Introduction

Osteoporotic vertebral compression fracture (VCF) is one of the most common complications of osteoporosis affecting estimated 1.4 million people worldwide annually [1]. Conservative management cannot reverse the kyphotic deformities that cause biomechanical changes in the spinal segment and increase the incidence of adjacent vertebral fractures [1, 2]. Moreover, VCFs result in prolonged hospitalization. The minimally invasive percutaneous vertebral augmentation techniques of vertebroplasty (VP), balloon kyphoplasty (KP) and stentoplasty (SP) comprising the injection of polymethylmethacrylate (PMMA) are effective procedures for immediate postoperative pain alleviation caused by VCF [3]. VP and KP are widely accepted treatments with reported pain relief in approximately 70–100 % [4].

However, augmentation techniques using PMMA must be done with caution. Potential serious complications may occur with the intraosseous injection of cement. The risk of extraosseous cement leakage in various series ranged between 3 and 74 % [58], with resultant neurological deficits such as radiculopathy and cord compression occurring in 0–3.7 and 0–0.5 %, respectively [5, 6, 810]. The risk of pulmonary cement embolism lies between 3.5 and 23 % [1114]. In addition to unwanted cement leakage also pulmonary fat embolism occurs as a result of the displacement of intraosseous bone marrow by the cement resulting in the occlusion of blood vessels downstream, most notably in the lungs. The acute consequences of fat embolism may be hypotension, cardiac arrest, and, in serious cases, sudden death [1517]. Bone marrow particles cause microembolization of the arterioles and capillaries of the lungs resulting in an increase in pulmonary arterial pressure, a systemic arterial hypotension, and a drop in heart rate [17, 18]. Moreover, hypotension can be result of cement monomers and vasoactive mediators gaining access to the circulation [19, 20].

To decrease these types of complication the number of augmented vertebrae should be limited to six levels per session or 25–30 cc of PMMA according to Heini et al. [21], knowing that the cement volume is the most important modifiable predictor for pain relief in KP [22]. This cut-off was empirically determined and means a higher risk for pulmonary problems when injecting larger volumes. On the other hand, Aebli et al. demonstrated in an animal study that already smaller volumes of injected PMMA lead to measurable cardiovascular changes as hypotension drop of cardiac output and pulmonary hypertension that may be fatal in older and/or pulmonary impaired patients [20]. They also showed a cumulative increase of fat embolism in lung biopsies in sheep after multiple level augmentations [20].

The cement viscosity issue is also a crucial parameter that is associated with the risk of cement leakage [23, 24]. Injecting PMMA at a low viscosity leads to a scattered, fingerlike, uncontrolled cement distribution pattern and, as a consequence, an increased chance of cement leakage. By the injection of the cement at higher viscosity, a uniform and better-controlled filling pattern can be achieved. During polymerization of commonly used PMMA cements, the viscosity steadily increases. Hence, the cement should be injected at the latest time point possible to have cement that is viscous enough to prevent leakage and extravasation [2527].

Based on these considerations a lavage technique for vertebroplasty was developed, which enables the surgeon to wash out the intertrabecular bone marrow of the vertebrae before PMMA injection. This leads theoretically to a reduction of fat embolism by reducing the intravertebral resistance and injection forces [28]. In a cadaveric feasibility study Benneker et al. tested the lavage technique in 24 intact vertebral bodies harvested from five osteopenic or osteoporotic human cadaver spines [29]. The study showed a reduction of injection forces that allows the use of more viscous PMMA cement. This lowers potentially the risk of cement leakage and fat embolization and makes vertebral augmentation safer [29]. It could be shown in a sheep model that lavage of the bone marrow prior to VP prevents cardiovascular complications, reduces injection forces and allows a better control of cement distribution pattern with less leakage [30, 31]. The advantages of lavage technique were not yet assessed in clinical studies.

In our hospital a simplified lavage technique prior to vertebral augmentations using PMMA cement in vertebral fractures is used since 2012.

The objective of this study was to assess radiologically the filling patterns of the cement and the cement leakage rates in patients treated with VP, KP or stentoplasty (SP) with and without lavage.

Materials and methods

Patient sample

The local medical ethical committee of the authors’ institution approved the study.

