Introduction

Diagnostic discrimination of mediastinal tumors is crucial for planning appropriate treatment [1, 2]. Different pathological results lead to different treatments. For example, tumors, such as thymoma, are often treated with surgery, whereas lymphoma or metastatic lesions would be treated with chemotherapy or radiotherapy [3]. CT-guided percutaneous needle biopsy (PNB) has been widely performed in the diagnosis of mediastinal nodules [4, 5]. However, the conventional CT guidance has limitations in the lack of real-time monitoring and gantry tilting for a more accessible needle pathway to the target lesion [6, 7]. Along with the development of flat detector cone-beam CT (CBCT), a novel technique for TNB guidance recently emerged. It combines advanced needle path planning with 3D CBCT images. The CBCT systems offer the real-time visualization of TNB procedures and more flexibility in the orientation of the detector system around the patient compared with the traditional CT systems [8, 9]. Therefore, the PNB procedures assisted by advanced 3D needle guidance systems can be performed in a sterile workspace with flexible system angulation capability as well as instantaneous fluoroscopic feedbacks [10]. Indeed, owing to this system’s ability to facilitate more accurate and safer needle placement, it has already been shown to provide excellent diagnostic accuracy for biopsy of the lung lesions [2, 8, 9, 1119] (Table 1). However, there are few literatures about CBCT-guided percutaneous needle biopsy of mediastinal lesions [20]. This paper describes our preliminary experience with PNB for mediastinal lesions biopsy under the CBCT guidance system.

Table 1 Recent studies on C-am cone-beam CT-guided biopsy procedures in the field of chest intervention

Methods and materials

Patients

This retrospective study was approved by the institutional review board of The First affiliated Hospital of Zhengzhou University with waiver of patient informed consent. From January 2010 to January 2015, a total of 100 patients with 100 solid mediastinal lesions were retrospectively enrolled to undergo PNB procedures. 100 consecutive patients (65 males and 35 females, mean age 52.00 ± 11.87 years; age range, 24–80 years) with enhanced CT or PET-CT confirmed solid mediastinal lesions were retrospectively enrolled in this study. The lesion size was recorded as the maximum diameter in the image data by one chest radiologist (Han XW, 20 years of experience in image-guided PNB).

Image acquisition

3D CBCT images of the patients were acquired with a rotational angiographic system (Artis Zeego, 30 × 40 cm FD detector, Siemens Healthcare, Forchheim Germany). The resulting raw projection images were then automatically transferred to a workstation (Syngo X Workplace, Siemens Healthcare) for 3D volume reconstruction. As a result, the CBCT images were reconstructed with 1 mm thickness and presented in axial, sagittal, and coronal orientations. The time from the end of the data acquisition to the presentation of multiplanar images on the workstation ranged between 43 and 45 s.

Needle path planning and guidance procedure

As the second step, the needle path was planned on the same workstation using the commercially available software (Syngo iGuide, Siemens Healthcare). Figures 1, 2 and 3 demonstrated this procedure for a 2.50 cm lesion in middle mediastinum. Figure 4 demonstrated this procedure for a 3.50 cm lesion in posterior mediastinum. Figure 5 demonstrated this procedure for a 1.80 cm lesion in middle mediastinum. The reconstructed 3D volume was first loaded. In the orthogonal multiplanar images, the skin entry point and target lesion positions were manually selected and marked by a cross and a circle, respectively. A virtual path was then generated with its angulations and length calculated and displayed. All three multiplanar images were automatically aligned to the defined path to provide in-plane views (Fig. 1b–d). Then, the virtual path was then projected and superimposed onto the live fluoroscopic images and displayed on a dedicated live monitor (Fig. 2). First, the C-arm rotated to the Bull’s Eye View, where the C-arm was angulated in the way that the cross and the circle displayed on the live monitor completely matched and the central X-ray beam was aligned with the planned path (Fig. 2a). Thus, the skin entry point could be determined (Fig. 2a). The needle orientation was adjusted until both the tip and hub of the needle in the fluoroscopic image were superimposed and located at the center of the circle and the cross. Second, after the skin entry point and needle orientation were determined, the needle was advanced under fluoroscopy until the planned target lesion position was reached. The C-arm was rotated back and forth to two different angles subsequently to monitor the needle progression. These two angles provided lateral views (Progression View) of the planned needle path and helped to ensure that the needle was advanced along it (Figs. 2b, 5d). Third, a 3D scan was acquired to confirm the final position of the needle (Fig. 3).

