Abstract
The number of shoulder arthroplasty procedures has increased dramatically in recent years, with the primary indication being osteoarthritis (OA). Thus, morphology and subchondral bone changes associated with OA may be important factors to consider when choosing a replacement component. For surgical treatment, many implant options exist and survivability is often dependent on patient age, activity level, and progression of OA. In the placement of these replacement components, patient-specific guides now exist to improve component positioning, with the goal to improve long-term survivability by ensuring that intra-operative placement meets component design.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Shoulder arthroplasty procedures have been shown to reduce pain and improve joint function when non-surgical options have been exhausted [1, 2]. The primary indication for anatomic total shoulder arthroplasty is glenohumeral osteoarthritis (GHOA) [3•], with secondary indications including inflammatory arthritis and post-traumatic arthritis. A retrospective study of hemi-arthroplasty and total shoulder arthroplasty (TSA) performed in the USA from 1993 to 2008 found GHOA as the primary indication for both hemi-arthroplasty and TSA procedures (hemi 44 %; TSA 77 %) [4]. A systematic review of 3853 TSAs performed between 1976 and 2007 also found GHOA to be the primary indication in 74 % of cases [3•].
The prevalence of shoulder replacements has been seen to have a dramatic increase in the past two decades [4, 5]. Additionally, with an ageing population and the typical age of patients ranging from 65 to 84 years [4], the prevalence of revision surgeries is also projected to increase dramatically [5]. The cost of revision surgery is much greater than that of a primary procedure and is associated with increased risks due to advanced age, potential for infection, and poor bone quality, resulting in a significant burden on health-care systems. Improving the survivorship of shoulder replacement components could dramatically reduce this burden.
A particularly difficult problem to treat in GHOA is glenoid bone loss. In cases of acquired bone loss, anatomic total shoulder arthroplasty may be conducted in association with bone grafting or with the selection of an augmented glenoid component [6•]. Understanding the pathoanatomy of glenoid bone loss due to GHOA and the post-operative biomechanics of the available implant solutions may assist with improving our understanding of implant survivability.
Glenoid bone loss
Cartilage degradation and bone remodelling in osteoarthritis, with associated soft tissue alterations, are believed to lead to glenoid bony erosion. This results in pain and loss of joint function due to bone-to-bone contact and osteophyte formation. It is unknown whether soft tissue abnormalities lead to the progression of osteoarthritis (OA) or if OA and the associated joint inflammation lead to soft tissue changes. At the time of arthroplasty, static subluxation of the humeral head is frequently observed, which increases the complexity of the procedure [7, 8]. A static subluxation is usually identified with a shift of the articulation, which implies an alteration of normal joint kinematics and contact forces, which leads to glenoid morphological changes characterized by wear and subchondral bone remodelling.
Glenoid morphology
Walch et al. classified glenoid erosion into five types (A1, A2, B1, B2, and C) based on erosion morphology [9•]. Types A1 and A2 classify symmetric glenoid erosions to minor and major degrees, respectively. Type B1 exhibits subluxation with a narrowed joint space but without bone loss. Types B2 and C refer to glenoids with posterior (asymmetric) bone loss, with type C being a dysplastic glenoid retroverted more than 25° [9•]. Walch type B2 biconcave glenoids are further characterized as having two articular facets: the posterior neoglenoid, which represents the region of new glenohumeral contact, and the anterior paleoglenoid, the remaining anterior native glenoid articular surface [10•, 11].
The B2 arthritic triad (glenoid biconcavity, acquired glenoid retroversion, and humeral head posterior subluxation) presents one of the most challenging and controversial osteoarthritis patterns when managed with joint replacement [12•, 13]. The recent literature has suggested that in these biconcave asymmetrically eroded glenoids, maximum bone loss occurs in a posteroinferior direction with the erosion being directed towards the 8 o’clock position in a right shoulder [14–17]. The surgical options in these cases include hemi-arthroplasty, eccentric reaming, glenoid bone grafting, augmented glenoid components, or reverse shoulder arthroplasty [6•, 12•]. Eccentric reaming with a standard glenoid component aims to restore the native glenohumeral joint position by reaming down the anterior non-eroded ‘high side’, approaching the level of the neoglenoid, in order to achieve full backside contact of the component. This results in a reduction of available bone stock and medializes the native surface. In an effort to preserve a glenoid bone stock, implant manufacturers have developed glenoid components augmented with deeper posterior backsides, adapted to characteristic B2 morphology (Fig. 1).
