Introduction

Myocardial viability studies are important tools in the evaluation of patients with ischemic heart disease.1 These studies provide valuable information on the volume of both scar and hibernating tissues through evaluations of myocardial perfusion imaging and myocardial metabolism.1 However, the quantitative accuracy of myocardial examinations is affected by image blurring from respiratory motion and cardiac contraction.

Gating techniques, which divide the cardiac contraction and/or respiration into several phases, have been suggested to limit cardiac motion effects. In cardiac PET, electrocardiogram (ECG) gating, using 3-lead ECG monitoring of the patient, has become the standard in clinical routine.2 Traditionally, respiratory motion is detected using external markers, such as respiratory belts, infrared sensors or measurements of the nasal air temperature.2,3 However, respiratory external markers frequently require time-consuming calibrations4 and bear the risk of being inadequate throughout the acquisition due to changes in the respiratory baseline5 or malfunctions during the acquisition.6 Moreover, their use may be uncomfortable for patients. Therefore, there is a growing interest in obtaining the respiratory signal without the use of external devices and with as little patient instructions as possible. In this context, data-driven methods, which derive the respiratory signal directly from the acquired PET data (i.e., self-gating methods) or through the use of MR navigators in combined PET/MR systems, are promising alternatives to the use of external markers.2,7,8,9,10,11,12

However, gating approaches have the disadvantage of using only fractions of the acquired data, thus leading to increased noise levels in the resulting images.13 This is particularly critical for myocardial PET-imaging, where corrections for both respiratory and cardiac motion are desired.14 For this reason, motion compensation (MoCo) approaches have generated substantial interest. To date, two main approaches have been proposed: the reconstruct-transform-average (RTA) method, which relies on co-registrations of gated image-reconstructions,15 and the motion-compensated image reconstruction, which is based on the inclusion of the motion information in the system matrix during the reconstruction.16,17,18 Despite promising results,19 these methods are not yet widely adopted in clinical routine due to complex setup procedures that involve user-defined reconstruction algorithms, which challenges a comparison to the results obtained in clinical settings.

Here, we propose a data-driven, projection-based respiratory MoCo (DPR-MoCo) approach for PET data acquired in PET-only, PET/CT or PET/MR systems. The proposed approach models the motion of each event in projection space and performs the correction event-by-event in the PET-listmode data, prior to the image reconstruction process.

Materials and Methods

The proposed DPR-MoCo technique was implemented and validated through analytical simulations and phantom experiments and subsequently tested in clinical patients who underwent myocardial viability examinations.

Analytical Simulations

A 4D extended cardiac-torso (XCAT) phantom20 was used for the analytical simulations based on the Siemens Biograph mMR PET/MR system. In contrast to an earlier approach by Tsoumpas and Cal-Gonzalez21,22 the analytical simulation in our study generated a listmode file instead of gated sinograms (Supplementary File 1). Photon attenuation, random and scatter coincidences, as well as Poisson noise, were included in the simulations to mimic a realistic patient acquisition. Three different parametric models of cardio-respiratory motion were included in our simulations: (i) a 2 cm superior-inferior 1-dimensional respiratory motion (0.2 Hz); (ii) a 2 cm superior-inferior and 1 cm anterior-posterior respiratory motion (0.2 Hz); and (iii) the respiratory motion described in (ii) combined with a cardiac rate of 1 bps (Figure 1A). Note that the cardiac motion in (iii) was simulated, but not accounted for in the corrections. The following standardized uptake values (SUV) were used in both simulations: soft tissue = 2, liver = 7, myocardium = 18, lumen = 5, ribs = 3.

