Abstract
The LEGEND collaboration has been developing a 76Ge-based double-beta decay experimental program where precise radiopurity measurements of ultraclean materials are crucial. Ultralow concentrations of thorium and uranium, the main contributors to the detector background via their decay products, can be determined by inductively coupled plasma mass spectrometry (ICPMS) and accelerator mass spectrometry (AMS). Here we shall present recent developments in thorium and uranium mass spectrometry methods, together with basics of separation chemistry applied to process different samples. The new possibilities to measure 232Th and 238U by ICPMS and AMS at the Comenius University in Bratislava are discussed as well.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In recent years, several large-scale physics experiments have been designed or put in operation to search for neutrinoless double beta decay, which, if observed, would enable scientists to answer the question if neutrino is a so-called Majorana particle and what its mass is [1,2,3,4,5,6]. One of such efforts is the world LEGEND (Large Enriched Germanium Experiment for Neutrinoless Double Beta Decay) experiment which uses high-purity germanium detectors made of material with the enriched level of 76Ge [7, 8]. The first phase, LEGEND-200, is a successor of the GERDA experiment [9], operating with its infrastructure at the Gran Sasso National Laboratory (LNGS) in Italy and also implementing the knowledge obtained from the development of another germanium double beta decay detector, MAJORANA DEMONSTRATOR [10]. The LEGEND-200 will comprise detectors containing roughly 200 kg of enriched Ge, reaching the sensitivity for neutrinoless double beta decay at the level of > 1027 yr after 5 years of data taking, while having the overall background below 2.5 × 10–4 counts keV−1 kg−1 yr−1 [11]. Next, the LEGEND collaboration plans to develop and construct an experiment that will use detectors with around 1000 kg of germanium with the total background of less than 1.0 × 10–5 counts keV−1 kg−1 yr−1 at the same time. LEGEND-1000 is expected to run for at least ten years, resulting in a limit for the half-life of neutrinoless double beta decay for 76Ge beyond 1028 yr [12].
The background of complex and sensitive detector consists of various components from different sources, naturally occurring thorium 232Th (T1/2 = 1.40 × 1010 yr) and uranium 238U (T1/2 = 4.47 × 109 yr) and their short-lived progenies being one of the most important contributors. In the case of LEGEND, radiation from daughter radioisotopes of thorium and uranium will make up to ~ 20–30% of the overall background, depending on the phase of the experiment [12]. Natural thorium and uranium are part of each material used for construction of the experiment, thus the information about their content is crucial for correct estimation of the background. Some materials, which are situated far from the sensitive volume of detector (e.g., cables, optical fibers), can be screened by non-destructive gamma spectrometry. More critical parts, such as germanium crystals and detector mounts, usually require the application of mass spectrometry techniques, which are generally much more sensitive for long-lived radionuclides than conventional radiometric methods [13,14,15]. Inductively coupled plasma mass spectrometry (ICPMS) has become a routine method for radiopurity measurements of high-purity materials, offering relatively quick and inexpensive analysis down to 10–15 g g−1 [16,17,18]. An even greater potential has accelerator mass spectrometry (AMS), whose exceptional limit of detection is given by the removal of interfering ions due to accelerator-based stripping and multiple mass separation steps. However, if one does not deal with specific materials [19], sample preparation is obligatory for both ICPMS and AMS.
The scientific program of the Comenius University in Bratislava has been devoted to – among others – nuclear physics and environmental studies that have been based on the determination of low-level activities of radionuclides and concentrations of elements (stable isotopes). At the Faculty of Mathematics, Physics and Informatics, two shielded HPGe detectors are in operation, which have been used for low-background gamma spectrometry measurements of natural and anthropogenic radionuclides in samples of various origin and rare events search [20,21,22,23,24]. Centre for Nuclear and Accelerator Technologies (CENTA) was established at the university ten years ago, comprising a tandem accelerator laboratory that has been used intensively for ion beam and mass spectrometry analyses [25,26,27,28], including AMS measurements of 10Be [29] and testing of uranium target matrices [30]. On top of that, the application of ICPMS for elemental composition determinations has been frequently performed at the Faculty of Natural Sciences of the Comenius University [31, 32].
