Abstract
To validate the accuracy of coronary artery calcium score (CACS) using photon-counting detector (PCD) CT under various scanning settings and explore the optimized scanning settings considering both the accuracy and the radiation dose. A CACS phantom containing six hollow cylindrical hydroxyapatite calcifications of two sizes with three densities and 12 patients underwent CACS scans. For PCD-CT, two scanning modes (sequence and flash [high-pitch spiral mode]) and five tube voltages (90kV, 120kV, 140kV, Sn100kV, and Sn140kV) at different image quality (IQ) levels were set for phantom, and patients were scanned with 120kV at IQ19 using flash mode. All acquisitions from PCD-CT were reconstructed at 70keV. Acquisitions in sequence mode at 120kV on an energy-integrating detector CT (EID-CT) was used as the reference. Agatston, mass, and volume scores were calculated. The CACS from PCD-CT exhibited excellent agreements with the reference (all intraclass correlation coefficient [ICC] > 0.99). The root mean square error (RMSE) between the Agatston score acquired from PCD-CT and the reference (5.4–11.5) was small. A radiation dose reduction (16–75%) from PCD-CT compared with the reference was obtained in all protocols using flash mode, albeit with IQ20 only at sequence mode (22–44%). For the patients, ICC ( all ICC > 0.98) and Bland–Altman analysis of CACS all showed high agreements between PCD-CT and the reference, without reclassifying CACS categories(P = 0.317). PCD-CT yields repeatable and accurate CACS across diverse scanning protocols according to our pilot study. Sn100kV, 90kV, and 120kV using flash mode at IQ20 are recommended for clinical applications considering both accuracy and radiation dose.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Coronary artery calcium (CAC) presents as a distinctive coronary artery disease manifestation, which is a potent predictor of cardiovascular morbidity and mortality [1,2,3]. The CAC quantification, accomplished through coronary artery calcium scores (CACS), including the Agatston, mass, and volume scores, holds significant clinical value and the Agatston score is particularly vital for cardiovascular risk stratification and patient management [4,5,6].
Conventionally, standard CACS is derived from fixed scanning parameters, such as 2.5- or 3-mm slice thickness with 120 kV using sequence mode [7], which have remained largely unaltered amidst advances in CT technology, making it difficult to further reduce radiation exposure. While qualitative and quantitative assessments of CAC in non-gated CT have been explored to mitigate radiation dose, these approaches have not yet fully supplanted ECG-gated non-contrast cardiac CT in terms of accuracy [8]. The prospect of dose reduction through tube voltage reduction is promising, though the requisite standardization and implementation of corresponding threshold adjustments to get comparable CACS as standard CACS acquisition present challenges [9,10,11].
Recent studies in CT technology have ushered the era of photon-counting detector CT (PCD-CT) [12, 13]. In contrast to the conventional energy-integrating detector CT (EID-CT) involving the conversion of X-ray photons into visible light and then into an electrical signal − a process that can interfere with adjacent detector elements, the direct conversion of X-ray photons into an electrical signal in PCD-CT can avoid such interference [14]. Furthermore, unlike conventional CT using energy-integrating detectors that only acquire the total amount of energy across multiple photons concurrently, the photon-counting detector of PCD-CT can discriminate individual photon according to its energy from the X-ray beam, then making it possible to acquire high spatial resolution, less electronic noise, fewer beam-hardening artifacts and more stable Hounsfield unit (HU) numbers [15]. The potential to enhance quantitative CT applications lies in the consistency and standardization of CT values secondary to dual-source PCD-CT [16]. The accuracy and stability of CACS obtained through various scanning parameters, including the potential of reducing radiation dose and maintaining CACS stability on PCD-CT, remains unknown.
To test whether PCD-CT can result in accurate and stable CACS with different scanning protocols, we aimed to systematically evaluate CACS of PCD-CT using multiple tube voltages and scanning modes at different radiation dose levels in phantom and patients by using acquisitions on EID-CT as the reference.
