Introduction

To date, the design and development phase of new cardiovascular devices is mainly based on benchmark tests and animal experiments, leading to long iterative development cycles. However, in the end, it still remains unclear how the final device will perform in diseased human anatomies, resulting in significant risks for the first patients that undergo treatment with a new device, as well as for the manufacturers of these devices. Using patient-specific computational modelling and simulation during the design and development phase could help to address these challenges.25 This approach allows to assess the interaction between devices and realistic human anatomies, opening the door towards in-silico clinical trials that could be conducted prior to the actual first-in-human clinical study. In this way, risks for patients and manufacturers can be significantly reduced. In addition, combining patient-specific computational modelling and simulation with device optimization techniques can help to develop better devices in a shorter time period.

Many computational modelling and simulation optimization studies have been performed to find optimal stent geometrical parameters based on mechanical or flow-related objectives: radial strength,26,27 fatigue resistance,1,18,19 frame expansion,18,19 flow disturbances and wall shear stress.12,23,28 However, the majority of those studies evaluated solely unit stent models,1 thereby neglecting the interactions with the arterial tissues. Others have used three-dimensional models12,17,20 and accounted for interactions with simplified vessel geometries.18,28 Very recently Bosi et al.5 combined patient-specific modeling and optimization techniques to tune the material parameter of the aortic root in 14 patients, to improve simulation results (i.e., prediction of post-implantation device deformation and paravalvular gaps).

To the best of our knowledge, this is the only available study that combines patient-specific modelling and optimization techniques. However, a device optimization study based on patient-specific modelling has not yet been performed. Such approach could be useful for a wide range of procedures and devices, but this study focuses on transcatheter aortic valve implantation (TAVI).

Clinical studies9,14,21 have demonstrated the efficacy of TAVI to treat patients with aortic stenosis at extreme- and intermediate-risk. However, complications as paravalvular aortic regurgitation (AR) and conduction abnormalities are considered the Achille’s heel of this treatment. AR has been associated with increased mortality,7,15,16 whereas conduction abnormalities leading to implantation of permanent pacemaker are associated with increased costs, longer hospital stays and increased patient’s morbidity. Considering the prospective of treating younger patients at low surgical risk, those complications are inacceptable and need to be addressed by improved devices and procedures.

To be effective and prevent migration, TAVI devices need to anchor at or below the annulus of the aortic root. A good apposition between the device and the aortic wall plays a key role in avoiding paravalvular AR. On the other hand, the contact between the stent and the aortic root might generate high pressure on the atrioventricular conduction system which, in turn, might be responsible of conduction abnormalities.

Previous studies have reported that patient-specific computer simulation can be used to predict paravalvular AR and conduction abnormalities after implantation of a CoreValve Revalving System (Medtronic).4,6,8,22,29

In this proof-of-concept study, we propose a framework to improve geometric parameters of a representative TAVI device frame through patient-specific computer models able to predict post-operative outcomes.

Materials and Methods

Valve Geometry and Material Properties

A parametric model of a 29-mm CoreValve-like transcatheter aortic valve was created. The base geometry of the frame was modeled starting from the planar ‘unwrapped’ configuration in pyFormex (www.nongnu.org/pyformex). The ‘unwrapped’ frame design can be seen as a repetition of 15 diamond-like cells along the circumferential axis and 4 diamond-like cells along the longitudinal axis (Fig. 1a). Each cell was described with Cubic Bezier spline interpolation.11 The profile of the frame of a CoreValve 29 mm (Fig. 1b) was used to generate an offset of the ‘planar’ frame (i.e., varying radius) (Fig. 1c). Finally, the planar frame was ‘wrapped’ to obtain a CoreValve-like shape, in the expanded configuration (Fig. 1d), by applying a coordinate transformation to the nodes of the frame. Geometric parameters of the reference frame design are summarized in Table 1. Briefly, the frame’s height was 46.0 mm, the diameter at the ventricular inflow was 29.0 mm and the diameter at the aortic outflow was 38.0 mm. The strut width and thickness (0.25 mm, 0.45 mm) were in line with literature.6,13 The frame was modeled with Timoshenko beam elements (B31 in the Abaqus nomenclature) and WELD connectors were used at the interpolation points between adjacent cells (Fig. 1e). A mesh sensitivity analysis was performed on the device radial force during the crimping and expansion phases. The final stent mesh counted 2670 elements. The material properties of the Nitinol alloy were taken from literature24 (Table 2).