LEGEND-200 shall start acquiring data soon, which means that all important parts of the detectors have been analyzed for the content of critical radionuclides and have met the required limits. However, one can reasonably anticipate that some of the parts may need replacement or upgrade as the experiment proceeds with time. Moreover, the LEGEND-1000 phase is only in its early development stage; many of the materials selected for the commissioning of the experiment will first have to be screened. Thus, the radiopurity assessment program of LEGEND is far from being over and should be intensified with the increasing demand for (ultra)clean materials. Here we present recent developments in thorium and uranium determinations by ICPMS and AMS, together with basics of separation chemistry applied to process different sample matrices. The new possibility to measure 232Th and 238U by ICPMS and the development of AMS at the Comenius University in Bratislava is also discussed.
Developments in sample preparation for natural Th and U mass spectrometry analysis
In collaboration with LNGS [33, 34], we have developed a simple extraction chromatography method based on the simultaneous separation of Th and U from metallic matrices with the use of TRU Resin (TrisKem International) that can now be utilized in the CENTA sample preparation laboratory. In short, sample is rinsed/etched to remove surface contamination and dissolved in ultrapure concentrated inorganic acid (HNO3/HCl). Diluted sample is loaded on a pre-treated TRU Resin chromatographic column which is then washed to separate matrix. Consequently, thorium and uranium are eluted by ammonium oxalate solution, which is then analyzed directly by ICPMS. Another option is to further process solution to obtain a target for AMS which we indeed exploited in our next development phase (see below). The method described above has been proven to reliably extract thorium and uranium with a yield of ≥ 90%, regardless the type of sample and its processed amount. Various types of solid materials (powders, scobs, wires, and pellets) have been treated to efficiently recover Th and U from metallic materials such as copper or aluminum; with slight modifications, the method could be usable for an even wider spectrum of metals/alloys, e.g., iron or steel, which are often exploited in experiments probing for rare nuclear processes. Procedural blanks on the level of ≤ 0.8 pg g−1 for 232Th and 238U can be achieved if one takes suitable countermeasures to prevent contamination of samples during their treatment (e.g., working in a clean environment, using only ultrapure chemicals, water and whole-body protection, extensive rinsing of glassware/labware, control of extraction chromatography resins). Specific precautions are needed for AMS; target holders (cathodes) and targets wheel, which are sputtered together with samples during the ion production, should be made of high-purity material. Very good quality was confirmed in the case of copper AMS target holders manufactured by National Electrostatic Caorporation (NEC, Middleton, USA), showing mean 232Th and 238U concentrations ~ 2 pg g−1.
Polymeric (plastic) materials are quite common parts of large-scale experiments such as LEGEND but generally require a different approach in treatment for radiopurity measurements than metallic components. We have tested a procedure for processing various polymeric foils (three Mylar and two polyethylene samples), with a main goal of determining the memory effect of containers used for combustion and its contribution to overall 232Th and 238U blanks. Samples were combusted in fused quartz crucibles (Thermo Fisher Scientific) in a furnace at 650 °C for 6 h; residual material was dissolved by the addition of 10 mL of 10% HNO3 to the crucible, which was then heated to 80 °C for 1 h. Consequently, two volumes of 10 mL 10% HNO3 were heated under the same conditions in the empty crucibles prior to each combustion and after the last one to obtain blank solutions that were later analyzed by ICPMS at LNGS, together with sample solutions. The results showed that the memory effect of the crucibles is more or less negligible and that they can contribute to contamination by thorium and uranium with an average value of 1.2–2.7 pg g−1 (Fig. 1) if we consider only the first two rinses as they were not affected by the subsequent combustions. Interestingly, the fused quartz crucibles seemed to be a greater source of 232Th than 238U.
Developments of AMS for natural Th and U determination
The determination of ultralow concentrations of 232Th and 238U by AMS is one of the very important topics in radiopurity assessment due to the prevailing demand for high-purity materials that can only be effectively screened by mass spectrometry techniques. Therefore, we have taken the first steps in the development of this method in collaboration with the CIRCE laboratory, which has a long history of AMS measurements of heavy isotopes [35, 36]. For testing purposes, we prepared a series of targets from solutions from the previous development step that contained known concentrations of natural thorium and uranium (verified by ICPMS at LNGS) and that were spiked with the 229Th tracer. The preparation was based on co-precipitation of radionuclides of interest on the iron oxide matrix. This was achieved by addition of Fe3+ carrier, purified from 232Th and 238U using UTEVA Resin (TrisKem International), to the solutions and raising their pH to 8–10 with concentrated NH4OH. The obtained precipitates were centrifuged, dried, baked at 550 °C and pressed in copper target holders (from NEC) which were then analyzed by AMS.
