Abstract
Multidimensional gas chromatography (MDGC) is now established as a technique which can help resolve most of the co-elution problems presenting with conventional gas chromatography for highly complex samples. Essential oils (EOs) are often used in the optimisation and development of novel MDGC methods and related technologies, to demonstrate and assess performance. In this review, recent trends and technical developments in the optimisation of MDGC, modulation and system configuration, pertinent and applied to EO analysis, will be critically discussed. Optimisation of MDGC will be discussed with reference to different column configurations, modulation periods and detection. Attention is given to novel modulation systems, development of multiplex MDGC systems and new approaches which combine heart-cut and comprehensive modes within one system. A section of this review will be dedicated to the preparative application of the MDGC and its application in the isolation of less abundant compounds from complex EO matrices.
We’re sorry, something doesn't seem to be working properly.
Please try refreshing the page. If that doesn't work, please contact support so we can address the problem.
Avoid common mistakes on your manuscript.
Introduction
Essential oils (EOs) are mixtures of volatile and semivolatile compounds obtained by hydrodistillation, steam distillation, dry distillation, or by some suitable cold-pressing method such as in the case of citrus oils. Due to the volatility and complexity of most natural oils, gas chromatography (GC) is the most widely used technique for EO separation, identification, and quantitative characterisation [1]. GC with a single column is still currently the method of choice for the routine analysis of EOs. However, due to the limited peak capacity of a single column, complete separation of all compounds by monodimensional GC is not always possible [2]. However, adding a second column with a different stationary phase, connected to the first dimension column (1D) through a transfer device, can provide the necessary increased selectivity, and a significant increase in peak capacity and resolution. This technique is now well known as multidimensional gas chromatography (MDGC) and can be performed in either heart-cut (GC–GC) or comprehensive (GC × GC) modes [1, 2]. An essential element of MDGC is the efficient functioning of the transfer device, which acts to transfer fractions eluting from the first column onto the second dimension (2D) column. During this process, these fractions can be focused into a narrow band and sent to 2D column, which can provide increased sensitivity, and help with the identification of trace-level components.
In GC–GC, only selected, discrete fractions from the 1D column separation are sent to the 2D column [3], whilst in GC × GC, there is continuous modulation (fractionation) of the 1D separation, where each fraction is transferred to the short, typically narrow-bore, 2D column. Due to the separation of the whole sample on both columns, GC × GC offers an increase of overall resolution, total peak capacity and selectivity [4]. In terms of EO analysis, compared to GC–GC, GC × GC can analyse the whole sample on both columns in a single analysis. Therefore, in the last 10 years, GC × GC has been a popular choice for advanced EO characterisation [1, 5]. Several general reviews have been published in recent years focusing on advances in MDGC [1,2,3,4, 6,7,8,9,10,11,12,13,14,15,16,17,18], some of which have in their scope the application of MDGC in EO analysis [1, 4, 7, 8, 10, 13, 15, 18].
Technical innovations continue to generate new capabilities in MDGC, that act to increase the separation power of this technique. In addition, ongoing optimisation of current systems also continues to provide increasing performance. In this review (Part 1 of a two-part series), we will limit the scope to the technical advances in MDGC affecting EO characterisation over the past 10 years. All the instrumental parameters—including size and type of columns, modulators, and detectors will be critically discussed, as all are important components within an optimised MDGC method. In addition, systems that combine comprehensive and heart-cut modes, multiplexed systems with dual columns in one of the two dimensions, and MDGC applied for preparative purposes, are also reviewed. The application of MDGC in the characterisation of EOs analysis will not be discussed herein, as this will be covered within the second part of this mini review series.
MDGC Applied to Essential Oil Characterisation
To discuss MDGC separations of EOs, it is first necessary to understand the general nature of EOs. EOs typically range from moderate to highly complex mixtures and contain classes of organic compounds of varying volatility. These compounds, products of the secondary metabolism processes within the plants [7, 8, 10, 13], are mainly terpenes (mono-, sesqui- or diterpenes) and terpenoids, the latter having additional functional groups, like alcohol, aldehyde, ketone, ether, ester or acid. Many compounds have chiral centres, forming enantiomeric pairs, where one of them can be predominant or the only one present in an EO [7]. The diversity and varying degree of complexity of EOs is discussed in much greater detail the Part 2 of this mini review series, where the focus is on MDGC applications.
GC separations, as performed upon a single column, predominantly reflect two fundamental characteristics of the column, its selectivity and efficiency. Column efficiency contributes to peak capacity, a theoretical maximum value which can be defined as the number of peaks which can be separated, at a given minimum resolution (typically Rs = 1) across the entire chromatogram. For a standard 30 m × 0.25 mm ID GC column, peak capacity is typically ~ 400. However, for most real samples the compounds rarely elute as evenly distributed peaks across the entire chromatogram, but rather more often elute as peak clusters and co-eluting peaks, separated by windows of empty baseline [18]. To predict the number of observable peaks and to statistically describe overlap in 1D and 2D chromatographic system, Giddings and Davis postulated statistical-overlap theory, which defines the probability of single peak formation as \(p={{\text{e}}^{{\raise0.7ex\hbox{${ - 2m}$} \!\mathord{\left/ {\vphantom {{ - 2m} n}}\right.\kern-0pt}\!\lower0.7ex\hbox{$n$}}}}\), where an m component mixture is analysed by a chromatographic system with maximum number of components that can be separated at certain resolution, defined as n [19, 20]. Therefore, peak capacity of the system needs to greatly exceed the number of components in the mixture, \(\alpha =m{\text{/}}n\) with α << 1 [21, 22]. For the EOs, with couple of hundreds of compounds this means that it is necessary for introduction of 2D column to increase peak capacity of the chromatographic system.
The correct choice of column chemistry (selectivity) substantially helps in this regard and for relatively simple EOs, proper selection of stationary phase can directly provide the overall resolution required [18]. However, for the more complex EOs, where the individual components can number several hundred, the solution lies in applying a multidimensional approach [3, 18]. Here overall peak capacity is equal to the product of the peak capacity achieved within each dimension, such that theoretically at least (and provided orthogonal selectivity between columns), an order of magnitude or more of an increase in peak capacity can be achieved [15]. The actual mode of MDGC applied—heart-cut or comprehensive—as well as the choice of modulator and modulation period, in the case of GC × GC, may depend on the specific application, herein the complexity and nature of the EO.
Each of the above options, together with related conditions and parameters, technologies and capabilities will be explained in detail within this review. As a reference Tables 1 and 2 are provided as an upfront summary of the MDGC configurations, columns, transfer devices and detectors applied to EO analysis over the past decade.
Off-line Pre-separation Using Liquid Chromatography
Despite the remarkable resolving power of MDGC, in some cases a fractionation step prior to the MDGC separation has been applied, to decrease the complexity of the EOs and further improve overall resolution. To our knowledge, in the past 10 years which align with the scope of this review, this has been done only by HPLC. In this scenario, HPLC is used as an orthogonal technique to GC, performing polarity-based chemical-class separations. In this way, a sample for analysis using GC becomes less complex and more homogenous, allowing for the GC separation of one or a few ‘classes’ of compounds at a time [13, 46]. Several authors have applied this hyphenated technique to citrus EOs. Tranchida et al. [45] used this concept for a more thorough chemical profiling of sweet orange and bergamot EOs (Fig. 1). The HPLC pre-separation gave two fractions: hydrocarbons and oxygenated compounds. Subsequent GC × GC analysis with a quadrupole mass spectrometer (qMS) as the detector, allowed the identification of many compounds reported for the first time. Figure 2 shows one of the reported chromatograms from this work. However, from this figure it is clear that in the sample shown, which contains the pre-fractionated sesquiterpenes, the peaks elute with similar retention times in 2D, suggesting that standard monodimensional GC would have provided essentially equal resolution for this fraction. Greater separations were achieved by applying GC × GC to the fractions containing oxygenated compounds Fig. 3a–d. Later, Zoccali et al. [46] used HPLC off-line to obtain fractions containing sesquiterpenes from several cold-pressed citrus EOs, which were then separated and identified by GC × GC. Compounds found in all samples can be used as general markers for citrus oils, while the ones identified only in a specific citrus EO can be used as the markers to distinguish different chemotypes [47].
Column Configuration
In the substantial majority of publications on EO analysis by MDGC in the last 10 years, the column used in the first dimension (1D) has been non-polar, either (100%) dimethylpolysiloxane or (5%) diphenyl (95%) dimethyl polysiloxane, typically with dimensions of 30 m × 0.25 mm × 0.25 µm, while in the second dimension (2D) (50%) diphenyl (50%) dimethylpolysiloxane or polyethylene glycol (PEG) columns were used (Tables 1, 2). The column configuration with a non-polar column in 1D and a more polar 2D allows separation in the 1D mainly due to dispersion interaction and in the second dimension according to the polarity of the components. Another reason for the application of this arrangement is that linear temperature programmed retention indices (LTPRI) obtained from the non-polar column are more rugged [65], providing higher similarity to the LTPRIs from data libraries. Methods for the determination of LTPRIs on the 2D column fall within the scope of the Part 2 of this mini review series.