Figure 1
figure 1

Valve geometry. (a) unwrapped configuration; (b) profile of the frame of a CoreValve 29 mm; (c) unwrapped configuration with offset; (d) expanded valve configuration, (e) detail of the valve strut. In red the geometrical frame parameters under investigation: X1 is the diameter at 4 mm above the ventricular inflow, X2 is the height of the first row of cells.

Table 1 Frame reference adopted geometrical parameters.
Table 2 Nitinol material parameters for modeling the frame behavior24.

Patient-Specific Aortic Root Model

Three-dimensional aortic root models were retrospectively reconstructed from anonymized pre-operative MSCT images of 12 patients, who underwent TAVI in two European centers (Rigshospitalet in Copenhagen, Denmark; Erasmus medical Center in Rotterdam, The Netherlands). Selected patients (3 males, 9 females), between 74 and 89 years old, were considered representative of the TAVI population (e.g., patients with calcified tricuspid or bicuspid aortic valve, heavy calcified valve or sub-annular calcifications). All patients had comparable annular diameters ranging from 23.0 to 26.0 mm. More details of the selected patients are reported in Table 3. Three-dimensional aortic root models consisted of the left-ventricular outflow tract (LVOT), the sinotubular junction and calcified aortic leaflets and ascending aorta (Fig. 2a). Linear elastic models rather than a more realistic but complex hyperelastic anisotropic description were adopted to describe the aortic root tissues. Modelling parameters were derived in a previous study31 with an iterative process of back-calculations using pre and postoperative MSCT of 39 patients. The aortic wall had an elastic modulus of 2 MPa and a uniform thickness of 2 mm. For the aortic leaflets, elastic modulus of 0.6 MPa and uniform thickness of 1.5 mm was adopted.3,10 Calcifications were modeled using a stiffer elastic material with perfect plasticity (E = 4 MPa, ν = 0.3, yield stress = 0.6 MPa).29,31 Spring elements were added at each node of the aortic wall to incorporate the impact of surrounding structures in the model.

Table 3 Patients selected for the study.
Figure 2
figure 2

Computer simulation strategy. (a) Patient-specific 3D reconstruction of the aortic root and the region of the atrioventricular conduction system (in red the 3 anatomical landmarks identified using pre-operative CT images); (b) positioning of the crimped device within the aortic root; (c) deployment of the device using FE computer simulation; (d) extraction of contact pressure output within the region of the atrioventricular conduction system; (e) CFD simulation and predicted regurgitant jets. CT = computed tomography. NCC = noncoronary cusp. RCC = right-coronary cusp.

In each 3D aortic root model, the region of the atrioventricular conduction system was identified as previously described.29 Briefly, the inferior border of the membranous septum was identified in the pre-operative CT images through three dedicated landmarks (Fig. 2a). Two of these landmarks were selected at the beginning and at the end of the inferior border of the membranous septum, namely p1 and p3, with p1 closer to the non-coronary cusp (NCC) and p3 closer to the right coronary cusp (RCC). An additional point (p2) was selected in between to better track the course of the inferior border of the membranous septum as this is often not a straight line. Starting from these anatomical landmarks, the region of the atrioventricular conduction system was defined by the area between the inferior border of the membranous septum (extended towards the RCC by a 25° angle) and the plane 15 mm below the annulus (Fig. 2a). The model of the patient-specific anatomy was pre-processed using TAVIguide (FEops, Gent, Belgium).

Computer Simulation Strategy and Output Variables

Finite-element (FE) simulations were performed using Abaqus/Explicit (v2017, Dassault Systemes Simulia Corp, Johnson, RI). The computer simulation strategy can be subdivided in three steps. In the first step, the device was crimped into the catheter (6 mm diameter), using a cylindrical surface.

In the second step, the device was positioned coaxially within the aortic root (Fig. 2b), and the catheter was retracted leading to the expansion of the device. The default general contact with finite sliding between all the surfaces was used, assuming a coefficient of friction of 0.7 between the valve frame and the aortic model. In all patients, the device was implanted at approximately 4 mm below the aortic annulus (Fig. 2c). In order to evaluate the risk of occurrence of new conduction abnormalities, the maximum contact pressure and the contact pressure index (i.e., the percentage of this region of interest subjected to contact pressure) within the region of the atrioventricular conduction system were extracted (Fig. 2d). The contact pressure index was calculated according to the following formula:

$$\frac{{Area_{{{\text{ROI}}\left( {{\text{where}}\;CP > 0} \right)}} }}{{Area_{\text{ROI}} }} \times 100,$$
(1)

where ROI is the region of interest for the atrioventricular conduction system and CP is the contact pressure.

