Introduction

In response to the COVID-19 pandemic, ultralow-temperature refrigeration is urgently required throughout the development, storage, transportation, and distribution of Sars-CoV-2 vaccines, and the number of ultralow-temperature freezers has expanded dramatically around the world. Some pharmaceutical companies, such as Pfizer-BioNTech, have claimed that their vaccines must be stored at temperatures ranging from -60 oC to -80 oC. This means that the COVID-19 vaccinations have stringent criteria for low-temperature storage freezers. According to American Society of Heating, Refrigerating, and Air-Conditioning Engineers (ASHRAE) (2014), ultralow-temperature refrigeration can be described as the storage of equipment at or below -50 oC and can also extend to the evaporation temperature of -100 oC (Mota-Babiloni et al. 2020).

CRS appears to be a better option for producing ultralow temperature. In general, the CRS operates with two or more serially disposed cycles and has a freezing temperature of < -60 oC. According to certain researchers (Mumanachit et al. 2012, Pan et al. 2020, Mateu-Royo et al. 2021), two-stage CRS is more efficient at low-temperature refrigeration, particularly in commercial supermarkets and industrial sectors. Chlorofluorocarbons (CFCs) refrigerants were mostly phased out due to their high Ozone Depletion Potential (ODP) and Global Warming Potential (GWP), particularly in developed countries. Hydrofluorocarbons (HFCs), a CFC substitute refrigerant, emit greenhouse gases. A transition to natural refrigerants appears to be unavoidable in the near future. Countries all over the world are focusing on natural refrigerants to help with global warming. Carbon dioxide R744 and N2O (R744A) are the most effective natural refrigerants. The fundamental disadvantage of R744 and R744A as low-temperature refrigerants is that both have a high operating pressure. Despite this, many researchers continue to believe that both are potential refrigerants for low-temperature refrigeration due to their zero ODP and low GWP.

Researchers conducted numerous theoretical analyses and experimental tests by employing R744 as a low-temperature refrigerant (LTR) in CRS. Lee et al. (2006) carried out a thermodynamic analysis on a R744-ammonia (R717) CRS in order to increase the COP and prevent exergy destruction in refrigeration system by regulating the condensing temperature of the cascade heat exchanger (CHE). It was discovered that the ideal COP increases with low-temperature cycle (LTC) evaporating temperature and decreases with high-temperature cycle (HTC) condensing temperature. To optimize the R744-R717 CRS's operating parameter, Getu and Bansal (2008) performed a thorough thermodynamic analysis. To estimate the ideal COP and refrigerant mass flow ratio between HTC and LTC, a multilinear regression analysis was used with the consideration of subcooling, superheating, evaporating and condensing temperatures, and CHE temperature difference. Bingming et al. (2009) experimentally evaluated the performance of a R744-R717 CRS with a twin-screw compressor. They discovered that the R744-R717 CRS had the maximum COP when the evaporation temperature was less than -40 oC. Dopazo and Fernández-Seara (2011) investigated the performance of a R744-R717 CRS with a refrigeration capacity of 9 kW and an evaporation temperature of -50 oC in an experimental study. The highest variation is roughly 2.4% when compared to optimum condensing temperature computed using different calculation correlations. Llopis et al. (2015) investigated the operation of four direct two-stage systems and a CRS with R744 as LTR. They discovered that direct CO2 transcritical two-stage systems are not suitable for usage in warm nations, however the CRS using R744 as LTR has good application prospects for commercial refrigeration in hot climates. Eini et al. (2016) performed a multi-objective optimization on a CRS from perspectives of economics, exergy, and the environment. The results revealed that R744-R717 and R744-R290 CRSs have nearly same economic and exergy efficiency, whereas R744-R717 CRS has intrinsic safety. Cabello et al. (2017) compared the performance of a CRS using R744 as LTR and R134a and R152a as high-temperature refrigerants (HTRs). The results revealed that the operation performance of the R134a-R744 CRS is comparable to that of the R152a-R744 CRS. That is, in a CRS, R134a can be replaced by R152a. Turgut and Turgut (2019) investigated CRS's performance with refrigerant pairs R744-R717, R744-R134a, and R744-R1234yf. R744-R1234yf CRS was discovered to have a high efficiency and a low annual cost. Amaris et al. (2019) examined the energy and exergy of a standard R744 booster refrigeration unit, a R744 parallel-compressor booster refrigeration unit, and a R717-R744 CRS. They discovered that the R717-R744 CRS operates better with an ambient temperature of up to 26 °C. Khalilzadeh et al. (2019) presented a novel integrated system powered by solar energy that combines a R744-R717 CRS and an Organic Rankine Cycle. The suggested system was evaluated using energy, exergy, exergoeconomic, and exergoenvironmental assessments. It was discovered that the COP of the integrated system is approximately 5.74 times greater than that of the standard CRS. Sánchez et al. (2019) conducted an experimental comparison of an R134a-R744 CRS and an indirect CRS with R744 as LTR and R152a, R1234ze(E), R290, and R1270 as HTRs. The results showed that the indirect CRS has a 2.5–17.1% rise in energy consumption, a 58.5% reduction in refrigerant mass charge, and a 30% reduction in environmental effect. Adebayo et al. (2021) performed the performance comparison of a CRS with R744 as LTR. It was determined that the refrigerant pair R717-R744 has the highest COP, and HFE7000 is a promising refrigerant that might be used in place of R134a. Unlike prior research, Alkhulaifi and Mokheimer (2022) used water as HTR and R744, R744A, R41, R717, R290, and R1270 as LTRs, and then performed a complete thermodynamic analysis on a CRS. They discovered that the refrigerant pair R717-R718 performs better, with a COP boost of 2.9–8.6%. Furthermore, they stated that there is no risk of freezing when utilizing water in CRS.