Recent clinical [19•, 20, 21], experimental [22–26], and computational [27–31] studies addressing asymmetric posterior glenoid erosion have focused on eccentric reaming and/or retroversion correction. It has been suggested that retroversion correction greater than 15° cannot be achieved with eccentric reaming alone [26, 32]. When correcting version using eccentric reaming, bone grafting, or augmented glenoid components, it is important to recognize the degree of component medialization required for full backside contact of the glenoid component. In order to accommodate the design of some augmented glenoid components, glenoid vault peg perforation may occur [29, 30]. Perforation may also occur when using standard components with eccentric reaming in patients with large degrees of acquired retroversion and may be one contributing factor to the reported maximum version correction. Peg perforation, however, may not be the most relevant clinical endpoint as it does not necessarily portend a poor outcome.
Although retroversion correction is often the main focus in the management of arthritis-related glenoid erosion, the degree of inclination correction for normal joint function and component stability is also important. A relatively large number of studies have addressed the improved accuracy of retroversion measurements using 3D versus 2D techniques [30, 33–35]; however, few studies have addressed 3D measurements of glenoid inclination. Habermeyer et al. [36] presented a glenoid inclination classification system based on the degree of inclination observed in 2D radiographs. All types are relative to ‘the coracoid base line’ (a vertical line drawn from the lateral aspect of the coracoid). Type 0 refers to a glenoid inclination line parallel to the coracoid base line, type 1 the coracoid base line and glenoid inclination line that intersect below the glenoid rim, type 2 the coracoid base line and glenoid inclination line that intersect between the inferior and centre aspects of the glenoid, and type 3 the coracoid base line and glenoid inclination line that intersect above the coracoid base. This classification system is useful in determining the degree of inclination correction to reduce the potential for glenoid component loosening due to ‘the rocking-horse effect’ [37]. However, in cases of asymmetric erosion of type B2 glenoids having biconcave surfaces, 2D radiographs may not capture the degree of inclination correction required on either the anterior or posterior side.
Osteoarthritic glenoid morphology has a dramatic effect on surgical technique and may result in varying clinical outcomes [20]. Although patients with asymmetric glenoid erosion are assumed to have suboptimal outcomes, a recent study of 196 patients with symmetric glenoid erosion compared to 148 patients with asymmetric glenoid erosion found similar clinical improvements in both groups with a 98 % revision-free survivorship at a mean of 51 months [19•]. All patients had a similar treatment method with all-polyethylene glenoid components and press-fit humeral components. Similar studies with long-term results and studies assessing the results of various implants (reverse TSA, hemi-arthroplasty, TSA, augmented TSA) would provide valuable insight into the effectiveness of these devices in the management of GHOA.
Subchondral bone changes
The altered glenohumeral contact mechanics of the osteoarthritic joint results in extensive bone remodelling of the glenoid subarticular bone. In regions of increased bone-to-bone contact, bone remodelling may result in dense sclerotic bone areas. This alteration in the density of underlying bone may alter the load transfer from glenoid components, affecting component stability.
