Introduction

The study of the phenolic composition of natural products is of interest in light of the bioactive roles that these compounds play in plants and as part of the human diet [1]. Due to the complexity of plant phenolics, however, conventional high performance liquid chromatography (HPLC) often does not provide sufficient separation for these samples. Comprehensive two-dimensional liquid chromatography (LC × LC) has for this reason found increasing application in the study of plant phenolics in recent years [2,3,4]. While the availability of commercial instrumentation has greatly improved the implementation of LC × LC in many laboratories, the complexity of LC × LC method development continues to limit the more widespread use of the technique. This is because method development in LC × LC requires consideration of a large number of intricately related experimental parameters and instrumental restraints and their effects on achieving mutually incompatible goals in terms of peak capacity, analysis time and dilution.

To avoid the pitfalls associated with trail-and-error LC × LC method development, which is often counter-intuitive because of the complex relationship between experimental parameters, several step-wise LC × LC method optimisation schemes have been reported [5,6,7,8,9,10,11]. A comprehensive kinetic optimisation protocol for online LC × LC was reported by Vivó-Truyols et al. [8], who used multi-parameter Pareto-optimisation to derive optimal LC × LC experimental conditions. A similar approach has subsequently been used to derive optimal conditions for the RP-LC × RP-LC separation of peptides [10, 11]. Recently, Pirok et al. [12] reported a method that also allows for the optimisation of mobile phase gradients to maximise resolution and orthogonality.

In the context of phenolic analysis, the combination of hydrophilic interaction chromatography (HILIC) and reversed-phase LC (RP-LC) has been shown to be especially beneficial due to the high degree of orthogonality associated with this combination for many classes of phenolics [13, 14]. Method development for online HILIC × RP-LC is, however, particularly challenging due to the devastating effect injection of organic-rich HILIC mobile phase can have on the performance of the second dimension (2D) RP-LC separation. In cases where the mismatch between first dimension (1D) mobile phase (effectively the sample solvent for transferred fractions) and 2D separation is less severe (such as RP-LC × RP-LC systems for example), well known equations to model injection band broadening have been used with some success in method optimisation protocols [5, 8, 10, 11]. However, we have recently shown that under extreme sample solvent mismatch conditions such as encountered in HILIC × RP-LC, predictions using these equations, both for direct transfer of undiluted fractions and large volume injections of diluted fractions, are inaccurate [15]. In such cases, more sophisticated methods, such as reported by Stoll et al. [16], may be required to accurately model the effect of injection band broadening. Unfortunately, incorporation of such approaches into method development protocols is currently not feasible due to impractical computational power requirements.

A number of strategies have been devised to circumvent problems arising from mobile phase mismatches in LC × LC. These so-called ‘active modulation’ approaches include basic solutions such as splitting the 1D effluent [9, 17] and diluting the 1D effluent with a weak 2D mobile phase [18], as well as more technically advanced options such as the use of trapping columns [19, 20], thermally assisted modulation [21], membrane evaporation [22] or vacuum evaporation [23]. While the latter four methods can greatly improve the performance of LC × LC separations, they are difficult to implement and require dedicated method development protocols. On the other hand, simpler active modulation strategies such as splitting and/or dilution of the 1D effluent can effortlessly be incorporated into method optimisation schemes [9,10,11, 15].

Since accurate prediction of injection band broadening in HILIC × RP-LC is not feasible using current theoretical models, an alternative approach is to first establish experimental conditions where injection band broadening is effectively avoided. Provided that these conditions are met, injection effects can then be disregarded during method optimisation. We recently reported a Pareto-based kinetic optimisation program for HILIC × RP-LC separations [15] which incorporates dilution of the 1D effluent to negate the effect of injection band broadening in the 2D under previously established conditions. The program uses the theoretical relationships between various chromatographic parameters and analyte properties to predict performance for any set of conditions and, by comparing the performance of numerous theoretical combinations, is able to derive optimal experimental conditions for any given separation in terms of analysis time, peak capacity and dilution. This program was use to compare the performance of different column dimensions and to derive optimal conditions for the online HILIC × RP-LC separation of procyanidins. In the current contribution, we demonstrate the application of this kinetic optimisation protocol in combination with fixed active modulation requirements to derive a generic HILIC × RP-LC methodology for the analysis of complex natural phenolic mixtures. Hyphenation of optimised HILIC × RP-LC methods to UV and high resolution mass spectrometric (HR-MS) detection is shown to provide a powerful analytical methodology for the detailed analysis of a wide range of phenolic compounds, including monomeric and oligomeric flavonoids, phenolic acids and chalcones in several natural product extracts.

