Abstract
A common advanced radiochemical technique is the use of the AG® 1-X8 anion exchanger, Cl2− form (from Bio Rad), to separate uranium and/or thorium from a sample by ion exchange. This method is used to separate elements by chemical elution from an ion exchange column via a precipitate of substances (co-precipitation), with Nd3+ as a thin layer of smooth fluoride particles on a membrane filter, then using α-spectrometry to measure uranium and thorium. The obtained data showed that the column could be reused, at least twelve times, safely in separating uranium and thorium from environmental samples, before observing any change in the performance of the exchanger.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Appropriate management of radioactive wastes and their subsequent effect both on nature and on human welfare requires analytical procedures for delicate and reliable identification and determination of a wide assortment of radionuclides in a wide assortment of sample matrices. Other researchers have developed methodologies for isolating, precipitating, and mounting actinides for α-spectrometry [1,2,3], either separately or mixed, as either fluorides or hydroxides [4, 5]. In these methods, actinides were isolated from all other components of the sample and from each other by chromatography of ion exchange, co-precipitated with Nd3+ as fine fluoride particles, and mounted on membrane filters for α-spectrometry [6,7,8].
Since ion exchange columns are only used once, it is common for many of organic ion exchangers to be expended in these studies. An anion exchanger (AG1-X8, chloride form, 100–200 mesh, from Bio-Rad) is routinely used in radio-analytical techniques in analyzing uranium, thorium, and plutonium radionuclides in different types of environmental samples [1, 2, 6, 7, 9,10,11]. Large quantities of ground water samples are typically received for investigation. Radionuclides of uranium and radium, in addition to radon-222, are the main radionuclides of interest in ground water [12,13,14,15] and the use of the exchanger is mainly due to the routine work of uranium analysis in ground water [16,17,18]. In the experimental, the equal amounts of the same sample were used in each run of the same column, and a blank sample was added to wash the exchanger after every three sample runs in the same renewal column to prepare it for measurement by using α-spectrometry. The performance of the washed exchanger was examined by examining changes in the sample recovery and the blank sample counting rate. In a similar experiment to determine the suitability of this method for separating thorium, we used the same level of aqueous solutions using 8 M HNO3 media; the column could be reused at least 12 times without degradation of the measurement for U and Th.
The aim of the present article is to study the possibility of renewing the exchanger and reusing the column, with the goal of reducing the resin consumption rate and reducing cost.
Low-level samples require 24 h of counting time [4, 19,20,21]. In the experimental setup described means that only 8 samples can be counted per day, even though more samples may be prepared. This leads investigators to store samples for some time while waiting for counting. Storing samples in a dry atmosphere may lead to changes in the physical properties of the fluoride particles of the thin source, and hence may result in spectrum degradation [8, 22, 23]. However, the stability of the fine fluoride particles with time is also tested in this work.
Methodology and experiments
Standard reference materials
For this study, we used 232U and 229Th standard reference materials, purchased from the National Bureau of Standards, USA, under the code numbers STM 4324 and STM 4328, respectively. The activities for 232U and 229Th are 0.1557209291 Bq/ml (reference date 14/2/2002), and 0.332834375 Bq/ml (reference date 7/5/1984) respectively. The reference materials were diluted and used as a spiking tracer for analyzing uranium and thorium standard solutions [24, 25].
Uranium and thorium standard solutions
Standard radioactive solutions are used for chemical yield determinations during the chemical separation process [27]. They are usually supplied in a form of 5 g aqueous solutions, composed of the radionuclide salt in an acidic medium contained in glass ampule. Uranium and thorium standard solutions were prepared in the laboratory from uranyl chloride and thorium nitrate salts, respectively [28, 29]. For quality control, aliquots of the prepared standard solutions were spiked with the respective radiotracer and analyzed by α-spectrometry, which is shown in Fig. 1.
Uranium samples
In a 100 mL measuring flask, we combined 5 mL of seawater, an aliquot of uranium standard (13.5 pCi 238U), and enough concentrated HCl to get 10 M, and diluted it to volume with distilled water. This solution contained almost 2.7 pCi of 238U/20 mL of 10 M HCl; it was designated U-I. The second uranium solution was prepared exactly as U-I, except that 28 pCi 238U were added instead of 13.5 pCi. This solution contained almost 5.6 pCi 238U/20 mL of 10 M HCl and was designated U-II. The fixed amount of seawater was added to each solution to bring ions in solution comparable to those in the environmental samples. Seawater does add a constant amount of natural uranium and thorium to all the samples; however, this amount of activity is negligible compared to the concentration of uranium or thorium in the prepared samples.