In symmetrically eroded glenoids, humeral contact is uniformly distributed across the glenoid articulation. Bone remodelling, as cartilage degrades in these cases, results in uniform glenoid bone density in all regions of the glenoid [38, 39]. In asymmetrically eroded glenoids, humeral articular contact is concentrated on the posterior neoglenoid region. This region typically represents around 50 % of the total glenoid surface area [15], thereby concentrating the region of load transfer and contact over a decreased area. The result of this is substantially denser and less porous bone in the neoglenoid [38, 39]. In contrast, the paleoglenoid no longer experiences joint loading, as the humeral head is subluxated posteriorly and articulates with the neoglenoid. In response, the paleoglenoid bone becomes substantially less dense and more porous. In asymmetrically eroded biconcave glenoids, it has been suggested that preservation of the dense sclerotic bone in the neoglenoid region may reduce implant migration and early implant failure [11, 38]. All of these factors are important to consider when assessing implant fixation in the B2 glenoid.
Arthroplasty options
The degree of osteoarthritis-related bony erosion, patient age, activity level, and soft tissue deficiencies are all factors affecting the choice of surgical procedures for joint replacement. Younger active patients with minimal bone loss may be considered for hemi-arthroplasty procedures [41]. Total shoulder arthroplasty is utilized for advanced arthritis with deformity that is correctable with reaming and a standard implant. For patients with substantial glenoid bone loss, uncorrectable with reaming, total shoulder arthroplasty with a posteriorly augmented glenoid component may be indicated. In patients with a high degree of bone loss that is not correctable or with substantial joint instability, reverse total shoulder arthroplasty (RTSA) is often used. To address these difficult deformities, patient-specific instrumentation has gained traction recently [42–47], with companies beginning to explore patient-specific implants as surgical options to replace population-based implant designs (Stanmore Implants, Zimmer, Mobelife). The following sections address the specific arthroplasty options.
Total shoulder arthroplasty
The survivorship of total shoulder arthroplasty (TSA) at 15 years has been shown to be greater than 85 %; however, glenoid component loosening remains a leading cause of component failure [48]. In a large systematic review comparing metal-backed (MB) (n = 1571) and all-polyethylene (n = 3035) components (mean follow-up of 5.8 years in MB and 7.3 years in all-polyethylene), 77 % of revisions in all-polyethylene cases were due to component loosening [49•]. A comprehensive review of the modes of glenoid component failure reported mechanical failure, inadequate bone surface preparation and component seating, suboptimal cement techniques, component rim loading, aseptic loosening and osteolysis, and complex pathoanatomy as common reasons for component failure [50]. The authors suggest patient selection, patient counselling, technique optimization, restoration of the normal glenohumeral relationship, component design and the identification of pathologic erosions, and malalignment as ways to minimize the risk of glenoid component loosening. Their review also showed that round-backed and all-polyethylene glenoid components using peg fixation performed better than flat-backed, metal-backed, or keeled glenoid components.
In cases of more complex pathoanatomy, alternative glenoid components have been developed to account for bone deficiencies. A recent study reported on non-standard glenoid component designs for bone deficiency [21]. This study compared three components: an angled keel, an extra-thick (6 mm) component, and an augmented metal-backed component. The authors reported a large percentage of unsatisfactory results, with the effectiveness of these components compromised by loosening. In other cases of bone deficiency, such as asymmetric erosion, eccentric reaming and use of a standard component are preferred by some authors to account for acquired retroversion [6•, 10•, 12•, 32]. However, as previously mentioned, glenoid retroversion greater than 15° cannot be corrected by eccentric reaming alone [26, 32], making augmented glenoid components an attractive choice to preserve bone in patients with greater retroversion.
Augmented glenoid components account for posterior glenoid bone loss by using a thickened posterior step or wedge to address the missing bone (Fig. 1). These designs act to preserve underlying bone and correct acquired retroversion by minimizing bone removal. However, existing designs orient the augment along the central axis of the implant, assuming that maximum bone loss is directed towards the 9 o’clock position in a right shoulder. As previously stated, the recent literature has found bone loss to occur in a posteroinferior direction towards the 8 o’clock position [14–17]. This may result in significant bone removal in order to achieve full backside seating [27, 28]. Furthermore, when using either augmented or standard components in patients with asymmetric erosion, significant differences occur in the bone density and quality of underlying bone [27, 28], which may further complicate the stability and fixation of glenoid components. Both eccentric reaming and augmented components are best utilized in patients with mild to moderate glenoid erosion. In more severe cases, with erosion progressing past the centre of the glenoid, alternative approaches should be considered. Standard components will likely result in peg perforation after eccentric reaming, and current augmented component designs do not account for the increased region of bone loss [15].