Experimental

Reagents and Materials

HPLC grade methanol (MeOH), formic acid and diethyl ether were purchased from Sigma-Aldrich (Steinheim, Germany) and HPLC grade acetonitrile (ACN) from ROMIL (Waterbeach Cambridge, England). Deionised water was obtained using a Milli-Q water purification system (Millipore, Milford, MA, USA). Grape seed extracts were provided by the Institute of Wine Biotechnology (IWBT, Stellenbosch University), dry rooibos tea samples by the Central Analytical Facility (CAF, Stellenbosch University) and experimental grape and wine samples (both Shiraz, 2016 vintage) were supplied by the Department of Viticulture and Oenology (Stellenbosch University).

Sample Preparation

Grape Seeds

Grape seeds were separated from the berries and washed with water. The seeds were crushed using a homogenizer and extracted with a 2:1 (v/v) acetone/water solution in a ratio of 1:10 (sample:extraction solvent) for 24 h at 4 °C [24]. After freeze-drying, the sample was reconstituted in MeOH (1000 ppm), filtered through a 0.45-µm hydrophilic PVDF filter membrane (Millipore) and diluted to 75% ACN.

Rooibos Tea

Dried tea leaves [2 g each of three unfermented and one fermented sample(s)] were soaked overnight in a 50% MeOH solution containing 1% formic acid (15 mL) [25]. The mixtures were sonicated for 1 h and centrifuged for 5 min. Each sample was diluted to 80% ACN prior to injection.

Wine

100 mL of the wine sample was extracted with 3 × 100 mL diethyl ether and evaporated under vacuum at room temperature until less than 1 mL remained [26]. The remaining liquid was reconstituted to 1 mL with MeOH and diluted to 80% ACN.

Grapes

Frozen whole grapes (50 g) were pitted and crushed using a mortar and pestle. The crushed pulp and skins were sonicated for 2 h in 100 mL 0.1% formic acid methanol/water (30/70, v/v) and centrifuged for 5 min. The supernatant was extracted with 3 × 100 mL diethyl ether and evaporated under vacuum at room temperature to less than 0.5 mL. The remaining liquid was reconstituted to 0.5 mL with MeOH and diluted to 80% ACN.

Instrumentation

A schematic illustration of the instrumental configuration used is presented in Fig. S1 (Supporting Information, SI). HILIC × RP-LC analyses were performed using a Waters CapLC 920 pump and autosampler (equipped with a 2-µL loop) in the 1D (Waters, Milford, MA, USA), an Agilent 1290 Infinity II binary pump in the 2D and a 2-position/8-port valve equipped with two 80-µL loops as interface between the two dimensions (Agilent Technologies, Waldbronn, Germany). An Agilent 1100 isocratic pump was used to dilute the 1D eluent before the valve and the 2D column was placed in an Agilent 1200 column oven. UV–Vis spectra were recorded between 190 and 640 nm (80 Hz) using an Agilent 1290 Infinity II DAD detector (1 µL flow cell). The system was coupled to a Waters Synapt G2 quadrupole time-of-flight (Q-TOF) mass spectrometer operated in negative ionisation mode. The following ionisation parameters were used: capillary voltage: 2.5 kV; cone voltage: 15 V; source temperature: 120 °C; cone gas flow (N2): 50 L h−1; desolvation gas flow (N2): 650 L h−1; desolvation temperature: 275 °C. The system was operated in MSE mode, which allows the acquisition of both low (4 eV) and high (ramped from 10 to 30 eV) collision energy data alternately in a single analysis. Data were recorded in the range of 150–2000 amu using a scan time of 0.2 s. Ion mobility (IM) separations were achieved using nitrogen as drift gas at a flow of 90 mL min−1 (3.27 mbar), with a He flow of 180 mL min−1 (1410 mbar) in the He cell. The mobility T-Wave velocity and wave height were 448 m s−1 and 37.1 V, respectively, and the transfer velocity was 380 m/s. For further details on the IM separation, the reader is referred to [27].