Figure 1
figure 1

A Top—illustration of the simulated XCAT phantom (reference image used for the analytical simulations). Bottom—motion models included in the simulations: (i) 1D axial motion of 2 cm; (ii) 2D motion in the axial direction (2 cm) and in the anterior-posterior direction (1 cm); (iii) the respiratory motion described in (ii) combined with cardiac beating, 1 beat per second. Note that the cardiac motion in (iii) was simulated but not considered in the corrections. B Top—schematic design of the experimental phantom used for the validation of the data-driven respiratory motion detection. Bottom—CT image of the phantom illustrating the myocardial compartment and the three plaque-type lesions. C Flow-chart of dual-tracer PET/MRI protocol. Patients underwent simultaneous acquisitions of PET-based myocardial perfusion imaging employing 13N-NH3 and cine-acquisitions in the MR system. Myocardial viability was assessed through simultaneous acquisitions of PET employing 18F-FDG and rest and late-gadolinium enhancement imaging protocols

Five respiratory gates were created from the simulated listmode data. In the simulated dataset (iii), cardiac motion was included in the simulation but not considered in the corrections. The reconstruction of the simulated data was performed using the Software for Tomographic Image Reconstruction (STIR).23

Phantom Experiments

Two phantom studies were performed on a Siemens Biograph mCT PET/CT system at a collaborating site in Münster (Germany) (Figure 1B).24 Three spherical plaque-type lesions of 36 mm, 31 mm and 18 mm3 were attached to the myocardium inside the phantom (Figure 1B). The studies were acquired either with or without phantom motion (No-Motion). The No-Motion acquisition was acquired for 10 minutes, while the study with motion effects was acquired over 30 minutes. The latter experiment simulated cardiac contractions every 1.9 seconds with a respiratory cycle of 4.6 seconds, as described by Cal-Gonzalez et al.22 The myocardium-to-lumen ratio in the experimental phantom acquisition was 5-to-1. A respiratory signal obtained using a respiratory belt (Anzai AZ-733V) was used for validation of the data-driven respiratory detection.

Patient Studies

Seven patients (Table 1) undergoing myocardial viability examinations prior to revascularization procedures were enrolled and followed a combined PET/MR imaging protocol using a fully integrated PET/MR system (Siemens Biograph mMR, Siemens Healthineers, Erlangen, Germany).

Table 1 Patient demographics

All patients fasted for 4 hours prior to the 18F-FDG examination in order to switch the patients to glucose metabolism. All patients were administered Acipimox (250 mg) to stimulate the uptake of the 18F-FDG. Patients underwent dynamic acquisitions of myocardial perfusion imaging employing 13N-NH3 and metabolism studies employing 18F-FDG in a combined PET/MR imaging protocol (Figure 1C). The study was approved by the institutional review board (IRB No.: 1832/2016).

The proposed motion compensation was evaluated on PET-emission data from the last 10 min of the acquisitions (13N-NH3: (10-20) min, 18F-FDG: (30-40) minutes) to ensure a high tracer-uptake in the myocardium, as suggested in guidelines for myocardial metabolism imaging by the American Society of Nuclear Cardiology (ASNC).25 All acquisitions were reconstructed using DIXON-attenuation correction maps in the clinical reconstruction tools (OSEM 3i21s, 4mm Gaussian filter).26

Data-Driven, Projection-Based Respiratory Motion Compensation (DPR-MoCo)

Standard clinical routine assessments involve image reconstruction of the acquired PET-data without any motion limiting post-processing (Figure 2A). The proposed DPR-MoCo technique involves two steps, (1) extraction of the respiratory signal and respiratory gating of the PET-images; (2) extraction of motion vector fields (MVF) and projection based respiratory MoCo.

Figure 2
figure 2

Data-driven motion detection. A Standard clinical reconstruction set-up employs the acquired listmode file with all stored events (E1, E2, …, EN) during the acquisition. The resulting reconstruction has respiratory motion embedded in images, thus, degrading the image-quality. B The proposed DDG extracts the respiratory signal from the listmode file, followed by a respiratory gating of the acquired data

In brief, in step (1) the respiratory signal was extracted from the PET-listmode file and used for an amplitude data-driven gating (DDG) of the obtained PET-list data (see subsection: DDG) (Figure 2B). In step (2), the MVFs were obtained from non-rigid, projection-based co-registrations of the myocardium for all respiratory gated images (Figure 3A) (see “Extraction of the motion vector fields”). Following their extraction, the MVF’s were used to correct all individual events in the acquired listmode data (Figure 3B), thus providing a data-driven, projection-based motion compensation (DPR-MoCo) (see “Projection-based respiratory motion compensation”).