As third step, a static computational fluid-dynamics (CFD) simulation was performed using the ico-FOAM solver available in OpenFOAM (OpenFOAM, v2.1.1, OpenCFD Ltd., Bracknell, UK). Ico-FOAM is a transient solver for laminar flow. Briefly, the deformed geometries (aortic root and device) were extracted from the FE analysis and a skirt was added to the TAVI device. The fluid domain was extended with cylindrical flow extensions of 8 diameters at the inlet (aortic side) and 3 diameters at the outlet (ventricular side). The extended flow domain was discretized using hexahedral elements and refined within the region of interest (i.e., the lumen within the rigid aortic root wall). The blood was modeled as Newtonian fluid (ρ = 1060 kg/m3, μ = 3.5 × 10−3 Pa s). No slip condition was described at the aortic wall. A fixed transaortic pressure difference of 32 mmHg was imposed to calculate the inlet flow and the location and the velocity of regurgitant jets (Fig. 2e). The imposed pressure difference value is the average of the post-operative end-diastolic transaortic pressure difference retrospectively observed in a sample of 20 patients.8 This value is in line with the post-operative end-diastolic transaortic pressure difference observed by Sinning et al.32 in a cohort of 146 patients (33.8 ± 12.3 mmHg). This analysis was previously validated for prediction of paravalvular AR after implantation of a CoreValve System.8

Results from finite-element and fluid-dynamics simulations were post-processed using the TAVIguide software.

Experimental Design

The central composite design (CCD) was applied to determine the best combination of the frame parameters to reduce conduction abnormalities and paravalvular AR after TAVI. Given the fact that the inflow portion of the device frame plays a crucial role in the anchoring and apposition to the aortic root as well as in the above-mentioned clinical outcomes, we selected two geometrical parameters in this region of the frame. These two parameters are: the valve diameter at 4 mm above the ventricular inflow (X1, mm) and the height of the first row of cells at the ventricular inflow (X2, mm) (Fig. 1b). Initial value for X1 is 29.0 mm and for X2 is 8.0 mm. The initial design space was limited by X1 ± 10% and X2 ± 20%. The parameter levels are coded as − α (low axial level), − 1 (low level), 0 (central point), + 1 (high level) and + α (high axial level), with \(\alpha \;{\text{equals}}\;\sqrt 2\). Independent variables and their levels for the CCD used in the initial design space are shown in Table 4. The designs of the frame obtained with the different combination of the geometrical parameters (X1 and X2) based on the initial design space are shown in Fig. 3. A qualitative analysis of the effects of the two parameters on the outputs showed that high X1 and X2 parameters were associated with low values of the outputs. Detailed results obtained from the analysis in the initial design space can be found in Online Appendix 1. Based on the effects of the two selected geometrical frame parameters on the outputs, subsequent refinement of the design space was performed. Independent variables and their levels for the CCD used in the refined design space are shown in Table 5.

Table 4 Tested levels of experimental variables X1 and X2 in the initial design space.
Figure 3
figure 3

Frame designs obtained with the experimental variables X1 and X2 from the initial design space. Geometrical parameters X1 and X2 are represented in red.

Table 5 Tested levels of experimental variables X1 and X2 in the refined design space.

Response Surface Methodology

A second-order polynomial response surface model (RSM) was used to fit the predicted outputs that are the maximum contact pressure and the contact pressure index (FE outputs) and the paravalvular AR (CFD output). As the study aim was to minimize the outputs predicted from the implantation of each device in all patients, their value averaged over all patients was considered. The fitting was based on the minimization of the residual error measured by sum of squared deviations between the actual and the estimated response. The second-order polynomial equation has been assumed as:

$$Y = B_{0} + \mathop \sum \limits_{i = 1}^{2} B_{i} X_{i} + \mathop \sum \limits_{i = 1}^{2} B_{ii} X_{i}^{2} + \mathop \sum \limits_{i < j}^{2} B_{ij} X_{i} X_{j} \pm \varepsilon_{\text{r}}$$
(2)

with Y the value of one output parameter averaged over the 12 patients, B the coefficients, X the experimental variables and εr the error associated with the model. Only the most significant terms from the polynomial equation were selected using the Exhaustive Search method.