R744, in addition to being an LTR, can also be used as HTR in CRS. Colorado et al. (2012) investigated the thermodynamic performance of CRS in HTC with R744 and LTC with R717, R134a, butane R600, and propane R290. They discovered that the COP of R600-R744 CRS is 7.3% higher than that of R717-R744 CRS. Bhattacharyya et al. (2009) investigated the thermodynamics of a CRS with R744A as LTR and R744 as HTR. The effects of various design and operation parameters on CRS's performance were compared. It was demonstrated that adding an internal heat exchanger has no discernible influence on system performance improvement.

In addition to R744, refrigerant R23 and natural refrigerants R41 and R170 are also utilized as LTRs in CRS (Mota-Babiloni et al. 2020). Kilicarslan and Hosoz (2010) investigated the energy performance and irreversibility of a CRS using R23 as LTR. After comprehensive comparisons, refrigerant couples R152a-R23 and R404a-R23 were proposed as replacements for refrigerants R717-R23 and R507-R23. Sarkar et al. (2013) developed a theoretical model to assess the performance of a CRS containing R170, R1150, and R744A in LTC and R717 in HTC. They discovered that the R717-R170 CRS has the best operation performance, while the R717-R744A CRS has the best volumetric cooling capacity. Sun et al. (2016) compared the performance of an R41-R404A CRS to an R23-R404A CRS using the first and second laws of thermodynamics. The discharge temperature, coefficient of performance (COP), exergy destruction, and exergy efficiency were determined and compared. They suggested that the refrigerant pair R41-R404A could be a preferable substitute for R23-R404A in CRS. Sun et al. (2019) then evaluated the energy and exergy of low GWP refrigerants in CRS using R23, R41, and ethane R170 as LTRs. They came to the conclusion that R41 and R170 are better LTRs. Furthermore, they discovered that the R41-R161 and R41-R717 CRS have better operation performances. Roy and Mandal (2019) compared the operation performance of R41-R404A and R170-R161 CRSs. The R170-R161 CRS has a better ideal COP and exergetic efficiency, according to the researchers. Furthermore, Roy and Mandal (2020) conducted a thorough investigation of a CRS from energetic, exegetic, economic, and environmental perspectives. The refrigerant pair R41-R161 attained the highest exergetic efficiency of CRS by using R41 and R170 as LTRs and R404A and R161 as HTRs. Aktemur et al. (2021) conducted energy and exergy analyses on CRS using refrigerant R41 in LTC and R1243zf, R423A, R601, R601A, R1233zd (E), and RE170 in HTC. The refrigerant pair R41-RE170 was discovered to have the best operation COP and exergy efficiency, and RE170 can be used as an alternate refrigerant for R423A. Aktemur and Ozturk (2021) investigated the performance of a subcritical CRS with an internal heat exchanger. R41 is employed as an LTR, and R601, R602A, and cyclopentane are used as HTRs. Following rigorous evaluations, they discovered that the refrigerant pair R41-R601 has the best thermodynamic performance among three refrigerant pairs. Mouneer et al. (2021) used R170 as LTR and performed thermodynamic analysis on a CRS with R32, R123, R134A, R404A, R407C, R410, and R290 as HTRs to generate -80 oC low temperature. The top three best exegetic efficiencies were discovered to be produced by refrigerant pairs R123-R170, R290-R170, and R134-R170, with values of 40%, 38%, and 35%, respectively. As a result, natural refrigerants R290-R170 are advised for CRS in order to create an ultralow-temperature zone. Udroiu et al. (2022) built a thermodynamic model using R170 as LTR and R290 as HTR to compare the performance of 42 different refrigeration system setups. Rodriguez-Criado et al. (2022) refitted a package R290 refrigeration unit with R170 to achieve ultralow-temperature freezing. The COP of the retrofitted refrigeration unit ranges from 0.6 to 1.6 with evaporation temperatures ranging from -65 °C to -80 °C.

It is clear from the introduction above that R744, R23, R744A, and R170 are frequently chosen as LTC fluids in CRS. Furthermore, less research focuses on CRS at ultralow temperatures. The natural refrigerant R170 is regarded as a viable LTR employed in CRS to realize ultralow temperatures because of the high operating pressure of R744 and R744A and the high GWP of R23. The little amount of associated experimental research on R170 as LTR in CRS is primarily concentrated on energy and exergy analysis. Despite the fact that Rodriguez-Criado (2022) used R170 in experiments, the evaporation temperature is between -65 °C and -80 °C. There has not yet been an experimental research of CRS at refrigeration temperatures below -86 °C. The low temperature of -80 °C was experimentally measured in a low-temperature freezer by Liu et al. (2023) and Tan et al. (2023) utilizing a cascade refrigeration cycle with refrigerants R170 and R290. Their research demonstrated that it is both feasible and appropriate to produce low temperatures of -80 °C utilizing R290-R170 CRS.