Figure 3
figure 3

Definition of the projection-based motion vector fields (MVF) and the DPR-MoCo method. A Respiratory motion vector fields (MVF) are extracted from forward-projections of segmentations of the myocardium obtained for each respiratory phase (see Figure 2B). B The extracted MVFs are used to correct for the respiratory motion directly in the listmode file. An additional normalization listmode file is created to compensate for changes in the normalization when moving the individual prompts. A fully-respiratory motion compensated image is reconstructed using the vendor provided software

Respiratory motion detection (DDG)

To ensure accurate detection of the respiratory signal, we propose to extract the respiratory signals from the myocardium only. This was achieved by using a bounding-box that covered the myocardial outline plus 2.5 cm in each direction and forward projecting the bounding-box to projection space (sinogram space). In this study, we used a center-of-mass-based assessment of the count-rates obtained within the bounding box for 100 ms long time-intervals (Eq. 1).

$$ {\text{Center }}\,{\text{of }}\,{\text{mass }}\,\left( t \right) \, = \, \frac{{\mathop \sum \nolimits_{i} i \cdot T(i,t)}}{{\mathop \sum \nolimits_{i} T(i,t)}} $$
(1)

where T is the histogram for the time segment t and i the slice number of the myocardial slice evaluated. The extracted respiratory signal was filtered using a moving average filter (0.3 seconds) to remove the 1Hz overlay signal introduced by the cardiac contractions (Figure 2B). Following filtering, the respiratory signal was used for an 4-bin amplitude-based gating (10-95-percentile of the respiratory range) of the PET listmode data and subsequent reconstruction in the vendor provided reconstruction toolbox (e7-tools, Siemens) 27 (Figure 2B).

Extraction of the motion vector fields

We propose to extract the MVF’s using projections of the gated images, to limit the risk of artifacts in the motion compensation (Figure 3A). The MVFs were extracted using three steps: first the myocardium was segmented for all respiratory gated images, which improved the definition of the MVF by suppression of the activity in the sub-diaphragm regions28 (Figure 3A). Then the segmented images were forward projected by using a Siddon’s ray tracing algorithm, to obtain ideal, noise-free, gated projections from which the respiratory MVFs were calculated (Figure 3A). These ideal projections were obtained using the forward projector available in the STIR-toolbox.23 Third, the sinogram-based MVFs were obtained from non-rigid co-registrations of the projected data using the Medical Imaging Registration Toolbox (MIRT)29 (Figure 3A). All gates were co-registered to the end-expiratory breath-hold position using a sum-of-square difference assessment.

Projection-based respiratory motion compensation

The extracted respiratory MVFs were used to create a motion compensated listmode (LM) file (Figure 3B). Motion compensation was achieved by applying the defined MVF’s to each event in the original listmode file. To ensure correct normalization of the data, the ratios of the normalization of the original and the motion-compensated positions were calculated and stored in an independent pseudo-LM file (Figure 3B). The respiratory MoCo sinograms were obtained from a combined binning of these two listmode files, followed by reconstruction using the vendor provided reconstruction tools.

Reconstruction Transformation Average (RTA-MoCo)

The DPR-MoCo method was validated against an image-based RTA-MoCo, in which the four amplitude-based reconstructions were co-registered to end-expiratory breath-hold position, as used for the DPR-MoCo.15 The image-based MVFs used for the RTA-MoCo were obtained in the MIRT software using the same segmentations as employed for the DPR-MoCo method, thus ensuring comparable results.

Data Analysis

A total of four different image-reconstructions were evaluated for the simulations, the phantom study and the patient examinations: reconstructions of the acquired data without any motion compensation (No-MoCo), a reference gate (Gated), an RTA based reconstruction (RTA-MoCo) as well as the DPR-rMoCo images (Table 2). In case of the simulated data and phantom acquisitions, we also included the reference images without any motion (No-Motion). In addition, two ECG-gated reconstructions were performed for the patient datasets, using the acquired data (No-MoCo-ECG) and the DPR-rMoCo (DPR-ECG). The ECG-gated datasets PET datasets were compared to the cardiac MR functional assessment of the myocardium using Medis MR Suite/Qmass toolbox (Medis, Leiden, the Netherlands).

