Introduction

Endothelial shear stress (ESS) is the hemodynamic force that flowing blood exerts on the vessel wall [1]. This force is continuously sensed by mechanoreceptors that are attached to the cytoskeleton of endothelial cells [2]. Processes involved in endothelial cell turnover, inflammation, coagulation, extracellular matrix degradation, and production of vasoactive substances such as nitric oxide are all tightly regulated by ESS [14]. In vitro and animal studies showed that protracted decrease in ESS and high variation of ESS during the cardiac cycle cause endothelial receptor and gene expression to switch to an atherogenic phenotype [14]. Vascular imaging studies corroborate these findings, showing a causal relationship between ESS and atherosclerosis development [58].

Assessing ESS in vivo is technically challenging, which is the reason why human studies on ESS remain scarce, and are typical performed in small sample sizes. Various methods are used, including computational fluid dynamics, velocity gradient modelling, and methods assuming Poiseuille flow [9]. Computational fluid dynamics can solve the Navier–Stokes equations of fluid flow [9], while velocity gradient modelling is based on the spatial gradient of the velocities close to the artery wall using second-order curve fitting of the velocity profile [5, 9, 10]. Although accurate methods, their downsides pertain to the fact that high resolution information is needed on the local velocity vector throughout the cycle, respectively as boundary condition for the Navier–Stokes equations and as base for the accurate fitting of the velocity gradient perpendicular to the wall. Obtaining such data is labor intensive and dedicated software is needed for the complex data analysis, which is not freely available. In contrast, Poiseuille based methods are simpler as they merely require measurement of artery diameter and flow rate or centerline velocity followed by a simple calculation [9]. Estimation of ESS dynamics are then based on the instantaneous hemodynamics during the cycle. However, Poiseuille flow requires among others steady flow in straight rigid tubes. These conditions are violated in vivo in human arteries, resulting in deviation from the assumed parabolic velocity profile [9, 11]. Despite the inherent inadequacy of Poiseuille based methods, the majority of current literature relies on this approach due to its simplicity and availability [68, 12, 13].

Therefore, we investigated whether an equally simple method based on similar parameters (diameter and flow rate or centerline velocity) provides a better alternative to the Poiseuille based method. This method is based on Womersley flow [14], which in contrast to Poiseuille covers pulsatile flow components. We compared the Womersley method to the Poiseuille based method as well as to velocity gradient modelling from high resolution phase contrast (PC)-MRI. We correlated ESS of each method with carotid artery wall thickness, assuming that higher correlations are obtained with more accurate ESS estimations [15]. We also compared reproducibility of each of the methods. All measurements were performed in the common carotid arteries by 3T-MRI.

Methods

Subject population

Forty-five subjects (age range 19–79 years) were enrolled prospectively. The population consisted of 30 volunteers and 15 patients with cardiovascular disease who were selected from the outpatient clinic of the Department of Vascular Medicine of the Academic Medical Center, Amsterdam, The Netherlands. In all subjects repeat 3T-MRI scans were acquired on three separate occasions, 1–3 weeks apart. All image analyses were done off-line. Prior to the studies, approval was obtained from the institutional review board of the Academic Medical Center, Amsterdam, The Netherlands. All subjects gave written informed consent.

Image acquisition

Magnetic resonance imaging scans were obtained bilaterally on a 3T-MRI scanner (Intera, Philips Medical Systems, Best, The Netherlands) using a single-element microcoil (Philips, Hamburg, Germany) with a diameter of 5 cm. Axial magnetic resonance angiography images were acquired using a time of flight sequence to localize the left and right common carotid artery and position the scan planes perpendicular to the vessel.

For hemodynamic assessments, axial gradient echo phase-contrast images were acquired with a temporal resolution of 17 ms, and temporal interpolation was performed by the scanner software to 60 phases per heartbeat (retrospective electrocardiography gating). The scan plane was positioned 27 mm proximal to the carotid flow divider. Sequence parameters were: slice thickness 3 mm, non-interpolated pixel size 0.65 × 0.65 mm, field of view 60 × 60 mm, velocity encoding 150 cm s−1 (unidirectional), repetition time 8.1 ms, echo time 5 ms, flip angle 10°, number of signal averages 2, total scan time 3–4 min. Both magnitude images and velocity encoded phase images were reconstructed.

To assess carotid wall thickness, axial T1-weighted Turbo Spin Echo image stacks were acquired at end-diastole using double inversion recovery black blood preparation applying active fat suppression (spectral attenuated inversion recovery technique) as described previously [5]. Sequence parameters were: slice thickness 3 mm, imaging matrix size 240, field of view of 60 × 60 mm, noninterpolated pixel size 0.25 × 0.25 mm, echo time 9 ms, repetition time according to the subjects’ heart rate (approximately 900 ms), echo train length 7, echo train duration 63 ms. All imaging was performed with cardiac gating.