Hemi-arthroplasty
Hemi-arthroplasty is an option for younger patients with sufficient glenoid bone stock. Replacement of the proximal humerus with a humeral component and conservative concentric glenoid reaming, without implantation of a glenoid component, removes the risk of glenoid component failure due to loosening [13, 51]. This procedure has also been found to be effective in the treatment of GHOA and may be an effective option in younger patients in order to retain bone for future revision surgeries [13, 51–53]. In type B2 biconcave cases, conservative reaming to restore the glenoid to a single concavity, with the use of a humeral hemi-arthroplasty, may be an effective treatment for the arthritic triad [13]. Unfortunately, disadvantages of the hemi-arthroplasty include glenoid-sided arthritic pain, progressive glenoid erosion, and recurrent instability.
Reverse total shoulder arthroplasty
With the FDA’s approval of reverse total shoulder arthroplasty (RTSA) in 2004, RTSA has become increasingly popular in the management of rotator cuff tear arthropathy, showing improved outcomes as compared to hemi-arthroplasty [2]. Short, intermediate, and long-term outcomes of RTSA have shown improvements in joint function, in both younger patients (less than 60) [54] and older patients (greater than 60) [55–58]; however, younger patients tend to report lower satisfaction [54] compared to older patients [55–58]. Reverse total shoulder arthroplasty provides an effective solution for older patients with increased bone loss, and soft tissue imbalance, who require a more constrained prosthesis [12•].
One recurrent issue with RTSA is scapular notching [57, 59]. Scapular notching is controversial as some studies have demonstrated no effect on outcomes while others have shown decreased outcome scores with greater degrees of notching. A classification system for scapular notching was devised by Sirveaux et al., who describes the extent of notching in relation to the inferior screw [58]. This classification describes four levels of scapular notching: 0 = no notch, 1 = small notching short of the inferior screw, 2 = medium notch reaching the inferior screw, and 3 = large notch extending beyond the inferior screw. To reduce scapular notching, a bone graft between the baseplate (with an extended post) and glenoid has been suggested to lateralize the joint [60]. This procedure, termed bony-increased offset reverse shoulder arthroplasty (BIO-RSA), has been suggested in the treatment of asymmetric glenoid erosion with a biconcave glenoid, in order to reconstruct the joint line and reduce scapular notching [10•, 61, 62]. Long-term clinical outcomes of this procedure have yet to be reported, and limitations in standard clinical imaging to assess bone graft resorption at the baseplate should be considered in this procedure [63].
Patient-specific options
The focus of current patient-specific literature in the shoulder has been on the increased accuracy of component placement with patient-specific positioning guides and 3D pre-operative templating. Improved glenoid component positioning accuracy, compared to traditional techniques, has been reported for standard and augmented components [47, 64–66] and in baseplate positioning in RTSA procedures [67]. These devices are possible due to advancements in medical imaging software, 3D printing technologies, and improved knowledge of the osteoarthritic joint. Many implant manufacturers now offer custom software with the use of 3D reconstructions of patient anatomy to assist in the computational pre-operative placement of arthroplasty components. This software also allows for the 3D printing of patient-specific drill guides to assist in drill placement intra-operatively.
Conclusions
Glenoid morphology and subchondral bone changes, as the result of progressive osteoarthritis, cause difficulty in the surgical management of the glenohumeral joint. The level of bony erosion and the corresponding bone adaptations, as well as patient age and activity level, should be considered when choosing joint replacements for patients with advanced OA. To improve previous suboptimal outcomes of glenoid component loosening, patient-specific guides may improve the accuracy of component placement, with the potential to improve the long-term survivability of these components.