The chemical nature and column manufacturing process for polar phases, involving different levels of cross-linking, can produce a higher variation in the resultant stationary phase properties. This has been particularly noted for polyethylene glycol-type (PEG) columns [87, 88]. These observations are in agreement with those made by Morrison et al. [49], whose evaluation of polar columns revealed that (50%) diphenyl (50%) dimethyl polysiloxane columns were more thermally stable compared to wax-based stationary phases, and less prone to degradation and change of performance with use. They also showed that PEG columns were well-suitable for the separation of polar components in EOs [49], and thus, the column of choice in the 2D (Tables 1, 2). However, it is important to also comment that the selection of a polar column for the 2D can lead to potential wrap-around effects, due to the high retention of most polar compounds, leading to a problem in their full separation and characterisation [65].
In their 2015 study, Cordero et al. [56] evaluated five different column and system configurations using a volatile model mixture (Table 3). Each setup was evaluated using several performance parameters to characterise GC × GC separations, namely re-injection pulse width (σ2i), net separation measure (SGC×GC), modulation ratio (MR), used separation space, and corresponding pixel-based area ratios (pixel counts). The initial column configuration evaluated was the most commonly applied combination, also being the manufacturer’s suggested configuration (Table 3), exhibiting very low 1D average carrier velocity resulting in longer analysis times and limited separation power. To increase the carrier gas velocity, a shorter column with smaller internal diameter and film thickness was used keeping the same phase ratio. Co-elution of several peaks was resolved by increasing the column loadability, increasing the film thickness, and increasing the length of the 2D column. Increasing the orthogonality of the column system using a PEG column in 2D provided a clear improvement on the performance, with this configuration having the highest value of utilised area and separation space, whilst at the same time producing the narrowest peaks and having the MR value above three (Table 3), which, according to the authors is required to obtain a good degree of confidence for area determination for trace solutes [56]. Thus, this setup (Configuration 4, Table 3) was used for the analysis of a very complex vetiver EO. Using Configuration 5, a polar column in 1D and a mid-polar in 2D was applied, which led to a very different peak distribution in the 2D contour plot. Compared to Configuration 4, there was a significant increase in the SGC×GC, with a narrower σ2i, MR being above 3, and at the same time decreasing used separation space and area (see Table 3). This configuration was applied for the determination of mint and lavender EOs [56].
In the most efficient separations obtained, the dimensions of the 1D column in the work of Cordero et al. were 10 m × 0.10 mm × 0.40 µm. However, columns used in the majority of studies published in the past 10 years were longer, with a larger internal diameter and thinner coating (30 m × 0.25 mm × 0.25 µm) (see Table 1). The advantage of the shorter column is the faster analysis, while the columns with a narrower diameter and thinner film thickness require more diluted samples or the application of a higher split ratio during injection. Ma et al. [23] performed a qualitative analysis of Artemisia annua L. using columns with higher retention factors and peak capacities, specifically DB-Petro 50 m × 0.20 mm × 0.5 µm, 100% dimethylpolysiloxane and DB-17ht 2.6 m × 0.10 mm × 0.1 µm (50%-phenyl) methyl polysiloxane (see Table 1). They tentatively identified 303 compounds, out of 700 peaks, with 33 having a relative abundance greater than 0.2%. Qin et al. [39] managed to characterise Citrus reticulata Blanco (Chenpi in Chinese, dry tangerine skin) pericarpium, a herbal medicine mostly used in Korea, China and Japan, using GC × GC/high resolution (HR)-time of flight mass spectrometry (TOFMS), with a 15 m × 0.25 mm × 0.25 µm low-polarity column in 1D, and a shorter 1.0 m BPX50 column with standard coating thickness in the second dimension. In this way, the speed of analysis was increased, with excellent peak identification using the HR-TOFMS. Using a GC × GC/qMS with a 60 m-long column in the 1D, Saucier et al. [40], managed to identify 130 compounds, out of which 109 were identified in the artichoke EO for the first time. As well as using most common column configurations, Namara et al. [75] and Robbat et al. [89] applied similar GC–GC approaches, with polyethylene glycol column in 1D and a non-polar column in 2D, to separate compounds present in juniper berry EOs, and EOs used for the development of gin, with the aim of building libraries of all compounds present.
Chiral Columns
Chiral compounds are commonly present in the EOs and the determination of their ratios, defined as enantiomeric excess (EE), which is specific for the species and chemotype, can thus be used for the assessment of the purity and potential adulteration of the EOs. Limitations in classical enantioselective 1D GC to distinguish targeted chiral compounds from the other interfering peaks, has led to the more widespread application of enantioselective MDGC, where one of the columns has chiral selectivity. Due to the limited number of chiral compounds within the samples, which in some cases elute within a narrow elution window [74], application of GC–GC is the more common approach, compared to GC × GC (Table 2). In the last decade the most common configuration has placed the enantioselective column in 2D, which allows the separation of compounds on the achiral 1D column and then transfer of the targeted compounds to 2D for the chiral separation (see Table 2) [79].
Application of the enantioselective column in second dimension of a GC × GC system can suffer from limited separation of chiral compounds on the short 2D column. Alternatively, GC × GC with enantioselective column in 1D requires that peaks should be sufficiently resolved to be determined in two or more different modulation periods, avoiding wrap-around, to obtain more reliable quantification. Separation must be optimised to obtain greater MR, which means increased data density, and minimising the unresolved overlapped region, allowing better reconstruction of the 1D peaks and better quantification of the enantiomeric pairs [79].
Wong et al. [79] applied GC–GC with an enantioselective column in the 2D, delivering the separation of chiral compounds within 25 min. However, utilising GC × GC with the enantioselective column in the 1D, the same chiral analysis was completed in under 20 min by optimising the column and oven temperature program. Further applications of enantioselective MDGC in the EO analysis will be covered in more detail explained in Part 2 [97].
Transfer Devices in GC–GC
To perfrom GC–GC analysis, a transfer device in the system needs to transfer selected fractions of the primary column effluent to the head of the 2D column. For this purpose, multiport valves have long been used as transfer devices in GC–GC, where the Deans Switch, and variants thereof, are by far the most commonly used transfer device for heart-cutting. Transfer devices based upon the Deans Switch contain two components, one within the sample path and one outside of the sample path. The part on the sample path contains no moving part and is located inside of chromatography oven, while the other, which contains movable parts (normally a solenoid valve) is located outside the oven. A Deans Switch uses an auxiliary flow of carrier gas to control the transfer of effluent from the 1D to 2D columns [2].
There have been several transfer devices recently introduced for heart-cutting. Agilent have developed their capillary flow technology (CFT) [90], and both PerkinElmer [91] and SGE [92] have introduced microchip devices which combine individual switch components into single small deactivated microchip plates. Such microfluidic designs help eliminate previous chromatographic problems by maintaining inert surfaces which prevent peak tailing or loss of solutes. In addition, low thermal mass allows the Deans Switch to equilibrate to the oven ramp temperatures quickly, which helps reduce peak broadening [3, 4]. The CFT Deans Switch was applied by Wong et al. [79] for the enantioselective analysis of tea tree EO and by Sasamoto et al. [83] in a 1D/2D GC/MS system applied to the analysis of lemon EO. Shimadzu have introduced a multi-Deans switching technology [93], with a heated transfer line connecting the two columns and with deactivated internal surfaces, and pressure control applied to both sides. The system is said to provide improved retention time reproducibility, and allows multiple transfers of different fractions during a single run. Gerstel’s multi-column switching approach [94] is based on positioning the exit of 1D column next to the entrance of the 2D column in a stream of auxiliary carrier gas, inside a glass cap [3, 4].
Development of Modulators
Modulation is the heart of GC × GC, providing accumulation and trapping, refocusing, and rapid release of the components eluting from the first column, providing for fast and efficient separations [63]. The most frequently applied modulators in GC × GC analysis of EOs (Table 1) are thermal, valve-based, and flow modulators (FM). Thermal modulators predominate, due to the fact that GC × GC systems with these modulators have higher sensitivity compared to the systems based upon flow modulators, despite their relative complexity. Advances in the technology of transfer devices in MDGC has been the topic of many previous reviews to which the reader is here directed [2,3,4, 6, 12, 16, 17].