Fig. 1
figure 1

55-year-old patient, preoperative CT showed a 2.50 cm middle mediastinal lesion (white arrow) (a). CBCT orthogonal multiplanar images with graphics showing planned needle path (yellow line) into target lesion (red circle) (b, c). The cross indicated the skin entry site, and the circle indicated the target lesion site. The needle position relative to the anatomical structures was displayed in 3D using the volume-rendering technique (d)

Fig. 2
figure 2

Real-time fluoroscopic images in Bull’s eye view (a) and progression views (b). The needle was advanced along the planned needle path (dotted line) from skin entry site (white cross) to target lesion site (white circle)

Fig. 3
figure 3

CBCT scan confirmed the needle position in multiplanar (ac) and volume-rendering (d) images

Fig. 4
figure 4

58-year-old patient, preoperative CT showed a 3.5 cm lesion in posterior mediastinum (a). CBCT images with graphics showing planned needle path (green line) into target lesion (green circle) (b). CBCT scan confirmed the needle position (c)

Fig. 5
figure 5

80-year-old patient, preoperative PET-CT showed a 1.8 cm lesion in middle mediastinum (a). CBCT orthogonal multiplanar images with graphics showing planned needle path (yellow line) into target lesion (red circle) (b, c). Real-time fluoroscopic images in progression views, the needle was advanced along the planned needle path (dotted line) (d). CBCT scan confirmed the needle position (e, f)

PNB technique

PNBs were performed by or under the supervision of one chest radiologist (Han XW, 20 years of experience in image-guided PNB). Prior to the procedure, previously acquired diagnostic contrast-enhanced chest CT or PET-CT was carefully reviewed to determine the most appropriate needle route (Figs. 1a, 4a, 5a). The mean length of time from acquisition of the diagnostic CT image or PET-CT to the biopsy procedure was 7.6 ± 5.3 days (range 1–30 days). The patient was placed in either supine or prone position according to location of the lesion, and local anesthesia was given (1 % lidocaine, ≤10 ml). A 16-gauge needle (Quick-Core, Cook Medical Inc., Bloomington, IN, USA) was advanced along the planed path under real-time fluoroscopy. For sampling, the stylet was removed from the guiding needle and replaced by a biopsy needle, and then approximately 1.0–2.0 cm sample tissue was taken. A pathologist was not present, and our criteria were that the total samples length was more than 1.0 cm to meet the needs of the pathological sections. After sufficient tissue samples were obtained, the coaxial introducer was removed. Thereafter, post-procedure CT images were acquired to identify procedure-related complications. The patients remained in the hospital for 24 h for observation after the procedure.

Lesion characteristics and procedural records

Operators documented lesion characteristics as well as the procedural records in our database. We recorded the lesion size, location in the mediastinum, and lesion feature (solid, cystic, or mixed solid and cystic). We recorded several factors during TNB; the patients’ positions during PNB, skin-to-target distance, the number of biopsies, the number of CT acquisitions, needle approach, total procedure time (defined as the duration from local anesthesia injection to the end of post-procedure CBCT). Technical success was defined as appropriate location of the coaxial needle within target nodules on CBCT images and adequate tissue sampling on visual inspection. We recorded the total coaxial introducer indwelling time and effective radiation exposure dose during the entire procedure (fluoroscopy dose and CBCT dose).

