Introduction

Flavonoids are widely distributed in fruits and vegetables. They have great antioxidant capacity and other various health benefits [1,2,3]. Rutin (3,3 4′5,7-pentahydrohyflavone-3-rhamnoglucoside, RU) is a well-known radical scavenger and strong antioxidant [4]. It is also known as vitamin P. Due to its good biological and physiological properties, it has raised considerable attention in food, clinical drug and cosmetic industries, etc. [5]. Therefore, it is crucial to build a simple approach for detection of rutin content in pharmaceuticals and various real samples. Up to know, there are several methods to detect rutin, such as high performance liquid chromatography (HPLC) [6], chemiluminescence [7], capillary electrophoresis [8], UV-vis spectrophotometers [9], flow injection analysis (FIA) [10], sequential injection analysis [11] and electrochemical method [12]. Among these methods, electrochemical method is regarded as a good applicable, on account of convenient, save time, low cost and detection limy and good sensitivity [13]. Furthermore, rutin is an electro-active molecule, which means that it can be easily oxidized and reduced on surface of working electrodes. The determination of rutin by voltammetric methods has become more hot spots.

Wang et al. [14] have described a new method for electrochemical detection of rutin. A nanocomposite was synthesized by a reduced graphene oxide (RGO) as a template for immobilizing zeolitic imidazolate framework-8 (ZIF-8) crystals loaded with the electrochemical probe Methylene Blue (MB), under the optimal conditions, the sensor has an excellent amperometric response towards rutin, with a 20 nM detection limit (at an S/N ratio of 3). Liu et al. [15] describe a method for electrochemical synthesis of a nanocomposite consisting of carboxylated multi-walled carbon nanotubes, polythionine and platinum nanoparticles (denoted as cMWCNT@pTh@Pt). The composite was placed on a glassy carbon electrode (GCE) to obtain a sensor for simultaneous determination of myricetin and rutin and successfully applied to the simultaneous determination of myricetin and rutin in spiked juice samples, and satisfactory results were obtained.

Carbon materials have high chemical, thermal and structural stability with good biocompatibility. It’s widely used in biomedicine fields, redox catalyst, ferrofluid, and wave shielding materials [16, 17]. In order to improve the electrochemical property of carbon materials, it is crucial to synthesize more efficient nanocomposites. Nowadays, carbon-based nanomaterials encapsulated with various kinds of metal nanocrystals are new kind of carbon/metal nanocomposite. Among them, magnesium-aluminum alloys and silicon were more functional material, has a number of unique property, such as excellent magnetic, catalytic properties and superior machinability [18, 19]. The combination of metal alloys and carbon materials has the high adsorption capacities and environmental friendliness [20].

Taraxacum-like Mg-Al-Si@PC nanocluster was successfully synthesed by simple solvo-thermal reaction and followed by the high-temperature annealing method. Afterwards, the glass carbon electrode modifies with Mg-Al-Si@PC nanoclusters and applies for electrochemical detection of rutin. The electrochemical sensor display good electro catalytic activity and sensing property towards rutin with the satisfactory linear range from 1 μM up to 10 μM, with low detection limit of 0.01 μM, good stability and interference ability, which is favorable for practical application of fast quantification of rutin.

Experimental

Apparatus

A CHI-760E electrochemical workstation (Shanghai Chenhua Instrument Co. LTD, China) equipped with a three-electrode system was used. Platinum wire as counter electrode, saturated calomel electrode (SCE) as reference electrode and Mg-Al-Si@PC modified galassy carbon electrode set as working electrode, respectively. Phosphate buffer (PB, 0.1 M, pH = 2.0) was used as electrolyte solution for the rutin detection. In the cyclic voltammetry (CV) test the working potential set as 0.2–0.9 V (vs. ref. electrode) and scan rate was 100 mV s −1; differential pulse voltammetry (DPV) performed scan range was from 0.2 V to 0.9 V, and the pulse width and pulse period were 0.05 s and 0.5 s, respectively, and the scanning rate was 100 mV s −1. In all experiment, the reported results for every electrode were the mean of three parallel measurements.

Scanning electron microscope (SEM, Carl Zeiss, Germany), transmission electron microscope (TEM, Titan G2 60–300) were apply to characterization of the micro zone structure of the synthesis nanocomposite. All experiments were carry on the under pure nitrogen bubbling for the aim to remove the interference of dissolved oxygen.