Measurements of thorium and uranium targets were conducted with the 3 MV tandem accelerator system at CIRCE. The tuning of the AMS system was performed using 238U from standard cathodes containing pure uranium oxide of a natural origin, however, we decided to detect 235U ions from samples instead as their 238U signal outreached the countrate limit of the ionization chamber; the 238U levels were calculated according to the natural isotopic ratio 235U/238U ≈ 7.26 × 10–3. Due to the unavailability of a suitable uranium tracer (e.g., 233U), we determined the isotopic ratios 229Th/232Th and 229Th/235U while applying corrections for different production rates of thorium and uranium ions in the source. For injection, molecular oxides Th16O− and U16O− were selected which were then stripped to single atoms with a charge state of 5 + and accelerated to the energies 16.70, 16.50 and 16.28 MeV for 229Th5+, 232Th5+ and 235U5+ ions, respectively (Fig. 2).
A comparison of the results from the AMS and ICPMS measurements of 232Th and 238U in the testing samples is shown in Fig. 3. The values represent the total amounts of 232Th and 238U in the AMS targets and in the solutions analyzed by ICPMS, discussed in the previous chapter. Unlike for 232Th, where its chemical yield was controlled by 229Th, the transfer of 238U from solutions to targets was only anticipated to be efficient. However, a reasonably good correlation between the methods (r = 0.9967 if obvious outliers are excluded) suggests that the uranium yield was indeed very high. Except for one case, the outliers represent samples whose analysis might have been affected by their contamination during the pressing, or by crosstalk and memory effect of the target wheel; the issue was addressed in the case of later samples. The relative statistical uncertainties of the AMS determinations were in the range of 7.2–10.3% for 232Th and 6.8–19.1% for 238U; slightly higher values in the case of uranium were probably caused by the absence of a tracer. By measuring one of the procedural blank solutions, we determined the limit of detection to be 0.4 pg of actinide per target. From the total detected counts for this sample, we estimate the instrumental limit of detection to be one order of magnitude lower.
The aforementioned experiments have the potential to serve as a solid basis for further development of AMS analysis of natural thorium and uranium. In August 2022, the CENTA tandem laboratory was upgraded with a new AMS line supplied by NEC, which comprises a 90° analyzing electromagnet with a high resolution (ME/Z2 = 176 MeV-amu), two 45° electrostatic analyzers, and a 4-anode ionization chamber for detection of ions of interest, including actinides (Fig. 4). The entrance window of the ionization chamber can be mounted by foils of different materials to suppress isobaric interferences. Another important part of the installation was a fast-bouncing system that enables switching between injected ions in a matter of microseconds, which is essential for precise measurements of isotopic ratios. Moreover, the accelerator has been equipped with an electrostatic quadrupole that significantly improved the possibility to focus the ion beam after its acceleration. After the necessary setting and optimization of the upgraded tandem system, we shall follow up on the already obtained results and finalize the development of the AMS method for ultra-low determination of 232Th and 238U for material screening in the near future.
ICPMS capabilities for natural Th and U determination
Several companies have been developing and commercializing stand-alone ICPMS instruments for elemental and isotopic analysis which, in comparison to AMS, have had a lower complexity and a simpler operation. At the Faculty of Natural Sciences, there are two ICPMS machines in operation that can be utilized for the determination of thorium and uranium. The first one is a quadrupole mass spectrometer PerkinElmer Elan 6000 with a simultaneous extended dynamic range detector (dual-stage discrete dynode electron multiplier). The instrument is capable of effectively measuring almost all isotopes (up to 270 amu), and it is equipped with a HF-resistant Scott-type spray chamber, which can be especially useful when dealing with samples that require the application of highly concentrated hydrofluoric acid for their dissolution. (e.g., pure germanium oxide intended for the production of detectors). Even though these types of spectrometers have only one mass analyzer and do not contain any reaction/collision cell for suppression of isobaric interferences, their limit of detection for 232Th and 238U determinations (~ 1 pg g−1) is generally sufficient for radiopurity measurements of non-critical components.