Leucine enkephalin (m/z = 554.2615) was used as the lock mass for accurate mass determination. The CapLC was controlled by MassLynx software (v4.0, Waters); the Q-TOF was controlled by MassLynx software (v4.1, Waters) and the rest of the modules (including the valve) by OpenLab CDS software (Agilent). Contour plots were constructed using LC Image (v2.6, GC Image LLC, Nebraska, USA).

Chromatographic Conditions

All 1D separations were performed on a Xbridge Amide column (150 × 1.0 mm, 1.7 µm, Waters) using mobile phases consisting of (A) 0.1% formic acid in ACN and (B) 0.1% formic acid in water at a flow rate of 11 µL min−1. Different 1D gradients were used for each sample, as outlined in Table 1. The 1D eluent was diluted with 0.1% formic acid in water at a flow rate of 99 µL min−1 (10 × dilution) prior to entering the modulation valve.

Table 1 Gradient conditions used for the HILIC × RP-LC separations of each of the investigated samples

In the 2D, a Kinetex C18 column (50 × 3.0 mm, 1.7 µm, Phenomenex, Torrance, USA) was used for the analysis of the grape seed sample, and a Zorbax Eclipse Plus C18 column (50 × 3.0 mm, 1.8 µm, Agilent) was used for analysis of the tea, wine and grape samples. In 2D, the mobile phases consisted of (A) 0.1% formic acid in water and (B) ACN. All 2D columns were thermostatted at 50 °C. A modulation period of 0.56 min was used for all analyses with the 2D flow rate and gradients as indicated in Table 1. For the grape seed sample, the 2D column effluent was split 1:5 after the UV detector, resulting in a flow of 0.5 mL min−1 to the MS. For all other samples, the effluent was split 1:4 between the MS and UV detectors.

Calculations

All computations were performed using in-house written Matlab R2010a (The Mathworks, Natick, USA) scripts, with the results exported to Excel. For a detailed description of the equations used to derive the optimum chromatographic conditions in terms of peak capacity, analysis time and total dilution factor, the reader is referred to our previous study [15]. A list of all the values used in the optimisation procedure is provided in Table S1.

Results and Discussion

Derivation of Optimal HILIC × RP-LC Separation Conditions

Chromatographic conditions for the HILIC × RP-LC separations were derived using a predictive kinetic optimisation program previously described [15]. As with most kinetic optimisation protocols, this methodology requires prior selection of stationary and mobile phases in both dimensions to optimise selectivity prior to kinetic optimisation. Alternative approaches have been reported to optimise mobile phase gradients and therefore selectivity in LC × LC [12]; in the present work, these were briefly optimised in 1D mode prior to kinetic optimisation of their combination in LC × LC. In addition, the protocol requires information regarding the retention behaviour (LSS parameters) and kinetic performance (plate height parameters) of the target analytes under the investigated conditions. For the purposes of this work, analyte constants for phenolic compounds were obtained from previous studies, as described below.

In [15], it was shown that the extent of injection band broadening in the 2D was negligible for weakly retained flavanols (2kw = 11.8 and 32.9, where kw is the retention factor in pure weak solvent) as long as the 1D HILIC mobile phase was diluted 10 times with water and the injection (fraction) volume was limited to less than 30% of the 2D column void volume (2V0). Since similar retention behaviour was expected for most other classes of phenolics in the studied samples, the same fixed active modulation requirements were used in the derivation of the proposed generic methodology (Table S1). The validity of this assumption is confirmed by the experimental results reported below. It should be noted that active modulation requirements such as the dilution factor depend strongly on the retention behaviour of compounds in the 2D, and that these should therefore ideally be determined for the target analytes under investigation. While the current findings can be considered generally applicable to phenolics, they are not generic for all compound classes.