Reagents and materials

The experiment chemicals were analytical grade without any treatment. Glyme, SiCl4, Mg-Al alloy powder were purchase from Aldrich, Aladdin Chemicals, 5% Nafion stock solution purchase from Sigma-Aldrich. Rutin was supply by MACKLIN (Shanghai). All aqueous electrochemical testing electrolyte agents were prepare by using deionizer water (≥18.2 MΩ cm). Phosphate buffer with different pH values (from 2.0 to 4.0) were prepared by mixing 0.1 KCl, 0.1 M KH2PO4 and 0.1 M K2HPO4 and adjusting the pH with HCl or NaOH.

Synthesis of mg-Al-Si@PC nanocluster

Mg-Al-Si@PC nanoclusters were synthesis through the one-pot solvo-thermal reaction and then followed with high-temperature annealing method [7, 21, 22], which was as shown in Scheme 1. The solvo-thermal reaction was performed mixing the a certain amount of Mg-Al alloy powder (0.5–1.0 g) and SiCl4 (2–3 ml) in N,N-dimethyl formamide (DMF, 30 ml) in a stainless autoclave with teflon liner, then at 170 °C reaction for 10 h. After cooling to room temperature, the product was immersed in ethanol and kept stirring for 15 h, and then centrifuged (5000 r min−1, 10 min), collected by dried at 90 °C in vacuum, for 8 h. Afterwards, the composite was treated with annealing at 800 °C for 2 h, then fast cooling 3 min under Ar flow. The quick cooling of the product may maintain the morphological structure of product. After the product ultrasonic treatment in aqueous solution for 30 min and filtering, subsequently washed for three times with water and ethanol, respectively, then was dried at 80 °C in vacuum, for 10 h. The particles become highly fluffy [23], due to the removal of MgCl2. Moreover, it is not a necessary to use strong acid for the post-treatments, which alleviate serious environmental concerns about the disposal of these acids. The safe and cheap reagents, conventional equipment, and simple operation to prepare porous structured nanoclusters make the synthesis attractive [24, 25].

Scheme 1
scheme 1

Schematic presentation of synthesis process of Mg-Al-Si@PC nonoclusters and electrochemical detection of rutin

Fabrication of working electrode

GCE was usually polished with aluminum slurries (1.0, 0.3 and 0.05 μm), for the purpose to get the mirror-like fresh surface. Sequentially, the GCE wash with (ethanol and deionizer water) under ultrasonication for 30 s to remove the aluminum oxide attached at the GCE surface. Finally, the clean GCE was dry with nitrogen flowing in air. Mg-Al-Si@PC/GCE was prepared by casting 5 μL of the synthesis Mg-Al-Si@PC suspension on the GCE surface and then drying at room temperature.

Results and discussion

Choice of material

Carbon nanomaterials a promising material, have gained widespread attention due to their unique structural properties. Among them, porous carbons offer numerous remarkable properties, such as high specific surface area, high conductivity, chemical stability and tunable surface chemistry. That allows the development of a range of strategies for the introduction of functional groups, and consequently, for the anchoring of metal and metal complexes. The synthesis of Mg-Al-Si@PC is simple, low costs and the precursor of nanoclusters Mg-Al alloy and SiCl4 were common chemicals, easy to get, also excellent material for synthesis of functionalized nanocomposite. In addition Mg-Al-Si@PC has the special morphological structure of taraxacum like porous structure. Therefore Mg-Al-Si@PC used as electrode material for electrochemical detection of rutin.

Characterization of Mg-Al-Si@PC nanoclusters

The morphological and structural properties of Mg-Al-Si@PC were characterized by SEM and TEM. SEM image of the Mg-Al-Si@PC confirms that taraxacum like cluster structure with the numerous holes was investigated in Fig. 1a. The further TEM image of the nanocomposite and the related high-angle annular dark field (HAADF) image (Fig. 1b and c) undoubtedly reveal the taraxacum like cluster structure of this material at same. The EELS mapping of the select area in Fig. 1d-1h illustrates that the Mg-Al-Si@PC was consist from aluminum, silicon, magnesium, carbon and oxygen elements. These elements were homogenously distribute in nanoclusters, as shown in Fig. 1 (i) displays the overlay signals in the element map texture.