References
Papers of particular interest, published recently, have been highlighted as: • Of importance
Norris TR, Iannotti JP. Functional outcome after shoulder arthroplasty for primary osteoarthritis: a multicenter study. J Shoulder Elbow Surg. 2002;11:130–5.
Leung B, Horodyski M, Struk AM, Wright TW. Functional outcome of hemiarthroplasty compared with reverse total shoulder arthroplasty in the treatment of rotator cuff tear arthropathy. J Shoulder Elbow Surg. 2012;21:319–23.
Papadonikoakis A, Neradilek MB, Matsen F. Failure of the glenoid component in anatomic total shoulder arthroplasty. J Bone Joint Surg Am. 2013;95(24):2205–12. A review of glenoid component failure in anatomic TSA.
Kim SH, Wise BL, Zhang Y, Szabo RM. Increasing incidence of shoulder arthroplasty in the United States. J Bone Joint Surg Am. 2011;93:2249–54.
Day JS, Lau E, Ong KL, et al. Prevalence and projections of total shoulder and elbow arthroplasty in the United States to 2015. J Shoulder Elbow Surg. 2010;19:1115–20.
Sears BW, Johnston PS, Ramsey ML, Williams GR. Glenoid bone loss in primary total shoulder arthroplasty: evaluation and management. J Am Acad Surg. 2012;20:604–13. A review of glenoid bone loss, classifications, and surgical management.
Gerber C, Costouros J, Sukthankar A, et al. Static posterior humeral head subluxation and total shoulder arthroplasty. J Shoulder Elbow Surg. 2009;18:505–10.
Walch G, Ascani C, Boulahia A, et al. Static posterior subluxation of the humeral head: an unrecognized entity responsible for glenohumeral osteoarthritis in the young adult. J Shoulder Elbow Surg. 2002;11(4):309–14.
Walch G, Badet R, Boulahia A, Khoury A. Morphologic study of the glenoid in primary glenohumeral osteoarthritis. J Arthroplasty. 1999;14(6):756–60. Glenoid erosion classification.
Denard PJ, Walch G. Current concepts in the surgical management of primary glenohumeral arthritis with a biconcave glenoid. J Shoulder Elbow Surg. 2013;22:1589–98. Management of primary glenohumeral osteoarthritis with a biconcave glenoid.
Walch G, Moraga C, Young A, Castellanos-Rosas J. Results of anatomic nonconstrained prosthesis in primary osteoarthritis with biconcave glenoid. J Shoulder Elbow Surg. 2012;21:1526–33.
Hsu JE, Ricchetti ET, Huffman GR, et al. Addressing glenoid bone deficiency and asymmetric posterior erosion in shoulder arthroplasty. J Shoulder Elbow Surg. 2013;22:1298–308. A review of asymmetric erosion and the surgical options.
Matsen FA, Warme WJ, Jackins SE. Can the ream and run procedure improve glenohumeral relationships and function for shoulders with the arthritic triad? Clin Orthop Relat Res. 2014;473(6):2088–96.
Churchill R, Spencer E, Fehringer E. Quantification of B2 glenoid morphology in total shoulder arthroplasty. J Shoulder Elbow Surg. 2015;24(8):1212–7.
Knowles NK, Keener J, Ferreira LF, Athwal GS. Quantification of the position, orientation, and surface area of bone loss in type B2 glenoids. J Shoulder Elbow Surg. 2014;24(4):503–10.
Beuckelaers E, Jacxsens M, Van Tongel A, De Wilde LF. Three-dimensional computed tomography scan evaluation of the pattern of erosion in type B glenoids. J Shoulder Elbow Surg. 2014;23:109–16.
Terrier A, Ston J, Larrea X, Farron A. Measurements of three-dimensional glenoid erosion when planning the prosthetic replacement of osteoarthritic shoulders. Bone Joint J. 2014;96-B:513–8.
Knowles NK, Osteoarthritis induced glenoid morphology and bone quality: an evaluation of augmented glenoid components 2015. Electronic thesis and dissertation repository. Paper 2752. http://ir.lib.uwo.ca/etd/2752/.
