Introduction

Optical coherence tomography (OCT) was first established for cross-sectional retinal imaging in 1991 by a Massachusetts Institute of Technology team. Then, OCT was first applied in intracoronary imaging in humans and showed promising capability as a tool for diagnostic imaging for arterial wall pathologies in 2001. Since then, OCT/optical frequency domain imaging (OFDI), a catheter-based imaging modality, has been widely used in the clinical catheterization laboratory for the imaging of the coronary artery wall. With an axial resolution of approximately 10 μm [1], OCT is capable of allowing for the visualization of the coronary artery microstructure at a resolution that is 10-times better than that of intravascular ultrasound (IVUS). Thus, OCT has become one of the commonly used intracoronary imaging techniques during a percutaneous coronary intervention (PCI) procedure. OCT provides valuable information that can be used clinically to optimize stent deployment and avoid PCI-related complications [2]. Pre-procedural measurement of lumen and vessel dimensions can facilitate accurate stent sizing [3]. Moreover, pre-procedural assessment of the tissue characteristics of the target lesion can guide optimal treatment strategies. However, OCT has evolved without existing standards for the measurement and the interpretation of images. In this review, we provide a short description of a consistent approach to OCT analysis to assist both clinicians and researchers.

Quantitative measurements

Measurements should be avoided if motion artifacts are present or incomplete flushing of the intra-coronary blood. To provide accurate measurements, the image should be correctly calibrated for z-offset and refractive index. Because coronary artery disease often appears to be more extensive on OCT than on angiography, it is quite difficult to accurately define reference segments in some cases. However, in general, the proximal reference is defined as the site with the largest lumen area, proximal to a stenosis but within the same segment (usually within 10 mm of the stenosis with no major intervening branches). Similarly, the distal reference is the site with the largest lumen area distal to a stenosis but within the same segment (usually within 10 mm of the stenosis with no intervening branches). The reference segment can be used which of these three reference sites (proximal, distal, or average). Quantitative measurements from JACC IVUS Consensus Document have been normally adopted for OCT measurement [4].

Lumen measurements

Lumen measurements are accomplished using the interface between the lumen and the leading edge of the intima. The following parameters can be measured in each cross section (mostly automatically):

  • Lumen cross-sectional area (CSA): the area bounded by the luminal border.

  • Minimum lumen diameter: the shortest diameter through the center of the lumen.

  • Maximum lumen diameter: the longest diameter through the center of the lumen. This line does not have to be orthogonal to the line of minimum lumen diameter.

  • Lumen eccentricity: (maximum lumen diameter minus minimum lumen diameter) divided by maximum lumen diameter.

  • Percent area stenosis: (reference lumen CSA minus minimum lumen CSA)/reference lumen CSA. The reference lumen CSA is generally used average lumen CSA of both proximal and distal reference site.

Vessel (internal elastic membrane) measurements

Vessel and plaque areas are not usually measurable using OCT because OCT has limited depth of penetration, which diminishes the ability to visualize the internal elastic membrane at the cross section with a large plaque area. For reference segments in which the internal elastic membrane can be identified, can the vessel measurements be made. Because the leading edge of the internal elastic membrane is well delineated on OCT, OCT-based measurements can be used to determine the true histological atheroma area.

Stent measurements

Metallic stent struts have a strong reflection coefficient in relation to the light signal used in OCT. Therefore, stents appear as highly reflective surfaces and cast shadows on the vessel wall behind. The following stent-related parameters can be measured using OCT:

  • Stent CSA: the area bounded by the luminal border of the stent.

  • Minimum stent diameter: the shortest diameter through the center of the stent.

  • Maximum stent diameter: the longest diameter through the center of the stent. This line does not have to be orthogonal to the line of minimum stent diameter.

  • Stent symmetry: (maximum stent diameter minus minimum stent diameter) divided by maximum stent diameter.

  • Percent stent expansion: the minimum stent CSA compared with the predefined average reference area.

  • Stent malapposition: when the stent strut is determined to not be fully attached to the vessel wall by visual assessment, the position of a stent strut relative to the vessel wall should be measured by magnifying the individual stent strut to maximize accuracy. Because OCT can show only the luminal surface of the strut owing to the limited depth of penetration through the metal, strut thickness should be considered in evaluating stent apposition for each type of stent design. First, one cursor is positioned on the adluminal surface of the strut, followed by another cursor at the surface of the estimated vessel wall within the stent strut shadow (Fig. 1) [5]. It is important to note that the right position of the adluminal surface of the stent strut falls in the center of the blooming [6]. The measurement line should be as perpendicular to the strut and vessel wall as possible. Stent malapposition can be defined as a distance between the adluminal surface of the strut and the estimated vessel wall greater than the thickness of each stent strut. Some investigators have estimated the location of the adluminal surface of the strut by drawing a line from the surface of the blooming toward the artery wall.