In 2016, the university acquired a new ICPMS machine Thermo Scientific iCap Q which works with a single quadrupole, a He collision cell, and a dual-mode discrete dynode electron multiplier for ion detection. The spectrometer has been shown to provide precise and accurate results for natural thorium and uranium at μg L−1/μg kg−1 levels, however, the declared limits are by a few orders of magnitude lower. To test them, we prepared and analyzed a series of calibration solutions containing 0–250 pg mL−1 of thorium and uranium. As expected, we obtained a sensitivity of ~ 1700 and 3300 cps per 10 pg mL−1 of 232Th and 238U in the standard mode (STD), respectively, while these values were found to be approximately 30% higher when the helium collision cell was used. According to the calibration curves shown in Fig. 5, we were able to achieve the instrumental limit of detection of 0.39 pg mL−1 for 232Th and 0.023 pg mL−1 for 238U in the standard mode. Thorium concentrations were determined with rather high standard deviations; however, this can be addressed by setting a longer time for stabilization of the detected signal.
Outlook
LEGEND represents the state-of-art experiment that has undoubtedly a potential to prove the existence of neutrinoless double beta decay. The success of the collaboration depends on many things, one of them being the ability to design and construct a detector that operates with virtually no background radiation. This strict demand has been fulfilled in the case of LEGEND-200, which should provide first results in the upcoming years. On the other hand, the task will be more challenging for LEGEND-1000 which is planned to work with five times more mass of germanium while at the same time having the background lower by one order of magnitude. After the start of its main part, the screening campaign for LEGEND-1000 will take some time before all appropriate materials will be identified, even if it is expected that at least twenty gamma spectrometry, neutron activation analysis and mass spectrometry facilities could participate in radiopurity measurements. Due to their ubiquity and relatively high abundance, naturally occurring thorium and uranium will be one of the most important radionuclides to target.
Comenius University in Bratislava has recently invested in the development of mass spectrometry techniques which have become mandatory for 232Th and 238U analyses in the case of large physical experiments searching for rare nuclear processes. Because these methods require samples to be treated and transformed into solutions or targets, we have tested procedures for the decomposition of metallic and polymeric materials and the consequent separation of thorium and uranium. Regarding the analytical infrastructure, two ICPMS instruments are now available to screen samples with actinide concentrations of > 1 pg g−1 or even less, depending on the quality of simultaneously prepared blanks. It is assumed that this limit could be further enhanced by utilizing the AMS line that has been recently installed at the CENTA tandem laboratory, bringing to the material screening program of the LEGEND collaboration another possibility for ultra-trace analysis of natural thorium and uranium. It is planned that the CENTA AMS laboratory will be the first world laboratory devoted to the radiopurity research.
Data availability
No additional data sets were generated or analyzed during the current study.
References
Arnold R et al (2010) Probing new physics models of neutrinoless double beta decay with SuperNEMO. Eur Phys J C 70:927–943. https://doi.org/10.1140/epjc/s10052-010-1481-5
Alfonso K et al (2015) Search for neutrinoless double-beta decay of 130Te with CUORE-0. Phys Rev Lett 115:102502. https://doi.org/10.1103/PhysRevLett.115.102502
Ebert J et al (2016) The COBRA demonstrator at the LNGS underground laboratory. Nucl Instrum Methods Phys Res Sect A Accel Spectrom Detect Assoc Equip 807:114–120. https://doi.org/10.1016/j.nima.2015.10.079
Albert JB et al (2018) Sensitivity and discovery potential of the proposed nEXO experiment to neutrinoless double-β decay. Phys Rev C 97:065503. https://doi.org/10.1103/PhysRevC.97.065503