Fig. 1
figure 1

SEM image of Mg-Al-Si@PC (a); TEM images of Mg-Al-Si@PC (b), (c) high-angle annular dark-field scanning transmission electron microscopy (HAADF-STEM) image of Mg-Al-Si@PC and (d–i) electronic emitter location system EELS map of aluminum, silicon, magnesium, carbon and oxygen

Electrochemical behavior of modified electrodes

CV of the ferricyanide system is a useful tool to characterize the electrochemical behaviors of the electroactive species on the modified electrode surface. Figure 2 shows a comparison of the electroactivity among different modified electrodes investigated by CV in 0.1 M KCl solution containing 5 mM K3Fe(CN) 6, at the scan rate 100 mV s−1. For bare GCE (curve a), a couple of well-defined redox peaks are observed with pea-to-peak separation (ΔEp) of 127 mV. For Mg-Al-Si@PC /GCE (curve b), peak currents increases and the ΔEp value decreases to 106 mV. The result indicate that the surface property of the Mg-Al-Si@PC /GCE has been significantly changed. This confirms elctrocatalytic activity of the Mg-Al-Si@PC /GCE. The available surface are for reaction of species in solution can be calculated by the Randles-Sevcik eq. (1) [26, 27]:

$$ {\mathbf{I}}_{\mathbf{p}}=\mathbf{2.69}\times {\mathbf{1}\mathbf{0}}^{\mathbf{5}}{\mathbf{n}}^{\mathbf{3}/\mathbf{2}}\mathbf{A}\ {\mathbf{D}}^{\mathbf{1}/\mathbf{2}}{\mathbf{C}}_{\mathbf{0}}\kern0ex {nu}^{\mathbf{1}/\mathbf{2}} $$
(1)
Fig. 2
figure 2

CV curves of bare GCE (a) and Mg-Al-Si@PC/GCE (b) in 0.1M KCl solution containing 1 mM K3Fe(CN)6. Scan rate = 100 mV s−1

Fig. 3
figure 3

A: DPV curves of Mg-Al-Si@PC/GCE with different rutin concentrations (0, 1,2, 3, 4, 5, 6, 7, 8, 9, 10 μM) in phosphate buffer of pH = 2.0, Scan rate = 100 mV s−1. a: The plots of the oxidation currents versus the rutin concentration

Where, Ip is the peak current in A. C0 is the concentration of the electroactive species (mol cm−3), n is the number of electrons involved, D is the diffusion coefficient (cm2 S−1), υ is the scan rate (V s−1) and A is the electroactive surface area (cm2). For Mg-Al-Si@PC /GCE the electroactive surface area is maximum (0.028cm2) as compared with bare GCE (0.015cm2).

Electrochemical determination of rutin by using different electrodes

The electrochemical behavior of the prepared material towards rutin was surveyed by voltammetric method such as cyclic voltammetry (CV) in phosphate buffer of pH = 2.0 containing 10 μM rutin. Electrochemical behavior of rutin at bare GCE and Mg-Al-Si@PC modified electrode is displayed in Fig. S1 (in Electronic Supporting Material). CV scan survey showed that bare GCE performed a pair of weak redox peaks (curve a), the redox peak potential appear at 562mV (Epa) and 543 mV (Epc), (vs.SCE). The redox peak currents were located as 4.03 μA and 3.122μA for Ipa and Ipc respectively, the ratio of Ipa/Ipc was calculated to be 1.32. While for the Mg-Al-Si@PC modified electrode, pair of visible redox peaks was observed, showing the sensor has better electrochemical response towards rutin. The anodic peak potential and cathodic peak potential was 576 mV and 531 mV, respectively (vs. SCE). The Ipa was 20.52 μA and the Ipc was 16.37 μA, which were 5 fold higher than unmodified GCE. The ratio of redox peak current (Ipa/Ipc) was calculated to be 1.25, cathode branch and the anode branch were nearly coincident, indicating more reversible electrochemical redox process of rutin on the modified electrode [9]. The results indicate that, the synthesis Mg-Al-Si@PC nanoclusters can act as a more effective medium for promoting electrochemical redox of rutin. Therefore the Mg-Al-Si@PC modified electrode shows better electrochemical sensing property towards rutin determination.

