly due to electronic noise, the cable and also depends on the temperature of the medium. The designed detectors provide a stable performance and high confidence that meets the requirements needed for the medical applications tests. This holds promises for developing industry-compatible sensitive self-powered detectors.
References
R.F. Barth, M.G.H. Vicente, O.K. Harling, W.S. Kiger, K.J. Riley, P.J. Binns, F.M. Wagner, M. Suzuki, T. Aihara, I. Kato, S. Kawabata, Current status of boron neutron capture therapy of high grade gliomas and recurrent head and neck cancer. Radiat. Oncol. 7, 1–21 (2012). https://doi.org/10.1186/1748-717X-7-146
G.F. Knoll, Radiation Detection and Measurement (Wiley, 2010)
M.E. Miller, M.L. Sztejnberg, S.J. González, S.I. Thorp, J.M. Longhino, G. Estryk, Rhodium self-powered neutron detector as a suitable on-line thermal neutron flux monitor in BNCT treatments. Med. Phys. 38, 6502–6512 (2011). https://doi.org/10.1118/1.3660204
M.E. Miller, L.E. Mariani, M.L.S. Gonçalves-Carralves, M. Skumanic, S.I. Thorp, Implantable self-powered detector for on-line determination of neutron flux in patients during NCT treatment. Appl. Radiat. Isot. 61, 1033–1037 (2004)
M. shikawa, N. Unesaki, T. Kobayashi, Y. Sakurai, K. Tanaka, S. Endo, M. Hoshi, Real-time estimation of neutron flux during BNCT treatment using plastic scintillation detector with optical fiber. in Research and Development in Neutron Capture Therapy. Monduzzi Editore, Bologna, pp 443–447 (2002)
M. Sajedi, M. Jafarzadeh, Z. Kargar, Construction of a prototype silver self-powered neutron detector and simulation of a developed SPND using MCNP4C code. Iran. J. Sci. Technol. Trans. A Sci. 43, 681–686 (2019)
N.S.M. Ali, K. Hamzah, F.M. Idris, M.H. Rabir, Thermal neutron flux measurement using self-powered neutron detector (SPND) at out-core locations of TRIGA PUSPATI Reactor (RTP). in IOP Conference Series: Materials Science and Engineering, Vol. 298. No. 1. (IOP Publishing, 2018). https://iopscience.iop.org/article/10.1088/1757-899X/298/1/012027/meta
V. Verma, L. Barbot, P. Filliatre, C. Hellesen, C. Jammes, S.J. Svärd, Self powered neutron detectors as in-core detectors for sodium-cooled fast reactors. Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip. 860, 6–12 (2017)
P. Raj, Development and testing of self-powered detectors for nuclear measurements in fusion reactors (2019). https://inis.iaea.org/collection/NCLCollectionStore/_Public/52/029/52029866.pdf?r=1
V.A. Trivillin, E.C.C. Pozzi, L.L. Colombo, S.I. Thorp, M.A. Garabalino, A.M. Hughes, S.J. González, R.O. Farías, P. Curotto, G.A. Santa Cruz, Abscopal effect of boron neutron capture therapy (BNCT): proof of principle in an experimental model of colon cancer. Radiat. Environ. Biophys. 56, 365–375 (2017)
E.C.C. Pozzi, J.E. Cardoso, L.L. Colombo, S. Thorp, A.M. Hughes, A.J. Molinari, M.A. Garabalino, E.M. Heber, M. Miller, M.E. Itoiz, Boron neutron capture therapy (BNCT) for liver metastasis: therapeutic efficacy in an experimental model. Radiat. Environ. Biophys. 51, 331–339 (2012)
S. Fiore, O. Aberle, M. Angelone, M. Calviani, F. Di Giambattista, L. Lepore, M. Nyman, M. Pillon, A. Plompen, Self powered neutron detectors with high energy sensitivity. in EPJ Web of Conferences, vol. 225, (EDP Sciences, 2020), p. 02001. https://www.epj-conferences.org/articles/epjconf/pdf/2020/01/epjconf_animma2019_02001.pdf
M. Fares, M.Y. Debili, M. Messaoudi, S. Begaa, K. Negara, A. Messai, Boron-10 lined proportional counter development for thermal neutron detection. Radiat. Detect. Technol. Methods. (2021). https://doi.org/10.1007/s41605-020-00226-5