Hussey MM, Steen BM, Cusick MC, Cox JL, Marberry ST, Simon P, et al. The effects of glenoid wear patterns on patients with osteoarthritis in total shoulder arthroplasty: an assessment of outcomes and value. J Shoulder Elbow Surg. 2015;24(5):682–90. Clinical outcomes of TSA with various glenoid wear patterns.
Klein SM, Dunning P, Mulieri P, et al. Effects of acquired glenoid bone defects on surgical technique and clinical outcomes in reverse shoulder arthroplasty. J Bone Joint Surg Am. 2010;92:1144–54.
Cil A, Sperling JW, Cofield RH. Nonstandard glenoid components for bone deficiencies in shoulder arthroplasty. J Shoulder Elbow Surg. 2014;23:149–57.
Wang T, Abrams G, Behn A, et al. Posterior glenoid wear in total shoulder arthroplasty: eccentric anterior reaming is superior to posterior augment. Clin Orthop Relat Res. 2015;473(12):3928–36.
Iannotti JP, Greeson C, Downing D, et al. Effect of glenoid deformity on glenoid component placement in primary shoulder arthroplasty. J Shoulder Elbow Surg. 2012;21:48–55.
Iannotti JP, Lappin KE, Klotz CL, Reber EW, Swope SW. Liftoff resistance of augmented glenoid components during cyclic fatigue loading in the posterior-superior direction. J Shoulder Elbow Surg. 2013;22:1530–6.
Gupta R, Lee TQ. Positional-dependent changes in glenohumeral joint contact pressure and force: possible biomechanical etiology of posterior glenoid wear. J Shoulder Elbow Surg. 2005;14(1):105S–10.
Nowak DD, Bahu MJ, Gardner TR, et al. Simulation of surgical glenoid resurfacing using three-dimensional computed tomography of the arthritic glenohumeral joint: the amount of glenoid retroversion that can be corrected. J Shoulder Elbow Surg. 2009;18:680–8.
Knowles NK, Ferreira LM, Athwal GS. Augmented glenoid component designs for type B2 erosions: a computational comparison by volume of bone removal and quality of remaining bone. J Shoulder Elbow Surg. 2015;24(8):1218–26.
Kersten AD, Flores-Hernandez C, Hoenecke HR, D’Lima DD. Posterior augmented glenoid designs preserve more bone in biconcave glenoids. J Shoulder Elbow Surg. 2015;24(7):1135–41.
Sabesan V, Callanan M, Sharma V. Guidelines for the selection of optimal glenoid augment size for moderate to severe glenohumeral osteoarthritis. J Shoulder Elbow Surg. 2014;23:974–81.
Sabesan V, Callanan M, Sharma V, Iannotti JP. Correction of acquired glenoid bone loss in osteoarthritis with a standard versus an augmented glenoid component. J Shoulder Elbow Surg. 2014;23:964–73.
Hermida JC, Flores-Hernandez C, Hoenecke H, D’Lima D. Augmented wedge-shaped glenoid component for the correction of glenoid retroversion: a finite element analysis. J Shoulder Elbow Surg. 2014;23(3):347–54.
Clavert P, Millett PJ, Warner JJP. Glenoid resurfacing: what are the limits to asymmetric reaming for posterior erosion? J Shoulder Elbow Surg. 2007;16:843–8.
Ganapathi A, McCarron JA, Chen X, Iannotti JP. Predicting normal glenoid version from the pathologic scapula: a comparison of 4 methods in 2- and 3-dimensional models. J Shoulder Elbow Surg. 2011;20:234–44.
Hoenecke HR, Hermida JC, Dembitsky N, et al. Optimizing glenoid component position using three-dimensional computed tomography reconstruction. J Shoulder Elbow Surg. 2008;17:637–41.
Kwon YW, Powell K, Yum JK, et al. Use of three-dimensional computed tomography for the analysis of the glenoid anatomy. J Shoulder Elbow Surg. 2005;14:85–90.