The quality of the model was assessed by coefficients of determination: R2, adjusted R2, predicted R2, and residuals. R2 represents a pure correlation between measured and predicted values and is indicative of the response variation explained by the model. Adjusted R2 is used to compare the explanatory power of the model, and its value increases only if an added term improves the model. The predicted R2 assesses the predictive power of the model for new observations. This coefficient was calculated by systematically removing each observation from the data set, estimating the regression equation and determining how well the model predicted the removed observation.2 Residual plots present the difference between the actual and the predicted values and were used to assess whether the coefficients predictions were biased. All analyses were performed in Isight (Dassault Systemes Simulia Corp, Johnson, RI).

Optimization Strategy

The Nonlinear Programming by Quadratic Lagrangian (NLPQL) was adopted as optimization algorithm, under the assumption that the objective functions are continuously differentiable.30 The three objective functions were rescaled to have a value between 0 and 1 (Eq. (3)) and combined into a single-objective function using a weighted sum method (Eq. (4)):

$$y_{i} = \frac{{x_{i} }}{{{ \hbox{max} }(x)}}$$
(3)
$$\mathop \sum \limits_{i = 1}^{3} w_{i} O_{i} \left( {X_{1} ,X_{2} } \right)$$
(4)

In Eq. (3) xi is a value of one output and yi is the corresponding rescaled value.

In Eq. (4), wi is the weight of each objective and Oi are the objectives (RSM functions). In this study, the single-objective function built to minimize the maximum contact pressure (O1), the contact pressure index (O2) and the paravalvular AR (O3) was:

$$0.5 O_{1} \left( {X_{1} ,X_{2} } \right) + 0.5 O_{2} \left( {X_{1} ,X_{2} } \right) + 1.0 O_{3} \left( {X_{1} ,X_{2} } \right)$$
(5)

The weight factors accounted for the fact that maximum contact pressure (O1) and contact pressure index (O2) are both predictors of new conduction abnormalities.

Lastly, the performance of the device with the optimal parameters was compared with the performance of the initial device in each patient. The one-tailed Wilcoxon rank test in SPSS (version 22.0, IBM Corporation, New York) was performed to verify whether the improvement in the predicted outcomes was statistically significant. The predicted maximum contact pressure, contact pressure index and paravalvular AR are reported as mean (standard deviation).

Figure 4 summarizes the workflow of the method used in this study.

Figure 4
figure 4

Pipeline that has been followed in this study.

Results

In total, 16 devices were created in this study: 9 devices based on the initial design space and 7 additional ones based on the refined design space. Each device was implanted in 12 patients (FE simulation) and, subsequently, a CFD simulation was performed for each patient to derive the paravalvular AR (Fig. 4). Therefore, this study counted 192 FE simulations and same amount of CFD simulations.

Average time for FE simulation was 53 min on a cluster with 16 computing cores (@ 4.0 GHz and 63.0 GB RAM for each node). All CFD simulations were performed on 2 cores (@ 3.2 GHz and 15.2 GB RAM for each node). Average time was 3 h and 50 min.

Goodness of Fit

The CCD was used to provide good fitting with a second-order polynomial function. RSM polynomial models for maximum contact pressure (O1), contact pressure index (O2) and paravalvular AR (O3) are respectively:

$$O_{1} = 3.57 - 0.08 X_{1} - 0.01 X_{2}^{2}$$
$$O_{2} = 70.38 - 3.92X_{2} - 0.02 X_{1}^{2}$$
$$O_{2} = 0.23 + 4.15X_{2} - 0.11X_{1} X_{2}$$

The coefficient of determination R2, adjusted R2 and predicted R2 of the models are reported in Table 6.

Table 6 Quality of the RSM models.

The coefficient of determination R2 for all models was higher than 0.89, which indicates a good approximation of the actual function at the design points used for its construction (Fig. 5a). The predicted R2 [0.74–0.86] was in reasonable agreement with the adjusted R2 [0.86–0.93], indicating good predictive power of the model. Residual plots (Fig. 5b) indicated that the coefficients prediction was not biased.

Figure 5
figure 5

(a) Plot of the actual and predicted responses and (b) of the actual and residual errors at each design point (black dots). Residuals are defined as the difference between the actual and the predicted responses.

The models have high R2, adjusted R2, predicted R2 and random distribution of residual errors. These results indicate a good fit of the experimental data and high precision in predicting maximum contact pressure, contact pressure index and paravalvular AR. Therefore, the polynomial models were used for further analysis.