M. Fares, A. Messai, S. Begaa, M. Messaoudi, K. Negara, M.Y. Debili, Design and study of the characteristics of a versatile ionization chamber for gamma-ray dosimetry. J. Radioanal. Nucl. Chem. 326, 1405–1411 (2020). https://doi.org/10.1007/s10967-020-07446-5
M. Fares, A. Messai, S. Begaa, M. Salem, K. Negara, M.Y. Debili, M. Messaoudi, 3 He proportional counter development for thermal neutron detection. Radiat. Detect. Technol. Methods. 5(2), 1–9 (2021)
M. Fares, A. Messai, M. Messaoudi, K. Negara, M.Y. Debili, S. Begaa, Initial and volume recombination losses in gamma versatile ionization chamber VGIC for gamma ray dosimetry, Appl. Radiat. Isot. 175, 109794 (2021). https://doi.org/10.1016/j.apradiso.2021.10979
A. Williams, R.I. Sara, Parameters affecting the resolution of a proportional counter. Int. J. Appl. Radiat. Isot. 13, 229–238 (1962)
E. Browne, C. J M, D. J M, L. C M, M.A. Sharp, S. V S, Table of isotopes project. in Annual Report for the period, vol. 155 (1979). https://escholarship.org/content/qt28m305b5/qt28m305b5.pdf#page=170
J.W. Hilborn, Self-powered neutron detectors for reactor flux monitoring. Nucleonics (US) Ceased publication, 22(2) (1964). https://www.osti.gov/biblio/4089624
T. Takeuchi, N. Ohtsuka, A. Shibata, K. Tsuchiya, Development of a self-powered gamma detector. J. Nucl. Sci. Technol. 51, 939–943 (2014)
D.C. Phillips, S.G. Burnay, Polymers in the nuclear power industry. Irradiat. Eff. Polym. (1991). https://inis.iaea.org/search/search.aspx?orig_q=RN:23024974
R. Van Nieuwenhove, L. Vermeeren, Online gamma dose-rate measurements by means of a self-powered gamma detector. IEEE Trans. Nucl. Sci. 49, 1914–1918 (2002)
A. Charlesby, Introductory talk: survey of recent developments in polymer radiation research. Polymer (Guildf). 5, 381–382 (1964)
S.R. Ahmad, A. Charlesby, Effect of 60Co-γ radiation on crystal structure of normal high paraffins. Radiat. Phys. Chem. 11, 29–37 (1978)
M. Alex, K.R. Prasad, S.K. Kataria, Development of bismuth self-powered detector. Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip. 523, 163–166 (2004)
W.H. Todt, Characteristics of self-powered neutron detectors used in power reactors, in Proc. a Spec. Meet. In-Core Inst. React. Core Assessment, NEA Nucl. Sci. Comm. (1996)
M. Angelone, A. Klix, M. Pillon, P. Batistoni, U. Fischer, A. Santagata, Development of self-powered neutron detectors for neutron flux monitoring in HCLL and HCPB ITER-TBM. Fusion Eng. Des. 89, 2194–2198 (2014)
Etude de la Sensibilite de Collectrons. [Study of the Sensitivity of Self-Powered Detectors.] - 0. Erben, C. Le Tanno and C. Morin - CEA-N-2010 (5 Dec 1977) (In French) (INIS-412014)
J.F. Villard, C. Destouches, L. Barbot, D. Fourmentel, V. Radulovic, Improvements in neutron and gamma measurements for material testing reactors, in RRFM/IGORR 2016, RRFM/IGORR 2016 (2016)
A.F. Dykhuis, Capturing irradiation-enhanced corrosion of zircaloy-4 with a charge-based diffusion/drift phase field model (2018)
O. Moreira, H. Lescano, Analysis of vanadium self powered neutron detector’s signal. Ann. Nucl. Energy. 58, 90–94 (2013)
Acknowledgements
This work has been carried out at Es-Salam Nuclear Research Centre of Birine, and the authors would like to thank Mr. Idir Abdellani, Director General of CRNB. Special thanks also are extended to the team of Nuclear Instrumentation Development Division.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Fares, M., Messai, A., Mameri, S. et al. Development of self-powered neutron detectors used in nuclear medicine for the measurement of neutron flows during treatment of boron neutron therapy. Radiat Detect Technol Methods 5, 459–465 (2021). https://doi.org/10.1007/s41605-021-00271-8
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s41605-021-00271-8