In order to comply with the requirements of the European Union F-Gas Regulation 517/2014, the environmental-friendly refrigerants R170 and R290 are selected to be LTR and HTR in CRS, and are employed in this study to create -86 °C ultralow-temperature. The operation performance of a -86 °C freezer is experimentally examined under two different ambient circumstances with a focus on the storage and transportation of COVID-19 vaccinations. The pull-down and steady operation processes of the -86 °C freezer are both studied. The power usage of CRS is assessed. Some significant conclusions are reached. The current study is important for gaining a thorough understanding of the operation performance of -86 °C CRS freezer and achieving global environmental sustainability.

System description and experimental setup

Experimental schematic diagram

Figure 1 depicts the schematic diagram (Liu et al. 2023, Tan et al. 2023) of the CRS for a -86 °C ultralow-temperature freezer. As can be seen, the CRS is made up of an HTC and an LTC. The HTC consists primarily of a high-temperature compressor, a condenser, a high-temperature capillary, an anti-condensation loop, a drying filter, and a liquid reservoir, whereas the LTC consists of a low-temperature compressor, a precooled condenser, a drying filter, a low-temperature capillary, an evaporator, and an oil separator. The HTC and LTC are connected by a cascade heat exchanger (CHE). Different components are connected in sequence by copper pipe.

Fig. 1
figure 1

Schematic diagram of CRS (Tan et al. 2023; Liu et al. 2023)

In this experiment, a precooled condenser is placed before CHE in LTC. This enhancement could lower the heat exchange burden of the CHE and significantly cool the LTR. As a result, LTC may readily generate extra-low temperatures. Furthermore, an anti-condensation loop is installed after the HTC compressor and is placed beneath the sealing area of the freezer's exterior door. HTR gas exits the HTC compressor and enters the anti-condensation loop. Condensation of water vapor close to the freezer's external door can be avoided when heated by HTR gas. Meanwhile, the anti-condensation loop can significantly cool the HTR gas, resulting in a noticeable temperature decrease.

Different monitors are configured to gather fluid temperature variations in order to investigate the operation performance of the CRS freezer. Two thermocouples are typically installed at the inlet and outlet of key components. In LTC, monitors T1 and T2 are installed at the inlet and outlet of low-temperature compressor, monitors T2’ and T2’’ are installed at the inlet and outlet of precooled condenser, and monitor T3 and T3’ is set at the inlet of CHE and LTC capillary. As for HTC, monitors, installed at the inlet and outlet of the HTC compressor, anti-condensation loop, condenser, capillary, CHE and liquid reservoir, are labelled T5’, T6, T6’, T6’’, T7, T7’ and T5, respectively. Based on the experimental rig, the related P–h diagram of the CRS is shown in Fig. 2. Hence, the thermal dynamic states of various temperature test points can be easily found from the P–h diagram.

Fig. 2
figure 2

P–h diagram of CRS

Along with above monitors, three thermocouples labeled TLTC, THTC and TOS are mounted to the chamber walls of two compressors and an oil separator. Furthermore, because the CHE is a small heat exchanger and the LTC evaporator is encased in foam layer, it is difficult to set monitors to measure fluid temperature at the input of both devices. As a result, T4 and T8 monitors after capillary are not established in experiments.

Experimental setup

The experimental apparatus of a -86 °C ultralow-temperature freezer and associated test instruments are shown in Fig. 3. The test is performed in a Type-laboratory, where varied dry bulb and wet bulb temperatures can be established. The test freezer is made up of an inner box and an outside box, both are composed of iron sheet. The size of the inner box is 588 × 710 × 1310 mm (W × L × H), and the size of the outer box is 873 × 968 × 1980 mm (W × L × H). To decrease heat leakage from external environment, the space between the inner and outer boxes is filled with foam. The foam layer in freezer's top, left, back, and right sides is 90 mm thick, while the foam layer at the bottom of the inner box is 130 mm thick to decrease heat invasion. The inner box contains 552 L of storage space divided into four cabinets. Each cabinet has a capacity of 138 L. As indicated in Fig. 3(a), four interior doors and a separate external door are installed to reduce cold loss from freezer. Internal and external doors have thicknesses of 34 mm and 81 mm, respectively. Apart from LTC evaporator and capillary, the primary components of CRS are located at the bottom of the inner box. This is due to the fact that two compressors, a condenser, a CHE, and other associated components take up too much space and have a high mass. Putting these components together and arranging them beneath the inner box can save more refrigeration space while also increasing freezer's operational stability. The HTC evaporator is handcrafted from copper tube and is coiled around the inner box's wall from top to bottom, as shown in Fig. 3(b). The evaporator is linked to low-temperature capillary, which is embedded in foam layer.