Armengaud E et al (2020) The CUPID-Mo experiment for neutrinoless double-beta decay: performance and prospects. Eur Phys J C 80:44. https://doi.org/10.1140/epjc/s10052-019-7578-6
Manly S et al (2021) Deep underground neutrino experiment (DUNE) near detector conceptual design report. Instruments 5:31. https://doi.org/10.3390/instruments5040031
Abgrall N et al (2017) The large enriched germanium experiment for neutrinoless double beta decay (LEGEND). AIP Conf Proc 1894:020027. https://doi.org/10.1063/1.5007652
Patrignani C et al (2016) Review of particle physics. Chinese Phys C 40:100001. https://doi.org/10.1088/1674-1137/40/10/100001
Agostini M et al (2016) Search of Neutrinoless Double Beta Decay with the GERDA Experiment. Nucl Part Phys Proc 273–275:1876–1882. https://doi.org/10.1016/j.nuclphysbps.2015.09.303
Abgrall N et al (2014) The MAJORANA DEMONSTRATOR neutrinoless double-beta decay experiment. Adv High Energy Phys 2014:1–18. https://doi.org/10.1155/2014/365432
Burlac N, Salamanna G (2021) Present and Future of 0ν2β Searches with Germanium. Universe 7:386. https://doi.org/10.3390/universe7100386
LEGEND Collaboration (2021) LEGEND-1000 Preconceptual Design Report. http://arxiv.org/abs/2107.11462
Laubenstein M, Hult M, Gasparro J, Arnold D, Neumaier S, Heusser G, Köhler M, Povinec P, Reyss J-L, Schwaiger M, Theodórsson P (2004) Underground measurements of radioactivity. Appl Radiat Isot 61:167–172. https://doi.org/10.1016/j.apradiso.2004.03.039
Povinec PP (2019) Ultra-sensitive radionuclide analyses: new frontiers in radioanalytics. J Radioanal Nucl Chem. https://doi.org/10.1007/s10967-019-06707-2
Laubenstein M, Lawson I (2020) Low background radiation detection techniques and mitigation of radioactive backgrounds. Front Phys 8:1–7. https://doi.org/10.3389/fphy.2020.577734
Povinec PP (2018) New ultra-sensitive radioanalytical technologies for new science. J Radioanal Nucl Chem 316:893–931. https://doi.org/10.1007/s10967-018-5787-3
Ferella F, Nisi S, Balata M, Grabmayr P, Laubenstein M, Schwingenheuer B, Schönert S (2023) Enriched high purity germanium detectors for the LEGEND-200 experiment: purification and characterization by quadrupole and high resolution inductively coupled plasma mass spectrometry (ICP-MS). Appl Radiat Isot 200:110904. https://doi.org/10.1016/j.apradiso.2023.110904
Gileva O, Choe J, Kim Y, Lee MH, Leonard DS, Shin K, Yeon H (2023) Thorium and uranium trace ICP-MS analysis for AMoRE project. Appl Radiat Isot 194:110673. https://doi.org/10.1016/j.apradiso.2023.110673
Famulok N, Faestermann T, Fimiani L, Gómez-Guzmán JM, Hain K, Korschinek G, Ludwig P, Schönert S (2015) Ultrasensitive detection method for primordial nuclides in copper with accelerator mass spectrometry. Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 361:193–196. https://doi.org/10.1016/j.nimb.2015.03.042
Sýkora I, Ješkovský M, Janik R, Holý K, Chudý M, Povinec PP (2008) Low-level single and coincidence gamma-ray spectrometry. J Radioanal Nucl Chem 276:779–787. https://doi.org/10.1007/s10967-008-0632-8
Ješkovský M, Lištjak M, Sýkora I, Slávik O, Povinec PP (2018) Anthropogenic 137Cs on atmospheric aerosols in Bratislava and around nuclear power plants in Slovakia. J Environ Radioact 184–185:77–82. https://doi.org/10.1016/j.jenvrad.2018.01.007
Povinec PP, Kontuľ I, Lee S-H, Sýkora I, Kaizer J, Richtáriková M (2020) Radiocarbon and 137Cs dating of wines. J Environ Radioact 217:106205. https://doi.org/10.1016/j.jenvrad.2020.106205
Ješkovský M, Frekers D, Kováčik A, Povinec PPP, Puppe P, Staníček J, Sýkora I, Šimkovic F, Thies JHH (2015) A search for double-electron capture of 74Se to excited levels using coincidence/anticoincidence gamma-ray spectrometry. Nucl Instrum Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip 795:268–275. https://doi.org/10.1016/j.nima.2015.05.071
Povinec PP, Sýkora I, Porubčan V, Ješkovský M (2009) Analysis of 26Al in meteorite samples by coincidence gamma-ray spectrometry. J Radioanal Nucl Chem 282:805–808. https://doi.org/10.1007/s10967-009-0211-7