The study included 142 consecutive patients with osteoporotic VCF who were treated in the period from January 2012 to August 2014 in our hospital. Osteoporotic VCF was diagnosed primarily on the basis of simple a.-p. and lateral standing X-ray of the spine and confirmed in computed tomography.

Four surgeons were involved in the study, and, overall, 215 levels were augmented in the 142 patients. Two surgeons consequently applied the newly developed lavage technique prior to cement application in 64 patients (45.1 %) with 116 vertebrae (1.8 vertebrae per patient). These were assigned to the interventional “lavage” group. Two other surgeons used a conventional cement injection technique and treated 78 patients (54.9 %) with 99 vertebrae (1.3 vertebrae per patient), who were assigned to “controls”.

Surgical technique

Three types of vertebral augmentation were used: VP, KP and SP. All procedures took place under general anesthesia. The reduction of the fractures was supported by prone patient positioning on a special mattress on the operating table to aggravate the lordosis and promote partial height restoration via ligamentotaxis. A bipedicular approach with the Jamshidi needles under fluoroscopic imaging was used in all patients. Cement preparation was performed according to the manufacturer’s manual (Vertecem V+, Synthes, Bettlach, Switzerland), injection of cement was stopped when either a sufficient filling of the vertebra was achieved (endplate to endplate) or cement leakage was observed.

In the kyphoplasty group a standard kyphoplasty device (SYNFLATE—Vertebral Balloon System, Synthes Bettlach, Switzerland), for the stentoplasty a standard stentoplasty device (VBS—Vertebral Body Stenting System, Synthes, Bettlach, Switzerland) was used according to the manufacturer’s manual. All reviewed cases were done with and bipedicular approach. After reduction of the fracture, cement injection was performed under fluoroscopic imaging as usual.

Lavage technique

The lavage procedure was performed with two 20 ml standard syringes (Injekt™ Syringe 20 ml LL, B. Braun Medical AG, Sempach, Switzerland) connected to the top of the each Jamshidi needle (Fig. 1). One of the syringes was filled with 10 ml saline solution. By the application of suction to the contralateral syringe, a negative pressure inside the vertebral body was produced. This negative pressure sucked in the saline solution from the filled syringe through the vertebral body. Thus, the intravertebral bone marrow was washed out. After filing the empty syringe with saline-bone-marrow solution the syringe was discarded. Then the lavage procedure was performed the other way around changing the positions of the filled and empty syringes. Two lavages in each direction resulted in a total of four lavages for each vertebral body. Usually after the 3rd passage the aspirated fluid is nearly clear.

Fig. 1
figure 1

Lavage technique: lavage procedure with two 20 ml syringes: a attachment of one empty syringe and one filled syringe to the 8 Ga Jamshidi-Biopsy needles located in vertebral pedicules; b application of suction (negative pressure) to the empty syringe; c suction in of the intravertebral bone marrow and the saline solution from the contralateral syringe

After the lavages the augmentation of the vertebral body was performed as usual, aiming for a large cement volume (“endplate-to-endplate” distribution). In the kyphoplasty and the stentoplasty groups lavage was performed prior to balloon dilatation (after creating the working channel, prior to ballon insertion).

Radiological assessment and outcome measures

Preoperative a.-p. and lateral X-ray were taken prior to surgery. Postoperative X-ray was performed on the first postoperative day. The X-rays were assessed on the PACS Sectra Workstation in the DICOM format (IDS7, Sectra, Sweden). All fractures were classified according to the AO-Magerl classification [32]. The fracture level and type of augmentation (VP, KP and SP) were assessed.

Cement filling pattern was analyzed on the postoperative X-ray images and classified based on a four-gradient scale into insufficient filling pattern (<25 % of the vertebra), medium filling pattern (25–50 %), sufficient filling pattern (50–75 %) and endplate-to-endplate filling pattern (100 %). Any cement located outside the vertebra was considered leakage and was classified based on their location: via the segmental or basivertebral vein (=venous leakage), through a cortical defect or through the endplate. If the leakage was into the spinal canal or into the intervertebral foramen it was judged as “posterior leakage”. Both cement filling pattern and cement leakage were assessed by an independent evaluator who was blinded to the group assignment.