Fig. 3
figure 3

Experimental rig of low-temperature freezer (a) and evaporation coil (b)

Two SC21CNX.2 LBP compressors are adopted. R290 and R170 charge quantities are 155 g and 103 g, respectively. A finned tube heat exchanger is used as condenser. Driven by a fan, external air exchanges heat with HTR gas. HTR gas is eventually condensed to pure liquid after being cooled by external air. In addition, a small finned tube condenser is installed in LTC. Cooled by external air, R170 gas has a noticeable temperature decrease through precooled condenser, which is significant for reducing CHE heat load and achieving a significantly lower evaporation temperature. A casing heat exchanger is chosen as CHE. Table 1 lists the detailed specifications of the CRS's main components (Liu et al. 2023, Tan et al. 2023).

Table 1 Specifications of the main components

The power source has a voltage of 220 V and a frequency of 50 Hz. The power consumptions of two compressors and fans are measured using a wattmeter. The power meter's measurement accuracy is 0.3%FS. Several test stations are set up to monitor variations in system pressure and temperature to assure the safe operation of the CRS. When overpressure and overheating occur, the emergency cut-off safeguard is activated. Furthermore, several temperature test points are established in freezer to control the start-stop switch of CRS.

The T-type thermocouple is used to measure fluid temperature at the entrance and exit of different components, as well as air temperature in freezer. As seen in Fig. 3, a multichannel data acquisition instrument (Liu et al. 2019, Liu and Tan 2019) is used to collect temperature data from several monitors. The measure uncertainty of the T-type thermocouple is ± 0.5 °C with a temperature range of -150—350 °C. The shallow groove in pipe is formed to improve the test accuracy, and the thermocouple is pasted to groove and coated with dielectric adhesive tapes. As Fig. 1 shows, 13 thermocouples, 6 for LTC and 7 for HTC, are installed in CRS to monitor temperature fluctuations of refrigerants R290 and R170. Furthermore, three temperature monitors are attached to the chamber walls of HTC compressor, LTC compressor, and oil-separator, respectively.

In addition to various state monitors, 11 temperature monitors are set up in freezer to analyze detailed air temperature distribution. As Fig. 4 shows, 4 cabinets are labeled No.1-No.4 from top to bottom. It contains a core temperature monitor and a back temperature monitor for each compartment. A copper bar is used to fix the thermocouple to test the core temperature in each compartment. The back monitor is directly tied to the inner box's wall. Tcore1-Tcore4 are the compartment numbers for four core monitors, and Tback1-Tback4 are the compartment numbers for four back monitors. In addition, three thermocouples, labeled Tleft, Tright and Ttop, are attached to the left, right, and top sides of compartment No.1.

Fig. 4
figure 4

Temperature test point distribution in freezer

Results and discussion

The operation performance of the -86 °C freezer has been tested experimentally. The experiment was conducted in a Type-laboratory at two different dry bulb and wet bulb temperatures. In the Type-laboratory, the relative humidity is set as 65%.

Dry bulb temperature of 25 °C

As Fig. 5 shows, the CRS operates for approximately 1200 min with a dry bulb temperature Tdry of 25.0 °C and a wet bulb temperature Twet of 20.2 °C. The associated temperature variations of different monitors are described.

Fig. 5
figure 5

Variations of dry bulb and wet bulb temperature

Temperature variation of different state monitors of HTC and LTC

Figure 6 displays the temperature changes of different state monitors in LTC. It is easy to see different monitors all experience initial temperature drops while the CRS is activated. Generally, the pull-down process can be separated into three phases, including early start-up (ESU), rapid temperature drop (RTD) and stable temperature drop (STD) phases. The three phases were also observed in Wang’s experiments (Wang et al. 2020). Three phases take 25 min, 95 min, and 412 min, respectively. During the first 25 min, suction temperature T1 drops somewhat, whereas discharge temperature T2 rises initially and reaches a maximum temperature of 102.8 °C in 25 min. Temperature drops also occur in other monitors, including T2’, T2’’, T3 and T3’. Distinct monitors have different temperature profiles and temperature changes during ESU period. After 25 min, the freezer enters RTD phase, and different monitors suffer rapid temperature reductions, as illustrated in Fig. 6. This phase lasts 95 min. Temperature drops of monitors T1-T3’ are 21.4 °C, 12.0 °C, 14.2 °C, 3.2 °C, 2.2 °C and 2.8 °C, respectively, with temperature drop rates of 3.52 °C/hour, 7.58 °C/hour, 8.97 °C/hour, 2.02 °C/hour, 1.39 °C/hour and 1.77 °C/hour. Following that, the CRS enters STD period, which lasts 532 min. Temperatures of different monitors, with the exception of monitor T1, slowly decrease during STD phase. When -86 °C is reached, the final values of monitors T2-T3’ are 85.3 °C, 53.2 °C, 29.3 °C, 29.2 °C and 27.3 °C, respectively. When it comes to monitor T1, its temperature fluctuates dramatically. The primary cause is as follows. The temperature rise of T1 is caused by environmental heat leaking through suction pipe between evaporator and LTC compressor. Furthermore, bits of liquid droplets are combined in LTC compressor's suction. The liquid droplet evaporates, lowering the refrigerant temperature. As a result, T1 experiences noticeable temperature oscillations and drops to -45.9 °C while the CRS ceases.