Thermal Modulators
The most frequently used thermal modulators (TMs) in EO analysis are the cryogenic ones, longitudinally modulated cryogenic system (LMCS) and jet modulators (Table 1). LMCS is the oldest thermal modulator type, developed by Marriot et al. [95] and is able to increase sensitivity and peak capacity compared to conventional GC by trapping the peaks of interest using liquid CO2 and then rapidly heating up the the portion of the column holding the trapped components, providing fast release of the focused peaks into the 2D column (Fig. 4A, B-a). The shortcomings of this system are the inability to trap very volatile species due to the relatively high trapping temperature of − 50 °C, plus the necessity of moving parts, making this system less robust. However, since EOs tend to have as their most volatile organic compounds the monoterpenes, the use of LMCS is still applicable for most EO analysis and so applied by many researchers (Table 1). Compared to LMCS, jet-based modulators use either carbon dioxide (Fig. 4B-b) or liquid nitrogen, and have simple valves as the only moving parts (Fig. 4B-c, d). Most commonly applied are those with 4 jets (Table 1; Fig. 4B-d). For the analysis of very volatile compounds, with boiling points lower than hexane, the application of liquid nitrogen for the trapping is necessary due to the significantly lower trapping temperature of − 160 °C [6, 12, 16]. A great disadvantage of this system is the high consumption of liquid N2, up to the 100 L day−1, making this technique very expensive [12, 16].
Valve-Based Modulators
Valve Modulators
Advances in valve modulators have been described previously by Seeley [2]. Valve-based modulators began with diaphragm-based valve modulators to divert the effluent from the 1D column to the head of a short 2D column. The main limitations of these early systems were that only 10% of the primary effluent reached the 2D column, which obviously decreased the sensitivity of the analysis, and a temperature limit of 175 °C. The temperature limit could lead to the condensation of less volatile compounds in the valve [2]. This limitation was overcome by the introduction of perfluoroelastomer-based O-rings, instead of temperature-sensitive ones used in the first designs, allowing reliable operation of the system up to 325 °C [16].
The advantage of valve modulators, compared to thermal modulators, is the efficacy in capturing very volatile compounds, ease of use, and no consumption of liquid nitrogen, which make valve modulators a much better choice for the analysis of volatile compounds in remote locations using portable systems [53]. Lidster et al. [53] utilised a rotary-valve modulated (Fig. 5) GC × GC system to analyse gasoline and tea tree EO.
Flow-Switching Modulators
Flow-switching modulators (FMs) are based on collecting the 1D effluent in a collection loop during the fill cycle, and then sending the collected fraction to the 2D column by changing the direction of the auxiliary flow, during the flush cycle. FMs offer an advantage over thermal modulators, especially when applied in remote field operations, as there is no necessity for cooling by a cryogen [49, 50]. There have been reports by users of unsatisfactory performance because of difficulties in obtaining rugged modulation conditions [72]. However, it has been shown that by optimising the GC × GC system, the efficiency of FM can be increased. Application of a long narrow-bore column in the 1D [72], increases the resistance and enables efficient stop-flow modulation. In addition, the application of a long wider bore column in the 2D (e.g. 10 m × 0.32 mm × 0.20 µm) is common due to the high gas flow and the widths of injected chromatographic bands, normally associated with FM, which allows the use of lower pressure for flushing the loop having the flow exiting the chromatographic system compatible with some kind of MS detectors [18]. Thus, the correct optimisation of the column dimensions can increase the performance of the FM-based GC × GC system to provide satisfactory results [18].
In forward fill/flush (FFF) mode, a flow modulator uses a loop channel, positioned in between the two columns, to collect the effluent from 1D during the fill cycle, and the loop is then flushed in the same direction onto the 2D column [49]. FFF modulator has been found by users to give satisfactory results when concentrations of the target compounds do not differ significantly. Separation of highly abundant compounds in 2D can lead to peak tailing and co-elution with less abundant compounds, which is also a common problem with thermal modulators. This effect has been overcome using a modified reverse fill/flush (RFF) flow modulator (Fig. 6), by introduction of a variable collection channel and a restrictor, which works as an outlet for the carrier gas through the collection channel during the fill cycle, and allows to reverse the flow direction during the flush cycle. This setup offers less band broadening in the 2D and reduced overloading, without losing separation power and resolution [49]. RRF provides improved reproducibility of 1D and 2D peak retention times, with Cordero et al. reporting the highest value of variation being 6.58% (neomenthol) on the 2D column [56].
Tranchida et al. [50] developed a new flexible loop-type FM with efficiency comparable to a cryogenic FM, showing comparable peak widths (4σ) of 250–300 ms in 2D for (+)-β-pinene, isomenthone, menthol, and caryophyllene. However, the high gas flow associated with FM is a disadvantage for the flexible loop-type FM system (Fig. 7), especially when it is hyphenated with MS. Therefore, it was demonstrated that by using different restrictor lengths in the connections linking the modulator to the auxiliary pressure source, it was possible to increase the duration of re-injection and greatly reduce gas flows, down to 4 mL min−1, to make the system compatible with MS detection [51]. By optimising the diameter of the modulation loop column preceding the 2D column, and decreasing it to 0.10 mm, there was a reduction in the volume of the final segment of the loop leading to limited band broadening and a decrease in peak widths. Therefore, there was a significant increase in the resolution and decrease in the plate height of the system with FM [37].
Modulation Period
Modulation period is one of the most important parameters to be optimised in GC × GC. Because of the typically short length of the 2D column in GC × GC, a long modulation period can deliver a large number of compounds, which need to be separated on the 2D column. Effluent coming from 1D column, containing numerous compounds, can exceed the limited 2D column peak capacity, leading to co-elution in the 2D and decrease in already achieved resolution in the 1D and in overall performance of the GC × GC system. Also, longer modulation period leads to lower MR which is only good for screening or semi-quantitative analysis, where for the quantification is necessary MR of at least 3 [79] (already explained in “Chiral columns”). However, if the modulation period is too short, some compounds retained longer on the 2D column and do not elute from 2D column before the next modulation cycle starts causing wrap-around [24]. Ideally the 2D separation should be finished before the next injection of the subsequent fraction of the 1D effluent [4]. In the majority of studies performed over the last 10 years, the reported value for a modulation period was 6 s, and the values varied over the range of 2–10 s. In GC–GC, the time of the heart-cut is similarly important, as it is necessary that a whole section of interest be transferred quantitatively to the 2D column, to perform proper qualitative and quantitative characterisation.
Detection Systems
The detectors utilised in the past 10 years have mostly been either flame ionization detectors (FID), qMS, or TOFMS (Fig. 8; Tables 1, 2). For qualitative analysis qMS and TOFMS are the detectors of choice. FID is often coupled with qMS since it offers higher acquisition frequency, thereby improving the accuracy of 2D peak areas used for quantitative purposes [64]. Applied in multiplex systems, FID has also been shown to give more repeatable results, compared to qMS alone, probably again due to the same reason [57]. TOFMS, with its high data acquisition frequency across a broader mass range, spectral deconvolution of overlapping peaks, and selectivity, is a widely used hyphenated detector with MDGC [4, 17]. Von Muhlen et al. [43] showed that TOFMS had better peak resolution and definition, compared to qMS, with a higher signal-to-noise ratio, with TOFMS identifying 15% more compounds than qMS, which is important for the analysis of complex matrices such as EOs. TOFMS showed another advantage over qMS, in that it gives a higher similarity with the MS library than qMS. The explanation given for this was that TOFMS was less affected by mass spectral distortion than qMS. Also, the higher acquisition frequency needed to perform GC × GC is easier to obtain using TOFMS [43].
An advantage of the triple quadrupole mass spectrometer (QqQMS) is the option to monitor the presence and abundance of certain specific ions in the system, using multiple reaction monitoring (MRM) and selected ion monitoring (SIM) operational modes. SIM mode can be very useful for the identification and quantification of trace-level compounds co-eluting with more abundant ions. Also, since there are many compounds coming from the same chemical class, the application of MRM mode can be a useful tool in monitoring the levels of compounds containing the same functional group (targeted analysis), as well as in the trace analysis. These operational modes offer specific, selective, sensitive, and quick characterisation of target peaks, managing also to perform untargeted analysis of EOs [66] and increasing the identification capabilities of the MDGC system. The disadvantage of QqQMS is that spectral similarity between experimentally obtained mass spectra and library spectra is lower compared to qMS. This comes from the fact that MS library, FFNSC 2.0 used in Tranchida study [66], was built with GC/qMS systems, which can differ significantly from QqQMS. MSMS systems coupled with GC × GC can analyse the high peak density produced by GC × GC, which cannot be followed up by switching of ion selection in the quadrupole section. This can be overcome using cold-electron ionization instead of hard-electron ionization [96]. Also, there is a decrease in spectral similarity observed for GC × GC/MSMS, compared to GC/MSMS, due to the differences in the flow of effluent entering ion source (e.g. 8.8 mL min−1 vs. 1.3 mL min−1, respectively), which is the case in flow modulated GC × GC. A similar observation was made for GC × GC/qMS and GC/qMS [66].