Statistical analysis

For the descriptive statistics, means and standard deviations were calculated and minimum and maximum values extracted. The non-parametric Wilcoxon rank-sum test was used for comparison of continuous and the Chi-square test for categorical variables.

A binomial multivariate logistic regression model was built to find out predictors for (1) an overall leakage, (2) for a venous leakage, (3) for cortical defect leakage and (4) for a leakage into intervertebral disc based on all cemented levels. Stepwise selection was used in these models. Furthermore, a multivariate linear regression was calculated to find predictors. Following covariates were included in the models: patient age, sex, fracture type, grouped segments (Th1–Th10, Th11–Th12, L1–L5), augmentation technique (KP, ST, VP) and the treatment group (lavage, control).

The significance level was set to 0.05 throughout the study. All statistical analyses were conducted using SAS 9.4 (SAS Institute, Inc., Cary, NC, USA).

Results

Patient characteristics

Patient characteristics are shown in Table 1. The distributions of patient characteristics was similar between the groups with the exception of the proportions of the fracture types. Prophylactic augmentation was more common in the lavage group while the fracture pattern A1.2 was more common in the controls (p < 0.001) (Table 1). Due to the same number of balloon-assisted procedures (VP and SP) in both groups (25.8 % in the lavage vs. 25.3 % in the non-lavaged group) we did not further split up the groups regarding the augmentation techniques.

Table 1 Patient characteristics of the treatment groups

Outcome parameters

Leakage rate

There were 44 vertebrae (20 patients) with 49 leakages in the lavage group, and 83 vertebrae (68 patients) with 109 leakages in the control group (vertebra-based p < 0.001; Table 2).

Table 2 Outcome measures (filling pattern, leakage rate and extent of leakage) in the groups

The proportion of overall, venous and cortical defect leakages were different in favor of the lavage group (p ≤ 0.001; Table 2), while the proportion of intradiscal leakages was similar (p = 0.51; Table 2). Posterior leakage was very rare (n = 5).

Overall leakage

The multivariate logistic regression analysis for an overall leakage based on all cemented levels showed lavage and fracture type, but not age, sex and grouped segments, to be independent predictors. Moreover, augmentation technique (VP, KP, SP) could not be shown to be a predicting factor for leakage. A prophylactic cement augmentation was associated with 0.39-times [95 confidence intervals (95 % CI) 0.16–0.99; p < 0.001] lower likelihood for an overall leakage in comparison to fracture. Lavage was associated with 0.13-times (95 % CI 0.07–0.27; p = 0.005) lower likelihood for an overall leakage in comparison to control.

Venous leakage

The multivariate logistic regression analysis for a venous leakage also based on all cemented levels showed lavage to be the only independent predictor. Lavage group was associated with 0.33-times (95 % CI 0.16–0.65; p = 0.001) lower likelihood for such a leakage in comparison to control.

Cortical leakage

The multivariate logistic regression analysis for a cortical leakage revealed fracture type and lavage to be the independent predictors. Lavage group was associated with 0.15-times (95 % CI 0.08–0.30; p < 0.001) lower likelihood for cortical leakage in comparison to control.

Filling pattern

The distributions of cement filling patterns were different between the groups in favor of the lavage group (p < 0.001; Table 2). In terms of the extent of leakage the groups were similar (p = 0.10; Table 2). No difference in filling pattern could be found between the different augmentation techniques (VP, KP, SP).

Discussion

The objective of this study was to investigate whether lavage used in PMMA augmentation techniques in the treatment of VCFS has a positive effect on the cement filling pattern and can reduce the risk of cement leakage.

The developed easy and cost-effective lavage technique with two syringes showed to be feasible and reproducibly applicable in the augmentation of osteoporotic and osteopenic fractured and non-fractured vertebral bodies.

So far few experimental and animal studies focused on the effect of bone-marrow removal of the vertebral body prior to PMMA application. It could be shown that (jet)-lavage does not cause cardiovascular reaction in combination with contralateral application of vacuum and can significantly decrease the embolic load in sheep [30]. Similar results, significant reductions of cement leakage, were found by Mohamed et al. [33], using an unipedicular aspiration technique (without lavage) in a vertebroplasty model.

