Introduction

Liquid scintillation counting (LSC) is a widely used radiometric technique for the measurement of alpha and beta emitters present in environmental samples and materials coming from decommissioning activities [1]. This technique makes use of liquid scintillation cocktails and one of the important components of these cocktails is the surfactant [2]. Commonly, phenol ethoxylates are used as surfactants in the commercially available cocktails [3]. These chemicals are however recently added in the REACH (Registration, Evaluation, Authorization and Restriction of Chemicals)—list of amendments to Annex XIV, as chemicals which can’t be used anymore [4] for routine laboratory work. A first study which reported the performance of an NPE-free cocktail was published by Varlam et al. in 2019 [3]. They report about the performance of Prosafe LT + cocktail for 3H determination in aqueous samples and concluded that this cocktail fulfilled the criteria needed for their routine analyses [3]. Radionuclide metrology is very much relying on liquid scintillation counting where two models, the CIEMAT/NIST approach and the TDCR method are commonly used in radionuclide standardization [5].

Recently, also another approach in scintillation counting is becoming promising which is based on plastic scintillation microspheres/resins (PSm and PSresin respectively). Several publications report the application of PSm/PS-resins for quantification of the different radionuclides in the environmental samples [6,7,8,9,10].The advantages and disadvantages of PSm are described in detail in Handbook of Radioactivity Analyses, 2020 [11]. Reducing the waste costs, high counting efficiencies also for alpha-emitter radionuclides, possibility of reuse them, possibility to impregnate them with different extractants and then combine the separation and the measurement are only a few of the many advantages [11].

At SCK CEN (LRM) an extensive study was performed in order to find/select the best available cocktail(s) that are NPE (nonylphenol ethoxylates) free, to replace NPE based cocktails currently used for routine measurements. A total of nine cocktails from several producers were selected and investigated in this study. Different parameters were tested, such as homogeneity/compatibility with the sample material, inherent background level and the influence on counting efficiency. The specific media obtained after chemical separations of the sample, resulting in a fraction containing a single radionuclide were investigated for several radionuclides (such as 90Sr, 210Pb, 63Ni, 99Tc, 14C in Carbosorb E), but also common sample materials (such as water, clay water, urine) where direct measurement of several radionuclides (such as 3H or 14C) after mixing with the cocktail were investigated. This paper will give an overview of obtained results leading to the selection of the best cocktail(s) for our routine applications in order to replace the Optiphase HiSafe 3 cocktail that is currently in use and selected based on a similar study of Verrezen et al. [12].

Experimental

Materials

Cocktails

The overview of the different types of cocktails used in this study is given in Table 1. A total number of 9 cocktails was bought from different producers. All of them are NPE-free cocktails and only one of them is DIN—free, namely Pico Fluor plus. Since one of the most important components of the liquid scintillation cocktail is the surfactant, the mixture of these chemicals present in the studied cocktails are given also in Table 1. The role of the surfactant is to ensure a homogeneous mixture between the organic phase of the cocktail and the aqueous phase where the radionuclide to be determined is present by forming a micro-emulsion [13].

Table 1 Overview of the different types of cocktails used in this study

Tracers and matrices

All tracers used in this study were traceable to SI unit and an overview is given in Table 2 together with the type of matrix specific to each investigated radionuclide. Traceability of the radioactivity (Bq) was assured by using working solutions resulting from gravimetrically made dilutions of the certified reference solutions.

Table 2 Overview of the tracers and type of matrices investigated in this study

All chemicals used in this study, such as acids or chemicals used for preparation of final measurement sources (e.g. Sr or Pb—oxalate) are of analytical purity.

All the samples were measured using a low-level liquid scintillation counter, Quantulus 1220™.

Liquid scintillation vials

Low potassium borosilicate vials were used for visual evaluation of the homogeneity and polyethylene vials were used for measurements.

Methods

The approach followed in this study for the final selection of best NPE-free cocktail(s) for our routine analyses is presented in Fig. 1. The first step was identification of the candidate cocktails (see Table 1), then several parameters were considered, such as homogeneous mixture with a direct sample (urine, clay water, RO-water) or with a specific media resulting from a radiochemical separation of the considered radionuclide (see Table 2). Another parameter investigated was FOM (Figure of Merit), defined as E2/B (E is counting efficiency, B is background level) [2]. The final volume of all the mixtures was 20 ml and the measurement temperature 10 °C. For background level determination the counting time was 100 min and for counting efficiency 30 min.