With a high acquisition range of 15–1500 amu, with ability to provide peak purity, and perform real-time calibration, HR-TOFMS is able to provide mass accuracy of ± 0.002 amu. These features enable a HR-TOFMS system to provide enhanced mass spectral data sets for chromatograms obtained in MDGC, enabling deconvolution of co-eluting compounds [36].
Chemometric Techniques in MDGC Optimisation
Besides their application in data analysis, chemometric techniques can also be employed in the optimisation of MDGC processes. Omar et al. [63] applied experimental design to optimise the GC × GC parameters. Central Composite Design (CCD) and Multisimplex were utilised to optimise several parameters, including modulation period, discharge time, first column flow, and second column flow, using the peak volume of each analyte as the response, which is directly related to sensitivity and resolution [63]. A full review of the application of chemometric techniques to MDGC-generated data will be provided in Part 2 [97].
Novel MDGC Approaches
To increase the resolution of EO separation, many authors have tried to take simultaneous advantage of both heart-cut and comprehensive modes. This principle has been developed in two different ways. The first is that the effluent from the 1D column can be split in two 2D columns, one long, and a second short column. All compounds are transferred to both 2D columns, without separating and targeting certain regions and compounds [77]. The second principle is that several selected regions are focused with cryotraps and transfered, using a Deans Switch to a long column (GC–GC), while the other regions are sent to a short column (GC × GC). Since the 2D column dimensions are different, even though both have the same stationary phase, there is a difference in retention times and resolution. Furthermore, the application of the Deans Switch can influence retention times, giving a time lag for the compounds separated in the heart-cuts. Application of this system on lavender EO showed good reproducibility, while the use of a cryotrap enabled an increase in peak heights due to the refocusing of the re-injected fraction [52].
Sasamoto et al. [83] developed a 1D/2D GC/MS system using a CFT Deans Switch, and a dual low thermal mass GC, to perform a quick and easy switch between one-dimensional and two-dimensional systems without any change in instrumental setup (see Fig. 9). One of the options of this system was to perform a preliminary FID analysis to select specific heart-cuts for further and more in-depth analysis, as well as monitoring the total ion chromatogram for the 1D column during the run with the GC–GC system.
Multiplexed Dual Column Comprehensive Systems
GC × GC is state-of-the-art technology for the analysis of EOs, and using short and narrow-bore columns in the second dimension, separation can occur in a few seconds. However, when using thermal modulation, a flow mismatch can occur between the two dimensions due to the column serial configuration. Also, since the column in 2D is short and narrow, with a thinner stationary phase, in most cases having same phase ratio, it can easily be overloaded. This decreases the resolution on the 2D column, particularly where minor peaks present are in close proximity to more abundant compounds and thus co-elute. The solution is to build a multiplex system which has two parallel 2D columns (GC × 2GC). The other advantage of a dual-column system in the 2D is the synergism of dual detectors. MS is known to provide a fundamental contribution to analyte identification, while FID offers a continuous data collection, thereby improving the accuracy of 2D peak areas used for quantitative purposes. This can be observed in the example of the analysis of the EO obtained from A. umbelliformis Lam. Two of the most abundant compounds in this EO are α- and β-thujone which, in the case of standard GC × GC co-elute with the minor compounds, nonanal and 2-methylbutyl isovalerate. By applying a GC × 2GC system, these two minor compounds were separated and identified.
The second great advantage of this system is that it provides consistent results, in terms of analyte characterisation, offering internal cross-validation of quantification and accuracy [64]. Nicolotti et al. reported upon the characterisation of a GC × 2GC system, performed through validation parameters such as linearity, precision, and accuracy against a reference standard. Using a loop-type thermal modulator, cryotrapping and refocusing of the sample at the head of the 2D was achieved, offering higher loading capacity with an increase in peak capacity, resolution, and overall orthogonality. Results for the most common authenticity markers for Mentha spp., with an absolute quantitation error smaller than 13%, were reported as being in agreement or even better than the standard [8]. Such results can be achieved as GC × 2GC offers better chromatographic resolution, which leads to better separation and easier quantification of desired compounds [57].
In comparison to the multiplex system with the two columns in the second dimension, Savareear et al. [65] set up a system with two columns with different stationary phase polarities in the first dimension (see Fig. 10). This was possible because of a contra-directional modulation approach, according to which two 1D columns were aligned so neither one of them blocked the jet of cool gas onto the other column. The injected sample splits into two 1D columns and later recombines into one stream on the 2D column. As a result, two 2D contour plots are obtained containing complementary information. The great advantage of this approach was that different interactions in the first dimension between solutes and stationary phases lead to different contributions in the 2D column. One of the disadvantages of this system was that quantitation of the analytes needed to be performed by normalization of peak areas, as the split between the two columns in 1D was not necessarily always even.
Preparative MDGC
Besides application in the characterisation of complex samples, as a separation technique MDGC can be applied to the isolation and purification of compounds of interest from complex mixtures. 1D GC has already been established as a preparative method [98]. However, the low peak capacity of one column leads to significant co-elution and comparatively low purity of the isolated compound. To increase the amount of the purified compound in preparative gas chromatography, samples large enough are injected to overload the column, which significantly decreases the separation power of the system. This is a big problem for 1D GC system because the purity of the final product is even lower. However, overloading is necessary to increase the amount of isolated compound and decrease total run time. Thus, mega-bore columns, with 0.53 mm inner diameter and 0.5–1.0 µm coating, which can manage high sample volumes, are widely used in preparative chromatography.
The use of additional columns, as in MDGC, with different stationary phases providing some degree of orthogonality, allows isolation of pure compounds while at the same time decreasing overloading effects [99,100,101]. Thus, in recent years, MDGC has increased in importance for preparative purposes.
Eyres et al. [99, 100] developed microscale preparative GC–GC by trapping compounds eluting from the 1D column with a LMCS and applying a microfluidic Deans Switch to selectively isolate the compound of interest eluting from the outlet of the 2D column, instead of transferring it to a 2D detector. By doing 100 injections they managed to isolate 77.6 µg of geraniol with 75.5% recovery from a mixture of peppermint, spearmint, and lavender EOs with a purity of 99.3%. The amount obtained was enough for nuclear magnetic resonance (NMR) spectroscopy analysis. For the isolation of 1,4-dimethoxybenzene from peppermint EO using the same system, the recovery, with cryotrap set on − 20 °C, was 52.2%. Decreasing the temperature of the cryotrap, from − 20 to − 25 °C and − 30 °C, led to recoveries of 63.4% and 76.8%, respectively, showing the importance of cryotrap optimisation. The same system was applied by Ruhle et al. [102] separating pairs of compounds from a mixture of peppermint, spearmint and lavender EOs and obtaining them with high purity. However, separated carvone and linalyl acetate showed lower purity due to the overlapping peaks and overloading of the GC–GC system. Better recovery without the presence of interfering peaks can be achieved with more precise oven settings in the critical target region.
Compared to previous preparative 2D GC systems with cryotraps and Deans switches, Sciarrone et al. [101] applied a three-dimensional heart-cut GC (GC–GC–GC) preparative system, equipped with three Deans Switch devices to enable efficient consecutive heart-cuts, where after each injection the isolated compound was collected with a low-cost lab-constructed collection system. Sciarrone et al. used three mega-bore columns with the thickest non-polar column coating in 1D, 5 µm, guaranteeing the greatest sample capacity, while in 2D and 3D the column coating thicknesses were 2.0 and 0.8 µm. The inner diameter and coating of the columns applied by Sciarrone are thicker compared to those applied by Eyres. This enabled the application of three times higher sample volumes than in microscale preparative GC, obtaining at the same time higher amounts of isolated compounds. After three consecutive heart-cut steps, the purity of carotol was 99.3%, which was much better than 1D preparative GC, where carotol was co-eluting with three more interfering compounds, decreasing its purity to 70%.