Table 2 Motion detection and compensation techniques employed for the acquired PET-data (phantom acquisitions and patient examinations) and their respective characteristics

A total of 10 figures-of-merit (FoM), consisting of both mathematical and clinical evaluations (Cedars-Sinai QPS/QGS 2013 version, Cedars-Sinai Medical Center, Ca, USA), were evaluated (Table 3). The mathematical FoM included: (1) the target-to-background ratios of the left ventricle (TBRLV). Here the myocardium was segmented using a 41% SUV-max assessment, while the background was calculated from a spherical VOI (2.5 cm3) inserted in the left atrium. (2) The noise in the images, which was assessed using the coefficient of variation (CoVliver) in a 15 cm3 spherical VOI inserted in the liver (3). The axial translations measured in the myocardium for the experimental phantom and patient acquisitions. The following FoM were used for the clinical evaluation: the lumen of left ventricle, the scar-size (18F-FDG data), the total perfusion defect (TPD) (13N-NH3 data), end-systolic volume (ESV), end-diastolic volume (EDV), the stroke-volume (SV) and the ejection fraction (EF).

Table 3 Figures of merit (FoM) for the evaluation of the proposed DPR-MoCo method

Statistical Analyses

The Gated, RTA-MoCo and DPR-MoCo images were compared to the No-MoCo reconstructions using one-way analyses of variance (1-way ANOVA) for all FoM’s. The statistical analyses were corrected using a Dunnet correction to compensate for the multiple comparisons.30

Results

Analytical Simulations

Figure 4 shows selected coronal views of the reconstructed images of the XCAT phantom with corresponding line profiles for different motion-patterns introduced in the simulations (1D, 2D). Upon visual inspection of the reconstructed images the DPR-MoCo and RTA-MoCo yield improved image-quality, with higher contrast and similar levels of noise. Upon quantitative evaluation, DPR-MoCo and RTA-MoCo images have TBRLV and CoVliver values comparable to the No-Motion images (Table 4). The No-MoCo images had lower TBRLV values, while the single-gate reconstructions presented much higher CoVliver values (Table 4).

Figure 4
figure 4

Simulated XCAT phantom. A Comparison of DPR-MoCo, RTA-MoCo, Gated and No-MoCo images for the simulations with 1D axial respiratory motion model. B Same images, but with a 2D axial + anterior/posterior motion model. The corresponding line profiles through the myocardium are depicted in the right side of the figure. A significant improvement in image contrast, while preserving noise properties, can be observed for the DPR-MoCo and the RTA-MoCo images. C Same images, but with a 2D axial + anterior/posterior + cardiac motion model. Note, the cardiac motion in (iii) was simulated but not accounted for in the corrections

Table 4 CoVliver and the TBRLV values obtained from images reconstructed with DPR-MoCo compared with the values obtained from images reconstructed without MoCo (No-MoCo), gated images (Gated), RTA-MoCo and acquisitions without motion (No-Motion)

Phantom Experiments

A high correlation (99% detection rate) was observed for the respiratory signal obtained using the proposed DDG and external markers (Figure 5). Both signals provided a strong respiratory peak in the Fourier spectrum for a frequency of 0.216 Hz, which corresponds to a respiratory signal with period of 4.6 seconds. The corresponding MVFs obtained for the segmented myocardium recovered 75% of the motion exerted to the myocardium in the axial direction (Table 4). Comparable noise levels were observed for the No-MoCo, RTA-MoCo and DPR-MoCo images (Table 4). Improved TBRLV were seen in the DPR-MoCo and RTA-MoCo images, when compared to the No-MoCo reconstructions (Figure 5). The TBRLV values for the DPR-MoCo and RTA-MoCo images were, however, not recovering the values obtained for the No-Motion images of the static phantom.