Fig. 1
figure 1

The decision approach considered in this study for selection of NPE-free cocktail(s)

Homogeneity using direct mixing and specific media with the cocktails

Blank Reverse Osmosis (RO)—water, urine and clay water samples were directly mixed with the cocktails in different sample to cocktail ratios (sample load). The sample load was defined by the ratio: V(sample)/V(total), where V(total) = V(sample) + V(cocktail). The mixing was performed by hand shaking using glass vials. After mixing, the samples were placed in the LS counter for cooling down for at least 1 h. Afterwards the homogeneity of the resulting mixture was visually evaluated (phase separation, turbidity of the mixture, etc.) and the corresponding loading capacity of the cocktail was determined (as the ratio of the volume of the sample in the mixture to the total volume of the mixture). The specific matrices (obtained after radiochemical separation), were also mixed with the cocktails and the homogeneity of the resulting mixtures was evaluated.

Figure of merit (FOM) evaluation

The FOM is defined as E2/B (E is counting efficiency, B is background level) [2]. FOM was chosen as a parameter to compare the performance of the investigated cocktails. The samples which successfully fulfilled the condition of being a homogeneous mixture, were measured using liquid scintillation counting in order to determine the background level. A similar mixture was prepared and spiked with the corresponding tracers in order to determine the counting efficiency.

All the results were compared with the presently used cocktail Optiphase HiSafe 3 (contains NPE). This cocktail was selected in the past based on a similar study described by Verrezen et al. [12].

Results and discussion

Homogeneity using direct mixing and loading capacity

Clay water samples

In Fig. 2, the degree of homogeneity of the investigated cocktails with clay water samples is compared with OptiPhase HiSafe 3. In order to keep the detection limit (DL) for 3H (13.76 Bq l−1) and 14C (3.57 Bq l−1) in clay water samples comparable with the one that is obtained currently using Optisafe HiSafe 3 cocktail, a minimum sample load of 25% is needed. As we can see from Fig. 2, six cocktails fulfill this condition. Two of them, namely Prosafe FC +, Prosafe TS + gave a milky mixture and one of them, namely ProSafe +, gave a turbid but still homogeneous mixture. Our findings are in agreement with the data provided by the manufacturer [14]. Mixtures with a milky appearance or with the occurrence of phase separation are considered to be not acceptable for LSC counting.

Fig. 2
figure 2

Overview of the homogeneity and loading capacity of the 9 investigated cocktails using clay water sample

Urine samples

A loading capacity of at least 20% is needed in order to maintain the same detection limits achieved with OptiPhase HiSafe 3. In Fig. 3, the homogeneity degree of the 9 selected cocktails for urine samples is presented. As can be observed five of the nine cocktails meet the condition of 20% loading capacity. Two of them, namely QuickSafe flow2 + and ProSafe FC +, only reach a loading capacity of 15 and 10%, respectively. ProSafe + and ProSafe TS +, can reach the 20% sample load, however at this value the mixture is turbid but still homogeneous. Our findings are in agreement with the data provided by the manufacturer for ProSafe HC +, Prosafe TS + and ProSafe + cocktails [14].

Fig. 3
figure 3

Overview of the homogeneity and loading capacity of the 9 investigated cocktails using urine sample

Reversed osmosis water samples

For reversed osmosis water a maximum loading capacity is needed for our laboratory in order to obtain a comparable detection limit with the samples mixed with OptiPhase HiSafe 3. In Fig. 4 the homogeneity degree of different sample load is shown for each of the 9 selected cocktails. As can be observed, only one cocktail, namely QuickSafe flow2 +, meets the condition of 45% loading capacity. ProSafe LT + and PicoFluor Plus, achieve a 40% loading capacity with a homogeneous mixture but a turbid appearance. Aqualight + cocktail achieved a 35% loading capacity with a homogeneous but turbid sample. The remaining five cocktails: ProSafe +, ProSafe FC +, Proflow G +, ProSafe HC +, ProSafe TS +, do not produce a homogeneous sample when loaded with at least 35% sample load. Our findings are in agreement with the data provided by the manufacturer, exept for ProSafe HC + cocktail, where we obtained a loading capacity using RO-water of 25% [14].