Habermeyer P, Magosch P, Luz V, Lichtenberg S. Three-dimensional glenoid deformity in patients with osteoarthritis: a radiographic analysis. J Bone Joint Surg Am. 2006;88:1301–7.
Walch G, Boulahia A. Primary glenohumeral osteoarthritis: clinical and radiographic classification. The Aequalis Group. Acta Orthop. 1997;64:46–52.
Walch G, Young A, Boileau P. Patterns of loosening of polyethylene keeled glenoid components after shoulder arthroplasty for primary osteoarthritis. J Bone Joint Surg. 2012;94(2):145–50.
Simon P, Gupta A, Pappou I, et al. Glenoid subchondral bone density distribution in male total shoulder arthroplasty subjects with eccentric and concentric wear. J Shoulder Elbow Surg. 2015;24(3):416–24.
Knowles NK, Athwal GS, Keener JD, Ferreira LM. Regional bone density variations in osteoarthritic glenoids: a comparison of symmetric to asymmetric (type B2) erosion patterns. J Shoulder Elbow Surg. 2015;24(3):425–32.
Matsen F. The ream and run: not for every patient, every surgeon or every problem. Int Orthop. 2015;39:255–61.
Verborgt O, Vanhees M, Heylen S, et al. Computer navigation and patient-specific instrumentation in shoulder arthroplasty. Sports Med Arthrosc. 2014;22:e42–9.
Walch G, Vezeridis P, Boileau P. Three-dimensional planning and use of patient-specific guides improve glenoid component position: an in vitro study. J Shoulder Elbow Surg. 2015;24(2):302–9.
Hendel MD, Bryan JA, Barsoum WK, et al. Comparison of patient-specific instruments with standard surgical instruments in determining glenoid component position. J Bone Joint Surg. 2014; 2167–75.
Subramanya S, Herald J. Reverse shoulder arthroplasty with patient-specific glenoid implant positioning guides. Tech Shoulder Elbow Surg. 2014;15:122–9.
Verborgt O, De Smedt T, Vanhees M, et al. Accuracy of placement of the glenoid component in reversed shoulder arthroplasty with and without navigation. J Shoulder Elbow Surg. 2011;20:21–6.
Lewis GS, Stevens NM, Armstrong AD. Testing of a novel pin array guide for accurate three-dimensional glenoid component positioning. J Shoulder Elbow Surg. 2015;24(12):1939–47.
Bohsali KI, Wirth MA, Rockwood Jr CA. Complications in total shoulder arthroplasty. J Bone Joint Surg Am. 2013;95:563–9.
Papadonikolakis A, Matsen F. Metal-backed glenoid components have a higher rate of failure and fail by different modes in comparison with all-polyethylene components. J Bone Joint Surg Am. 2014;96(12):1041–7. A systematic review of metal-backed versus all-polyethylene glenoid components.
Matsen F, Clinton J, Lynch J, et al. Glenoid component failure in total shoulder arthroplasty. J Bone Joint Surg Am. 2008;90:885–96.
Matsen F, Lippitt S. Current technique for the ream-and-run arthroplasty for glenohumeral osteoarthritis. J Bone Joint Surg Essent Surg Technol. 2012;2:e20.
Gilmer BB, Comstock BA, Jette JL, et al. The prognosis for improvement in comfort and function after the ream-and-run arthroplasty for glenohumeral arthritis: an analysis of 176 consecutive cases. J Bone Joint Surg Am. 2012;94:e102.
Mercer D, Gilmer B. A quantitative method for determining medial migration of the humeral head after shoulder arthroplasty: preliminary results in assessing glenoid wear at a minimum of two years after hemiarthroplasty with concentric glenoid reaming. J Shoulder Elbow Surg. 2011;20(2):301–7.
Muh SJ, Streit JJ, Wanner JP, et al. Early follow-up of reverse total shoulder arthroplasty in patients sixty years of age or younger. J Bone Joint Surg. 2013;95(20):1877–83.