Fig. 6
figure 6

Temperature distribution of different test points of LTC

The CRS enters a periodic start-stop switching phase after 532 min. Based on Fig. 6, it is clear that, aside from monitor T1, other monitors all have a tendency to certain temperatures. T2 is held at 84.2 °C, the input and outlet gas temperatures of precooled condenser are kept at 52.9 °C and 29.3 °C, respectively, and the inlet temperatures of the CHE and LTC capillaries are 29.2 oC and 27.3 °C, respectively. According to comprehensive investigation, the temperature drop from the output of LTC compressor to the inlet of the precooled condenser is 31.3 °C, whereas the temperature drop via precooled condenser is 23.6 °C. That is to say the long connecting pipe has a greater cooling impact than the precooled condenser. This is mostly due to the fact that the environment temperature is fixed at 25 °C and the discharge temperature of the compressor surpasses 80 °C. The considerable temperature difference leads to the high-temperature discharge gas being greatly cooled. T2’’ and T3 monitors practically have the same temperature with only 0.1 °C difference. It means that the temperature difference between the precooled condenser outlet and the CHE can be ignored. While refrigerant R170 flows through CHE and arrives at the inlet of capillary, its temperature remains almost constant at 27.3 °C.

Figure 7 depicts temperature variations of various HTC monitors. It is apparent that the HTC likewise goes through an initial adjustment. In contrast to LTC, temperature variations in HTC exhibit minor temperature decreases or rises. Aside from HTC compressor discharge gas monitor T6, all other monitors show early temperature drops. T6 experiences an initial temperature rise and achieves the maximum temperature 106.8 °C in 50 min. The initial pull-down operation takes 120 min. Since then, different monitors remain constant. The HTC achieves stable functioning 412 min earlier than the LTC. Furthermore, it appears that the pull-down process has slight effect on the operation performance of the HTC. The CRS enters periodic start-stop switch 412 min later. Monitor T5’ has the most noticeable fluctuation profile. This phenomena also occurs at LTC compressor's inlet for similar reasons. While the HTC is stable, the average temperatures of monitors T5-T7’ are -11.7 °C, 105.5 °C, 60.4 °C, 35.2 °C, 31.8 °C and 31.7 °C, respectively. T5’ changes constantly, but its average temperature is 23.8 °C. When high-temperature R290 gas flows from HTC compressor to anti-condensation loop, it is considerably cooled by cold air from outside, with a temperature drop (T6-T6’) of 45.1 °C. R290 is chilled in anti-condensation loop by penetration cold capacity from the freezer and external air flow. The temperature drop (T6’-T6’’) through the anti-condensation loop is about 25.2 °C, which is lower than the temperature drop through the long connecting pipe. The pipeline length from HTC compressor to anti-condensation loop can be correctly decreased to improve the temperature reduction effect created by anti-condensation loop. Following that, HTR gas passes through condenser and toward capillary. It is obvious that a temperature difference of 3.4 °C exists between the condenser's inlet and outlet. The primary causes are stated. (1) There is still a long pipe connecting T6’’ monitor to HTC condenser. (2) The refrigerant at the condenser's exit may be subcooled. (3) A visible pressure drop occurs when refrigerant runs through HTC condenser. All of the following factors contribute to the difference between T6’’ and T7. Only 0.1 °C temperature difference is generated as the refrigerant fluid travels to LTC capillary. As a result, the temperature of the refrigerant is nearly constant from the condenser outlet to the capillary inlet.

Fig. 7
figure 7

Temperature distribution of different test points of HTC

Figure 8 depicts the fluctuations in wall temperature of two compressors and an oil separator. Both HTC compressor and oil separator have obvious wall temperature reductions. The oil separator experiences an early wall temperature drop as a result of CRS startup, while high-temperature compressor experiences a rapid wall temperature rise in the first 25 min. Following that, both have a quick temperature drop and a slow temperature reduction since 25 min. The wall temperature of low-temperature compressor chamber increases rapidly in the first 120 min, then gradually slows and lasts to 532 min. The CRS enters periodic start-stop switch once -86 °C is reached. The average temperatures of TLTC, THTC and TOS after 11 operating cycles are 61.6 °C, 40.2 °C and 28.8 °C, respectively. Furthermore, TOS is always lower than TLTC. Generally, lubricant oil has significant cooling effect on low-temperature compressor with a temperature difference of 32.8 oC.