Povinec PP, Masarik J, Kúš P, Holý K, Ješkovský M, Breier R, Staníček J, Šivo A, Richtáriková M, Kováčik A, Szarka J, Steier P, Priller A (2015) A new IBA-AMS laboratory at the Comenius University in Bratislava (Slovakia). Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 342:321–326. https://doi.org/10.1016/j.nimb.2014.10.011
Povinec PP, Masarik J, Ješkovský M, Kaizer J, Šivo A, Breier R, Pánik J, Staníček J, Richtáriková M, Zahoran M, Zeman J (2015) Development of the accelerator mass spectrometry technology at the Comenius University in Bratislava. Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 361:87–94. https://doi.org/10.1016/j.nimb.2015.02.021
Ješkovský M, Kaizer J, Kontuľ I, Kvasniak J, Pánik J, Zeman J, Povinec PP (2022) Recent developments in IBA analysis at CENTA, Bratislava. EPJ Web Conf 261:01002. https://doi.org/10.1051/epjconf/202226101002
Povinec PP, Masarik J, Ješkovský M, Breier R, Kaizer J, Pánik J, Richtáriková M, Staníček J, Šivo A, Zeman J (2016) Recent results from the AMS/IBA laboratory at the Comenius University in Bratislava: preparation of targets and optimization of ion sources. J Radioanal Nucl Chem 307:2101–2108. https://doi.org/10.1007/s10967-015-4406-9
Ješkovský M, Steier P, Priller A, Breier R, Povinec PP, Golser R (2015) Preliminary AMS measurements of 10Be at the CENTA facility. Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 361:139–142. https://doi.org/10.1016/j.nimb.2015.04.072
Povinec P, Svetlik I, Ješkovský M, Šivo A, John J, Špendlíková I, Němec M, Kučera J, Richtáriková M, Breier R, Fejgl M, Černý R (2015) Joint Bratislava-Prague studies of radiocarbon and uranium in the environment using accelerator mass spectrometry and radiometric methods. J Radioanal Nucl Chem 304:67–73. https://doi.org/10.1007/s10967-014-3618-8
Bujdoš M, Hagarová I, Matúš P, Canecká L, Kubová J (2012) Optimization of determination of platinum group elements in airborne particulate matter by inductively coupled plasma mass spectrometry. Acta Chim Slovaca 59:124–128
Duborská E, Urík M, Bujdoš M, Matulová M (2019) Influence of physicochemical properties of various soil types on iodide and iodate sorption. Chemosphere 214:168–175. https://doi.org/10.1016/j.chemosphere.2018.09.041
Nisi S, Copia L, Dafinei I, Di Vacri ML (2017) ICP-MS measurement of natural radioactivity at LNGS. Int J Mod Phys A 32:1743003. https://doi.org/10.1142/S0217751X17430035
Kaizer J, Nisi S, Povinec PP (2019) Radiopurity measurements of aluminum, copper and selenium materials for underground experiments and mass spectrometry development. J Radioanal Nucl Chem 322:1447–1454. https://doi.org/10.1007/s10967-019-06857-3
De Cesare M, Fifield LK, Sabbarese C, Tims SG, De Cesare N, D’Onofrio A, D’Arco A, Esposito AM, Petraglia A, Roca V, Terrasi F (2013) Actinides AMS at CIRCE and 236U and Pu measurements of structural and environmental samples from in and around a mothballed nuclear power plant. Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 294:152–159. https://doi.org/10.1016/j.nimb.2012.05.020
De Cesare M, Gialanella L, Rogalla D, Petraglia A, Guan Y, De CN, D’Onofrio A, Quinto F, Roca V, Sabbarese C, Terrasi F (2010) Actinides AMS at CIRCE in Caserta (Italy). Nucl Instrum Methods Phys Res Sect B Beam Interact Mater Atoms 268:779–783. https://doi.org/10.1016/j.nimb.2009.10.029
Acknowledgements
This publication is the result of support under the Operational Program Integrated Infrastructure for the project: Advancing University Capacity and Competence in Research, Development and Innovation (ACCORD, ITMS2014+:313021X329), co-financed by the European Regional Development Fund. The authors also acknowledge the support provided by the Slovak Research and Development Agency (projects no. APVV-21-0377 and APVV-15-0576).
Funding
Open access funding provided by The Ministry of Education, Science, Research and Sport of the Slovak Republic in cooperation with Centre for Scientific and Technical Information of the Slovak Republic.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Kaizer, J., Bujdoš, M., Buompane, R. et al. Mass spectrometry developments of 232Th and 238U radiopurity measurements for LEGEND. J Radioanal Nucl Chem 333, 3431–3437 (2024). https://doi.org/10.1007/s10967-024-09352-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-024-09352-6