For accurate performance prediction using the Pareto-optimisation method, information relating to the target analytes (i.e., plate height terms, diffusion coefficients (Dm’s) and linear solvent strength (LSS) parameters) is required. However, to experimentally determine these parameters for all target analyte classes in the investigated samples on different stationary phases would be impractical, as is indeed the case in many LC × LC studies. It was, however, shown [15] that although the accuracy of the predicted performance depends on the analyte constants used as input, the optimal chromatographic conditions do not differ greatly as a function of these values. This means that approximated analyte constants can be used as input to derive kinetically optimal conditions. Therefore, since the goal of kinetic optimisation in the current study was to derive optimal experimental conditions—as opposed to predicting the attainable performance—analyte constants for a series of oligomeric procyanidins [degree of polymerisation (DP) 1–5] previously determined on diol HILIC and superficially porous [15] and fully porous [28] C18 RP-LC columns were used for method optimisation in the present work (Table S1). These compounds, as the main constituents of grape seed tannins, are ideal reference compounds for this particular sample. However, they are also suitable model compounds for phenolics in general since they share chemical properties with flavonoids and include high molar mass compounds (up to 1442) to accommodate the chromatographic behaviour of oligomeric and highly glycosylated species of similar masses. Furthermore, because the required gradient span (Δφ) depends on the retention behaviour (LSS parameters) of the target analytes, the gradients required on the columns used to determine the employed LLS parameters were utilised in the optimisation. The total analysis time used for kinetic method optimisation was set to 70 min, which is near the maximum time the MS could record drift time filtered MS data.

For method optimisation, two 1-mm internal diameter (i.d.) 1D columns and four 2D columns of different dimensions and packed with superficially or fully porous particles were selected. These columns were selected primarily based on their availability and the suitability of their dimensions for high resolution HILIC × RP-LC separations [15]. For a detailed description of the columns considered during optimisation, refer to Section S1 in the SI. The resultant Pareto fronts presented in Fig. S2 show the optimal performance attainable for different combinations of the studied columns in terms of practical peak capacity (corrected for 1D undersampling) as a function of total dilution for a 70-min analysis time. Based on the Pareto fronts for these columns (Fig. S2) calculated using the parameters listed in Table S1, we selected the 150 × 1.0 mm, 1.7 µm column in the 1D, since this column provided better performance than the 250 × 1.0 mm, 5 µm column for all column combinations due to the smaller particle size. An amide column of the former dimensions was used for all 2D analyses, since this phase has been shown to offer good performance for anthocyanins [29] and procyanidins [30]. In the 2D, two different columns were selected: a Kinetex C18 (50 × 3.0 mm) column packed with 1.7 µm superficially porous particles was used for the grape seed analysis, and a Zorbax Eclipse Plus C18 (50 × 3.0 mm) column packed with 1.8 µm fully porous particles for the analysis of the tea, wine and grape samples. Although the combination of the 1.7 µm superficially porous column with the 1 mm 1.7 µm 1D column provides the best predicted performance (Fig. S2, green line), the lifetime of the former column was found to be limited under LC × LC conditions (~ 30 h). The superficially porous column was therefore only used for the analysis of the most complex sample investigated here, the grape seed sample. In contrast, a column of the same dimensions but packed with 1.8 µm fully porous particles, although less performant than the superficially porous phase, was found to be more stable and was therefore used for the separation of the less complex grape, wine and tea samples. The experimental conditions selected for analysis of the samples (“Chromatographic conditions” section) correspond to the two marked points (labelled #1 for the grape seed sample and #2 for tea, wine and grape samples) lying on the relevant Pareto fronts for the 150 × 1.0 mm, 1.7 µm × 50 × 3.0 mm, 1.7 µm and 150 × 1.0 mm, 1.7 µm × 50 × 3.0 mm, 1.8 µm column combinations (Fig. S2). These conditions were selected based on maximum separation performance (practical peak capacities of ~ 2000 to 2600) attainable for an acceptable total dilution factor of ~ 6×. The optimal experimental conditions selected for both column combinations are surprisingly similar (Table 1, Fig. S2), with identical 1D conditions and modulation times, although the 2D flow rate was slightly higher on the 1.7 µm superficially porous phase. The conditions selected for the configuration incorporating the 1.8 µm fully porous 2D column involved a lower 2D flow rate to provide the same total dilution, which means that the effective 2D gradient time was therefore slightly shorter on this column. The longer 2D gradient time, coupled with the good plate height behaviour of the 1.7 µm superficially porous column, is responsible for the higher predicted peak capacity compared to the configuration with the 1.8 µm fully porous column of the same dimensions (2670 vs. 2040).