Fig. 8
figure 8

Wall temperature variations of two compressors and one oil separator

Temperature variation of different monitors in freezer

The temperature changes of different core monitoring and back monitors in freezer are shown in Fig. 9. It is clear that once the CRS is activated, all core monitors suffer a rapid temperature drop. The ESU of CRS appears to have minor influence on the quick temperature reductions of different monitors. Different core monitors entere a consistent temperature drop period since 120 min. That is to say the core monitors only undergo two temperature drop phases. The initial temperatures of Tcore1-Tcore4 are -50.3 °C, -48.3 °C, -40.5 °C and -35.3 °C, respectively. Furthermore, as shown in Fig. 9(a), it takes 356 min, 248 min, 182 min and 151 min for monitors Tcore1-Tcore4 to reach the desired temperature of -86 °C. The related temperature reduction rates are 6.02 °C/hour, 9.12 °C/hour, 15.0 °C/hour and 20.14 °C/hour, respectively. It is clear to see that monitor Tcore4 has the highest temperature reduction rate and monitor Tcore1 has the lowest temperature reduction rate. This is mostly due to the fact that monitor Tcore1 is only cooled by evaporator coil, but monitor Tcore4 receives cold capacity from both evaporator coil and cold air sink from  freezer's top. Hence, when compared to other monitors, Tcore4 has a faster temperature reduction rate and a lower temperature. 532 min later, the freezer enters a periodic start-stop switch phase, and different monitors show reasonably constant temperature fluctuations. After 640 min start-stop cycle operation, the average temperatures of Tcore1-Tcore4 are -86.4 °C, 87.5 °C, -88.4 °C and -88.8 °C, respectively. It can be seen that the temperature of the core monitor reduces from top to bottom.

Fig. 9
figure 9

Temperature variations of different core monitors (a) and back monitors (b)

As Fig. 9(b) shows, different back monitors also experience RTD in the first 120 min, and then undergo STD. Different from core monitors, four back monitors dropped to -86 °C in the first 86 min. Due to heat transfer temperature difference and thermal resistance, the air temperature in the freezer is still higher than -86 °C. When the interior temperature of the freezer reaches -86 °C, temperatures of four back monitors are -92.1 °C, -91.9 °C, -91.7 °C and -90.8 °C, respectively. Since then, temperatures of four back monitors enter periodic fluctuation variations. Moreover, it is easy to find the temperature of back monitors rises from top to bottom, which is opposite to the temperature variation trend of core monitors. This is mostly due to the fact that the temperatures of back monitors reflect temperature changes of refrigerant within evaporation coil. While core monitors display air temperature in the freezer, the air temperature is determined by a mix of cooling from evaporation coil and natural convection within the freezer. As a result, the temperature of back monitors steadily falls in the flow direction. Temperatures of various monitors are generally below -86 °C, with the lowest temperature of -92.8 °C.

Power consumption

Figure 10 presents the variation of power consumption P of CRS compressors and fans. When the CRS is turned on, P rises to 1176.1 W. During pull-down process, P drops rapidly for the first 120 min and then undergoes stable decrease. Once the CRS ends in 532 min, the consumed power is about 865.6 W. Thereafter, P experiences cyclical fluctuations and ranges in 865.6–925.5 W. The average power consumption is about 872 W.

Fig. 10
figure 10

Power consumption of CRS

Dry bulb temperature of 32 °C

With Tdry = 32 °C and Twet = 26.5 °C, the experimental test on the ultralow-temperature freezer lasts for 1440 min, as shown in Fig. 11. The associated temperature variations of different test points are monitored and introduced.

Fig. 11
figure 11

Variations of dry bulb temperature and wet bulb temperature

Temperature variation of different state monitors of HTC and LTC

Figure 12 displays the temperature variations of various monitors in LTC. Here, two operation start-stop cycles are tested with Tdry = 32 °C and Twet = 26.5 °C. Cycle 1 lasts around 960 min and cycle 2 operates for 292 min.

Fig. 12
figure 12

Temperature distribution of different test points of LTC

As Fig. 12 shows, the CRS enters stable operation after 10 min of adjustment. Aside from suction gas temperature T1, other monitors all have stable operation temperature values. For example, the discharging vapor temperature T2 of LTC compressor is typically 94.1 °C, while the inlet temperature T2’ and outlet temperature T2’’ of precooled condenser are kept to 63.2 °C and 37.1 °C, respectively. The inlet gas temperature T3 of CHE after passing through drying filter, is approximately 36.8 °C with a temperature reduction of 0.3 °C, as shown in Fig. 12(b). Flowing through a long connecting pipe, refrigerant temperature T3’ before low-temperature capillary is kept to 35.0 °C. It signifies that a temperature difference of 1.8 °C exists between the CHE exit and the inlet of LTC capillary. As for the suction temperature of LTC compressor, T1 exhibits significant temperature fluctuations, shown as Fig. 12(a). Generally, T1 nearly increases from -54.2 °C to -44.6 °C with noticeable temperature fluctuation during cycle 1. When it comes to cycle 2, monitor T1 has a minor temperature reduction, while other monitors remain at the same temperatures.