Pathological results, diagnostic accuracy

The pathological reports of biopsy specimens, surgical specimens, or follow-up images were reviewed to evaluate the final diagnosis of target nodules. Final diagnosis was confirmed in four ways. (1) If the patient underwent surgical resection, the pathological report decided the final diagnosis. (2) If the pathological result of biopsy demonstrated a malignant or a specific benign pathology, such as thymoma or sarcoidosis, it was accepted as the final diagnosis. (3) In the cases of non-specific benign pathology (negative for malignancy, chronic inflammation, etc.), follow-up CT helped to decide whether the lesion would be truly or falsely benign, or indeterminate. If the lesion decreased 20 % or more in diameter, we considered the final diagnosis as benign. (4) If a nodule of non-specific benign pathology did not show a sufficient interval decrease in size nor had follow-up images, its final diagnosis was defined as indeterminate. Interminate lesions were not included in the calculation of diagnostic accuracy. Any PNB-related complication, such as pneumothorax or haemoptysis, was also recorded. Pneumothorax was evaluated with post-procedural CBCT and follow-up chest radiographs during the hospitalization.

All data analyses were performed using the Excel 2010 (Microsoft, Redmond, WA) and SPSS software (version 13.0; SPSS, Chicago). Numeric data are reported as the mean ± standard deviation.

Results

Nodule characteristics, procedural records, and radiation doses

A detailed description of the lesion characteristics and procedure records is summarized in Table 2. The mean size of nodules was 4.4 ± 1.8 cm, ranging from 1.8 cm to 9.0 cm. Among the 100 nodules, 81 lesions were solid, 2 lesions were cystic, and 17 showed mixed solid and cystic features. 76 lesions were located in the anterior mediastinum, 12 lesions in the middle mediastinum, and 12 in the posterior mediastinum. The patients underwent TNB in the supine position in 85 cases and in the prone position in 15 cases. The number of biopsies and CBCT acquisitions were 2.6 ± 1.2 (range 1–6) and 3.0 ± 1.1 (range 2–8), respectively. The mean total procedure time was 11.7 ± 3.4 min (range 6–27 min). With regard to the mediastinal needle approach used, the parasternal approach was used for 78 lesions, the paravertebral approach for 12 lesions, and transpulmonary approach for 10 lesions. The technical success rate of TNB under iGuide CBCT virtual navigation system was 100.0 % (100/100). The total coaxial introducer dwelling time in our study was 8.2 ± 3.4 min. The overall procedural time ranged from 6 to 27 min, and the mean total procedure time was 11.7 ± 3.4 min, resulting in a mean effective exposure dose of 9.7 ± 4.3 mSv.

Table 2 Lesion characteristics and procedure records of flat detector cone-beam CT-guided percutaneous needle biopsy of mediastinal lesions

Pathological results, diagnostic accuracy, and complications

Detailed pathological results are listed in Table 3. Among the 100 nodules, 83 nodules were malignant, 12 benign, and 5 indeterminate lesions. The final diagnosis of malignancy was made on the basis of surgical pathology (n = 44), including the conventional open surgery (n = 24) and endoscopic surgery (n = 20), and specific malignant biopsy results (n = 39). The 83 malignant lesions consisted of thymoma (n = 22), non-Hodgkin’s lymphoma (NHL) (n = 17), thymic carcinoma (n = 11), metastases (n = 10), Hodgkin’s disease (HD) (n = 7), thyroid carcinoma (n = 5), malignant mesenchymal tumor (n = 4), squamous cell carcinoma (n = 3), fibrosarcoma (n = 2), and liposarcoma (n = 2). The final diagnosis of benign lesions was made on the basis of surgical pathology (n = 4), including the conventional open surgery (n = 2) and endoscopic surgery (n = 3), and specific malignant biopsy results (n = 7). The 21 malignant lesions consisted of chronic granulomatous (n = 3), fibroma (n = 3), mature teratoma (n = 1), sarcoidosis (n = 1), Schwannoma (n = 1), inflammation with necrosis (n = 1), and tuberculosis (n = 1). One patients’ initial biopsy result was fibrous connective tissue; the nodule was enlarged more than 20 % in size on follow-up chest CT after 2 months. TNB procedures were performed again, and the patient was confirmed with fibrosarcoma. The other three patients (one with HD and two with NHL) had completed the entire treatments, and tumor volumes were reduced more than 80 % compared with pretreatment. However, there were still persistence of soft tissue. Residual mediastinal tissue pathology results were coexists with fibrotic and/or necrotic tissue or thymic hyperplasia. Among them, two patients received the conventional open surgery and were confirmed with residual tumors. Another patent with NHL with high suspicion of tumor residue, TNB procedures were performed again, and the patient was confirmed with the residual tumor of NHL. Finally, the sensitivity, specificity, and accuracy of TNB of mediastinal lesions under the Siemens CBCT virtual navigation guidance were 95.1 % (79/83), 100 % (12/12), and 95.7 % (91/95), respectively.