Following selection of experimental conditions based on kinetic considerations, 1D gradient profiles for each sample were tuned based on the peak distribution(s) obtained from one or two scouting experiments employing linear gradients of differing initial and final mobile phase compositions at the optimised 1D flow rate of 11 µL min−1. A generic 2D gradient (1–45% ACN) was used for all samples. Although further optimisation of the respective gradients for each sample may improve the 2D separations reported here, this would come at the expense of increased method development time and was therefore not explored.

Results obtained for the analysis of each of the studied samples using these kinetically optimised conditions are discussed in the following sections. Examples of the UV and MS contour plots obtained for each of these samples are presented in Fig. S3 (SI). Comparison of these contour plots shows a noticeable loss in resolution for the MS data, which is due to both unavoidable post-column band broadening in the MS flowpath and the relatively slow acquisition rate (5 Hz) of the MS compared to UV detection (80 Hz). Although the TOF-MS instrument is capable of faster acquisition rates, for the mass range used here higher scan speeds result in a significant decrease in signal-to-noise ratios and were therefore not used. Nevertheless, using the drift time filtered MS data, we were able to obtain clean low and high collision energy MS data for partially co-eluting compounds, which simplified compound identification; this aspect is addressed in more detail in another contribution [27]. Examples of the UV and MS spectra used to tentatively identify selected compounds in each of the samples are given in Fig. S4 (SI).

Grape Seed

Grape seeds are a rich source of oligomeric procyanidins. The analysis of these compounds is particularly challenging due to the large number of isomeric species involved [31,32,33]. While grape seed procyanidins are comprised of only four isomeric flavanol building blocks—catechin and epicatechin and their respective galloylated derivatives—variation in the type of interflavan linkage (C4–C6 or C4–C8), the order of connecting units as well as the degrees of polymerisation and of galloylation leads to the extreme structural diversity of grape seed condensed tannins (Fig. 1) [31]. Indeed, although grape seed procyanidins have been the subject of several LC × LC studies [32, 34, 35], their complete characterisation remains an elusive goal.

Fig. 1
figure 1

Chemical structures of the procyanidins detected in grape seeds

The UV contour plot (280 nm) obtained for the HILIC × RP-LC separation of grape seed procyanidins is shown in Fig. 2. The necessity for higher resolving power than provided by 1D LC is evident from the complexity of the contour plot. In the HILIC dimension procyanidins are separated according to DP (as indicated by the circles in Fig. 2), while RP-LC provides isomeric separation the 2D. This structured elution pattern is one of the advantages of HILIC × RP-LC separation for procyanidins [32]. However, it is also clear that separation in both dimensions becomes progressively worse with increasing DP due to the exponential increase in the number of possible isomeric structures. This is evident from the unresolved polymeric ‘hump’ observed in the high MW region at the end of the HILIC separation.