Figure 13 depicts the temperature variations of various HTC monitors. It is easy to see the CRS comes into stable functioning after a short-term adjustment, and different monitors incline to certain temperature values. The inlet temperature T5 and outlet gas temperature T6 of HTC compressor remain constant at -10 °C and 113.4 °C, respectively. HTR gas goes through a long connection pipe to anti-condensation loop, and the inlet temperature T6’ of anti-condensation loop is around 72.6 °C. It is easy to locate 40.8 °C temperature decrease forms. This is primarily due to the discharge gas temperature being higher than 110 °C and the ambient temperature being set at 32 °C. Because of large temperature difference, the high-temperature discharge gas is cooled by externally chilled air. The outlet gas temperature T6’’ is approximately 45.2 °C when R170 gas flows through the anti-condensation. That means the temperature drops by only 27.4 °C. The anti-condensation loop's temperature drop is primarily cooled by cold air from the freezer. Following that, a high-temperature refrigerant gas runs through the condenser with an exit temperature T7 of 39.5 °C. It is seen that 4.7 °C temperature difference forms between the inlet and outlet of the condenser. A similar explanation was given above. A temperature difference of 0.1 °C is generated by refrigerant fluid flowing from the CHE exit to the LTC capillary.

Fig. 13
figure 13

Temperature distribution of different test points of HTC

Figure 14 depicts the temperature variations on the wall of two compressors and an oil separator. When the CRS is activated, the wall temperature TLTC of LTC compressor rises from 49.4 °C to 71.3 °C and stays there, whereas THTC rises from 35.0 °C and goes to 47.2 °C during stable operation. The chamber wall of oil separator has low temperature, and TOS rises from 32.6 °C to 36.2 °C. Throughout whole procedure, TOS is lower than TLTC, with an average temperature difference of 35.1 °C. The excellent cooling effect of lubricating oil has been confirmed once more. In cycle 2, three monitors also undergo a fast temperature increase and tend to be constant.

Fig. 14
figure 14

Temperature variations of two compressors and an oil separator

Temperature variation of different monitors in freezer

Figure 15 displays the temperature distributions of air in freezer. Monitors in freezer also have stable temperature changes, which are influenced by stable temperature distribution of different state monitors depicted in Figs. 1213. As indicated in Fig. 15(a), when the freezer is turned on, four core monitors exhibit initial modest amplitude temperature rises, then slightly decrease and tend to be steady after 180 min. Temperatures of four core monitors tend to stabilize after roughly 180 min. Similar temperature variations also occur in cycle 2. During stable operation, the average temperatures of Tcore1-Tcore4 finally maintain at -84.6 °C, -86.1 °C, -87.1 °C and -87.4 °C, respectively. The temperature of the air in the freezer drops from top to bottom. Furthermore, save from some core space in compartment No.1, temperatures in the rest of the freezer all drop to or below -86.0 °C. To realize the entire space of the freezer being cooled down -86.0 °C, some optimism measurements and winding adjustments on the evaporation coil should be performed, such as decreasing the coil interval of the evaporator in compartment No.1 or raising the mass flow rate of refrigerants.

Fig. 15
figure 15

Temperature variations of different core monitors (a) and back monitors (b)

The temperature distributions of four monitors placed at the back of the freezer are shown in Fig. 15(b). Four back monitors exhibit abrupt temperature increases at first, then decrease and stabilize. The final temperature values of monitors Tback1-Tback4 after 180 min adjustment are -90.2 °C, -90.1 °C, -89.9 °C and -89.2 °C, respectively. The temperatures of back monitors rise from top to bottom and are lower than that of core monitors. Furthermore, based on the temperatures of four rear monitors, it is possible to determine that the evaporation temperature of LTC must be less than -90 oC.

Power consumption

The change in P of CRS with Tdry = 32 °C and Twet = 26.5 °C is shown in Fig. 16. It is obvious that when the CRS is activated, P increases rapidly at first and reaches 1137 W. Following then, the CRS experiences short-term adjustment, and P tends to be 935 W during stable operation. P climbs to 1048 W in cycle 2 before remaining constant at 935 W.

Fig. 16
figure 16

Power consumption of CRS

Discussion and analysis

Based on preceding testing results, the impact of dry bulb and wet bulb temperatures on the operation performance of -86 °C ultra-low-temperature freezer is analyzed. The stable operating temperatures of various state monitors are recorded and compared, with the average temperature of monitor T1 being adopted due to its decrease during operation cycle 1. Table 2 displays the stable operating temperatures of various state monitors in the freezer, revealing that T1 drops from -21.6 °C to -49.4 °C as Tdry increases from 25 °C to 32 °C and Twet increases from 20.2 °C to 26.5 °C. However, the temperatures of other monitors increase alongside T1 with Tdry/Twet rising from 25/20.2 °C to 32/26.5 °C, resulting in temperature differences between two cases varying from 7.7–10.3 °C. As the ambient temperature increases, more energy is required to pull down to the desired temperature, leading to a noticeable rise in power consumption. The pressure ratio also increases as the temperatures of other monitors rise. The temperature drop through the long connecting pipe between the freezer compressor and precooled condenser is also calculated, with temperature drops in both cases being nearly identical, differing by only 0.4 °C. Additionally, the precooled condenser has a better cooling effect with a temperature drop of 26.1 °C when Tdry = 32 °C and Twet = 26.5 °C, whereas the connecting pipe has a slightly better effect with Tdry = 25 °C and Twet = 20.2 °C. Overall, the length of the connecting pipe can be increased appropriately to reduce the size of the precooled condenser.