Effects of the Geometrical Parameters

Two-dimensional contour plots of the RSM models are shown in Fig. 6. High values of maximum contact pressure were observed for low X1 and X2. The diameter at the ventricular inflow of the device (X1) influences its profile. Small diameters determine a sharp profile of the frame, while large diameters make the profile rounder (Fig. 7a). Therefore, small diameters are associated with local spots of contact with high maximum contact pressure. In addition, short cells (X2) at the ventricular inflow of the frame increase the radial force, contributing to high maximum contact pressure on the aortic wall at the region of the ventricular inflow of the device.

Figure 6
figure 6

Contour plot of the response surface models of the maximum contact pressure, contact pressure index and paravalvular AR, respectively from the left to the right. AR = aortic regurgitation.

Figure 7
figure 7

Representative examples of the effect of the geometrical parameters on maximum contact pressure and contact pressure index (a), and paravalvular AR (b). At the top, a schematic representation of the CCD. (a) A device with low X1 and low X2 (A) results in high maximum contact pressure and contact pressure index. Whereas, a device with high X1 and X2 (B) results in low contact within the region of the atrioventricular conduction system. (b) A device with low X1 (C) results in high paravalvular AR, whereas high X1 (D) prevents paravalvular AR. AR = aortic regurgitation. CCD = central composite design.

The RSM model of the contact pressure index showed a comparable behavior as for the maximum contact pressure. High values of contact pressure index were observed for low X1 and X2. The diameter (X1) and, therefore, the profile of the frame, strongly influences the level at which the device anchors to the aortic annular region. Large diameters (round profile) favor a higher anchoring level compared to small diameters (sharp profile). As the contact pressure index is a measure of the area of contact between the device and the aortic wall within the region of the atrioventricular conduction system, large diameters (high anchoring level) reduce the area of contact and are associated with low values of contact pressure index. On the contrary, the devices with small diameter (sharp profile) often anchor at the level of the atrioventricular conduction system, resulting in high contact pressure index value (Fig. 7a). Also, short cells (X2), associated with higher radial force, favor the expansion of the frame, contributing to an increase in contact and, hence, the contact pressure index.

With respect to the paravalvular AR, large diameters (X1) and short cells (X2) favor a better apposition of the device to the aortic root, reducing paravalvular AR (Fig. 7b). However, the diameter seems to have a larger influence on the output compared to the height of the cells.

Optimization Results

Figure 8 reports the optimization history with NLPQL algorithm for the three objectives O1, O2, O3. The orange line indicates the position where the minimum is located (# Run 13). The starting point corresponds to the initial design parameters: X1 = 29.0 mm and X2 = 8.0 mm. The optimal parameters are: X1 = 31.9 mm and X2 = 9.6 mm. The optimized device is shown in Fig. 9.

Figure 8
figure 8

Optimization history of the objectives-function. The location of the minimum is indicated by the orange line.

Figure 9
figure 9

Optimal frame design.

Figure 10 reports the maximum contact pressure, contact pressure index and paravalvular AR obtained in each patient after implantation of the device with the initial and the optimized design.

Figure 10
figure 10

Comparison of performance between the optimized and initial designs. Continuous grey lines represent a decrease in maximum contact pressure, contact pressure index or paravalvular AR in the same patient. Dotted grey lines represent an increase in maximum contact pressure, contact pressure index or paravalvular AR in the same patient. Mean (standard deviation) are represented by the red dot and the blue bar respectively.

The optimized device was associated with lower maximum contact pressure and contact pressure index in all patients. The mean value of maximum contact pressure associated with the optimized device was almost threefold lower compared to the initial device (0.33 vs. 0.88 MPa). Similarly, the mean value of contact pressure index was twofold lower (10% vs. 19%). The standard deviation slightly decreased as well with the optimized device. With regard to paravalvular AR, the average value decreased from 6.9 to 4.9 ml/s with the optimized device. Although in some patients paravalvular AR slightly increased with the optimized device, the Wilcoxon rank test confirmed that improvement in maximum contact pressure, contact pressure index as well as paravalvular AR is statistically significant (p value was found ≤ 0.001 and ≤ 0.05 for the contact pressure-related parameters and for AR respectively).