Material and methods
Phantom study
In this study, a commercial static anthropomorphic chest phantom (QRM, Moehrendorf, Germany) was used (anterior posterior diameter, 20cm; lateral diameter, 30cm). The simulated coronary arteries (water-equivalent and CaHA material) contained six cylindrical inserts with different diameters (3 and 5 mm) and various densities (hydroxyapatite at 200, 400, and 800 mg/cm3, indicating low, medium, and high density, respectively).
Phantom scan acquisition and reconstruction
For the reference acquisition, scanning was first performed on a conventional EID-CT (SOMATOM Force; Siemens Healthineers, Forchheim, Germany) using sequence mode with 120 kV to obtain the the gold standard CACS. Other detailed parameters are listed in Table 1. For PCD-CT (NAEOTOM Alpha; Siemens Healthineers GmbH) acquisition, 20 protocols using two scanning modes (sequence and “flash” [high-pitch spiral: pitch factor 3.2]) with five tube voltages (90 kV, 120 kV, 140 kV, 100 kV with tin filter [Sn100], and 140 kV with tin filter [Sn140]) and automatic tube current modulation (CARE Dose4D) at two image quality (IQ) levels optimized for non-contrast (low-IQ20; high-IQ42 or IQ50) were adapted. IQ 42 was used only for Sn100 because the highest IQ level of Sn100 is 42 and IQ 50 for the other tube voltage. Each of the scanning protocols was repeated three times. IQ level indicates the effective mAs that are applied for the reference water-equivalent diameter specific to the protocol along with a CT geometry correction, specifically for the focal spot to iso-center distance effect. As a result, the image quality level offers a definition of image quality that is independent of systems and reconstruction.
All acquisitions were then reconstructed via virtual monoenergetic imaging at 70 keV according to a previous study [17] with 3-mm slice thickness and 1.5-mm increments. The quantum iterative reconstruction (QIR) level was set at QIR2 by default setting.
All scanning parameters are listed in Table 1. For simplicity, we will denote the scan mode as “Scan mode, tube voltage_IQ level” next.
Patient study
Our study was approved by Ruijin Hospital Ethics Committee Shanghai Jiaotong University School of Medicine (Number 2023-186). All volunteers who were suspected of coronary artery disease (CAD) and indicated by clinicians for CCTA were recruited in September 2023, and all patients provided written informed consent. The inclusion criteria were as follows: (1) age > 18 years; and (2) previous chest CT or coronary CT angiography (CCTA) examination indicating coronary artery calcification.The exclusion criteria were as follows: (1) pregnant or lactating women; (2) coronary stents or coronary artery bypass graft surgery; (3) severe arrhythmia; and (4) chest CT or CCTA examination performed within a year (to avoid excessive radiation exposure). We recruite 15 patients but 12 patients were finally included after excluding two with severe atrial fibrillation and one with coronary stents.
Patient scan acquisition and reconstruction
All patients did not use any medical agents to control heart rate before CT examination. First, the EID-CT mentioned above in phantom scan was also used to scan the patients with sequence mode to obtain the referred ECG-gated CACS, followed by a PCD-CT ECG-gated CACS scanning on the same day. For PCD-CT scanning, the tube voltage was 120kV and current were determined using the automated tube current modulation (CARE Dose4D) with the IQ level set to 19. Then, the PCD-CT images were reconstructed at 70 keV according to the previous study[17]. The detailed scanning parameters are shown in Table 1. Volumetric Computed Tomography Dose Index (CTDIvol) and Dose-Length Product (DLP) were recorded from dose report automatically produced by CT system.
CACS measurement
Using a dedicated semi-automatic software (CT CaScoring, Syngo.Via, Version VB70; Siemens Healthineers), The CACS acquired from all scanning protocols was calculated on two separate occasions, with a 2-week interval between each calculation, by a radiologist with 3-year experience in cardiovascular imaging. The software identified all pixels with a density > 130 HU as calcification and the observer can correct the wrong segmentations manually. The factor used to calculate mass score was 0.743. For phantom, the Agatston, mass, and volume scores were calculated for each calcification (Fig. 1a). For the patients, the Agatston, mass, and volume scores were calculated at single lesion and patient level (Fig. 1b and 1c). CACS categories based on the Agatston score are as follows: 0, 1–10, 11–100, 101–400, and > 400 [7].