Preparative GC–GC can be applied to determine the structure of unknown compounds as well as to confirm their structure via 1H NMR analysis. This way it is possible to identify compounds for which a match cannot be found in a MS library [99, 100]. Chin et al. [78] applied a peak enrichment technique to separate a compound from an EO. This technique provides on-line and in situ enrichment of the compounds that are heart-cut using a microfluidic Deans Switch device, cryotrapped, and stored after each injection. The main reason for developing this technique was to isolate and detect compounds that were present in trace amounts, or have a low S/N ratio, or for detectors that require a larger injected mass of compounds to provide adequate identification [78]. This technique can be also applied for the isolation of compounds for their quantification or for use as reference standards. This is of importance, because many compounds present in EOs cannot be found as commercial pure reference standards. Such was the case with α-amorphene and β-vetivone, for which standards could not be found, so Sciarone et al. [103] isolated them from Haitian vetiver EO applying a GC–GC–GC system. The addition of LC to a GC–GC system, was used as a pre-separation step with the aim of providing class-separation of sesquiterpenes and oxygenated sesquiterpenoids, allowing introduction of 5–10 times higher amounts of samples, due to the increased sample capacity of LC compared to GC.
Panto et al. applied [104] 3D LC–GC–GC and GC–GC–GC systems (Fig. 11) for the isolation of seven highly pure sesquiterpenoid alcohols from sandalwood EO. With a suitable choice of column configuration and heart-cuts, compounds co-eluting in the 1D were purified and collected individually. Application of LC, with all its aforementioned advantages when coupled with a GC–GC system, was exploited for the purification of low abundance compounds such as α-bisabolol, (Z)-lanceol, and (Z)-nuciferol [104]. Sciarrone et al. [105] showed that the application of an appropriate column configuration can lead to separation of one compound in each dimension with high levels of purity. By setting (5%) diphenyl (95%) dimethylsiloxane as the 1D column, patchouli alcohol was isolated, while after a 2D (100%) polyethylene glycol column α-bulnesene was isolated, and after a custom-made ionic liquid-based 3D column, α-guaiene was purified from the patchouli EO. In this way, in only one process, three compounds were isolated, each collected individually without any interfering compounds.
Concluding Remarks and Future Perspectives
The work performed in the past years shows that MDGC clearly has higher separation power when compared to conventional GC. This is especially observed in the analysis of complex matrices, such as EOs. In EO analysis, when untargeted separation is preferable, GC × GC has a clear advantage over the heart-cut technique. However, in targeted analysis, such as the enantioselective analysis of EOs, when only a few pairs of compounds need to be characterised, GC–GC has the advantage because of the longer 2D column, which provides higher peak capacity compared to the short column in GC × GC. Novel approaches in MDGC take the advantage of both the heart-cut and comprehensive modes, combining them into one system. In addition, several multiplex systems have been developed to increase the power of separation on the 2D column, decreasing the loaded effluent from the 1D column to the 2D and obtaining two different separation profiles when two different columns were applied in 1D. This clearly shows that in the future even more information about the EOs can be obtained from the chromatographic process, helping provide faster and more accurate peak identification, EO authentication, and quality assurance.
In addition to technological advances, increased performance of MDGC can be obtained by optimisation of existing systems. By introducing an off-line pre-separation step, the complexity of certain EOs can be decreased, which leads to increased resolution in subsequent MDGC analysis. This can be applied in EO characterisation and also for preparative purposes. In the past years, only LC was used as a pre-separation step in EOs analysis. In the future, it can be expected that other extraction techniques will be connected with the MDGC system to determine some compounds present at trace levels, such as pesticides, allergens, sensitizers, etc. As already established, for the most efficient separations, the stationary phases in the two dimensions must have different selectivity, while the dimensions of the columns must be optimised according to the complexity of the EO. Fast separations can be performed in a system containing a shorter column, and hyphenated with advanced MS systems such as QqQ or HR-TOFMS, which can perform both targeted and untargeted qualitative and quantitative characterisation and spectral deconvolution of the overlapped peaks. Thus, application of high-performance detectors can improve the quality of chromatograms, and increase characterisation capabilities of the whole MDGC system in EO analysis. In addition, modulation period has a great impact on the quality of separation and must be very well optimised. Due to the complexity of the optimisation process in MDGC, in the future it is expected that experimental design will become more important in the optimisation process.
References
Rubiolo P, Sgorbini B, Liberto E, Cordero C, Bicchi C (2010) Essential oils and volatiles: sample preparation and analysis. A review. Flavour Fragr J 25:282–290. https://doi.org/10.1002/ffj.1984
Seeley JV (2012) Recent advances in flow-controlled multidimensional gas chromatography. J Chromatogr A 1255:24–37. https://doi.org/10.1016/j.chroma.2012.01.027
Tranchida PQ, Sciarrone D, Dugo P, Mondello L (2012) Heart-cutting multidimensional gas chromatography: a review of recent evolution, applications, and future prospects. Anal Chim Acta 716:66–75. https://doi.org/10.1016/j.aca.2011.12.015
Marriott PJ, Chin S-T, Maikhunthod B, Schmarr H-G, Bieri S (2012) Multidimensional gas chromatography. Trends Anal Chem 34:1–21. https://doi.org/10.1016/j.trac.2011.10.013
dos Santos AL, Polidoro AdS, Schneider JK, da Cunha ME, Saucier C, Jacques RA, Cardoso CAL, Mota JS, Caramão EB (2015) Comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry (GC × GC/TOFMS) for the analysis of volatile compounds in Piper regnellii (Miq.) C. DC. essential oils. Microchem J 118:242–251. https://doi.org/10.1016/j.microc.2014.07.007
Adahchour M, Beens J, Brinkman UA (2008) Recent developments in the application of comprehensive two-dimensional gas chromatography. J Chromatogr A 1186:67–108. https://doi.org/10.1016/j.chroma.2008.01.002
Bicchi C, Cagliero C, Rubiolo P (2011) New trends in the analysis of the volatile fraction of matrices of vegetable origin: a short overview. A review. Flavour Fragr J 26:321–325. https://doi.org/10.1002/ffj.2059
Bicchi C, Liberto E, Matteodo M, Sgorbini B, Mondello L, Zellner BdA, Costa R, Rubiolo P (2008) Quantitative analysis of essential oils: a complex task. Flavour Fragr J 23:382–391. https://doi.org/10.1002/ffj.1905
Chin S-T, Marriott PJ (2015) Review of the role and methodology of high resolution approaches in aroma analysis. Anal Chim Acta 854:1–12. https://doi.org/10.1016/j.aca.2014.06.029
Do TKT, Hadji-Minaglou F, Antoniotti S, Fernandez X (2015) Authenticity of essential oils. Trends Anal Chem 66:146–157. https://doi.org/10.1016/j.trac.2014.10.007
Mostafa A, Edwards M, Gorecki T (2012) Optimization aspects of comprehensive two-dimensional gas chromatography. J Chromatogr A 1255:38–55. https://doi.org/10.1016/j.chroma.2012.02.064
Edwards M, Mostafa A, Gorecki T (2011) Modulation in comprehensive two-dimensional gas chromatography: 20 years of innovation. Anal Bioanal Chem 401:2335–2349. https://doi.org/10.1007/s00216-011-5100-6
Marriott P, Shellie R, Cornwell C (2001) Gas chromatographic technologies for the analysis of essential oils. J Chromatogr A 936:1–22. https://doi.org/10.1016/S0021-9673(01)01314-0
Nolvachai Y, Kulsing C, Marriott PJ (2017) Multidimensional gas chromatography in food analysis. Trends Anal Chem 96:124–137. https://doi.org/10.1016/j.trac.2017.05.001