Figure 5
figure 5

Experimental phantom study. A No-Motion, B No-MoCo, C Gated, D DPR-MoCo and E RTA-MoCo. One of the inserted plaque-type lesions is clearly visible in the images shown in panels A, C, D and E. F Line profile through the myocardium shows resolution recovery for the motion limited data when compared to the static images. G Respiratory signal obtained using the DDG (red) and external markers (blue), which shows an excellent overlap between them

Patient Studies

DPR-MoCo: static myocardial examination

Significant respiratory motion was observed for the patient acquisitions, with translations of up to 31 mm in the axial direction (Table 5). Improved resolution recoveries were observed when applying motion limiting (Gated) or motion compensation (DPR-MoCo, RTA-MoCo) techniques (Figure 6, Table 6), resulting in contour sharpening of TPD and better scar delineations on the polar-maps. Assessments of the image-quality reveal statistically significantly increased noise-levels for the Gated and RTA-MoCo images, while the DPR-MoCo have slightly better image-quality across the patients, although the differences were not statistically significant (Table 6). The individual assessment of CoVLiver, left ventricular volume and CBRLV for the different image reconstruction methods are provided in the supplementary data (Figure SM1). The reconstructed images for two patients are also provided in the supplementary data (Figures SM2 and SM4).

Table 5 Respiratory translations (mm) observed for the respiratory gated images
Figure 6
figure 6

Comparison of image-quality obtained for the 13N-NH3 (top) and 18F-FDG (bottom) scans for a patient #7. (A) No-MoCo, (B) Gated, (C) DPR-MoCo and (D) RTA-MoCo images are shown. Improved contrast recovery is seen for the Gated, DPR-MoCo and RTA-MoCo images, as improved delineation of the left-ventricular lumen (white arrow) and sharpening of the myocardial wall (green arrow)

Table 6 Mathematical FoM obtained for the patient analyses averaged across all seven patients

DPR-MoCo + ECG-gating

Diastolic and systolic phases of the DPR-MoCo and standard LM-based reconstructions showed marked differences when the polar maps were analyzed using the clinical evaluation software (Figure 7, Supplementary Figures 3, 5). The patient displayed in Figure 7 had confirmed 3-vessel disease as visualized by a general reduced radio-tracer uptake (13N-NH3 and 18F-FDG) in the myocardial PET images and the corresponding polar maps. A slight increase in the tracer-uptake is seen for the anterior part of the myocardium when applying the DPR-MoCo.

Figure 7
figure 7

Comparison of End-Diastolic and End-Systolic images obtained for patient #7, Top: images and polar maps from the last 10 minutes of the 13N-NH3 scan. Bottom: images and polar maps from the last 10 min of the 18F-FDG scan. No-MoCo ECG reconstructions are shown on the left while DPR-MoCo ECG-reconstructions are shown on the right

Grouped analyses of the ESV, EDV, SV and EF revealed improved delineations of the left-ventricular wall after motion compensation across the patients, shown as reduced ESV and EDV values. Furthermore, the DPR-MoCo images revealed higher correlation with the cardiac MRI extracted EF values (Table 7).

Table 7 Functional values obtained for the ECG-gated reconstructions of the No-MoCo and DPR-MoCo image reconstructions as well as the cardiac MR functional parameters, averaged across patients

Discussion

In this study, we describe the feasibility of a data-driven, projection-based motion detection and correction method (DPR-MoCo) for cardiac PET imaging. The DPR-MoCo method was evaluated in simulations, phantom studies and clinical patients. In all cases, we observed a clear improvement in the reconstructed PET images when employing the proposed DPR-MoCo, with increased TBRLV values and similar noise levels to the ones obtained in No-MoCo images (Table 4). Furthermore, relative changes between No-MoCo and DPR-MoCo reconstructions were reported for several parameters of clinical interest, such as the ESV, the EDV, and the EF.

The performance of the DPR-MoCo was evaluated by means of analytical simulations of the XCAT phantom. We demonstrated a good performance of the proposed DPR-MoCo (Figure 4 and Table 4). A high number of acquired counts was simulated (500 million counts per gated sinogram), leading to reconstructed images with very low levels of noise (Table 4).