Figure 1a shows our phantom containing six hollow cylindrical hydroxyapatite calcifications of various sizes (large and small) and densities (200, 400, and 800 mg/cm3). Figure 1b and 1c demonstrates the CACS of a 65-year-old male patient scanned using energy-integrating detector computed tomography (EID-CT) and photon-counting detector CT (PCD-CT), respectively.
Statistical analysis
In this study, two commercial statistical programs (SPSS, version 26 and R, version 4.1.3) were used. To test continuous data for normality, the Kolmogorov–Smirnov test was used. Data that conform to a normal distribution are expressed as mean ± standard deviation, whereas those with a non-normal distribution are reported as median (interquartile range). CACS and radiation dose differences between PCD-CT and EID-CT were compared using paired-samples t tests or Wilcoxon rank sum tests, depending on distribution. Agreements between PCD-CT- and EID-CT-derived CACS were assessed through intraclass correlation coefficient (ICC), Bland–Altman with 95% confidence interval limits of agreement (LoA), and root mean square error (RMSE) analysis. ICC was interpreted as follows: 0.0 to 0.3, lack of agreement; 0.31 to 0.5, weak agreement; 0.51 to 0.7, moderate agreement; 0.71 to 0.9, strong agreement; and 0.91 to 1.00, very strong agreement. Statistical significance was set at a P < 0.05.
Results
Radiation dose
In vitro
The CTDIvol of the phantom using the standard CACS scan protocol on EID-CT served as reference was 2.4 mGy and DLP was 31.1 mGy·cm, recorded form dose report provided by CT system, and SSDE was 4.2 mGy, calculated by the formula[18] CTDIvol, DLP and SSDE of all PCD-CT protocols were recorded in Supplementary materials-Table S1.
A radiation dose reduction (16–75%) from PCD-CT compared with the reference was obtained in all protocols using flash mode, with Flash spiral Sn100kV IQ20 got the lowest radiation dose (75% dose reduction). For sequence mode, only low IQ level (IQ20) got a radiation dose reduction (22–44%) while the high IQ level got a radiation dose increase(5–94%) compared with the reference, with Sn140 kV IQ50 got the highest radiation dose (94% increase) as shown in Fig. 2.
In vivo
A total of 12 patients (age, 69.83 ± 9.27 years, body mass index, 21.34 ± 1.72 kg/m2) with 28 calcified coronary arteries were finally recruited. The CTDIvol of the reference and PCD-CT were 2.21(1.94, 2.68) and 1.03(0.83, 1.51) mGy, respectively.
CACS analysis
In vitro
All intra-observer ICC of CACS were > 0.99. As shown in Table 2, there is no difference between Agatston and volume scores from all PCD-CT scans and the references (all P > 0.05), except for the volume scores acquired from Flash Spiral 120 kV/90 kV_IQ50, as well as sequence 140 kV/Sn 140kV_IQ20 and sequence 90 kV_IQ50. However, mass scores were significantly higher (P < 0.05) for most PCD-CT settings than the references except Flash Spiral 90 kV_IQ20, sequence 90 kV_IQ20/50, and sequence 140 kV_IQ20 (all P > 0.05). ICC analysis showed strong agreements between Agatston, mass, and volume scores obtained with PCD-CT using various settings and references, with all ICC > 0.99 (Table 2).