Image analysis

For assessment of hemodynamic parameters in the carotid arteries, off-line semi-automated qualitative and quantitative image analysis was performed using software written in Matlab developed at the Academic Medical Center, Amsterdam, The Netherlands. The lumen area (LA, mm2) of all 60 phases per cardiac cycle was assessed by automated tracing of the lumen-wall boundaries on the gradient echo images [16]. The velocity at each pixel in the artery lumen was measured, which enabled the calculation of the volumetric flow rate (Q, cm3 s−1) and centerline velocity (v, cm s−1) of each phase in the cardiac cycle. The measured lumen area, flow rate, and centerline velocity were used to calculate the Poiseuille and Womersley based endothelial shear rate (ESR, s−1). The calculations are shown in the Supplementary Methods. While Poiseuille and Womersley should predict identical mean ESS, a very minor difference exists in the data due to taking respectively an actual and a time-averaged diameter for the calculations. For the assessment of ESR based on the velocity gradient modelling method, we used dedicated software written in Matlab developed at the Academic Medical Center, Amsterdam, The Netherlands. We have described the method in detail previously [5]. Firstly, the lumen area (LA, mm2) of all 60 phases per cardiac cycle was assessed as described above. Secondly, for each pixel the shortest distance to the artery wall was calculated. The lumen-wall boundaries were projected on the velocity encoded images. The ESR was assessed by determining the spatial gradient of the velocities close to the artery wall using second order curve fitting of the velocity profile. The pixels located from 0.31 to 1.56 mm from the wall were included for the analysis. The fit was forced to include the position of the artery wall with a velocity of zero to conform to the zero-slip condition.

For all methods (Poiseuille, Womersley and velocity gradient modelling), ESR was multiplied with the blood viscosity to calculate the ESS throughout the cardiac cycle. Blood viscosity was assumed to be 3.2 cP [17]. Poiseuille and Womersley methods were either based on flow rate (ESSpq, ESSwq) or centerline velocity (ESSpv, ESSwv).

From these dynamics, we calculated the mean ESS of all phases in the cardiac cycle as well as the shear pulsatility index (SPI) and the oscillatory shear index (OSI). SPI was defined as difference between the maximum and minimum ESS in the cardiac cycle divided by their sum. The oscillatory shear index represents the proportion of shear stress deviating from its predominant direction during the cardiac cycle [18].

For the carotid wall thickness quantification, semiautomated image analysis was performed using VesselMass software (Leiden University Medical Center, Leiden, The Netherlands). VesselMass performed automated tracing of the lumen-wall boundaries and the outer wall boundaries. The software algorithm for boundary detection and analysis methods are described elsewhere [19, 20].

Statistical analysis

Continuous variables are expressed as means ± standard deviations (SD). We used multiple linear regression analysis to assess the association between the ESS values and MWT with MWT as the response variable and ESS as the explanatory variable. We determined by stepwise analysis with backward elimination that age and systolic blood pressure were the only significant confounders among all measured cardiovascular risk factors (patient characteristics and biochemistry as listed in Table 1). We therefore adjusted for these confounders. The agreement between successive ESS estimations was assessed using intraclass correlation coefficients (ICC) and the 95 % confidence intervals of the ICCs. For all statistical analyses Statistical Package for the Social Sciences version 20.0 for Windows was used.

Table 1 Volunteer characteristics, hemodynamics and ESS estimates

Results

Three MRI scans were performed in all 45 subjects. The population consisted of 20 females and 25 males. Their characteristics are shown in Table 1. Figure 1 shows two examples of wall thickness (Fig. 1a, d), measured peak systolic flow velocity (color scales in Fig. 1b, c, e, f), and estimations of peak systolic flow velocity based on Poiseuille (Fig. 1b, e) and Womersley (Fig. 1c, f) with flow rate as input. As can be seen, both vessels were approximately circular and measured flow velocity was relatively axisymmetric in both examples. The Womersley-based prediction far more closely matched the measured profile as compared to the Poiseuille-based estimate. The Womersley but not the Poiseuille model predicted flow reversal at the wall (OSI > 0), while in these cases the thicker wall was associated with a larger OSIwq. Average data for common carotid lumen area, mean wall thickness (MWT, mm), hemodynamic parameters, and ESS parameters are shown in Table 1. Mean ESS is by definition equal when based on Poiseuille or Womersley flow, and these estimates were comparable to the estimate from velocity gradient modelling. SPI for the Womersley based methods was around twice as high as that for the other methods. The amount of negative shear stress as assessed with OSI was low for all methods with Womersley based methods exhibiting the highest values (Table 1). In particular, the OSIpv was constantly zero such that its reproducibility and correlation with MWT could not be assessed. Figure 2 shows the different ESS estimates throughout the cardiac cycle averaged for all subjects.