Fig. 2
figure 2

Reprinted (adapted) with permission from [27]. Copyright 2018 American Chemical Society

UV contour plot (280 nm) for the HILIC × RP-LC separation of grape seed procyanidins. The top figure shows the full contour plot with the marked section enlarged below. Circles indicate retention windows for procyanidins of the specified degrees of polymerisation (DPx). In the notation YXZ,X indicates the degree of polymerisation, Y the degree of galloylation (omitted if Y = 0) and Z distinguishes isomers. Labelled peaks correspond to Table 2 with the retention times provided in Table S2 (SI).

Table 2 Summary of procyanidin species detected in a grape seed extract

UV data can be used to distinguish procyanidins from other compounds present in the grape seed sample. However, since the UV–Vis spectra for all procyanidins are similar, they cannot be used to differentiate between procyanidin species. Mass spectrometry data are therefore indispensable for the identification of individual procyanidin molecular species. Hyphenation of HILIC × RP-LC separation to HR-MS detection was therefore used to tentatively identify 72 procyanidin species present in the grape seed extract based on accurate mass data (Table 2). These include procyanidins of a DP up to 7, with varying degrees of galloylation. The HILIC and RP-LC retention times of individual compounds identified in this manner are listed in Table S2 (SI). While the presence of higher DP compounds could be established based on MS data, these were not resolved as individual peaks [32].

Rooibos Tea

Rooibos tea products have attracted growing attention due to their acclaimed health-promoting benefits [36]. This herbal tea originates from the indigenous South African rooibos plant, Aspalathus linearis, and is produced from either the fermented (oxidised) or unfermented (green) plant material [37]. Rooibos tea contains a diverse range of phenolic compounds, including flavones, flavanones, flavonols and dihydrochalcones, with some compounds being unique to rooibos (Fig. 3). This phenolic complexity has led to several 1D LC, LC × LC and even capillary zone electrophoresis (CZE) methods being reported for the analysis of rooibos tea phenolics [25, 36, 38,39,40,41,42]. In addition to the species used for commercial rooibos production, several subforms or ecotypes adapted to different geographical locations have been identified [25]. In this study, two of these ecotypes originating from the Wupperthal district in the Western Cape province of South Africa [25] were analysed, along with commercial fermented and unfermented samples.

Fig. 3
figure 3

Structures of some of the phenolic compounds identified in rooibos. Numbers correspond to the labels in Fig. 4. C-glc C-linked glucoside, O-glc O-linked glucoside

Fig. 4
figure 4

Figure 4d reprinted (adapted) with permission from [27]. Copyright 2018 American Chemical Society

Contour plots obtained at 280 nm for the HILIC × RP-LC separation of the phenolic compounds in four rooibos tea samples: unfermented (a) and fermented (b) commercial samples, and two unfermented tea samples from the Wupperthal district (c, d). The inserts in the top right corner show the complete contour plots with the marked sections being enlarged. Peak numbers correspond to Table 3 and Fig. 3.

Table 3 Characteristics of the phenolic compounds tentatively identified in the rooibos tea samples by HILIC × RP-LC‒DAD-HR-MS

The HILIC × RP-LC-UV contour plots at 280 nm for the four rooibos tea samples are presented in Fig. 4. Excellent separation of rooibos phenolics is again evident from this figure, with highly glycosylated species generally grouped by their high retention in the HILIC dimension. For these samples, some evidence of an inverse relationship between HILIC and RP-LC retention is evident from the contour plots. The separation could therefore be further improved using reverse shifted gradients [43], where the % organic modifier in RP-LC is systematically decreased during the 1D gradient, although this was not attempted in the present work since the goal was to illustrate how the kinetically optimised HILIC × RP-LC methods are applicable to the analysis of a wide range of phenolic mixtures. Tentative identification of the separated compounds was based on comparison of retention times, UV spectra, accurate mass MS data and MSE fragmentation patterns with literature reports [25, 38, 39] (Table 3).