    Fig. 1
    figure 1

    Assessment of strut apposition, and normal coronary artery. a, b The distance between the abluminal surface of the strut and the estimated vessel wall is measured, extending and joining the contours of the wall on either side of the strut shadow with the measurement line as perpendicular as possible to the strut and vessel wall. Stent malapposition is defined as a distance between the abluminal surface of the strut and the estimated vessel wall (a) greater than the thickness of each stent strut plus polymer (a′). c, d Normal coronary artery reveals three layers, comprising a high-intensity intima, homogeneous low-backscattering media, and heterogeneous high-intensity adventitia. The internal elastic membrane should exit between the intima and media while the external elastic membrane exits between the media and adventitia

  • Neointimal thickness: the distance between the endoluminal surface of the neointima and the luminal surface of the stent strut.

  • Uncovered strut: struts with a measured neointima thickness equal to 0 mm.

  • Percentage of uncovered strut: The number of struts without distinct overlying tissue, in which the luminal reflection of the strut surface is directly interfacing with the lumen, divided by the total number of analyzed stent struts.

Qualitative assessment

The normal coronary artery consists of three layers: the intima, consisting of the endothelium and collagen fibers; the media, consisting of smooth muscle cells and elastic fibers; and the adventitia, consisting of collagen fibers and surrounding fat. Because OCT measures the intensity of light returning from a tissue, tissues with a higher heterogeneity in the refractive optical index exhibit stronger optical scattering, and therefore, a stronger OCT signal. The intima in the normal arterial wall is usually represented by the high-intensity signal from the collagen fibers. The media appears as a homogeneous dark layer because of the presence of less collagen fibers and abundant smooth muscle cells and extracellular matrix. The adventitia is also represented by the high-intensity signal from the collagen fibers (Fig. 1).

Plaque morphology

Optical coherence tomography is capable of differentiating lipid-rich plaque from fibrous plaque. Because OCT uses near-infrared light and cross-sectional images are generated by measuring the echo time delay and intensity of light that is reflected or backscattered from the arterial wall, OCT can characterize tissue morphology by measuring the backscattered infrared light. In general, plaques can be characterized as fibrous, lipid rich, or calcified, according to the plaque characterization criteria developed by histology-validation studies [7,8,9] (Fig. 2).

Fig. 2
figure 2

Fibrous, lipid-rich, and calcified plaques. a Fibrous plaque appears as a homogeneous high-intensity tissue with a gradual signal attenuation. b Lipid-rich plaque appears as low-signal-intensity regions with diffuse borders. c Calcified plaque appears as low-signal-intensity areas with sharply delineated borders

Fibrous plaques

These plaques consist of bundles of collagen fibers, smooth muscle cells, and extra-cellular matrix such as proteoglycan, and appear as a high-signal-intensity tissue because there is much reflected OCT light signal returning from the collagen fibers. Fibrous plaques represent the majority of atherosclerotic lesions. In general, the large amount of collagen fibers does not allow the visualization of the internal and external elastic membrane.

Lipid-rich plaques

These plaques appear as low-signal-intensity regions with diffuse borders because of considerable scattering of light of approximately 1000-nm wavelength from lipid components [10]. Fibroatheroma, which contains a necrotic core, cholesterol crystals, and foam cells, is considered as a representative tissue of a lipid-rich plaque. It should be noted that it is difficult to distinguish between the necrotic core and foam-cell accumulation on a single cross-sectional OCT image because these two tissues have similar OCT attenuation coefficients [11,12,13]. It has been postulated that thin-cap fibroatheroma (TCFA), which is characterized by a large necrotic core with an overlying thin fibrous cap, measuring < 65 µm, is the precursor plaque composition of plaque ruptures [14]. For OCT analysis, TCFA is usually defined as a plaque with lipid content with a fibrous cap measuring < 65 µm (Fig. 3). The cap thickness should be measured at its thinnest region within the fibrous cap. However, it should be noted that imaging geometries with a small angle between the line of sight and the tangent to the lumen contour may lead to tangential signal dropout because the imaging beam cannot penetrate the vessel wall owing to strong scattering along an oblique line-of-sight (Fig. 3c). Therefore, such sections should be interpreted with care.