Electrochemical parameters of rutin at Mg-Al-Si@PC/GCE

It’s well-known that the scanning rate affects the oxidation peak current of rutin at Mg-Al-Si@PC/GCE. So, the effect of scan rates on rutin redox currents were firstly investigated with Mg-Al-Si@PC/GCE by CV technique. It can be seen from the Fig. S2 (in Electronic Supporting Material), the peak current increased gradually by increasing of scan rates, and then peak currents are proportional with the scan rate of potential range. This indicates that the electrode process of rutin at Mg-Al-Si@PC/GCE surface are a adsorption-controlled process [28], this attributed to the large surface area of the modified electrode. The linear regression equations are expressed as Ipa(μA) = 0.164V(mV s−1) − 1.923, (R = 0.999) and Ipc(μA) = −0.134V(mV s−1) + 1.758 (R = 0.999). With the increasing of scan rate, the oxidation peak potential of rutin positively shifts while the reduction peak potential negatively shifts, resulting in the increasing of ∆Ep value. The relationship of the redox peak potentials with lnʋ were further obtained with two linear regression equations as Epa (V) = 0.016lnv (mV s−1) + 0.506 (R = 0.996) and Epc(V) = −0.012lnʋ (mV s−1) + 0.583 (R = 0.992). Based on Laviron’s equation, the slope of the line for Epa and Epc can be expressed as RT/ (1-a)nF and -RT/anF, respectively. The specific calculation formulas as follows (2–4):

$$ Epa={E}^{0\prime }+\frac{\mathrm{RT}}{\left(1-\mathrm{a}\right)\mathrm{nF}}\ln \nu $$
(2)
$$ {\mathrm{E}}_{\mathrm{pc}}={E}^{0\prime }-\frac{\mathrm{RT}}{\mathrm{anF}}\ln \nu $$
(3)
$$ \nu \upsilon {\mathrm{lgk}}_{\mathrm{s}}= alg\left(1-a\right)+\left(1-\mathrm{a}\right)\mathrm{lga}-\lg \frac{\mathrm{RT}}{\mathrm{nF}\nu }-\mathrm{a}\left(1-\mathrm{a}\right)\frac{\mathrm{nF}\Delta {\mathrm{E}}_{\mathrm{p}}}{2.303\mathrm{RT}} $$
(4)

Where (a) is electron-transfer coefficient, (n) is number of electron transfer, (υ) is scan rate, (k) is apparent heterogeneous electron transfer rate constant, (∆Ep) is peak-to-peak potential separation, (F) is Faraday’s constant, (R) and (T) represent their usual meaning. According to linear regression equations from lnʋ and Ep, (n) and (a) were calculated to be 2 and 0.57, respectively, based on eq. (4), when the scan rate was 100 mV s−1, the (ks) was calculated to be 1.56. It can be concluded that a 2e transfer were involved in electrochemical reaction [29].

The effect of the acidity of the buffer solution on peak current of rutin also tested by differential pulse voltammetry (DPV) in phosphate buffer containing 10 μM rutin, as shown in Fig. S3 (in Electronic Supporting Material). When increasing pH value from 2.0 to 4.0, the oxidation peak current of rutin decrease and peak potentials negatively shifted, indicating that protons take part in reaction [30]. The linear regression equation of oxidation peak potential was obtained as Epa = −0.059pH + 0.681 (R = 0.991) a slope of −59 mVpH−1 was the same as the Nernst system of −59 mV/pH at 25 °C, demonstrate that an equal number of protons and electrons took part in the electrochemical redox process [31] . At pH 2.0 buffer has a high concentration of protons, this can be increases the electrode reaction. In addition, maximum peak current was appear at pH 2.0, and then was decrease with further increase of pH value, considering that the general situation pH 2.0 was select in the test. It can be inferred that the redox mechanism of rutin at the Mg-Al-Si@PC/GCE is illustrated in Fig. S4.

The effect of the accumulation time towards detection of rutin was shown in Fig. S5 (in Electronic Supporting Material), as can be seen from Fig. S5 the redox peak currents of the rutin increased rapidly with the increase of the incubation time in beginning and reached maximum at 180 s, indicating that the adsorption equilibrium is reached. Therefore, 180 s was chosen as optimal time for rutin analysis [32].