Thorium samples
Two thorium solutions (2.5 and 5.1 pCi 232Th in 20 mL of 8 M HNO3, respectively) were prepared in the same manner as the uranium solutions except that 232Th and HNO3 were used instead of 238U and HCl, respectively. These solutions were respectively designated Th-I and Th-II.
Apparatus
For this project, we used an ORTEC Octete Plus high resolution α-spectrometry system from ORTEC. The device was connected to a vacuum pump and contained eight 450 mm2 silicon surface barrier detectors situated in isolated chambers. The background count rate ranged from 0 to 2 counts within 18 h under the energy peaks of 238U or 234U, producing an ultra-low detection limit of " < 0.01/Bq/L". The efficiency of the detectors ranged from 20 to 21.5%.
Analysis
Uranium and thorium were analyzed by chemical separation and co-precipitation as a thin fluoride sources for α-spectrometry according to the method described by Moghissi (1969) [6], which is shown in Fig. 2.
Briefly, in water samples, aliquots were first spiked with the radiotracer. Next, the solution evaporated to reach dryness. Then, the resulting salt was transformed into the suitable form.
To isolate uranium, the material in the chloride structure, which was spiked with 232U radiotracer (1 cpm/mL), was loaded in 20 mL 10 M HCl. A 10 mL sample was placed in the separation column (from Bio Rad), filled with AG® 1-X8 anion exchanger (5 mL bed volume); 20 mL of 10 M HCl was passed to precondition the column. Next, the uranium was washed with another 20 mL of 10 M HCl and eluted from the column by passing 20 mL of 0.1 M HCl. To prepare the uranium source, 50 μL of Nd3+ solution (1 mg Nd3+/mL) were added to the isolated uranium (the eluted sample), followed by an additional measure of TiCl3 (15%), drop by drop, to change the solution color to purple; at that point an additional 1 mL was added to keep uranium in the tetravalent state. The next step was to add the HF solution (5 mL of 40% of HF) to the sample and leave it for 30 min before filtering it with a 0.1 μL polypropylene membrane filter. The precipitate was washed with 3 mL of an 80% ethanol solution. Finally, the sample was dried at 50° C for five minutes, placed on stainless steel disc, and measured using α-spectrometer [3, 6, 26].
The thorium was isolated in the same manner as the uranium except that 8 M and 1 M HNO3 was used in loading and eluting the thorium instead of the 10 M and 0.1 M HCl used with the uranium, respectively. The source of thorium was also prepared as the uranium was, except there was no need for the addition of a TiCl3 solution as shown in Fig. 2.
Results and discussion
Equal fractions of 20 mL of each of the uranium and thorium solutions (U-I, U-II, Th-I, and Th-II) were repeatedly applied to each of four fixed columns (numbered I, II, III, and IV, respectively) packed with AG® 1-8X anion exchanger, 100–200 mesh, Cl2− form. In other words, equal fractions of the same sample were applied in each run for the same column.
As mentioned above, dilute acid was used to elute the retained uranium or thorium and regenerate the column for the next run. The eluted fraction was prepared for counting by α-spectrometry. Twenty-two runs were carried out through each column with the same solution. The column was regenerated between runs, and a blank sample was applied to each column after every three successive sample runs to follow the change in counting rate of the blank samples due to repeated column reuse. The performance of the column could be evaluated by following the change in the chemical yield of the eluted uranium or thorium fractions, in addition to the change in the counting rate of the eluted fraction of the blank sample.
The change in chemical yield with repeated column reuse is illustrated in Figs. 3 and 4. Two different concentrations were tested for uranium and for thorium to evaluate the effect of concentration on the performance of the reused column. The Figures showed that the chemical yield was almost the same along all 22 runs in the four columns (90.8 ± 6.2) % and (81.4 ± 6.2) % for the U-I and U-II solutions, respectively. For both thorium solutions (Th-I and Th-II), the chemical yield was almost equal (87.4 ± 10.5)%. This means that the ion exchange was unaffected, although the resin may suffer from contact with the reused concentrated mineral acids.
Results of the counting rate of the blank samples are given in Tables 1 and 2. It can be concluded that the slight increase in count rate appeared after run number 15, indicating slight contamination.
The background level was measured for each detector after every five sample runs (Tables 3 and 4). This was to test whether the slight change in counting rate of the blank sample was due to column contamination or detector recoil contamination. The results showed a negligible recoil contamination. Accordingly, these results indicated that the columns can be regenerated at least 12 times for safe reuse.