Fig. 3
figure 3

A representative image of thin-cap fibroatheroma. a OFDI image indicates signal-poor lesions with an overlying signal-rich band. b A magnified image of a. The minimum fibrous cap thickness was 50 µm. c A small angle between the line of sight (red line) and the tangent to the lumen contour (yellow line) could lead to a tangential signal dropout

Calcified plaques

These plaques appear as low-signal-intensity areas with sharply delineated borders. Because the dimensions of the individual particles of calcium hydroxyapatite are smaller than the wavelength of near-infrared light, there is little reflected light returning from these tissues. Thus, OCT images of dense calcium show weaker optical scattering and therefore, a lower OCT signal intensity.

In addition to fibrous, lipid-rich, and calcified plaques, the following morphology has been frequently found in patients with acute coronary syndrome on OCT.

Plaque rupture

A ruptured plaque is defined as a plaque with intimal tearing, disruption, or dissection of the cap. On injection of optically transparent crystalloid or radiocontrast media, these defects may have little or no OCT signal and may appear as cavities.

Erosion

Optical coherence tomography-derived erosion could be composed of intracoronary thrombus attaching to the luminal surface without detectable signs of fibrous cap rupture. However, it should be noted that the resolution of OCT may be insufficient for directly visualizing one layer of endothelial cells in the form of a 5-μm-thick cellular monolayer, although OCT has a higher resolution than any other imaging modality,

Calcified nodule

A calcified nodule, which also has the potential to develop into coronary thrombosis, is defined as a high-backscattering mass protruding into the lumen with a strong signal attenuation and an irregular surface [15] (Fig. 4). The calcified nodule often contains fibrin between the bony spicules, along with osteoblasts, osteoclasts, and inflammatory cells. It should be noted that the OCT image of a calcified nodule is similar to that of a red thrombus, although a calcified nodule usually coexists with a calcified sheet in the same segments.

Fig. 4
figure 4

A representative image of a calcified nodule. a OFDI image showing a high-backscattering protruding mass with an irregular surface, following a low-intensity area with a diffuse border (arrowheads) such as a red thrombus. b A magnified image of A

The other plaque morphologies can be evaluated using OCT.

Thrombus

A thrombus appears as an intramural mobile mass attached to the luminal surface or floating within the lumen. OCT can be used to classify a thrombus as a red or white thrombus. A red thrombus is a red blood cell-rich thrombus, which appears as an intramural mobile mass protruding into the lumen with high backscattering and attenuation (Fig. 5a). A white thrombus is a platelet-rich thrombus that is defined as an intramural mass with homogeneous backscattering and low attenuation (Fig. 5b). Because red thrombi could be misinterpreted as calcified nodules on OCT, the patient’s background history and lesion morphology (e.g., acute coronary syndrome, on hemodialysis) should be taken into accounts for their interpretation.

Fig. 5
figure 5

Representative image of a red and white thrombus, and microvessel. a Red thrombus appears as an intramural mobile mass protruding into the lumen with a high backscattering and attenuation. b White thrombus appears as an intramural mass with homogeneous backscattering and low attenuation. c Microvessel appears as a no-signal tubuloluminal structure without a connection to the vessel lumen in multiple contiguous cross sections

Cholesterol crystals

Cholesterol crystals on OCT might appear as a thin, linear region of high signal intensity within a lipid-rich plaque. However, further histological validation data would be necessary for the accurate detection of cholesterol crystals.

Microvessels

A microvessel appears as a no-signal tubuloluminal structure without a connection to the vessel lumen. These microvessels usually have a sharply delineated border and can be followed in multiple contiguous cross sections (Fig. 5c). Owing to the limited penetration depth and axial resolution of OCT, the capability of OCT to reliably detect tiny microvessels in vulnerable atheromatous plaques merits future research. Recanalization of organized thrombus can be identified as multiple large channels, described as having “honeycomb-like appearance”, “swiss cheese appearance”, or “lotus root appearance”.

Dissection

Optical coherence tomography is commonly employed to detect dissections after interventional stent implantation. The classification of dissections into five categories is recommended:

  • Intimal: limited to the intima or plaque, and not extending to the media.

  • Medial: extending into the media.

  • Adventitial: extending through the external elastic membrane.

  • Intramural hematoma: an accumulation of flushing media within the medial space, displacing the internal elastic membrane inward and external elastic membrane outward.

  • Intra-stent: separation of neointimal hyperplasia from stent struts, usually seen only after treatment of in-stent restenosis.