Fig. 4
figure 4

Overview of the homogeneity and loading capacity of the 9 investigated cocktails using RO water sample

Based on the evaluations described above, it was decided that three cocktails, namely ProSafe +, ProSafeFC +, ProSafe TS + were not considered for further testing using specific media. These cocktails didn’t fulfill the requirements for the minimum loading capacity. The remaining six cocktails were further tested for homogeneity using specific media resulting from a radiochemical separation. However, three of the cocktails, namely ProSafe LT +, Pico-Fluor plus and Aqualight +, even if they gave a turbid, but homogeneous, mixture were considered for further testing since obtaining a slightly lower counting efficiency did not compromise the final DL. Due to the turbidity, the counting efficiency was mainly affecting 3H determinations.

As can be observed in Table 1, the main surfactant used in all the LS cocktails is based on ethoxylates. As co-surfactants, different chemicals were used, such as sodium dioctyl sulphosuccinate (in 5 of the studied cocktails), phosphate esters (in 7 of the studied cocktails) and diethanolamine salt (in 3 of the studied cocktails). The sulphosuccinates are known to improve the micro-emulsion formation, however the performance of the phosphate esters is better especially for difficult matrices [12]. When the two best performing cocktails from Fig. 4 are compared, the highest sample load is obtained with Quicksafe Flow 2 + followed by, Prosafe LT +. The first cocktail contains phosphate ester, while the second one contains dioctyl sulphosuccinate. Pico-Fluor plus cocktail contains both chemicals, phosphate ester and dioctyl sulphosuccinate, but the sample load is lower than Quickflow 2 +.

Homogeneity using specific media (after radiochemical separation) with the cocktails

In Table 2 (see “Materials” and “Methods” sections) the specific media used for this study are given for each radionuclide investigated. The results of these tests are given in Table 3. The cocktails are selected based on visual evaluation of the homogeneity of the mixture. As can be seen, Carbosorb E, Quicksafe flow2 + and AquaLight + cocktails can’t be used since a turbid mixture is obtained. Picofluor plus cocktails gave turbid mixtures with the final medium for 99Tc (5 ml of 0.5 M HNO3) and 210Pb (1 ml 6 M HNO3). ProSafe HC + was also not compatible with the media of 90Sr (0.05 M HNO3). All the other cocktails meet the homogeneity requirement. All six cocktails were considered further for determination of the background level and counting efficiency.

Table 3 Overview of the selected cocktails which fulfill the homogeneity requirement

Figure of merit (FOM) evaluation

Based on measurements of a blank sample and the counting efficiency, the FOM parameter was determined for different matrices and radionuclides. The uncertainty components which are taken into account for the final expanded uncertainty calculation are: weighing of the tracer, activity of the tracer (from the certificate), dilution factor, counting statistics of the blanc and the sample. The obtained values are compared with the values that we presently use in our routine procedure. In order to take a decision, we selected the three best performing cocktails.

Clay water samples

In Figs. 5 and 6, FOM for clay water samples and 3H and 14C (sample load 25%), respectively, are presented. We decided to select the AquaLight +, ProSafe LT +, ProSafe HC + (see Table 4), since the FOM values are close to the one obtained with OptiPhase Hisafe 3.

Fig. 5
figure 5

Overview of FOM performance for 3H in clay water samples (sample load 25%) using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Fig. 6
figure 6

Overview of FOM performance for 14C in clay water samples (sample load 25%) using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Table 4 Overview of the selected cocktails based on FOM evaluation for different matrices

Urine samples

In Figs. 7 and 8, FOM for urine samples containing 3H and 14C, respectively, at 20% sample load are presented. We decided to select the Pico-fluor plus, ProSafe LT +, AquaLight + (see Table 4), since the FOM values are close to the one obtained with OptiPhase Hisafe 3.