Bland–Altman analysis (Table 2) revealed excellent agreements of CACS between PCD-CT and the reference. For Agatston score, the bias ranged from − 1.7 to 4.95, with only a single dot outside the LoA (Flash Spiral 90kV_IQ50) and Sequence Sn100kV_IQ42 obtained the narrowest LoA (12.72). For the mass scores, the bias spanned from − 2.02 to − 0.27, with only three dots outside the LoA (Flash Spiral Sn140kV_IQ20) and Sequence 90kV_IQ50 obtained the narrowest LoA (1.33). For the volume score the bias extended from − 1.35 to 4.20, with only one dot outside the LoA (Flash Spiral 120kV_IQ50 and Sequence 140kV_IQ20) and Sequence 90kV_IQ50 obtained the narrowest LoA (3.73).
For all PCD-CT scannnig modes, Fig.S1 illustrates that the RMSE of the Agatston score between PCD-CT and the reference ranged from 5 to 11.5. For the 200, 400, and 800 mg/cm3 calcification, RMSE (Fig.S2-4) ranged from 1.4 to 10.9, 6.8–18.3, and 1.8–10.9, respectively. Bland–Altman and ICC analysis also showed high Agatston score agreements between all PCD-CT scanning settings and the reference, with all dots in the LoA (bias range, 0.6/10.1, − 14.8/ − 0.4, − 10.8/0.2, respectively) and all ICC > 0.90.
The Agatston score of calcifications with different sizes and densities scanned with PCD-CT using various scanning modes presented good agreements with the references (Fig. 3), except for the calcification with both low density and small size (200 mg/cm3, 3mm).
In-vivo
All intra-observer ICC of CACS at the single calcified plaque level was > 0.99. Agatston, volume, and mass scores of the patients were 494.45(142,45, 1187.05), 417.45(136.25, 1046), and 114.46(23.87, 250.13) for the EID-CT and 514.9(175.33, 1302.6), 454.95(156.98, 1111.63), and 123.58(29.91, 286.03) for the PCD-CT, respectively. The radiation dose of PCD-CT was significantly lower than that of EID-CT (CTDIvol, 1.03[0.83, 1.51] vs. 2.21[1.94, 2.68]mGy, p < 0.001). Table 3 shows the CACS of EID-CT and PCD-CT. At both the single lesion and patient levels, Bland–Altman analysis (Fig. 4) and ICC (from 0.991 to 0.998) showed good agreements between the PCD-CT and EID-CT. No significant difference was found for the CACS categories (P = 0.317), except for one patient who upgraded from 100 to 400 to > 400.
Discussion
In our in vivo and in vitro study, we comprehensively evaluated the accuracy of CACS with various scan settings acquired from PCD-CT. In the in vitro settings, we ascertained that the CACS measurements acquired from PCD-CT, spanning varying tube voltages, scanning modes, and IQ levels reconstructed at 70 keV, remain accurate and reproducible and the results were also validated in patients.
CACS has been widely used routinely in clinics to quantify coronary calcification. Precise and reproducible CACS measurement is imperative for guiding clinical patient management [4,5,6]. However, CACS is susceptible to scan parameters, particularly varying tube voltages [19,20,21], which may compromise CACS reproducibility and affect the monitoring of coronary artery sclerosis progress. The standard scanning tube voltage for CACS is 120 kV [7], which is insufficient to satisfy the requirement of radiation exposure reduction. No widely acknowledged thresholds complementary to the low tube voltages exist currently, although many studies have put forward different adjusted thresholds for varying tube voltages and reconstruction algorithms [22,23,24].
Since the combination of dual-source and PCD techniques enables spectral imaging with high temporal resolution and spatial resolution [25], which may provide high CT value accuracy at low-dose levels, and the consistency and standardization of CT values provided by PCD-CT improved the screening and monitoring of calcium plaques. This property has been validated by several phantom studies [26, 27] and holds great potential in advancing quantitative application of CT such as CACS. Confirming whether the parameter measurements of PCD-CT now extensively employed in clinical practice such as CACS are accurate is therefore requisite.