Table 2 Temperature comparison of different state monitors of LTC

Table 3 presents the stable temperatures of different monitors in HTC. In contrast to LTC, the temperatures of different monitors in HTC increase with Tdry and Twet. The suction temperature and discharge temperature exhibit temperature changes of 1.7 °C and 6.8 °C, respectively. In this case, the temperature drops through the long pipe connecting the outlet of HTC compressor and the inlet of anti-condensation loop, and through anti-condensation loop, are compared. It is apparent that a higher temperature decline of 46.1 °C occurs through the loop pipe with Tdry = 25 °C and Twet = 20.2 °C, while a 27.4 °C temperature drop is observed in the anti-condensation loop with Tdry = 32 °C and Twet = 26.5 °C. The temperature drop through the long pipe is larger than that in the anti-condensation loop. Properly reducing the length of the connecting pipe can enhance the anti-condensation effect.

Table 3 Temperature comparison of different state monitors of HTC

To evaluate the cooling performance of the CRS in two different environmental conditions, Table 4 displays the stable air temperatures inside the freezer, as well as the lowest temperatures recorded by four back monitors. In both scenarios, the temperatures of four core monitors decrease from top to bottom. When Tdry is set as 25 °C and Twet is 20.2 °C, the entire freezer is chilled to -86.0 °C, and the lowest temperature of the four back monitors reaches -92.8 °C. However, when Tdry increases to 32 °C and Twet to 26.5 °C, cabinet No.1 cannot be cooled down to -86.0 °C, while the rest of the freezer reaches the desired temperature. The lowest temperature of the back monitor is -90.2 °C. The heat transfer temperature differences between the evaporator coil and the inner space of the freezer are approximately 6.8 °C and 4.2 °C.

Table 4 Temperature comparison of different monitors in the freezer

Based on test results, it is found that the power consumption of the -86 °C ultralow-temperature freezer during stable operation increases from 872 to 935 W when Tdry increases from 25 °C to 32 °C and Twet increases from 20.2 °C to 26.5 °C. The rise in power consumption is primarily due to the drop in LTC suction temperature and the increase in the discharge temperatures of both LTC and HTC. This results in a higher temperature difference between the inlet and exit of two compressors, causing the increase of the pressure ratio and, in turn, leading to higher power consumption. In addition, high dry bulb and wet bulb temperatures cause the increase of heat load in condensers, which increases fan power consumption.

Conclusions

An experimental study was carried out to investigate the operation performance of a -86 °C ultralow-temperature freezer using natural refrigerants R290-R170 under two dry bulb and wet bulb temperatures. Various state monitors and temperature monitors were set up to examine the temperature changes of R290 and R170, and the temperature distribution in freezer during the pull-down and stable operation stages. Some of the most important discoveries are summarized.

Generally, the temperature fluctuations of different state monitors in LTC can be separated into three stages during the pull-down operation, namely EST, RTD, and STD. With Tdry = 25 °C and Twet = 20.2 °C, the EST phase lasts 25 min, during which different monitors undergo initial temperature adjustments. During the RTD phase (25–120 min), different monitors experience rapid and significant temperature reductions. The longest of the three stages is the STD phase, where various monitors experience small temperature drops. The HTC has no discernible influence on temperature fluctuations across different monitors. The core and back monitors undergo rapid temperature drops before entering the STD phase. The temperature drop rates of the inlet and outlet of LTC compressor are 13.52 °C/hour and 7.58 °C/hour, respectively. Temperature decrease rates of four core monitors are 6.02 °C/hour, 9.12 °C/hour, 15.0 °C/hour, and 20.14 °C/hour, respectively. While Tdry is set as 32 °C and Twet as 26.5 °C, the CRS freezer experiences 10 min adjustment initially and then comes into stable operation, so the initial effect is not introduced here.

Once -86 °C is achieved, the CRS operates periodically and different monitors tend to be steady. Temperature rises are observed across all monitors with Tdry ranging from 25 °C to 32 °C and Twet ranging from 20.2 °C to 26.5 °C, except for the suction temperature T1 of LTC compressor, which decreases as Tdry and Twet increase. As a result, the power consumption has increased by 7.2%. The length of the LTC and HTC connecting pipelines should be adjusted suitably to improve the effects of the precooled condenser and anti-condensation loop. The wall temperature of the oil separator is always lower than that of the LTC compressor, confirming the cooling impact of the lubricating oil on LTC compressor.

The entire freezer space can be cooled to -86 °C with Tdry = 25 °C and Twet = 20.2 °C. When Tdry rises to 32 °C and Twet rises to 26.5 °C, some space in cabinet No.1 remains uncooled at -84.6 °C, but the air temperature in cabinets No.2–4 has dropped to -86 °C. Back monitors have the lowest temperatures of -92.8 °C and -90.2 °C in two occasions. This indicates that the current ultralow-temperature freezer can meet the demand for COVID-19 vaccine storage at an environmental temperature of 25 °C. However, improvements should be made to reduce the air temperature in the top cabinet of the freezer when Tdry reaches 32 °C.