Marriott PJ, Eyres GT, Dufour JP (2009) Emerging opportunities for flavor analysis through hyphenated gas chromatography. J Agric Food Chem 57:9962–9971. https://doi.org/10.1021/jf9013845
Prebihalo SE, Berrier KL, Freye CE, Bahaghighat HD, Moore NR, Pinkerton DK, Synovec RE (2018) Multidimensional gas chromatography: advances in instrumentation, chemometrics, and applications. Anal Chem 90:505–532. https://doi.org/10.1021/acs.analchem.7b04226
Tranchida PQ, Franchina FA, Dugo P, Mondello L (2016) Comprehensive two-dimensional gas chromatography-mass spectrometry: recent evolution and current trends. Mass Spectrom Rev 35:524–534. https://doi.org/10.1002/mas.21443
Tranchida PQ, Franchina FA, Mondello L (2017) Analysis of essential oils through comprehensive two-dimensional gas chromatography: general utility. Flavour Fragr J 32:218–227. https://doi.org/10.1002/ffj.3383
Davis JM, Giddings JC (1983) Statistical theory of component overlap in multicomponent chromatograms. Anal Chem 55:418–424. https://doi.org/10.1021/ac00254a003
Davis JM, Giddings JC (1985) Statistical method for estimation of number of components from single complex chromatograms: application to experimental chromatograms. Anal Chem 57:2178–2182. https://doi.org/10.1021/ac00289a002
Liu S, Davis JM (2006) Dependence on saturation of average minimum resolution in two-dimensional statistical-overlap theory: peak overlap in saturated two-dimensional separations. J Chromatogr A 1126:244–256. https://doi.org/10.1016/j.chroma.2006.05.064
Chin ST, Marriott PJ (2014) Multidimensional gas chromatography beyond simple volatiles separation. Chem Commun 50:8819–8833. https://doi.org/10.1039/c4cc02018a
Ma C, Wang H, Lu X, Li H, Liu B, Xu G (2007) Analysis of Artemisia annua L. volatile oil by comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry. J Chromatogr A 1150:50–53. https://doi.org/10.1016/j.chroma.2006.08.080
Wang K, Zhu R-z, Qu R-f, Li Z-y (2012) Comprehensive two-dimensional gas chromatography–time-of-flight mass spectrometry for the analysis of volatile components in Neroli essential oil. Mendeleev Commun 22:45–46. https://doi.org/10.1016/j.mencom.2012.01.018
Baharum SN, Bunawan H, Ghani MA, Mustapha WA, Noor NM (2010) Analysis of the chemical composition of the essential oil of Polygonum minus Huds. using two-dimensional gas chromatography-time-of-flight mass spectrometry (GC-TOF MS). Molecules 15:7006–7015. https://doi.org/10.3390/molecules15107006
Pripdeevech P, Chumpolsri W, Suttiarporn P, Wongpornchai S (2010) The chemical composition and antioxidant activities of basil from Thailand using retention indices and comprehensive two-dimensional gas chromatography. J Serb Chem Soc 75:1503–1513. https://doi.org/10.2298/jsc100203125p
Pripdeevech P, Wongpornchai S, Marriott PJ (2010) Comprehensive two-dimensional gas chromatography-mass spectrometry analysis of volatile constituents in Thai vetiver root oils obtained by using different extraction methods. Phytochem Anal 21:163–173. https://doi.org/10.1002/pca.1173
Brokl M, Fauconnier ML, Benini C, Lognay G, du Jardin P, Focant JF (2013) Improvement of ylang-ylang essential oil characterization by GCxGC-TOFMS. Molecules 18:1783–1797. https://doi.org/10.3390/molecules18021783
Kirkin C, Mitrevski B, Gunes G, Marriott PJ (2014) Essential-oil analysis of irradiated spices by using comprehensive two-dimensional gas chromatography. ChemPlusChem 79:798–803. https://doi.org/10.1002/cplu.201300430
Qiu Y, Lu X, Pang T, Zhu S, Kong H, Xu G (2007) Study of traditional Chinese medicine volatile oils from different geographical origins by comprehensive two-dimensional gas chromatography-time-of-flight mass spectrometry (GC × GC-TOFMS) in combination with multivariate analysis. J Pharm Biomed Anal 43:1721–1727. https://doi.org/10.1016/j.jpba.2007.01.013
Filippi JJ, Belhassen E, Baldovini N, Brevard H, Meierhenrich UJ (2013) Qualitative and quantitative analysis of vetiver essential oils by comprehensive two-dimensional gas chromatography and comprehensive two-dimensional gas chromatography/mass spectrometry. J Chromatogr A 1288:127–148. https://doi.org/10.1016/j.chroma.2013.03.002
Fidelis CHV, Augusto F, Sampaio PTB, Krainovic PM, Barata LES (2012) Chemical characterization of rosewood (Aniba rosaeodora Ducke) leaf essential oil by comprehensive two-dimensional gas chromatography coupled with quadrupole mass spectrometry. J Essent Oil Res 24:245–251. https://doi.org/10.1080/10412905.2012.676770
Fidelis CHV, Sampaio PTB, Krainovic PM, Augusto F, Barata LES (2013) Correlation between maturity of tree and GC × GC-qMS chemical profiles of essential oil from leaves of Aniba rosaeodora Ducke. Microchem J 109:73–77. https://doi.org/10.1016/j.microc.2012.03.034
Santos TG, Fukuda K, Kato MJ, Sartorato A, Duarte MCT, Ruiz ALTG, de Carvalho JE, Augusto F, Marques FA, Sales Maia BHLN (2014) Characterization of the essential oils of two species of Piperaceae by one- and two-dimensional chromatographic techniques with quadrupole mass spectrometric detection. Microchem J 115:113–120. https://doi.org/10.1016/j.microc.2014.02.014
Polidoro AdS, Scapin E, Malmann M, do Carmo JU, Machado ME, Caramão EB, Jacques RA (2016) Characterization of volatile fractions in green mate and mate leaves (Ilex paraguariensis A. St. Hil.) by comprehensive two-dimensional gas chromatography coupled to time-of-flight mass spectrometry (GC × GC/TOFMS). Microchem J 128:118–127. https://doi.org/10.1016/j.microc.2016.03.019
Cai H, Cao G, Zhang HY (2017) Qualitative analysis of a sulfur-fumigated Chinese herbal medicine by comprehensive two-dimensional gas chromatography and high-resolution time of flight mass spectrometry using colorized fuzzy difference data processing. Chin J Integr Med 23:261–269. https://doi.org/10.1007/s11655-015-1966-z
Tranchida PQ, Zoccali M, Franchina FA, Cotroneo A, Dugo P, Mondello L (2013) Gas velocity at the point of re-injection: an additional parameter in comprehensive two-dimensional gas chromatography optimization. J Chromatogr A 1314:216–223. https://doi.org/10.1016/j.chroma.2013.09.025
Cao G, Cai H, Cong X, Liu X, Ma X, Lou Y, Qin K, Cai B (2012) Global detection and analysis of volatile components from sun-dried and sulfur-fumigated herbal medicine by comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry. Analyst 137:3828–3835. https://doi.org/10.1039/c2an35543d
Qin K, Zheng L, Cai H, Cao G, Lou Y, Lu T, Shu Y, Zhou W, Cai B (2013) Characterization of chemical composition of pericarpium Citri reticulatae volatile oil by comprehensive two-dimensional gas chromatography with high-resolution time-of-flight mass spectrometry. Evid Based Complement Alternat Med, 2013: 237541–237552. https://doi.org/10.1155/2013/237541
Saucier C, Polidoro AdS, dos Santos AL, Schneider JK, Caramão EB, Jacques RA (2014) Comprehensive two-dimensional gas chromatography with mass spectrometry applied to the analysis of volatiles in artichoke (Cynara scolymus L.) leaves. Ind Crops Prod 62:507–514. https://doi.org/10.1016/j.indcrop.2014.09.023
Nasim N, Ray A, Singh S, Jena S, Sahoo A, Kar B, Sandeep IS, Mohanty S, Nayak S (2017) Characterization of Kewda volatile components by comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry. Nat Prod Res 31:853–856. https://doi.org/10.1080/14786419.2016.1269099
Ray A, Dash B, Sahoo A, Nasim N, Panda PC, Patnaik J, Ghosh B, Nayak S, Kar B (2017) Assessment of the terpenic composition of Hedychium coronarium oil from Eastern India. Ind Crops Prod 97:49–55. https://doi.org/10.1016/j.indcrop.2016.11.063
von Muhlen C, Zini CA, Caramao EB, Marriott PJ (2008) Comparative study of Eucalyptus dunnii volatile oil composition using retention indices and comprehensive two-dimensional gas chromatography coupled to time-of-flight and quadrupole mass spectrometry. J Chromatogr A 1200:34–42. https://doi.org/10.1016/j.chroma.2008.05.070
Sousa FL, Santos M, Rocha SM, Trindade T (2014) Encapsulation of essential oils in SiO2 microcapsules and release behaviour of volatile compounds. J Microencapsul 31:627–635. https://doi.org/10.3109/02652048.2014.911376
Tranchida PQ, Zoccali M, Bonaccorsi I, Dugo P, Mondello L, Dugo G (2013) The off-line combination of high performance liquid chromatography and comprehensive two-dimensional gas chromatography–mass spectrometry: a powerful approach for highly detailed essential oil analysis. J Chromatogr A 1305:276–284. https://doi.org/10.1016/j.chroma.2013.07.042