In the current clinical study we did not investigate the aspirated fluid. The analysis of the relation of fat and blood in the aspirate is actually subject of an ongoing study.

We could show that with the newly developed bipedicular lavage technique a significant reduction of radiological visible cement leakage can be achieved, especially for the potentially dangerous venous leakage, for which lavage was the only independent predictor. Reasons, therefore, may be that by decreasing of the intravertebral resistance the injection of higher viscous cement is possible and the initial path of least resistance is no longer the segmental vein but the vertebral body itself.

For cortical defect leakage fracture type could be shown to be independent predictor additionally to lavage. While other studies found a higher complication and leakage rate in the augmentation of burst fractures [34], in our study population A1.2 fractures (wegde-impaction) showed a higher leakage rate compared to other fracture forms. The cause for this may be iatrogenically higher PMMA insertion pressures in an attempt to achieve endplate-to-endplate filling in this fracture type, in which one endplate is still intact.

It is known that by the use of higher viscous cement a more uniform cement filling with an increased leakage rate is achievable [25]. Moreover, it has been shown that removing bone marrow fat by lavage allows injecting more cement leading to a more homogenous distribution [28]. It might be possible that reduced injection pressures after lavage increase both effects, better filling pattern and less leakage. Our findings support those results, showing that the filling pattern in the lavage group is better in terms of a significant higher rate of endplate-to-endplate fillings in the lavage group, which is favorable [22]. Unfortunately, we did not record the amount of injected cement in detail, but it can be assumed that the better filling pattern of the lavage group is due to higher cement volumes at lower leakage rates. In addition to the missing recording of cement volumes also no measurement of injection pressures have been made and should be seen as limitations of the study.

The risk of extraosseous cement leakage in various series ranges between 3 and 74 % [57, 35] Even so it is hard to compare leakage rates between studies, mainly due to different patient inclusion criteria, methodology (leakage determined by CT, radiographs or intraoperative) as well as cement parameters (type, percentage of filling), in a review of 69 clinical studies Hulme et al. [5] reported an average leakage rate of 41 % for vertebroplasty and of 9 % for kyphoplasty. Compared to this data our lavage group showed a low leakage rate (12.9 %), especially for the potentially life threatening venous leakage.

Interestingly we could not find any difference in leakage rate and cement distribution between vertebroplasty, kyphoplasty and stentoplasty. Higher leakage rates have been reported for single-group cohort vertebroplasty studies compared to kyphoplasty studies [5]. Nevertheless, for kyphoplasty it has been shown [36] that attempting to fill the whole vertebra, rather than just the cavity, results in higher leakage rates similar to rates observed in vertebroplasty (33 %). Our attempt to reach an “endplate-to-endplate” filling pattern in all three different augmentation techniques may contribute to our findings.

We could not find any complications directly associated with the new lavage technique. Theoretically there might be a minimal risk of increased blood loss due to the procedure. As we lavaged four times with a 20-ml syringe the maximum additional blood loss could be 80 ml. Anyway in our patients we did not have problems with postoperative anemia. So far we have not done a qualitative analysis of the aspirate. Previous not-published animal studies suggest the percentage of fat/blood varies largely between 5 and 15 % for the first passage. Actually a prospective study is ongoing investigating this topic specifically.

There are some limitations of the study to be mentioned. Fracture age, which may affect the augmentation process, was not available in all patients and, therefore, could not be integrated into the multiregression analysis. Nevertheless, from our experience it does not affect the lavage process per se. The comparison of two vs. two independent surgeons may be one other limitation potentially causing bias. Anyway, as all surgeons in the department have the same philosophy, used identical cement and cannulas and underwent the same surgical training, we believe it to be valid to compare all the four.

However, in combination with other techniques, which help to prevent cement leakage, like sequential application [4] and use of ultra-high viscous cement, the lavage technique is promising to reduce potentially life threatening cardiopulmonary events.

Conclusions

Removing intravertebral bone marrow by lavage is inexpensive, safe, feasible and has the potential to reduce fat embolism and PMMA leakage in vertebroplasty or ballon kyphoplasty. Therefore, it may be useful for preventing potentially life threatening complications in patients with impaired cardiopulmonary function. Moreover, a better filling pattern can be achieved.