DPV determination of rutin

The electrochemical sensing performance of the Mg-Al-Si@PC/GCE towards rutin was carried out by DPV under optimal conditions. As shown in Fig. 3, the oxidation peak current (Ip, μA) of rutin increased with increasing rutin concentration and the Ipa was linearly dependent to the rutin concentration (c, μM) from 1.0 μM to 10 μM. The linear regression equations was described as follows: Ip (μA) = 4.212C (μM)-1.228 with a linear correlation coefficient of 0.991. Based on the signal-to-noise ratio of 3 (S/N = 3), the detection limit is calculated to be 0.01 μΜ.

Table 1 summarizes the comparison of the electrochemical performance of the Mg-Al-Si@PC/GCE with other reported electrochemical methods for rutin determination. The modified electrode show a reasonable linear range and the detection limit was lower than or equal to most of the previous reports, which can be attributed to the current nanomaterial with highly porous structure, high surface area and good electro-conductivity. In addition, the fabrication of our nanomaterial sensor was more controllable and convenient through the strategy of electrochemical synthesis.

Table 1 Comparison of analytical parameters of different modified electrodes for rutin analysis

Interferences study

To investigate the interference performance of the proposed electrochemical sensor, selectivity study was investigated by introducing other interfering compounds while the rutin concentration was fixed to 10 μM. The limit of potential interfering substances was defined as ±5% relative error in the determination of rutin. The results were showed in Table S1 (in Electronic Supporting Material). The result indicates that various potentially interfering ions have tiny influences to the signals of the rutin (signal change less than 5%, n = 3). In addition, in the electrochemical test, appear the response currents for small drug molecules (ascorbic acid, glucose, dopamine, uric acid) and luteolin, indicate that the analytical method had good selectivity and can be applied for the electrochemical detection of some other flavonoids and small drug molecules.

Reproducibility and stability of the modified electrode

The reproducibility and stability of the modified electrode are the important characteristics for the future application. The reproducibility of the electrochemical sensor was evaluated by detecting rutin through CV method for ten times under the same testing condition. The relative standard deviation (R.S.D) is calculated to be 4.63%, indicating that the electrode is highly reproducible. The stability was investigating through a common process by storing the modified electrode at a 4 °C refrigerator for a week and then used for test its sensing performance towards rutin again; the result revealed that the peak currents of rutin still remained similar to their initial values. Hence, the sensor shows the stability for rutin analysis.

Analytical application

In order to verify the application of the this method, the troxerutin tablet (label amount of 20 mg per tablet, C151104, Yabao Pharmaceutical Group Co. Ltd., China) was detected with the optimal experimental procedure. Before the measurement, a certain number of drug tablets were ground into powder, and then the powder was accurately weigh, subsequently dissolved in ethanol prepared for stock solution. In the test, the appropriate volume of stock solution was diluted with phosphate buffer of pH = 2. It can be seen from the Table S2 (in Electronic Supporting Material), the recovery was calculated as range from 99.8–100.6%, indicate that the standard addition method for rutin detection more visible. In the same, in order to proof the accuracy of the this method, HPLC was apply to the detection of rutin in real samples, the rutin contents were detected for 20.0 ± 0.08 mg per tablet for the troxerutin tablet, which were closed to the result of electrochemical method, so it suggests that Mg-Al-Si@PC/GCE is suitable for the testing for rutin in commercial pharmaceutical samples. However, the current method is not suitable for rutin detection in both body fluids and food samples like juice, owing to thousands of complicated components such as biological macromolecules, small molecules, and high-concentration inorganic salts that coexist in these samples. There for further pretreatments are necessary for these real body fluids and food samples before they can be subjected to rutin detection.

Conclusions

An electrochemical sensor based on the use of Mg-Al-Si@PC nanoclusters is successfully fabricated for the sensitive detection of rutin. The sensor exhibits excellent electro catalytic ability for the sensitive detection of rutin due to the special structure (porous structure) and synergetic effect of the Mg, Al, Si atoms and porous carbon (superior electrical conductivity, good electro catalytic performance and excellent electron transfer mediator). The electrochemical sensor has good analytical performance for rutin with wide linear range, lower LOD, excellent reproducibility selectivity and stability. The modified electrode is further used to analyze rutin in tablet samples with excellent practical abilities. These results indicate that the Mg-Al-Si@PC sensor may have great potential applications in electrochemical sensing.