Table 3 Pathologic outcomes of the patients who underwent PNB procedure with Siemens CBCT needle guidance system

Complications occurred in 5 of 100 procedures (5.0 %), all of which were pneumothorax (n = 3), subcutaneous hematoma (n = 1), and hemoptysis (n = 1). However, pneumothorax volume in every patient was below 5 %, and no specific treatment was given, as the patients were asymptomatic and clinically stable. Post-operative hemoptysis occurred in one patient, but the symptom was self-limiting and disappeared within 1 day. An old patient with giant mediastinal lesion (size 8.3 cm) and upper vena cava compression syndrome occurred subcutaneous hemorrhage after procedure. Puncture point was oppressed for 20 min, and bleeding was stopped.

Discussion

Imaging-guided PNB is regarded as an adequate technique for characterizing mediastinal lesions to be used before invasive surgical diagnostic procedures (mediastinoscopy, thoracoscopy, anterior mediastinotomy, and exploratory thoracotomy) [21]. CT-guided mediastinal lesion biopsy is well described in the literature with technical success ranging from 77 to 96.4 % [2224]. The technical success rate of mediastinal lesion biopsy under CBCT virtual navigation was 100 % with accuracy rate 95.7 %, which falls within the aforementioned range. The favorable outcome achieved in our preliminary study suggested that the CBCT-guided biopsies can be as safe and accurate as CT-guided procedures. This finding is not surprising, because CBCT imaging guidance primarily affects the needle pass, but tissue sampling itself is essentially comparable to CT-guided procedures. Moreover, The CBCT systems offer real-time visualization on TNB procedures and more flexibility in the orientation of the detector system around the patient compared with the traditional CT systems [25]. Busser et al. [26] demonstrated that the CBCT was the preferred modality over CT, irrespective of the level of operator experience, in terms of accurate needle placement in difficult guidance procedures requiring double-angulated needle paths. As the mediastinum is a complex body part comprising the heart, trachea, esophageal, and multiple vascular structures, elaborate needle placement is essential. The study suggested that the CBCT virtual navigation-guided biopsy may be an exceptionally well-suited procedure for lesions in the mediastinum.

Several sampling methods are available. The simplest is fine-needle-aspiration cytology, which uses 20- to 25-gauge needles and provides a cytological sample consisting of exfoliated cells. Another sampling system is core biopsy, which relies on the use of special semiautomatic or automatic cutting needles with a quick-release mechanism and a 14- to 18-gauge caliber. This system provides microhistological samples (frustules). The larger amount of tissue harvested allows sophisticated laboratory investigations, such as electronic microscopy, immunohistochemistry, and tumor-marker analysis, and all factors improve diagnostic specificity [27]. Considering the diversity of the mediastinum tumors and classification necessity of lymphomas, we used 16-gauge biopsy needle advancing along the planed path under real-time fluoroscopy to get enough sample, even if the lesion closed to blood vessels (Fig. 5). The real-time imaging CBCT guidance can make the biopsy procedures faster, and the total coaxial introducer dwelling time in our study was only 11.7 ± 3.4 min, which was shorter than that under the conventional CT guidance (mean 14.4–23.8 min) [2830].