Fig. 1
figure 1

Two examples of measured mean wall thickness (MWT, a, d) and peak systolic flow velocity profiles with their Poiseuille- and Womersley-based estimates. The phase contrast MRI-based velocity profiles are shown as color-coded image (b, c, e, f). Estimated profiles are indicated as contour plots. b, e Poiseuille-based with flow rate as input, c, f Womersley-based with flow rate as input. OSI: estimated oscillatory shear index

Fig. 2
figure 2

Common Carotid Endothelial Shear Stress (ESS, N m−2) during the cardiac cycle averaged for all subjects for all estimation methods (red velocity gradient modelling method, blue Poiseuille flow modelling based on centerline velocity, interrupted blue Poiseuille flow modelling based on flow rate, grey Womersley flow modelling based on centerline velocity, interrupted black Womersley flow modelling based on flow rate). Per cardiac cycle, 60 image frames were made where the first frame coincided with the R wave of electrocardiogram. ESS was quantified for all 60 frames per heartbeat

Partial regression coefficients between ESS, SPI and OSI values and MWT, adjusted for the potential confounders age and systolic blood pressure, are shown in Table 2. The mean ESS and SPI of the velocity gradient modelling and Womersley flow rate based modelling methods correlated equally well with MWT, while only the mean ESS of the Poiseuille flow rate based method correlated with MWT. The OSI of both Womersley methods (either based on flow rate or centerline velocity) correlated with MWT. The Poiseuille centerline velocity based method did not show any good correlations with MWT. The interscan variabilities of the different methods are shown in Table 3. Womersley and Poiseuille based methods tended to have higher intraclass correlation coefficients than the velocity gradient modelling methods, indicating better interscan reproducibility.

Table 2 Relation between endothelial shear stress and carotid wall thickness
Table 3 Interscan measurement variability of endothelial shear stress estimations

Discussion

In the present study we show that for estimation of dynamic ESS from flow rate or centerline velocity in the common carotid artery, methods based on Womersley outperform Poiseuille based methods. Furthermore we show that flow rate-based methods perform better than centerline velocity-based methods. We base these conclusions on the correlation of predicted ESS with the mean wall thickness of the vessel. A third method, based on the velocity gradient near the wall, also predicted ESS that correlates with MWT, but this method requires detailed information on the flow pattern, which is not commonly available. All methods had good interscan reproducibility. Our findings implicate that the easily applicable Womersley based method enables better ESS, SPI and OSI estimation in future clinical studies and can replace the currently frequently used Poiseuille based methods.

Womersley theory is far from new [14, 21]. Yet, it is not commonly applied for estimation of ESS parameters [22] with few in vivo applications [2326], and studies on the correlation with MWT in the common carotid arteries were lacking. The fundamental difference between Poiseuille and Womersley flow is that the latter takes the inertia of blood into account. This effect of inertia depends on the flow dynamics and the geometry and is known to be substantial for common carotid arteries [27]. For the mean ESS, inertia is not relevant and Poiseuille and Womersley provide identical estimates.

A low time-averaged ESS is considered to be atherogenic [3, 68, 2830]. We indeed observed that low mean ESS, when using the flow rate as input, or based on the velocity gradient method, correlated with a thicker wall. We did find differences in estimates for mean ESS between the methods. In particular, using centerline velocity as input predicted lower values, which also were no longer correlated with MWT. This may be caused by more inherent noise in centerline velocity as compared to the flow rate, as well as by systematic deviations of the time-averaged velocity field from an axisymmetric parabolic profile, towards a flatter profile. For flow rate as input, the difference between the Womersley-based prediction and the velocity gradient method was limited to ~13 %. Such systematic deviation would not affect the relevance for clinical studies, as demonstrated by the significant correlation of mean ESS with MWT. Taken together, the data suggest that flow rate should be preferred over centerline velocity as input for estimation of mean ESS.

Temporal variations in ESS and notably a change in direction during part of the cycle have strong effects on endothelial biology towards a pro-atherogenic phenotype [1, 3, 28, 29]. The more than twofold difference in estimated SPI between Poiseuille and Womersley flow and the lack of correlation between Poiseuille-based SPI and OSI estimation and MWT demonstrate the necessity to account for blood inertia in pulsatility assessment. This is notably the case in peak systole, where a major deviation from a parabolic flow profile occurred. The supplementary material analysis shows the consistency of flow profiles under Womersley and Poiseuille.