A previous study found that CACS reconstructed with 70 keV obtained from various voltages (70kV, 90kV, Sn100kV, Sn140kV, and 140 kV) at the same IQ levels yielded comparable results to a 120 kV reference [28]. However, there is no systematic analysis was not conducted for CACS obtained from PCD-CT with different tube current and scanning modes. Furthermore, the agreement and variability of CACS between conventional EID-CT and PCD-CT were absent, and the established risk catergories of CACS was based on EID-CT, making it necessary to validate the comparability of CACS between PCD-CT and EID-CT. Similarly, several previous phantom studies explored the accuracy of PCD-CT CACS at different tube voltages; however, none of these studies validated their findings in patients [28,29,30]. Although Eberhard M’s study encompassed both the phantom and patients, but they only analyzed the accuracy of PCD-CT CACS acquired using 120 kV in the phantom [31]. Our study presents a comprehensive exploration of the accuracy of CACS using different scan modes and tube voltages at different radiation dose levels in the phantom, as well as a validation of these findings in a small cohort of patients by using 120 kV sequence scanning on EID-CT as reference.
In our study, we assessed not only Agatston scores but also mass and volume scores by using simulating calcification with different densities in detail. Our findings were inconsistent with the study by van der Werf NR [30], which demonstrated clinically relevant differences from the reference in low-density calcification when using reduced tube current protocols, whereas our study demonstrates that PCD-CT achieves a minor Agatston score RMSE relative to the reference for low-density calcification. However, our study found that the mass scores derived from PCD-CT demonstrated a relatively greater deviation from references than the Agatston and volume scores. The mass score is calculated as the product of the calibration factor, volume, and mean attenuation values for each calcification [8], whereas the traditional CACS is calibrated and converted using 120 kV tube voltage. In the context of our study, the decision to employ 70 keV images was based on that 70 keV images are the most similar to 120 kV images among the mono-energetic images, which was also adapted by a prior Agatston score phantom study [17]. The calibration factor and this specific choice of energy level may need further be elaborated for accurate mass score calculation on PCD-CT.
This study has the following limitations: (1) Our phantom classifications were stationary and homogeneous; however, we subsequently performed patient validation, which to some extent compensated for this limitation. (2) The number of included patients was small, which might reduce the statistical power to assess whether PCD-CT led to CACS reclassification. Moreover, the sample size needs to be expanded in the future. (3) The effect of different QIR and keV levels of PCD-CT on CACS was not explored because our study purpose was to explore the effect of scanning parameters and not reconstruction parameters on CACS, the latter of which has been investigated in previous studies.
Conclusions
In conclusion, PCD-CT employing different tube voltages, scanning modes, and IQ levels can generate an accurate and repeatable CACS, improving atherosclerotic disease screening and monitoring. Sn100kV, 90kV, and 120kV using flash mode at IQ20 are recommended for clinical applications considering both accuracy and radiation dose.
Data availability
No datasets were generated or analysed during the current study.