The resolution of the resultant α-spectrometry is similar to that obtained by electrodeposition onto cleaned steel plates; however, the technique discussed here is much quicker, more reliable, and generally gives a higher chemical yield [15].
Due to the possibility of storing the prepared sample sources in covered Petri dishes for days or weeks before counting, or of needing to recount older samples after some period, the physical properties of the fluoride particles (such as porosity, surface area, or particle size) may change with time due to the dry atmosphere of the laboratory, resulting in spectrum degradation. However, the stability of the prepared uranium or thorium sources were tested by repeated counting of two sources on the same detectors occasionally over a 100-day period to examine the change in resolution of the α-spectrometry peaks with time, which are shown Fig. 5. The results showed that the resolution of the Spectra peaks was almost constant over the studied period.
Conclusion
An AG® 1-X8 anion exchanger was used to separate uranium or thorium from all other components of a sample by ion exchange chromatography, a commonly used radiochemical technique.
The concept of renewing the exchanger for column reuse was tested, with the goal of reducing the costs in mind. The obtained data showed that the column could be regenerated safely at least 12 times for reuse before observing any change in the performance of the exchanger.
The stability of the uranium or thorium fluoride sources that were prepared for counting by α-spectrometry were also tested by constructing peak resolution–aging time curves. The data obtained showed that even after 100 days of storage, the resolution of the source peaks is virtually unaffected with time.
Availability of data and materials
All the experiments were carried out at radiochemical analysis laboratory at the center of Radiation Protection & Training, King Abdul-Aziz University. The data were measured by using α-spectrometry and all data saved in the system, which are available at Nuclear Engineering Department. All the data are available from the corresponding author (Dr. O. Fallatah) of this manuscript.
Abbreviations
- Th:
-
Thorium
- U:
-
Uranium
- HNO3 :
-
Nitric acid
- HCl:
-
Hydrochloric acid
- pCi:
-
Picocurie
- TiCl3:
-
Titanium (III) chloride
- HF:
-
Hydrofluoric acid
References
Abbasisiar F, Hosseini T, Fathivand A, Heravi Gh (2004) Determination of uranium isotopes (234U, 238U) and natural uranium (U-nat) in water samples by alpha Spectrometry Iran. J Radiat Res 1:1–6
Acena M, Crespo M, Galan M, Gascon J (1994) Determination of isotopes of uranium and thorium in low-level environmental samples. Nucl Instrum Method Phys Res A 339:302–308
Galindo C, Mougin L, Nourreddine A (2007) an improved radiochemical separation of uranium and thorium in environmental samples involving peroxide fusion. Appl Radiat Isot 65:9–16
Saidou M, Francois B, Jean P, Kwato M, Pascal F (2008) A comparison of alpha and gamma Spectrometry for environmental natural radioactivity surveys. J Appl Radiat Isot 66:215–222
Shabana EI, Al-Mogabes KS, Al-Najem KN, Farouk MA (1999) Radioactivity in some gas-flow lantern mantles produced by different manufacturers. J Appl Radiat Isot 51:609
Moghissi AA, Kelley HL, Regnier JE, Carter MW (1969) Low-level counting by liquid scintillation—I Tritium measurement in homogeneous systems. Int J Appl Radiat Isot 145:156
Oh YH, Yoon YY, Koh DC, Ko KS (2019) Measuring cosmogenic 35 S in natural waters using large-volume liquid scintillation counting. J Radioanal Nucl Chem 1739:1745
Sill CW (1987) Precipitation of actinides as fluorides or hydroxides for high-resolution alpha Spectrometry. Nucl Chem Waste Manage 201:215
Kemnitz E, Mahn S, Krahl T (2020) Nano metal fluorides: small particles with great properties. ChemTexts 1:27
Shabana EI, Al-Hobaib AS (1990) Activity concentration of natural radium, thorium and uranium isotopes in ground water of two different regions. Radiochimica Act 87(12):41
Shabana EI, Al-Shammari HL (2001) Assessment of the global fallout of plutonium isotopes and americium-241 in the soil of the central region of Saudi Arabia. J Environ Radioactivity 57:67
Beunon H, Chernonozhkin SM, Mattielli, N, Goderis S, Doucet LS, Debaille V, & Vanhaecke F (2020) Innovative two-step isolation of Ni prior to stable isotope ratio measurements by MC-ICP-MS: application to igneous geological reference materials. J Anal Atomic Spect
Cho BW, Choo CO (2019) Geochemical behavior of uranium and radon in groundwater of jurassic granite area, Icheon, middle Korea. Water 11(6):1278
Ibrahiem N, Pimple M (1994) Uranium concentrations in sediments of the Suez Canal. Appl Radiat Isot 919:921
Shabana EI, Kinsara AA (2014) Radioactivity in the groundwater of a high background radiation area. J Environ Radioact 181:189
Avila-Rodriguez MA, Aguilar-Ortiz EA, Jalilian AR, Manrique-Arias JC, Zarate-Morales A, Flores-Moreno A (2019) A simple and efficient method to recover isotopically enriched Ni-64 from electrolytic solutions. Appl Radiat Isot 55:58
Ferreira FT, Mesquita LS, Mesquita RB, Rangel AO (2020) Improved sequential injection method for phosphate quantification within a wide dynamic range with in-line pre-concentration to monitor soil leachates. Talanta Open 2:100015
García-León M, Madurga G: (Eds) (1991) Low-level Measurements of Man-made Radionuclides in the Environment-Proceedings of the 2nd International Summer School. World Scientific.