Fig. 7
figure 7

Overview of FOM performance for 3H in urine samples using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Fig. 8
figure 8

Overview of FOM performance for 14C in urine samples using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Reversed osmosis water

In Figs. 9 and 10, FOM for RO water samples, 3H and 14C, and the selected cocktails based on homogeneity study are presented. We decided to select the Picofluor plus, ProSafe LT + and Aqualight + cocktails(see Table 4), since the FOM values are close to the one obtained with OptiPhase Hisafe 3.

Fig. 9
figure 9

Overview of FOM performance for 3H in RO water using four cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2). For every cocktail the loading capacity is mentioned in the figure

Fig. 10
figure 10

Overview of FOM performance for 14C in RO water using four cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2). For every cocktail the loading capacity is mentioned in the figure

Specific media and radionuclides

When evaluating the FOM for specific matrix and radionuclides we observed two distinct groups of radionuclides. Based on this, we decided to perform the evaluation of FOM in two groups: (1) 14C in Carbosorb E, 63Ni and 36Cl and (2) 99Tc, 90Sr and 210Pb.

Evaluation of FOM for 14C in Carbosorb E, 63Ni and 36Cl

The results of the FOM for 14C, 63Ni and 36Cl are presented in Figs. 11, 12, and 13. For 14C in Carbosorb, the only cocktail which gave comparable results with the currently used OptiPhase Hisafe 3 is the Picofluor plus cocktail. The rest of the cocktails gave much lower results. In case of 63Ni and 36Cl comparable results with OptiPhase Hisafe 3 were obtained for Pico-fluor plus, ProSafe LT +, AquaLight +.

Fig. 11
figure 11

Overview of FOM performance for 14C trapped in Carbosorb E using four cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Fig. 12
figure 12

Overview of FOM performance for 63Ni using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Fig. 13
figure 13

Overview of FOM performance for 36Cl using six cocktails (grey bars) and compared with routinely used OptiPhase Hisafe3 (black bar). Uncertainty is given as expanded uncertainty (k = 2)

Evaluation of FOM for 99Tc, 90Sr and 210Pb

The results of the FOM for 99Tc, 90Sr and 210Pb are presented in Figs. 14, 15, and 16. For 99Tc comparable results with OptiPhase Hisafe 3 were obtained for AquaLight +, ProSafe LT +, Quicksafe flow 2 +. For 90Sr, AquaLight +, Pico-fluor plus and Prosafe HC + cocktails were the three giving results close to the OptiPhase Hisafe 3, while for 210Pb, Proflow G +, Prosafe HC + and Quicksafe flow 2 + gave results much better than the one currently used.

Fig. 14
figure 14

Overview of FOM performance for 99Tc using five cocktails (grey bars) and compared with routinely used OptiPhase HiSafe3 (black bar). Uncertainty is given as expanded unceratinty (k = 2)

Fig. 15
figure 15

Overview of FOM performance for 90Sr using five cocktails (grey bars) and compared with routinely used OptiPhase HiSafe3 (black bar). Uncertainty is given as expanded unceratinty (k = 2)

Fig. 16
figure 16

Overview of FOM performance for 210Pb using six cocktails (grey bars) and compared with routinely used OptiPhase HiSafe3 (black bar). Uncertainty is given as expanded unceratinty (k = 2)

Conclusions

Based on the Table 4 where all the “best performing” cocktails for the different matrices and radionuclides are summarized, we conclude that ProSafe LT + and Aqualight + cocktails fulfill best our criteria (minimum sample load, type of matrix, needed detection limit). The composition of these two cocktails is similar, both contain dioctyl sulphosuccinate as co-surfactant. Picofluor plus was selected as the best performing cocktail to be used with Carbosorb E. However, for 90Sr and 210Pb we observe that the candidate cocktail ProSafe LT + performance is inferior than the one presently used in routine measurements. Even so, we decided to use Prosafe LT + for all our routine analyses. Picofluor plus cocktail also can be considered as one of the best performing scintillation cocktails. Since the legislation is becoming more and more restrictive regarding the chemicals to be allowed for use in the production of the liquid scintillation cocktails, maybe another alternative can be considered for this type of measurements, such as plastic scintillation microspheres (PSm) or plastic scintillations resins (PS-resin).