The contour plot of the fermented commercial tea sample (Fig. 4b) is noticeably different from the unfermented sample (Fig. 4a), with the former containing overall lower levels of phenolics (based on peak area). During fermentation, aspalathin (20) is converted to orientin (18) and isoorientin (19) via the intermediate flavanones eriodictyol-8-C-glucoside (11) and eriodictyol-6-C-glucoside (12 and 13) [44]. While peak areas for orientin (18) and isoorientin (19) between the fermented and unfermented samples were roughly similar, the levels of aspalathin (20) and the aspalathin dimer (34) in the fermented sample were significantly lower, while those of the three flavanones (11, 12 and 13) were higher. The lower concentration of rutin (8) in the fermented sample, as well as the detection of the quercetin aglycone (29), may be explained by the partial hydrolysis of rutin during fermentation [45]. Compared to the commercial tea sample, the unfermented samples of two ecotypes from Wupperthal are clearly differentiated based on their high contents of phloridzin (31) and the Sieboldin positional isomer (32). These two compounds are unique to tea species from the Wupperthal district [25].

Wine and Grape Samples

Phenolic compounds are important constituents of wine since they contribute to organoleptic properties such as astringency, bitterness, colour and ageing potential [46], and are also associated with the health benefits related to moderate wine consumption [47, 48]. The phenolic composition of wines is largely determined by that of the grapes used for production, as well as by winemaking practices. Phenolic profiles of grapes differ between cultivars, but are also influenced by environmental factors such as geographical location and microclimatic conditions [49].

In this study, Shiraz grapes of the 2016 harvest and red wine produced from these grapes were analysed to confirm the applicability of the proposed methodology for both types of samples. An ether extraction was performed to concentrate the compounds of interest and to remove the unwanted anthocyanins and higher molecular weight polyphenols. Since this extraction also removed compounds highly retained in the HILIC dimension, the 1D HILIC analysis time for these samples was shortened from 70 to 50 min (Table 1).

The UV contour plots (280 nm and 360 nm) obtained for the HILIC × RP-LC‒MS analysis of the grape and wine samples are presented in Fig. 6. In total 49 compounds were assigned, of which 45 (21 flavonols, 8 flavanonols, 8 phenolic acids, 3 stilbenes and 5 procyanidins) were tentatively identified based on the UV spectra, retention times, high resolution MS data and MSE fragmentation patterns (Table 4).

Table 4 Characteristics of the phenolic compounds tentatively identified in the wine and grape samples by HILIC × RP-LC‒DAD-HR-MS
Fig. 5
figure 5

Structures of the phenolic compounds tentatively identified in the wine and grape samples. Numbers correspond to Fig. 6 and Table 4. O-glc O-linked glucoside, DH dihydro

The contour plots presented in Fig. 6 also allow for the evaluation of differences in phenolic contents between grapes and wine. As expected, the phenolic profile of the wine differed significantly from that of the grapes used for its production (Fig. 6a, b) [50]. Flavonols, one class of grape flavonoids which mainly occur in the berry skins [46], is of particular interest in this regard [50]. For visual comparison of the flavonol profiles of the samples, contour plots at 360 nm are presented in Fig. 6c, d, since this wavelength is selective for flavonols (and anthocyanins, not present in the ether extracts analysed here). From these figures, the previously reported [51] group-type separation of flavonols by HILIC × RP-LC is also evident. The wine sample contained noticeably higher levels of the flavonol aglycones (16), flavanonol aglycones (1925) and free phenolic acids (2633, Fig. 6a, c). On the other hand, the grapes contained much higher levels of flavonol-glycosides (718) and di-glycosylated flavonol species (4347), which were not detected in the wine (Fig. 6b, d). This can be explained by the acid-catalysed hydrolysis of the flavonol-glycoside bond during wine production, which leads to higher concentrations of the aglycones in the wine compared to the grapes [50].