The proposed method (DPR-MoCo) was validated also in phantom studies by comparing the data-driven extracted respiratory signals to the respiratory signals obtained from using a respiratory belt (Anzai AZ-733V) (Figure 5). Here, nearly perfect correlations between the respiratory signals were observed for the DDG technique, which translated into improvements of reconstructed images employing either gating or MoCo (Figure 5, Table 4). As expected, image noise levels did increase for the single-gate reconstructions when compared to image reconstructions using the full-acquired data: No-MoCo, DPR-MoCo and RTA-MoCo (Tables 4 and 6).

Further, the clinical effect of the DPR-MoCo was assessed also in seven patients. We observed relative improvements in the visual quality of the reconstructed images after applying the DPR-MoCo (Figure 6, Table 6). However, these improvements were lower than expected and not statistically significant (Table 6). Nevertheless, we observed a higher impact of our proposed DPR-MoCo approach on the 13N-NH3 scans, due to increased respiratory translations introduced by the simultaneous acquisition of MR-cine sequences that require multiple breath-holds (Figure 1, Table 5).31

An important benefit of the proposed DPR-MoCo technique is the ability of performing ECG-gating of the listmode data including correction for respiratory motion. Evaluations of the ECG-gated PET-images employing the DPR-MoCo and No-MoCo datasets provided notably more consistent EF measures when compared to the cardiac MR measures following DPR-MoCo. The mitigated difference between the cMR and DPR-MoCo was introduced by enhanced delineations of the myocardial wall (Figure 7, Supplementary Figures 1-5) owing to the correction of the respiratory blur in the PET images. The relative improvement in the image-quality was correlated with the motion observed for the myocardium across patients. In cases with significant respiratory motion, such as the 13N-NH3 studies (introduced by multiple breath-holds, required by the cMR), we obtained higher image quality improvement after applying the MoCo techniques.

The proposed method has certain advances in terms of translation into clinical practice, as the method can be fully automatized. The current implementation offers a semiautomatic approach, which only requires manual entries in terms of segmentation of the myocardium. Automatic delineations of the myocardium can further improve the proposed technique to become fully automated, which currently is work in progress.

Despite being tested in a clinical setting with high count statistics, it is believed that this technique also works for other studies with low-count statistics such as coronary PET-studies for two reasons.32 First, the DDG requires high contrast between the myocardium and the lungs, a requirement that is fulfilled for most tracers following the first-pass phase. The second reason is that the MVFs are extracted using ideal sinograms, obtained from projections of the respiratory gated images. These ideal projections contain is stable in terms of noise and are, thus, not affected by even low count statistics as can be expected in some applications.

Limitations of the DPR-MoCo Method

The sinogram-based definition of the MVFs is inherent to the proposed DPR-MoCo approach and can be argued as its major limitation since motion in image space is only partly reflected as motion in sinogram space. In this work, we assumed that the respiratory motion in the heart can be approximated by a simple motion pattern in 2D (axial and anterior-posterior), and therefore, the loss of information due to moving from image-to-sinogram space does not yield significant degradation of the accuracy of the motion model. However, for more complex motion patterns, e.g., cardio-respiratory motion, the proposed method will be challenged. Further investigations of the range of applicability of the DPR-MoCo approach are required to understand its pitfalls.

Another limitation is the use of only four bins in the amplitude-based respiratory gating, which is known to possess a significant amount of intra-frame motion.27 The effect of the DPR-MoCo is expected to be improved if using a higher number of respiratory bins, however, the increase in noise in the gated reconstructions may also reduce the accuracy of the resulting DPR-MoCo images. Further studies are required to find the optimal settings of this technique.

Finally, in view of the small size of the analyzed patient-cohort, a follow-up study related to the clinical evaluation of the proposed method based on a larger patient cohort has been initiated as well.

New Knowledge Gained

In this study, we demonstrated that a projection-based motion modeling and compensation technique can provide an image-quality of the heart comparable to already established methods.

Conclusion

This study proposes a data-driven MoCo technique for PET imaging. The proposed DPR-MoCo method applied to respiration and yields significant improvements in the image quality and quantification of the myocardial wall for ECG-gated reconstructions. To date, the methods have been validated for simple motion models and well-defined sources of tracer distribution in cardiac PET imaging.