Young SW, Everts NM, Ball CM, et al. The SMR reverse shoulder prosthesis in the treatment of cuff-deficient shoulder conditions. J Shoulder Elbow Surg. 2009;18:622–6.
Wall B, Nové-Josserand L, O’Connor DP, et al. Reverse total shoulder arthroplasty: a review of results according to etiology. J Bone Joint Surg Am. 2007;89:1476–85.
Boileau P, Watkinson D, Hatzidakis AM, Hovorka I. Neer Award 2005: the Grammont reverse shoulder prosthesis: results in cuff tear arthritis, fracture sequelae, and revision arthroplasty. J Shoulder Elbow Surg. 2006;15:527–40.
Sirveaux F, Favard L, Oudet D, et al. Grammont inverted total shoulder arthroplasty in the treatment of glenohumeral osteoarthritis with massive rupture of the cuff. Results of a multicentre study of 80 shoulders. J Bone Joint Surg (Br). 2004;86:388–95.
Lévigne C, Boileau P, Favard L, et al. Scapular notching in reverse shoulder arthroplasty. J Shoulder Elbow Surg. 2008;17:925–35.
Boileau P, Moineau G, Roussanne Y, O’Shea K. Bony increased-offset reversed shoulder arthroplasty minimizing scapular impingement while maximizing glenoid fixation. Clin Orthop Relat Res. 2011;469:2558–67.
Mesiha M, Boileau P, Walch G. Technique for reverse total shoulder arthroplasty for primary glenohumeral osteoarthritis with a biconcave glenoid. J Bone Joint Surg Essent Surg Technol. 2013;3:e21.
Mizuno N, Denard PJ, Raiss P, et al. Reverse total shoulder arthroplasty for primary glenohumeral osteoarthritis in patients with a biconcave glenoid. J Bone Joint Surg Am. 2013;95:1297–304.
Ferreira LM, Knowles NK, Richmond DN, Athwal GS. Effectiveness of CT for the detection of glenoid bone graft resorption following reverse shoulder arthroplasty. Orthop Traumatol Surg Res. 2015;101:427–30.
Iannotti JP, Weiner S, Rodriguez E, et al. Three-dimensional imaging and templating improve glenoid implant positioning. J Bone Joint Surg. 2015;97(8):651–8.
Iannotti J, Baker J, Rodriguez E, et al. Three-dimensional preoperative planning software and a novel information transfer technology improve glenoid component positioning. J Bone Joint Surg. 2014;71:1–8.
Throckmorton TW, Gulotta LV, Bonnarens FO, et al. Patient-specific targeting guides compared with traditional instrumentation for glenoid component placement in shoulder arthroplasty: a multi-surgeon study in 70 arthritic cadaver specimens. J Shoulder Elbow Surg. 2015;24:965–71.
Levy JC, Everding NG, Frankle M, Keppler LJ. Accuracy of patient-specific guided glenoid baseplate positioning for reverse shoulder arthroplasty. J Shoulder Elbow Surg. 2014;23:1563–7.
Acknowledgments
Nikolas K. Knowles is supported in part by the Natural Sciences and Engineering Research Council of Canada and in part by a Transdisciplinary Bone and Joint Training Award from the Collaborative Training Program in Musculoskeletal Health Research at the University of Western Ontario.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Nikolas K. Knowles and Louis M. Ferreira declare that they have no conflict of interest.
George S. Athwal reports receiving grants from DePuy Mitek, Exactech, and Tornier during the conduct of this study.
Human and animal rights and informed consent
This article does not contain any studies with human or animal subjects performed by any of the authors.
Additional information
This article is part of the Topical Collection on Shoulder Arthroplasty
Rights and permissions
About this article
Cite this article
Knowles, N.K., Ferreira, L.M. & Athwal, G.S. The arthritic glenoid: anatomy and arthroplasty designs. Curr Rev Musculoskelet Med 9, 23–29 (2016). https://doi.org/10.1007/s12178-016-9314-2
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12178-016-9314-2