Importantly, in Womersley flow, phases of negative ESS can co-exist with continuous forward flow and forward centerline velocity. In Poiseuille flow, this is not possible. Indeed, Womersley flow predicted a non-zero OSI in part of the cases, while for Poiseuille flow OSI was zero, in concordance with the absence of flow reversal in the studied non-stenotic common carotid arteries. The occurrence of a non-zero OSI in the Womersley predictions depends primarily on the balance of mean flow and the amplitude of the pulsatile components. Also, the SPI was defined as this balance. The correlations of the Womersley-predicted OSI and SPI with MWT strongly suggest that this balance is clinically relevant. Vascular stiffening and increased pulse pressure are conditions that would increase the pulsatile flow components and cause a non-zero OSI. A high mean flow would reduce the likelihood of a non-zero OSI. It remains to be established whether the correlation of low mean flow with thicker walls that we observed is in fact the consequence of reversal of ESS during the heart cycle.

A substantial difference remained between Womersley-based and gradient-based estimates of ESS pulsation. Extensive work on software phantoms has evaluated the accuracy and precision of a comparable velocity gradient technique as a function of scanning and resolution parameters [31], providing arguments for the accuracy of this method. However, in the absence of a golden truth in the current data, there is no straightforward way of quantitating accuracy of any of the methods. We suspect that the bias is also in the gradient method. ESR is defined as the velocity gradient in an infinitesimal small space above the endothelial cells. This gradient quickly falls further away from the endothelium in notably dynamic flow. The gradient method, having a finite resolution, might thus underestimate ESS pulsatility and also miss local reversal of the direction. Womersley provides a continuous solution for velocity as function of distance from the endothelium, which under the assumption of no-slip should give an inherently better estimate of ESS pulsatility. The Womersley-based and gradient-based SPI correlated equally well with MWT, indicating that these differences do not affect the usefulness of either SPI estimate in clinical studies. However, the OSI, which relies on subtle changes of the flow in the boundary layer, only correlated with MWT for Womersley-based methods.

Applicability

The current approach uses vascular diameter and either centerline velocity or flow rate as input, without requirement for more detailed data on the spatial velocity profile. Techniques that can provide such data include Echo-Doppler scanning and phase contrast MRI. The ability to estimate ESS from Echo-Doppler scanning allows implementation as pseudo-endpoint in large clinical studies. MRI methods based on direct measurement of local velocity vectors in 3D, such as 3D PC-MRI, may form a good approach for dedicated studies in complex geometries [32]. Yet, the current findings suggest that a much simpler implementation of Womersley-based estimation of shear parameters with consequently much shorter scanning times is a good alternative strategy in clinical studies on common carotid arteries.

All applied methods had good reproducibility. Reproducibility of the flow-based Womersley estimates for mean ESS and pulsatility exceeded that of the gradient modelling. While flow rate should be preferred over centerline velocity for mean ESS, for pulsatility (SPI and OSI) centerline velocity also provides a good correlation with MWT and good reproducibility, supporting the use of ultrasound or simple MRI protocols for pulsatility quantification.

Limitations of the study

It is clear that the requirements for Womersley flow were violated to some degree. Recently, Mynard et al. [33] evaluated the consequences of carotid curvature on Poiseuille- and Womersley-based estimates for mean and peak ESS. In that study, computational fluid dynamics based on the Navier–Stokes equations were used as the gold standard. Their results indicate that even a slight curvature causes substantial bias and circumferential heterogeneity of these estimates. The effect of curvature on SPI or OSI was not tested. In the current study, we did not apply computational fluid dynamics, and it remains to be established to what extent deviations from a straight tube influence the pulsatile ESS parameters and their correlation with MWT. Computational fluid dynamics methods are, apart from being time-consuming, affected by errors in boundary conditions and geometry definitions, as well as by assumptions on wall rigidity and blood rheology, precluding their general use in large clinical studies.

In the absence of a gold standard for ESS in large clinical studies, evidence for usefulness of any ESS method comes from correlation with wall biology and thickening, and it seems reasonable to evaluate the accuracy of the ESS estimates from the correlation with MWT. The current correlation of Womersley-based mean ESS, SPI and OSI with MWT as a marker for atherosclerotic burden is not necessarily causal; longitudinal studies on predictive value of shear parameters for cardiovascular events should allow evaluating their usefulness as pseudo-endpoints in clinical studies.

Conclusions

In the present study we developed a reproducible and accurate Womersley flow based method to estimate the endothelial shear stress dynamics in the common carotid arteries. We showed that this method is equally good as the more complex velocity gradient modelling method and superior to the simpler Poiseuille flow based method to estimate ESS. The Womersley based method we present solely requires the measurement of flow rate or centerline velocity and carotid artery diameter. This renders the method applicable for large clinical studies without the need for complex analyses and computations. While having its limitations, the Womersley based method should therefore replace the frequently used Poiseuille based method, enabling better assessment of the relationship between ESS, atherosclerosis and cardiovascular disease in future clinical studies.