Discussion

This proof-of-concept study presents an optimization workflow based on patient-specific computer simulation to improve the TAVI device performance in reducing the predicted paravalvular AR and conduction abnormalities after TAVI. The central composite design along with the NLPQL algorithm were adopted to optimize two geometrical parameters of a CoreValve-like device: the diameter at 4 mm above the ventricular inflow and the height of the first row of cells. The effect of the design parameters on the objectives was studied using RSM. Results suggest that large diameters and high height of the first row of cells favor a better apposition of the device to the aortic wall avoiding paravalvular AR, and lead to an anchoring level above the location of the atrioventricular conduction system preventing conduction abnormalities. The optimized device showed improved performance in all patients. Maximum contact pressure and contact pressure index that are predictive for conduction abnormalities lowered with the optimized device. Predicted paravalvular AR was also reduced in 83% of the patients (10/12). Furthermore, good improvement in the design was achieved with only small changes to the geometry which will not change the mechanics or add difficulties in manufacturing.

Most of the existing optimization studies applied to stent geometries, employed computer simulation to improve mechanical properties (stress distribution, radial strength, fatigue resistance) or hemodynamic properties (flow disturbances, wall shear stress). However, optimization was based on mechanical tests performed on solely unit stent models,1 or simplified 3D stent models12,17,20 neglecting the interaction with the arterial tissues. Recently, Putra et al.28 adopted computational fluid dynamics simulation to optimize a simplified and idealized two-dimensional stent geometry under the influence of the vessel wall deformation. Li et al.18 accounted for the interaction of the stent with simplified plaque and artery in the optimization of a stent life prediction and stent expansion. Recently, Bosi et al.,5 combined patient-specific modeling with optimization techniques to tune the material parameter of the aortic root in their TAVI simulation workflow.

Though, to the best of our knowledge, this is the first study that adopts full three-dimensional patient-specific models to optimize stent geometrical parameters of a TAVI device, minimizing the predicted clinical complications in the average population. With this approach, the performance of the device can be evaluated in models resembling anatomical variability among patients eligible for TAVI, (aortic root variability, amount and location of calcifications, tricuspid or bicuspid aortic valve) before entering the clinical phase.

This proof-of-concept study shows that computer simulations based on patient-specific anatomies and population-based parameters on a representative TAVI population can be used to optimize the design of a TAVI device reducing the costs and the time of the design and development phase of a new device, as well as the level of risk during the clinical studies.

This is a valuable approach with the potential to be extended to medical devices in general. The objectives can also be extended to any other clinical outcome as long as computational modelling and simulation can accurately predict those outcomes. In this study, the optimized device is quite similar to the CoreValve system, for which the strategy for predicting paravalvular AR and conduction abnormalities was previously validated.8,29,31 Due to the similarity between the CoreValve and the device variants modelled in this study, we believe the previous validation work remains valid for these new devices.

While in this study we focused on three key functional post-deployment variables (maximum contact pressure, contact pressure area and paravalvular regurgitation), this approach can be used to optimize any geometrical parameter that plays a relevant role in the performance of the device (i.e., stress distribution, durability of the device). Also, the amount of geometrical parameters under investigation can be increased.

Study Limitations

Although device optimization based on patient-specific computer simulation has many advantages in the stent design, this study has certain limitations. First, significant computational resources are required to perform this kind of study. A total of 16 devices were created and virtually implanted in 12 patients, resulting in 192 FEA simulations and same amount of CFD simulations. Additionally, iterative repositioning of the device was necessary in some cases, to reach comparable implantation depth of the device in each patient. In fact, the obtained optimal device is implantation depth-specific. Whether it remains the optimal solution at different implantation depths needs to be verified. Second, a limited number of patients was enrolled in the study. Despite the fact that only 12 patients were used, they were representative of the population eligible for TAVI, and assumed sufficient for a proof-of-concept study. In future, automatization of the entire process is planned, with inclusion of more patients to ensure optimal device performance. Last, this study focused on the prediction of three post-procedural outcomes, but other parameters that play an important role in the performance and durability of the device should be evaluated. For example, the crimping feasibility of those devices should be checked to ensure that the device is not damaged. A mechanical analysis of the stress distribution on the device frame and device leaflets should be performed to guarantee high performance and device durability. Also, mechanical factors that might affect the stress/strain distribution on the aortic wall to avoid negative outcomes (i.e., annulus rupture) should be included in the optimization study.

Conclusion

In this study, we proposed a framework to improve the design of a representative TAVI frame using patient-specific computer models able to predict post-operative outcomes. Results show that this approach can effectively evaluate and optimize the performance of a TAVI device on a virtual TAVI population. The proposed approach can help to speed up the initial design and development phase of new devices and lower the risk during clinical studies.