Fig. 6
figure 6

Contour plots obtained at 280 nm for the HILIC × RP-LC separation of the phenolics in ether extracts of a a Shiraz red wine sample, and b the grapes from which the wine was produced. The corresponding contour plots at 360 nm are shown in (c) and (d). Peak numbers correspond to Table 4 and Fig. 5

Performance of the Kinetically Optimised HILIC × RP-LC-MS Methodology

To assess the performance of the HILIC × RP-LC separations, the practical peak capacities and orthogonality were calculated for each sample. Details on how these calculations were performed can be found in Sect. S2 in the SI, and a summary of the values obtained is presented in Table S3.

Experimentally determined 1D peak capacities corrected for undersampling (1nc) were significantly higher, especially for the grape seed and rooibos tea samples, than the predicted value of 43 (Fig. S2, SI). This underestimation of 1D performance was expected, since plate height data measured on a diol column were used in the kinetic prediction, while a more efficient amide column was used in the experiments. Slightly larger average 1D peak widths were measured for the grape and wine samples, likely due to the shallow gradient employed in these analyses; much lower 1nc are also due to the shorter 1D gradients used for these samples. In the 2D, experimentally determined peak capacities for the superficially porous 1.7 µm and fully porous 1.8 µm columns were only 11% and 14% lower, respectively, than the values predicted using plate height data for procyanidins (Fig. S2, SI). The relatively good agreement between predictions and measured performance for the rooibos tea, wine and grape samples confirms the suitability of oligomeric procyanidins as reference compounds to derive optimal experimental conditions for a much wider range of phenolics. Practical 2D peak capacities corrected for undersampling (c,2D) ranged from 2200 for 50 min analyses to 4400 for 70 min analyses (Table S3). These values do not provide a complete estimate of separation performance, however, since orthogonality is not accounted for. Therefore, the orthogonality of the separations reported in this work was estimated using both the asterisk [52] and convex hull methods [53]. Orthogonality values determined in this manner ranged between 61 and 69% for the asterisk method, while surface coverage values obtained by the convex hull method were between 50 and 71% (Table S3), confirming the suitability of HILIC × RP-LC systems for phenolic analysis. Combining c,2D and surface coverage values, the average practical peak capacity corrected for undersampling and finite surface coverage [53] is 1900 for the methodology reported here, which represents excellent performance for analysis times in the order of an hour. Note though that a generic 2D gradient was used in the present work to minimise method development time. Further fine-tuning of the 2D gradient according to the composition of each sample, including the possibility of using shifting gradients [15, 43], is recommended to improve kinetically optimised configurations for given applications.

Conclusions

In this study, a predictive optimisation protocol incorporating fixed modulation requirements to limit injection band broadening was used to derive kinetically optimised experimental conditions for the online HILIC × RP-LC-DAD-MS analysis of phenolic compounds. To demonstrate the general applicability of the proposed methodology, several natural products (grape seeds, rooibos tea, grapes and wine) containing diverse phenolic constituents were analysed. The results confirm that excellent separation of condensed tannins, flavonols, flavones, flavanones, flavanonols, (dihydro)chalcones, stilbenes and phenolic acids can be obtained using the proposed methodology, as confirmed by practical peak capacities in the order of 2000 for 50–70 min analysis times.

Hyphenation of optimised HILIC × RP-LC separation to diode array and HR-MS detection enabled the detection of 156 phenolic compounds (72 in grape seeds, 35 in rooibos tea and 49 in wine and grapes), of which 149 were tentatively identified based on UV spectra, HILIC and RP-LC retention times, high resolution MS data and MSE fragmentation patterns. The presence of a relatively large number of unidentified minor constituents in each of the investigated samples highlights the complexity of natural phenolic mixtures, and the importance of developing more powerful analytical tools to investigate these. UV contour plots, furthermore, allow visual comparison of the phenolic profiles, thereby facilitating identification of differences between samples.

In summary, the kinetically optimised HILIC × RP-LC‒DAD-HR-MS methodology based on dilution of the 1D effluent and large volume injection offers a powerful and relatively fast methodology for the detailed screening of phenolic content in a range of natural products. While further fine-tuning of the methodology for particular samples might be required for detailed investigation of these, the reported protocol represents a suitable generic starting point for such studies also.