Zoccali M, Bonaccorsi IL, Tranchida PQ, Dugo P, Mondello L, Dugo G (2015) Analysis of the sesquiterpene fraction of citrus essential oils by using the off-line combination of high performance liquid chromatography and gas chromatography-based methods: a comparative study. Flavour Fragr J 30:411–422. https://doi.org/10.1002/ffj.3262
Zoccali M, Tranchida PQ, Bonaccorsi IL, Dugo P, Mondello L, Dugo G (2016) Detailed profiling of the volatile oxygenated fraction of mandarin essential oils by using the off-line combination of high-performance liquid chromatography and comprehensive two-dimensional gas chromatography-mass spectrometry. Food Anal Method 10:1106–1116. https://doi.org/10.1007/s12161-016-0668-2
Morrison PD, Marriott PJ, Poynter SDH, Shellie RA (2010) Selection of columns for GCXGC analysis of essential oils. LC GC Eur 23:76–81
Krupcik J, Gorovenko R, Spanik I, Sandra P, Giardina M (2016) Comparison of the performance of forward fill/flush and reverse fill/flush flow modulation in comprehensive two-dimensional gas chromatography. J Chromatogr A 1466:113–128. https://doi.org/10.1016/j.chroma.2016.08.032
Tranchida PQ, Purcaro G, Visco A, Conte L, Dugo P, Dawes P, Mondello L (2011) A flexible loop-type flow modulator for comprehensive two-dimensional gas chromatography. J Chromatogr A 1218:3140–3145. https://doi.org/10.1016/j.chroma.2010.11.082
Franchina FA, Maimone M, Tranchida PQ, Mondello L (2016) Flow modulation comprehensive two-dimensional gas chromatography-mass spectrometry using approximately 4 mL min(-1) gas flows. J Chromatogr A 1441:134–139. https://doi.org/10.1016/j.chroma.2016.02.041
Maikhunthod B, Morrison PD, Small DM, Marriott PJ (2010) Development of a switchable multidimensional/comprehensive two-dimensional gas chromatographic analytical system. J Chromatogr A 1217:1522–1529. https://doi.org/10.1016/j.chroma.2009.12.078
Lidster RT, Hamilton JF, Lewis AC (2011) The application of two total transfer valve modulators for comprehensive two-dimensional gas chromatography of volatile organic compounds. J Sep Sci 34:812–821. https://doi.org/10.1002/jssc.201000710
Kutlular Ö, Özel MZ (2009) Analysis of essential oils of Origanum onites by superheated water extraction using GCxGC-TOF/MS. J Essent Oil Bear Plants 12:462–470. https://doi.org/10.1080/0972060x.2009.10643745
Van der Wat L, Dovey M, Naudé Y, Forbes P (2013) Investigation into the aroma of Rosemary using multi-channel silicone rubber traps, off-line olfactometry and comprehensive two-dimensional gas chromatography–mass spectrometry. S Afr J Chem 66:21–26
Cordero C, Rubiolo P, Cobelli L, Stani G, Miliazza A, Giardina M, Firor R, Bicchi C (2015) Potential of the reversed-inject differential flow modulator for comprehensive two-dimensional gas chromatography in the quantitative profiling and fingerprinting of essential oils of different complexity. J Chromatogr A 1417:79–95. https://doi.org/10.1016/j.chroma.2015.09.027
Sgorbini B, Cagliero C, Boggia L, Liberto E, Reichenbach SE, Rubiolo P, Cordero C, Bicchi C (2015) Parallel dual secondary-column-dual detection comprehensive two-dimensional gas chromatography: a flexible and reliable analytical tool for essential oils quantitative profiling. Flavour Fragr J 30:366–380. https://doi.org/10.1002/ffj.3255
Cao G, Cai H, Jiang J, Yao L, Tu S, Wang L, Ma X, Cai B (2014) Chemical differentiation of volatile compounds in crude and processed Atractylodis Macrocephalae Rhizoma by using comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry combined with multivariate data analysis. J Sep Sci 37:1194–1198. https://doi.org/10.1002/jssc.201301376
Nsuala BN, Kamatou GP, Sandasi M, Enslin G, Viljoen A (2017) Variation in essential oil composition of Leonotis leonurus, an important medicinal plant in South Africa. Biochem Syst Ecol 70:155–161. https://doi.org/10.1016/j.bse.2016.11.009
Schaffer M, Groger T, Putz M, Zimmermann R (2013) Forensic profiling of sassafras oils based on comprehensive two-dimensional gas chromatography. Forensic Sci Int 229:108–115. https://doi.org/10.1016/j.forsciint.2013.03.046
de Godoy LA, Hantao LW, Pedroso MP, Poppi RJ, Augusto F (2011) Quantitative analysis of essential oils in perfume using multivariate curve resolution combined with comprehensive two-dimensional gas chromatography. Anal Chim Acta 699:120–125. https://doi.org/10.1016/j.aca.2011.05.003
He M, Yang ZY, Yang TB, Ye Y, Nie J, Hu Y, Yan P (2017) Chemometrics-enhanced one-dimensional/comprehensive two-dimensional gas chromatographic analysis for bioactive terpenoids and phthalides in Chaihu Shugan San essential oils. J Chromatogr B 1052:158–168. https://doi.org/10.1016/j.jchromb.2017.03.029
Omar J, Alonso I, Olivares M, Vallejo A, Etxebarria N (2012) Optimization of comprehensive two-dimensional gas-chromatography (GC × GC) mass spectrometry for the determination of essential oils. Talanta 88:145–151. https://doi.org/10.1016/j.talanta.2011.10.023
Nicolotti L, Cordero C, Bressanello D, Cagliero C, Liberto E, Magagna F, Rubiolo P, Sgorbini B, Bicchi C (2014) Parallel dual secondary column-dual detection: a further way of enhancing the informative potential of two-dimensional comprehensive gas chromatography. J Chromatogr A 1360:264–274. https://doi.org/10.1016/j.chroma.2014.07.081
Savareear B, Jacobs MR, Shellie RA (2014) Multiplexed dual first-dimension comprehensive two-dimensional gas chromatography–mass spectrometry with contra-directional thermal modulation. J Chromatogr A 1365:183–190. https://doi.org/10.1016/j.chroma.2014.09.014
Tranchida PQ, Franchina FA, Zoccali M, Panto S, Sciarrone D, Dugo P, Mondello L (2013) Untargeted and targeted comprehensive two-dimensional GC analysis using a novel unified high-speed triple quadrupole mass spectrometer. J Chromatogr A 1278:153–159. https://doi.org/10.1016/j.chroma.2012.12.066
Qiu Y, Lu X, Pang T, Ma C, Li X, Xu G (2008) Determination of radix ginseng volatile oils at different ages by comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry. J Sep Sci 31:3451–3457. https://doi.org/10.1002/jssc.200800253
Cai H, Cao G, Li L, Liu X, Ma XQ, Tu SC, Lou YJ, Qin KM, Li SL, Cai BC (2013) Profiling and characterization of volatile components from non-fumigated and sulfur-fumigated Flos Lonicerae Japonicae using comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry coupled with chemical group separation. Molecules 18:1368–1382. https://doi.org/10.3390/molecules18021368
Costa R, Ragusa S, Russo M, Certo G, Franchina FA, Zanotto A, Grasso E, Mondello L, Germano MP (2016) Phytochemical screening of Artemisia arborescens L. by means of advanced chromatographic techniques for identification of health-promoting compounds. J Pharm Biomed Anal 117:499–509. https://doi.org/10.1016/j.jpba.2015.10.006
Krupcik J, Gorovenko R, Spanik I, Sandra P, Armstrong DW (2015) Enantioselective comprehensive two-dimensional gas chromatography. A route to elucidate the authenticity and origin of Rosa damascena Miller essential oils. J Sep Sci 38:3397–3403. https://doi.org/10.1002/jssc.201500744
Krupcik J, Gorovenko R, Spanik I, Armstrong DW, Sandra P (2016) Enantioselective comprehensive two-dimensional gas chromatography of lavender essential oil. J Sep Sci 39:4765–4772. https://doi.org/10.1002/jssc.201600986
Harvey PM, Poynter SDH, Shellie RA (2011) GC × GC with fluidic modulation for enantioselective essential oil analysis. LC GC Eur 24:548–555
Schmarr H-G, Engel K-H (2012) Analysis and stereodifferentiation of linalool in Theobroma cacao and cocoa products using enantioselective multidimensional gas chromatography. Eur Food Res Technol 235:827–834. https://doi.org/10.1007/s00217-012-1812-x
Lo Presti M, Sciarrone D, Crupi ML, Costa R, Ragusa S, Dugo G, Mondello L (2008) Evaluation of the volatile and chiral composition in Pistacia lentiscus L. essential oil. Flavour Fragr J 23:249–257. https://doi.org/10.1002/ffj
Namara KM, Howell J, Huang Y, Robbat A Jr (2007) Analysis of gin essential oil mixtures by multidimensional and one-dimensional gas chromatography/mass spectrometry with spectral deconvolution. J Chromatogr A 1164:281–290. https://doi.org/10.1016/j.chroma.2007.07.042