References
Shemesh J, Henschke CI, Shaham D et al (2010) Ordinal scoring of coronary artery calcifications on low-dose CT scans of the chest is predictive of death from cardiovascular disease. Radiology 257(2):541–548. https://doi.org/10.1148/radiol.10100383
Lei MH, Wu YL, Chung SL et al (2021) Coronary artery calcium score predicts long-term cardiovascular outcomes in asymptomatic patients with type 2 diabetes. J Atheroscler Thromb 28(10):1052–1062. https://doi.org/10.5551/jat.59386
Church TS, Levine BD, McGuire DK et al (2007) Coronary artery calcium score, risk factors, and incident coronary heart disease events. Atherosclerosis 190(1):224–231. https://doi.org/10.1016/j.atherosclerosis.2006.02.005
Patel J, Pallazola VA, Dudum R et al (2021) Assessment of coronary artery calcium scoring to guide statin therapy allocation according to risk-enhancing factors: the multi-ethnic study of atherosclerosis. JAMA Cardiol 6(10):1161–1170. https://doi.org/10.1001/jamacardio.2021.2321
Ferket BS, Hunink MGM, Masharani U et al (2022) Lifetime cardiovascular disease risk by coronary artery calcium score in individuals with and without diabetes: an analysis from the multi-ethnic study of atherosclerosis. Diabetes Care 45(4):975–982. https://doi.org/10.2337/dc21-1607
Parcha V, Malla G, Kalra R et al (2021) Coronary artery calcium score for personalization of antihypertensive therapy: a pooled cohort analysis. Hypertension 77(4):1106–1118. https://doi.org/10.1161/HYPERTENSIONAHA.120.16689
Leipsic J, Abbara S, Achenbach S et al (2014) SCCT guidelines for the interpretation and reporting of coronary CT angiography: a report of the society of cardiovascular computed tomography guidelines committee. J Cardiovasc Comput Tomogr 8(5):342–358. https://doi.org/10.1016/j.jcct.2014.07.003
Alluri K, Joshi PH, Henry TS et al (2015) Scoring of coronary artery calcium scans: history, assumptions, current limitations, and future directions. Atherosclerosis 239(1):109–117. https://doi.org/10.1016/j.atherosclerosis.2014.12.040
Grani C, Vontobel J, Benz DC et al (2018) Ultra-low-dose coronary artery calcium scoring using novel scoring thresholds for low tube voltage protocols-a pilot study. Eur Heart J Cardiovasc Imaging 19(12):1362–1371. https://doi.org/10.1093/ehjci/jey019
Nakazato R, Dey D, Gutstein A et al (2009) Coronary artery calcium scoring using a reduced tube voltage and radiation dose protocol with dual-source computed tomography. J Cardiovasc Comput Tomogr 3(6):394–400. https://doi.org/10.1016/j.jcct.2009.10.002
Kim S, Lee CJ, Han K et al (2021) Histogram-derived modified thresholds for coronary artery calcium scoring with lower tube voltage. Sci Rep 11(1):17450. https://doi.org/10.1038/s41598-021-96695-9
Rajendran K, Petersilka M, Henning A et al (2022) first clinical photon-counting detector CT system: technical evaluation. Radiology 303(1):130–138. https://doi.org/10.1148/radiol.212579
Si-Mohamed SA, Boccalini S, Lacombe H et al (2022) Coronary CT angiography with photon-counting CT: first-in-human results. Radiology 303(2):303–313. https://doi.org/10.1148/radiol.211780
Flohr T, Petersilka M, Henning A et al (2020) Photon-counting CT review. Phys Med 79:126–136. https://doi.org/10.1016/j.ejmp.2020.10.030
Willemink MJ, Persson M, Pourmorteza A et al (2018) Photon-counting CT: technical principles and clinical prospects. Radiology 289(2):293–312. https://doi.org/10.1148/radiol.2018172656
McCollough CH, Rajendran K, Leng S (2023) Standardization and quantitative imaging with photon-counting detector CT. Invest Radiol 58(7):451–458. https://doi.org/10.1097/RLI.0000000000000948
van der Werf NR, Greuter MJW, Booij R et al (2022) Coronary calcium scores on dual-source photon-counting computed tomography: an adapted Agatston methodology aimed at radiation dose reduction. Eur Radiol 32(8):5201–5209. https://doi.org/10.1007/s00330-022-08642-5
Boone JM, Strauss KJ, Cody DD et al (2011) Report No. 204 - Size-specific dose estimate in pediatric and adult body CT examinations. American Association of Physicists in Medicine. https://doi.org/10.37206/143