Gascoyne M (1981) A simple method of uranium extraction from carbonate groundwater and its application to disequilibrium studies. J Geochem Explor,199:207
Sill CW, Williams RL (1981) Precipitation of actinides for α-spectrometry without electrodeposition. Anal Chemica Acta 53:412
Tran QT, Pierre S, de Sanoit J, Pomorski M, Bergonzo P (2019) Electro-Precipitation of actinides on boron-doped diamond thin films for solid sources preparation for high-resolution alpha-particle spectrometry. Appl Sci 9(7):1473
Totland M, Jarvis K (1997) Assessment of Dowex 1-X8-based anion-exchange procedures for the separation and determination of ruthenium, rhodium, palladium, iridium, platinum and gold in geological samples by inductively coupled plasma mass Spectrometry. Analyst 19:26
Pimpl M, Yoo B, Yordanova I (1992) Optimization of a radio analytical procedure for the determination of uranium isotopes in environmental samples. J Radioanal Nucl Chem 437:441
Kumar D, Singh A, Kumar P, Jha RK, Sahoo SK, Jha V (2020) Sobol sensitivity analysis for risk assessment of uranium in groundwater. Environ Geochem Health 1:13
Park J, Kim JY, Lee K, Kim MS, Kim MJ, Choi JW (2020) Comparison of acid extraction and total digestion methods for measuring Cd isotope ratios of environmental samples. Environ Monit Assess 1:10
Stucker V, Ranville JN, Peacock M, Cho A, Hatfield K J (2011) Evaluation and application of anion exchange resins to measure groundwater uranium flux at a former uranium mill site. Water Res 4866:4876
Vesterlund A, Tovedal A, Nygren U, Ramebäck H (2009) Uncertainty assessment of methods for chemical yield determination in measurement of radioactive strontium. J Radioanal Nucl Chem 951:95
Rohwer H, Rheeder N, Hosten E (1997) Interactions of uranium and thorium with arsenazo III in an aqueous medium. Anal Chim Acta 263:268
Truscott JB, Bromley L, Jones P, Evans EH, Turner J, Fairman B (1999) Determination of natural uranium and thorium in environmental samples by ETV-ICP-MS after matrix removal by on-line solid phase extraction. J Anal At Spectrom 627:631
Acknowledgements
This work was funded by the Deanship of Scientific Research (DSR), King Abdulaziz University, Jeddah, under grant No. ( D-072-135-1442). The authors, therefore, acknowledge with thanks DSR technical and financial support. The authors also thank the Editor and the anonymous reviewers of journal of Radioanalytical and Nuclear Chemistry.
Funding
This work was financially supported by the Deanship of Scientific Research ( DSR), KAU, Jeddah, under grant No. (D-072-135-1442).
Author information
Authors and Affiliations
Contributions
O. Fallatah will oversee the overall objectives of the article at radiochemical analysis laboratory in Radiation Protection & Training Centre, King Abdul-Aziz University. He’s responsible for the supervision of organizing and overall management of the components of the research activities, ordering required materials, calibration of the Alpha spectrometer, and also be responsible for publication and presentation of results. M. M. T. Qutub carried out analysis, chemical separation, and measurement of results by using α-spectrometry.
Corresponding author
Ethics declarations
Conflict of interest
There is no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Fallatah, O., Qutub, M.M.T. Radiochemical technique optimization to measure uranium and thorium by α-spectrometry. J Radioanal Nucl Chem 328, 1077–1083 (2021). https://doi.org/10.1007/s10967-021-07706-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-021-07706-y