Mobin M, De Lima SG, Almeida LTG, Takahashi JP, Teles JB, Szeszs MW, Martins MA, Carvalho AA, Melhem MSC (2016) MDGC-MS analysis of essential oils from Protium heptaphyllum (Aubl.) and their antifungal activity against Candida specie. Rev Bras Plant Med 18:531–538. https://doi.org/10.1590/1983-084X/15_110
Yang SO, Kim Y, Kim HS, Hyun SH, Kim SH, Choi HK, Marriott PJ (2011) Rapid sequential separation of essential oil compounds using continuous heart-cut multi-dimensional gas chromatography–mass spectrometry. J Chromatogr A 1218:2626–2634. https://doi.org/10.1016/j.chroma.2011.02.060
Chin ST, Maikhunthod B, Marriott PJ (2011) Universal method for online enrichment of target compounds in capillary gas chromatography using in-oven cryotrapping. Anal Chem 83:6485–6492. https://doi.org/10.1021/ac200973z
Wong YF, West RN, Chin ST, Marriott PJ (2015) Evaluation of fast enantioselective multidimensional gas chromatography methods for monoterpenic compounds: authenticity control of Australian tea tree oil. J Chromatogr A 1406:307–315. https://doi.org/10.1016/j.chroma.2015.06.036
Sciarrone D, Ragonese C, Carnovale C, Piperno A, Dugo P, Dugo G, Mondello L (2010) Evaluation of tea tree oil quality and ascaridole: a deep study by means of chiral and multi heart-cuts multidimensional gas chromatography system coupled to mass spectrometry detection. J Chromatogr A 1217:6422–6427. https://doi.org/10.1016/j.chroma.2010.08.019
Sciarrone D, Costa R, Ragonese C, Tranchida PQ, Tedone L, Santi L, Dugo P, Dugo G, Mondello L (2011) Application of a multidimensional gas chromatography system with simultaneous mass spectrometric and flame ionization detection to the analysis of sandalwood oil. J Chromatogr A 1218:137–142. https://doi.org/10.1016/j.chroma.2010.10.117
Eyres G, Marriott PJ, Dufour JP (2007) The combination of gas chromatography-olfactometry and multidimensional gas chromatography for the characterisation of essential oils. J Chromatogr A 1150:70–77. https://doi.org/10.1016/j.chroma.2006.07.019
Sasamoto K, Ochiai N (2010) Selectable one-dimensional or two-dimensional gas chromatography-mass spectrometry with simultaneous olfactometry or element-specific detection. J Chromatogr A 1217:2903–2910. https://doi.org/10.1016/j.chroma.2010.02.045
Bonaccorsi I, Sciarrone D, Schipilliti L, Dugo P, Mondello L, Dugo G (2012) Multidimensional enantio gas chromatography/mass spectrometry and gas chromatography-combustion-isotopic ratio mass spectrometry for the authenticity assessment of lime essential oils (C. aurantifolia Swingle and C. latifolia Tanaka). J Chromatogr A 1226:87–95. https://doi.org/10.1016/j.chroma.2011.10.038
Barba C, Toledano RM, Santa-Maria G, Herraiz M, Martinez RM (2013) Enantiomeric analysis of limonene and carvone by direct introduction of aromatic plants into multidimensional gas chromatography. Talanta 106:97–103. https://doi.org/10.1016/j.talanta.2012.11.050
Greule M, Hänsel C, Bauermann U, Mosandl A (2007) Feed additives: authenticity assessment using multicomponent-/multielement-isotope ratio mass spectrometry. Eur Food Res Technol 227:767–776. https://doi.org/10.1007/s00217-007-0785-7
Mateus E, Barata RC, Zrostlíková J, Gomes da Silva MGR, Paiva MR (2010) Characterization of the volatile fraction emitted by Pinus spp. By one- and two-dimensional chromatographic techniques with mass spectrometric detection. J Chromatogr A 1217:1845–1855. https://doi.org/10.1016/j.chroma.2010.01.045
Babushok VI, Zenkevich IG (2008) Retention indices for most frequently reported essential oil compounds in GC. Chromatographia 69:257–269. https://doi.org/10.1365/s10337-008-0872-3
Robbat A Jr, Kowalsick A, Howell J (2011) Tracking juniper berry content in oils and distillates by spectral deconvolution of gas chromatography/mass spectrometry data. J Chromatogr A 1218:5531–5541. https://doi.org/10.1016/j.chroma.2011.06.053
Agilent Technologies Inc (2018) https://www.agilent.com/en/products/gas-chromatography/gc-gc-ms-technologies/capillary-flow-technology. Accessed 2 Oct 2018
Perkin-Elmer (2018) https://www.perkinelmer.com/lab-solutions/resources/docs/TCH_SwaferInletSwitching.pdf. Accessed 2 Oct 2018
SGE Analytical Science (2018) http://www.sge.com/products/silflow-stainless-steel-micro-fluidic-platform. Accessed 2 Oct 2018
Shimadzu Corp (2018) https://www.shimadzu.com/an/gc/multidimgc/multidimmdgc.html. Accessed 2 Oct 2018
Gerstel GmbH & Co. KG (2018) http://www.gerstel.com/en/multidimensional-gc-mcs.htm. Accessed 2 Oct 2018
Marriott PJ, Kinghorn RM (1997) Longitudinally modulated cryogenic system. A generally applicable approach to solute trapping and mobilization in gas chromatography. Anal Chem 69:2582–2588. https://doi.org/10.1021/ac961310w
Nolvachai Y, Kulsing C, Marriott PJ (2015) Pesticides analysis: advantages of increased dimensionality in gas chromatography and mass spectrometry. Crit Rev Environ Sci Technol 45:2135–2173. https://doi.org/10.1080/10643389.2015.1010431
Lebanov L, Tedone L, Kaykhaii M, Linford MR, Paull B (2018) Multidimensional gas chromatography in essential oil analysis Part 2. Application to characterisation and identification (submitted to Chromatographia)
Park HE, Yang SO, Hyun SH, Park SJ, Choi HK, Marriott PJ (2012) Simple preparative gas chromatographic method for isolation of menthol and menthone from peppermint oil, with quantitative GC–MS and (1) H NMR assay. J Sep Sci 35:416–423. https://doi.org/10.1002/jssc.201100670
Eyres GT, Urban S, Morrison PD, Dufour JP, Marriott PJ (2008) Method for small-molecule discovery based on microscale-preparative multidimensional gas chromatography isolation with nuclear magnetic resonance spectroscopy. Anal Chem 80:6293–6299. https://doi.org/10.1021/ac8007847
Eyres GT, Urban S, Morrison PD, Marriott PJ (2008) Application of microscale-preparative multidimensional gas chromatography with nuclear magnetic resonance spectroscopy for identification of pure methylnaphthalenes from crude oils. J Chromatogr A 1215:168–176. https://doi.org/10.1016/j.chroma.2008.10.102
Sciarrone D, Panto S, Ragonese C, Tranchida PQ, Dugo P, Mondello L (2012) Increasing the isolated quantities and purities of volatile compounds by using a triple Deans-switch multidimensional preparative gas chromatographic system with an apolar-wax-ionic liquid stationary-phase combination. Anal Chem 84:7092–7098. https://doi.org/10.1021/ac3013829
Ruhle C, Eyres GT, Urban S, Dufour JP, Morrison PD, Marriott PJ (2009) Multiple component isolation in preparative multidimensional gas chromatography with characterisation by mass spectrometry and nuclear magnetic resonance spectroscopy. J Chromatogr A 1216:5740–5747. https://doi.org/10.1016/j.chroma.2009.06.006
Sciarrone D, Panto S, Tranchida PQ, Dugo P, Mondello L (2014) Rapid isolation of high solute amounts using an online four-dimensional preparative system: normal phase-liquid chromatography coupled to methyl siloxane-ionic liquid-wax phase gas chromatography. Anal Chem 86:4295–4301. https://doi.org/10.1021/ac404078u
Panto S, Sciarrone D, Maimone M, Ragonese C, Giofre S, Donato P, Farnetti S, Mondello L (2015) Performance evaluation of a versatile multidimensional chromatographic preparative system based on three-dimensional gas chromatography and liquid chromatography-two-dimensional gas chromatography for the collection of volatile constituents. J Chromatogr A 1417:96–103. https://doi.org/10.1016/j.chroma.2015.09.039
Sciarrone D, Panto S, Donato P, Mondello L (2016) Improving the productivity of a multidimensional chromatographic preparative system by collecting pure chemicals after each of three chromatographic dimensions. J Chromatogr A 1475:80–85. https://doi.org/10.1016/j.chroma.2016.11.013
Acknowledgements
Authors would like to thank University of Tasmania, Australia and Plant Therapy Inc., Twin Falls, Idaho, USA, for their generous provision of a jointly funded PhD scholarship and project support for Leo Lebanov.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Authors declare that they have no conflict of interest.
Ethical approval
This article does not contain any studies with human participants or animals performed by any of the authors.
Additional information
Published in the topical collection Rising Stars in Separation Science, as part of Chromatographia’s 50th Anniversary Commemorative Issue.