Zhang D, Scott A, Lee C et al (2021) Coronary artery calcium scoring at lower tube voltages—dose determination and scoring mechanism. Eur J Radiol 139:109680. https://doi.org/10.1016/j.ejrad.2021.109680
Allio IR, Caobelli F, Popescu CE et al (2022) Low-dose coronary artery calcium scoring compared to the standard protocol. J Nucl Cardiol. https://doi.org/10.1007/s12350-022-03120-3
Fujioka C, Funama Y, Kiguchi M et al (2012) Coronary artery calcium scoring on different 64-detector scanners using a low-tube voltage (80 kVp). Acad Radiol 19(11):1402–1407. https://doi.org/10.1016/j.acra.2012.07.003
Marwan M, Mettin C, Pflederer T et al (2013) Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols. J Cardiovasc Comput Tomogr 7(1):32–38. https://doi.org/10.1016/j.jcct.2012.11.004
Thomas CK, Muhlenbruch G, Wildberger JE et al (2006) Coronary artery calcium scoring with multislice computed tomography: in vitro assessment of a low tube voltage protocol. Invest Radiol 41(9):668–673. https://doi.org/10.1097/01.rli.0000233324.09603.dd
Kamani CH, Huang W, Lutz J et al (2022) Impact of adaptive statistical iterative reconstruction-v on coronary artery calcium scores obtained from low-tube-voltage computed tomography—a patient study. Acad Radiol 29(Suppl 4):S11–S16. https://doi.org/10.1016/j.acra.2020.10.025
Euler A, Higashigaito K, Mergen V et al (2022) High-pitch photon-counting detector computed tomography angiography of the aorta: intraindividual comparison to energy-integrating detector computed tomography at equal radiation dose. Invest Radiol 57(2):115–121. https://doi.org/10.1097/RLI.0000000000000816
Liu LP, Shapira N, Chen AA et al (2022) First-generation clinical dual-source photon-counting CT: ultra-low-dose quantitative spectral imaging. Eur Radiol 32(12):8579–8587. https://doi.org/10.1007/s00330-022-08933-x
Symons R, Reich DS, Bagheri M et al (2018) Photon-counting computed tomography for vascular imaging of the head and neck: first in vivo human results. Invest Radiol 53(3):135–142. https://doi.org/10.1097/RLI.0000000000000418
Dobrolinska MM, van der Werf NR, van der Bie J et al (2023) Radiation dose optimization for photon-counting CT coronary artery calcium scoring for different patient sizes: a dynamic phantom study. Eur Radiol. https://doi.org/10.1007/s00330-023-09434-1
Mergen V, Higashigaito K, Allmendinger T et al (2021) Tube voltage-independent coronary calcium scoring on a first-generation dual-source photon-counting CT-a proof-of-principle phantom study. Int J Cardiovasc Imaging. https://doi.org/10.1007/s10554-021-02466-y
van der Werf NR, van Gent M, Booij R et al (2021) Dose reduction in coronary artery calcium scoring using mono-energetic images from reduced tube voltage dual-source photon-counting CT data: a dynamic phantom study. Diagnostics (Basel). https://doi.org/10.3390/diagnostics11122192
Eberhard M, Mergen V, Higashigaito K et al (2021) Coronary calcium scoring with first generation dual-source photon-counting CT-first evidence from phantom and in-vivo scans. Diagnostics (Basel). https://doi.org/10.3390/diagnostics11091708
Funding
National Key Research and Development Program of China, 2022YFC2401604.
Author information
Authors and Affiliations
Contributions
Guarantors of integrity of entire study, W. Yang, F. Yan; study design, W. Yang, F. Yan, M. Wang; data acquisition, X. Zhang, H. Dong, J. Li; data interpretation, all authors; manuscript drafting, M. Wang, X. Zhang; manuscript revision, W. Yang and M. Wang; Technical support, Z. Xu approval of final version of submitted manuscript, all authors; agrees to ensure any questions related to the work are appropriately resolved, all authors; statistical analysis, M. Wang, Z. Xu; and manuscript editing, all authors.
Corresponding author
Ethics declarations
Competing interests
Zhihan Xu is an employee of Siemens Healthineers. The remaining authors declare that they have no conflict of interest with each other.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Wang, M., Zhang, X., Li, J. et al. Quantification accuracy in photon-counting detector CT for coronary artery calcium score: a pilot study. Int J Cardiovasc Imaging (2024). https://doi.org/10.1007/s10554-024-03209-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10554-024-03209-5