The previous studies on CT-guided mediastinal biopsy have reported a wide range of complication rates, from 3.8 to 25.5 % [22, 24, 29, 30]. The most frequently complication of mediastinal biopsy has been pneumothorax. In our study, the pneumothorax rate 3 % (3/100) was found to be lower than CT-guided TNB procedure. We believe that this low incidence of pneumothorax was possible, as the virtual navigation system enabled us to select a safer and more accurate targeting route in navigating the needle approach to the target. As for puncture approach, the transpulmonary approach was used for ten lesions to avoid damage to the lung tissue. In addition, the coaxial needle technique also played a significant role in the reduction of complications by avoiding repeated pleural punctures or passages. Of the three patients who had pneumothorax, all patients had pulmonary emphysema along the needle pathway, and this may suggest that considering emphysema in the needle pathway would be more likely to lead to the occurrence of pneumothorax. Considering the intrinsic disadvantage of CBCT, in which the large cone angle increases scattered radiation as well as noise, which in turn can disturb the discrimination of low-contrast objects, vascular injury during CBCT-guided PNB is of significant concern. However, after scrupulous review of the diagnostic contrast-enhanced CT data taken prior to the procedure, the estimation of the lesion boundary and the safe range for needle advancement was feasible [20]. An old patient with giant mediastinal lesion and upper vena cava compression syndrome occurred subcutaneous hemorrhage after procedure. Obstruction of superior vena cava leads to an increase in venous pressure, and injury to collateral vessels by puncture during procedure leads to subcutaneous bleeding. Puncture point was oppressed for 20 min, and bleeding was stopped.

There were four cases (4.3 %) misdiagnosed in the study. One patients’ initial biopsy result was fibrous connective tissue; the nodule was enlarged more than 20 % in size on follow-up chest CT. TNB procedures were performed again, and the patient was confirmed with fibrosarcoma. The other there patients (one with HD and two with NHL) had completed the entire treatments, and tumor volumes were reduced more than 80 % compared with pretreatment. However, there were still persistence of soft tissue. Residual mediastinal tissue pathology results were coexists with fibrotic and/or necrotic tissue or thymic hyperplasia. Among them, two patients received the conventional open surgery and were confirmed with residual tumors. Another patient with NHL with high suspicion of tumor residue, TNB procedures were performed again and the patient was confirmed with residual tumor of NHL. An intrathoracic residual mass is present in 20 % of patients after treatment of an HD or an NHL. Eighteen percent of these patients experience a relapse at the level of the residual mass [31]. In our study, all patients were done with an enhanced chest CT before biopsy. CT alone does not allow one to discriminate between residual tumor and fibrosis or necrosis easily, especially for Lymphoma. According to most recent studies, CT has a poor sensitivity with respect to the diagnosis of residual disease. MRI has been shown to be slightly more reliable with respect to the diagnosis of fibrosis [32]. Preoperative MR combined with CT can better target mediastinal lesions and improve the correct rate. In fact, a negative result always possesses a diagnostic challenge. A non-specific benign result does not exclude malignancy, and further evaluation is required. The lesion may be malignant, but the sample obtained may lie outside the nodule borders or come from a necrotic area, thus preventing pathologists from establishing a correct diagnosis. Clinical and radiographic follow-ups are warranted. If further growth occurs after a non-specific benign diagnosis, it is obtained with TNB; repeat biopsy or resection is indicated [33].

In terms of radiation dose, a mean exposure dose was 9.7 ± 4.3 mSv. In Chiara et al. [18] study that dealt with pulmonary nodules with C-arm cone-beam CT, the total dose equals 11.62 mSv. Braak et al. [16] performed 92 biopsies in 88 patients with a similar guidance and reported a dose value equal to 9.95 mSv. On the basis of our results and the data from the literature regarding lung biopsies performed with CBCT guidance, despite the limited amount of currently available data and the lack of homogeneity, a slight reduction in dose could be hypothesized with the use of iGuide guidance. It could reduce the total dose delivered, because it can reduce the number of CBCT scans needed. The radiation dose of cone-beam CT-guided biopsy may not be a substantial problem, although a continuous effort must be made to reduce the radiation dose through use of a small field of view or collimation.

The limitations of this technique should also be mentioned. As patient movements may result in motion artifacts and affect the registration accuracy of the projected path, the key factor for achieving successful needle guidance is to generate high-quality 3D CBCT images. To this aim, during the imaging procedure, the patient should be well stabilized. However, this requirement may not be fulfilled and limits its applications for some patients who have difficulty sustaining a breath hold for the duration of imaging. The article did not have a control group, and it is difficult to assess the real advantages of CBCT-guided PNB for mediastinal lesions. Another main limitation of the study is the small sample size that limits the statistical power. Future studies examining a larger population using this system are warranted in this regard.

In conclusion, flat detector cone-beam CT-guided PNB is an accurate and safe diagnostic method for mediastinal lesions.